Search Results
Found 11 results
510(k) Data Aggregation
(159 days)
Ask a specific question about this device
(93 days)
VitruvianScan is indicated for use as a magnetic resonance diagnostic device software application for non-invasive fat and muscle evaluation that enables the generation, display and review of magnetic resonance medical image data.
VitruvianScan produces quantified metrics and composite images from magnetic resonance medical image data which when interpreted by a trained healthcare professional, yield information that may assist in clinical decisions.
VitruvianScan is a standalone, post processing software medical device. VitruvianScan enables the generation, display and review of magnetic resonance (MR) medical image data from a single timepoint (one patient visit).
When a referring healthcare professional requests quantitative analysis using VitruvianScan, relevant images are acquired from patients at MRI scanning clinics and are transferred to the Perspectum portal through established secure gateways. Perspectum trained analysts use the VitruvianScan software medical device to process the MRI images and produce the quantitative metrics and composite images. The device output information is then sent to the healthcare professionals for their clinical use.
The metrics produced by VitruvianScan are intended to provide insight into the composition of muscle and fat of a patient. The device is intended to be used as part of an overall assessment of a patient's health and wellness and should be interpreted whilst considering the device's limitations when reviewing or interpreting images.
Here's an analysis of the acceptance criteria and the study provided in the document for the VitruvianScan (v1.0) device:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document unfortunately does not explicitly state the specific quantitative acceptance criteria for each performance aspect (e.g., a specific percentage for repeatability, a particular correlation coefficient). Instead, it states that "All aspects of the performance tests met the defined acceptance criteria." and that the device "successfully passed the acceptance criteria with no residual anomalies."
However, based on the described performance tests, we can infer the types of acceptance criteria that would have been defined. The document also lacks specific numerical results for the device performance that directly map to these criteria.
| Performance Aspect | Inferred Acceptance Criteria (Example) | Reported Device Performance |
|---|---|---|
| Repeatability of metrics | Coefficient of Variation (CV) or Intraclass Correlation Coefficient (ICC) for various metrics (Visceral Fat, Subcutaneous Fat, Muscle Area) to be within a pre-defined threshold for the same subject, scanner, field strength, and day. | "met the defined acceptance criteria" |
| Reproducibility of metrics | Coefficient of Variation (CV) or Intraclass Correlation Coefficient (ICC) for various metrics (Visceral Fat, Subcutaneous Fat, Muscle Area) to be within a pre-defined threshold for the same subject, scanner (different field strength), and day. | "met the defined acceptance criteria" |
| Inter-operator variability | Low variability (e.g., high ICC or low CV) in metric measurements between different trained operators using VitruvianScan. | "Characterization of inter-operator variability" met acceptance criteria |
| Intra-operator variability | Low variability (e.g., high ICC or low CV) in metric measurements by the same trained operator over repeated measurements. | "Characterization of intra-operator variability" met acceptance criteria |
| Benchmarking against reference device | Established equivalence or non-inferiority in metric measurements when compared to a validated reference regulated device (OSIRIX MD). | "Results...compared with the results from testing using reference regulated device 'OSIRIX MD' for benchmarking performance" met acceptance criteria |
| Comparison to gold standard (human experts) | High agreement (e.g., high ICC, low mean absolute error) between device output and the gold standard (mean of 3 radiologists' results). | "Comparative testing between the operators' results and the gold standard (mean of 3 radiologists results)" met acceptance criteria |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the test set.
The data provenance (country of origin, retrospective/prospective) is not explicitly stated. However, the context of an FDA submission for a device used in clinical settings suggests the data would likely be from a clinical or research environment, potentially multi-center, but this is an inference.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: 3 radiologists
- Qualifications of Experts: The document states "3 radiologists results" but does not provide specific qualifications (e.g., years of experience, subspecialty).
4. Adjudication Method for the Test Set
The adjudication method used for the test set is implicitly a "mean (average) of 3 radiologists results" for establishing the gold standard. This suggests a form of consensus, where the average of their interpretations serves as the reference. It's not a typical "X+1" method (like 2 out of 3, or 3 out of 1 for disagreement), but rather a central tendency measure of their assessments.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A MRMC comparative effectiveness study was NOT explicitly done in the sense of comparing human readers with AI assistance versus without AI assistance.
The study did involve "Comparative testing between the operators' results and the gold standard (mean of 3 radiologists results)," which evaluates the device's output and how operators use it against a human expert consensus. However, it doesn't describe a scenario where human readers improve with AI assistance in their own diagnostic performance compared to their performance without the AI. The stated use case is that "Perspectum trained analysts use the VitruvianScan software medical device to process the MRI images and produce the quantitative metrics and composite images," and then these are sent to "trained Healthcare Professionals who then utilize these to make clinical decisions." This suggests the device provides quantitative data to healthcare professionals, rather than directly assisting their image interpretation to improve their diagnostic accuracy from images alone.
6. Standalone (Algorithm Only) Performance Study
Yes, a standalone performance assessment was conducted. The "Comparative testing between the operators' results and the gold standard (mean of 3 radiologists results)" and the "benchmarking performance" against OSIRIX MD inherently assess the algorithm's output (via the trained analysts) against a reference, which signifies a standalone evaluation of the device's quantitative capabilities.
7. Type of Ground Truth Used
The primary type of ground truth used for the comparative testing was expert consensus (mean of 3 radiologists results).
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size used for the training set for the VitruvianScan algorithm. This information is typically crucial for understanding the generalizability and robustness of an AI/ML device.
9. How the Ground Truth for the Training Set Was Established
The document does not provide any information on how the ground truth for the training set was established.
Ask a specific question about this device
(90 days)
MRCP+ v2 is indicated for use as a software-based image processing system for noninvasive, quantitative assessment of biliary system structures.
MRCP+ v2 calculates quantitative three-dimensional biliary system models that enable the measurement of bile duct widths and quantification of strictures and dilatations. MRCP+ v2 includes tools for interactive segmentation and labelling of the biliary system, including the gallbladder.
MRCP+ v2 provides metrics for interpretation by trained physicians to assist in the assessment of the biliary system. These metrics support physicians in the visualization, evaluation and reporting of hepatobiliary structures.
MRCP+ v2 is designed to utilize DICOM-compliant MRCP datasets acquired on supported MR scanners using supported MRCP acquisition protocols.
MRCP+ v2 is a standalone software medical device. The purpose of the MRCP+ v2 device is to assist a trained operator with the quantitative evaluation of biliary system structures acquired from magnetic resonance (MR) images from a single time-point (a patient visit). A Perspectum-trained operator loads previously acquired magnetic resonance cholangiopancreatography (MRCP) data as input into the MRCP+ v2 device. A structured summary report is generated as output by the device, which includes quantitative analysis results and geometric characteristics of the biliary system and pancreatic ducts. The MRCP+ v2 report is intended to facilitate reporting by a radiologist for subsequent interpretation and to aid diagnosis by a physician as part of a panel of testing, including conventional radiological tools.
MRCP+ v2 is designed to utilize Digital Imaging and Communications in Medicine (DICOM)-compliant MRCP datasets acquired on supported MR scanners to display the fluid-filled tubular structures in the abdomen, including intra- and extra-hepatic biliary tree, gallbladder, and pancreatic duct system. Analysis and evaluation tools in MRCP+ v2 include segmentation of structures utilizing user input of seeding points, interactive labelling of segmented areas, and quantitative measurement derived from segmentation results. MRCP+ v2 calculates quantitative three-dimensional biliary system models that enable measurement of bile duct widths and automatic detection of regions of variation in the width of tubular structures, including those quantified as strictures and dilations. MRCP+ v2 includes tools allowing for regional volumetric analysis of segmented tree-like, tubular structures and the gallbladder.
Here's a breakdown of the acceptance criteria and study information for the Perspectum MRCP+ version 2 (MRCP+ v2) device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally expressed as Bland-Altman 95% Limits of Agreement (LoA) or maximum bias and other precision metrics. The document states that the device successfully met all predetermined acceptance criteria. Therefore, the "Reported Device Performance" effectively matches the "Acceptance criteria" as the device passed.
Table: Acceptance Criteria for MRCP+ v2 Performance
| Metric | Acceptance Criteria (from Tables 2, 3, 4, 5, 6, 7, 8, 9) |
|---|---|
| Individual Duct Metrics (Stricture & Dilatation Phantom) | |
| Duct length (mm) | Max bias: ±9.5, LoA: ±81.3 |
| Stricture length max (mm) | Max bias: ±1.1, LoA: ±14.6 |
| Stricture length mean (mm) | Max bias: ±1.1, LoA: ±14.3 |
| Stricture length sum (mm) | Max bias: ±1.8, LoA: ±17.8 |
| Dilatation length sum (mm) | Max bias: ±2.2, LoA: ±23.2 |
| Dilatation diameter max (mm) | Max bias: ±0.7, LoA: ±6.0 |
| Stricture absolute severity sum (mm) | Max bias: ±2.0, LoA: ±3.8 |
| Dilatation absolute severity sum (mm) | Max bias: ±3.3, LoA: ±5.5 |
| Dilatation absolute severity max (mm) | Max bias: ±1.1, LoA: ±3.8 |
| Stricture relative severity sum (%) | Max bias: ±18.6, LoA: ±76.7 |
| Dilatation relative severity sum (%) | Max bias: ±407.5, LoA: ±520.6 |
| Dilatation relative severity max (%) | Max bias: ±209.6, LoA: ±281.2 |
| Stricture score sum (mm) | Max bias: ±1.95, LoA: ±8.6 |
| Dilatation score sum (mm) | Max bias: ±33.3, LoA: ±49.6 |
| Number of strictures | Max bias: ±1.0, LoA: ±3.0 |
| Number of dilatations | Max bias: ±1.0, LoA: ±3.0 |
| Tube Width Measurements (Tube Width & Clinical Biliary Phantoms) | |
| Accuracy of algorithm - synthetic data: % points stably matched | > 80% |
| Accuracy of algorithm - synthetic data: Absolute value of mean bias | ≤ 0.4 mm |
| Accuracy of algorithm - synthetic data: Largest absolute LoA | ≤ 1.4 mm |
| Accuracy of algorithm - synthetic data: Slope of bias | ≤ 0.1 mm per mm |
| Accuracy of device - acquired data: % points stably matched | > 80% |
| Accuracy of device - acquired data: Absolute value of mean bias | ≤ 0.5 mm |
| Accuracy of device - acquired data: Largest absolute LoA | ≤ 1.5 mm |
| Accuracy of device - acquired data: Slope of bias | ≤ 0.1 mm per mm |
| Repeatability: % points stably matched | > 80% |
| Repeatability: Absolute value of mean bias | ≤ 0.4 mm |
| Repeatability: Largest absolute LoA | ≤ 1.4 mm |
| Repeatability: Slope of bias | ≤ 0.15 mm per mm |
| Reproducibility: % points stably matched | > 80% |
| Reproducibility: Absolute value of mean bias | ≤ 0.6 mm |
| Reproducibility: Largest absolute LoA | ≤ 1.6 mm |
| Reproducibility: Slope of bias | ≤ 0.15 mm per mm |
| Substantial equivalence (v1 vs v2): % points stably matched | > 80% |
| Substantial equivalence (v1 vs v2): Absolute value of mean bias | ≤ 0.6 mm |
| Substantial equivalence (v1 vs v2): Largest absolute LoA | ≤ 1.6 mm |
| Substantial equivalence (v1 vs v2): Slope of bias | ≤ 0.15 mm per mm |
| In Vivo Performance (Summary Whole Tree Metrics - LoA) | |
| Total number of ducts | ±116.2 |
| Total number of strictures | ±16.5 |
| Total number of dilatations | ±33.8 |
| Total number of ducts with a stricture or dilatation | ±27.4 |
| Total number of ducts with a stricture and dilatation | ±9.6 |
| Number of ducts with a stricture | ±12.4 |
| Number of ducts with a dilatation | ±24.5 |
| Duct length mean (mm) | ±12.3 |
| Duct length sum (mm) | ±2082.7 |
| Stricture length sum (mm) | ±121.4 |
| Dilatation length sum (mm) | ±215.8 |
| Abnormal length sum (mm) | ±326.3 |
| Stricture absolute severity sum (mm) | ±32.0 |
| Dilatation absolute severity sum (mm) | ±78.7 |
| Dilatation absolute severity max (mm) | ±4.5 |
| Stricture relative severity sum (%) | ±708.0 |
| Dilatation relative severity sum (%) | ±2437.6 |
| Dilatation relative severity max (%) | ±165.3 |
| Stricture score sum (mm) | ±51.7 |
| Dilatation score sum (mm) | ±179.2 |
| Dilatation diameter max (mm) | ±6.3 |
| Biliary tree volume | ±20 mL |
| Gallbladder volume | ±30 mL |
| Percentage of ducts with median width less than 3 mm | ±40% |
| Percentage of ducts with median width greater than 3 mm up to 5 mm | ±40% |
| Percentage of ducts with median width greater than 5 mm up to 9 mm | ±40% |
| Percentage of ducts with median width greater than 9 mm | ±40% |
| Precision: Repeatability and Reproducibility (Whole Tree Metrics - LoA) | (Same as In Vivo Performance) |
| Precision: Repeatability and Reproducibility (Single Duct Metrics - LoA) | |
| Total number of strictures | ±2.0 |
| Total number of dilatations | ±2.0 |
| Duct length (mm) | ±71.8 |
| Stricture length sum (mm) | ±16.0 |
| Dilatation length sum (mm) | ±21.0 |
| Stricture length mean (mm) | ±13.2 |
| Stricture length max (mm) | ±13.5 |
| Stricture absolute severity sum (mm) | ±1.8 |
| Dilatation absolute severity sum (mm) | ±2.2 |
| Dilatation absolute severity max (mm) | ±2.7 |
| Stricture relative severity sum (%) | ±58.1 |
| Dilatation relative severity sum (%) | ±113.1 |
| Dilatation relative severity max (%) | ±71.6 |
| Stricture score sum (mm) | ±6.6 |
| Dilatation score sum (mm) | ±16.3 |
| Dilatation diameter max (mm) | ±5.3 |
| Median duct width (mm) | ±3 (Repeatability), ±3.5 (Reproducibility) |
| Duct width interquartile range (mm) | ±3.5 (Repeatability), ±4.0 (Reproducibility) |
| Maximum duct width (mm) | ±4.5 (Repeatability), ±4.8 (Reproducibility) |
| Minimum duct width (mm) | ±3.8 (Repeatability), ±3.8 (Reproducibility) |
| Intra and Inter-Operator Assessment (Whole Tree Metrics - LoA) | (Same as In Vivo Performance) |
| Intra and Inter-Operator Assessment (Single Duct Metrics - LoA) | |
| Total number of strictures | ±2.0 |
| Total number of dilatations | ±2.0 |
| Duct length (mm) | ±71.8 |
| Stricture length sum (mm) | ±16.0 |
| Dilatation length sum (mm) | ±21.0 |
| Stricture length mean (mm) | ±13.2 |
| Stricture length max (mm) | ±13.5 |
| Stricture absolute severity sum (mm) | ±1.8 |
| Dilatation absolute severity sum (mm) | ±2.2 |
| Dilatation absolute severity max (mm) | ±2.7 |
| Stricture relative severity sum (%) | ±58.1 |
| Dilatation relative severity sum (%) | ±113.1 |
| Dilatation relative severity max (%) | ±71.6 |
| Stricture score sum (mm) | ±6.6 |
| Dilatation score sum (mm) | ±16.3 |
| Dilatation diameter max (mm) | ±5.3 |
| Median duct width (mm) | ±2.0 |
| Duct width interquartile range (mm) | ±2.0 |
| Maximum duct width (mm) | ±3.0 |
| Minimum duct width (mm) | ±2.5 |
| Equivalence Testing to Predicate Device (LoA) | |
| Median duct width (mm) | ±3 |
| Duct width interquartile range (mm) | ±3.5 |
| Maximum duct width (mm) | ±4.5 |
| Minimum duct width (mm) | ±3.8 |
| Biliary tree volume | ±20 mL |
| Gallbladder volume | ±30 mL |
| Percentage of ducts with median width less than 3 mm | ±40% |
| Percentage of ducts with median width greater than 3 mm up to 5 mm | ±40% |
| Percentage of ducts with median width greater than 5 mm up to 9 mm | ±40% |
| Percentage of ducts with median width greater than 9 mm | ±40% |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly mentions the use of "previously acquired data" for substantial equivalence testing with MRCP+ v1 and "acquired phantom data" for accuracy, repeatability, and reproducibility. Additionally, "volunteer data" was used for clinical precision and operator variability assessments.
-
Phantoms:
- Types: Stricture and Dilatation phantom, Tube Width and Clinical Biliary phantoms, 3D printed phantoms, digital synthetic phantoms.
- Data Provenance: The phantoms were "scanned" on various MRI systems (Siemens 1.5T/3T, GE 1.5T/3T, Philips 1.5T/3T). This indicates controlled, prospective data acquisition specifically for testing. Digital synthetic phantoms are likely "in silico" generated.
- Sample Size: Not explicitly stated as a single number for all phantom tests. The document refers to "the metrics introduced in MRCP+ v2" and comparison of "results from scanned 3D printed phantoms" suggesting multiple phantom uses. For substantial equivalence, it mentions "previously acquired data from Tube Width and Clinical Biliary phantoms."
-
Clinical Data (for Precision, Repeatability & Reproducibility, Intra/Inter-Operator Assessment):
- Types: "Volunteers scanned under the same measurement conditions (twice, on the same scanner, a short time apart for repeatability) and under different measurement conditions (on more than one scanner, compared to a reference scanner for reproducibility)." This confirms prospective data acquisition from healthy individuals or similar.
- Data Provenance: Not explicitly stated (e.g., country of origin), but implies a controlled clinical setting for volunteer scanning.
- Sample Size: Not explicitly stated. The text notes "volunteer data used for the precision experiments was also used to assess inter-operator and intra-operator variability."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
-
Phantom Data:
- Ground Truth: For accuracy assessments, the ground truth was the "specification used to generate the phantom (i.e., the specification file used to create the phantom)" and the "ground truth specification" for digital synthetic phantoms. This implies a known, engineered ground truth rather than expert-derived ground truth.
- Number/Qualifications of Experts: Not applicable, as the ground truth was based on phantom specifications, not expert interpretation.
-
Clinical Data:
- The document implies that "Perspectum-trained operator loads previously acquired MRCP data" and "A structured summary report is generated as output by the device, which includes quantitative analysis results and geometric characteristics of the biliary system and pancreatic ducts." These operators were also involved in inter/intra-operator assessment.
- Ground Truth: The document does not explicitly state how ground truth was established for the clinical precision/variability studies or if a separate, expert-derived ground truth was used for assessing the accuracy of metrics against real anatomy. The focus for clinical testing was on device precision (repeatability and reproducibility) and operator variability, implying a comparison of the device's output against itself under different conditions rather than a comparison to an independent "true" clinical assessment of duct characteristics.
4. Adjudication Method (for the test set)
- Phantoms: Not applicable, as ground truth was derived from phantom specifications.
- Clinical Data (Precision/Operator Variability):
- Adjudication: For inter-operator variability, "two internal operators separately processed the same cases and the metrics obtained by each operator were compared statistically." For intra-operator variability, "one internal operator processed the same set of cases twice... and the metrics from each analysis were compared statistically." The operators were "blinded to all analyses conducted by other operators and origins of datasets." This indicates a statistical comparison rather than a consensus-based adjudication of ground truth.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not describe a Multi Reader Multi Case (MRMC) comparative effectiveness study evaluating human reader improvement with AI assistance. The study focuses on the standalone performance characteristics of the MRCP+ v2 device itself, including its accuracy, precision, repeatability, reproducibility, and operator variability, and its substantial equivalence to its predicate device. It assesses the device's ability to generate quantitative measurements, not its impact on human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance assessment was conducted, particularly for the "Accuracy of algorithm - synthetic data" (Table 3). This involved comparing the algorithm's output on digital synthetic phantoms directly against the known ground truth specifications of those phantoms. The device is also described as a "standalone software medical device" and the tests for accuracy, repeatability, and reproducibility of the device are essentially standalone assessments of the product's output.
7. The Type of Ground Truth Used
- Phantoms: The ground truth for phantom studies was the specification used to generate the phantom (e.g., the CAD file or design parameters for 3D printed phantoms, or the configuration files for digital synthetic phantoms). This is a precise, known, and engineered ground truth.
- Clinical Data (Precision/Operator Variability): For the clinical precision studies, the ground truth was implicitly the device's own output from repeated measurements or different operators. The goal was to assess consistency of the device, not necessarily its accuracy against an independent clinical ground truth (like pathology or clinical outcomes). The document does not mention ground truth being established by pathology, outcomes data, or expert consensus for the clinical performance evaluation of the metrics themselves.
8. The Sample Size for the Training Set
The document does not specify the sample size used for the training set of the MRCP+ v2 algorithms. The provided text is a 510(k) summary, which focuses on validation and comparison to a predicate, rather than detailed algorithm development or training data.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. Again, this level of detail is typically not included in a 510(k) summary.
Ask a specific question about this device
(29 days)
CoverScan is a medical image management and processing software package that allows the display, analysis and post-processing of DICOM compliant medical images and MR data.
CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart, lungs, liver, spleen, pancreas and kidney.
CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal relaxation time (T1, srT1, T2), fat content (proton density fat fraction or PDFF) and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change on deep inspiration).
These metrics derived from the images, when interpreted by a licensed physician, yield information that may assist in diagnosis, clinical management and monitoring of patients.
CoverScan is not intended for asymptomatic screening. This device is intended for use with Siemens 1.5T and Siemens 3T MRI scanners.
CoverScan is a post-processing software system comprised of several software modules. It uses acquired MR data to produce metrics of quantified tissue characteristics of the heart, lungs, liver, kidneys, pancreas and spleen.
Metrics produced by CoverScan can be used by healthcare professionals (HCPs) in a clinical setting for the purposes of assessing multiple organs.
CoverScan v1.1 is a new version of CoverScan that supports the use of Siemens 3T scan data to enable more scan sites to utilise CoverScan.
The provided text is a summary of the FDA 510(k) clearance for CoverScan v1.1. It primarily focuses on demonstrating substantial equivalence to a predicate device (CoverScan v1.0) due to an expanded compatibility with Siemens 3T MRI scanners. The document outlines the device's intended use, functions, and the types of data it processes (T1, T2, PDFF, etc.).
However, the document does not contain the detailed study information required to fully answer your request regarding acceptance criteria and performance study specifics. It mentions "performance testing" and "extensive validation testing" but does not provide the specific metrics, sample sizes, ground truth establishment methods, or human reader study details you've asked for.
It states: "The change to add the Siemens 3T Scan data was made and performance testing using well established methods demonstrated equal performance between the subject and the predicate. In summary, CoverScan v1.1 is substantially equivalent to the predicate device CoverScan v1.0 (K212565)." This implies that the performance was compared to the predicate, but the specific performance results and acceptance criteria are not elaborated upon in this summary document.
Therefore, I cannot populate the table or answer most of your detailed questions directly from the provided text.
Here's what can be extracted and what cannot:
What can be extracted:
- Device Name: CoverScan (CoverScan v1.1)
- Predicate Device: CoverScan v1.0 (K212565)
- Purpose of the change (v1.0 to v1.1): To support Siemens 3T scan data.
- General mention of testing: "Performance testing using well established methods demonstrated equal performance between the subject and the predicate." and "Internal verification and validation testing confirms that the product specifications are met."
- Ground Truth Type (Inferred, not explicitly stated as "ground truth"): Device performance was assessed with "purpose-built phantoms and in-vivo acquired data from volunteers covering a range of physiological values for sT1, cT1, T1 and PDFF." This suggests comparison against known phantom properties and physiological values, but the direct "ground truth" establishment method for patient data is not detailed.
- Standards referenced: IEC 62304, IEC 62366, DICOM 3.0, ISO 14971, ISO 13485.
What cannot be extracted (due to lack of information in the document):
- Table of Acceptance Criteria and Reported Device Performance: No specific quantitative acceptance criteria or reported performance values are provided.
- Sample sizes: No sample size is given for the test set or training set.
- Data provenance: Not specified (e.g., country of origin, retrospective/prospective).
- Number of experts and qualifications for ground truth: Not mentioned.
- Adjudication method: Not mentioned.
- MRMC study: Not mentioned.
- Standalone (algorithm only) performance: Not explicitly detailed with metrics.
- How ground truth for the training set was established: Not mentioned.
In summary, the provided document is an FDA clearance letter focusing on substantial equivalence based on implied equivalent performance rather than a detailed performance study report with specific metrics and methodologies.
Ask a specific question about this device
(263 days)
LiverMultiScan v5 (LMSv5) is indicated for use as a magnetic resonance diagnostic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.
LMSv5 is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.
LMSv5 provides several tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessment of selected regions includes the determination of triglyceride fat fraction in the liver (PDFF), T2*, LIC (Liver Iron Concentration) and iron corrected T1 (cT1) measurements.
These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
LiverMultiScan v5 (LMSv5) is a standalone software device. The LiverMultiScan v5 device is to assist a trained operator with the evaluation of information from Magnetic Resonance (MR) images from a single time-point (a patient visit).
LiverMultiScan is a post-processing software device, a trained operator uses tools within the device interface to quantify liver tissue characteristics from parametric maps. LiverMultiScan v5 includes automatic processing functionality based on machine-learning to assist in the quantification of metrics during analysis, such as automatic artefact detection and automatic segmentation of the liver. A summary report from the analysis conducted is generated for interpretation by a clinician.
LiverMultiScan v5 is not intended to replace the established procedures for the assessment of a patient's liver health by a clinician, providing many opportunities for competent human intervention in the clinical care of patients.
The metrics are intended to be used as an additional diagnostic input to provide information to clinicians as part of a wider diagnostic process. It is expected that in the normal course of clinical care, patients with clinical symptoms or risk factors which may indicate liver disease. The interpreting clinician needs to take into consideration the device's limitations and accuracy during clinical interpretation.
Information gathered through existing diagnostic tests and clinical evaluation of the patient, as well as information obtained from LiverMultiScan v5 metrics, may contribute to a diagnostic decision.
LiverMultiScan v5 is not a computer-aided diagnostic device and can only present imaging information which must be interpreted by a qualified clinician. LiverMultiScan v5 is an aid to diagnosis and treatment decisions remain the responsibility of the clinician.
Here's an analysis of the acceptance criteria and study information for LiverMultiScan v5 (LMSv5) based on the provided text, structured according to your requested points:
1. Table of Acceptance Criteria and Reported Device Performance
The provided FDA 510(k) summary document does not explicitly list quantitative acceptance criteria in a dedicated table format. Instead, it states that "All product specifications were verified and the overall ability of the product to meet user needs was validated." and "The accuracy and precision of device measurements was assessed using purpose-built phantoms...". It also mentions "The performance testing conducted demonstrates that LiverMultiScan v5 is at least as safe and effective as the predicate devices."
While specific numerical acceptance criteria (e.g., "PDFF accuracy must be within X%", "LIC precision must be within Y%") are not detailed in this public document, the general acceptance can be inferred from the statement that the device meets safety and effectiveness standards comparable to its predicates. The performance claims are therefore that the device provides accurate and precise measurements of liver triglyceride fat fraction (PDFF), T2*, Liver Iron Concentration (LIC), and iron-corrected T1 (cT1), comparable to or exceeding its predicate devices.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| Accuracy and precision of PDFF measurements | Assessed using purpose-built phantoms, revealing accuracy and precision corresponding to physiological ranges. In-vivo volunteer data also used to assess precision across supported scanners. |
| Accuracy and precision of T2* measurements | Assessed using purpose-built phantoms, revealing accuracy and precision corresponding to physiological ranges. In-vivo volunteer data also used to assess precision across supported scanners. |
| Accuracy and precision of LIC measurements | Assessed using purpose-built phantoms, revealing accuracy and precision corresponding to physiological ranges. In-vivo volunteer data also used to assess precision across supported scanners. (Note: LMSv5 introduces LIC quantification, which was not available in the primary predicate LMSv4, but was available in the secondary predicate FerriScan. The performance is assessed to be at least as safe and effective as the predicates.) |
| Accuracy and precision of cT1 measurements | Assessed using purpose-built phantoms, revealing accuracy and precision corresponding to physiological ranges. In-vivo volunteer data also used to assess precision across supported scanners. |
| Inter and Intra-operator variability | Assessed. (Specific metrics not provided in this document). |
| Functionality (installation, licensing, labeling, features) | Met design requirements. |
| Overall safety and effectiveness | Demonstrated to be at least as safe and effective as the predicate devices. Differences from predicates do not result in a new intended use or raise new questions of safety and effectiveness. |
| Artifact detection and delineation (new feature in LMSv5) | Automatically detects artifacts and delineates them on parametric maps, recommending slices for quantitative output. (Functionality and impact on quantification/interpretation assessed). |
| Compatibility with DICOM 3.0 compliant MR image datasets | Designed to utilize them. |
| Independence of reported metrics from MRI equipment vendor | Datasets imported into LiverMultiScan (LMSv5) are DICOM 3.0 compliant, reported metrics are independent of the MRI equipment vendor. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: The document mentions "in-vivo volunteer data was used" for assessing precision across supported scanners and for inter/intra-operator variability. It also refers to "purpose-built phantoms" for accuracy and precision of device measurements. Specific numerical sample sizes for either phantom or in-vivo data are not provided in this summary document.
- Data Provenance: Not explicitly stated from which country the "in-vivo volunteer data" originated. It is generally understood that studies supporting medical device submissions to the FDA can involve multinational data, but this document does not specify. The data would be prospective if volunteers were recruited and scanned specifically for this validation, or retrospective if existing de-identified patient data was used. The document does not clarify this.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This document does not provide details on the number or qualifications of experts used to establish ground truth for the test set. For quantitative MRI biomarkers like PDFF, T2*, cT1, and LIC, ground truth is typically established through direct physical measurements (e.g., chemical extraction for fat, histology/biopsy for iron/fibrosis, though biopsy has limitations) or well-validated reference standards (phantoms or highly accurate established MRI methods). The document mentions "purpose-built phantoms containing vials with different relaxation times corresponding to the physiological ranges of tissue values" were used for accuracy and precision, implying these phantoms served as a form of ground truth for those measurements.
4. Adjudication method for the test set
The document does not describe any expert adjudication methods (e.g., 2+1, 3+1) for the test set within this summary. For this type of quantitative imaging device, the focus is more on technical accuracy and precision against established physical or biological references rather than subjective expert consensus on image interpretation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not report an MRMC comparative effectiveness study where human readers' performance with and without AI assistance (LMSv5) was evaluated. The device is described as "post-processing software" that provides "additional diagnostic input" and is "an aid to diagnosis." It includes "automatic processing functionality based on machine-learning to assist in the quantification of metrics... such as automatic artefact detection and automatic segmentation of the liver." However, the study described focuses on the device's standalone performance (accuracy, precision, variability) and not on its impact on human reader performance in a controlled comparative setting.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance assessment was done. The testing described focuses on the device's intrinsic capabilities:
- "The accuracy and precision of device measurements was assessed using purpose-built phantoms..."
- "To assess the precision of LiverMultiScan v5 measurements across supported scanners, in-vivo volunteer data was used."
- "Inter and intra operator variability was also assessed." (While this involves human operators, it assesses the device's consistency under different human interactions, not the human's diagnostic performance).
- The device is a "standalone software device" and "All operations are directly controlled by the LiverMultiScan device."
This indicates that the technical performance of the algorithm itself, in generating the quantitative metrics (PDFF, T2*, LIC, cT1), was evaluated.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The primary ground truth described in this summary are:
- Physical Phantoms: "purpose-built phantoms containing vials with different relaxation times corresponding to the physiological ranges of tissue values expected to be seen in-vivo" were used to assess the accuracy and precision of measurements. This provides a quantifiable, objective ground truth for the relaxation times and component fractions (like fat).
- For LIC, the device "uses the measured T2* value and uses them to characterise iron loading in the liver which is then transformed by a defined calibration curve to provide a quantitative measure of liver iron concentration in vivo." This implies the ground truth for establishing that calibration curve would likely have been against biopsy-derived liver iron concentration data or other highly validated reference methods in earlier studies (though not explicitly stated for this particular submission's testing itself, it's the basis for LIC quantification).
8. The sample size for the training set
The document does not specify the sample size used for the training set. It mentions that LMSv5 includes "automatic processing functionality based on machine-learning to assist in the quantification of metrics during analysis, such as automatic artefact detection and automatic segmentation of the liver." This clearly indicates machine learning was used, and thus a training set was necessary. However, the size and characteristics of that training set are not included in this 510(k) summary.
9. How the ground truth for the training set was established
The document does not detail how the ground truth for the training set (used for machine learning components like automated segmentation and artifact detection) was established. Typically, for such features, ground truth would be established through:
- Expert manual segmentation/annotation: Radiologists or trained medical image analysts would manually delineate liver boundaries and artifacts on a large dataset of MR images.
- Consensus from multiple experts: To reduce individual bias, several experts might review and agree upon segmentations or artifact labeling.
However, these specifics are omitted from this summary.
Ask a specific question about this device
(279 days)
CoverScan is a medical image management and processing software package that allows the display, analysis and postprocessing of DICOM compliant medical images and MR data.
CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart, lungs, liver, spleen, pancreas and kidney.
CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal relaxation time and rate (T1, SR-T1, cT1, T2), fat content (proton density fat fraction or PDFF) and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change on deep inspiration).
These metrics derived from the interpreted by a licensed physician, vield information that may assist in diagnosis, clinical management and monitoring of patients.
CoverScan is not intended for asymptomatic screening. This device is intended for use with Siemens 1.5T MRI scanners.
CoverScan is a post-processing software system comprised of several software modules. It uses acquired MR data to produce metrics of quantified tissue characteristics of the heart, lungs, liver, kidneys, pancreas and spleen.
Metrics produced by CoverScan can be used by licensed physicians in a clinical setting for the purposes of assessing multiple organs.
The provided documentation describes the CoverScan v1 device, which is a medical image management and processing software. While it mentions internal verification and validation testing, and that product specifications were met, it does not explicitly state specific quantitative acceptance criteria or detailed results of a study proving the device meets those criteria, especially in a comparative effectiveness context (MRMC).
The document primarily focuses on establishing substantial equivalence to predicate devices through a qualitative comparison of intended use, technological characteristics, and performance features. It indicates that "bench testing included functional verification to ensure software installation, licensing, labeling, and feature functionality all met design requirements" and that "The accuracy and precision of device measurements was assessed using purpose-built phantoms and in-vivo acquired data from volunteers." However, it does not provide the specific quantitative results of these assessments against defined acceptance criteria.
Therefore, much of the requested information cannot be directly extracted from the provided text. I will explain what information is available and highlight what is missing.
Here's an attempt to structure the information based on the provided text, with clear indications where the information is not present:
Acceptance Criteria and Device Performance Study (CoverScan v1)
The provided document indicates that CoverScan v1 underwent internal verification and validation testing to confirm it met product specifications and its overall ability to meet user needs was validated. However, specific, quantitative acceptance criteria for metrics like accuracy, sensitivity, specificity, etc., are not explicitly defined in the provided text. Similarly, the reported numerical device performance against such criteria is not detailed. The document broadly states that "The accuracy and precision of device measurements was assessed using purpose-built phantoms and in-vivo acquired data from volunteers."
Missing Information: A detailed table of acceptance criteria with numerical targets and the corresponding reported device performance values.
1. A table of acceptance criteria and the reported device performance
| Metric / Category | Acceptance Criteria (Quantitative) | Reported Device Performance (Quantitative) | Source/Test Type |
|---|---|---|---|
| Accuracy of measurements (cT1, T1, PDFF, T2) | Not explicitly defined in the document | "Assessed using purpose-built phantoms and in-vivo acquired data from volunteers covering a range of physiological values for cT1, T1 and PDFF." "Inter and intra operator variability was also assessed." | Bench testing, Phantom studies, In-vivo volunteer data |
| Precision of measurements | Not explicitly defined in the document | "Assessed using purpose-built phantoms... To assess the precision of CoverScan v1 measurements across supported scanners, in-vivo volunteer data was used." | Bench testing, Phantom studies, In-vivo volunteer data |
| Functional Verification | "Software installation, licensing, labeling, and feature functionality all met design requirements." | "Bench testing included functional verification to ensure software installation, licensing, labeling, and feature functionality all met design requirements." | Bench testing |
| Stress Testing | "System as a whole provides all the capabilities necessary to operate according to its intended use." | "All of the different components of the CoverScan software have been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use." | Stress testing |
2. Sample sized used for the test set and the data provenance
- Test Set Sample Size: The document mentions "in-vivo acquired data from volunteers" and that "Volunteers participating in the performance testing were representative of the intended patient population." However, the specific number of cases or volunteers used in the test set is not provided.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It only mentions "in-vivo acquired data from volunteers," implying prospectively collected data for assessment, but this is not explicitly stated.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The text indicates that the device produces quantitative metrics that are "interpreted by a licensed physician" to "assist in diagnosis, clinical management and monitoring of patients." However, it does not describe how ground truth was established for the performance testing, nor the number or qualifications of experts involved in that process.
4. Adjudication method for the test set
This information is not provided in the document. There is no mention of an adjudication process (e.g., 2+1, 3+1, none) for the test set's ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not indicate that a MRMC comparative effectiveness study was performed. The device is described as a "post-processing software system" that provides "quantified metrics" and does not describe AI assistance for human readers in a diagnostic workflow. The primary method of performance assessment mentioned is the accuracy and precision of the measurements themselves using phantoms and volunteer data, not reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The performance testing described involves the device generating quantitative metrics. The phrase "The accuracy and precision of device measurements was assessed" suggests a standalone performance assessment of the algorithm's output (measurements) against some reference (phantoms, in-vivo data). While the final interpretation is by a physician, the core performance reported relates to the device's ability to produce these measurements consistently and accurately, which aligns with a standalone assessment of the algorithms.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the "accuracy and precision of device measurements" was established using:
- Purpose-built phantoms: These phantoms "containing vials with different relaxation times corresponding to the physiological ranges of tissue values." This provides a known, controlled physical ground truth.
- In-vivo acquired data from volunteers: For this type of data, the text does not specify how ground truth was established (e.g., through other validated methods, clinical outcomes, or expert consensus on a final diagnosis based on all available information). It only mentions that the studies "covering a range of physiological values for cT1, T1 and PDFF."
8. The sample size for the training set
This information is not provided in the document. The document describes CoverScan v1 as software that takes acquired MR data and processes it; it does not detail any machine learning training processes or associated datasets.
9. How the ground truth for the training set was established
As there is no mention of a specific training set in the provided text, the method for establishing its ground truth is also not described. The document implies that the device is a measurement and processing tool rather than a machine learning model that requires a dedicated training set.
Ask a specific question about this device
(67 days)
Hepatica (Hepatica v1) is a post-processing medical device software that presents quantified metrics which may contribute to the assessment of a patient's liver health.
Hepatica (Hepatica v1) uses image visualisation and analysis tools to process DICOM 3.0 compliant magnetic resonance image datasets to produce semi-automatic segmented 3D models of the work of Couinaud and the Brisbane 2000 terminology. For each identified Couinaud segment, volumetric data is determined and reported.
Hepatica v1) may also report iron corrected-T1 (cT1) and PDFF calculated using the IDEAL method from multi-slice acquisitions, on a per segment basis, over the whole liver. Both metrical values of different fundamental liver tissue characteristics that can be used as measures of liver tissue health.
Hepatica (Hepatica v1) provides trained clinicians with additional information to evaluate the volume and health of a patient's liver on a segmental basis. It is not intended to replace the established procedures for the assessment of a patient's liver health. However, information gathered through existing diagnostic tests, clinical evaluation of the patient, as well Hepatica (Hepatica v1), may support surgical decision making.
Hepatica v1 is a standalone software device that imports MR datasets encompassing the abdomen, including the liver. Visualisation and display of T1-weighted MR data which can be analysed, and quantitative metrics of tissue characteristics and liver volume are then reported. Datasets imported into Hepatica are DICOM 3.0 compliant and reported metrics are independent of the MRI equipment vendor. It allows for the 3D visualisation of the liver and quantification of metrics (cT1, PDFF and volumetry) from liver tissue and exportation of results and images to a deliverable report. Hepatica v1 supports semi-automatic liver segmentation of T1-weighted volumetric data. Liver segmentation in Hepatica v1 requires the placement of anatomical landmarks to define the outer contours of the liver and can be adjusted by the operator, where necessary. Where available, whole liver and segmental cT1 and PDFF quantitative metrics derived from the predicate device may be presented in the final report. Hepatica uses volumetric datasets to create 2D anatomical views from all supported scanners. Where available, cT1 and PDFF parametric maps are derived from the predicate device. Quantified metrics and images derived from the analysis of liver volume and tissue characteristics are collated into a report for evaluation and interpretation by a clinician.
Here's a breakdown of the acceptance criteria and study information for the Hepatica (Hepatica v1) device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a quantitative format (e.g., "accuracy must be >X%"). Instead, it states that the performance testing "demonstrates that Hepatica v1 is at least as safe and effective as the predicate device and does not introduce any new risks."
However, it provides "Upper and Lower Limits of Agreement" from a Bland-Altman analysis, which can be interpreted as the range within which the device's measurements agree with the gold standard. A tighter range indicates higher accuracy. The precision results are also presented as "Upper and Lower limits of Agreement" for repeatability and reproducibility.
The table below summarizes the reported device performance, which implies that these values met the internal acceptance parameters for demonstrating substantial equivalence:
| Metric | Type of Measurement | Reported Device Performance (Upper and Lower Limits of Agreement) |
|---|---|---|
| Volume (% of total liver volume) | ||
| Segment 1 | Accuracy | -0.49% to 0.95% |
| Segment 2 | Accuracy | -3.09% to 5.06% |
| Segment 3 | Accuracy | -5.01% to 3.9% |
| Segment 4a | Accuracy | -4.60% to 4.26% |
| Segment 4b | Accuracy | -5.50% to 2.56% |
| Segment 5 | Accuracy | -1.54% to 3.38% |
| Segment 6 | Accuracy | -4.34% to 4.29% |
| Segment 7 | Accuracy | -3.30% to 1.79% |
| Segment 8 | Accuracy | -3.86% to 5.54% |
| Whole liver | Accuracy | -4.16% to 0.54% |
| Segment 1 | Repeatability | -0.72% to 0.65% |
| Segment 2 | Repeatability | -3.06% to 3.24% |
| Segment 3 | Repeatability | -2.67% to 3.13% |
| Segment 4a | Repeatability | -2.48% to 2.43% |
| Segment 4b | Repeatability | -1.82% to 1.96% |
| Segment 5 | Repeatability | -4.45% to 4.45% |
| Segment 6 | Repeatability | -3.60% to 4.10% |
| Segment 7 | Repeatability | -3.32% to 3.33% |
| Segment 8 | Repeatability | -4.99% to 3.81% |
| Whole liver | Repeatability | -6.15% to 3.78% |
| Segment 1 | Reproducibility | -1.39% to 0.90% |
| Segment 2 | Reproducibility | -3.10% to 3.15% |
| Segment 3 | Reproducibility | -2.41% to 2.06% |
| Segment 4a | Reproducibility | -2.54% to 2.58% |
| Segment 4b | Reproducibility | -1.70% to 1.74% |
| Segment 5 | Reproducibility | -4.97% to 5.94% |
| Segment 6 | Reproducibility | -3.69% to 5.40% |
| Segment 7 | Reproducibility | -4.39% to 3.59% |
| Segment 8 | Reproducibility | -6.23% to 5.04% |
| Whole liver | Reproducibility | -16.6% to 6.95% |
| cT1 | ||
| Segment 1 | Accuracy | -1.13% to 0.61% |
| Segment 2 | Accuracy | -2.38% to 1.56% |
| Segment 3 | Accuracy | -1.51% to 1.31% |
| Segment 4a | Accuracy | -0.77% to 1.10% |
| Segment 4b | Accuracy | -1.32% to 1.13% |
| Segment 5 | Accuracy | -1.11% to 0.87% |
| Segment 6 | Accuracy | -1.00% to 0.83% |
| Segment 7 | Accuracy | -0.88% to 0.64% |
| Segment 8 | Accuracy | -0.91% to 1.09% |
| Whole liver | Accuracy | 0.00% to 0.00% (This suggests perfect agreement or rounding issues) |
| PDFF | ||
| Segment 1 | Accuracy | -0.26% to 0.21% |
| Segment 2 | Accuracy | -0.33% to 0.38% |
| Segment 3 | Accuracy | -0.16% to 0.17% |
| Segment 4a | Accuracy | -0.30% to 0.23% |
| Segment 4b | Accuracy | -0.16% to 0.14% |
| Segment 5 | Accuracy | -0.16% to 0.18% |
| Segment 6 | Accuracy | -0.16% to 0.26% |
| Segment 7 | Accuracy | -0.12% to 0.18% |
| Segment 8 | Accuracy | -0.24% to 0.32% |
| Whole liver | Accuracy | -0.02% to 0.02% |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document states that the performance testing used "previously acquired in-vivo data from healthy and non-healthy volunteers." It also mentions "Volunteers participating in the performance testing were representative of the intended patient population." However, a specific number for the sample size (N) of these volunteers or images in the test set is not provided.
- Data Provenance: "Previously acquired in-vivo data." The country of origin is not explicitly stated. It can be inferred that the data is likely from the UK, given the submitter's address (Oxford, UK). The data is retrospective, as it was "previously acquired."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: The document refers to the gold standard as "radiologists." It does not specify the number of radiologists involved in establishing the ground truth.
- Qualifications of Experts: The qualification is "radiologists." No further details on their years of experience or sub-specialty are provided.
4. Adjudication Method for the Test Set
The document states that the "gold standard" is "radiologists." It does not describe any specific adjudication method (e.g., 2+1, 3+1 consensus) used to establish this ground truth. It implies that the radiologists' readings were considered the definitive truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compared the device's measurements to a "gold standard" (radiologists' assessments), not the performance of human readers with vs. without AI assistance. Therefore, there is no effect size reported for human readers improving with AI assistance.
6. If a Standalone Study Was Done
Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The performance testing section directly reports on the "Accuracy" and "Precision" of Hepatica v1's measurements (cT1, PDFF, and volumetry) when compared to the gold standard. The device operators are "trained Perspectum internal operators," but the reported metrics are explicitly from the device's output. The statement "The variation introduced by operator measurements are well within the acceptance criteria" also suggests an understanding of the device's standalone performance separate from human interpretation of the reports.
7. The Type of Ground Truth Used
The primary ground truth used for accuracy comparison is expert consensus/interpretation, specifically "radiologists." For cT1 and PDFF, these are quantitative measurements derived from imaging, which radiologists would interpret or measure. For volumetry, it's also based on radiological assessment.
8. The Sample Size for the Training Set
The document does not provide the sample size for the training set. It only mentions the "previously acquired in-vivo data from healthy and non-healthy volunteers" used for performance testing.
9. How the Ground Truth for the Training Set Was Established
Since the training set sample size and its specifics are not mentioned, how its ground truth was established is not described in the provided document.
Ask a specific question about this device
(60 days)
LiverMultiScan (LMSv4) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.
LiverMultiScan (LMSv4) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.
LiverMultiScan (LMSv4) provides a number of tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessments of selected regions include the determination of triglyceride fat fraction in the liver (PDFF), T2* and iron-corrected T1 (cT1) measurements. T2* may be optionally computed using the DIXON or LMS MOST methods.
These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
LiverMultiScan is a standalone software device. The purpose of the LiverMultiScan device is to assist a trained operator with the evaluation of information from Magnetic Resonance (MR) images from a single time-point (a patient visit). LiverMultiScan is a post-processing software device, a trained operator uses tools such as automatic liver segmentation and region of interest placement upon previously acquired MR images, from which a summary report is generated. The summary report is subsequently sent to an interpreting clinician at the acquiring site.
LiverMultiScan is not intended to replace the established procedures for the assessment of a patient's liver health by an interpreting clinician, providing many opportunities for competent human intervention in the interpretation of images and information displayed.
The metrics are intended to be used as an additional diagnostic input to provide information to clinicians as part of a wider diagnostic process. It is expected that in the normal course of liver disease diagnosis, patients with clinical symptoms or risk factors which may indicate liver disease. The interpreting clinician needs to take into consideration the device's limitations and accuracy during clinical interpretation.
Liver function tests, blood tests, ultrasound scanning as well as liver biopsy are all expected to be used at the discretion of a qualified clinician in addition to information obtained from the use of LiverMultiScan metrics. The purpose of LiverMultiScan metrics is to provide imaging information to assist in characterizing tissue in the liver, in addition to existing methods for obtaining information relating to the liver. LiverMultiScan metrics are not intended to replace any existing diagnostic source of information but can be used to identify patients who may benefit most from further evaluation, including biopsy.
Information gathered through existing diagnostic tests and clinical evaluation of the patient, as well as information obtained from LiverMultiScan metrics, may contribute to a diagnostic decision.
LiverMultiScan is not a computer-aided diagnostic device and can only present imaging information which must be interpreted by a qualified clinician. LiverMultiScan is an aid to diagnosis and treatment decisions remains the responsibility of the clinician.
In consequence, the product is considered to have no adverse effect on health since the results represent only a part of the information that the user will utilize for final interpretation. In this regard, LiverMultiScan presents a moderate level of concern with respect to patient safety.
Here's a detailed breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the performance test results demonstrating acceptable levels of accuracy, repeatability, and reproducibility. The document states that the observed variations were "well within the prescribed acceptance criteria," but the numerical values for the criteria themselves are not explicitly listed. Instead, the reported performance values are provided.
| Metric (Type) | Performance Aspect | Acceptance Criteria (Implicit) | Reported Device Performance (LMSv4) |
|---|---|---|---|
| cT1 (Bench) | Accuracy (1.5T) | Within acceptable range (not explicitly stated, but negative bias in line with literature is acceptable) | -189.5 to -35.11 ms (Consistent with literature-reported underestimation of ground truth T1 using MOLLI techniques) |
| cT1 (Bench) | Accuracy (3T) | Within acceptable range (not explicitly stated, but negative bias in line with literature is acceptable) | -187.0 to -19.12 ms (Consistent with literature-reported underestimation of ground truth T1 using MOLLI techniques) |
| T2* (Bench) | Accuracy (1.5T) | Accurate over expected physiological range of values (not explicitly stated numerically) | -0.68 to 0.64 ms (Accurate over the expected physiological range of values) |
| T2* (Bench) | Accuracy (3T) | Accurate over expected physiological range of values (not explicitly stated numerically) | -0.30 to 0.39 ms (Accurate over the expected physiological range of values) |
| IDEAL PDFF (Bench) | Accuracy (1.5T) | Accurate over expected physiological range of values (not explicitly stated numerically) | -3.80 to 6.08% (Accurate over the expected physiological range of values) |
| IDEAL PDFF (Bench) | Accuracy (3T) | Accurate over expected physiological range of values (not explicitly stated numerically) | -1.39 to 5.58% (Accurate over the expected physiological range of values) |
| cT1 (ROI) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -43.25 to 26.77 ms (Highly repeatable) |
| cT1 (Segmentation) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -40.75 to 25.02 ms (Highly repeatable) |
| T2* (DIXON) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -5.21 to 6.01 ms (Highly repeatable) |
| T2* (MOST) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -3.17 to 3.25 ms (Highly repeatable) |
| IDEAL PDFF (ROI) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -1.48 to 1.42% (Highly repeatable) |
| IDEAL PDFF (Segmentation) (Clinical) | Repeatability | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -1.31 to 1.34% (Highly repeatable) |
| cT1 (ROI) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -103 to 91.8 ms (Reproducible between scanners and field strengths) |
| cT1 (Segmentation) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -102.3 to 93.69 ms (Reproducible between scanners and field strengths) |
| T2* (DIXON) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -1.74 to 0.35 ms (Reproducible between scanners and field strengths) |
| T2* (MOST) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -2.40 to 2.15 ms (Reproducible between scanners and field strengths) |
| IDEAL PDFF (ROI) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -2.88 to 2.53% (Reproducible between scanners and field strengths) |
| IDEAL PDFF (Segmentation) (Clinical) | Reproducibility | Well within prescribed acceptance criteria (Limits of Agreement - not numerically specified) | -2.94 to 2.53% (Reproducible between scanners and field strengths) |
| Operator Variability | Repeatability/Reproducibility | Well within prescribed acceptance criteria (not numerically specified) | Variation introduced by operator measurements (both ROI and segmentation) is well within the prescribed acceptance criteria. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not explicitly state the numerical sample size for the test set. For the bench testing, it mentions "purpose-built phantoms" and for clinical testing, "in-vivo acquired volunteer data." The number of phantoms or volunteers is not specified.
- Data Provenance:
- Bench Testing: Data was acquired from "purpose-built phantoms." The country of origin is not specified, but the applicant (Perspectum Ltd) is based in the UK.
- Clinical Testing: "in-vivo acquired volunteer data" from "volunteers participating in the performance testing were representative of the intended patient population." The country of origin for these volunteers is not specified. The study appears to be prospective for the gathered "volunteer data" in the sense that data was acquired for the purpose of testing, but the details are sparse.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not mention the use of experts to establish ground truth for the test set.
- Bench Testing: Ground truth for phantom accuracy was based on the "gold standard" to which the LMSv4 values were compared. This "gold standard" is implied to be the known physical properties or reference measurements of the phantoms.
- Clinical Testing: The clinical performance relates to repeatability and reproducibility of the device's measurements themselves, not against a human-established ground truth. Operator variability was assessed, implying trained operators were involved, but their qualifications and number are not detailed beyond "trained internal Perspectum operators."
4. Adjudication Method for the Test Set
No adjudication method for the test set is mentioned. The performance testing focuses on the device's intrinsic accuracy, repeatability, and reproducibility, rather than agreement with human interpretation that would typically require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. The document states: "No clinical investigations or studies were conducted during performance testing of LMSv4." The device is positioned as an "aid to diagnosis" and "not intended to replace any existing diagnostic source of information," suggesting its role is as a tool rather than a standalone diagnostic.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, a standalone performance study was clearly done. The "Performance Testing - Bench" section directly assesses the "accuracy and precision of device measurements" using phantoms. The "Performance Testing - Clinical" section assesses the "precision of LMSv4 measurements" (repeatability and reproducibility) using volunteer data. While operator variability was assessed, the core measures are of the algorithm's output. The device itself is described as a "post-processing software device" where an "operator uses tools such as automatic liver segmentation and region of interest placement," implying the algorithm performs the quantification and the operator interfaces with it.
7. Type of Ground Truth Used
- Bench Testing: The ground truth for accuracy was established using a "gold standard" derived from the "purpose-built phantoms" that contained vials with known relaxation times.
- Clinical Testing: For repeatability and reproducibility, the "ground truth" is essentially the device's own consistent measurement across repeated scans, different scanners/field strengths, and different operators. It is not externally validated against pathology or clinical outcomes in this document.
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size for the training set. It focuses solely on the performance testing of the device.
9. How the Ground Truth for the Training Set Was Established
The document does not provide any information regarding the training set or how its ground truth was established, as it does not describe the development or training of the LMSv4 algorithm. The emphasis is on proving substantial equivalence to a predicate device through performance benchmarking.
Ask a specific question about this device
(175 days)
LiverMultiScan (LMSv3) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.
LiverMultiScan (LMSv3) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.
LiverMultiScan (LMSv3) provides a number of tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessment of selected regions include the determination of triglyceride fat fraction in the liver (PDFF), T2* and iron-corrected T1 (cT1) measurements. PDFF may optionally be computed using the LMS IDEAL or three-point Dixon methodology.
These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
LiverMultiScan (LMSv3) is a standalone software application for displaying 2D Magnetic Resonance (MR) medical image data acquired from compatible MR Scanners. LiverMultiScan runs on general-purpose workstations with a colour monitor, keyboard and mouse.
The main functionality of LiverMultiScan (LMSv3) includes:
- Reading DICOM 3.0 compliant datasets stored on workstations, and display of the data acquisition information
- Post-processing of MRI data to generate parametric maps of Proton Density Fat Fraction PDFF), T2*, T1 and ironcorrected T1 (cT1) of the liver.
- Quantification, and calculation of PDFF, T2* and cT1 metrics using tools such as automatic liver segmentation and ROI (region of interest) placement.
- Generation of a summary report demonstrating the quantitative assessment results of fat fraction in the liver (PDFF), T2* and iron-corrected T1 (cT1).
Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with corresponding reported performance for specific metrics like sensitivity, specificity, accuracy, etc., as one might expect for a diagnostic device. Instead, the performance testing focuses on the accuracy, repeatability, reproducibility, and inter-/intra-operator variability of quantitative measurements, and particularly on demonstrating substantial equivalence to a predicate device (LMSv2).
The acceptance criteria are implicitly defined by the reported performance ranges and the conclusion that the device performs "as well as its predicate" and that "all the testing results are well within the acceptance criteria."
| Metric / Test Type | Acceptance Criteria (Implicit from Study Conclusion) | Reported Device Performance (Summary) |
|---|---|---|
| Phantom Testing | ||
| T1 Accuracy | T1 measurements consistent with literature-reported underestimation for MOLLI techniques. | Up to 18.89% lower to ground truth (95% CI Limits of Agreement) |
| T2* Accuracy | Accurate over expected physiological range. | -9.31% to 7.53% of ground truth (95% CI Limits of Agreement) |
| DIXON PDFF Accuracy (<30%) | Relatively accurate over expected physiological range; minor deviations due to known fat bias. | -7.37% to 1.72% (95% CI Limits of Agreement) |
| DIXON PDFF Accuracy (>30%) | Relatively accurate over expected physiological range; minor deviations due to known fat bias. | -28.93% to 6.83% (95% CI Limits of Agreement) |
| IDEAL PDFF Accuracy (<30%) | Accurate over expected physiological range. | -1.17% to 1.43% (95% CI Limits of Agreement) |
| IDEAL PDFF Accuracy (>30%) | Accurate over expected physiological range. | -5.05% to 10.70% (95% CI Limits of Agreement) |
| T1 Repeatability (same scanner) | Highly repeatable. | -13.88 to 14.47 ms (95% CI Limits of Agreement) |
| T2* Repeatability (same scanner) | Highly repeatable. | -0.89 to 1.43 ms (95% CI Limits of Agreement) |
| DIXON PDFF Repeatability (<30%) | Highly repeatable. | -0.66 to 0.82 % (95% CI Limits of Agreement) |
| DIXON PDFF Repeatability (>30%) | Highly repeatable. | -2.11 to 1.96% (95% CI Limits of Agreement) |
| IDEAL PDFF Repeatability (<30%) | Highly repeatable. | -1.27 to 0.87% (95% CI Limits of Agreement) |
| IDEAL PDFF Repeatability (>30%) | Highly repeatable. | -3.80 to 1.93 % (95% CI Limits of Agreement) |
| T1 Reproducibility (different scanners) | Reproducible between different scanners. | -2.66 to 10.78% (95% CI Limits of Agreement) |
| T2* Reproducibility (different scanners) | Reproducible between different scanners. | -3.43 to 2.42 ms (95% CI Limits of Agreement) |
| DIXON PDFF Reproducibility (<30%) | Reproducible between different scanners. | -1.86 to 5.95% (95% CI Limits of Agreement) |
| DIXON PDFF Reproducibility (>30%) | Reproducible between different scanners. | -8.64 to 23.52% (95% CI Limits of Agreement) |
| IDEAL PDFF Reproducibility (<30%) | Reproducible between different scanners. | -1.99 to 2.80% (95% CI Limits of Agreement) |
| IDEAL PDFF Reproducibility (>30%) | Reproducible between different scanners. | -13.46 to 6.98% (95% CI Limits of Agreement) |
| In-Vivo Testing | ||
| cT1 Repeatability | Highly repeatable. | -94.38 to 63.38 ms (ROI); -76.93 to 59.39 ms (Segmentation) |
| T2* Repeatability | Highly repeatable. | -6.07 to 5.70 ms (ROI) |
| DIXON PDFF Repeatability | Highly repeatable. | -1.77 to 3.64 % (ROI); -1.20 to 1.06% (Segmentation) |
| IDEAL PDFF Repeatability | Highly repeatable. | -1.92 to 1.54% (ROI); -1.83 to 1.28 % (Segmentation) |
| cT1 Reproducibility (between scanners) | Reproducible between scanners. | -89.70 to 120.58 ms (ROI); -84.91 to 121.79 ms (Segmentation) |
| T2* Reproducibility (between scanners) | Reproducible between scanners. | -3.68 to 6.35 ms (ROI) |
| DIXON PDFF Reproducibility (between scanners) | Reproducible between scanners. | -6.21 to 2.63% (ROI); -3.14 to 0.88% (Segmentation) |
| IDEAL PDFF Reproducibility (between scanners) | Reproducible between scanners. | -2.66 to 2.77% (ROI); -1.74 to 1.21% (Segmentation) |
| cT1 Intra-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -27.38 to 28.33ms (ROI); -20.81 to 13.06ms (Segmentation) |
| T2* Intra-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -2.29 to 2.91 ms (ROI) |
| DIXON PDFF Intra-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -0.78 to 1.90 % (ROI); -0.29 to 0.45% (Segmentation) |
| IDEAL PDFF Intra-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -1.26 to 1.05% (ROI); -0.16 to 0.14% (Segmentation) |
| cT1 Inter-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -48.05 to 39.89ms (ROI); -37.84 to 26.51ms (Segmentation) |
| T2* Inter-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -2.64 to 4.90 ms (ROI) |
| DIXON PDFF Inter-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -2.27 to 4.57% (ROI); -0.55 to 1.22% (Segmentation) |
| IDEAL PDFF Inter-Operator Variability | Variation well within prescribed criteria; minor additional variation for ROI method. | -2.09 to 1.82 % (ROI); -0.37 to 0.26% (Segmentation) |
| cT1 Worst-Case Variability | Highly reproducible. | -126.52 to 104.19 ms (ROI); -65.27 to 120.27 ms (Segmentation) |
| T2* Worst-Case Variability | Highly reproducible. | -3.68 to 6.35 ms (ROI) |
| DIXON PDFF Worst-Case Variability | Highly reproducible. | -2.04 to 0.76 % (ROI); -2.72 to 1.24% (Segmentation) |
| IDEAL PDFF Worst-Case Variability | Highly reproducible. | -3.75 to 2.83% (ROI); -1.92 to 1.35% (Segmentation) |
| Substantial Equivalence (LMSv3 vs. LMSv2.1) | Performs as well as its predicate. | |
| Phantom T1 | Negligible difference. | -1.96 to 2.09ms (95% CI Limits of Agreement) |
| Phantom T2* | Negligible difference. | -0.08 to 0.08ms (95% CI Limits of Agreement) |
| Phantom DIXON PDFF (< 30%) | Within 1% of predicate. | -0.18 to 0.10 % (95% CI Limits of Agreement) |
| Phantom DIXON PDFF (≥ 30%) | Within 2% of predicate. | -1.62 to 1.02 % (95% CI Limits of Agreement) |
| In-vivo T1 | cT1 values within 30ms of predicate. | -28.08 to 28.73ms (95% CI Limits of Agreement) |
| In-vivo T2* | T2* values within 2ms of predicate. | -0.43 to 1.69ms (95% CI Limits of Agreement) |
| In-vivo DIXON PDFF | Negligible difference. | -0.18 to 0.10 % (95% CI Limits of Agreement) |
2. Sample Sizes Used for the Test Set and Data Provenance
- Phantom Testing: The sample size for phantom testing is not explicitly stated as a number of phantoms, but it involved phantoms "designed to mimic the human data but provide a wider range" and covered "worst-case scenarios." The data provenance is controlled laboratory conditions, using prepared phantoms.
- In-Vivo Testing (Clinical): The study used "in-vivo volunteer data." The precise number of volunteers is not specified. The data provenance is implicitly prospective, as it refers to "volunteer scans." No country of origin is explicitly mentioned, but the submitter (Perspectum Diagnostics Ltd) is based in the United Kingdom.
- Substantial Equivalence Testing: Used both "phantom measurements" and "in-vivo measurements." The specific sample sizes for these comparisons are not detailed.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not describe the establishment of a "ground truth" through expert consensus for the test set, specifically in the context of diagnostic accuracy. The performance testing focuses on the accuracy and variability of the quantitative measurements (cT1, T2*, PDFF) derived by the software rather than a diagnostic outcome.
For phantom testing, ground truth values are inherent to the precisely calibrated phantoms. For in-vivo testing, variability is measured, and for substantial equivalence, the comparison is against the predicate device's measurements. There is no mention of human experts establishing a ground truth for the quantitative values directly measured by the device for these performance tests.
4. Adjudication Method for the Test Set
Not applicable. The performance testing described does not involve human adjudication of diagnostic outcomes or image interpretation in the way a clinical study for sensitivity/specificity for a diagnosis would. It focuses on the quantitative output of the software.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document describes performance testing for device accuracy, repeatability, reproducibility, and substantial equivalence to a predicate device. It does not mention an MRMC comparative effectiveness study involving human readers with and without AI assistance or an effect size of improvement.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, the performance testing described is primarily standalone performance testing. LiverMultiScan (LMSv3) is explicitly described as a "standalone software application" and a "post-processing, standalone software device." The tests assess the algorithms' performance in generating quantitative metrics (accuracy, repeatability, reproducibility) without requiring human interpretation as part of the core measurement. Operators (trained PD operators) are involved in using the device to generate reports and place ROIs, and inter/intra-operator variability is assessed, but the fundamental measured values are algorithmically derived from the MR data.
7. Type of Ground Truth Used
- Phantom Testing: The ground truth used was established by the precisely characterized properties of the phantoms, which were designed to mimic human data over a wide range of physiological values.
- In-Vivo Testing: For the in-vivo volunteer data, the concept of "ground truth" for the measured parameters (cT1, T2*, PDFF) refers to the true physiological values within the volunteers, against which the device's precision and variability are assessed. While not explicitly stated how this "true" value would be independently confirmed, the focus is on self-consistency (repeatability, reproducibility, operator variability) of the device's measurements rather than comparison to an external gold standard like pathology.
- Substantial Equivalence Testing: The ground truth for this comparison was the performance and measurements of the legally marketed predicate device, LiverMultiScan (LMSv2.1).
8. Sample Size for the Training Set
The document does not explicitly state the sample size (or any details) for a "training set." This type of detail is typically associated with AI/machine learning models where a dataset is used to train the algorithm. While LMSv3 includes "New algorithms" like Automatic Liver Segmentation, the document does not elaborate on how these algorithms were developed or if they involved a distinct training phase with a specific dataset.
9. How the Ground Truth for the Training Set Was Established
Since a "training set" with established ground truth is not detailed, this information is not provided in the document.
Ask a specific question about this device
(57 days)
MRCP+v1 is indicated for use as a software-based image processing system for non-invasive, quantitative assessment of biliary system structures by facilitating the generation, visualisation and review of three-dimensional quantitative biliary system models and anatomical image data.
MRCP+v1 calculates quantitative three-dimensional biliary system models that enable measurement of bile duct widths and automatic detection of regions of variation (ROV) of tubular structures. MRCP+v1 includes tools for interactive segmentation and labelling of the biliary system and tubular structures. MRCP+v1 allows for regional volumetric analysis of segmented tree-like, tubular structures and the gallbladder.
Combining image viewing, processing and reporting tools, the metrics provided is designed to support physicians in the visualization, evaluation and reporting of hepatobiliary structures. These models and the physical parameters derived from the models, when interpreted by a trained physician, yield information that may assist in biliary system assessment.
MRCP+v1 is designed to utilize DICOM compliant MRCP datasets, acquired on supported MR scanners using supported MRCP acquisition protocols.
MRCP+v1 is suitable for all patients not contra-indicated for MRI.
MRCP+v1 is a standalone software device. The purpose of the MRCP+v1 device is to assist the trained operator with the evaluation of information from Magnetic Resonance (MR) images from a single time-point (patient visit). A trained operator, typically a radiographer or technician trained in radiological anatomy, loads as input into the MRCP+v1 device previously acquired MRCP data. The device is intended to be used by a trained operator to generate metrics in the form of a summary report to enable reporting by a radiologist for subsequent interpretation and diagnosis by a clinician.
MRCP+v1 is designed to utilize DICOM 3.0 compliant MRCP datasets, acquired on supported MR scanners, to display the fluid-filled tubular structures in the abdomen, including the intra- and extrahepatic biliary tree, gallbladder and pancreatic ductal system.
Analysis and evaluation tools include:
- Segmentation of structures utilising user input of seeding points.
- Interactive labelling of segmented areas.
- Quantitative measurement derived from segmentation and labelling results.
MRCP+v1 calculates quantitative three-dimensional biliary system models that enable measurement of bile duct widths and automatic detection of regions of variation (ROV) of tubular structures. MRCP+v1 includes tools for interactive segmentation and labelling of the biliary system and tubular structures. MRCP+v1 allows for regional volumetric analysis of segmented tree-like, tubular structures and the gallbladder.
The radiologist may use existing radiological tools to report on the biliary tree. The reviewing radiologist needs to take into consideration the device's limitations and accuracy during review. The MRCP+v1 report is intended to supplement a conventional radiology report, for interpretation by a clinician.
MRCP+v1 does not replace the usual procedures for assessment of the biliary system by a reviewing radiologist or interpreting clinician, providing many opportunities for competent human intervention in the interpretation of images and information displayed.
The metrics are intended to be used as an additional input to radiologists and clinicians in addition to that provided by conventional MRCP.
The provided FDA 510(k) document for MRCP+ v1.0 describes its acceptance criteria and the studies performed to demonstrate its performance.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in terms of specific performance thresholds that must be met. Instead, it reports the observed performance metrics and concludes that the device is "highly accurate," "highly repeatable," and "reproducible," and that "the variation introduced by operator analysis is well within the prescribed acceptance criteria set for performance testing."
However, we can infer the acceptance criteria are related to the reported 95% Confidence Intervals (CI) for the Limits of Agreement, with the implicit goal being that these limits should be acceptably narrow.
Here's a table summarizing the reported device performance:
| Performance Metric | Reported Device Performance (95% CI Limit of Agreement) |
|---|---|
| Algorithmic Accuracy – Digital Synthetic Data | |
| Clinical Phantom | Upper: 0.9mm, Lower: -0.9mm |
| Tube Width Phantom | Upper: 1.3mm, Lower: -0.6mm |
| Device Accuracy – Physical Phantoms | |
| Clinical Phantom | Upper: 1.0mm, Lower: -1.1mm |
| Tube Width Phantom | Upper: 0.7mm, Lower: -0.9mm |
| Precision (Repeatability) – Physical Phantoms | |
| Clinical Phantom | Upper: 0.4mm, Lower: -0.4mm |
| Tube Width Phantom | Upper: 0.3mm, Lower: -0.3mm |
| Precision (Reproducibility) – Physical Phantoms | |
| Clinical Phantom | Upper: 0.5mm, Lower: -1.1mm |
| Tube Width Phantom | Upper: 0.7mm, Lower: -0.8mm |
| Precision (Repeatability) – In-vivo Data | |
| Tree Volume (ml) | Upper: 3.4, Lower: -3.5 |
| Gallbladder Volume (ml) | Upper: 4.4, Lower: -9.4 |
| 3-5mm (%) | Upper: 17.7, Lower: -19.2 |
| 5-7mm (%) | Upper: 8.5, Lower: -8.0 |
| Greater than 7mm (%) | Upper: 1.6, Lower: -1.8 |
| Less than 3mm (%) | Upper: 20.2, Lower: -19.2 |
| Duct Median (mm) | Upper: 1.7, Lower: -2.0 |
| Duct Minimum (mm) | Upper: 1.9, Lower: -2.2 |
| Duct Maximum (mm) | Upper: 3.2, Lower: -2.9 |
| Duct IQR (mm) | Upper: 1.8, Lower: -1.7 |
| Precision (Reproducibility) – In-vivo Data | |
| Tree Volume (ml) | Upper: 4.7, Lower: -6.9 |
| Gallbladder Volume (ml) | Upper: 13.8, Lower: -18.4 |
| 3-5mm (%) | Upper: 27.8, Lower: -15.8 |
| 5-7mm (%) | Upper: 10.8, Lower: -11.8 |
| Greater than 7mm (%) | Upper: 2.6, Lower: -3.1 |
| Less than 3mm (%) | Upper: 30.8, Lower: -23.4 |
| Duct Median (mm) | Upper: 2.6, Lower: -2.8 |
| Duct Minimum (mm) | Upper: 2.6, Lower: -2.8 |
| Duct Maximum (mm) | Upper: 4.0, Lower: -3.5 |
| Duct IQR (mm) | Upper: 1.9, Lower: -2.3 |
| Intra-Operator Performance | |
| Tree Volume (ml) | Upper: 0.9, Lower: -0.6 |
| Gallbladder Volume (ml) | Upper: 0.0, Lower: 0.0 |
| 3-5mm (%) | Upper: 7.4, Lower: -7.7 |
| 5-7mm (%) | Upper: 5.0, Lower: -5.1 |
| Greater than 7mm (%) | Upper: 3.9, Lower: -4.7 |
| Less than 3mm (%) | Upper: 11.7, Lower: -10.7 |
| Duct Median (mm) | Upper: 0.4, Lower: -0.4 |
| Duct Minimum (mm) | Upper: 0.5, Lower: -0.4 |
| Duct Maximum (mm) | Upper: 2.0, Lower: -1.6 |
| Duct IQR (mm) | Upper: 0.5, Lower: -0.4 |
| Inter-Operator Performance | |
| Tree Volume (ml) | Upper: 3.6, Lower: -6.8 |
| Gallbladder Volume (ml) | Upper: 0.6, Lower: -0.7 |
| 3-5mm (%) | Upper: 16.8, Lower: -11.1 |
| 5-7mm (%) | Upper: 6.3, Lower: -6.3 |
| Greater than 7mm (%) | Upper: 1.4, Lower: -1.7 |
| Less than 3mm (%) | Upper: 12.3, Lower: -17.4 |
| Duct Median (mm) | Upper: 1.1, Lower: -1.8 |
| Duct Minimum (mm) | Upper: 1.0, Lower: -1.3 |
| Duct Maximum (mm) | Upper: 2.0, Lower: -3.2 |
| Duct IQR (mm) | Upper: 1.6, Lower: -1.9 |
2. Sample Size Used for the Test Set and Data Provenance
- Algorithmic Accuracy (Digital Synthetic Data): "a dataset of digital synthetic data" - no specific number of samples is provided.
- Device Accuracy (Physical Phantoms): "Two different types of phantom, a clinical and tubewidth" - no specific number of scans or repetitions is provided. One would infer 'multiple' scans given the reporting of CI, but the exact count isn't in the provided text.
- Precision (In-vivo Data): "in-vivo data acquired from volunteers" - no specific number of volunteers or scans is provided.
- Intra- and Inter-Operator: No specific sample size (number of cases or readings) is provided.
Data Provenance: The provenance is described as:
- "digital synthetic data"
- "phantom scans"
- "in-vivo (healthy volunteer and patients with suspected hepatobiliary disease) scans"
- MR systems from Siemens, GE, and Philips across 1.5T and 3T field strengths were used for in-vivo testing (and presumably for phantoms as well).
- The document implies the company is based in the United Kingdom ("OXFORD, OX1 2ET OXFORDSHIRE UNITED KINGDOM"), suggesting the data could originate from there or other international sites. The text doesn't specify if the data was retrospective or prospective, though "acquired from volunteers" might suggest prospective data collection for the precision study if these were specifically for validation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not explicitly state the number of experts or their qualifications used to establish ground truth for the test set. The data provenance for "Algorithmic Accuracy – Digital Synthetic Data" mentions "the duct file specification of the physically printed phantoms," which suggests an objective, engineered ground truth rather than expert consensus on images for this part of the testing. For in-vivo data, the ground truth establishment method isn't detailed, nor are the experts.
It does state that the device is to be used by a "trained operator, typically a radiographer or technician trained in radiological anatomy" to generate reports for "reporting by a radiologist for subsequent interpretation and diagnosis by a clinician." This describes the user, but not the experts for ground truth.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for establishing ground truth for the test set.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No Multi Reader Multi Case (MRMC) comparative effectiveness study evaluating human reader improvement with AI assistance versus without AI assistance is mentioned in the provided text. The "Intra- and Inter-Operator" section assesses the variability of the device's output when operated by different or the same human operators, not directly the diagnostic performance improvement of human readers using the device.
6. Standalone (Algorithm-Only) Performance Study
Yes, a standalone performance study was conducted. The section "Algorithmic Accuracy – Digital Synthetic Data" describes testing the algorithm's accuracy "against the duct file specification of the physically printed phantoms, a dataset of digital synthetic data." The conclusion clarifies this: "It can be said that the algorithmic accuracy of MRCP+v1 when used to quantify tubular structures from acquired MRCP data is high when not presented with the variability introduced by phantom or in-vivo scanning." This directly speaks to the algorithm's performance independent of scanning variability, which constitutes a standalone assessment.
7. Type of Ground Truth Used
- Algorithmic Accuracy: "Duct file specification of the physically printed phantoms" (an engineered/objective ground truth).
- Device Accuracy (Physical Phantoms): The document implies the "physical phantoms" themselves serve as the ground truth or have known parameters against which the device's measurements are compared.
- Precision (In-vivo Data) and Intra-/Inter-Operator: The ground truth for these measurements (Tree Volume, Gallbladder Volume, Duct Median, etc.) would be the presumed true values derived from the MRCP images, but the method of establishing this ground truth (e.g., expert manual segmentation, another reference standard) is not explicitly described. It measures consistency and agreement between measurements rather than agreement with an external gold standard for these particular sections.
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size used for the training set.
9. How the Ground Truth for the Training Set Was Established
The document does not provide any information on how the ground truth for the training set was established.