Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203727

    Validate with FDA (Live)

    Date Cleared
    2021-08-05

    (227 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Harmony® Latex & the Harmony® Polyisoprene Dams are used as a barrier when engaging in oral/vaginal and oral/ anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections.

    Device Description

    The Harmony® Latex and Harmony® Polyisoprene Dams are rectangular shaped non-porous, intact polymer film that are placed over the vagina or anus during oral sexual contact. The dam provides coverage to the external female genitalia or to the anal acts as a physical barrier to infectious agents responsible for the transmission of sexually transmitted infections (STIs). Both Harmony® Latex and Harmony® Polyisoprene Dams are 152.4 mm (6") x 254 mm (10") and 0.12 mm thick. The Harmony® Latex Dam contains natural polyisoprene rubber and the Harmon® Polyisoprene Dam contains synthetic polyisoprene rubber. The dams are slightly scented with vanilla fragrance and are provided in natural color.

    AI/ML Overview

    This document is a 510(k) summary for the Harmony® Latex Dams and Harmony® Polyisoprene Dams. It outlines the acceptance criteria and supporting studies for these devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the international standard ISO 29942:2011 Prophylactic dams -- Requirements and test methods and the 2020 FDA guidance Use of International Standard ISO 10993-1. The specific ranges or thresholds for these criteria are not provided in the document, but it states that "All test results were acceptable."

    Acceptance Criteria CategorySpecific Tests / PropertiesReported Device Performance
    Physical Performance (ISO 29942:2011)Dimensional TestingAcceptable (Implies meeting specified dimensions: 152.4 mm x 254 mm, Thickness- 0.12 mm)
    Tensile Testing and Elongation at BreakAcceptable
    Tear Resistance and Tearing ForceAcceptable
    Freedom from Holes and Visual DefectsAcceptable
    Barrier IntegrityViral Barrier PropertiesAcceptable
    Packaging RobustnessPackaging Integrity TestingAcceptable
    Biocompatibility (ISO 10993-1, -5, -10, -11)Cytotoxicity (ISO 10993-5:2009)Acceptable
    Irritation Testing (ISO 10993-10:2010)Acceptable
    Sensitization (ISO 10993-10:2010)Acceptable
    Acute Systemic Toxicity (ISO 10993-11:2017)Acceptable

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for each test in the acceptance criteria section. The data provenance is also not specified in terms of country of origin or whether it was retrospective/prospective. The testing appears to be prospective bench testing performed on manufactured Harmony® Latex and Harmony® Polyisoprene Dams.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the device is a physical barrier (dam) and the testing involves non-clinical bench tests (mechanical, barrier, biocompatibility), not interpretation of medical images or patient data that would require expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    This section is not applicable for the same reasons as #3. Bench testing results are typically objectively measured against specified standards, not adjudicated by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This section is not applicable. An MRMC study is relevant for diagnostic devices that involve human interpretation of results, especially in the context of AI assistance. The Harmony® Dams are physical barrier devices.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This section is not applicable. There is no algorithm or AI component in the Harmony® Dams. The performance evaluated is the physical and biological properties of the device itself.

    7. Type of Ground Truth Used

    The ground truth used for the non-clinical tests is established by international standards and FDA guidance. For example:

    • ISO 29942:2011 specifies requirements and test methods for prophylactic dams.
    • ISO 10993 series outlines biological evaluation of medical devices.
      The ground truth is therefore defined by the specifications and thresholds outlined in these recognized standards.

    8. Sample Size for the Training Set

    This section is not applicable. There is no "training set" as this device is not an AI/ML algorithm or a diagnostic tool that learns from data.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1