Found 2 results

510(k) Data Aggregation

    K Number
    K172583

    Validate with FDA (Live)

    Date Cleared
    2018-05-10

    (255 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthoPediatrics PediNail Intramedullary Platform is used for pediatric and stature adult patients as indicated to stabilize fractures of the femoral shaft; subtrochanteric fractures; psilateral neck/shaft fractures; prophylactic nailing of impending pathologic fractures; nonunion; fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

    Additional indications includes simple long bone fractures; severely comminuted spiral, large oblique and segmental fractures; polytrauma and multiple fractures; reconstruction, following tumor resection and grafting; supracondy|ar fractures; bone lengthening; fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.

    The OrthoPediatrics PediNail Intramedullary Platform is for single use only.

    Device Description

    The OrthoPediatrics PediNail Intramedullary Nailing System (K083726) consists of a rigid stainless steel nail and stainless steel screws. The proposed OrthoPediatrics PediNail Intramedullary Platform seeks to broaden the reach of the PediNail brand by offering additional nail diameters, screws, nails designed specifically for child and adolescent, and improved instrumentation. The proposed device is composed of the intramedullary fixation rods (nails) with screw holes at either end for fixation to bone, screws, interlocking pegs, and end caps coupled with device-specific instruments including the Modular Necks, the Targeting Guides, the Attachment Bolts, the Derotation Dial Targeting Guide Attachment, the AP Nail Positioning Jig, the AP Position Templates, the Impactors, and the Nail Extractor Adaptor.

    The OrthoPediatrics PediNail Intramedulary Platform nails have a complex 3-dimensional geometry resulting in an anatomically appropriate design. The smaller diameter of the nail allows it to be inserted in patients with a narrow medullary canal and allows for easier insertion without the need for excessive reaming.

    AI/ML Overview

    The provided text describes the OrthoPediatrics PediNail Intramedullary Platform, a medical device, and its 510(k) premarket notification. The document focuses on demonstrating substantial equivalence to a predicate device (OrthoPediatrics PediNail Intramedullary Nailing System K083726) rather than a study validating a specific acceptance criterion for an AI/CADe system.

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, and MRMC studies for an AI/CADe device is not applicable to the provided text. The submission is for a physical medical device (intramedullary fixation rod), not an AI algorithm.

    However, I can extract the information that is present regarding the non-clinical performance testing of the device:

    1. A table of acceptance criteria and the reported device performance:

    Test MethodAcceptance Criteria (Implied by equivalence)Reported Device Performance
    Static Four-Point Bend TestingEquivalent to predicate (K083726)Testing concluded that the proposed OrthoPediatrics PediNail Intramedullary Platform will perform equivalently to the predicate OrthoPediatrics PediNail Intramedullary Nailing System.
    Static Torsion TestingEquivalent to predicate (K083726)Testing concluded that the proposed OrthoPediatrics PediNail Intramedullary Platform will perform equivalently to the predicate OrthoPediatrics PediNail Intramedullary Nailing System.
    Torsional Yield Strength TestingEquivalent to predicate (K083726)Testing concluded that the proposed OrthoPediatrics PediNail Intramedullary Platform will perform equivalently to the predicate OrthoPediatrics PediNail Intramedullary Nailing System.
    Driving Torque TestingEquivalent to predicate (K083726)Testing concluded that the proposed OrthoPediatrics PediNail Intramedullary Platform will perform equivalently to the predicate OrthoPediatrics PediNail Intramedullary Nailing System.
    Full Construct Testing (Fatigue)Survival of 150,000 cycles at 75kg (165lbs)"Three such constructs survived 150,000 cycles at a load of 75kg (165lbs)." This demonstrates "no additional risk is conferred to the device from the presence or use of the dynamic slot."
    MR Environment (MR Conditional)MR Conditional StatusThe OrthoPediatrics PediNail Intramedullary Platform was evaluated in an MR Environment and was determined to be MR Conditional.

    2. Sample sized used for the test set and the data provenance:

    • Sample Size for Full Construct Testing: 3 constructs (described as "system worst case" with a 7mm nail locked distally and a specific fracture/osteotomy).
    • Data Provenance: The tests were non-clinical, performed by the manufacturer (OrthoPediatrics). Country of origin is not specified, but the company is based in Warsaw, Indiana, USA. The data is non-retrospective as it describes custom-built test setups.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable. This is not an AI/CADe device that requires expert ground truth establishment for diagnostic performance. The ground truth for mechanical testing is based on physical scientific principles and engineering standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable, as this is mechanical performance testing, not diagnostic assessment.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/CADe device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an AI algorithm. The performance described is the standalone mechanical performance of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests, the "ground truth" is defined by the physical properties of the materials, the mechanical stress and strain limits, and the ability to withstand simulated physiological loads based on established engineering principles and standards (e.g., ASTM F138 for material). For the fatigue testing, the "ground truth" for survival was reaching 150,000 cycles at a specific load without failure.

    8. The sample size for the training set:

    • Not applicable. There is no training set as this is not an AI/CADe device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no training set as this is not an AI/CADe device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172425

    Validate with FDA (Live)

    Date Cleared
    2017-12-06

    (118 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OrthoPediatrics Wrist Fusion Plates are intended for wrist arthrodesis and fractures of other small bones in children, adolescents, and small stature adults. Specific indications include post-traumatic arthritis of the wrist; rheumatoid wrist deformities requiring restoration; complex carpal instability; post-septic arthritis of the wrist; severe unremitting pain related to motion; brachial plexus nerve palsies; tumor resection; and spastic deformities.

    Device Description

    OrthoPediatrics Wrist Fusion Plates are provided non-sterile, are available in two lengths, 99mm and 112mm, to accommodate a patient with or without a Proximal Row Carpectomy (PRC). The Wrist Fusion Plates accept 2.7mm screws both distally and in the capitate, and 3.5mm screws proximally. These screws were previously cleared in the PediLoc Fragment System 510(k), K140431. In additional to the two wrist fusion plates, bending templates are included to assist with intra-operative contouring of the plate into either dorsiflexion or palmar flexion.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the OrthoPediatrics Wrist Fusion Plate. However, it does not describe:

    • A device that relies on AI or an algorithm.
    • A study that establishes acceptance criteria for "device performance" in terms of sensitivity, specificity, accuracy, or similar metrics for diagnostic or AI-driven devices.
    • Ground truth establishment, expert consensus, or multi-reader multi-case (MRMC) studies.

    Instead, this document pertains to a Class II surgical implant (bone fixation appliance) and its substantial equivalence to predicate devices. The "performance data" discussed is primarily non-clinical mechanical testing (four-point bend testing) to show equivalence in terms of physical properties and function.

    Therefore, I cannot extract the information requested as it is not present in the provided document. The questions you've asked are typically relevant for diagnostic devices, imaging systems, or AI/ML-driven medical software, not for a mechanical bone fixation plate.

    To summarize why the requested information is absent:

    • No AI/Algorithm: The device is a physical wrist fusion plate, not an algorithm or AI.
    • No "Performance" in the Diagnostic Sense: The "performance" for this type of device relates to mechanical strength, material compatibility, and similar physical attributes, not diagnostic accuracy or human reader improvement.
    • No Clinical Study for Equivalence: The document explicitly states "Clinical Tests: None provided as a basis for substantial equivalence." The equivalence is based on similar intended use, indications, materials, design features, and function, supported by non-clinical bench testing.

    If you have a document describing a diagnostic device or an AI-driven medical device, I would be happy to analyze it for the criteria you provided.

    Ask a Question

    Ask a specific question about this device