510(k) Data Aggregation

    K Number
    K200642

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-10

    (30 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthAlign Plus® system is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instructures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® system facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length discrepancy in Total Hip Arthroplasty.

    Example orthopedic surgical procedures include but are not limited to:

    • Total Knee Arthroplasty
    • · Total Hip Arthroplasty: Anterior/Posterior
    • · Unicompartmental Knee Arthroplasty: Tibial transverse resection

    The OrthAlign HipAlign® Sterile Pin Pack is indicated for use with the OrthAlign Plus® System.

    Intended Use: The OrthAlign® HipAlign® Sterile Pin Pack is intended to be used with the OrthAlign Plus® system specific to Total Hip Arthroplasty.

    Device Description

    The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display anqular and positional measurement changes in a triaxial format. The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes,

    The OrthAlign HipAliqn® Sterile Pin Pack is a set of sterile, single use pins used in conjunction with the OrthAlign Plus® system of instruments and electronics. The OrthAlign® HipAlign® Pin Pack includes:

    • . Pelvic Fixation pins: 4.0mm in diameter and 90mm to 140mm in length
    • Pelvic Fixation pins: 5mm in diameter and 170mm in length
    • . Femoral registration markers: 4.0mm to 4.5mm in diameter and 16mm to 25mm in length
    AI/ML Overview

    This document describes a Special 510(k) Premarket Notification for the OrthAlign HipAlign® Sterile Pin Pack, which is a modification to the previously cleared OrthAlign Plus® System (K171780). The focus of this notification is on the sterility and packaging of the pins and markers used with the system, rather than a a comprehensive AI/ML device. Therefore, the details requested in the prompt regarding AI/ML device performance, such as MRMC studies, training set ground truth establishment, and expert adjudication, are not applicable or available in this document.

    However, based on the provided text, we can describe the acceptance criteria and study that proves the device meets the acceptance criteria as it relates to the modification of the device (sterility and packaging).

    Device: OrthAlign HipAlign® Sterile Pin Pack, used with the OrthAlign Plus® System.

    Purpose of this Submission: To demonstrate substantial equivalence of the modified sterile, single-use pins and markers to the previously cleared non-sterile, reusable pins and markers.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this submission is for a modification primarily related to sterility and packaging, the acceptance criteria and performance are focused on these aspects, ensuring the modified components are as safe and effective as the predicate.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from document)Reported Device Performance
    Packaging IntegrityPackaging materials confirm intended use; verified and validated per procedures.Confirmed: "Performance testing of the packaging materials confirm that the OrthAlign® Sterile Pin Packs can be used according to its intended use. The packaging has been verified and validated according to OrthAlign's procedures for product design and development."
    Sterilization EfficacySterilization of the product is validated.Validated: "The sterilization of the product has been validated."
    Environmental RobustnessAbility to withstand environmental conditioning and distribution simulation.Demonstrated: "Performance testing included: Environmental conditioning and distribution simulation."
    Package Seal IntegrityMaintain integrity during visual inspection, dye penetration test, and peel strength testing.Demonstrated: "Package integrity testing (visual, dye penetration test, and peel strength testing)."
    Safety & Effectiveness EquivalenceModified pins and markers are as safe and effective as the predicate device.Demonstrated: "This testing demonstrates that the pins and markers are as safe and effective as the predicate device."
    Substantial EquivalenceModified device is substantially equivalent to the legally marketed predicate device for its intended use.Confirmed: "This testing regime demonstrates that the modified device is substantially equivalent to the legally marketed predicate device, for its intended use in detecting, measuring, and displaying angular and positional measurement changes in a triaxial format. The information provided by OrthAlign in this 510(k) notification confirms that the modified OrthAlign Plus® instruments are substantially equivalent to the predicate device, the OrthAlign Plus® System (K171780)."
    Shelf LifeEstablish a defined shelf life for the sterile product.Established: "7-year shelf life"

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for the performance testing (e.g., number of packages tested for integrity, number of sterilization cycles validated). The testing described relates to device components (packaging and sterilization), not patient data or clinical performance studies.

    • Sample Size: Not explicitly stated as numerical values for packaging and sterilization tests. It is implied that sufficient samples were used to meet validation standards for these types of tests.
    • Data Provenance: The data is generated from in-house laboratory performance testing conducted by OrthAlign, Inc., as part of their design and development procedures. This is neither retrospective nor prospective patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The "ground truth" here is compliance with engineering and quality standards for sterile medical devices (e.g., package integrity, sterilization efficacy). These are established through validated test methods and industry standards, not typically through human expert adjudication of medical images or diagnoses.


    4. Adjudication Method for the Test Set

    Not applicable. This type of submission does not involve adjudication of clinical data or expert consensus. It relies on objective engineering and microbiological test results.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for AI/ML diagnostic or assistive devices where human reader performance (e.g., radiologists interpreting images) is being augmented by AI. This submission is for surgical instrumentation components (pins and markers) and their sterility/packaging.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical surgical instrument set. It is not an algorithm, and therefore, standalone performance in the context of AI is irrelevant. The OrthAlign Plus® System itself is a "computer-controlled system intended to assist the surgeon," implying a human-in-the-loop system, but this specific submission focuses on the sterile pins and markers, not the performance of the computer-controlled system's algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for this specific submission's performance evaluation is based on:

    • Engineering Standards and Test Methods: For packaging integrity (e.g., ASTM standards for dye penetration, peel strength), and environmental conditioning.
    • Sterilization Validation Standards: For demonstrating the effectiveness of the sterilization process (e.g., ISO, AAMI standards for EtO sterilization).
    • Biocompatibility Standards: Implicit in the use of standard medical-grade materials.

    These are objective, quantitative criteria, not subjective expert consensus or clinical outcomes data in the traditional sense of diagnostic AI.


    8. The Sample Size for the Training Set

    Not applicable. There is no software algorithm or AI model being "trained" in this submission. The device is a physical product (sterile pins and markers).


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172462

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-22

    (100 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty. Ligament balancing is provided by the OrthAlign Plus® System in primary or revision Total Knee Arthroplasty.

    Example orthopedic surgical procedures include but are not limited to:

    • Total Knee Arthroplasty
    • Total Hip Arthroplasty: Anterior/Posterior
    • Unicompartmental Knee Arthroplasty: Tibial transverse resection
    • Ligament Balancing
    Device Description

    The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format. The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes. The OrthAlign Plus® System comprises a single use computer module and reusable instrumentation.

    AI/ML Overview

    The OrthAlign Plus® System is a computer-controlled surgical navigation system intended to assist surgeons in determining reference alignment axes during stereotactic orthopedic surgical procedures. It facilitates accurate implant positioning and aids in controlling leg length and offset discrepancies in Total Hip Arthroplasty (THA), and provides ligament balancing in primary or revision Total Knee Arthroplasty (TKA).

    Here's an analysis of the provided information regarding acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for accuracy. However, it does describe the performance testing conducted and concludes that the device performs "as well as the predicate devices." The key performance metric highlighted is accuracy in gap distance and angle measurement.

    Acceptance Criteria (Implied)Reported Device Performance
    Gap distance measurement accuracyBench testing with mechanical fixtures verified accuracy.
    Angle measurement accuracyBench testing with mechanical fixtures verified accuracy.
    Software functionality and reliabilitySoftware verification and validation ensured integrity, functionality, and reliability.
    Instrument mechanical requirementsSystem hardware verification/validation confirmed instruments meet mechanical requirements.
    BiocompatibilityAssessed per ISO 10993-1 (2009).
    Usability and user needsValidated in simulated use cadaver testing with an advising surgeon.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated with a numerical value. The document mentions "bench testing with mechanical fixtures" and "simulated use testing in cadaver." The number of mechanical tests or cadavers used is not specified.
    • Data Provenance: The cadaver testing implies a prospective experimental setting. The location of the test (e.g., country of origin) is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: One "advising surgeon" was involved in the simulated use cadaver testing.
    • Qualifications: "Advising surgeon" is the only qualification provided. Specific experience or subspecialty is not mentioned.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set in the traditional sense (e.g., 2+1, 3+1). For the cadaver study, an "advising surgeon" was present, suggesting a direct observation and evaluation of the system's performance in a simulated surgical environment. For bench testing, the ground truth would be objectively measured by the mechanical fixtures themselves.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described. The study focuses on evaluating the device's standalone performance and its equivalence to predicate devices, not on comparing human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done. "Bench testing with mechanical fixtures to verify gap distance and angle measurement accuracy" specifically evaluates the algorithm's ability to measure accurately without direct human interpretation or intervention affecting the measurement output. The software verification and validation also fall under standalone testing. The "simulated use testing in cadaver" does involve a human surgeon, but the primary focus is on the system's output and meeting user needs, rather than comparing human performance.

    7. The Type of Ground Truth Used

    • Bench Testing: Objective measurements from mechanical fixtures were used as ground truth for gap distance and angle measurement accuracy.
    • Simulated Use Testing: The "advising surgeon" would have implicitly provided a form of expert consensus/observation regarding the system's ability to meet user needs and usability in a simulated surgical environment. However, this is more for usability and functionality validation rather than a quantitative ground truth for accuracy.

    8. The Sample Size for the Training Set

    The document does not provide any information about a training set since this is a 510(k) summary for a medical device which is largely a hardware and software system, not an AI/Machine Learning model that undergoes "training" in the traditional sense of data-driven learning. While algorithms are involved, they appear to be rule-based or deterministic rather than machine learning algorithms requiring a distinct "training set."

    9. How the Ground Truth for the Training Set was Established

    As no training set is mentioned or implied for a machine learning model, the method for establishing ground truth for a training set is not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171780

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-11

    (26 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instructures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty.

    Example orthopedic surgical procedures include but are not limited to:

    • Total Knee Arthroplasty
    • · Total Hip Arthroplasty: Anterior/Posterior
    • · Unicompartmental Knee Arthroplasty: Tibial transverse resection
    Device Description

    The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format. The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.

    The OrthAlign Plus® System comprises a single use computer module and reusable instrumentation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the OrthAlign Plus System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria (e.g., "accuracy must be within X degrees"). Instead, it describes performance testing that "confirms that the OrthAlign Plus® System can be used according to its intended use" and "meets design input requirements." The performance is reported in terms of general accuracy and reliability.

    Acceptance Criteria (Implied)Reported Device Performance
    Integrity of software code and functionalitySoftware verification and validation ensured integrity, functionality, and reliability.
    System meets design input requirements (clinical utility)Customer requirements validated with an advising surgeon.
    Accuracy of leg length and offset measurementBench testing with mechanical fixtures and foam models verified leg length and offset measurement accuracy.
    Accuracy of scale reader measurementBench testing with mechanical fixtures and foam models verified updated scale reader measurement accuracy.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for the test set used in the performance studies. It mentions "bench testing with mechanical fixtures and foam models," which implies in-vitro testing rather than human subject data.

    • Sample Size (Test Set): Not specified (implied to be an unspecified number of mechanical fixtures and foam models).
    • Data Provenance: In-vitro (bench testing). No country of origin is specified, but the applicant is US-based. The testing is retrospective in the sense that it's performed on existing models, not prospective patient studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document mentions "Customer requirements validation with an advising surgeon," implying at least one expert was involved in validating clinical utility. However, it does not specify the number of experts for establishing ground truth for the accuracy tests (leg length/offset, scale reader). For the bench tests, the "ground truth" would be the known, precisely measured values of the mechanical fixtures and foam models, rather than expert consensus on anatomical structures.

    • Number of Experts: At least one "advising surgeon" for customer requirements validation.
    • Qualifications of Experts: "Advising surgeon" (specific qualifications like years of experience or subspecialty are not provided).

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the performance testing. For bench testing with mechanical fixtures, adjudication in the sense of reconciling differing expert opinions on a common case would not be applicable, as the expected values are mechanically determined. For the customer requirements validation, it's likely a qualitative assessment of whether the system met the surgeon's expectations.

    • Adjudication Method: Not applicable for accuracy testing; not specified for customer requirements validation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or described. The performance data focuses on system accuracy and validation against design requirements, not on the improvement of human readers (surgeons) with or without AI assistance.

    • MRMC Study Done: No.
    • Effect Size with AI vs. Without AI Assistance: Not applicable, as no MRMC study was performed.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    The performance testing described, particularly the "System accuracy testing: bench testing with mechanical fixtures and foam models to verify leg length and offset and updated scale reader measurement accuracy," represents a form of standalone testing of the algorithm's output against a known physical ground truth. While the overall system is "computer-assisted surgical navigation" (implying human-in-the-loop during actual surgery), the accuracy verification of its measurements on models is essentially a standalone evaluation of its core calculation capabilities.

    • Standalone Performance Done: Yes (accuracy testing on mechanical fixtures and foam models).

    7. The Type of Ground Truth Used

    • Ground Truth Type:
      • Mechanically determined values: For the "system accuracy testing" of leg length, offset, and scale reader measurements, the ground truth was derived from the precisely known measurements of mechanical fixtures and foam models.
      • Design input requirements/Clinical utility: For "customer requirements validation," the ground truth was the satisfaction of predefined design input requirements and the system's utility as assessed by an advising surgeon.

    8. The Sample Size for the Training Set

    The document primarily describes a device modification (Special 510(k)) and performance testing for that modification. It does not provide information about a "training set" for the algorithms. Given the nature of the device as a "computer-assisted surgical navigation system" that detects, measures, and displays angular and positional changes using inertial sensors, microcontrollers, and digital signal processors, it's less likely to involve a large-scale machine learning training dataset in the same way an image-based AI diagnostic device would. Its "algorithms" convert raw sensor outputs into spatial coordinates.

    • Sample Size (Training Set): Not specified.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is explicitly mentioned or described, the method for establishing its ground truth is also not provided. The system's operation appears to be based on physical principles and sensor data processing rather than learning from a large, annotated dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163379

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-02

    (91 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KneeAlign® system is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical structures during stereotactic surgical procedures. The KneeAlign system facilitates the accurate positioning of implants and instrumentation, relative to these alignment axes.

    Orthopedic surgical procedures include but are not limited to:

    • · Total Knee Arthroplasty
    • · Unicompartmental Knee Arthroplasty Tibial Transverse Resection
    Device Description

    The KneeAlign® 2 System is a non-invasive computer assisted surgical navigation system for use in knee arthroplasty procedures. The KneeAlign® 2 System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format.

    The KneeAlign® 2 System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.

    In total knee arthroplasty procedures, the device assists the surgeon in:

    • . Establishing the mechanical axis of the femur, determining the varus/valgus angle and the flexion/extension angle of the cutting block relative to the femur.
    • Establishing the mechanical axis of the tibia, determining the varus/valgus . angle and the posterior slope angle of the cutting block relative to the tibia.

    In unicompartmental knee arthroplasty procedures, the device assists the surgeon in:

    • Establishing the mechanical axis of the tibia, determining the varus/valgus . angle and the posterior slope angle of the cutting block relative to the tibia, for the transverse resection.
      The KneeAlign® 2 System comprises a single use computer module and reusable instrumentation.
    AI/ML Overview

    Acceptance Criteria and Study for KneeAlign 2 System

    This device, the KneeAlign 2 System, is a computer-assisted surgical navigation system. Based on the provided FDA 510(k) summary, the substantial equivalence to predicate devices (KneeAlign 2 System K103829 and OrthAlign Plus System K153237) is demonstrated through technological comparison and performance testing. The key addition to the subject device is the ability to navigate tibial transverse resections in unicompartmental knee arthroplasty.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA summary does not explicitly list quantitative "acceptance criteria" for accuracy performance in a tabular format with corresponding "reported device performance" values. Instead, it states that "System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy" was performed.

    For the purpose of this request, we can infer the acceptance criteria for the added functionality (tibial transverse resection in unicompartmental knee arthroplasty) by the assertion that the device is "as safe and effective as the predicate devices." The predicates were already cleared for their accuracy. Since the only performance data mentioned directly relates to this specific functionality, the "acceptance criteria" here is met by demonstrating the system's ability to accurately navigate these resections.

    Acceptance Criteria (Inferred)Reported Device Performance
    Navigated resection plane angular accuracy for tibial transverse resections in unicompartmental knee arthroplasty sufficient for intended use"System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy." The 510(k) states: "This testing regime demonstrates that the subject device is as safe and effective as the predicate devices. This testing regime demonstrates that the subject device is substantially equivalent to the legally marketed predicate devices, for its intended use in the accurate navigation of tibial resection planes in unicompartmental knee arthroplasty."
    Navigated resection plane depth accuracy for tibial transverse resections in unicompartmental knee arthroplasty sufficient for intended use(Same as above)
    System hardware meets mechanical requirements"System hardware verification/validation testing to ensure the instruments meet their mechanical requirements."
    Instrumentation cleaning, sterilization, and shipping validations met"Instrumentation cleaning, sterilization and shipping validations for the specified processes."
    Navigation device sterilization, packaging, shelf life, environmental conditions, and shipping validations met"Navigation device sterilization, packaging, shelf life, environmental conditions and shipping validations for the specified ranges of conditions involved in each process. (Summary data for the identical predicate device is referenced.)"
    System components biocompatibility meets ISO 10993-1 (2009)"System components biocompatibility assessment per ISO 10993-1 (2009)."
    Customer requirements / usability validation in cadaver with an advising surgeon meets design input requirements in a simulated use environment"Customer requirements / usability validation in cadaver with an advising surgeon to validate the system meets design input requirements for its functions in a simulated use environment."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions "bench testing with mechanical fixtures and foam models" for accuracy testing. It does not provide specific numerical sample sizes for these tests (e.g., number of resections, number of models). For the simulated use testing, it refers to a "prospective cadaver validation" but again, no specific number of cadavers is given.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but given the company (OrthAlign, Inc.) is based in Aliso Viejo, California, USA, and the submission is to the U.S. FDA, it is highly likely the testing was conducted in the USA.
      • Retrospective or Prospective: The cadaver validation was "prospective." The bench testing with foam models would also be considered prospective as it involves specifically designed tests for the purpose of validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: For the "Customer requirements / usability validation," it states "with an advising surgeon." This implies at least one surgeon was involved in the simulated use ground truth establishment.
    • Qualifications of Experts: The qualification mentioned is "surgeon." No further details on years of experience, sub-specialty, or board certification are provided in this summary.

    4. Adjudication Method for the Test Set

    • The document implies that the ground truth for "customer requirements/usability" was established by a single "advising surgeon" during the cadaver validation. There is no mention of an adjudication method involving multiple experts (e.g., 2+1, 3+1). For bench testing, the "ground truth" would be the known and controlled parameters of the mechanical fixtures and foam models.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a surgical navigation system, not an AI diagnostic tool primarily aimed at improving human reading proficiency for imaging. The performance testing focuses on the device's accuracy in assisting the surgeon with measurements and positioning during a procedure, not on radiologists interpreting images.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, in essence, standalone performance (in a controlled environment) was done for accuracy. The "System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy" evaluates the device's algorithmic and sensor performance independent of a live surgical scenario, under controlled conditions where the "ground truth" (correct angles and depths) is preset. The "human-in-the-loop" (a surgeon) is then involved in the usability and simulated use validation.

    7. The Type of Ground Truth Used

    • Bench Testing: The ground truth for angular and depth accuracy was established through the known parameters of "mechanical fixtures and foam models." This represents a controlled, engineered ground truth.
    • Simulated Use (Cadaver Validation): The ground truth was established by the "advising surgeon" determining if the system "meets design input requirements for its functions." This is a clinical expert consensus/assessment ground truth within a simulated environment.

    8. The Sample Size for the Training Set

    • This document is a 510(k) summary for a surgical navigation system, not an AI/machine learning device that typically requires a large 'training set'. The principles of operation are based on inertial sensors and calculation of spatial coordinates (not deep learning). Therefore, the concept of a "training set" for an algorithm to learn from data (as in AI) does not apply in the context of this device's validation as described. The device utilizes algorithms based on established physics and geometry.

    9. How the Ground Truth for the Training Set Was Established

    • As explained in point 8, the concept of a "training set" and its associated ground truth establishment is not applicable to the validation of this device as presented in the 510(k) summary. The algorithms are based on known mathematical principles for converting sensor outputs into spatial coordinates. Their accuracy is verified through empirical testing against known standards (bench testing), not "trained" on a dataset.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162962

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-01-23

    (91 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty.

    Example orthopedic surgical procedures include but are not limited to:

    • · Total Knee Arthroplasty
    • · Total Hip Arthroplasty: Anterior/Posterior
    • · Unicompartmental Knee Arthroplasty: Tibial transverse resection
    Device Description

    The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format. The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state numerical acceptance criteria in a table format, nor does it provide specific numerical performance metrics for the device. Instead, it describes "performance testing" and states that the "OrthAlign Plus® System can be used according to its intended use" and "performs as well as or better than the predicate devices."

    However, we can infer the tested capabilities and the reported validation of those capabilities.

    Acceptance Criteria (Implied)Reported Device Performance
    Accuracy of Leg Length and Offset Measurement (THA, Anterior)- Bench testing with mechanical fixtures and foam models verified leg length and offset measurement accuracy. - Simulated use testing in cadavers validated system accuracy vs. the gold standard of radiographic measurement. - 40 data points (on 5 hips) from cadaver validation showed accurate measurement of intraoperative change in leg length and offset.
    Accuracy of Acetabular Shell Insertion Navigation (THA, Anterior, relative to anterior pelvic plane adjusted for pelvic tilt)- Bench testing with mechanical fixtures and foam models verified updated scale reader measurement accuracy (likely contributing to this). - Simulated use testing in cadavers validated system accuracy vs. the gold standard of radiographic measurement. - 38 data points (on 5 hips) from cadaver validation showed accurate navigation of cup placement angles.
    Software Integrity and FunctionalitySoftware verification and validation ensured the integrity of the code and functionality and reliability of the software in various use sequences.
    Hardware Mechanical RequirementsSystem hardware verification/validation testing ensured the electronics hardware meets its mechanical requirements.
    Electrical SafetyElectrical safety testing to IEC 60601-1:2005 + Corr. 1 (2006) + Corr. 2 (2007) + AM1 (2012) or IEC 60601-1:2012.
    Electromagnetic CompatibilityElectromagnetic compatibility testing to IEC 60601-1-2: 2015.
    Laser Product SafetyLaser product safety verification to IEC 60825-1:2014.
    Instrumentation Cleaning, Sterilization, and ShippingValidations for the specified processes.
    Navigation Device Sterilization, Packaging, Shelf Life, Environmental, ShippingValidations for the specified ranges of conditions involved in each process. (Summary data for the identical predicate device is referenced for some validations.)
    BiocompatibilitySystem components biocompatibility assessment per ISO 10993-1 (2009).
    User Requirements Validation (Simulated Use)Customer requirements validation with an advising surgeon validated the system meets design input requirements for its functions in a simulated use environment.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (for clinical performance aspects):
      • Intraoperative change in leg length and offset: 40 data points (on 5 hips)
      • Acetabular shell insertion navigation: 38 data points (on 5 hips)
    • Data Provenance: Prospective cadaver validation in a simulated operating room environment. The country of origin is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: A "surgeon" conducted the procedures in the simulated use cadaver testing. It mentions "an advising surgeon" for customer requirements validation. The exact number of independent experts contributing to ground truth establishment for the cadaver study is not explicitly stated beyond "a surgeon."
    • Qualifications of Experts: The individual conducting the cadaver procedures is referred to as a "surgeon." No specific years of experience or subspecialty (e.g., orthopedic surgeon) are mentioned, though the context implies an orthopedic surgeon familiar with the procedures.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set in the sense of multiple reviewers resolving discrepancies. The ground truth (radiographic evaluation) was compared against the device's measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed or described. This device is a surgical navigation system, not an AI-assisted diagnostic tool that aids human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the cadaver study described is a standalone performance test of the algorithm and system's accuracy in a simulated surgical environment. The "surgeon" conducts the procedure using the device, and the device's output (measurements and navigation guidance) is then compared to a ground truth (radiographic evaluation). This assesses the system's ability to accurately determine surgical parameters independently, under realistic conditions.

    7. The Type of Ground Truth Used

    • Radiographic measurement: Used for validating the accuracy of leg length and offset changes, and cup placement angles in the hip arthroplasty procedures on cadavers. This is considered a "gold standard" in the context of the study.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" or "training data" as this device is a rule-based surgical navigation system rather than a machine learning/AI model that requires training data in the conventional sense. The "performance data" section focuses on verification and validation of the system's design and functionality.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed or implied for a machine learning model, this question is not applicable based on the provided text. The system relies on its inherent algorithms and sensor data, validated against known physical principles and clinically accepted measurement techniques (like radiography) rather than being "trained" on a dataset with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153237

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-03-01

    (113 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.

    Example orthopedic surgical procedures include but are not limited to:

    • · Total Knee Arthroplasty
    • · Total Hip Arthroplasty: Anterior/Posterior
    • · Unicompartmental Knee Arthroplasty: Tibial transverse resection
    Device Description

    The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format.

    The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.

    The OrthAlign Plus® System comprises a single use computer module and reusable instrumentation.

    AI/ML Overview

    The provided text describes modifications to an existing orthopedic surgical navigation system, the OrthAlign Plus® System (K153237), to add functionality for unicompartmental knee arthroplasty (UKA) tibial transverse resections. The submission claims substantial equivalence to its prior version (K140331) and the Aesculap OrthoPilot Next Generation (K141694).

    However, the document does not contain specific acceptance criteria with numerical targets (e.g., minimum accuracy of X degrees or Y mm) or detailed results from a study demonstrating the device meets those criteria. Instead, it describes general categories of performance testing.

    Therefore, I cannot populate the table with acceptance criteria and reported device performance as they are not explicitly stated with specific numerical values in the provided text. I also cannot provide detailed answers for many of the questions as the specific study details (e.g., sample sizes for test/training, expert qualifications, adjudication methods, MRMC studies) are not present.

    Here's a breakdown of what can be extracted and what information is missing:


    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria (Specific Targets)Reported Device Performance (Specific Results)
    Not explicitly stated in the documentNot explicitly stated with numerical results in the document
    Example: Navigated resection plane angular accuracy: within X degrees of targetExample: Achieved Y degrees of angular accuracy for navigated resection plane
    Example: Navigated resection plane depth accuracy: within Z mm of targetExample: Achieved W mm of depth accuracy for navigated resection plane

    Missing Information: The document states "System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy." However, it does not provide the specific acceptance criteria (i.e., "within X degrees/mm of target") nor the actual measured performance results.


    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not specified. The document mentions "bench testing with mechanical fixtures and foam models" for system accuracy and "cadaver with an advising surgeon" for usability validation, but no sample sizes for these tests are provided.
    • Data Provenance: The cadaver study would typically be prospective for that specific test run, but the location/country of origin is not mentioned. Bench testing is laboratory-based.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: "an advising surgeon" (singular) for the cadaver study.
    • Qualifications of Experts: The qualification is "advising surgeon." No further details on experience level or specialization are provided.

    4. Adjudication method for the test set

    • Adjudication Method: Not specified. For the cadaver study, an "advising surgeon" was involved in validating usability, but the method for establishing ground truth or resolving discrepancies is not detailed. For bench testing, mechanical fixtures and foam models would likely have a pre-defined "ground truth" based on their design.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study involving human readers or AI assistance in that context was not mentioned or performed. The device is a surgical navigation system, not an imaging interpretation AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The described "System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy" can be considered a form of standalone testing for the device's accuracy in measuring and displaying angles/positions. However, it's not a purely "algorithm-only" test as it involves hardware components interacting with physical models. The document doesn't explicitly delineate "algorithm-only" performance metrics separate from the integrated system.

    7. The type of ground truth used

    • Ground Truth Type:
      • For system accuracy testing, the ground truth would be based on the known, precise measurements of the mechanical fixtures and foam models.
      • For usability validation in cadaver, the "ground truth" relates to whether the system meets design input requirements for its functions, likely assessed by the advising surgeon against surgical standards.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable/not specified. This device is a surgical navigation system, which primarily relies on algorithms and hardware for real-time measurements rather than a "training set" in the machine learning sense for image recognition or similar AI applications. The algorithms are developed and calibrated, but a "training set" for performance evaluation is not referenced.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable/not specified. As above, the concept of a "training set" for this type of device is not directly addressed in the document. The algorithms would be designed and tested based on engineering principles and physical measurement accuracy.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1