Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101083

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-02

    (74 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor: pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid is intended to be used with autograft and/or allograft.

    The DSS™ Stabilization System - Slotted is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, kyphosis, and failed previous fusion (pseudarthrosis).

    In addition, the DSS™ Stabilization System - Slotted is indicated for use in patients:
    Who are receiving fusions with autogenous graft only;
    Who are having the device fixed or attached to the lumbar or sacral spine;
    Who are having the device removed after the development of a solid fusion mass.

    Note: The Rigid Coupler and Slotted Coupler are not intended to be used together.

    Device Description

    The DSS™ Stabilization System -- Rigid is comprised of a variety of pedicle screws sizes, and couplers that act as longitudinal spacers that are uniquely fitted for each individual case. The pedicle screws and couplers are manufactured from medical grade titanium alloy (Ti6Al4V). This submission presents the addition of a 20mm long rigid coupler.

    AI/ML Overview

    The provided text describes a medical device, the DSS™ Stabilization System -- Rigid, and its 510(k) premarket notification (K101083). The submission focuses on demonstrating substantial equivalence to predicate devices based on non-clinical mechanical testing.

    Here's an analysis of the provided information against the requested categories:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Material: Medical grade titanium alloy (Ti6Al4V)Device is manufactured from medical grade titanium alloy (Ti6Al4V).
    Mechanical Performance (Static Flexion): Equivalent to predicate devices based on ASTM F1717.Testing performed in conjunction with ASTM F1717. Performance indicates equivalence to predicate devices.
    Mechanical Performance (Static Extension): Equivalent to predicate devices based on ASTM F1717.Testing performed in conjunction with ASTM F1717. Performance indicates equivalence to predicate devices.
    Mechanical Performance (Static Axial Rotation): Equivalent to predicate devices based on ASTM F1717.Testing performed in conjunction with ASTM F1717. Performance indicates equivalence to predicate devices.
    Mechanical Performance (Dynamic Compression Bending): Equivalent to predicate devices based on ASTM F1717.Testing performed in conjunction with ASTM F1717. Performance indicates equivalence to predicate devices.
    Indications for Use: Same as predicate devices.The device has the same indications for use as predicate devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes non-clinical mechanical testing, not a clinical study involving human patients or data provenance in that sense. The "sample size" refers to the number of test articles used in the mechanical tests. The document states: "Static and dynamic mechanical testing was performed in conjunction with ASTM F1717." While it confirms that testing was done, the exact number of samples (test articles) used for each specific test (static flexion, extension, axial rotation, dynamic compression bending) is not explicitly stated.

    The data provenance is from non-clinical laboratory testing of the device components, not from human subjects or clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to this type of submission. The "ground truth" for mechanical testing is established by engineering standards (ASTM F1717) and the physical properties and behavior of the materials and device constructs, not by expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods are typically used in clinical studies for resolving discrepancies in expert opinions on clinical outcomes or image interpretations. For mechanical testing, the results are derived from physical measurements and engineering analyses, not subjective interpretation requiring "adjudication."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is a type of clinical study used to evaluate diagnostic imaging devices or AI algorithms where multiple human readers interpret cases. This submission is for a spinal implant, which undergoes mechanical testing and not typically MRMC studies for its premarket clearance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This device is a physical spinal implant, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this submission's non-clinical testing is based on established engineering standards and material science principles, specifically ASTM F1717. The performance of the device is measured against the expected mechanical properties and behavior defined by this standard and compared to the performance of predicate devices.

    8. The sample size for the training set

    This question is not applicable. This is not an AI/machine learning device that requires a training set. The "training" for a mechanical device is its design and manufacturing process according to engineering principles.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no "training set" in the context of this traditional medical device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1