510(k) Data Aggregation

    K Number
    K181854

    Validate with FDA (Live)

    Date Cleared
    2019-05-07

    (300 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OssBuilder System is used for space maintenance and stabilization of bone grafting material in an alveolar bone defect site.

    Device Description

    OssBuilder System is consisted of metal devices, OssBuilder, Healing Cap, Cover Cap, OB Anchor, and Tenting Screw, and is intended for use to stabilize and support of bone graft in dento-alveolar bony defect sites.

    Tenting Screw, a product in the OssBuilder System product group, is a temporary dental implant that can be implanted in the bone defect site of the alveolar bone to fix the guided tissue regeneration material.

    Healing Cap is used by connecting with the Tenting Screw to bind OssBuilder during osseointegration.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the OssBuilder System, an intraosseous fixation device. It asserts substantial equivalence to predicate devices based on non-clinical performance testing. Due to the nature of this submission, which focuses on demonstrating substantial equivalence through non-clinical testing rather than clinical study, many of the typical elements expected for evaluating AI/algorithm performance (e.g., ground truth establishment with experts, MRMC studies, training/test set details) are not applicable or provided.

    Here's a breakdown of the available information regarding acceptance criteria and device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this specific device, the "acceptance criteria" are not reported as specific performance metrics like accuracy, sensitivity, or specificity, which would be common for diagnostic AI/software. Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to legally marketed predicate devices through:

    • Identical or similar design, operating principle, and intended use as predicates.
    • Identical or similar materials.
    • Successful non-clinical performance testing (bench tests) demonstrating comparable mechanical properties.
    • Validated sterilization and shelf-life.
    • Biocompatibility evaluation.

    The reported device performance heavily relies on comparisons to predicate devices in the following aspects:

    Acceptance Criterion (Implicitly for SE)Device Performance (Reported)
    Intended UseOssBuilder System: "used for space maintenance and stabilization of bone grafting material in an alveolar bone defect site." This is compared to the predicate's intended use for fixing and stabilizing non-resorbable barrier membranes for tissue regeneration. The submission argues equivalence despite linguistic differences, viewing them as a whole system.
    Principle of OperationOssBuilder System (Tenting Screw): "is placed in maxilla or mandible bone area to be combined with OssBuilder (non-resorbable titanium membrane) and Healing Cap or Cover Cap. It helps to support and prevent the mobility of the membrane during the bone regeneration." This is compared to the predicate's "temporary dental implant ... to be combined with titanium membrane and cover screw or cover cap. It helps to support the membrane." Again, linguistic differences are noted but equivalence asserted.
    MaterialOssBuilder System: Ti-6Al-4V ELI (ASTM F136). Predicate (Neo GBR): Ti-6Al-4V ELI (ASTM F136). (Identical)
    Dimensions (Tenting Screw)OssBuilder System: Head Diameter: 3.0mm; Major Thread Diameter: 2.1mm; Length: 8.5, 10.0, 11.5, 13.0mm. Predicate (Neo GBR): Head Diameter: 2.85mm; Major Thread Diameter: 2.0mm; Length: 7.2, 8.7, 10.2, 11.7, 13.2, 15.2, 18.2mm. (Similar, differences noted but deemed acceptable).
    Dimensions (Healing Cap)OssBuilder System: Diameter: 4.0, 5.0mm; Height: 7.0, 8.0mm. Predicate (K172354): Diameter: 4.0, 5.0mm; Height: 7.0, 8.0mm, 4.3mm (for Cover Cap). (Identical for healing cap).
    SterilizationOssBuilder System: Gamma Sterilization. Predicate: Gamma Sterilization. (Identical)
    Shelf LifeOssBuilder System: 8 years. Predicate (Neo GBR): Unknown. Predicate (K172354): 8 years. (Similar/Identical to one predicate)
    BiocompatibilityEvaluation conducted per ISO 10993-1. Deemed necessary as the OssBuilder System is made of same material, chemical composition, manufacturing process, and body contact with the predicate devices. (No additional testing required due to substantial equivalence in these factors).
    Sterilization Validation & Shelf-lifeConsidered by leveraging data from prior submissions.
    Mechanical PropertiesBench tests conducted per ASTM F543 (fracture load, insertion torque, axial pullout strength, torsional strength, breakage angle, removal torque, self-tapping force). Results found substantially equivalent to the predicate device for worst-case insertion in block material representing hard/soft bone.

    2. Sample Size for the Test Set and Data Provenance

    This submission does not involve an AI/algorithmic device that would typically have a "test set" of patient data for performance evaluation. The "tests" performed are bench tests on the physical device components.

    • Sample Size: Not explicitly stated as a number of devices/units tested, but refers to "Bench tests evaluated for OssBuilder System according to ASTM F543" for "worst-case insertion in block material." The assumption is that standard engineering testing practices for mechanical properties were followed.
    • Data Provenance: The tests are laboratory-based bench tests, not patient data from a specific country. They are likely conducted at the manufacturer's facility or a contracted testing lab. The submission is made by a company based in the Republic of Korea (Osstem Implant Co., Ltd.) with a US correspondent (Hiossen Inc.).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. As this is a physical medical device submission demonstrating substantial equivalence through non-clinical testing, there is no "ground truth" derived from human experts interpreting medical images or data. The "ground truth" for mechanical properties is established by the specifications defined in standards like ASTM F543 and the performance of the predicate devices.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human interpretation or subjective assessment that would require an adjudication method. Mechanical properties are measured objectively.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device, and therefore, no MRMC comparative effectiveness study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic device.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission are the performance specifications and characteristics of the legally marketed predicate devices and generally accepted industry standards (e.g., ASTM F543) for mechanical properties testing, along with established biocompatibility guidelines (ISO 10993-1). The manufacturer demonstrates that the new device meets or is "substantially equivalent" to these established benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or AI, and therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172354

    Validate with FDA (Live)

    Date Cleared
    2018-03-26

    (235 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.

    Device Description

    OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. It is consisted of OssBuilder, Healing Cap, Cover Cap, and OB Anchor. The OssBuilder System is not indicated for permanent implantation.

    AI/ML Overview

    The provided text is a 510(k) summary for the OssBuilder System, a dental device intended to stabilize and support bone grafts. It does not describe a clinical study in the traditional sense, but rather a set of non-clinical performance tests to demonstrate substantial equivalence to predicate devices. Therefore, many of the requested categories for clinical studies will not be applicable.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific quantitative thresholds. Instead, it states that the mechanical properties of the OssBuilder System were found to be "substantially equivalent" to predicate devices. The acceptance criteria are implied to be that the proposed device performs at least as well as the predicate device in relevant mechanical tests and maintains biocompatibility and sterility.

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    BiocompatibilityMaterials (Titanium, Ti-6Al-4V) must be biologically compatible for dental use, consistent with predicate devices."Biocompatibility evaluation for OssBuilder System is not considered because the materials used for manufacturing OssBuilder System are titanium and titanium alloy which have been generally and widely used as a dental material such as implant for a long time. OssBuilder System is made of same material, chemical composition, and body contact with the predicated devices..."
    Sterilization ValidationSterility must be confirmed according to ISO 11137, with worst-case testing among subject devices."For the sterilization validation, it was performed accordance with worst-case chosen among subject devices and following ISO 11137." Acknowledged as performed.
    Shelf-lifePackaging materials and device integrity must be validated for the stated shelf-life (8 years for components)."For the shelf-life of subject devices, we considered their validation of packaging materials by leveraging the data that of our prior submissions." New shelf life of 8 years for OssBuilder, Healing Cap, Cover Cap, and OB Anchor compared to 5 years for predicates, supported by leveraging prior submission data.
    Mechanical PropertiesTensile Strength, Yield Strength, and Elongation of components must be comparable to, or better than, predicate devices."Bench tests evaluated for OssBuilder System included Tensile Strength, Yield Strength, and Elongation. Mechanical properties from bench tests of OssBuilder System found substantially equivalent to the predicate device." The document explicitly states that mechanical properties were found "substantially equivalent" to predicate devices, and that the differences in design (e.g., shape, height, connection structure) or material (for some Cover Cap models) do not affect function or indication for use.

    2. Sample size used for the test set and the data provenance

    For the non-clinical performance testing (biocompatibility, sterilization, shelf-life, mechanical properties), specific sample sizes are not explicitly mentioned in the provided text. The data provenance is also not detailed beyond stating they are "non-clinical testing data" and "bench tests." It can be inferred that these tests were conducted by the manufacturer, Osstem Implant Co., Ltd., likely in their Korean facilities or outsourced to testing laboratories. The nature of these tests (e.g., material testing, sterilization validation) means they are not retrospective or prospective studies in the sense of clinical trials, but rather laboratory-based evaluations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the document describes non-clinical performance testing (material properties, sterilization, etc.), not a study requiring expert readers or ground truth establishment for diagnostic accuracy.

    4. Adjudication method for the test set

    This question is not applicable for the same reasons as #3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, human readers, or AI assistance in this document. The OssBuilder System is a physical dental device, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the OssBuilder System is a physical dental device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests described:

    • Biocompatibility: The ground truth is based on the well-established biocompatibility of titanium and Ti-6Al-4V alloys, which are widely used in medical implants, and the equivalence to predicate devices made of the same materials.
    • Sterilization Validation: The ground truth for sterility is established by industry standards (ISO 11137).
    • Shelf-life: The ground truth for shelf-life is established by internal testing and leveraging data from prior submissions for packaging materials durability.
    • Mechanical Properties: The ground truth for mechanical properties (Tensile Strength, Yield Strength, Elongation) is derived from standardized material testing methods and comparison to the known properties of the predicate devices.

    8. The sample size for the training set

    This question is not applicable as there is no "training set" in the context of a physical medical device's non-clinical performance testing.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170736

    Validate with FDA (Live)

    Date Cleared
    2017-06-07

    (89 days)

    Product Code
    Regulation Number
    872.3660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HySil Putty is to be used as preliminary materials for:

    • Two-step Putty-wash impression technique .
    • · One-step Putty-wash impression technique

    HySil Heavy is to be used as heavy-bodied materials for:

    • . One-step impression technique (simultaneous technique)
      using single or dual viscosities
    • · Two-step impression technique using dual viscosities
    • · Functional impressions

    HySil Mono is to be used as a medium-bodied tray or syringeable impression material for:

    • Taking impressions over fixed/removable restorations and implants (i.e., transferring .
    • impression posts and bridge components)
    • · Functional impressions
    • · Fabricating crown and bridgework or inlays
    • · Fabricating full or partial dentures
    • . Reline impressions
    • · Use in the simultaneous mixing technique as well as the putty-wash and triple tray techniques
    • · Transferring root posts when fabricating posts and cores indirectly

    HySil Light is to be used as syringeable impression materials for:

    • Two-step putty-wash impression technique .
    • · One-step putty-wash impression technique
    • · Two-step impression technique using dual viscosities
    • · Reline impressions
    • · Fabricating full or partial dentures

    HySil Extra Light is to be used as syringeable impression materials for:

    • Two-step putty-wash impression technique .
    • · One-step putty-wash impression technique
    • · Two-step impression technique using dual viscosities
    • · Reline impressions
    • Fabricating full or partial dentures

    HySil Bite is used for impression as below.

    • Taking occlusal surfaces
    • Confirming occlusal surfaces
    • · Recording after putting the articulator
    Device Description

    HySil Impression Materials are addition-curing, elastomeric materials with hydrophilic properties, high tear strength, dimensional accuracy, and resistance to permanent deformation. The HySil impression materials family consists of six different viscosities (putty, heavy, mono, light. Extra-light and Bite) available in an assortment of delivery systems: standard 1:1 50 ml automix cartridges, and traditional 1:1 putty jars.
    Putty, Heavy, Mono, Light and Extra-light of this proposed devices are classified into Type 0, Type 1, Type 2, and Type 3 according to ISO 4823, and Bite is classified as Dentistry - Jaw relation recording material according to DIN 13903 HySil Bite is an addition-curing silicone bite registration materials in a 50ml cartridge type to measure accurate bite. HySil Bite is suitable for bite registration as it has a short setting time, high final hardness after curing for user convenience, and less elastic deformation when eliminated.

    AI/ML Overview

    This is a 510(k) premarket notification for a dental impression material, not an AI/ML device. Therefore, the questions related to AI/ML specific criteria (e.g., sample sizes for test/training sets, expert adjudication, MRMC studies, standalone algorithm performance) are not applicable.

    The submission focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance testing and biocompatibility.

    Here's the information that can be extracted from the provided text, aligning with the categories where applicable for a non-AI/ML device:

    1. Table of Acceptance Criteria and Reported Device Performance

    For HySil Extra Light and HySil Putty (ISO 4823 and ISO 10993 family):

    Performance CharacteristicAcceptance Criterion (Predicate)Reported Device Performance (HySil Extra Light)Reported Device Performance (HySil Putty)
    Component ColourPassPassPass
    Weight (Volume) testPassPassPass
    PackagePassPassPass
    Consistency45.03mm (Light), 26.86mm (Putty)44.61mm31.27mm
    Working Time2.25 min (Light)3.6 min-
    Mixing Time25 sec (Putty)-38 sec
    Detail ReproductionPassPassPass
    Compatibility with GypsumPassPassPass
    Linear Dimensional change0.04%0.027%0.031%
    Elastic Recovery98.33% (Light), 97.78% (Putty)99.5%99.8%
    Strain-in-Compression3.87% (Light), 4.70% (Putty)2.6%1.6%
    CytotoxicityPassPassPass
    Oral IrritationPassPassPass
    Maximization and SensitizationPassPassPass

    For HySil Bite (DIN 13903 and ISO 10993 family):

    Performance CharacteristicAcceptance Criterion (Predicate)Reported Device Performance (HySil Bite)
    ColoursPassPass
    Working Time (Processing Time)PassPass
    Minimum Period Spent in the Mouth1 min. 30 sec1 min. 30 sec
    Recovery after Deformationmore than 0.1mmmore than 0.1mm
    Load Endurance during BendingPassPass
    Hardness Degree (HD)More than 20 HD (Predicate values: 74 HD, 72 HD, 70 HD, 64 HD, 70 HD)50 HD
    Linear Dimensional Changeshall be within ± 1.5 % (Predicate values: 0.00%, 0.00%, 0.10%, 0.10%, 0.02%)- 0.16 %
    CytotoxicityPassPass
    Oral IrritationPassPass
    Maximization and SensitizationPassPass

    2. Sample size used for the test set and the data provenance:

    • The document implies that testing was performed on the HySil Impression Materials as the "subject device" against predicate devices.
    • The exact sample sizes for each specific test (e.g., how many specimens were tested for linear dimensional change) are not explicitly stated in the provided text.
    • Data Provenance: The manufacturer is Osstem Implant Co., Ltd., located in Busan, Republic of Korea. The testing was conducted as part of their 510(k) submission, suggesting it was internal testing by the manufacturer or a contracted lab. The data would be considered prospective as it was generated to support this specific regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • This question is not applicable as the submission is for a material, not an AI/ML device that generates diagnoses or interpretations requiring expert ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This question is not applicable as this is not an AI/ML device involving human interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable as this is not an AI/ML device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • This question is not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the physical and mechanical properties, the "ground truth" or acceptance criteria are based on recognized international standards:
      • ISO 4823 for general impression materials.
      • DIN 13903 for jaw relation recording materials (HySil Bite).
    • For biocompatibility, the "ground truth" is based on ISO 10993 family standards (e.g., ISO 10993-5 for cytotoxicity, ISO 10993-10 for sensitization and irritation).
    • The device performance is compared directly to the specified requirements within these standards and the reported performance of legally marketed predicate devices.

    8. The sample size for the training set:

    • This question is not applicable as this is not an AI/ML device. No training set was used.

    9. How the ground truth for the training set was established:

    • This question is not applicable as this is not an AI/ML device. No training set was used.