510(k) Data Aggregation

    K Number
    K122744

    Validate with FDA (Live)

    Date Cleared
    2013-03-28

    (202 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

    Device Description

    PL009 and PL009A are electrically powered devices intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. It provides three selectable massage pulse styles with five auto operational modes through two channels with four attachment pads and shows graphic information about massage style, intensity and time remaining on a LCD based display which is incorporated with the device body. The device uses DC 6V power supplied by 4 AAA batteries. It is approximately 0.5lbs with dimensions 8" x 2" x 1" plus attachment pads. The only difference between PL009 and PL009A is the former has a button of "Repeat" representative of the function of repeat.

    PL029 is an electrically powered device intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. It provides eight selectable massage pulse styles with five auto operational modes through two channels with four attachment pads and shows textual message about massage style, intensity and time remaining on a LCD based display which is incorporated with the device body. The device uses DC 3V power supplied by 2 AAA batteries. It is approximately 0.5lb with dimensions 8" x 2" x 1" plus attachment pads.

    The differences between PL009 and PL029 include:

    (1) PL009's work voltage is 6V supplied by 4 AAA batteries, while PL029's work voltage is 3V supplied by 2 AAA batteries.

    (2) PL009 provides three operational modes, i.e., massage, beat and knead, which can be combined into six stimulation programs for WAIST, SHOULDER, JOINT, HAND-FOOT, REPEAT and SOLE, while PL029 provides eight operational, i.e., massage, beat, knead, slim, acupuncture, scraping, cupping and immune.

    (3) PL009 uses mechanical potentiometer to control power and intensity, while PL029 uses electronic switches to control power and intensity.

    (4) PL009's LCD display shows body parts in graphics when the device is in operation, while PL029's LCD display shows working modes in text only when the device is in operation.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary for the Prospera OTC TENS Electronic Pulse Massager, structured to address your specific questions.

    Important Note: The provided document is a 510(k) Summary, which is designed to demonstrate substantial equivalence to previously marketed devices, not necessarily to provide full clinical study reports with acceptance criteria and detailed performance metrics as one might find for a novel or high-risk device. The "study that proves the device meets the acceptance criteria" in this context refers to the non-clinical testing performed to show compliance with electrical safety and electromagnetic compatibility standards, and comparison to predicate devices, rather than a clinical trial evaluating pain relief efficacy with specific performance targets.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this device (a TENS unit), "acceptance criteria" are primarily established by compliance with recognized medical device standards for electrical safety, electromagnetic compatibility, and specific requirements for nerve and muscle stimulators. The "reported device performance" refers to the measurements taken during these non-clinical tests and the comparison of key technical characteristics with predicate devices.

    Parameter/Acceptance CriteriaReported Device Performance (Prospera OTC TENS)Basis for Acceptance (Standard/Predicate Comparison)
    Electrical Safety (IEC 60601-1)Compliance confirmed.Demonstrated by non-clinical testing against IEC 60601-1.
    Electromagnetic Compatibility (IEC 60601-1-2)Compliance confirmed.Demonstrated by non-clinical testing against IEC 60601-1-2.
    Safety of Nerve & Muscle Stimulators (IEC 60601-2-10)Compliance confirmed.Demonstrated by non-clinical testing against IEC 60601-2-10.
    Patient Leakage Current (Normal Condition)PL009: 1.8 µA; PL029: 2.0 µAWithin limits of recognized standards and comparable to predicate devices (K102598: 2.0 µA).
    Patient Leakage Current (Single Fault Condition)PL009: 3.3 µA; PL029: 3.0 µAWithin limits of recognized standards and comparable to predicate devices (K102598: 4.2 µA).
    Average DC Current through Electrodes (device on but no pulses)0 µA for all models.Meets expectation for TENS devices to prevent continuous DC current. Comparable to predicate devices (0 µA).
    General Construction and MaterialsEnclosure: ABSComparable to predicate devices (ABS, silicone).
    Software VerificationCarried out according to FDA guidance.Compliance with FDA Guidance for the Content of Premarket Submission for Software Contained in Medical Devices.
    Electrodes Safety RequirementsMeet safety requirements.Mentioned as compliant, specific standard not detailed in summary.
    Intended Use & Operational Principle"Temporary relief of pain associated with sore and aching muscles..."Substantially equivalent to predicate devices with similar indications and operational principles (generating electrical pulses through electrodes to activate nerves for pain relief).
    Technical Characteristics (e.g., Output Voltage, Current, Pulse Duration, Frequency)Values presented in Tables B and C (e.g., Max Output Voltage at 500Ω: PL009: 12.8V, PL029: 49.6V; Pulse Duration: PL009: 120-6800 µSec, PL029: 100-200 µSec)These values are compared to predicate devices (e.g., K102598: Max Output Voltage at 500Ω 62.4V, Pulse Duration 16.3-781mS). The argument is for substantial equivalence despite numerical differences, based on the overall effect and safety profile being similar. The summary states "The design differences... are insignificant and do neither affect the intended use, nor alter the operational principle."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a "test set" in the context of clinical data. For non-clinical (engineering/bench) testing, the sample size typically involves a small number of devices (often 1-3 units) tested against standards. The document does not specify the number of devices tested for compliance with the enumerated standards.
    • Data Provenance: The data is generated from non-clinical tests performed on the subject device. This refers to bench testing in a laboratory setting to evaluate electrical safety, EMC, and performance characteristics. There is no mention of country of origin for the non-clinical test data, but it would typically be conducted by a certified testing lab. The study is not a retrospective or prospective clinical study but rather a technical verification.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to this 510(k) submission.

    • Nature of the Device: The Prospera OTC TENS Electronic Pulse Massager is a hardware device for pain relief, and its approval relies on demonstrating electrical safety, functional performance, and substantial equivalence to existing predicate devices, not on the output of an algorithm or diagnosis by experts.
    • "Ground Truth": For this type of device, "ground truth" relates to the objective measurements of electrical parameters, safety compliance with standards, and functional specifications, not expert consensus on medical diagnoses or interpretations. These are typically established by objective physical measurements and adherence to engineering standards.

    4. Adjudication Method for the Test Set

    This information is not applicable to this 510(k) submission. Adjudication methods like 2+1 or 3+1 are used in clinical trials, especially when evaluating subjective endpoints or interpretations (e.g., radiological reads), to resolve discrepancies or ensure consistency in expert assessment. This submission relies on objective physical measurements and comparisons to predicate device specifications.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done or reported in this 510(k) summary.

    • Device Type: This is a Transcutaneous Electrical Nerve Stimulator (TENS), which is a direct-to-patient therapy device. It does not involve human readers interpreting output or an AI component that assists human readers in diagnosis or treatment planning.
    • Purpose of 510(k): The 510(k) pathway focuses on demonstrating substantial equivalence, primarily through non-clinical testing for devices like TENS units. Clinical studies of efficacy (e.g., "how much human readers improve with AI vs without AI assistance") are typically not required for such devices unless there are significant technological differences or new indications for use that cannot be supported by equivalence.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    No, a standalone "algorithm only" performance study was not done or reported in this 510(k) summary.

    • Nature of the Device: The Prospera OTC TENS is a hardware device with embedded software for controlling its electrical pulses and modes. It is not an "algorithm" in the sense of a diagnostic or predictive AI tool, and its function is not meant to be separated from the human user's interaction (e.g., selecting modes and intensity) for pain relief. Software verification was indeed performed, but this is to ensure the control software functions correctly and safely, not to evaluate its performance as a standalone interpretative algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Compliance with recognized electrical safety and EMC standards: This involves objective measurements against established thresholds defined in standards like IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10.
    • Comparison to predicate device specifications: The technical characteristics (e.g., voltage, current, frequency, pulse duration) of the subject device are measured and compared to those of legally marketed predicate devices to establish substantial equivalence.
    • Functional verification: Ensuring that the device operates as described (e.g., selection of massage styles, intensity, timer function).

    There is no mention of pathology, expert consensus, or outcomes data being used as ground truth for this particular 510(k) submission because it's a TENS device seeking approval via substantial equivalence based on technical and safety characteristics.


    8. Sample Size for the Training Set

    This information is not applicable to this 510(k) submission.

    • No AI/Machine Learning Training: The device does not employ machine learning or AI models that require a "training set." The embedded software controls the device's pre-programmed electrical pulse patterns and operational modes.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set for an AI/ML model for this device. The "ground truth" for the device's functionality is its design specifications and compliance with established engineering and medical device safety standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113544

    Validate with FDA (Live)

    Date Cleared
    2012-07-27

    (239 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Symptomatic relief of acute post traumatic pain, acute post surgical pain, and chronic intractable pain.
    Prescription Use

    Device Description

    The Electronic Pulse Massager, which includes Models PL009, PL009A, and PL029, is a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief. TENS can help to relieve the pain caused by trauma or continual strain, and is suitable to be used on the user's back, neck, shoulders, legs, arms, and feet. The proposed Electronic Pulse Massager, which is compact, portable, and microprocessor-controlled, delivers a gentle electrical pulse to the underlying nerve through the wires and electrode pads placed on the skin of users. According to the need of users, the pulse intensity can be adjustable on the front control panel of the massager. Both Models PL009 and PL009A have five automatic stimulation programs and three manual programs, while Massager PL029 contains five automatic stimulation programs and eight selectable programs. The only difference between Models PL009 and PL009A is the former has a "REPEAT" program function and the latter does not. All these programs as well as the pulse intensity and time remaining can be displayed on the liquid crystal display screen of both massagers.

    AI/ML Overview

    The provided K113544 510(k) Summary for the Electronic Pulse Massager, Models PL009, PL009A, and PL029, does not describe the use of a study with acceptance criteria and reported device performance in the typical sense of a clinical trial or algorithm performance evaluation.

    This submission is for a Transcutaneous Electrical Nerve Stimulator (TENS) device, which is a medical device and not an AI/ML algorithm. Therefore, the questions related to AI/ML specific aspects like training sets, test sets, ground truth establishment by experts, adjudication methods, and MRMC studies are not applicable.

    Instead, the submission demonstrates substantial equivalence to a predicate device (TENS 3000 Stimulator, K102014) through a comparison of technological characteristics and compliance with a set of voluntary non-clinical design standards.

    Here's an breakdown based on the information provided and addressing the nature of the device:

    1. A table of acceptance criteria and the reported device performance

    For this type of device (TENS), "acceptance criteria" and "reported device performance" are primarily demonstrated through compliance with recognized standards and comparison to a predicate device. The submission does not provide explicit numerical "performance" metrics/acceptance criteria like sensitivity/specificity for disease detection, but rather shows that its properties and functionalities meet expected standards for safety and effectiveness for a TENS device.

    Acceptance Criteria (Demonstrated through Standard Compliance & Predicate Comparison)Reported Device Performance (as demonstrated by compliance and comparison)
    Safety:
    Electromagnetic compatibility (Emissions)Complies with EN55014-1
    Electromagnetic compatibility (Immunity)Complies with EN55014-2
    Household electrical appliances safety (General requirements)Complies with EN60335-1
    Safety for massage appliances (Particular requirements)Complies with EN60335-2
    Measurement methods for electromagnetic fields (Human exposure)Complies with EN62233
    Medical electrical equipment safety (General requirements)Complies with IEC 60601-1
    Medical electrical equipment safety (Collateral Standard)Complies with IEC 60601-1-2
    Medical electrical equipment safety (Nerve and muscle stimulators)Complies with IEC 60601-2-10
    Compliance with 21 CFR 898 (presumably related to performance standards for diagnostic X-ray systems, though its relevance here for a TENS is unclear without further context, but claimed compliance)Complies with 21 CFR 898 (Yes)
    Effectiveness/Functionality:
    Intended UseSymptomatic relief of acute post traumatic pain, acute post surgical pain, and chronic intractable pain (Same as predicate)
    Power Source6V Battery (PL009/PL009A), 3V Battery (PL029) - Different from predicate (9V), but functional
    Output Channels2 (Same as predicate)
    Channel IsolationBy Enclosure (Same as predicate)
    Software/Firmware/Microprocessor ControlYes (Same as predicate)
    Automatic Overload/Over Current/No-Load TripNo (Same as predicate)
    Automatic Shut OffNo (Same as predicate)
    Patient Override ControlNo (Same as predicate)
    Indicator DisplayYes (Same as predicate)
    Timer Range15 min, continue (Same as predicate)
    WaveformBiphasic or Monophasic Rectangular Pulse (Same as predicate)
    Frequency2-60 Hz - Different from predicate (2-150 Hz), but within range for TENS
    Weight150g (PL009/PL009A), 125g (PL029) - Different from predicate (115g), but acceptable design difference
    Dimensions (HxWxT)200x55x33mm (PL009/PL009A), 203x63x13mm (PL029) - Different from predicate (95x65x23.5mm), but acceptable design difference
    Housing Materials & ConstructionEnclosure: ABS Plastic (Same as predicate)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This device is cleared through substantial equivalence based on non-clinical testing and comparison to a predicate, not clinical data or performance on a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for a test set is not established for this type of device clearance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For this device, "ground truth" is conceptually represented by the established safety and performance standards for TENS devices and the characteristics of the legally marketed predicate device.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.