510(k) Data Aggregation

    K Number
    K992424

    Validate with FDA (Live)

    Date Cleared
    2000-12-29

    (527 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Plastimed's Seldiflex® central venous catheter is intended for long-term vascular access and is indicated for use with patients who require central venous pressure monitoring, I.V. fluids, blood products, drugs, or parenteral nutrition solutions.

    Device Description

    The Seldiflex® is a single lumen, radiopaque polyurethane, long-term central venous catheter. Its accessories include a protective adapter female cap, introducer needle, needle obturator, double-ended J guidewire, scalpel, catheter clamp, rigid fastener, a vessel dilator, and silver Dacron® cuff.

    AI/ML Overview

    The provided 510(k) summary for the Seldiflex® central venous catheter describes a device that is exempt from a clinical study to prove acceptance criteria because it focuses on demonstrating substantial equivalence to predicate devices through functional and safety testing, rather than an AI/ML-driven device requiring performance metrics. Therefore, many of the requested categories related to AI/ML device performance studies are not applicable to this submission.

    Here's the information that can be extracted or deduced from the provided document:

    Acceptance Criteria and Device Performance (Not Applicable for AI/ML performance)

    This submission focuses on demonstrating substantial equivalence to predicate devices based on functional design, materials, and indications for use, rather than meeting specific quantifiable performance metrics typically associated with AI/ML devices. The "acceptance criteria" are implied by successful functional and safety testing and similarity to predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Functional operation under normal usage conditions"The results of the examination and testing were successful."
    Conformance to product specifications"The results of the examination and testing were successful."
    No new issues of safety and effectiveness compared to predicates"did not raise any issues of safety and effectiveness."
    Substantial equivalence to predicate catheters"substantially equivalent to the Hickman/Broviac central venous catheters..." and "ONECATH® long-term intravascular catheters..."

    Study Information (Relevant to the provided document)

    Since this is for a medical device (central venous catheter) that is not an AI/ML device, many of the requested study details (sample size for test set, data provenance, ground truth experts, adjudication, MRMC study, standalone performance, training set details) are not applicable in the context of this 510(k) submission.

    1. Sample size used for the test set and the data provenance: Not specified. The document states "Functional and safety testing... consisted of examination of the function of the device under conditions similar to those found in normal usage and testing to ensure conformance to product specifications." This implies in-house testing, likely with a sample size determined by internal quality standards for mechanical and material properties, rather than a clinical trial. Data provenance is not described.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For a central venous catheter, "ground truth" would relate to its physical properties, biocompatibility, and functional integrity, which are assessed through engineering and material science testing, not typically by clinical experts establishing ground truth in the way it's done for diagnostic AI/ML.

    3. Adjudication method for the test set: Not applicable. The testing described (functional and safety) would involve objective measurements and observations against pre-defined specifications without the need for an adjudication process.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device, so no MRMC study was conducted.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.

    6. The type of ground truth used: For this type of device, the "ground truth" for the functional and safety testing would be based on engineering specifications, material standards, and established performance benchmarks for similar medical devices. For example, testing for tensile strength (material property), flow rates (functional property), or biocompatibility (safety).

    7. The sample size for the training set: Not applicable. There is no AI/ML component, therefore no training set.

    8. How the ground truth for the training set was established: Not applicable. There is no AI/ML component or training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974642

    Validate with FDA (Live)

    Date Cleared
    1998-03-13

    (88 days)

    Product Code
    Regulation Number
    874.4680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The COMBICATH Catheter Model 58216.27 is intended for use as a sampling tool by performing mini bronchoalvestar lavage in mechanically ventilated patients to obtain bronchoalveolar samples for laboratory testing and diagnosis.

    The COMBICATH Catheter Models 58223.19, 58228.19, and 58229.19 are intended for use as sampling tools to capture protected distal bronchoalveolar samples from mechanically ventilated patients for laboratory testing and diagnosis.

    Device Description

    The COMBICATH Catheter, is a radiopaque, double plugged, telescoping catheter designed to capture bronchoalveolar secretion samples in mechanically ventilated patients.

    AI/ML Overview

    This 510(k) submission (K974642) is for a medical device called the COMBICATH Catheter, which is a physical device for sampling bronchoalveolar secretions, not an AI or software-driven diagnostic tool. Therefore, many of the requested categories related to AI performance metrics, ground truth establishment, expert adjudication, and comparative effectiveness studies are not applicable.

    The performance described is purely related to functional and safety testing of a physical device.

    Here's the relevant information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What was tested)Reported Device Performance (Results)
    Functionality of the device under conditions similar to normal usageSuccessful
    Conformance to product specificationsSuccessful
    Safety of the deviceNo issues raised regarding safety
    Effectiveness of the deviceNo issues raised regarding effectiveness

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a quantitative "sample size" in terms of number of catheters or patients for the functional and safety testing. It mentions "examination of the function of the device under conditions similar to those found in normal usage and testing to ensure conformance to product specifications." The data provenance is not explicitly stated but implies testing conducted by the manufacturer, Plastimed Laboratoire Pharmaceutique, as part of their product development and regulatory submission. It is likely prospective testing related to device validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a physical device. Ground truth as typically understood for diagnostic AI (e.g., presence or absence of a disease) is not relevant here. The "ground truth" during testing would be whether the device physically functioned as intended and met specifications.

    4. Adjudication Method for the Test Set

    Not applicable. This is a physical device. Adjudication methods are typically used for evaluating human or AI interpretations of data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study is focused on the functional and safety performance of a physical sampling catheter, not on the interpretation of medical images or data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's testing was its functional performance and adherence to engineering specifications. This includes successful capture of bronchoalveolar secretion samples, material compatibility, and overall device integrity during simulated use, as well as meeting manufacturing and design specifications.

    8. The Sample Size for the Training Set

    Not applicable. This device does not use a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.