Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K043090

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-12-23

    (45 days)

    Product Code
    Regulation Number
    874.3320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing or hearing impairments. The device is used with an associated transmitter microphone and speaker systems and does not provide coupling to the ear through either earphones or earmolds.

    These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTF hearing aids and/or personal FM devices. The talker's voice with or which bI E throughout the room to assist listeners, regardless of seating location, to consistently hear what the talker is saying.

    Device Description

    Sound Field group amplification systems consist of FM or IR transmitters and microphone used by a speaker/instructor to send auditory signals to FM or IR receiver/amplifier which noreesses the voice signals and broadcast the signals through one or more loudspeakers. These systems are generally installed under the guidance of an audiologist or a sound contractor. Sound Field systems consist of an on/off switch and volume controls. Specifications and user guider in cre provided for Sound Field installation and use.

    Sound Field systems are assembled from standard electronic components widely used by Group Hearing Aid and Group Auditory Trainer manufacturers. Sound Field wireless ampification systems amplify and broadcast the speaker/instructor's voice through strategically-mnounted.will and/or ceiling loud speakers placed throughout the listening area to distribute the final moder want signal (sound) of the speaker/instructor and are not directly connected or coupled to the piroener.

    Sound Field wireless amplification systems are intended for use as group auditory trainers used to communicate simultaneously with one or more listeners with normal hearing on he aring impairments. The device is used with an associated transmitter microphone and speakers systems and does not provide coupling to the ear through either earphones or earmolds.

    These systems use FM and/or infrared wireless technology which include receivers, transmitters, and strategic placement of high quality speaker systems and can be used with or without BTE hearing aids and/or personal FM devices. The talker's voice is mildly amplified and dispersed throughout the room to assist listeners, regardless of seating location, to consistently bear what the talker is saying.

    AI/ML Overview

    The provided text K043090 includes a 510(k) summary for Phonic Ear Sound Field Group Amplification Systems. However, it does not contain a study that proves the device meets specific acceptance criteria through performance metrics or statistical analysis.

    Instead, the document states:

    • "Many case studies, experiments, and trials exist that document the benefits and effectiveness of speech perception from Sound Field amplification." Crucially, it does not provide details of these specific studies that were conducted for this device to establish acceptance criteria and performance against them.
    • "Field studies were conducted in accordance with ANSI S12.60 Acoustical Performance Criteria, Design Requirements, and Guidelines for Schools." Again, specific results, acceptance criteria, or performance data from these particular field studies are not included.
    • "Test procedures to obtain FCC (Federal Communications Commission) approval are in accordance with good engineering practices and in accordance with FCC Rules, Part 15, Subpart A, 15, 19 (a) (3) and (C) Code of Federal Regulations (CFR 47)." This refers to regulatory compliance for radio frequency emissions, not clinical performance or efficacy.

    Therefore, based solely on the provided text, it is not possible to complete most of the requested sections as the necessary information (specific acceptance criteria, detailed study results, sample sizes for test/training, ground truth establishment, expert qualifications, etc.) is absent. The submission relies on establishing substantial equivalence to predicate devices (K974287) and generally referencing existing literature and compliance with technical standards (ANSI, FCC) rather than presenting a novel performance study with defined acceptance criteria for the current device.

    Here's a breakdown of what can and cannot be answered from the provided text:


    Description of Acceptance Criteria and Device Performance Study (Based on Provided Text)

    Unfortunately, the provided 510(k) summary for the Phonic Ear Sound Field Group Amplification Systems does not include specific acceptance criteria or a detailed study that measures the device's performance against such criteria. The submission focuses on establishing substantial equivalence to predicate devices and adherence to existing standards (ANSI S12.60 for acoustics and FCC for radio frequency emissions).

    The document mentions that "Many case studies, experiments, and trials exist that document the benefits and effectiveness of speech perception from Sound Field amplification," and "Field studies were conducted in accordance with ANSI S12.60." However, it does not present the results of these or any new studies conducted specifically for this device against defined performance acceptance criteria.

    Therefore, a table of acceptance criteria and reported device performance cannot be created from this document. The following sections are addressed where information is available, and marked as "Not provided in the text" otherwise.


    1. A table of acceptance criteria and the reported device performance

      Acceptance CriteriaReported Device Performance
      Not provided in the textNot provided in the text
      • The document implies adherence to "ANSI S12.60 Acoustical Performance Criteria, Design Requirements, and Guidelines for Schools" and FCC regulations, but it does not specify quantitative acceptance criteria derived from these standards for the device, nor does it provide performance data to demonstrate compliance.
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      • Sample Size (Test Set): Not provided in the text.
      • Data Provenance: Not provided in the text. The document refers generally to "Field studies were conducted" but gives no details about the participants, methodology, or geographical origin.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      • Not applicable/Not provided in the text. The type of study described (adherence to acoustical standards, and general benefit of sound field amplification) does not typically involve expert "ground truth establishment" in the same way an AI diagnostic device would.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      • Not applicable/Not provided in the text. No details of a test set and associated adjudication method are present.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This device is a sound amplification system, not an AI diagnostic tool designed to assist human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this device's function and was not performed.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

      • Not applicable. This device is an auditory trainer and amplifier, not a standalone algorithm. Its performance is inherently tied to its physical components and how it processes and broadcasts sound for human listeners.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • Not provided in the text. For a sound amplification system, "ground truth" would likely relate to objective acoustical measurements (e.g., sound pressure levels, speech intelligibility index, signal-to-noise ratio) and subjective listener perception. However, no specific ground truth methodology or data is detailed in the submission for this device's performance.
    8. The sample size for the training set

      • Not applicable. This device is a hardware-based amplification system, not a machine learning algorithm requiring a "training set."
    9. How the ground truth for the training set was established

      • Not applicable. As above, this device does not use a training set for machine learning.
    Ask a Question

    Ask a specific question about this device

    K Number
    K974658

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-02-25

    (72 days)

    Product Code
    Regulation Number
    874.3320
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Model PE 600 Series devices form the Impaired Group Amplification System which is used by adults and children in public halls, theaters, churches, courtrooms and other public gatherings and provide auditory assistance to hearing impaired persons.

    Device Description

    The Model PE 600 Series consists of:

    1. The Model PE 600E is a power amplifier which is combined with an infrared LED array. Audio signals from a sound system are processed and broadcast into the listening area by means of infrared light.
    2. The Model PE 600R is a body worn infrared receiver. It receives the infrared light signals from the PE 600E and converts these signals to high level audio signals required by the hearing impaired listener. The audio signals can be adjusted to meet the user.
    3. The PE 601R is an Infrared Headset Receiver. It receives the infrared light signals from the PE 600E and converts these signals to audio signals required by the hearing impaired listener. This devise is capable of stereo reception so that different audio signals can be received simultaneously and fed to each ear separately.
    4. The PE 602R is an Infrared Headset Receiver. It receives the infrared light signals from the PE 600E and converts these signals to audio signals required by the hearing impaired listener. The device is capable of receiving infrared light signals transmitted on two separate frequencies. A switch on the headset allows the user to select the program frequency desired. The selected program is heard in each ear.
    AI/ML Overview

    The provided text is a 510(k) summary for the Phonic Ear Model PE600 Series Infrared Group Amplification System. This device is an auditory assistance system for hearing-impaired individuals in various public settings.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary primarily focuses on demonstrating substantial equivalence to previously approved devices of the same type, rather than setting out specific numerical acceptance criteria. The information provided is descriptive of the system's capabilities and its equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety:
    Equivalent to approved systems of the same type."The safety of this Infrared Amplification System is equivalent to approved systems of the same type."
    Moderate sound pressure levels, adjustable by professional."While moderate sound pressure levels are produced by the PE600 Series receivers, this output can be limited to the output required by the user. Limiting can be accomplished by the setting of the volume controls in accordance with the tests made by a clinician, audiologist or hearing professional prior to use by the hearing impaired user."
    User instructions for malfunction."Instructions for the user to follow, should the system not function as expected, are contained in the user information."
    Technical specifications provided."Complete technical specifications are contained in the user information."
    Distributed to qualified personnel."The systems are distributed directly to qualified dispensers or others who are trained in the proper fitting and use of the system."
    Material safety (e.g., lack of side effects from casing)."The system elements are housed in cases made of cycolac plastic which has few known side effects caused by the material."
    Accessories description provided."A complete line of accessories to be used with the system elements are available from the manufacturer and a list of these accessories and a description of their use is contained in the user information."
    Service information available."Addresses and telephone numbers of authorized service centers are available to the user should any questions arise about the function of the system."
    Detailed use, maintenance, adjustment, and function descriptions."Detailed descriptions on the use, maintenance, adjustment and function of each element of the system are contained in the user information."
    Low voltage, battery compartment security, fusing."The system elements are powered by low voltage batteries which are contained within battery compartments that are fitted with tight snap fitting covers to prevent inadvertent opening. The battery circuits are protected by fusing devices to prevent over heating should a short circuit occur."
    Removable covers for hygiene."The user worn receivers are fitted with removable covers over the earphones that contact the skin and ears of the users, which can be removed for cleaning or exchange to prevent the spread or transmission of possible contamination."
    Effectiveness:
    Equivalent to approved systems of the same type."The effectiveness of this Infrared Amplification System is equivalent to approved systems of the same type."
    Clear reception from any angle."The infrared light gathering lenses on the receivers are designed to detect the sound signals from any angle providing for clear reception at any location within the installation."
    Reduced size/weight without loss of features."The use of solid state surface mount components and integrated circuits allow the size and weight of the system elements to be reduced without the loss of any of the beneficial features provided in the design. Units are less obtrusive to the user because of the smaller size and easier to use because of the reduction in weight."
    More natural sound with accessories."The use of several optional listening accessories provide a more natural sound to the user in different environments because the sounds are processed more like the sounds received by normal hearing persons."
    Wide choice of fitting possibilities."The ability of the system elements to utilize a variety of listening accessories allow the user a wider choice of fitting possibilities to better meet their personal needs."
    Battery charging indicator."The inclusion of a battery charging LED indicator on the PE600R receiver insures that the device's rechargeable batteries will receive a full charge by confirming to the user that a positive charging connection has been made."
    Flexible power source options."The ability of the receivers to operate on either rechargeable or throw-away batteries allow the user to select the power source most beneficial to their use."
    Selection of receiver types."The provision of optional receiver types provide the user with a selection which will best suit their own needs."
    Easier installation and adjustment of emitter."The combining of the power and amplifier circuits with the infrared LED circuits into one unit, provide easier installation and adjustment of the infrared emitter PE600E."
    Dual frequency operation (2.3/2.8 Mhz)."The ability of the PE600E to operate on both 2.3 Mhz and 2.8 Mhz simultaneously provides greater flexibility for the use of the system. Either stereo or 2-channel programs can be transmitted by the same system."
    Elimination of interference from electrical circuits/lighting."The use of the higher 2.3 Mhz and 2.8 Mhz transmission frequencies eliminates the interference which may be caused by electrical circuits and lighting, thus providing a clearer, interference free sound to the listener."
    Flexible mounting options for emitter."The ability of the PE600E Infrared Emitter to be mounted in several ways provide more flexibility for use as either a small portable system or in permanently installed systems for large auditoriums."
    Scalability for large installations."The ability of the systems to be enlarged by the addition of more infrared emitters provides full sound coverage even in very large installations."
    Compatibility with tele-coils in hearing aids."The ability of the infrared receivers to power the tele-coil found in most hearing aids enables the users to utilize their own hearing aids which provide more convenience of use as well as allowing the user to hear the same sounds they would normally hear using only their hearing aid."
    No dead spots or sound fade-outs."The design of the infrared detecting elements in the receivers allow the user to sit anywhere within the infrared environment without experiencing the dead spots or sound fade-outs usually found in most installations."
    Increased speech understanding and word discrimination."The use of an Infrared Group Amplification System provides increased speech understanding and word discrimination for the hearing impaired in listening environments such as school classrooms, concert halls, theaters or other similar situations where noises, reverberation or poor acoustical treatment may be found. The use of these systems also reduce the detrimental effects caused by the distance between the speaker or sound source and the hearing impaired listener."
    Dual-channel operation for language translation."The ability of the infrared system to operate on two channels simultaneously allows the system to be used for language translation thus integrating users of different languages into the same environment."
    Adaptability for secure sound environments."The PE600 Infrared Systems are especially adaptable for use in installations where security of the sound is desired, such as in court rooms or jury deliberation rooms."

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary does not detail a specific test set, its sample size, or its data provenance (e.g., country of origin, retrospective/prospective). The document primarily cites clinical studies on FM Auditory Training Systems to substantiate the effectiveness claims related to increased speech understanding and word discrimination. It does not describe a new clinical trial specifically for the PE600 Series.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The substantiation for effectiveness claims references "many clinical studies" by specific researchers (Dr. David Hawkins, Dr. Mark Ross, Robert Benoit, M.A., and Dr. Thomas Giolas) but does not outline how ground truth was established within those studies or if they were done for the PE600 series directly.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable to the Phonic Ear Model PE600 Series Infrared Group Amplification System. This device is a sound amplification system, not an AI-assisted diagnostic or interpretive tool that would involve "human readers" or AI assistance in the way typically seen in medical imaging or diagnostic studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the device is not an algorithm, but a hardware-based amplification system.

    7. The Type of Ground Truth Used

    For the claims related to "increased speech understanding and word discrimination," the document states: "Effectiveness claims outlined in paragraph 16 have been NOTE: substantiated by the results of many clinical studies made over the years since the introduction of FM Auditory Training Systems. Three such study reports by Dr. David Hawkins, Dr. Mark Ross, Robert Benoit, M.A. and Dr. Thomas Giolas form a part of this 510(k) submission."

    This indicates that the ground truth for effectiveness was established through prior clinical studies (outcomes data) on similar FM Auditory Training Systems, rather than a new, specific ground truth established for the PE600 series in this submission. The nature of these "clinical studies" would likely involve objective measures of speech recognition or discrimination in controlled listening environments.

    8. The Sample Size for the Training Set

    This information is not provided as this device does not involve a "training set" in the context of machine learning or AI models.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided as it is not applicable to the type of device being described.


    Summary of Limitations in the Provided Document for Your Request:

    The document is a 510(k) summary from 1997 for an infrared group amplification system. Its purpose is to demonstrate substantial equivalence to predicate devices, not to present a detailed clinical study with quantifiable acceptance criteria, specific test/training sets, or AI performance metrics. Therefore, much of the information you requested (especially around sample sizes, expert qualifications, adjudication, MRMC studies, and AI-specific ground truth) is not present in this type of regulatory submission. The "study that proves the device meets the acceptance criteria" is broadly defined as prior clinical research on similar FM Auditory Training Systems, used to support the general effectiveness claims.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974287

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-02-10

    (88 days)

    Product Code
    Regulation Number
    874.3320
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is a receiver for an FM group auditory trainer. It is intended to be used with the Phonic Ear PE 571t group auditory trainer transmitter. Auditory trainers are used primarily in institutions (schools, churches, theaters, etc.) to allow teachers, ministers, etc. to communicate with individuals with hearing deficits.

    This device can also be used with its built-in microphones or plug-in microphones as a stand alone assistive listening device.

    Device Description

    This device is a receiver for an FM group auditory trainer. It is intended to be used with the Phonic Ear PE 571t group auditory trainer transmitter. The device is powered by low voltage batteries which are contained within a battery compartment that is fitted with a locking cover to prevent inadvertent opening. The battery circuit is fused to prevent over heating should a short circuit occur. The devices' plastic housing is made of Cycolac plastic. The device includes controls for maximum power output, sound pressure level of built-in receiver microphones and FM signal, CHG/LOW BATT LED, NO FM LED, enlarging lens for FM frequency channel number, a 3 position switch for operating modes, a 7 position FM level switch for each ear, a 7 position SSPL switch for each ear, a 7 position GAIN switch for each ear, a 7 position TONE switch for each ear, and a 2 position microphone switch. It uses solid state surface mount components and integrated circuits. The receiver's FM receiving frequency is controlled by 2 rotary switches accessible under a latching cover.

    AI/ML Overview

    The provided document is a 510(k) notification for a medical device (Group Auditory Trainer, Model PE571R). It describes the safety and effectiveness of the device, but it does not contain the specific acceptance criteria or a study designed to prove the device meets these criteria in the typical sense of a clinical trial with predefined metrics.

    Instead, the document focuses on demonstrating substantial equivalence to previously approved devices, a requirement for 510(k) submissions. The "effectiveness" section lists various features and functionalities of the device and how they contribute to its intended use. For these effectiveness claims, the document refers to existing clinical studies that support the general concept of FM Auditory Training Systems, rather than a single study specifically designed for this particular device model and its acceptance criteria.

    Therefore, many of the requested information points cannot be directly extracted from this document in the way they would be from a clinical trial report.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly defined in terms of measurable performance metrics with thresholds (e.g., "sensitivity must be >X%"). The criteria for acceptance are implicitly that the device is safe and effective, and substantially equivalent to predicate devices.
    • Reported Device Performance: The document describes various features and functionalities as contributing to the device's effectiveness. These are qualitative descriptions of performance rather than quantitative measurements against specific criteria. For example, "The maximum power output of the receiver can be set to the appropriate level required by the user and then locked at that level to prevent over amplification..." describes a safety/effectiveness feature without providing a specific quantitative performance value or an acceptance threshold for that value.
    Acceptance Criteria (Implicit from Effectiveness Claims)Reported Device Performance (as described)
    Prevention of over-amplification/painMaximum power output can be set and locked.
    Proper balance between signal sourcesIndependent control of receiver microphones and FM transmitter signal levels.
    Indication of battery statusCHG/LOW BATT LED provides positive indication of charge and low battery.
    Indication of FM signal receptionNO FM LED indicates when no FM signal is received.
    Easy visibility of FM frequency channelEnlarging lens on front face.
    Multiple operating modes3-position switch for various operating modes.
    Attenuation of audio signals for emphasis7-position FM level switch for each ear.
    Limiting output to appropriate sound pressure level7-position SSPL switch for each ear.
    Adjustable amplification level7-position GAIN switch for each ear.
    Adjustable frequency response7-position TONE switch for each ear.
    Options for microphone input2-position microphone switch for built-in or BTE device microphone.
    Compact size and light weightUse of solid state surface mount components and integrated circuits.
    Natural sound in different environmentsUse of several optional listening accessories.
    Wide choice of fitting possibilitiesAbility to use a variety of listening accessories.
    Optimum performance in most listening environmentsSelection of operating modes (FM only, FM+mic, mic only).
    Easy change of FM channels2 rotary switches under a latching cover.
    Improved word discrimination in noisy environments, reduced effects of distance/reverberation"Effectiveness claims outlined in paragraph 17 have been substantiated by the results of many clinical studies made over the years since the introduction of FM Auditory Training Systems." (General claim, not device-specific study)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable for this document. The safety and effectiveness claims are based on substantial equivalence to predicate devices and general understanding of FM Auditory Training Systems, supported by existing literature. There is no specific "test set" for this particular device's performance in a clinical study detailed here. The technical specifications were obtained by testing devices per ANSI Standard ANSI.S3.22-1987, which is a standard for hearing aid characteristics, not a clinical test set in the usual sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No specific test set with ground truth established by experts for this device is detailed in the submission. The acceptance of the device relies on its technical specifications meeting standards and its features aligning with established principles of FM auditory trainers, as judged by the FDA.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No specific test set requiring adjudication is detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an auditory trainer, not an AI-powered diagnostic tool for interpretation by human readers. It's designed to assist hearing, not assist interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a hardware device with specific audio processing capabilities, not an algorithm in the sense of AI. Its "performance" is its ability to transmit and amplify sound according to its design, which is tested against engineering standards (like ANSI S3.22-1987).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the claims of improved word discrimination, reduced effects of distance/reverberation (paragraph 17), the document references "many clinical studies made over the years since the introduction of FM Auditory Training Systems." It specifically mentions "Three such study reports by Dr. David Hawkins, Dr. Mark Ross, Robert Benoit, M.A. and Dr. Thomas Giolas form a part of this 510K submission." These studies would likely use outcomes data (e.g., speech recognition scores in various conditions) as their ground truth for demonstrating the effectiveness of FM systems in general. However, the details of how ground truth was established in those specific studies are not provided in this document.
    • For the technical specifications, the "ground truth" is adherence to the ANSI Standard ANSI.S3.22-1987, which defines objective measurement methods for hearing aid characteristics.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device that would require a training set.

    9. How the ground truth for the training set was established

    • Not applicable.

    In summary, this 510(k) notification primarily relies on demonstrating that the device's design, features, and technical specifications are substantially equivalent to legally marketed predicate devices and adhere to relevant engineering standards. The effectiveness claims for FM auditory trainers in general are supported by existing clinical literature, rather than a new, dedicated clinical study specifically for this model with defined acceptance criteria and a test set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1