    Intended Use

    The Planer BT37 Incubator is intended to be used to provide an environment with controlled temperature at or near body temperature, CO2, O2 and N2 gases, and elevated humidity for the development of gametes and embryos during in vitro fertilization (IVF) / assisted reproductive technology (ART) treatments.

    Device Description

    The BT37 Incubator has been designed as a bench-top incubator to provide a electronically controlled environment (temperature and humidified-gas (premixed gas; typically 6% CO2, 5% O2, balance N2)) for use in cell culture as part of In Vitro Fertilization (IVF) / Assisted Reproductive Technology (ART) treatments. The BT37 Incubator overall dimensions are 420mm wide x 270mm deep x 210mm high, with a weight of 15,5 kg. The BT37 Incubator has two chambers each of which have been designed to hold a variety of dishes. In case of external power failure, the BT37 incubator is equipped with an internal backup battery to maintain functionality for up to 120 minutes. The BT37 Incubator provides an environment with elevated humidity by bubbling the incoming gas through a heated humidification bottle containing water. This gas this then passed to the left and right incubation chambers. In its default configuration, the BT37 normally provides gas at a constant flow rate of 30 mL/min. After closing the lids, an increased gas flow at 360 ml./min is provided for 3 minutes to purge the systems. After fitting a new bottle, a triple length purge of 360 ml /min for 9 minutes is provided. An alternative pulsed flow regime is available, where the normally constant flow of 30 mL/min can be replaced by a flow that pulses between a high and low rate. The default for these settings are normally 60 mL/min and 20 mL/min respectively with a time ratio designed to provide a mean flow of 30 mL/min. This flow approach is not normally used.

    AI/ML Overview

    The Planer BT37 Incubator is a bench-top incubator designed for use in In Vitro Fertilization (IVF) / Assisted Reproductive Technology (ART) treatments. It provides an electronically controlled environment (temperature and humidified-gas) for cell culture.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Function / ParameterAcceptance Criteria (Predicate or BT37)Reported Device Performance (BT37)
    Indication for UseStorage and preservation of gametes and/or embryos at or near body temperature.Provides an environment with controlled temperature, CO2, O2, N2 gases, and elevated humidity for the development of gametes and embryos during IVF/ART.
    Gas Supply BlendSimilar (Implicitly 6% CO2, 5% O2, 89% N2)6% CO2, 5% O2, 89% N2
    Gas Supply Pressure150 kPa +/- 15 kPa150 kPa +/- 15 kPa
    Gas Flow Rate Capability (per side)15 mL/min to 25 mL/min (normal)0 mL/min to 450 mL/min (fully adjustable)
    Gas Flow Rate Accuracy (normal flow)±15% of flow per chamber±10% of flow per chamber
    Gas Flow PatternNon-pulsed (constant)Pulsed or non-pulsed (constant)
    Chamber Temperature Capability35.0 °C to 40.0°C (with specific ambient temp ranges)(ambient + 5 °C) to (ambient + 20 °C), not to exceed 40 °C
    Chamber Temperature Accuracy±0.2 °C at calibration point±0.2 °C at calibration point
    Humidification SystemSimilar (Preheated bottle, gas bubbled through water)Preheated bottle, gas bubbled through water prior to chambers
    Accelerated Ageing (Humidification Bottle)Not explicitly stated (predicate)Justified shelf-life of 29 months at room temperature
    Software FunctionalityNot explicitly stated (predicate)All functions operated correctly with no significant anomalies
    Electrical SafetyCompliance with relevant standards (Implicitly)EN61010-1:2001 and BS EN 61010-2-010
    Electromagnetic Compatibility (EMC)Compliance with relevant standards (Implicitly)EN 61236-1:2006

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a separate "test set" in the context of clinical or evaluation studies with human data. The "tests" described are engineering and performance validation tests of the device itself.

    • Sample Size for performance tests: Not explicitly stated, as these are likely repeat measurements on the device itself.
    • Data Provenance: The tests were conducted internally by Planer plc and, for electrical safety and EMC, by accredited testing bodies. The testing is prospective in nature, as it's directly evaluating the performance of the BT37 Incubator. The country of origin of the data is United Kingdom (where Planer plc is located).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The "ground truth" for this device's performance tests would be established by validated scientific and engineering principles, and measurement standards, rather than expert consensus on medical images or diagnoses. No human experts were used to establish ground truth for a clinical "test set" in the context of the provided text.

    4. Adjudication Method for the Test Set

    Not applicable. There was no "adjudication" necessary as the tests were objective performance measurements against defined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study is not relevant for this type of device (an incubator). Such studies are typically performed for diagnostic devices (e.g., AI for medical imaging) to assess how human readers perform with and without AI assistance on a set of clinical cases.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence. The "tests and test results" section describes the standalone performance of the BT37 Incubator in achieving its specified operational parameters for temperature, humidity, gas concentration, and flow. These tests evaluated the device's inherent capabilities without human intervention or judgment as part of the performance measurement.

    7. The Type of Ground Truth Used

    The ground truth used for the performance tests was based on:

    • Engineering specifications and design requirements: The device was tested against its own design parameters for temperature control, gas flow, humidification, etc.
    • Industry and regulatory standards: Electrical safety (EN61010-1:2001, BS EN 61010-2-010) and EMC (EN 61236-1:2006) tests confirm compliance with established external benchmarks.
    • Physical measurements: Using calibrated equipment to measure temperature, gas flow, humidity, etc., provides objective ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for the device's control system would be part of its engineering design and calibration, not a data-driven learning process.

    9. How the Ground Truth for the Training Set Was Established

    As above, this concept is not applicable here as it refers to a device which is not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032086

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-19

    (256 days)

    Product Code
    Regulation Number
    884.6120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kryo ART brand of controlled rate freezers are intended to be used to freeze gametes or embryos at a user determined rate.

    Device Description

    The Kryo ART range of controlled rate freezers have been specifically designed and configured for use in the assisted reproduction field. The Kryo ART range of controlled rate freezers are available in three chamber sizes depending on the number of samples to be frozen; 1.7 liters, 3.3 liters and 16 liters. All models incorporate the critical features expected from a high specification biological chamber and the -180 degrees C to +40 degrees C temperature range allows flexibility for a wide range of applications and protocols. All of the chamber sizes are controlled by the MRV controller system, which has been created to offer multiple protocols while remaining simple to program and operate. All systems include power failure protection, which allows the controller to continue monitoring during mains power failures. All software is designed, developed and maintained in accordance with the Company's internal procedures. These procedures form part of the full Quality System, which is approved to ISO 9001:1994, ISO 13485:1996 and EN 46001:1996. Performance testing is complete. Units were tested according to a specific protocol. Testing demonstrates the functional suitability of the Kryo ART Controlled Rate Freezer for their intended purpose. The Kryo ART brand of controlled rate freezes are intended to freeze gametes or embryos at a user determined rate.

    AI/ML Overview

    The provided text is a 510(k) summary for the Kryo ART Controlled Rate Freezer. It outlines the device's classification, intended use, and substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria, detailed study designs, or performance data beyond a general statement that "Performance testing is complete. Units were tested according to a specific protocol. Testing demonstrates the functional suitability of the Kryo ART Controlled Rate Freezer for their intended purpose."

    Therefore, I cannot provide the requested information in the format specified because the document does not contain that level of detail.

    The document indicates that the device met the requirements for a finding of substantial equivalence to the predicate device K021042, CryoMed IVF Controlled Rate Freezer, marketed by Thermo Forma, Inc. The basis for this equivalence is stated as "basic design, materials, intended use and performance characteristics."

    Without further documentation, it is not possible to describe:

    1. A table of acceptance criteria and the reported device performance: This information is not present.
    2. Sample size used for the test set and the data provenance: Not specified.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified.
    4. Adjudication method: Not specified.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not specified, and unlikely for this type of device.
    6. If a standalone performance study was done: The document states "Performance testing is complete. Units were tested according to a specific protocol," implying standalone testing, but no details are provided.
    7. The type of ground truth used: Not specified.
    8. The sample size for the training set: Not applicable as this is a physical device, not an AI/ML algorithm.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1