    K Number
    K981834

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-09-29

    (126 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Small Organ (Breast, Thyroid, Testes), Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Musculo-skeletal Conventional.

    Device Description

    The Perception's GPS-6TBMD Ultrasound System is a PC (computer) based and controlled real time. two dimensional, mechanical sector and electronic array diagnostic ultrasound and pulsed Doppler imaging system which produces diagnostic ultrasonic images and blood flow spectral analysis through user friendly operation. User interface is via an alphanumeric keypad, trackball and Icon-driven graphical interface. The Perception Inc. GPS-6TBMD Diagnostic Ultrasound System may be operated in M /B /D modes of inspection. The Perception GPS-6TBMD Diagnostic Ultrasound System supports M. B. M&B, Dual B, Quad B, D and D/B, display modes. All probes currently intended for use with the Perception GPS-6TBMD Diagnostic Ultrasound System are either mechanical sector devices or electronic linear and curve array, and make use of a fluid filled design.

    AI/ML Overview

    This appears to be a 510(k) summary for the Perception Inc. GPS-6TBMD Diagnostic Ultrasound System, dated September 29, 1998. It details the device, its intended uses, and its substantial equivalence to predicate devices.

    However, the provided document explicitly states:

    • "8. Discussion of Clinical Test Performed: Not Applicable" (Page 3)

    This means that no clinical study was performed to prove the device meets acceptance criteria in a clinical setting. Instead, the submission focuses on non-clinical tests to demonstrate substantial equivalence to legally marketed predicate devices.

    Therefore, I cannot provide information on acceptance criteria related to clinical performance, how ground truth was established, sample sizes, or expert qualifications for a clinical study, as no such study is documented in this 510(k) summary.

    Here's what I can extract regarding the non-clinical acceptance criteria and tests:

    Acceptance Criteria and Device Performance (Non-Clinical)

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Acoustic OutputStandards: NEMA 1992, Acoustic output measurement standard for diagnostic ultrasound equipment, NEMA Standard UD-21992. FDA Center for Devices and Radiological Health 1985 510(k) Guide for Measuring and Reporting and Acoustic Output of Diagnostic Ultrasound Medical Devices (Revised, 1989, 1990, 1991,1993, 1994, and 1995). Specific Assessments: Maximum Acoustic Output Information, Estimated In-Situ Intensity, FDA In-Situ Intensity Limits, Acoustic Output Information for each system/transducer/mode combination.Acoustic output for each system/transducer/mode/application combination was measured and calculated per the above documents. The testing conducted "revealed satisfactory testing results and compliance to applicable standards." No design characteristics were found that violated the requirements of the September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, Center for Devices and Radiological Health," or resulted in any safety hazards. The contract testing laboratory concluded that the device met all relevant requirements of the aforementioned guidance testing requirements.
    Electrical SafetyUL-544, Third EditionSatisfactory testing results and compliance to applicable standards.
    Electromagnetic Compatibility (EMC)Emissions: - Radiated and Conducted Emissions per CISPR 11 - Magnetic Field Emissions per MIL-STD-482D, method RE101 Immunity: - Electrostatic Discharge Immunity per IEC 801-2 - Radiated Field Immunity (3 V/m, 26 MHz to 1 Ghz, 100% Square wave Modulation) - Steady State Voltage Fluctuations - Line Voltage Dropouts - Slow Line Voltage Sags and Surges - Fast Transients Bursts per IEC 801-4 - Fast Line Voltage Surges - Conducted Energy Immunity per MIL-STD-462D, Method CS114 - Magnetic Field Immunity per MIL-STD-482D, Method RS 101 - Quasi-Static Electric Field ImmunitySatisfactory testing results and compliance to applicable standards. No design characteristics were found that violated the requirements of the September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, Center for Devices and Radiological Health," or resulted in any safety hazards. The contract testing laboratory concluded that the device met all relevant requirements of the aforementioned guidance testing requirements.
    Biocompatibility(Implicitly tested as stated in the conclusion.)"Biocompatibility testing revealed no new questions of safety or effectiveness." (Specific standards or detailed results are not provided in this summary.)
    Substantial EquivalenceThe device should have the same intended use, method of operation, materials, and design, or comparable without raising new questions of safety or effectiveness, as legally marketed predicate devices.The Perception Inc. GPS-6TBMD Diagnostic Ultrasound System has the same intended use as a combination of all cited predicates. All non-clinical testing and biocompatibility testing revealed no new questions of safety or effectiveness. When compared to the predicate devices, the device "does not incorporated any significant changes in intended use, method of operations, material or design that could affect safety or effectiveness."

    Study Details (Non-Clinical Testing)

    Since no clinical study was performed, the following points relate to the non-clinical testing described:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • This information is not applicable for non-clinical testing of a diagnostic ultrasound system's technical specifications and safety. The "test set" in this context refers to the device itself and its components. The provenance would be the manufacturing site and testing laboratories.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the sense of clinical ground truth. The "experts" involved would be the contract testing laboratory personnel who conducted the engineering and safety tests, qualified to perform tests against the listed standards (e.g., UL, CISPR, IEC, MIL-STD, NEMA). Their specific number or qualifications are not detailed in this summary.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable for non-clinical engineering and safety testing. Compliance is determined by whether the device's measured performance meets the specified thresholds or requirements of the cited standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a diagnostic ultrasound system from 1998, which is a hardware device. There is no mention of AI or image analysis software that would involve human readers or MRMC studies.
    5. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

      • No. Not applicable for this type of device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical tests, the "ground truth" refers to the established technical standards and regulatory limits (e.g., maximum acoustic output levels, acceptable EMI/EMC performance thresholds).
    7. The sample size for the training set:

      • Not applicable, as there is no algorithmic component requiring a training set for machine learning.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K974880

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-06-08

    (160 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Abdominal, Cardiac, Peripheral Vascular.

    Device Description

    The Perception's GPS-RST Ultrasound System is a PC (computer) based and controlled real time, two dimensional, mechanical sector and electronic array diagnostic ultrasound and pulsed Doppler imaging system which produces diagnostic ultrasonic images and blood flow spectral analysis through user friendly operation. User interface is via an alphanumeric keypad, trackball and Icon-driven graphical interface. The Perception Inc. GPS-RST Diagnostic Ultrasound System may be operated in M /B /D modes of inspection. The Perception GPS-RST Diagnostic Ultrasound System supports M, B, M&B, Dual B, Quad B. D and D/B, display modes. All probes currently intended for use with the Perception GPS-RST Diagnostic Ultrasound System are either mechanical sector devices or electronic linear array, and make use of a fluid filled design.

    AI/ML Overview

    The provided 510(k) summary for the Perception Inc. GPS-RST Diagnostic Ultrasound System does not contain information regarding acceptance criteria for clinical performance or any clinical study to prove the device meets such criteria.

    The document states "Discussion of Clinical Test Performed: Not Applicable" {3}, indicating that no clinical tests were performed for this submission. The 510(k) focuses entirely on non-clinical testing to demonstrate substantial equivalence to predicate devices, primarily concerning acoustic output and electrical safety standards.

    Therefore, I cannot populate the requested tables and information regarding acceptance criteria and clinical study details. The document explicitly declares that no clinical tests were conducted.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972192

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-12-17

    (190 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GPS-XYZ device is a diagnostic ultrasound system which produces twodimensional diagnostic ultrasonic images. The following intended uses are identified for the transducer applications: General Radiology, Abdominal, Cardiac and Vascular, with the use of ultrasonic probes from 2.8-12.5 MHz. There is no transcranial application for this device.

    Device Description

    The GPS-XYZ Diagnostic Ultrasound System device is a diagnostic ultrasound system which produces two-dimensional diagnostic ultrasonic images. The following intended uses are identified for the transducer applications: General Radiology, Abdominal, Cardiac and Vascular, with the use of ultrasonic probes from 2.8-12.5 MHz. There are no transcranial applications for this device.

    User interface is via an alphanumeric keypad and integrated trackball. The Perception Inc. GPS-XYZ Diagnostic Ultrasound System may be operated in M and B modes of inspection. The Perception Inc. GPS -XYZ Diagnostic Ultrasound System supports M, B, M&B, Dual B and Quad B display modes.

    All probes currently intended for use with the Perception Inc. GPS -XYZ Diagnostic Ultrasound System are either mechanical sector devices or electronic curved array, and make use of a fluid filled design.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Perception Inc. GPS-XYZ Diagnostic Ultrasound System, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document is a 510(k) summary for a pre-market notification, which focuses on demonstrating substantial equivalence to predicate devices rather than specific performance metrics against pre-defined acceptance criteria for a new, novel device. Therefore, explicit quantifiable "acceptance criteria" and "reported device performance" in the typical sense of a clinical trial are not presented.

    Instead, the "acceptance criteria" for this device are implicitly met by demonstrating:

    1. Substantial Equivalence: The device's intended use, technological characteristics, and safety/effectiveness are comparable to legally marketed predicate devices.
    2. Compliance with Recognized Standards: The device meets applicable industry and regulatory standards for safety and acoustic output.

    The document states:
    "This 510(k) submission is intended as a Track 1 type submission. Acoustic Output Reporting was prepared utilizing the following documents.

    • NEMA 1992, Acoustic output measurement standard for diagnostic ultrasound . equipment, NEMA Standard UD-21992.
    • FDA Center for Devices and Radiological Health 1985 510(k) Guide for ● Measuring and Reporting and Acoustic Output of Diagnostic Ultrasound Medical Devices, December, 1985, Revised, 1989, 1990, 1991, 1993, 1994 and 1995. --"

    And:
    "The following testing was conducted by Intertec Testing Services:

    • UL-544, Third Edition .
    • Radiated and Conducted Emissions per CISPR 11 ●
    • Magnetic Field Emission per MIL-STD-482D, method RE101 ●
    • Electrostatic Discharge Immunity per IEC 801-2 .
    • Radiated Field Immunity (3 V/m, 26 MHZ to 1 Ghz, 100% Square Wave Modulation)
    • Steady State Voltage Fluctuations
    • Line Voltage Dropouts
    • Slow Line Voltage Sags and Surges .
    • . Fast Transients Bursts per IEC 801-4
    • . Fast Line Voltage Surges
    • Conducted Energy Immunity per MIL-STD-462D, Method CS114 .
    • Magnetic Field Immunity per MIL-STD-482D, Method RS101 ●
    • Quasi-Static Electric Field Immunity

    None of the testing demonstrated any design characteristics that violated the requirements of the "FDA November 1993 Draft Reviewer Guidance for Pre-Market notification submissions, DCRND" or resulted in any safety hazards. It was Intertec Testing Services' conclusion that the device tested met all relevant requirements of the aforementioned quidance testing requirements."

    Acceptance Criteria (Implied)Reported Device Performance
    A. Substantial Equivalence in Intended Use: Device's clinical applications align with predicate devices.The GPS-XYZ system has the same intended use as a combination of all cited predicate devices: General Radiology, Abdominal, Cardiac, and Vascular imaging using transducers from 2.8-12.5 MHz. (Explicitly stated: "The Perception Inc. GPS-XYZ Diagnostic Ultrasound System has the same intended use as a combination of all cited predicates.")
    B. Substantial Equivalence in Technological Characteristics: Device operates similarly to predicate devices.The device produces two-dimensional diagnostic ultrasonic images and supports M and B modes of inspection, M, B, M&B, Dual B, and Quad B display modes, similar to predicate ultrasound systems. (Implicit through description and comparison to existing ultrasound devices).
    C. Acoustic Output Safety: Compliance with relevant acoustic output standards and limits."Acoustic output of each system/transducer/mode/application combination was measured and calculated per... NEMA 1992 and FDA 510(k) Guide... which revealed satisfactory testing results and compliance to applicable standards." This included: Maximum Acoustic Output Information, Estimated In-Situ Intensity, FDA In-Situ Intensity Limits, and Acoustic Output Information for each system/transducer/mode combination.
    D. Electrical Safety and EMC: Compliance with electrical safety and electromagnetic compatibility standards.Testing by Intertec Testing Services included UL-544, Radiated and Conducted Emissions (CISPR 11), Magnetic Field Emission (MIL-STD-482D), Electrostatic Discharge Immunity (IEC 801-2), Radiated Field Immunity, Steady State Voltage Fluctuations, Line Voltage Dropouts, Slow Line Voltage Sags and Surges, Fast Transients Bursts (IEC 801-4), Fast Line Voltage Surges, Conducted Energy Immunity (MIL-STD-462D), Magnetic Field Immunity (MIL-STD-482D), and Quasi-Static Electric Field Immunity. "None of the testing demonstrated any design characteristics that violated the requirements of the 'FDA November 1993 Draft Reviewer Guidance for Pre-Market notification submissions, DCRND' or resulted in any safety hazards. It was Intertec Testing Services' conclusion that the device tested met all relevant requirements..."
    E. Biocompatibility: Materials in contact with patients are biocompatible."All non-clinical testing and biocompatibility testing revealed no new questions of safety or effectiveness." (No specific test results provided in this summary, just a statement of conclusion).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not Applicable. The document explicitly states "Discussion of Clinical Tests Performed: Not Applicable." This indicates that clinical trials with patient data were not conducted as part of this 510(k) submission. The evaluation was based on engineering and safety testing, and comparison to predicate devices.
    • Data Provenance: Not Applicable. Since no clinical tests were performed, there is no patient data, and therefore no country of origin or retrospective/prospective classification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not Applicable. No clinical studies were performed, thus no "test set" requiring ground truth established by experts.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No clinical studies were performed.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a submission for a diagnostic ultrasound system, not an AI-powered diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance was not conducted and is not relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This device is a diagnostic ultrasound system, which is inherently operated by a human. It is not an algorithm performing a standalone task.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable. No clinical studies were performed, and no ground truth from patient data was established. The "ground truth" for this 510(k) submission primarily relies on the established safety and performance records of the predicate devices and compliance with recognized engineering and safety standards.

    8. The sample size for the training set:

    • Not Applicable. This device is a traditional ultrasound imaging system, not an artificial intelligence or machine learning algorithm that requires a "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. As it is not an AI/ML device, there is no training set and no ground truth for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K972973

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-10-15

    (65 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GPS-TUV device is a diagnostic ultrasound system which produces twodimensional diagnostic ultrasonic images. The following intended uses are identified for the transducer applications: General Radiology, Abdominal, Cardiac and Vascular, with the use of ultrasonic probes from 3.0-12.5 MHZ. There is no transcranial application for this device.

    Device Description

    The GPS-TUV Diagnostic Ultrasound System device is a PC (Computer) based and controlled real time, two-dimensional, mechancial sector and electronic array diagnostic ultrasound system which produces diagnostic ultrasonic images through a user friendly operation. The following intended uses are identified for the transducer applications: General Radiology, Abdominal, Cardiac and Vascular, with the use of ultrasonic probes from 3.0-12.5 MHZ. There are no transcranial applications for this device. User interface is via an alphanumeric keypad, trackball and Icon-driven graphical interface. The Perception Inc. GPS-TUV Diagnostic Ultrasound System may be operated in M and B modes of inspection. The Perception Inc. GPS-TUV Diagnostic Ultrasound System supports M, B, M&B, Dual B and Quad B display modes. All probes currently intended for use with the Perception Inc. GPS -TUV Diagnostic Ultrasound System are either mechanical sector devices or electronic curved array, and make use of a fluid filled design.

    AI/ML Overview

    The provided 510(k) summary for the Perception Inc. GPS-TUV Diagnostic Ultrasound System (K972973) indicates that no clinical tests were performed to establish the device's performance against acceptance criteria. The submission is a "Track 1 type submission," which relies on non-clinical tests and a comparison to predicate devices to demonstrate substantial equivalence.

    Therefore, the following information is not available in the provided text:

    • Acceptance criteria table and reported device performance based on clinical studies.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for the test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study or its effect size.
    • Standalone (algorithm only) performance study.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for a test set.
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    However, the document does contain information about non-clinical tests performed to demonstrate substantial equivalence, which can be interpreted as fulfilling certain "acceptance criteria" for regulatory clearance based on NEMA and FDA guidance documents.


    Summary of Non-Clinical Testing (interpreted as meeting acceptance criteria for substantial equivalence):

    The device's safety and effectiveness were established through non-clinical tests and a comparison to predicate devices, rather than through direct clinical performance metrics against specific acceptance criteria in a clinical study. The relevant "acceptance criteria" for this type of submission are compliance with established standards and guidance documents.

    Acceptance Criteria and Reported Device Performance (Non-Clinical Focus):

    Acceptance Criteria (based on regulatory standards/guidance)Reported Device "Performance" (Compliance)
    Acoustic Output:
    - NEMA 1992 (UD-2) complianceSatisfactory testing results and compliance to applicable standards.
    - FDA CDHR 1985 510(k) Guide complianceSatisfactory testing results and compliance to applicable standards.
    - Maximum Acoustic Output InformationMeasured and calculated per NEMA and FDA documents.
    - Estimated In-Situ IntensityMeasured and calculated per NEMA and FDA documents.
    - FDA In-Situ Intensity LimitsDemonstrated compliance.
    - Acoustic Output Information for each system/transducer/mode combinationMeasured and calculated for each combination.
    Electrical Safety & EMC:
    - UL-544, Third Edition complianceSatisfactory testing results and compliance.
    - Radiated and Conducted Emissions per CISPR 11Satisfactory testing results and compliance.
    - Magnetic Field Emission per MIL-STD-482D, method RE101Satisfactory testing results and compliance.
    - Electrostatic Discharge Immunity per IEC 801-2Satisfactory testing results and compliance.
    - Radiated Field Immunity (3 V/m, 26 MHz to 1 GHz, 100% Square Wave Modulation)Satisfactory testing results and compliance.
    - Steady State Voltage FluctuationsSatisfactory testing results and compliance.
    - Line Voltage DropoutsSatisfactory testing results and compliance.
    - Slow Line Voltage Sags and SurgesSatisfactory testing results and compliance.
    - Fast Transients Bursts per IEC 801-4Satisfactory testing results and compliance.
    - Fast Line Voltage SurgesSatisfactory testing results and compliance.
    - Conducted Energy Immunity per MIL-STD-462D, Method CS114Satisfactory testing results and compliance.
    - Magnetic Field Immunity per MIL-STD-482D, Method RS101Satisfactory testing results and compliance.
    - Quasi-Static Electric Field ImmunitySatisfactory testing results and compliance.
    Overall Safety & Effectiveness:
    - Compliance with "FDA November 1993 Draft Reviewer Guidance for Pre-Market notification submissions, DCRND"No design characteristics violated requirements; met all relevant testing requirements.

    Detailed breakdown of unavailable information:

    1. Sample size used for the test set and the data provenance: Not applicable, as no clinical test set was mentioned. The non-clinical tests were performed on the device itself and its components.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no ground truth for a clinical test set was established. The "ground truth" for the non-clinical tests was adherence to the specified regulatory standards and guidance.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an ultrasound system, not an AI-assisted diagnostic tool in the modern sense.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of a clinical study. The "ground truth" for the non-clinical tests was the specified performance limits and requirements defined by NEMA, FDA, UL, CISPR, IEC, and MIL-STD standards.
    7. The sample size for the training set: Not applicable, as no algorithm training was described.
    8. How the ground truth for the training set was established: Not applicable.

    Conclusion from the 510(k) Summary:

    The submitter concluded that the Perception Inc. GPS-TUV Diagnostic Ultrasound System is substantially equivalent to predicate devices based on:

    • Having the same intended use.
    • Satisfactory results from all non-clinical testing (acoustic output, electrical safety, EMC).
    • No new questions of safety or effectiveness revealed by non-clinical and biocompatibility testing.
    • No significant changes in intended use, method of operation, material, or design compared to predicate devices that could affect safety or effectiveness.
    • The device being a PC-based real-time 2D ultrasound system with M and B modes, supporting various display modes and covering general radiology, abdominal, cardiac, and vascular applications with specific probes (3.0-12.5 MHz).

    The "study" that proves the device meets "acceptance criteria" in this context is the series of non-clinical engineering and performance tests conducted by a contract testing laboratory, which demonstrated compliance with recognized industry standards and FDA guidance for diagnostic ultrasound equipment at the time.