510(k) Data Aggregation

    K Number
    K131769

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-07

    (112 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CinchLock Knotless Suture Anchor and the NanoTack Suture Anchor 1.4mm are intended for the fixation of soft tissue to bone in the hip, shoulder, foot/ankle, hand/wrist, elbow, and knee. Please see individual indications for use statements.

    The NanoTack® Suture Anchor, previously cleared under 510(k) K110473 and K130351, is intended for fixation of soft tissue to bone in the hip, shoulder, foot/ankle, hand/wrist, elbow, and knee, in the following procedures: Hip capsule repair, Acetabular labrum reattachment, Capsular stabilization (Bankart repair, Anterior shoulder instability, SLAP lesion repairs, Capsular shift or capsulolabral reconstructions), Acromilavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs, Biceps tenodesis, Hallux valgus repairs, Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstructions, Midfoot reconstructions, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy, Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Extra-capsular repairs (Medial collateral ligament, Lateral collateral ligament, Posterior oblique ligament), Patellar realignment and tendon repairs (Vastus medialis obliquous advancement), Iliotibial band tenodesis.

    The CinchLock™ Knotless Anchor, previously cleared under 510(k) K123651, is intended for fixation of soft tissue to bone in the hip, shoulder, foot/ankle, hand/wrist, elbow, and knee, in the following procedures: Hip capsule repair, Acetabular labrum reattachment, Capsular stabilization, Bankart repair, Anterior shoulder Instability, SLAP lesion repairs, Capsular shift or capsulolabral reconstructions, Acromilavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs, Biceps tenodesis, Hallux valgus repairs, Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstructions, Midfoot reconstructions, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy, Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Extra-capsular repairs, Medial collateral ligament, Lateral collateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs, Vastus medialis obliquous advancement, Iliotibial band tenodesis.

    Device Description

    The Pivot Suture Anchors which are part of this expanded indications submission are as follows: Pivot CinchLock Knotless Suture Anchor, Pivot NanoTack Suture Anchor 1.4mm. These anchors are non-degradable suture anchors manufactured from PEEK-OPTIMA® LT1 polymer. The CinchLock and NanoTack Anchors are attached to / pre-assembled to a stainless steel Inserter. Non-degradable ultra-high molecular weight polyethylene (UHMWPE) blue co-braid #1 suture is provided in the sterile package with the CinchLock Knotless Suture Anchor and Inserter. The Pivot NanoTack Suture Anchor incorporates a non-degradable ultra high molecular weight polyethylene (UHMWPE) blue co-braid #1 suture. The CinchLock Knotless Suture Anchor and NanoTack Suture Anchor 1.4mm devices are provided as a single use sterile devices.

    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed information required to describe the acceptance criteria and the study proving the device meets those criteria, as it focuses on a medical device clearance (510(k) for suture anchors) rather than an AI/ML device.

    Specifically, the document discusses:

    • Device Names: CinchLock Knotless Suture Anchor, NanoTack Suture Anchor 1.4mm
    • Classification: Bone Anchor, Class II, Product Code MBI
    • Predicate Device: Smith and Nephew BioRaptor 2.3PK Suture Anchor - K071586
    • Device Description: Non-degradable suture anchors made from PEEK-OPTIMA® LT1 polymer, pre-assembled to a stainless steel Inserter, with UHMWPE suture.
    • Intended Use: Fixation of soft tissue to bone in various anatomical regions (hip, shoulder, foot/ankle, hand/wrist, elbow, and knee) for specified procedures.
    • Technological Characteristics: Identical to previous submissions; the current submission is for expanded indications.
    • Performance Testing: Mentions "insertion strength, anchor strength, suture strength, and biocompatibility testing" demonstrating substantial equivalence to the predicate device.
    • Substantial Equivalence: Claimed based on indications for use, technological characteristics, and comparison to the predicate device.
    • FDA Clearance: A letter from the FDA confirming substantial equivalence and providing clearance.
    • Expanded Indications for Use: Detailed lists of procedures for the NanoTack and CinchLock anchors across various joints.

    This information pertains to a traditional medical device (suture anchors) and its mechanical and biological performance, not an AI/ML diagnostic or prognostic device. Therefore, the questions regarding acceptance criteria in terms of performance metrics like sensitivity, specificity, AUC, sample sizes for test/training sets, ground truth establishment, expert consensus, MRMC studies, or standalone algorithm performance are not applicable to this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130351

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-14

    (30 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NanoTack Suture Anchor is intended for the fixation of soft tissue to bone in the hip, and is indicated for the reattachment of hip labrum to the acetabulum.

    Device Description

    The Pivot NanoTack Suture Anchor 1.4mm is a non-degradable suture anchor manufactured from PEEK-OPTIMA® polymer attached to a non-degradable ultra high molecular weight polyethylene (UHMWPE) blue co-braid #1 suture that is pre-assembled to a Flex stainless steel Inserter. The NanoTack Suture Anchor 1.4mm with Flex Inserter is provided as a single use sterile device.

    AI/ML Overview

    The Pivot Medical NanoTack Suture Anchor 1.4mm with Flex Inserter is a medical device designed for the fixation of soft tissue to bone in the hip, specifically for the reattachment of the hip labrum to the acetabulum. The device is a "Special 510(k)" submission, meaning it is substantially equivalent to a previously cleared device (NanoTack Suture Anchor 1.4mm - K110473) with minor modifications.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Insertion StrengthSubstantially equivalent to the predicate device (NanoTack Suture Anchor 1.4mm)
    Anchor StrengthSubstantially equivalent to the predicate device (NanoTack Suture Anchor 1.4mm)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states that "Pre-clinical testing includes insertion strength and anchor strength." However, it does not specify the sample size used for these tests. The data provenance is also not explicitly mentioned, but "pre-clinical testing" typically implies laboratory or cadaveric studies, not human clinical trials, and would likely be conducted in the country of origin of the manufacturer (USA, given the FDA submission). It is a prospective study as it evaluates the device performance for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The testing described (insertion strength and anchor strength) are biomechanical performance tests, not subjective evaluations requiring expert clinicians to establish ground truth in the same way as, for example, image interpretation. Therefore, "ground truth" would be established by the objective measurements obtained during the mechanical testing.

    4. Adjudication Method for the Test Set:

    This information is not applicable and is not provided. The performance testing involves objective measurements rather than subjective assessments that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret images or data, and the AI's impact on their performance is evaluated. The NanoTack Suture Anchor is a surgical implant, and its evaluation focuses on mechanical performance, not diagnostic accuracy.

    6. Standalone (Algorithm Only) Performance Study:

    No, a standalone (algorithm only) performance study was not done. The NanoTack Suture Anchor is a physical medical device, not an algorithm. Its performance is assessed through mechanical and material testing.

    7. Type of Ground Truth Used:

    The ground truth for the performance testing (insertion strength and anchor strength) was established through objective mechanical and material property measurements. For example, force gauges would be used to measure insertion force, and load cells would measure anchor pull-out strength.

    8. Sample Size for the Training Set:

    This information is not applicable and is not provided. The NanoTack Suture Anchor is not an AI/ML device that requires a training set. Its design and performance are based on engineering principles and materials science.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device