510(k) Data Aggregation

    K Number
    K212717

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-11-30

    (460 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Paneffort AAMI Level 3 Isolation Gowns are intended to be worn by healthcare personnel in isolations to provide moderate barrier protection for health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Paneffort AAM Level 3 Isolation Gowns meet the requirements of level 3 Liquid Barrier Performance as per AAMI PB70:2012 and are provided non-sterile and are single use only.

    Device Description

    The Paneffort AAMI Level 3 Isolation Gown is a surgical isolation gown with moderate barrier protection identified by Regulation 21 CFR 878.4040 under FDA product code FYC. The Paneffort AAMI Level 3 Isolation Gown is a single use, disposable medical device provided non-sterile. Paneffort AAMI Level 3 Isolation Gown is offered in one color (blue) and six different sizes of Small, Medium, Regular/Large, X-Large, 2X-Large, and 3X-Large. Each model is constructed of a nonwoven Spunbond-Meltblown-Spunbond (SMS) material that is tested according to ANSI/AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities and meets AAMI Level 3.

    The Paneffort AAMI Level 3 Isolation Gown consists of one critical zone throughout the entire gown including seams, but excluding cuffs, hems, and bindings as per AAMI PB70:2012. It is a protective tape sealed gown along with other design features like tieneck, elastic cuffs, and belt tie, that meets AAMI PB70 Level 3 barrier performance for protective apparel.

    AI/ML Overview

    This document describes the non-clinical testing performed on the Paneffort AAMI Level 3 Isolation Gown to demonstrate its substantial equivalence to a predicate device.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    S.No.Test PerformedAcceptance CriteriaReported Device PerformanceConclusion
    1.Bursting Strength≥ 40 kPa (≥5.80 psi)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    2.Basis Weight≥35 g/m²The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    3.Linting TestCo-coefficient of Linting ≤4.2The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    4.Air Permeability≥15 ft³/min/ft²The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    5.Water Resistance: Hydrostatic Pressure≥ 50cm H₂O (Ac: 3, Re: 4); AQL: 4% (Level 3 ANSI/AAMI PB70:2012)Lot 1: 3 out of 32 samples failed. Lot 2: 3 out of 32 samples failed. Lot 3: 2 out of 32 samples failed. Despite these failures, the overall conclusion states: "The test meets the Acceptance Criteria."The test meets the Acceptance Criteria
    6.Water Resistance: Impact Penetration≤1.0g (Ac: 3, Re: 4); AQL: 4% (Level 3 ANSI/AAMI PB70:2012)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    7.Tensile Strength/Breaking Strength≥30N (≥7 lbf)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    8.Tearing Strength≥10N (≥2.3 lbf)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    9.Seam Strength≥30N (≥7 lbf)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    10.FlammabilityClass I. Textile exhibiting normal flammability.The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    11.Water Vapour resistance0,0-1,0 kPa.m²/W (This appears to be a range, but typically there would be a specific limit or target for water vapor resistance for PPE)The test meets the Acceptance CriteriaThe test meets the Acceptance Criteria
    Biocompatibility: In Vitro CytotoxicityNon-cytotoxicUnder the conditions of the study, the device is non-cytotoxic.The test meets the Acceptance Criteria
    Biocompatibility: Skin IrritationNot an irritantUnder the conditions of the study, not an irritant.The test meets the Acceptance Criteria
    Biocompatibility: Skin SensitizationNot a sensitizerUnder the conditions of the study, not a sensitizer.The test meets the Acceptance Criteria

    Note on Water Resistance (Hydrostatic Pressure): The reported performance shows individual sample failures within each lot (3 out of 32 in Lot 1, 3 out of 32 in Lot 2, and 2 out of 32 in Lot 3). However, the "Conclusion" column for this test still states "The test meets the Acceptance Criteria." This implies that the passing condition is based on an Allowable Quality Level (AQL) or a specific sampling plan where a certain number of defects are permitted within the sample size while still meeting the overall acceptance criteria (Ac: 3, Re: 4; AQL: 4%).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Sizes:
      • Bursting Strength, Basis Weight, Linting Test, Air Permeability, Tensile Strength/Breaking Strength, Tearing Strength, Seam Strength, Flammability, Water Vapour resistance: 32 samples each from 3 different lots (total 96 samples per test).
      • Water Resistance: Hydrostatic Pressure and Impact Penetration: Lot 1: 74 pieces; Lot 2: 72 pieces; Lot 3: 72 pieces (total 218 pieces across 3 lots for each test).
    • Data Provenance: The document does not explicitly state the country of origin for the testing data. However, the manufacturer is Paneffort (Cambodia) Garment Co. Ltd., and the specification developer is Paneffort, LLC (USA). The testing standards are international (ASTM, AATCC, ISO, CFR), suggesting tests were conducted in reputable labs, but the specific location of the test facilities is not provided. The data is retrospective, as it was collected to support a 510(k) submission for a device that has already been manufactured and tested.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This study is a non-clinical performance testing of a physical medical device (isolation gown) against established engineering and material science standards. It does not involve human interpretation of data or diagnostic tasks where expert ground truth would be required. The "ground truth" here is objective measurement against specified criteria.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As no expert review or human interpretation for diagnosis is involved, there is no adjudication method in the traditional sense. The pass/fail criteria are defined by the testing standards and AQLs.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. An MRMC study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images). This submission is for a physical medical device (isolation gown) and does not involve human readers or diagnostic cases.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This is not an AI/algorithm-based device. It is a physical product. The performance reported is that of the gown itself, as tested in a lab environment.

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance is based on established industry and regulatory standards for protective apparel. This includes:
      • ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities (specifically Level 3).
      • Various ASTM, AATCC, ISO, and CFR standards for material properties and flammability.
      • ISO 10993 for biocompatibility.
    • The tests generate objective, measurable data that is compared directly to the numerical acceptance criteria defined in these standards.

    8. The Sample Size for the Training Set:

    • Not Applicable. This device does not use machine learning or AI, and therefore does not have a "training set." The performance is inherent to its physical design and materials, verified through non-clinical testing.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221409

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-25

    (70 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Paneffort 3-ply Surgical and Procedural Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

    Device Description

    The Paneffort 3-Ply Surgical and Procedural masks are flat pleated 3-ply device, which consist of three layers, i.e., Outer layer is made of Spunbond Polypropylene, Middle/Filter layer is made of Melt-blown Polypropylene and Inner layer is made of Spunbond Polypropylene. There are two options for the mask to be secured on the user, viz. earloops or tie-on. Earloops are made of spandex and polyamide and the tie-on model are made of spunbond polypropylene. The nosepiece is made up of Iron wire, polypropylene and Zinc, and is encased between Outer and Middle later. Paneffort 3-ply Surgical and Procedural masks are available in blue colour.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for Paneffort 3-Ply Surgical and Procedural Masks. It details the device's technical specifications, intended use, and comparative testing against predicate devices to demonstrate substantial equivalence.

    Based on the provided information, here's a description of the acceptance criteria and the study that proves the device meets them:

    Device: Paneffort 3-Ply Surgical and Procedural Masks
    Regulatory Class: Class II
    Product Code: FXX
    Regulation Name: Surgical Apparel (21 CFR 878.4040)

    The study focuses on non-clinical performance and biocompatibility testing, primarily against ASTM F2100 standards for surgical masks. It does not describe a study involving human readers or AI-assisted performance analysis, as the device is a physical medical mask, not a diagnostic AI system or image analysis tool. Therefore, questions 2, 3, 4, 5, 8, and 9 are largely not applicable as they pertain to clinical studies, AI performance evaluation, and ground truth establishment for such studies, which are not relevant to this type of device submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from ASTM F2100-19e1, an FDA-recognized consensus standard for medical face mask materials. The performance results show that the Paneffort masks meet or exceed these criteria for both Level 2 and Level 3 masks.

    ParametersAcceptance Criteria (ASTM F2100-19e1)Reported Device Performance (Paneffort 3-Ply Surgical/Procedural Masks)Comments
    Bacterial Filtration Efficiency (BFE) (ASTM F2101)≥98%Pass at >98% (for Level 2 and Level 3)Meets criteria. Table 5 (page 5) lists ">99.9 %" for subject device and predicate, while Table 3 (page 9) states ">98%". The most stringent stated passing result for the subject device is ">99.9 %". The crucial point is that it Passes at > 98%.
    Differential Pressure (Delta-P) (EN 14683 Annex C)<6mm H2O/cm²<6mm H2O/cm² (specifically 3.175 mmH2O/cm²)Meets criteria.
    Sub-micron Particulate Filtration Efficiency (PFE) (ASTM F2299)≥98%Pass at >99% (for Level 2 and Level 3)Meets criteria. Table 5 (page 5) lists "≥ 99.20%" for subject device and predicate, while Table 3 (page 9) states ">99%". The crucial point is that it Passes at > 98%.
    Resistance to Penetration by Synthetic Blood (ASTM F1862)≥120 mm Hg (for Level 2)Pass at ≥120 mm Hg (specifically ≥120 mm Hg, and tested at 160 mm Hg)Meets criteria for both Level 2 and Level 3. The test result for the subject device is "Pass at 160 mm Hg", exceeding the 120 mm Hg requirement for Level 2 and meeting the 160 mm Hg requirement for Level 3.
    Flammability (16 CFR Part 1610)Class 1Class 1Meets criteria.
    Biocompatibility - Cytotoxicity (ISO 10993-5)Non-cytotoxicPass - Non-cytotoxicMeets criteria.
    Biocompatibility - Irritation (ISO 10993-10)Non-irritatingPass - Non-irritatingMeets criteria.
    Biocompatibility - Sensitization (ISO 10993-10)Non-sensitizingPass - Non-sensitizingMeets criteria.

    2. Sample Size and Data Provenance

    • Sample Size: The document does not specify the exact sample size for each performance test (e.g., how many masks were tested for BFE, PFE, etc.). It only provides the acceptance criteria and whether the tested product met them.
    • Data Provenance: The data is gathered from non-clinical laboratory testing of the subject device (Paneffort 3-Ply Surgical and Procedural Masks). The manufacturer is Paneffort (Cambodia) Garment Co. Ltd. (Cambodia), and the legal manufacturer is PANEFFORT, LLC (USA). The testing was conducted in accordance with FDA-recognized consensus standards (ASTM and ISO). The document does not specify if the tests were performed retrospectively or prospectively, but typically, these are prospective tests conducted on representative samples of the manufactured product.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable. This is a non-clinical performance and biocompatibility study of a physical device (surgical mask), not an AI or image analysis system requiring expert review for ground truth establishment. The "ground truth" is the objective measurement defined by the standards (e.g., a BFE of >98%).

    4. Adjudication Method for Test Set

    • Not applicable. This is a non-clinical performance and biocompatibility study. Adjudication methods are relevant for clinical studies, especially those involving subjective assessments or disagreements among human readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not applicable. This study does not involve human readers or AI assistance. It is a technical performance assessment of a surgical mask.

    6. Standalone (Algorithm Only) Performance

    • Not applicable. This is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used

    • The "ground truth" for this device and study type is established by objective, standardized test methods and criteria defined by international consensus standards (ASTM F2100-19e1, ASTM F2101, EN 14683, ASTM F2299, ASTM F1862, 16 CFR Part 1610, ISO 10993-5, ISO 10993-10). The performance of the mask is measured directly against these established physical and biological benchmarks.

    8. Sample Size for Training Set

    • Not applicable. This is a physical medical device, not an AI model, so there is no "training set."

    9. How Ground Truth for Training Set was Established

    • Not applicable. As there is no training set for an AI model, this question is not relevant.