510(k) Data Aggregation

    K Number
    DEN120011

    Validate with FDA (Live)

    Date Cleared
    2012-07-10

    (60 days)

    Product Code
    Regulation Number
    880.6305
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Proteus Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relative to gravity, and time-stamped, patientlogged events, including events signaled by swallowing the Ingestion Event Marker (IEM) accessory. The Proteus Personal Monitor enables unattended data collection for clinical and research applications. The Proteus Personal Monitor may be used in any instance where quantiffable analysis of event-associated heart rate, activity, and body position is desirable.

    Device Description

    The Proteus Personal Monitor, also called the "Patch", is a body-worn sensor that collects physiological and behavioral metrics including heart rate, activity, body angle and time-stamped user-logged events generated when a user marks an event by swallowing an Ingestion Event Marker (IEM) or by manually pressing an event marker button on the Patch. The Patch stores and wirelessly sends the IEM data to a general computing device.

    The Proteus Personal Monitor Ingestion Event Marker system is comprised of three main subsystems; (1) the ingestion event marker (IEM), (2) the data recorder (Patch), and (3) the Proteus software.

    1. Ingestion Event Marker (IEM)
      The grain-of-sand sized IEM is designed to communicate the time-stamped confirmation of IEM device ingestion as a unique identifier to the Proteus Personal Monitor worn on the skin. The ingestion signal is communicated via volume conduction communication also known as intrabody communication. The IEM is attached to an inert pharmaceutical excipient tablet for ease of handling and swallowability.

    2. Proteus Personal Monitor (Patch)
      The Proteus Personal Monitor (Patch) receives, stores, and wirelessly sends ingestion confirmation data to a general computing device.

    3. Software
      The Proteus software is used to pair the Patch with a mobile computing device. The software organizes and displays ingestion events.

    AI/ML Overview

    Here's the breakdown of the acceptance criteria and the study details for the Proteus Personal Monitor including Ingestion Event Marker:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility/Non-toxicity: Demonstrate the device is biocompatible and non-toxic.IEM: - Cytotoxicity: No unintended compounds detected above stringent ICH reporting threshold. Realistic exposure levels are non-cytotoxic. - Copper (Cu) Toxicity: No risk of Cu toxicity with realistic exposure. Realistic exposure levels are non-cytotoxic.
    Mechanical Safety (Excretion & GI Injury): Ingested IEMs reliably excreted, and supra-normal doses do not inflict clinically significant injuries.Canine studies: Ingested IEMs reliably excreted. Supra-normal doses of IEMs do not inflict any clinically significant injuries.
    Electrical Safety (Tissue Stimulation, Leakage Current, Dielectric Withstand, etc.): No abnormal ECG morphology or arrhythmia; all applicable electrical safety tests passed.Canine studies: No abnormal ECG morphology or arrhythmia.
    In Vivo Toxicity: No evidence of toxicity even at high doses in relevant animal models.14-day rat oral gavage study: No evidence of toxicity in any dosing groups, including max dose (equivalent to 30,000 IEMs/day). Canine oral toxicology study: No evidence of IEM toxicity. Rodent oral toxicology study: No evidence of IEM toxicity (even in highest dose group, equivalent of 30,000 IEMs/day). IEM copper (Cu) human health assessment: Practical-use scenario (15 IEMs simultaneously, daily or twice-daily) poses no risk of copper toxicity. Extreme-use scenario (30 IEMs simultaneously, daily) poses no risk of systemic toxicity, but transient, non-systemic gastric upset could result (mitigated by intake with meal). Post-operative renal transplant patients not at greater risk. Physiological response to Cu in IEMs will not differ from Cu in food.
    Electromagnetic Compatibility (EMC): All applicable EMC tests passed.EMC testing (Patch and IEM): Passed EN 60601-1-2, EN 55011 (CISPR 11), and IEC 60601-1-2:2007 6.2.3.
    Device Performance (Event Recording): Characterize positive detection accuracy (PDA) and negative detection accuracy (NDA) for IEM ingestions.Clinical Studies: - Cumulative average Positive Detection Accuracy (PDA): 97.2% (95% CI) - Cumulative average Negative Detection Accuracy (NDA): 100% (95% CI) - Overall System Detection Accuracy: 99.3%
    Software: Software development processes, hazard analysis, and system performance testing conform to FDA guidance.Patch software reviewed in K093976 in conformance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005.

    Study Information

    1. A table of acceptance criteria and the reported device performance:
    See table above.

    2. Sample size used for the test set and the data provenance:

    • Human Clinical Test Set:

      • Number of subjects: 219 subjects ingested IEMs across multiple studies, with 254 subjects wearing the Proteus Personal Monitor (Patch) in total.
      • Number of IEM ingestions: 11,655 cumulative IEM ingestions.
      • Subject/days of system use: 3,811 cumulative subject-days.
      • Data Provenance: Not explicitly stated, but the mention of "The PROMITTER substudy was conducted on the Proteus campus and enrolled 5 healthy volunteer subjects" suggests some studies were conducted in the US. The overall clinical data likely comes from studies conducted by Proteus Biomedical, Inc. (based in Redwood City, CA). The data appears to be prospective as it involves active participation in clinical trials.
    • Animal Test Set:

      • Number of animals: Forty-two (42) in-vivo studies in rodent, canine, and porcine models. The specific number of animals per study is not detailed.
      • Data Provenance: Not explicitly stated, but likely from studies commissioned by Proteus Biomedical, Inc. and conducted in laboratory settings. These are prospective animal studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Clinical Studies (IEM Ingestion Detection): The document does not specify the number of experts or their qualifications for establishing the ground truth of IEM ingestion events in the clinical studies. Typically, this would involve direct observation, patient logs, or other objective measures as the ground truth for "ingestion event" rather than expert interpretation of a signal.
    • ECG Performance Testing: For the MIT-BIH arrhythmia database, the ground truth is established by cardiologists and is inherent to the database itself. For the PROMITTER substudy, it is not specified how the ECG ground truth was established, but typically this would involve a reference ECG device and interpretation by cardiology experts.

    4. Adjudication method for the test set:

    • The document does not explicitly describe an adjudication method (like 2+1 or 3+1) for disagreement among experts, which typically applies to subjective interpretations. For IEM ingestion events, the ground truth is likely directly observed or logged, rather than relying on interpretation that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers improving with or without AI assistance was not conducted or reported in this document. The device's primary function is to automatically detect ingestion events, not to assist human readers in interpreting medical images or other data that typically involves MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the core performance metrics "Positive Detection Accuracy (PDA)" and "Negative Detection Accuracy (NDA)" for IEM ingestion detection represent the standalone (algorithm only) performance of the Proteus Personal Monitor system. The system automatically detects and logs these events without direct human-in-the-loop intervention for each individual ingestion detection.

    7. The type of ground truth used:

    • IEM Ingestion Detection: The ground truth for IEM ingestion events in human clinical studies is implicitly based on documented ingestions by the subjects. This would likely involve patient logs, direct observation during controlled settings, or other objective confirmation of "patient-logged events" as described in the Indications for Use.
    • Physiological Measurements (ECG, Accelerometer, Respiratory Rate): For these, the ground truth was established using:
      • Standard databases: MIT-BIH arrhythmia database for ECG performance.
      • Reference measurements: Applied values for accelerometers, controlled breathing rates for respiratory rate.
    • Safety (Toxicity, Biocompatibility): Ground truth established through established laboratory testing protocols, histopathology, clinical observations, and established toxicology assessments by experts (e.g., Gradient Corp for metal toxicology).

    8. The sample size for the training set:

    • The document does not provide details on the sample size used for training the algorithms. This information is typically proprietary or not included in regulatory summaries focused on validation.

    9. How the ground truth for the training set was established:

    • The document does not provide details on how the ground truth for the training set was established. Similar to the test set, it would likely involve objective data related to IEM ingestion, physiological signals with known parameters, or established clinical diagnoses for specific outcomes targeted by any machine learning components.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093976

    Validate with FDA (Live)

    Date Cleared
    2010-03-25

    (91 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Raisin™ Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relatively to gravity, and time-stamped, patient-logged events. The Raisin™ Personal Monitor enables unattended data collection for clinical and research applications. The Raisin™ Personal Monitor may be used in any instance where quantifiable analysis of eventassociated heart rate, activity, and body position is desirable.

    Device Description

    The Raisin100 Personal Monitor (RPM) is a miniaturized, ambulatory, battery-operated datalogging device that is worn on the torso to record heart rate, activity, and patient-logged events. Patient-logged events can be extrinsic (e.g., dosing of a medication) or intrinsic (e.g., a symptom) and are time-stamped using a manual button on the device, in order to contextualize the physiologic measures. Subjective meaning of these events is assigned by the user. In addition to quantification of physical motion, signals from the device's accelerometer are used to determine body position relative to gravity. Electrode-to-electrode impedance is also measured to assess whether the device is attached properly to the user. RPM recorded data are transferred via Bluetooth telemetry to a general computing device for display and conversion for export to other programs. The RPM is available in two form factors to accommodate individual comfort preferences: one-piece and two-piece. The functionality, intended use, duration and location of wear, and fundamental scientific technologies are exactly the same between the two RPM form factors.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Raisin™ Personal Monitor, based on the provided text:

    Acceptance Criteria and Device Performance

    ParameterAcceptance Criteria (Expected Results)Reported Device Performance (Algorithm Results)
    Heart Rate R-wave Detection (Bench Testing - ANSV/AAMI EC 13 standard)
    Default ECG waveform80 bpm80.0 bpm
    T-wave rejection (R-wave 1 mV, T-wave 0.4 mV)80 bpm80.0 bpm
    Ventricular bigeminy80 bpm79.9 bpm
    Slow alternating ventricular bigeminy60 bpm60.5 bpm
    Rapid alternating ventricular bigeminy120 bpm119.8 bpm
    Bidirectional systoles90 bpm90.1 bpm
    Default ECG waveform (Pacing pulse 2 mV, 2 ms width)80 bpm80.0 bpm
    Heart Rate R-wave Detection (Arrhythmia Database)
    Positive Detection AccuracyNot explicitly stated (implied high)Median: 99.7%, Standard Deviation: 5.9%
    False Positive RateNot explicitly stated (implied low)Median: 0%, Standard Deviation: 1.7%
    Heart Rate R-wave Detection (Clinical Setting - Various Body Locations)Not explicitly stated (implied near 100%)Anterior Chest: 99.40%Xyphoid: 99.17%Stomach: 99.07%Lateral Chest: 98.82%(Average R-wave detection accuracy)
    Accelerometer (Bench Validation)Known acceleration applied against each of its three axes to demonstrate linearityShown in scatter plots with strong linear correlation between measured and applied acceleration for X, Y, and Z axes.
    Accelerometer (Clinical Validation)Capture expected features of subject movement (e.g., walking)Demonstrated data from a representative walking test showing expected acceleration fluctuations.

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Heart Rate R-wave Detection (Arrhythmia Database): All 48 test files from the MIT-BIH arrhythmia database.
      • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): Data from 4 subjects (Subject 1, Subject 2, Subject 4, Subject 5) across different body locations. The provenance is not explicitly stated but implies prospective clinical data collection for these subjects.
      • Accelerometer (Clinical Validation): Data from a "representative walking test" and "a representative subject." The exact number of subjects or tests is not explicitly stated.
      • Accelerometer (Bench Validation): Not applicable for test set size as it's a bench validation with known inputs.
      • Data Provenance: The MIT-BIH arrhythmia database is a publicly available, retrospective database. Clinical data for the heart rate and accelerometer validation appears to be prospectively collected (e.g., "representative walking test," "subject 1, chest location, sitting").
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Heart Rate R-wave Detection (Arrhythmia Database): Not specified in the provided text. The MIT-BIH arrhythmia database has widely accepted expert annotations, but the number and qualifications of the original annotators are not detailed here.
      • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): Not specified. The reference is to "automatically identified R-waves highlighted" against the captured ECG waveform, implying a comparison against the raw ECG, but the method for establishing the true R-wave locations for accuracy calculation (e.g., manual expert review) is not detailed.
      • Accelerometer (Bench Validation): Not applicable, as ground truth is the "known acceleration applied."
      • Accelerometer (Clinical Validation): Not specified. The "expected features" of walking would likely derive from general biomechanical understanding, not specific expert annotations for each test.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not specified in the provided text for any of the studies.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed or reported. This device is primarily a data logger and not an AI-assisted diagnostic tool for human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the reported performance metrics (e.g., R-wave detection accuracy, false positive rate, accelerometer linearity) are all indicative of standalone algorithm performance. The device is described as an "unattended data collection" system.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Heart Rate R-wave Detection (Arrhythmia Database): R-wave locations were "annotated R-wave locations" from the MIT-BIH arrhythmia database, which typically involves expert-reviewed annotations.
      • Heart Rate R-wave Detection (Bench Testing): "Expected Results (bpm)" based on the ANSV/AAMI EC 13 standard, implying predefined reference values for specific ECG waveforms.
      • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): "Automatically identified R-waves highlighted" implies comparison against captured ECG, but the true ground truth for accuracy calculation is not explicitly stated (e.g., manual expert annotation of raw ECG).
      • Accelerometer (Bench Validation): "Known acceleration applied" (physical inputs).
      • Accelerometer (Clinical Validation): "Expected features" of movement (e.g., walking), based on general physiological understanding of movement patterns.
    7. The sample size for the training set: Not specified. The studies describe validation testing but do not provide details on the training set used for developing the device's algorithms.

    8. How the ground truth for the training set was established: Not specified, as training set details are not provided.