Search Filters

Search Results

Found 28 results

510(k) Data Aggregation

    K Number
    K251146

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (186 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use:
    The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.

    Indications for Use
    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are only for use on one patient at a time.

    The monitors are not therapeutic devices.

    The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The IntelliVue NMT is intended to be used as an objective neuromuscular transmission monitor that measures the muscle response to electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

    The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

    The Edwards FloTrac solution offers continuous assessment of hemodynamic parameters. It is indicated to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is indicated for use in adult critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. The Edwards FloTrac solution is indicated to be used in the operating room, intensive care unit, and emergency room.

    The monitors are intended for use by trained healthcare professionals in a hospital environment.
    They are not intended for home use.
    The monitors are additionally intended for use in transport situations within hospital environments.

    Device Description

    The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.

    The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and gas analysis.

    The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals. The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display. The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233440

    Validate with FDA (Live)

    Date Cleared
    2024-07-02

    (258 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avalon CL Fetal & Maternal (F&M) Pod & Patch is a device indicated for use by healthcare professionals in a clinical setting for non-invasive monitoring of maternal heart rate (aHR), fetal heart rate (aFHR), and uterine activity (aToco) in women who are at >36 completed weeks, in labor, with singleton pregnancy, using surface electrodes on the maternal abdomen.

    Device Description

    The Avalon CL Fetal & Maternal (F&M) Pod and the Avalon CL Fetal & Maternal (F&M) Patch is a beltless battery-powered maternal-fetal monitoring system that non-invasively measures abdominal fetal heart rate (aFHR), abdominal uterine activity (aToco), and abdominal maternal heart rate (aHR). The Avalon CL Fetal & Maternal (F&M) Patch is a single-use disposable adhesive electrode patch designed to be affixed to the maternal abdomen. The Avalon CL Fetal & Maternal (F&M) Pod is a reusable device which, when connected to the Avalon CL Fetal & Maternal (F&M) Patch, picks up electrical signals and converts it to Short Range Radio (SRR). The Avalon CL Fetal & Maternal Pod communicates the data measurement values to the Avalon CL Base Station using Short-Range Radio (SRR). The Avalon CL Base Station in turn relays the information to the connected Philips Fetal-Maternal (FM) Monitor (i.e., FM20, FM30, FM40, and FM50).

    AI/ML Overview

    The provided FDA 510(k) summary for the Philips Avalon CL Fetal & Maternal (F&M) Pod & Patch focuses heavily on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of technical characteristics rather than a detailed clinical study report with specific acceptance criteria and performance metrics for the device's accuracy in monitoring FHR, MHR, and UA.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical performance (e.g., accuracy, sensitivity, specificity, agreement with ground truth for FHR, MHR, and UA) is not explicitly detailed in this document. The document primarily discusses non-clinical tests for safety, electrical performance, and biocompatibility.

    However, based on the information provided, here's what can be extracted and inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific clinical performance acceptance criteria (e.g., accuracy ranges for FHR) and reported device performance from an effectiveness standpoint. Instead, it details non-clinical technical acceptance criteria related to safety, electrical performance, and biocompatibility, which the device met.

    Criterion CategorySpecific Criterion / TestAcceptance Criterion (Implicit)Reported Device Performance (Implicit)
    BiocompatibilityCytotoxicity (ISO 10993-5)Met acceptance criteria as defined in test requirementsMet
    Sensitization (ISO 10993-10)Met acceptance criteria as defined in test requirementsMet
    Irritation (ISO 10993-10)Met acceptance criteria as defined in test requirementsMet
    Electrical SafetyANSI AAMI ES60601-1Compliance with standard for basic safety and essential performancePassed
    EMC/WirelessIEC 60601-1-2Compliance with standard for electromagnetic disturbancesPassed
    IEEE ANSI C63.27Compliance with standard for evaluation of wireless coexistencePassed
    IEC/TR 60601-4-2Compliance with standard for electromagnetic immunityPassed
    Alarm SystemsIEC 60601-1-8Compliance with standard for alarm systemsPassed
    Battery SafetyIEC 62133-2Compliance with standard for lithium systemsPassed
    Software/FirmwareFDA Guidance complianceCompliance with "Content of Premarket Submissions for Device Software Functions"Documentation provided and reviewed
    CybersecurityFDA Guidance complianceCompliance with "Cybersecurity in Medical Devices" guidanceDocumentation provided and reviewed
    Performance BenchInspection of labeling and pouch sealingN/A (Visual inspection)Met
    Impedance/tensile strength/pull-off force/noise level/conductivity/offset voltage/defibrillation overload (new and aged patches)Met acceptance criteria as defined in test requirementsMet
    In vivo testing: integrity, detachment/reattachment, and performance (impedance, noise level, MHR, conductivity) after shower and usageMet acceptance criteria as defined in test requirementsMet
    Peel-off force of each electrode and central stickerMet acceptance criteria as defined in test requirementsMet
    MHR/FHR/UA accuracy after storage at various temperaturesMet acceptance criteria as defined in test requirementsMet
    Signal transmission continuityMet acceptance criteria as defined in test requirementsMet

    Regarding MHR/FHR/UA accuracy, the document states for "Performance Bench" that "MHR/FHR/UA accuracy after stored in room (23℃), high (32℃) and low (2-8℃) temperature" were conducted and "met the acceptance criteria as defined in the test requirements." However, the specific numerical acceptance criteria for accuracy (e.g., mean absolute difference, percentage agreement, etc.) and the reported numerical performance regarding MHR/FHR/UA accuracy are not provided in this summary. This suggests that these accuracy tests were likely bench tests under controlled conditions, not a clinical trial comparing device readings to a clinical ground truth.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly mention a "test set" in the context of a clinical performance study with human subjects to evaluate the accuracy of FHR, MHR, and UA measurements. The in-vivo testing mentioned under "Performance Bench" refers only to "integrity, detachment/reattachment, and performance (impedance, noise level, MHR, conductivity) after shower and usage (8 hours/32 hours) for the patch (Novii Patch)." This does not sound like a large-scale clinical accuracy study.

    Therefore, based on the provided text alone:

    • Sample size for the test set: Not explicitly stated for clinical performance as commonly understood for device accuracy. The "in vivo testing" details are too limited to determine sample size or its direct relation to device accuracy claims.
    • Data provenance: Not explicitly stated. The type of testing suggests it might be internal company testing rather than an independent clinical trial.

    3. Number of Experts and Qualifications for Ground Truth

    Given the lack of a detailed clinical performance study report, there is no information provided regarding the number or qualifications of experts used to establish ground truth for a clinical test set for FHR, MHR, or UA.

    4. Adjudication Method

    Again, due to the absence of a detailed clinical performance study, there is no information provided on any adjudication method (e.g., 2+1, 3+1) for a clinical test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human readers or AI assistance in this context. This device appears to be a monitoring system for physiological parameters, not an AI-assisted diagnostic imaging or interpretation tool.

    6. Standalone Performance

    The device itself is a "standalone" monitoring system in the sense that it performs its measurements (aHR, aFHR, aToco) via its electrodes and pod, then relays this data to a Philips Fetal-Maternal (FM) Monitor for display. The performance tests ("Performance Bench") assess the device's ability to measure these parameters. However, the exact "standalone" clinical accuracy metrics (e.g., sensitivity, specificity, accuracy vs. a gold standard) are not provided. The phrase "standalone performance" is generally associated with diagnostic algorithms, which doesn't seem to be the primary claim here.

    7. Type of Ground Truth Used

    For the non-clinical performance "MHR/FHR/UA accuracy after stored in room (23℃), high (32℃) and low (2-8℃) temperature," the type of ground truth used is not specified. It likely refers to controlled laboratory measurements against calibrated reference standards, rather than clinical ground truth like pathology, expert consensus, or outcomes data. For clinical performance data (which is not detailed), common ground truths for FHR, MHR, and UA would be internal fetal monitoring (IUPC for UA, fetal scalp electrode for FHR) or expert interpretation of existing monitoring tracings (though this isn't mentioned).

    8. Sample Size for the Training Set

    No information is provided about a "training set." This term is typically associated with machine learning or AI algorithm development. While the device uses signal processing (template matching, filtering, confidence tagging) to identify fECG and mECG complexes, the document does not describe the development or training of such algorithms or any associated data sets used for this purpose.

    9. How Ground Truth for the Training Set Was Established

    As no training set is discussed, there is no information provided on how ground truth for a training set was established.

    In summary, the provided FDA summary focuses on demonstrating substantial equivalence through technical and non-clinical performance and safety testing. It lacks detailed clinical performance data (e.g., accuracy, sensitivity, specificity) against a clinical ground truth, specific sample sizes for clinical evaluations, or information about expert consensus or adjudication methods for such clinical data, which are typically found in clinical study reports for devices claiming diagnostic or interpretative capabilities.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use:

    The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.

    Indications for Use:

    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are only for use on one patient at a time.

    The monitors are not therapeutic devices.

    The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help quide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics >5 kg and <40 kg in healthcare environments.

    The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil. Desflurane. Fentanyl. Isoflurane. Nitrous Oxide. Propofol. Remifentanil. and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

    The monitors are intended for use by trained healthcare professionals in a hospital environment. They are not intended for use in transport situations.

    They are not intended for home use.

    Device Description

    All features and technological characteristics of the primary predicate IntelliVue Patient Monitors MX750 (866471), MX850 (866470) and IntelliVue 4-Slot Module Rack FMX-4 (866468) remain unchanged and are still applicable to the IntelliVue Patient Monitors, subject of this submission.

    The proposed modifications for the IntelliVue Patient Monitors MX750/MX850 and IntelliVue 4-Slot Module Rack FMX-4 include:

      1. A new software version, designated as IntelliVue Software P.01. The new software version introduces following new features:
      • . New alarm sounds designated as "Philips 2021" sounds, based on a joint development of Philips and Sen Sound researchers who focus on transforming the experience of sound in healthcare.
      • . New Visual Patient feature, which displays patients's vital signs in an animated patient avatar alongside the conventional vital sign waveforms and numerics.
      • . configurable alarm management, limiting changes in alarm settings to pre-defined authorized personnel.

    It further enhances following existing features

    • Manual Selection of QRS /Pulse Tone
    • Introduction of Lead Diagram for ECG
    • . Enhancement of Visual Presentation for "Global Alarm Off" State
    AI/ML Overview

    This document describes the Philips IntelliVue Patient Monitor MX750/MX850 and IntelliVue 4-Slot Module Rack FMX-4. The submission references a prior clearance (K221348) as the primary predicate device, indicating that the current submission is for modifications to an existing device.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for performance in a table format with numerical values. Instead, it details that the device passed various tests related to electrical safety, EMC, wireless coexistence, software integration, software functionality within the system, essential performance, and human factors. The acceptance criterion for each of these tests is implied to be "PASSED" against recognized consensus standards and internal requirements.

    Test NameTest DescriptionAcceptance Criteria (Implied)Reported Device Performance
    Electrical Safety TestingTested to recognized consensus standards: IEC 60601-1, IEC 60601-1-8, IEC 62304PASSEDPASSED
    EMC TestingTested to recognized consensus standard: IEC 60601-1-2; Tested against common EM emittersPASSEDPASSED
    Wireless CoexistenceTest sensitivity of patient monitors to common interferersPASSEDPASSED
    Software Integration TestingTest of new functionality integration into existing software infrastructurePASSEDPASSED
    Software TestingTest of new software functionality ("Test Report for System Requirements Testing," "Test Report for Safety Risk Assessment," "Test Report for Regression Testing," "Test Report for Security Risk Assessment," "Test Report for Compatibility Testing")PASSEDPASSED
    Essential Performance TestingTests verifying that essential performance requirements are met (for Patient Monitors)PASSEDPASSED
    Human FactorsTest of device usabilityPASSEDPASSED

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample sizes used for any of the tests mentioned (e.g., how many devices were tested for electrical safety, or how many users were involved in human factors testing). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for any of these non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The tests described are primarily engineering and software validation tests rather than clinical studies requiring expert ground truth for interpretation. For "Human Factors," it would involve users, but their qualifications and numbers are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided. The nature of the tests (compliance with standards, software functionality) suggests that a formal adjudication method like 2+1 or 3+1 for ground truth establishment, typical in clinical studies, would not be applicable here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a patient monitor with software enhancements, not an AI-assisted diagnostic tool that would typically involve human readers interpreting output.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The document indicates "Software Testing" was performed, which would involve testing the algorithms and functionalities of the software in a standalone manner. However, it's not a standalone diagnostic algorithm in the typical sense; it's the core operating software of the patient monitor. The new features like "new alarm sounds," "Visual Patient feature," and "configurable alarm management" were tested for their functionality and integration.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the tests described is primarily based on:

    • Recognized consensus standards: For electrical safety (IEC 60601-1, IEC 60601-1-8, IEC 62304), EMC (IEC 60601-1-2).
    • System requirements/specifications: For software integration, software functionality, and essential performance.
    • Usability principles: For Human Factors testing.

    There is no mention of expert consensus, pathology, or outcomes data being used as ground truth, as these are non-clinical hardware and software validation tests.

    8. The sample size for the training set:

    This information is not applicable as this document does not describe an AI or machine learning model that requires a training set. The changes described are software updates for existing functionalities and introduction of new display/alarm features, not the development of a predictive algorithm from data.

    9. How the ground truth for the training set was established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230604

    Validate with FDA (Live)

    Date Cleared
    2023-11-22

    (264 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use:

    The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.

    Indication for Use:

    The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitor is only for use on one patient at a time. The monitor is not a therapeutic device. Caution: The monitor is for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only.

    Warning: The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for use during patient transport inside the hospital environment. It is not intended for home use.

    Warning: The monitor is not intended for use in an MRI environment or in an oxygen-enriched environment (for example, hyperbaric chambers).

    Contraindications: There are no known contraindications for the IntelliVue MX100/X3 Patient Monitor/Multi-Measurement Module.

    Device Description

    The Multi-Measurement Module X3 and the IntelliVue Patient Monitor MX10
    0, acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The devices offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They can be located in the patient vicinity at the bedside or can be used mobile, during patient transport inside hospitals. The X3 and MX100 can interact with the Central Station via LAN or wireless link.

    The Multi-Measurement Module X3 and IntelliVue Patient Monitor MX100 are compact, rugged, lightweight monitors with built-in invasive and non-invasive measurements, namely ECG (including arrhythmia and ST), respiration, SpO2, NBP, dual invasive pressure, temperature, and CO2. It can further extend the measurement capabilities when connecting to the legally marketed IntelliVue Multi-Measurement Extensions

    The X3 can be used in two ways: as a multi-measurement module for the Philips IntelliVue family of patient monitors and as a stand-alone monitor.

    AI/ML Overview

    The provided document, an FDA 510(k) summary for the Philips IntelliVue Patient Monitor MX100 and IntelliVue Multi-Measurement Module X3, describes a device that has undergone technical modifications, primarily software updates, rather than being a novel AI-powered diagnostic or assistive tool.

    Therefore, the document does not contain the information requested regarding acceptance criteria for an AI-powered diagnostic device, a study proving it meets those criteria, sample sizes for test and training sets, expert usage for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance.

    The document focuses on demonstrating substantial equivalence to a predicate device, as per FDA 510(k) requirements. This typically involves showing that the device has the same intended use, similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness.

    Here's why the requested information is absent and what is discussed:

    1. Acceptance Criteria for AI Performance and Reported Performance: Not applicable as this is not an AI diagnostic device. The acceptance criteria mentioned are related to compliance with general medical device safety and performance standards (e.g., electrical safety, EMC, software lifecycle, usability, alarm systems).
    2. Sample Sizes and Data Provenance: Not applicable for AI model testing. The document states "No new issues of safety or effectiveness as compared to the predicate are introduced because of using this device," indicating that the focus was on validating the software changes within the existing framework of a patient monitor, not on an AI algorithm's diagnostic performance on a specific dataset. There is no mention of a "test set" in the context of AI performance.
    3. Number of Experts and Qualifications: Not applicable. Ground truth establishment for AI models is not relevant to this submission.
    4. Adjudication Method: Not applicable.
    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. The device is a patient monitor, not an AI-assisted diagnostic tool that would improve human reader performance.
    6. Standalone Algorithm Performance: Not applicable. There is no standalone AI algorithm being assessed.
    7. Type of Ground Truth: Not applicable in the context of AI. The "ground truth" for this device would be the accurate measurement of physiological parameters (e.g., ECG, SpO2, NBP), which is assessed through bench testing and compliance with recognized standards, not "expert consensus" or "pathology" in the AI sense.
    8. Sample Size for Training Set: Not applicable. There is no AI model training described.
    9. How Ground Truth for Training Set Was Established: Not applicable.

    What the document does demonstrate (in lieu of AI performance data):

    The document primarily focuses on software changes and the continued compliance with established medical device standards to prove substantial equivalence:

    • Software Changes: The key difference from the predicate device is a new software version (P.01) introducing new alarm sounds, configurable alarm management, enhancing Early Warning Score Validation, additional SpO2 fallback configuration, additional Lead Diagram presentation for ECG, and enhancement of the global "alarm off" state.
    • Non-Clinical Bench Tests and Harmonized Standards: The device has passed safety tests demonstrating compliance with recognized standards like ANSI AAMI ES60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (Usability), IEC 60601-1-8 (Alarm Systems), and IEC 62304 (Medical device software lifecycle).
    • No Clinical Studies: The submission explicitly states, "The subject devices, like the primary predicate devices, did not require clinical trials." This further indicates that the changes were not considered to introduce new safety or effectiveness concerns requiring human subject data for validation.
    • Substantial Equivalence: The conclusion is that the modified devices "do not raise different questions of safety and effectiveness when compared to the predicate, perform as intended, and have performance characteristics that are substantially equivalent to the predicate devices."
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are only for use on one patient at a time.

    The monitors are not therapeutic devices.

    The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The Intelli Vue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

    The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentani, Desflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

    The monitors are intended for use by trained healthcare professionals in a hospital environment. They are not intended for home use.

    The MX400/MX450/MX500/MX550 monitors are additionally intended for use in transport situations within hospital environments.

    Device Description

    The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and gas analysis. The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals. The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display. The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

    AI/ML Overview

    This document is a 510(k) Summary for the Philips IntelliVue Patient Monitor MX400, MX450, MX500, and MX550. It does not describe a study involving an AI/algorithm that requires detailed acceptance criteria, ground truth establishment, or specific human reader studies as typically seen for AI-enabled medical devices. Instead, this submission focuses on demonstrating substantial equivalence to a predicate device, primarily due to software and hardware modifications allowing integration with external measurement devices like the Masimo SedLine Sedation Monitor.

    Therefore, many of the requested points, particularly those related to AI/algorithm performance studies (e.g., sample size for test set/training set, expert adjudication, MRMC studies, standalone performance, specific effect sizes), are not applicable to this type of 510(k) submission.

    However, I can extract information related to the overall device performance and testing as described in the document.

    Here's an analysis of the provided text in relation to your request, with an emphasis on what is and is not present:

    Device: IntelliVue Patient Monitor MX400, MX450, MX500, MX550

    Study Type: This is a 510(k) premarket notification primarily demonstrating substantial equivalence (predicate pathway) for modifications to an existing patient monitor, specifically related to software updates and integration of an external monitoring module (Masimo SedLine Sedation Monitor). It is not a de novo submission for a new AI/algorithm.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a table of specific quantitative acceptance criteria for an AI/algorithm's performance (e.g., accuracy, sensitivity, specificity metrics) nor does it provide a "reported device performance" table in that context.

    Instead, the "acceptance criteria" can be inferred from the non-clinical tests performed to demonstrate compliance with recognized standards for medical electrical equipment. The "reported device performance" is that the device "passed all safety tests" and "perform as intended."

    Acceptance Criteria (Inferred)Reported Device Performance
    Compliance with IEC 62304 Ed 1.1 (Software life cycle processes)Passed all safety tests for demonstrated compliance with the standard
    Compliance with ANSI AAMI ES60601-1 (Basic safety & essential performance)Passed all safety tests for demonstrated compliance with the standard
    Compliance with IEC 60601-1-2 Ed 4.0 (Electromagnetic disturbances)Passed all safety tests for demonstrated compliance with the standard
    Compliance with IEC 60601-1-8 Ed 2.2 (Alarm systems)Passed all safety tests for demonstrated compliance with the standard
    Compliance with IEC 60601-1-6 Ed 3.1 (Usability)Passed all safety tests for demonstrated compliance with the standard
    No new issues of substantial equivalence introducedSubstantially equivalent to predicate devices; performs as intended

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable/not specified for an AI/algorithm test set. The testing described is primarily for system-level safety, performance, and software validation.
    • Data Provenance: Not applicable in the context of an AI/algorithm dataset. The document discusses compliance with various electrical, safety, software, and usability standards for the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth establishment by experts is typically for diagnostic AI/imaging algorithms. This submission is for a patient monitor system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for expert review of images or clinical data for AI algorithm validation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a patient monitor, not an AI-assisted diagnostic tool that would typically undergo MRMC studies to evaluate human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/algorithm in the sense of a diagnostic or predictive model being evaluated in standalone mode. It's a medical device system. While it has "software," the performance evaluation is centered on its functionality as a patient monitor and its compliance with relevant standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. The "ground truth" for this device's performance validation is its ability to correctly acquire, display, and alert on physiological parameters according to established benchmarks and safety standards (e.g., AAMI EC11 for ECG, IEC standards for electrical safety). It's not a diagnostic AI's ground truth based on clinical diagnoses or pathological findings.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML model generated from a training set in the typical sense of deep learning or machine learning algorithms.

    9. How the ground truth for the training set was established

    Not applicable. (See #8)


    Summary of what the document does discuss regarding validation and performance:

    The document emphasizes non-clinical bench tests and compliance with harmonized standards (e.g., IEC 62304 for software, ANSI AAMI ES60601-1 for basic safety, IEC 60601-1-8 for alarms, IEC 60601-1-6 for usability). The "performance data" section primarily lists these standards and states that the devices "passed all safety tests."

    Key statements from the document regarding performance:

    • "The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below." (See table in section 1.8 Performance Data)
    • "No new issues of substantial equivalence are introduced because of using this device." (Section "Non-clinical Bench Tests")
    • "The subject devices, like the primary predicate devices, did not require clinical trials. Any clinical studies performed for the Masimo SedLine parameter are still valid as the measurements are not modified; they are only being connected to an additional host patient monitor." (Section "Clinical Studies")
    • "Compliance to the FDA Quality System Regulations, FDA recognized standards, FDA guidance documents, harmonized standards, verification and validation, software validation, usability validation, and risk management activities have taken place for the subject devices." (Section "Clinical Studies")
    • "The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety and electromagnetic compatibility, software verification and validation, human factors, usability and interoperability testing, demonstrate that the modified devices do not raise different questions of safety and effectiveness when compared to the predicate, perform as intended, and have performance characteristics that are substantially equivalent to the predicate devices." (Section 1.9 CONCLUSIONS)

    In conclusion, this 510(k) submission for the Philips IntelliVue Patient Monitor MX series primarily relies on demonstrating compliance with recognized electrical safety, performance, and software development standards, along with a comparison to predicate devices, rather than an AI/algorithm performance study with specific metrics, expert adjudication, or MRMC studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210906

    Validate with FDA (Live)

    Date Cleared
    2021-10-07

    (195 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.

    The monitors are only for use on one patient at a time. They are not intended for use in transport situations. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The SSC Sepsis Protocol in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The 4-Slot Module rack FMX-4 is intended to connect up to four particular plug-in physiological measurement modules to the dedicated host patient monitors.

    Device Description

    The IntelliVue Patient Monitors MX750 and MX850 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. They operate with the external Measurement Modules and the IntelliVue 4-Slot Module Rack FMX-4, which establishes the connection between the individual plug-in measurement modules and the MX750 and MX850 monitors.

    The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, Sv02, ScvO2, EEG, BIS, NMT, and gas analysis.

    The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located in the patient vicinity at the bedside. These devices have a color display with a touch-screen as a primary input device. They also support keyboard and pointing devices such as a mouse.

    The monitor models MX750 and MX850 differ mainly in size. Whilst MX750 has a 19" flat panel display, MX850 has a 22" display.

    The hardware of the new 4-Slot Module Rack FMX-4 is very similar to that of its 8 years old predicate 4-Slot Module Rack FMS-4. The new FMX-4 has the same housing, which is made of the same material as FMS-4 but has a slightly different shade of color (same shade of color as the new MX750 and MX850 monitors). The FMX-4 has a new Printed Circuit Assembly with standard hardware components like CPU and memory being replaced by state of the art electronic components.

    The software modifications comprise the following changes:

    • Support of the new state-of-the-art electronic components.
    • Modified elements of the Graphical User Interface.
    • Implementation of a feature called Electronic Strip Recording. This feature allows electronic strips that can be reviewed on the monitor and printed out as a report on the printer.
    • Modification of the existing feature Remote Applications to support HTML5 and PDF format.
    AI/ML Overview

    This document is a 510(k) summary for the Philips IntelliVue Patient Monitors MX750 and MX850 and the IntelliVue 4-Slot Module Rack FMX-4. It states that these devices are substantially equivalent to previously cleared predicate devices (Philips IntelliVue Patient Monitor MX800 and Philips IntelliVue 4-Slot Module Rack FMS-4). The submission primarily focuses on hardware and software modifications to existing devices.

    The information provided does not detail specific acceptance criteria and device performance in the format of a table with numerical values for metrics like sensitivity, specificity, or accuracy, as would be common for an AI/ML device study. Instead, it describes compliance with recognized consensus standards and general V&V activities. This is likely because the device is a patient monitor, and the modifications are primarily hardware updates and minor software changes to support the new hardware and improve existing functionalities, not a new AI/ML algorithm requiring extensive clinical performance validation against a pre-defined ground truth for diagnostic accuracy.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC comparative effectiveness study, standalone performance) are not explicitly addressed in the provided text as they pertain more to novel diagnostic or AI algorithms rather than general patient monitoring device updates.

    I can, however, extract information about the types of tests conducted and general statements about meeting requirements.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific numerical acceptance criteria and corresponding device performance metrics (e.g., sensitivity, specificity, AUC) for the various physiological parameters monitored by the device. Instead, it states that "Test methods and acceptance criteria were the same as those for the predicate devices and test results showed substantial equivalence with respect to safety and effectiveness."

    The V&V activities focused on compliance with various international standards:

    Acceptance Criteria (Compliance with Standards)Reported Device Performance
    AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) (Electrical Safety)All applicable requirements have been met.
    IEC 60601-1-2:2014 (Ed. 4) (EMC)All applicable requirements have been met.
    IEC 60601-2-23:2011 (Ed.3) (Non-invasive sphygmomanometers)All applicable requirements have been met.
    IEC 60601-2-25:2011 (Ed. 2) (Electrocardiographs)All applicable requirements have been met.
    IEC 60601-2-27:2011(Ed. 3) (Electrocardiographic monitoring equipment)All applicable requirements have been met.
    IEC 80601-2-30:2009 (Ed. 1.1) and A1:2013 (Automated non-invasive sphygmomanometers)All applicable requirements have been met.
    IEC 60601-2-34:2011(Ed.3) (Invasive blood pressure monitoring equipment)All applicable requirements have been met.
    ISO 80601-2-55:2018 (Ed. 2) (Respiratory gas monitors)All applicable requirements have been met.
    ISO 80601-2-56:2017 (Ed. 2) and A1:2018 (Clinical thermometers)All applicable requirements have been met.
    ISO 80601-2-61:2017 (Ed. 2) and Corr1:2018 (Pulse oximetry equipment)All applicable requirements have been met.
    IEC 60601-1-8:2006 (Ed. 2.1) and A1:2012 (Alarm systems)All applicable requirements have been met.
    IEC 62304:2006 (Ed. 1.1) and A1:2015 (Software lifecycle processes)All applicable requirements have been met.
    IEC 60601-1-6:2010 (Ed.3.1) and A1:2013 (Usability)All applicable requirements have been met.
    Environmental Testing (Mechanical: Shock, vibration, free fall; Climatic: Temp, humidity)Specified test requirements have been met.
    Human Factors Engineering Testing (GUI evaluation)All specified test requirements have been met and no new hazards have been identified. User feedback led to iterative UI improvements.
    WLAN Coexistence TestingThe specified pass/fail criteria has been met.
    Software tests (Safety risk, Software specifications, Security risk)All tests have been passed.
    Overall ConclusionThe results demonstrate that the Philips IntelliVue Patient Monitors MX750, MX850 and the IntelliVue 4-Slot Module Rack FMX-4 meet all defined reliability requirements and performance claims.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in terms of subject count or data record count for performance evaluation in the context of, for example, diagnostic algorithm accuracy. The studies described are primarily engineering validation and verification tests against established standards. Therefore, information regarding data provenance (e.g., country of origin, retrospective/prospective) is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As there is no "test set" described for a diagnostic algorithm, this information is not applicable and not provided in the document. The human factors engineering testing involved "focus groups" and "expert reviews," but the number and qualifications of these experts are not specified, nor were they establishing ground truth for a diagnostic outcome.

    4. Adjudication method for the test set

    Not applicable, as no dedicated "test set" requiring adjudication for ground truth of a diagnostic outcome is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    No MRMC comparative effectiveness study is mentioned. The device is a patient monitor, and the submission concerns updates to an existing monitoring platform, not a new AI diagnostic aid that would typically warrant such a study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is less applicable to a patient monitoring device update. The performance evaluation focuses on the device's ability to accurately measure and display physiological parameters and generate alarms in compliance with safety and performance standards, rather than an autonomous diagnostic algorithm. The document states the monitors are "for use by trained healthcare professionals."

    7. The type of ground truth used

    For the various measurement parameters (ECG, SpO2, etc.), the "ground truth" implicitly aligns with the accuracy and performance specifications outlined in the referenced standards (e.g., AAMI EC11 for ECG, ISO 80601-2-61 for pulse oximetry). These standards define acceptable deviations from known physical or simulated physiological signals. The document does not describe the establishment of ground truth in the context of expert consensus, pathology, or outcomes data, as would be relevant for a diagnostic AI algorithm.

    8. The sample size for the training set

    This information is not provided. The document primarily describes hardware and software updates to an existing system, rather than the development and training of a new AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as the document does not describe the development or training of a new AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212208

    Validate with FDA (Live)

    Date Cleared
    2021-09-30

    (77 days)

    Product Code
    Regulation Number
    870.2450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for generation of a patient record.

    The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified measurements, Philips Patient Monitors and qualified 3rd party measurements that are connected through networks.

    Device Description

    The IntelliVue GuardianSoftware is a stand-alone 'Clinical Information Management System (CIMS)' software application with client-server architecture and designed to be used in professional healthcare facilities (i.e. hospitals, nursing homes) and is intended to be installed on a 'customer-supplied', compatible off-the-shelf (OTS) information technology (IT) equipment.

    The IntelliVue GuardianSoftware is a documentation, charting, and decision-support software that is configurable by the hospital to suit the needs of individual clinical units. The device collects data/vital signs from the following Philips compatible patient monitor/measuring devices.

    Using the collected data, the device provides trending, review, reporting and notification. The 'Guardian Early Warning Score (EWS)' application is integrated into the IntelliVue GuardianSoftware to provide the healthcare professional/provider basic assessment and the ability to recognize early signs of deterioration in patients.

    The IntelliVue GuardianSoftware is not an alarming device and displays alarms from patient monitors as supplemental information only.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not contain specific quantitative acceptance criteria or a side-by-side performance comparison of the new device versus these criteria. Instead, the document focuses on demonstrating substantial equivalence to a predicate device. The acceptance is implied by the successful software verification and validation, as well as human factors testing, which collectively show the device "met all safety and reliability requirements and performance claims."

    The closest we get to "performance" in the context of the comparison is in the "Substantial Equivalence Determination" column, which consistently states "IDENTICAL" or provides explanations for differences that do not affect substantial equivalence.

    Key areas assessed for equivalence (and thus, implicit "performance" against the predicate):

    FeaturePredicate Device (Rev. D.0)Subject Device (Rev. E.0X)Acceptance/Equivalence Determination
    Intended UseIdenticalIdenticalSubstantially Equivalent (IDENTICAL)
    Indications for UseCollection, storage, management of data from Philips specified measurements & Patient monitors.Collection, storage, management of data from Philips specified measurements, Patient Monitors, and qualified 3rd party measurements.Substantially Equivalent (Difference in indications for use does not affect substantial equivalence, HL7 testing verified 3rd party measurements.)
    System PlatformClient Server Architecture, Microsoft OS, OTS IT equipmentIdenticalSubstantially Equivalent (IDENTICAL)
    Operating System(s) & DatabaseWindows 7/8.1/10, Win Server 2008R2/2012R2/2016, SQL 2014/2016/2017, Android 4.4+Windows 8.1/10, Win Server 2012R2/2016/2019, SQL 2014/2016/2017, Android 5.0+Substantially Equivalent (Updates to OS versions (removal of unsupported, addition of newer) ensure continued support and do not affect substantial equivalence.)
    Programming LanguageMicrosoft® .NET C#, Microsoft® .NET C++, Java (mobile client)IdenticalSubstantially Equivalent (IDENTICAL)
    Maximum # of Supported Patients/Servers/ClientsPatients: 1200, Servers: 120, Clients: 240, SW Clients: 40IdenticalSubstantially Equivalent (IDENTICAL)
    Compatible DevicesPhilips IntelliVue Cableless Measurements, MP5/MP5SC, MX400/XG50, SureSigns VS3/VS4, Biosensor EarlySense Insight DeviceAdds Philips EarlyVue VS30 (K190624) and Philips Biosensor BX100 (K192875)Substantially Equivalent (Addition of new patient monitoring devices does not affect substantial equivalence.)
    Software Functionality (General Overview)Clinical Documentation, Patient Data Management, Reporting (SBAR), Calculations (Protocol Watch, EWS Scoring), Clinical decision support, StorageIdenticalSubstantially Equivalent (IDENTICAL)
    System Interfaces (IT Network Requirements)Hospital IT (W)LAN infra, HL7, ADT, Labs, PagingAdds HL7 data import extensionSubstantially Equivalent (Addition of HL7 data import expands compatibility to 3rd party systems and does not affect substantial equivalence.)
    Device InterfacesInternal interface for connection to measuring devices via hospital LANIdenticalSubstantially Equivalent (IDENTICAL)
    Remote Viewing/OperationIndependent display/operating interface, operations from host measuring device, PC UI (mouse/touchscreen), XDS Infrastructure ServiceIdenticalSubstantially Equivalent (IDENTICAL)

    2. Sample size used for the test set and the data provenance:

    The document does not specify a "test set" in terms of patient data. The evaluation relies on:

    • Software Verification and Validation Testing: This is typically performed on software builds and simulated environments, not directly on patient data.
    • Human Factors and Usability Testing: This involves human users interacting with the device. The sample size for this specific testing is not mentioned.
    • Data Provenance: Not applicable in the context of patient data testing, as no patient data was used for performance evaluation of the software itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. Since the evaluation focused on software verification/validation and human factors/usability, and not on clinical performance with patient data requiring expert ground truth, this type of detail is absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    No adjudication method is mentioned, as there was no test set requiring expert adjudication for clinical ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was conducted or mentioned. The device is a "Clinical Information Management System" and not an AI-assisted diagnostic tool for human readers. It collects, stores, and manages data, and provides decision support (e.g., Early Warning Score), but it doesn't appear to directly assist human readers in interpreting medical images or other complex data where "improvement" with AI would be measured.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is inherently a "Clinical Information Management System (CIMS) software application" that operates in a standalone manner as software. However, its function as data collection, storage, management, and providing decision support (like EWS) means it is intended for use by healthcare providers and is integrated into clinical workflows. The performance testing focuses on its software functionality, reliability, and human factors, rather than a diagnostic algorithm's standalone performance. The "Guardian Early Warning Score (EWS)" is an algorithm, and its performance would be assessed for accuracy in calculating scores, but the document doesn't provide details on its standalone performance metrics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the software verification and validation, the "ground truth" would be the expected functional behavior and output of the software as defined by its requirements and specifications. For human factors testing, the ground truth would be the expected safe and effective interaction of users with the device. There isn't an external clinical ground truth (like pathology or outcomes) applied to the software's performance itself in this submission.

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning model where a specific training set (of patient data) would be used. The software is developed based on engineering principles and regulatory requirements.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192137

    Validate with FDA (Live)

    Date Cleared
    2020-09-11

    (401 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.

    The monitors are additionally intended for use in transport situations within hospital environments.

    The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

    Device Description

    The IntelliVue Patient Monitors MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, invasive and noninvasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, spirometry, EEG, BIS, NMT, and gas analysis.

    The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals.

    The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display.

    The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Philips IntelliVue Patient Monitors MX500 and MX550 based on the provided FDA 510(k) summary:

    This submission is a 510(k) for modifications to an existing device, the IntelliVue Patient Monitors MX500 and MX550. The acceptance criteria and supporting studies primarily focus on demonstrating that the modifications do not introduce new questions of safety or effectiveness and that the modified device remains substantially equivalent to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is an update to an existing device focusing on integrating new measurement modules (Masimo O3, IRMA CO2, ISA CO2) and updating EMC standards, the "acceptance criteria" are primarily related to meeting established performance standards for patient monitors and demonstrating that the new modules function as intended without compromising the overall device safety or efficacy. The document doesn't present specific numerical acceptance criteria for all physiological parameters of the overall monitor, but rather confirms compliance with recognized standards.

    Acceptance Criteria CategoryDetails/Standard AdherenceReported Device Performance
    General Safety and PerformanceDemonstrated compliance with harmonized standards.Passed all safety tests for demonstrated compliance with harmonized standards.
    Software Life Cycle ProcessesIEC 62304:2015 Edition 1.1 (consolidated version) - Medical device software - software life cycle processes.Complies. Software Verification and Validation testing was conducted, and documentation was provided as recommended by FDA guidance. Software Level of Concern determined to be Major.
    Basic Safety and Essential PerformanceAAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) - Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.Complies. Electrical safety testing conducted.
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014 (Ed. 4.0) - Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.Complies. EMC testing conducted. Note: The predicate device conformed to Ed. 3; the subject device updates to Ed. 4.
    Alarm SystemsIEC 60601-1-8:2012 (Ed. 2.1) - Medical Electrical Equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems.Device passed tests for this standard.
    Respiratory Gas MonitorsISO 80601-2-55:2011 (Ed. 1) - Medical Electrical Equipment - Part 2-55: Particular Requirements For The Basic Safety And Essential Performance Of Respiratory Gas Monitors.Device passed tests for this standard.
    System Level Device SpecificationsBench testing for system level device specifications.Verified through bench testing.
    Mechanical & Electrical SpecificationsBench testing for mechanical and electrical specifications.Verified through bench testing.
    Packaging IntegrityBench testing for packaging integrity.Verified through bench testing.
    Shelf LifeNot applicable, as devices do not contain aging components.No shelf life specified.
    SterilityNot applicable, as devices are not sterile.Sterilization not evaluated.
    BiocompatibilityNot applicable, as devices do not have patient contact.Biocompatibility requirements are not applicable. Legally marketed medical accessories remain unchanged, so biocompatibility aspects are not affected.
    Functionality of Added Masimo O3 ModuleIntended use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region for adults ≥40 kg and pediatrics ≥5 kg and <40 kg. (Reference Device: K162603)The added Masimo O3 module's intended use fully aligns with the reference device, and its integration does not alter the substantial equivalence of the overall device. Received physiological data and forward it into existing SW infrastructure for presentation, data storage, user interface and alarming.
    Functionality of Added Masimo IRMA/ISA CO2Intended use as a gas analyzer for Carbon Dioxide. (Reference Devices: K123043, K171121)The added IRMA/ISA CO2 measurements align with the reference devices and their integration does not alter the substantial equivalence of the overall device. Received physiological data and forward it into existing SW infrastructure for presentation, data storage, user interface and alarming.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "The IntelliVue Patient Monitors MX500 and MX550, like the predicate device, did not require clinical trials."

    Therefore, there is no "test set" in the traditional sense of a clinical study involving patients or human readers. The evaluation was based on non-clinical bench testing and compliance with harmonized standards.

    • Sample Size for Test Set: Not applicable as no clinical test set was used.
    • Data Provenance: Not applicable. The data is from in-house bench testing and compliance evaluations against recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. As no clinical trials or human-in-the-loop performance studies were conducted for this submission, there was no need for experts to establish ground truth in the context of a test set involving patient data or diagnostic interpretations. The ground truth for the functionality of the integrated Masimo modules is established by their prior clearance as standalone devices (reference devices K162603, K123043, K171121).

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission states, "The IntelliVue Patient Monitors MX500 and MX550, like the predicate device, did not require clinical trials." This indicates that there was no study involving human readers or AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    While there isn't an "algorithm only" study defined in the context of AI, the performance of the integrated components (Masimo O3, IRMA CO2, ISA CO2 modules) in receiving and forwarding physiological data was evaluated. This can be seen as a form of standalone testing of the integration without human interpretation. The original Masimo devices (reference devices) would have undergone their own standalone validation studies prior to their clearances (K162603, K123043, K171121), which are implicitly relied upon here. This submission focuses on the integration into the Philips monitor.

    7. Type of Ground Truth Used

    For the integrated measurement modules:

    • Previous FDA clearances of reference devices: The performance and safety of the Masimo O3 Regional Oximeter System, Masimo Infrared Mainstream Gas Analyzer CO2, and Masimo Infrared Sidestream Gas Analyzer CO2 were established during their original 510(k) clearances (K162603, K123043, K171121). This forms the "ground truth" for their individual measurement capabilities.
    • Compliance with harmonized standards: For the overall monitor and its specific functions (e.g., electrical safety, EMC, alarm systems), the ground truth is defined by meeting the requirements of internationally recognized standards (e.g., IEC 60601 series, ISO 80601-2-55).
    • Bench testing results: In-house bench testing validated the system-level specifications, mechanical/electrical aspects, and software functionality according to pre-defined internal test protocols and specifications derived from design requirements.

    8. Sample Size for the Training Set

    Not applicable. This device is a patient monitor with physiological measurement capabilities, not an AI/machine learning device that requires a training set in the typical sense. Its functionality is based on established engineering principles and validated measurement technologies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

    The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.

    The MP5 and MP5SC monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor is for use during patient transport outside of the hospital environment. The MP5 and MP5SC when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.

    The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.

    The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

    ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

    The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment.

    The SSC Sepsis Protocol, in the Protocol Watch clinical decision support tool, is intended for use with adult patients only.

    The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

    The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

    The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

    BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

    Device Description

    The IntelliVue Patient Monitors MP5, MP5SC, MX100, MX400, MX430, MX450, MX500, MX550, MX700, MX800 and IntelliVue Multi-Measurement Module X3 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.

    The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and noninvasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, Sv02, ScvO2, spirometry, EEG, BIS, NMT, and gas analysis.

    The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located in the patient vicinity at the bedside. Several monitor models can also be used mobile, during patient transport inside or outside of hospitals.

    The monitors have a color display with touch-screen and/or keys and a navigation point as a primary input device. They also support further local input devices such as specialized remote control, keyboard, and mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display.

    The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

    With the current software Rev. N.0 the following modifications have been introduced:

    • Implementation of the existing feature Alarm Advisor into the IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3. Alarm Advisor provides feedback on recurring and continuous alarm limit violations based on configured criteria. The information provided by the Alarm Advisor supports device operator in adapting alarm limits more specifically. This is the same functionality as that in other IntelliVue Patient Monitors: MP5, MP5SC, MX400, MX430, MX450, MX500, MX550, MX700, and MX800 (cleared with K161531).
    • Modification of a few specific elements of the Graphical User Interface (GUI) of the IntelliVue Patient Monitors MP5, MP5SC, MX100, MX400, MX430, MX450, MX500, MX550, MX700, MX800, and the Intelli Vue Multi-Measurement Module X3:
      • The key 'Silence' has been renamed to 'Acknowledge',
      • A new default configuration for the Pause/Switch off of alarms via 'Acknowledge' key combined with a specific pop-up window has been added to all monitor models,
      • New default configurations of visual alarm indicators with other colors and flashing behavior of the alarm numerics and limits have been added.
    AI/ML Overview

    The provided document is a 510(k) premarket notification from Philips Medizin Systeme Boeblingen GmbH to the FDA for their IntelliVue Patient Monitors. This document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices and addressing minor software modifications, specifically the implementation of an "Alarm Advisor" feature and GUI changes related to alarm management.

    The document does not contain the specific details required to answer all parts of your request, particularly regarding clinical performance studies with specific acceptance criteria, sample sizes for test and training sets, expert consensus, or MRMC studies for an AI device. This is because the submission is for a patient monitor and its software updates, not an AI-powered diagnostic device in the sense that would require such extensive clinical validation to demonstrate improved human reader performance or standalone algorithm performance.

    However, I can extract information related to the acceptance criteria and study proving the device meets its acceptance criteria, based on the scope of this type of submission.

    Here's an analysis based on the provided text, while acknowledging the limitations for a full AI device performance study:

    Acceptance Criteria and Device Performance (as evident from this 510(k) submission):

    For this type of device (patient monitors with software updates), the "acceptance criteria" and "device performance" are primarily demonstrated through compliance with recognized standards, hazard analysis, and functional/regression testing rather than a clinical trial with a defined performance metric like sensitivity/specificity for a diagnostic AI.

    Acceptance Criterion (Implicit/Explicit)Reported Device Performance (Summary)
    Safety: Device operates without introducing new hazards or increasing existing risks."All specified pass/fail criteria have been met. The test results confirmed the effectiveness of the implemented design risk mitigation measures." "All specified test requirements have been met and no new hazards have been identified."
    Usability: User interface is intuitive and effective, especially for new features or changes."Evaluated and improved iteratively during the design phase conducting several formative usability evaluations. They included established usability engineering methods like focus groups, expert reviews and usability tests using user interface prototypes. User feedback was translated into iterative user interface improvements."
    Functionality: New features (Alarm Advisor, GUI changes) operate as designed and specified."Functional tests... with feature Alarm Advisor at System Level. All tests have been passed." "Functional tests... with the feature blinking behavior of numerics at System Level. All tests have been passed." "Functional tests... with the feature Alarm Acknowledge at System Level. All tests have been passed." "Functional tests... with the feature Pause/Switch off Alarms at System Level. All tests have been passed. The performed functional tests demonstrate that the new features... are correctly presented on the display, can be correctly operated, controlled, configured, and function as specified and according to the labeling claims."
    Reliability/Maintain Existing Functionality: Unchanged functions continue to work correctly after software modifications."Regression tests... All tests have been passed. The performed regression tests demonstrate that the unchanged and not affected functions also work correctly and in accordance with all specifications and labeling claims in the modified software."
    Compliance with Standards: Adherence to relevant medical device and software standards."Testing according to the recognized consensus standard: IEC 60601-1-8: 2012 (Ed. 2.1) (Alarms) - All applicable requirements have been met." "Compliance with the recognized consensus process standards: AAMI ANSI IEC 62304:2006 (Ed. 1) (Software life cycle processes); IEC 60601-1-6: 2013 (Ed. 3.1) (Usability). The modified devices are compliant with all applicable requirements of the above stated process standards."
    Substantial Equivalence: Device performs similarly to predicate devices."Test methods and acceptance criteria were the same as those for the predicate devices and test results showed substantial equivalence." "The results demonstrate that the Philips IntelliVue Patient Monitors... meet all defined reliability requirements and performance claims."

    Study Details (based on the provided document):

    Given that this is a 510(k) for patient monitors with minor software updates, the "study" is primarily a series of verification and validation (V&V) activities focused on engineering and software aspects, rather than a clinical trial for an AI diagnostic.

    1. Sample Size Used for the Test Set and Data Provenance:

      • The document does not specify a "test set" in the context of a dataset for an AI model. Instead, it refers to V&V activities that include:
        • Hazard Analysis Testing: "All specified pass/fail criteria have been met." (No specific sample size of incidents/scenarios mentioned, typically an engineering analysis).
        • Functional System Level Tests: Performed on a variety of monitor models (MP5, MP5SC, MX100, MX400, MX430, MX450, MX500, MX550, MX700, MX800, Multi-Measurement Module X3) for the new features (Alarm Advisor, blinking numerics, Alarm Acknowledge, Pause/Switch off Alarms). No specific number of test cases or "patients" is provided.
        • Regression Tests: Performed on the same range of monitor models to ensure unchanged functions still work.
      • Data Provenance: Not applicable in the context of clinical data for AI model evaluation. The tests are performed on the device itself and its software.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

      • Not described as "ground truth" establishment in this context. For usability testing, it mentions "expert reviews" as part of formative evaluations, but the specific number and qualifications of these experts are not detailed. These would typically be human factors engineers, clinical specialists, or design experts.
    3. Adjudication Method for the Test Set:

      • Not applicable as there is no "test set" of clinical cases requiring adjudication. The V&V activities are based on engineering specifications and standard compliance.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

      • No. An MRMC study is relevant for diagnostic imaging AI systems where human readers interpret medical images with and without AI assistance. This submission is for patient monitors and their alarm management and GUI features, which do not involve diagnostic interpretation in that manner. The "Alarm Advisor" provides feedback on alarm limit violations, which is a clinical decision support tool related to device settings, not an image interpretation aid.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

      • Yes, to an extent, in the sense that the new software features were functionally tested independently of human interaction (e.g., "Functional tests... All tests have been passed"). The "Alarm Advisor" functionality and GUI changes needed to perform as specified by the algorithm/software design. However, this is not a "standalone performance study" in the context of an AI diagnostic that might output a diagnosis or risk score. It's more about verifying software logic and output.
    6. The Type of Ground Truth Used:

      • Engineering Specifications and Standards Compliance. For the new features and existing functionalities, the "ground truth" is whether the software performs according to its design specifications, established industry standards (e.g., IEC 60601-1-8 for alarms), and internal quality requirements. For usability, "user feedback" (from focus groups, usability tests) serves as the basis for improvement.
    7. The Sample Size for the Training Set:

      • Not applicable. This device is not an AI model that undergoes a "training phase" from a clinical dataset in the traditional machine learning sense. The software development process involves design, coding, and V&V activities.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190624

    Validate with FDA (Live)

    Device Name
    EarlyVue VS30
    Date Cleared
    2019-10-27

    (230 days)

    Product Code
    Regulation Number
    870.1100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Early Vue VS30 is intended to measure, display, alarm and, record physiological information of adult, pediatric and neonatal patients in hospitals and in out-of-hospital patient care is administered by a healthcare professional (such as clinics, outpatient surgery facilities, long-term care facilities, and physician offices). It is not intended for use in mobile settings such as ambulances and aircraft. Clinical users may use the monitor during patient transport within a healthcare facility.

    The intended use of measurement parameters for each patient type is shown in table below (Parameter (patient types)): Non-invasive BP--NBP (Adult, Pediatric, Neonatal); SpO2 --Masimo rainbow SET or Philips FAST (Adult, Pediatric, Neonatal); Pulse Rate -- PR -- derived from SpO2 or NBP (Adult, Pediatric, Neonatal); Temperature-Predictive, Temporal (Adult, Pediatric, Neonatal); end tidal CO2--etCO2 (Adult, Pediatric, Neonatal); Respiration Rate: airway respiration--awRR (Adult, Pediatric, Neonatal); Respiration Rate: acoustic respiration--RRa (Adult, Pediatric);

    Non-invasive Total Hemoglobin--SpHb (Adult, Pediatric)

    Contra-indications: Not for transport outside the healthcare facility. Not for use in home setting.

    Device Description

    The subject device EarlyVue VS30 is a platform redesigned from the predicate device Suresign VS4 to provide the measurement of the same physiological parameters. It is a multi-parameter compact, portable monitor, offering several configurations and optional features that are intended to meet the customer's needs. The VS30 is designed with a simple, intuitive user interface for ease of operation. It is used to measure or monitor NBP, SpO2, SpHb, RRa, CO2 and optional temperature primarily in non-acute settings.

    The subject VS30 also provides the wireless option. The wireless functions for VS30 are for the monitors to communicate with the hospital EMR system or to communicate with Philips IGS software over the hospital's wireless infrastructure.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    Device: EarlyVue VS30 Vital Signs Monitor

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the EarlyVue VS30 are implied by its comparison to the predicate device, Philips SureSigns VS4. The document asserts that the VS30 has the "same fundamental technological characteristics" and that "performance data demonstrates the subject device Early Vue VS30 is as safe and effective as the predicate device for all the measurements." Therefore, the predicate device's specifications serve as the de facto acceptance criteria for each measurement parameter.

    ParameterAcceptance Criteria (Predicate Device VS4 Performance)Reported Device Performance (EarlyVue VS30)
    Non-invasive BP (NBP)Technique: Oscillometric, stepwise deflation. Measurement Range: Adult: Systolic 30-270mmHg, Diastolic 10-240mmHg, Mean 20-250mmHg; Pediatric: Systolic 30-180mmHg, Diastolic 10-150mmHg, Mean 20-160mmHg; Neonate: Systolic 30-130mmHg, Diastolic 10-100mmHg, Mean 20-120mmHg. Accuracy: Max Std. Dev: 8 mmHg, Max Mean Error: +/- 5 mmHg. Derived PR Accuracy: 40-100 bpm: +/- 5 bpm; 101-200 bpm: +/- 5% of reading; 201-300 bpm: +/- 10% of reading. Initial Cuff Inflation: Adult: 160 mmHg; Pediatric: 140 mmHg; Neonatal: 100 mmHg.Technique: Oscillometric, stepwise deflation. Measurement Range: Adult: Systolic 30-270mmHg, Diastolic 10-240mmHg, Mean 20-250mmHg; Pediatric: Systolic 30-180mmHg, Diastolic 10-150mmHg, Mean 20-160mmHg; Neonate: Systolic 30-130mmHg, Diastolic 10-100mmHg, Mean 20-120mmHg. Accuracy: Max Std. Dev: 8 mmHg, Max Mean Error: +/- 5 mmHg. Derived PR Accuracy: 40-100 bpm: +/- 5 bpm; 101-200 bpm: +/- 5% of reading; 201-300 bpm: +/- 10% of reading. Initial Cuff Inflation: Adult: 160 mmHg; Pediatric: 140 mmHg; Neonatal: 100 mmHg. Reported as "Identical"
    SpO2 (Philips SpO2)Measurement Range: SpO2: 0-100%, Pulse Rate: 30-300 bpm. PR Accuracy: Greater of +2% or +1 bpm. Resolution: SpO2: 1%. Accuracy (70-100%): Philips reusable sensors: +/- 2% (M1191B, BL, A, AL, M1192A), +/- 3% (M1191T, M1192T, M1193T adult, M1193A, M1194A, M1195A, M1196A, T, S), +/- 4% (M1193T neonatal). Philips disposable sensors: +/- 2% (M1132A, M1133A, M1134A), +/- 3% (M1131A).Measurement Range: SpO2: 0-100%, Pulse Rate: 30-300 bpm. PR Accuracy: Greater of +2% or +1 bpm. Resolution: SpO2: 1%. Accuracy (70-100%): Philips reusable sensors: +/- 2% (M1191B, BL, A, AL, M1192A), +/- 3% (M1191T, M1192T, M1193T adult, M1193A, M1194A, M1195A, M1196A, T, S), +/- 4% (M1193T neonatal). Philips disposable sensors: +/- 2% (M1132A, M1133A, M1134A), +/- 3% (M1131A). Reported as "Identical"
    SpO2 (Masimo SpO2)Measurement Range: 0-100%. Resolution: 1%. Accuracy: No motion: 60-80 +/- 3% (adult/ped/inf), 70-100 +/- 2% (adult/ped/inf), 70-100 +/- 3% (neonates). Motion: 70-100 +/- 3% (all). Low perfusion: 70-100 +/- 2% (all).Measurement Range: 0-100%. Resolution: 1%. Accuracy: No motion: 60-80 +/- 3% (adult/ped/inf), 70-100 +/- 2% (adult/ped/inf), 70-100 +/- 3% (neonates). Motion: 70-100 +/- 3% (all). Low perfusion: 70-100 +/- 2% (all). Reported as "Identical"
    Pulse Rate (Masimo SpO2)Range: 25-240 bpm. Resolution: 1 bpm. Accuracy: +/- 5 bpm motion or low perfusion (worst case).Range: 25-240 bpm. Resolution: 1 bpm. Accuracy: +/- 5 bpm motion or low perfusion (worst case). Reported as "Identical"
    SpHb (Masimo)Range: 0-25 g/dL. Accuracy: 8-17 g/dL (arterial/venous) adults/pediatrics: +/- 1 g/dL. Resolution: 0.1 g/dL.Range: 0-25 g/dL. Accuracy: 8-17 g/dL (arterial/venous) adults/pediatrics: +/- 1 g/dL. Resolution: 0.1 g/dL. Reported as "Identical"
    Respiration Rate (RRa - Masimo)Range: 0-70 bpm. Accuracy: 4-70 bpm: +/- 1 bpm (adults >30kg and adults/pediatrics >10kg). Resolution: 1 bpm.Range: 0-70 bpm. Accuracy: 4-70 bpm: +/- 1 bpm (adults >30kg and adults/pediatrics >10kg). Resolution: 1 bpm. Reported as "Identical"
    Temperature (Predictive)Measurement Range: 26.7–43.3°C (80–110°F). Accuracy (monitoring mode): +/- 0.1°C (+/- 0.2°F). Measurement Time: Oral: 4-6s, Adult axillary: 12-15s, Pediatric axillary: 10-13s, Rectal: 10-13s. Resolution: 0.1°C. Measurement Sites: Oral, rectal, axillary.Measurement Range: 26.7–43.3°C (80–110°F). Accuracy (monitoring mode): +/- 0.1°C (+/- 0.2°F). Measurement Time: Oral: 4-6s, Adult axillary: 12-15s, Pediatric axillary: 10-13s, Rectal: 10-13s. Resolution: 0.1°C (+/- 0.1°F). Measurement Sites: Oral, rectal, axillary. Reported as "Identical"
    Temperature (Temporal)Range: 16°C - 43°C. Clinical Accuracy: +/- 0.2°F or 0.1°C per ASTM E1112. Arterial heat balance range: 34.5°C - 43°C (Normal: 35.9°C to 37.5°C). Resolution: 0.1°C or 0.1°F. Response Time: <=1 second.Range: 16°C - 43°C. Clinical Accuracy: +/- 0.2°F or 0.1°C per ASTM E1112. Arterial heat balance range: 34.5°C - 43°C (Normal: 35.9°C to 37.5°C). Resolution: 0.1°C or 0.1°F. Response Time: <=1 second. Reported as "Identical"
    end tidal CO2 (etCO2)Measurement Range: 0-150 mmHg. Sampling Rate: 20 samples/second. Flow Rate: 50ml/min, +15ml/min, -7.5ml/min. Warm-up Time: 30 seconds (typical). Calibration Interval: Initial: 1200 hrs or 1 year; Subsequent: 1 year or 4000 hrs. Accuracy: 0-38 mmHg: +/- 2 mmHg; 39-150 mmHg: +/-(5% of reading + 0.08% for every 1 mmHg above 38 mmHg). Response Time: ~3.9 seconds.Measurement Range: 0-150 mmHg. Sampling Rate: 20 samples/second. Flow Rate: Same. Warm-up Time: 30 seconds (typical). Calibration Interval: Initial: 1200 hrs or 1 year; Subsequent: 1 year or 4000 hrs. Accuracy: 0-38 mmHg: +/- 2 mmHg; 39-150 mmHg: +/-(5% of reading + 0.08% for every 1 mmHg above 38 mmHg). Response Time: ~3.9 seconds. Reported as "Identical"
    Respiration Rate (awRR - CO2)Range: 0-150 respirations/minute. Accuracy: +/- 1 rpm (0-70 rpm); +/- 2 rpm (71-120 rpm); +/- 3 rpm (121-150 rpm).Range: 0-150 respirations/minute. Accuracy: +/- 1 rpm (0-70 rpm); +/- 2 rpm (71-120 rpm); +/- 3 rpm (121-150 rpm). Reported as "Identical"
    Battery operation time3 hours6 hours (Improved)
    User Interface8.4" resistive touchscreen, fixed key switches.10.1" LCD touchscreen, soft keys. Verified and validated, passed verification and validation.
    Monitor-to-monitor communicationNo (for predicate)Yes (Non-medical application, allows viewing saved patient records remotely, viewing last saved record for up to 100 patients, viewing trends. Fully verified and passed tests.)
    RFID scannerNo (for predicate)Yes (Non-medical application, entry of patient ID and user ID. Fully verified and passed tests.)
    Operating Temperature10°C to 40°C monitor (16°C to 33°C with tympanic, 0°C to 50°C Masimo, 16°C to 40°C Exergen, 10°C to 40°C WelchAllyn)10°C to 40°C monitor (10°C to 40°C WelchAllyn, 16°C to 40°C Exergen) Reported as "Same"
    Software Operating SystemWin CELinux (recoded software). Implied acceptance through comprehensive software verification and validation.
    Hardware ComponentsVS4 hardwareNew CPU, PCAs. Implied acceptance through comprehensive hardware verification and validation.
    Wireless ModuleVS4 wireless moduleDifferent from VS4, but from same supplier, has FCC certificate and ID. Tested to and passed ANSI C63.27-2017.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes (number of patients, number of measurements) used for performance testing (e.g., NBP accuracy, SpO2 accuracy). It generally refers to "verification and validation" and "performance data."

    For the SpO2 and SpHb measurements using Masimo technology, the document references the Masimo RRa respiration rate K110028 and the Oridion MicroMediCO2 module K123690. The accuracy claims for these integrated modules would presumably be based on studies performed during their original clearance, but specific details about those test sets are not in this document.

    The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information on the number of experts or their qualifications for establishing ground truth for the test set. For physiological measurements like NBP, SpO2, and temperature, ground truth is typically established using reference methods (e.g., arterial catheter for NBP, co-oximetry for SpO2, calibrated thermometer for temperature) rather than expert consensus on interpretive tasks.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method used for the test set, as this is typically applicable to subjective interpretations by multiple experts (e.g., radiology reads), not direct physiological measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable to this submission. The EarlyVue VS30 is a vital signs monitor that directly measures physiological parameters. It does not involve "human readers" interpreting medical images or other data with or without "AI assistance." The device's function is to provide direct measurements, not to aid in human diagnostic interpretations.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the primary function of the EarlyVue VS30 is as a standalone device that performs physiological measurements without a human-in-the-loop for the measurement process itself. The "performance data" referred to throughout the document would be standalone performance of the device's measurement functions.

    7. The Type of Ground Truth Used

    The ground truth for the physiological measurements (NBP, SpO2, Temperature, CO2) would typically be established using:

    • Reference Devices/Methods:
      • NBP: Invasive arterial blood pressure measurements or comparison to a validated reference sphygmomanometer.
      • SpO2: Co-oximetry for arterial oxygen saturation.
      • Temperature: Calibrated reference thermometers (e.g., mercury-in-glass, highly accurate electronic thermometers).
      • CO2: Calibrated gas analyzers.

    The document states that the technologies used are "identical" to the predicate device, which in turn relied on established measurement technologies (Philips picoNBP, Masimo Rainbow SET, Oridion MicroMediCO2 module). These underlying technologies would have their own validation against established ground truth methods.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for a training set. This is because the EarlyVue VS30 is not described as an AI/ML-based device that learns from a training set in the typical sense (e.g., for image classification or prediction). Its functions are based on established physiological measurement principles and algorithms. The software was "recoded" compared to the predicate, implying engineering-based development and verification, rather than a data-driven training process.

    9. How the Ground Truth for the Training Set Was Established

    Since no explicit training set is mentioned for an AI/ML context, this question is not directly applicable. The device's "ground truth" refers to its accuracy against established medical standards for physiological measurements. The document outlines a comprehensive verification and validation process (software, hardware, environmental, reliability, safety, performance, applicable consensus standards: IEC 60601 series, ISO 80601 series, etc.) to ensure the device performs as intended and is as safe and effective as the predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3