510(k) Data Aggregation

    K Number
    K231253

    Validate with FDA (Live)

    Date Cleared
    2023-07-07

    (67 days)

    Product Code
    Regulation Number
    888.3540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Overture Orthopaedics Patellofemoral System is intended to be used in cemented arthroplasty in patients with osteoarthritis limited to the distal patello-femoral joint, patients with a history of patellar fracture, and those patients with failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release, etc.) where pain, deformity or dysfunction persists.

    Device Description

    The Overture Orthopaedics Patellofemoral System is comprised of trochlear implants, and a set of ancillary instruments. The trochlear implants are titanium and feature a polished articulating surface with a titanium nitride (TiN) coating. The patellar implants are cross-linked ultra high molecular weight polyethylene (UHMWPE) and feature a titanium pin for X-ray visualization. The trochlear and patellar implants utilize cemented pegs for immediate fixation. The trochlear implants also feature porous titanium bone-contacting surfaces.

    The implants are provided in a variety of configurations and sizes to accommodate various patient anatomy. The trochlear implants are offered in long and round geometries. The long implants are offered in lengths ranging 25-40mm and widths ranging 20-35mm. The round trochlear implants are offered in diameters ranging 20-35mm. The patellar implants are offered in a domed geometry in diameters of 25mm and 30mm.

    The implants are provided gamma sterilized and individually packaged. The ancillary instruments are provided non-sterile and are to be sterilized by the end user.

    AI/ML Overview

    This document is a 510(k) summary for the Overture Orthopaedics Patellofemoral System. It describes the device, its indications for use, and a comparison to a predicate device to establish substantial equivalence.

    Based on the provided text, the device in question is a medical implant, specifically a patellofemoral system for knee joint arthroplasty. The acceptance criteria and the study that proves the device meets them are described in terms of performance testing for mechanical characteristics, not in terms of diagnostic accuracy or reader performance as the format of the request assumes.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

      The document does not explicitly list acceptance criteria in a quantitative table with corresponding reported performance for each criterion. Instead, it lists the types of performance testing conducted to demonstrate substantial equivalence. The document states:

      "In summary, rationales and mechanical testing of the Overture Orthopaedics Patellofemoral System indicated there are no new risks and demonstrated substantial equivalence in performance compared to a legally marketed predicate."

      This implies that the acceptance criterion for each test was likely "demonstrates substantial equivalence to the predicate device" or "meets established rationales without new risks." The reported device performance is summarized as having met this overall criterion.

      Acceptance Criteria (Implied)Reported Device Performance
      Demonstrate substantial equivalence to predicate device in terms of:
      - Patellofemoral Contact Area and Contact StressTesting indicated no new risks and substantial equivalence.
      - Patellofemoral ConstraintTesting indicated no new risks and substantial equivalence.
      - Range of MotionTesting indicated no new risks and substantial equivalence.
      - Fatigue StrengthRationale adopted, indicating substantial equivalence.
      - WearRationale adopted, indicating substantial equivalence.
      - Additive Manufacturing CharacteristicsRationale adopted, indicating substantial equivalence.
      - Characterization of UHMWPERationale adopted, indicating substantial equivalence.
      - Characterization of Porous CoatingRationale adopted, indicating substantial equivalence.
      - Modified Metallic SurfacesRationale adopted, indicating substantial equivalence.
      Absence of new risksNo new risks indicated.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      The document describes "bench testing" and "cadaveric validation lab." It does not specify sample sizes for these tests (e.g., how many implants were tested for fatigue, how many cadavers were used). It also does not mention country of origin or whether the data was retrospective or prospective, as these terms are generally not applicable to bench and cadaveric testing of medical implants.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      This question is not applicable to the type of device and testing described. The "ground truth" for mechanical performance tests is typically measured physical properties, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      This question is not applicable. Adjudication methods are typically used for establishing ground truth in clinical studies involving interpretation (e.g., by radiologists), not for bench testing of medical implants.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      This question is not applicable. An MRMC study with human readers and AI assistance is relevant for diagnostic AI devices. The Overture Orthopaedics Patellofemoral System is a surgical implant undergoing mechanical and material testing, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      This question is not applicable. The device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For the performance testing mentioned (e.g., contact area, stress, constraint, fatigue), the "ground truth" would be physical measurements and engineering data obtained from standardized mechanical tests or cadaveric studies. The basis for comparison is the performance of the legally marketed predicate device.

    8. The sample size for the training set

      This question is not applicable. There is no concept of a "training set" for a physical medical implant device in the context of its mechanical performance testing.

    9. How the ground truth for the training set was established

      This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221292

    Validate with FDA (Live)

    Date Cleared
    2023-03-10

    (310 days)

    Product Code
    Regulation Number
    888.3520
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Overture Resurfacing Uni Knee Resurfacing System is intended to be used in the partial replacement of the articulating surfaces of the knee in instances where, due to compartmental degenerative disease, post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity, or previous arthroplasty, only the one side of the joint is affected. This device is intended to be used with bone cement.

    Device Description

    The Uni Knee Resurfacing System is comprised of femoral implants, and a set of ancillary instruments. The femoral implants are titanium and feature a spherical polished articulating surface with a titanium nitride (TIN) coating. The tibial implants are comprised of a titanium tibial tray with an integrated ultra-high molecular weight polyethylene (UHMWPE) tibial insert. The femoral implants and tibial trays utilize cemented pegs and porous titanium bone-contacting surfaces to allow for fixation. The femoral implant articulates against the tibial insert in a non-constrained manner.

    The implants are provided in a variety of configurations and sizes to accommodate various patient anatomy. Femoral implants are offered in oblong and round configurations. The oblong femoral implants are offered in lengths ranging 25-40mm and diameters ranging 17.5-27.5mm, and the round femoral implants are offered in diameters ranging 17.5-27.5mm. The tibial implants are offered in diameters ranging 17.5-22.5mm.

    The implants are provided gamma sterilized and individually packaged. The ancillary instruments are provided non-sterile and are to be sterilized by the end user.

    AI/ML Overview

    This document is a 510(k) summary for the Uni Knee Resurfacing System. It focuses on demonstrating substantial equivalence to a predicate device based on design, materials, and performance testing, rather than an AI/ML-based device requiring extensive clinical study for performance metrics like accuracy, sensitivity, or specificity.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets these criteria in the context of AI/ML performance (e.g., sample size, data provenance, expert ground truth, MRMC studies) is not applicable to this submission.

    The "Performance Testing" section (page 4) details bench testing conducted for the Uni Knee Resurfacing System, which typically involves mechanical and material evaluations, not clinical performance metrics for an AI algorithm.

    To specifically address the prompts:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The document describes engineering and material-based performance tests, not AI-related performance metrics. The conclusion states that these tests "indicated no new risks and demonstrated substantial equivalence in performance compared to a legally marketed predicate." Specific numerical acceptance criteria for each bench test are not provided in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. This document refers to bench testing of physical implants, not an AI algorithm's performance on clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth as typically defined for AI validation (e.g., expert consensus on medical images) is not relevant to this device's testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI. The "ground truth" for the performance testing would be established by objective measurements and scientific standards for material properties, mechanical strength, wear, etc., as per the listed ASTM and other relevant engineering standards.

    8. The sample size for the training set: Not applicable. There is no AI algorithm being trained.

    9. How the ground truth for the training set was established: Not applicable.