Found 6 results

510(k) Data Aggregation

    K Number
    K102887

    Validate with FDA (Live)

    Date Cleared
    2011-05-16

    (228 days)

    Product Code
    Regulation Number
    876.4680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PercSys Accordion® Stone Management Device is intended to be used in endoscopic procedures to bypass, entrap and remove calculi and other foreign objects from the urinary tract, to prevent retrograde migration of calculi during lithotripsy, and to guide instrumentation within the ureter.

    Device Description

    The Accordion device consists of a film membrane preloaded onto a two-part wire guide with a removable handle. The device is advanced within the urinary system similar to a wire guide, then, once the film membrane is in the desired position, the film folds into a helical coil (a film-based occlusion) which occupies the lumen of the anatomy in which it resides. In this manner, the occlusion limits migration of stones and stone fragments during lithotripsy. The shaft of the device is 0.97mm (0.038 inch) in outer diameter, has a 150 cm working length, and the film occlusion can be formed into either a 7 mm or 10 mm diameter helical coil. The device is visible under fluoroscopic imaging due to marker bands on either side of the film component and radiopaque disc embedded in the film. The Accordion device is provided sterile and is a single-use only device.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (PercSys Accordion® Stone Management Device) and details its intended use, description, and substantial equivalence to predicate devices. It does not contain information about specific acceptance criteria, device performance metrics, or details of a study that explicitly proves the device meets such criteria.

    The document states: "Bench top and clinical data from the literature were submitted in support of the proposed performance claims for this 510(k) submission, as well as in support of documenting that the changes in device materials and sterilization method has not affected design specifications." and "The bench top and clinical data show the Accordion device is as safe and as effective as the predicate devices and support the proposed advertising claims."

    However, it does not provide the details of this data, making it impossible to answer the specific questions regarding acceptance criteria, sample sizes, ground truth establishment, or clinical study methodologies.

    Therefore, I cannot populate the table or answer the specific questions as the required information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082803

    Validate with FDA (Live)

    Date Cleared
    2008-11-13

    (50 days)

    Product Code
    Regulation Number
    876.4680
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Coaxial Accordion Stone Management Device Urological is intended to be used endoscopically to entrap and remove calculi and other foreign objects from the urinary tract and facilitate drainage and retrograde pyelogram.

    Device Description

    The Coaxial Accordion Stone Management Device consists of a film membrane attached onto a cannula with a removable handle.

    AI/ML Overview

    The provided text mentions "Performance Data: Not required" for the Coaxial Accordion Stone Management Device. This implies that no specific performance study with acceptance criteria was conducted or necessary for its 510(k) clearance based on substantial equivalence to predicate devices.

    Therefore, I cannot provide the requested information as it is not present in the given document. The device received clearance based on its "same intended use, indications for use, and principles of and very similar technological characteristics as the predicate devices."

    Ask a Question

    Ask a specific question about this device

    K Number
    K052134

    Validate with FDA (Live)

    Date Cleared
    2005-09-22

    (48 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CYSTOGLIDE DILATING INTRODUCER SHEATH is intended to facilitate the dilation of urethral strictures and the introduction of catheters or instruments into the urethra. The CYSTOGLIDE DILATING INTRODUCER SHEATH is indicated for use as a urethral dilator and as a guide for urological catheters or instruments and as a lubricious barrier between the urethral tissue and the catheter or instrument.

    Device Description

    The CYSTOGLIDE DILATING INTRODUCER SHEATH consists of a film, a sheath, a stabilizing hub and a dilator. The device can be front-loaded onto a urological guidewire.

    AI/ML Overview

    The provided 510(k) summary for the CYSTOGLIDE DILATING INTRODUCER SHEATH does not contain specific acceptance criteria or a detailed study proving the device meets explicit performance metrics.

    Instead, the submission focuses on demonstrating substantial equivalence to predicate devices based on "Performance data demonstrated no significant difference in the performance of the CYSTOGLIDE DILATING INTRODUCER SHEATH and the predicate device." This means the device's performance was compared to existing, legally marketed devices, rather than against predefined, quantitative acceptance criteria.

    Therefore, providing a table of acceptance criteria and reported device performance as requested, along with details about sample sizes, ground truth establishment, expert qualifications, and MRMC studies, is not possible based on the information given. The document does not provide these types of details because the approval pathway is substantial equivalence, not a demonstration of meeting specific performance endpoints.

    However, I can extract and summarize the available information related to performance and study references:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated as quantitative criteria with predefined pass/fail thresholds. The basis for acceptance is "no significant difference" compared to predicate devices."Performance data demonstrated no significant difference in the performance of the CYSTOGLIDE DILATING INTRODUCER SHEATH and the predicate device."

    2. Sample size(s) used for the test set and the data provenance

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The document only refers to "Performance data."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not specified. The document does not describe a study involving human expert evaluation for establishing ground truth or performance.

    4. Adjudication method for the test set

    • Not applicable/Not specified.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs. without AI assistance

    • Not applicable. This device is a physical medical device (dilating introducer sheath), not an AI/software device that would involve human readers or comparative effectiveness studies with AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable/Not explicitly stated. The "performance data" likely refers to bench testing or animal model studies assessing physical properties and functionality mimicking the intended use, rather than clinical ground truth as understood in diagnostic AI/imaging studies. The comparison is to a predicate device's performance.

    8. The sample size for the training set

    • Not applicable. This is for a physical device, not an algorithm requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable.

    Summary based on available information:

    The 510(k) submission for the CYSTOGLIDE DILATING INTRODUCER SHEATH demonstrates substantial equivalence based on a comparison to predicate devices, specifically stating "Performance data demonstrated no significant difference in the performance of the CYSTOGLIDE DILATING INTRODUCER SHEATH and the predicate device." The document does not provide details on quantitative acceptance criteria, specific study designs (e.g., sample sizes, provenance), expert involvement, or AI-related metrics as it pertains to a physical medical device rather than a diagnostic algorithm or AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042877

    Validate with FDA (Live)

    Date Cleared
    2004-11-16

    (29 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MICROVERTER Ureteral Access Device is intended to facilitate the introduction of endoscopes and other instruments into the urinary tract. The device is indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract, and as a lubricious barrier between the ureteral tissue and the endoscope or instrument.

    Device Description

    The MICROVERTER Ureteral Access Sheath consists of a film membrane covering an outer tube, and a pusher tube that provides a lumen for the introduction of endoscopes and instruments. The outer tube is pre-loaded with the film membrane.

    AI/ML Overview

    The provided text, a 510(k) summary for the MICROVERTER Ureteral Access Sheath, states that "Performance data demonstrated no significant difference in the performance of the MICROVERTER Ureteral Access Sheath and the predicate device." However, it does not explicitly list acceptance criteria nor detail the study that proves the device meets specific criteria.

    Instead, the submission relies on demonstrating substantial equivalence to predicate devices. This means that if the predicate devices were deemed to meet performance and safety requirements, and the new device is substantially equivalent, it is also considered to meet those requirements.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the provided document. The basis of acceptance is substantial equivalence to predicate devices."Performance data demonstrated no significant difference in the performance of the MICROVERTER Ureteral Access Sheath and the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample size used for any performance testing or the provenance (e.g., country of origin, retrospective/prospective) of the data. It merely mentions "Performance data."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not mention the involvement of experts to establish ground truth for any test set.

    4. Adjudication Method:

    The document does not describe any adjudication method as it does not detail a study involving expert assessment of a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention or describe a MRMC comparative effectiveness study. As such, no effect size for human reader improvement with AI assistance can be reported. The device itself is a medical tool, not an AI diagnostic system.

    6. Standalone Performance Study:

    The document does not explicitly describe a standalone (algorithm only) performance study. The "Performance data" mentioned is likely a comparison of the device's functional characteristics to the predicate device, rather than an "algorithm only" study, as this is a physical medical device.

    7. Type of Ground Truth Used:

    The document does not specify the type of ground truth used. Given the nature of the device (ureteral access sheath), "ground truth" would likely relate to objective measurements of physical properties, functionality (e.g., ease of introduction, lubricity, fluid injection capabilities, integrity), or potentially in-vivo performance in animal models, rather than expert consensus on images or pathology. However, specific details are not provided.

    8. Sample Size for the Training Set:

    The concept of a "training set" is generally applicable to machine learning or AI models. Since this is a physical medical device and the document doesn't indicate the use of AI, the notion of a training set as typically understood does not apply in this context. The document does not provide information on any "training set" sample size.

    9. How Ground Truth for the Training Set Was Established:

    As the concept of a training set is not applicable (see point 8), the document does not describe how ground truth for a training set was established.

    Summary of the Study (as inferred from the 510(k) Summary):

    The "study" referenced in the 510(k) summary is not described in detail but broadly refers to the generation of "Performance data." The primary objective of presenting this data was to demonstrate that the MICROVERTER Ureteral Access Sheath has "no significant difference in the performance" compared to its predicate devices (PSI SLIP Urology Introducer Sheath, Cook Flexor Ureteral Access Sheath, Boston Scientific Navigator Ureteral Access Sheath). This approach is standard for 510(k) submissions, where substantial equivalence to legally marketed predicate devices is the pathway to clearance. The specific methods, sample sizes, or detailed results of this performance evaluation are not provided in the publicly available summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040519

    Validate with FDA (Live)

    Date Cleared
    2004-05-20

    (83 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SLIP Urology Catheter is intended to provide an intermittent pathway for draining fluids from the bladder. The device is indicated for providing increased lubricity during the catheter's advancement.

    Device Description

    The SLIP Urology Catheter consists of a catheter, a sheath, a snap ring, and a guide ring. The catheter is pre-loaded with a membrane sheath for increased lubricity.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Substantial Equivalence to Predicate DeviceThe SLIP Urology Catheter is identical to the Memcath Urology Catheter (except for the trade name). It has the same intended use, indications for use, technological characteristics, and principles of operation. Therefore, it is considered substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    No specific test set or data provenance (country of origin, retrospective/prospective) is mentioned in the provided text, as no performance data was provided for clinical testing of the device itself. The primary argument for clearance is substantial equivalence to a predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. No direct performance data or test set requiring expert ground truth establishment was conducted for the SLIP Urology Catheter, as its clearance was based on substantial equivalence.

    4. Adjudication Method for the Test Set

    Not applicable. No direct performance data or test set requiring adjudication was conducted for the SLIP Urology Catheter.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission relies on substantial equivalence to a predicate device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device (urology catheter), not an algorithm or AI-driven system.

    7. The Type of Ground Truth Used

    Not applicable. The clearance is based on substantial equivalence to a predicate device, not on specific performance metrics established against a ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device, not an AI or algorithm-driven system that would require a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is a medical device, not an AI or algorithm-driven system that would require a ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040520

    Validate with FDA (Live)

    Date Cleared
    2004-05-20

    (83 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SLIP Urology Introducer Sheath is intended to facilitate the introduction of catheters or instruments into the urethra. The SLIP Urology Introducer Sheath is indicated for use as a guide for urological catheters or instruments inserted into the urethra and as a lubricious barrier between the urethral tissue and the catheter or instrument.

    Device Description

    The SLIP Urology Catheter consists of a sheath, pusher tube, a twist ring, and a guide ring. The pusher tube is pre-loaded with the sheath.

    AI/ML Overview

    This 510(k) summary is for a medical device (SLIP Urology Introducer Sheath) seeking substantial equivalence to a predicate device, not for an AI/ML-driven device that typically undergoes performance studies with acceptance criteria, test sets, and ground truth establishment.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth is not applicable to this document.

    Here's an explanation based on the provided text:

    • Acceptance Criteria and Reported Device Performance: Not applicable. The submission asserts "identicality" to a predicate device rather than presenting new performance data against specific criteria.
    • Study Details (Sample Size, Data Provenance, Experts, Adjudication, MRMC, Standalone): Not applicable. No performance study, as typically understood for new AI/ML devices, was conducted or presented.
    • Type of Ground Truth: Not applicable. This concept is relevant for performance studies where a reference standard is needed to evaluate the device's output. Since no new performance data was provided, no ground truth was established for the SLIP Urology Introducer Sheath.
    • Training Set Sample Size and Ground Truth Establishment (for training): Not applicable. This device is not an AI/ML model requiring a training set.

    Summary of what can be extracted from the document:

    The core of this 510(k) submission is based on substantial equivalence.

    • Device: SLIP Urology Introducer Sheath
    • Predicate Device: Memcath's Urology Introducer Sheath
    • Key Argument for Substantial Equivalence: The submitter states: "The SLIP Urology Introduction Sheath is identical to the Memcath Urology Introduction Sheath (except for the trade name). The SLIP Urology Introduction Sheath has the same intended use, indications for use, technological characteristics, and principles of operation."
    • Performance Data: "No performance data were provided" because the device is claimed to be identical to the predicate.

    In a traditional 510(k) for a device like this, the 'study' to prove it meets acceptance criteria is often a comparison to a legally marketed predicate device demonstrating that it's just as safe and effective, without raising new questions of safety or effectiveness. For an identical device, the "acceptance criteria" can implicitly be considered "is identical to the predicate."