510(k) Data Aggregation

    K Number
    K252792

    Validate with FDA (Live)

    Date Cleared
    2025-11-17

    (76 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PM2™ System with ECGuide™ Connector is indicated for the positioning of central venous access devices (CVADs) of at least 1Fr in size. It provides catheter tip location information by using the patient's cardiac electrical activity. The PM2™ System with ECGuide™ Connector is indicated for use as an alternative method to chest x-ray or fluoroscopy for confirmation of all central venous access devices.

    In adult patients and in adolescents (greater than 12 through 21 years of age), The ECGuide™ can be used with CVADs such as peripherally inserted central catheters (PICCs), centrally inserted central catheters (CICCs), implantable ports, and hemodialysis catheters; in children (greater than 2 to 12 years of age), ECGuide™ can be used with PICCs and with CICCs; in infants (greater than 1 month to 2 years of age) and in neonates (from birth to 1 month of age), ECGuide™ Can be used with PICCs and CICCs. In each specific age group, the CVAD type and size must be chosen and the CVAD must be used according to the CVAD's indications and instructions for use.

    Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-wave; including

    • Atrial fibrillation
    • Atrial flutter
    • Severe tachycardia
    • Pacemaker-driven rhythm
    • Chronic obstructive pulmonary disease (COPD)

    Such patients are easily identified prior to central catheter insertion. In these specific cases, use of an additional confirmation method is necessary to confirm catheter tip location.

    Device Description

    The purpose of this Special 510(k) is for the expansion of indications for use of the previously cleared PM2™ System with ECGuide™ Connector (K240486) from use in adults to pediatric populations. There have been no major changes to the device since initial clearance under K240486. The PM2™ System with ECGuide™ Connector is a device used by clinicians for bedside tip location confirmation of central venous catheters. The PM2™ System with ECGuide™ Connector provide real-time catheter tip location information by using intravascular ECG (ivECG). The ECGuide™ Connector is connected to the distal lumen hub of a CVC and primed with saline to create an ivECG electrode at the distal tip of the catheter and allow for ECG catheter tip confirmation. The ECGuide™ Connector is used only during CVC installation, and removed once the CVC tip position is confirmed to be at the target location. The ECGuide™ Connector is electrically connected to the PM2 system which processes, displays, and stores the ECG waveforms.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243484

    Validate with FDA (Live)

    Date Cleared
    2025-05-29

    (202 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PM2+ System and SmartPICC Stylet is indicated for guidance and positioning of commercially available central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording for catheter guiding and positioning, and intravascular ionic dilution for supplemental catheter navigation. The SmartPICC Stylet, when used with the PM2+ System, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When relying on the patient's ECG signal, the PM2+ System and SmartPICC Stylet is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. The SmartPICC Stylet is indicated for use with PICC lines having a minimum lumen diameter of 0.021".

    Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous catheter insertion the use of an additional method is required to confirm catheter tip location.

    Device Description

    The PM2+ System and SmartPICC Stylet is a device used by clinicians for bedside navigation and tip location confirmation of central venous catheters including PICCs. The PM2+ System and SmartPICC Stylet provide real-time catheter tip location information by using intravascular ECG (ivECG). The SmartPICC Stylet provides intravascular ECG for catheter tip location confirmation as well as an optional, ionic dilution navigation feature. The SmartPICC Stylet is used only during PICC installation, and removed once the PICC tip position is confirmed to be at the target location. The SmartPICC Stylet is electrically connected to the PM2+ system which processes, displays, and stores the ECG waveforms as well as the ionic dilution data. The PM2+ System is also compatible with the commercially available ECGuide Connector.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the PM2+ System and SmartPICC Stylet does not include acceptance criteria or a detailed study that proves the device meets specific acceptance criteria related to its primary function (catheter tip placement confirmation using ECG).

    The document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance metrics derived from a clinical trial.

    Here's an analysis based on the information provided, highlighting what is present and what is absent:


    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The clearance letter states that "Performance testing performed on the PM2+ System and SmartPICC Stylet verified that device performance requirements including tensile strength, insertion force, corrosion resistance, catheter compatibility, leak testing, infusion rate testing, ionic dilution performance testing, algorithm verification, electrical hardware verification, sterile barrier packaging testing, biocompatibility, electrical safety, EMC, software verification and validation, design validation and human factors validation. All tests were successfully completed." However, the specific acceptance criteria (e.g., "tensile strength must be >X Newtons") and the measured performance values for these non-clinical tests are not disclosed.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided for demonstrating the core performance of the device (catheter tip placement confirmation). The document explicitly states: "No human clinical data was provided to support substantial equivalence." This implies that there was no clinical test set for this specific submission to demonstrate the accuracy of the ECG-based tip localization. The performance testing mentioned in point 1 refers to non-clinical (engineering and benchtop) tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. As "No human clinical data was provided," there was no clinical ground truth established for a test set in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. As "No human clinical data was provided," there was no clinical test set requiring an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. The device is a "Tip Location System" and "Stylet," which assists in the placement of central venous catheters. It's not described as an AI system that improves human readers of medical images. The primary use case is direct guidance during placement, not interpretation of images post-placement. The clearance relies on substantial equivalence to predicate devices, where the ECG-based tip confirmation method is already established.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation of the algorithm's accuracy in determining catheter tip location is not explicitly detailed with specific metrics. While "algorithm verification" is listed as a completed test, the specific methodology, dataset, and performance criteria for this verification are not disclosed. The device description explicitly states it provides "real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information," which implies it has an embedded algorithm, but its standalone performance against a ground truth is not specified in this document.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests (tensile strength, etc.), the ground truth would be established by objective physical measurements. However, for the crucial function of accurate catheter tip placement confirmation using ECG, the document does not specify the type of ground truth used for algorithm verification, nor does it provide a clinical study where a ground truth (e.g., fluoroscopy, chest X-ray interpreted by experts, or MRI) would be established for comparison. The clearance implies that the underlying technology (ivECG for tip location) is already validated from the predicate devices.

    8. The sample size for the training set

    This information is not provided. Given "No human clinical data was provided" for the current submission, and the reliance on substantial equivalence, details about a training set for a new algorithm performance claim are absent. If the "algorithm verification" refers to the core ECG-based localization, the training would have been part of the predicate device development.

    9. How the ground truth for the training set was established

    This information is not provided. (See explanation for point 8).


    In summary, the FDA 510(k) clearance for the PM2+ System and SmartPICC Stylet does not contain the detailed acceptance criteria and study data you requested for its primary function (ECG-based catheter tip localization). The clearance is based on demonstrating substantial equivalence to existing legally marketed predicate devices (K200037 Piccolo Medical SmartPICC System and K240486 PM2 System and ECGuide Connector) which already use similar technology, and on successful completion of various non-clinical engineering and benchtop performance tests. The document explicitly states that "No human clinical data was provided to support substantial equivalence" for this specific submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240486

    Validate with FDA (Live)

    Date Cleared
    2024-09-08

    (201 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo Medical PM2™ System with ECGuide™ Connector is indicated for the positioning of central venous catheters including PICCs. It provides catheter tip location information by using the patient's cardiac electrical activity. The PM2™ System with ECGuide™ Connector is indicated for use as an alternative method to chest x-ray or fluoroscopy for confirmation of central venous catheter tip placement in adult patients.

    Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated. for patients where cardiac rhythms may change presentation of the P-wave; including

    -Atrial fibrillation

    -Atrial flutter

    -Severe tachycardia

    -Pacemaker-driven rhythm

    -Chronic obstructive pulmonary disease (COPD)

    Such patients are easily identified prior to central catheter insertion. In these specific cases, use of an additional confirmation method is necessary to confirm catheter tip location.

    Device Description

    The PM2™ System and ECGuide™ Connector is a device used by clinicians for bedside tip location confirmation of central venous catheters. The PM2™ System and ECGuide™ Connector provide real-time catherer tip location information by using intravascular ECG (ivECG). The ECGuide™ Connected to the distal lumen hub of a CVC and primed with saline to create an ivECG electrode at the distal tip of the catheter and allow for ECG catheter tip confirmation. The EGuide™ Connector is used only during CVC installation, and removed once the CVC tip position is confirmed to be at the target location. The ECGuide™ Connected to the PM2 system which processes, displays, and stores the ECG waveforms.

    AI/ML Overview

    The provided FDA 510(k) summary (K240486) for the "PM2 System and ECGuide Connector" does not contain information regarding acceptance criteria or a study proving the device meets acceptance criteria related to its clinical performance for catheter tip placement confirmation.

    The document states: "No human clinical data was provided to support substantial equivalence."

    Instead, the summary focuses on non-clinical performance testing for aspects such as electrical safety, EMC, software verification and validation, tensile insertion force, corrosion resistance, catheter compatibility, leak testing, electrical hardware verification, sterile barrier packaging testing, biocompatibility, design validation, and human factors validation. It concludes that "All tests were successfully completed. The passing results supported the determination of substantial equivalence with the predicate device."

    Therefore, I cannot provide the requested information as it is not present in the provided text.