Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K231753

    Validate with FDA (Live)

    Date Cleared
    2024-02-21

    (251 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Oneday Implant Abutment is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implants with diameters larger than 5 mm are intended to be used in the molar region.

    Device Description

    The Oneday Implant Abutment is a material for dental surgery and is an abutment used to support and maintain prosthetic restored teeth in case of partial or total loss of teeth. It is used in combination with a fixture implanted in the jawbone.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Oneday Implant Abutment," a dental device. The primary purpose of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, meaning it has the same intended use and the same technological characteristics, or if there are differences, those differences do not raise new questions of safety and effectiveness.

    The provided text details the device description, indications for use, comparison to predicate devices, and non-clinical test data. It is crucial to understand that this document does NOT describe the acceptance criteria and a study proving a device meets those criteria in the context of clinical performance or AI/software validation. Instead, it describes mechanical, material, and biocompatibility testing for a physical dental implant component to demonstrate its substantial equivalence to an existing device.

    Therefore, many of the requested items (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth establishment for AI systems, training set sample size) are not applicable to this type of device submission. This is a traditional medical device submission, not a submission for a software-as-a-medical-device (SaMD) or an AI/ML-driven device that would involve performance metrics related to diagnostic accuracy or clinical outcomes.

    Here's an analysis of the provided information relative to your request:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria here are standards for mechanical performance, biocompatibility, and sterilization, rather than performance metrics for an AI algorithm.

    Test TypeAcceptance Criteria (Standard Met)Reported Device Performance
    Fatigue TestingISO 14801:2016 (Implies meeting the fatigue strength requirements defined by the standard for dental implants/abutments)"The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device." Specifically, "The Fatigue Testing was performed under worst case scenario according to ISO 14801." (The exact quantitative performance is not provided in a summary table but is stated to meet the standard.)
    End User Sterilization ValidationANSI/AAMI ST79, ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2, and ISO 11138-1 (Implies demonstrating that the proposed sterilization method effectively sterilizes the device and maintains its properties)"The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device." "The end user sterilization test was performed on abutment that is not cleared but it can be leveraged for the subject device because the product category, material, manufacturing process, facility, and packaging of the both products are exactly same."
    Biocompatibility TestingISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010, and ISO 10993-11:2006 (Implies demonstrating the device is safe for biological contact according to the specified parts of the ISO 10993 series)"The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device." "The Biocompatibility Test was conducted on abutment made of Ti-6AI-4V ELI and leveraged for the subject device because both products are manufactured with same materials and manufacturing process. It demonstrates that the subject device is biocompatible and substantial equivalence with the predicate."
    Dimensional ToleranceWithin ± 1% range (for stated dimensions of various components like abutments, cylinders, etc.)Not explicitly stated as a 'reported performance' for the device itself, but assumed to be met for the manufactured products as part of quality control and design specifications. The document lists dimensions but not measured ranges.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for the non-clinical tests (e.g., number of abutments tested for fatigue, sterilization, or biocompatibility). This information is typically detailed in the full test reports referenced in the submission, not summarized in the 510(k) summary letter.
    • Data Provenance: The tests are non-clinical (laboratory-based) performed on "subject device" or "worst-case test article" samples. There is no patient data involved for these tests. The country of origin for the data generation (the testing laboratories) is not specified in this summary. These are prospective tests performed specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable. The "ground truth" for mechanical testing, sterilization, and biocompatibility is based on established ISO and ANSI standards. There are no "experts" establishing a "ground truth" in the clinical imaging or diagnostic sense. The evaluations are objective measurements against predefined thresholds.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. Adjudication methods are typically used in clinical studies, particularly for subjective assessments (e.g., image interpretation). These are objective non-clinical tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device is a physical dental implant abutment, not an AI-powered diagnostic or assistive tool. No MRMC study would be performed for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. There is no algorithm or software for which standalone performance would be relevant for this physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests, the "ground truth" is defined by the objective pass/fail criteria specified by the referenced ISO/ANSI standards for mechanical properties, sterility assurance levels, and biocompatibility endpoints. There is no clinical "ground truth" in the sense of pathology or outcomes data for this specific submission, as it relies on non-clinical testing for substantial equivalence.

    8. The sample size for the training set:

    • This is not applicable. This is a physical medical device, not a software/AI product requiring a training set for model development.

    9. How the ground truth for the training set was established:

    • This is not applicable. See point 8.

    In summary: The provided document is a 510(k) summary for a physical medical device, the Oneday Implant Abutment. It demonstrates substantial equivalence primarily through comparisons of design, materials, intended use, and adherence to established performance standards through non-clinical (laboratory) testing. It does not involve AI/ML technology or associated validation studies typically seen in software medical device submissions, and therefore, many of the detailed questions regarding AI performance criteria and study design are not relevant to this specific document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222946

    Validate with FDA (Live)

    Date Cleared
    2023-08-23

    (330 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oneday Mini Implant System is intended for two-stage surgical procedures in the following situations and with the following clinical protocols:

    • The intended use for the 3.0mm, 3.3mm diameter Oneday Mini Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.

    • Immediate placement in extraction situations with a partially or completely healed alveolar ridge.

    • It is intended for delayed loading.

    Device Description

    The Oneday Mini Implant System is used to replace missing teeth in various situations ranging from a single tooth loss to the complete loss of incisors teeth. This system is restricted to substitute the maxillary lateral incisors and mandibular incisors. It is two stage endosseous screw type implant with internal hexagonal connection. This system consists of the fixture, cover screw, and various abutments. Only the subject abutments can be used with the subject fixtures. The Fixture is made of Pure Titanium of ASTM F67 and the surface of the fixture is treated with the SLA(Sand-blasted, Large grit, Acid-etched surface). Fixture is provided sterile.

    AI/ML Overview

    This document is a 510(k) summary for the Oneday Mini Implant System, primarily focused on demonstrating substantial equivalence to a predicate device (UF(II) Narrow Implant System by DIO Corporation, K161987) rather than proving performance against specific acceptance criteria for an AI/ML device.

    Therefore, many of the requested details related to AI/ML device acceptance criteria, study design (like MRMC, training/test set ground truth, expert adjudication), and performance metrics (e.g., AUC, sensitivity, specificity) are not present in this document because it is for a traditional dental implant device, not an AI/ML powered device.

    However, I can extract the information relevant to a traditional medical device's acceptance criteria and performance proof, focusing on the equivalence to a predicate device.


    Acceptance Criteria and Study for the Oneday Mini Implant System (A Traditional Medical Device)

    1. Table of Acceptance Criteria and Reported Device Performance:

    For a traditional medical device like a dental implant, "acceptance criteria" are typically demonstrated through various non-clinical tests (mechanical, material, sterilization, biocompatibility) and comparison to a legally marketed predicate device. The performance is assessed by showing substantial equivalence to this predicate, meaning it is as safe and effective as the predicate.

    Acceptance Criteria CategorySpecific Criteria (Implied by Substantial Equivalence Documentation)Reported Device Performance/Comparison to Predicate
    Indications for UseMatch the predicate device's Indications for Use.Met: "The Oneday Mini Implant System is intended for two-stage surgical procedures in the following situations and with the following clinical protocols: - The intended use for the 3.0mm, 3.3mm diameter Oneday Mini Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors. - Immediate placement in extraction situations with a partially or completely healed alveolar ridge. - It is intended for delayed loading." This is identical to the predicate's indications.
    Design CharacteristicsSimilar fundamental design, with acceptable differences that do not raise new questions of safety or effectiveness.Met (with minor non-impactful differences): The subject device and primary predicate have the same characteristics such as indications for use, diameter, length, surface treatment, material, abutment connection, and sterilization method. The only stated difference is "fixture design", which the submitter claims "doesn't impact product's fundamental technologies." Minor dimensional differences are noted for some abutments (e.g., Cover Screw, Cemented Abutment, Healing Abutment, Abutment Screw), but these are deemed not to impact fundamental technologies.
    Material CompositionSame or equivalent biocompatible materials as the predicate.Met: Fixture: Pure Titanium of ASTM F67 (same as predicate). Other components (Cover Screw, Cemented Abutment, Solid Abutment, Angled Abutment, Healing Abutment, Abutment Screw): Ti-6Al-4V ELI (same as predicate).
    Surface TreatmentSame or equivalent surface treatment.Met: SLA (Sand-blasted, Large grit, Acid-etched surface) for fixtures (same as predicate). Surface roughness, surface composition analysis, and SEM imaging were provided (leveraged from K192294) to demonstrate substantial equivalence of the surface modification.
    Sterilization MethodSame or equivalent sterilization method with validation data.Met: Fixtures: Gamma Irradiation (same as predicate). Abutments: End User Sterilization (same as predicate). Validation for Gamma Sterilization (fixtures) was leveraged from K192294. Validation for End User Sterilization (abutments) was performed on the subject healing abutment and met standards (ANSI/AAMI ST79, ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2, and ISO 11138-1).
    BiocompatibilityBiocompatible materials and design, demonstrated through testing or leveraging predicate data.Met: Biocompatibility testing for both fixture and abutment materials was leveraged from predicate devices (K192294) because the materials and manufacturing processes are the same. Standards included ISO 10993-1, -3, -5, -6, -10, -11.
    Mechanical Performance (Fatigue)Meet established mechanical performance standards (e.g., ISO for dental implants).Met: Fatigue Testing performed under worst-case scenario according to ISO 14801:2016. Results met criteria.
    Shelf-LifeDemonstrated stability and integrity over the claimed shelf-life.Met: Shelf-Life Test on Fixtures leveraged from K192294 (ASTM F1980), as materials, sterilization, packaging, and manufacturing process are the same.
    Bacterial EndotoxinMeet limits for Bacterial Endotoxins.Met: Bacterial Endotoxin Test Report on Fixtures leveraged from K192294 (ANSI/AAMI ST72:2011, USP <161>, and USP <85>).
    MRI Safety (if applicable)Assessment of safely in MRI environments.Met: Non-clinical worst-case MRI review performed using scientific rationale and published literature, addressing magnetically induced displacement force and torque per FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment". The assessment covered all compatible implant bodies, dental abutments, and fixation screws, and material compositions.

    2. Sample Size for the Test Set and Data Provenance:

    • Test Set (for performance evaluation): For a traditional device showing substantial equivalence, there isn't a "test set" in the sense of a distinct dataset used to calculate performance metrics like sensitivity/specificity for an algorithm. Instead, the "testing" involves performing specific engineering and biological tests (e.g., fatigue testing on a sample of implants, sterilization validation on a sample of devices) and comparing the design specifications and materials to a predicate device.
      • Fatigue Testing: "Fatigue Testing under the worst-case scenario according to ISO 14801:2016." The standard dictates sample sizes.
      • End User Sterilization Validation: Performed on the "subject Healing Abutment Mini with the greatest surface area" to be representative. Standards dictate sample sizes.
      • Data Provenance: The document does not specify the country of origin for the testing data. Test results are generally from controlled laboratory environments using product samples. Most data is "leveraged" from previous 510(k) clearances (K192294 and K161987) for similar devices, suggesting a retrospective use of existing data to support the current claim of equivalence.

    3. Number of Experts and Qualifications for Ground Truth:

    • Not applicable in the context of this 510(k) summary for a traditional medical device. Ground truth for device performance is established through adherence to recognized international standards (e.g., ISO, ASTM) for material properties and mechanical testing, and through comparison to a device already cleared by the FDA (the predicate). There is no "expert panel" establishing a clinical ground truth for a diagnostic output as there would be for an AI/ML imaging device.

    4. Adjudication Method for the Test Set:

    • Not applicable. There is no "adjudication" of results in the sense of reconciling differing expert opinions, as this is not an AI/ML diagnostic device with human readers. Testing results are objective measurements against pre-defined engineering and biological standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic devices, particularly AI/ML-powered ones, to assess the impact of the AI on human reader performance. This document is for a physical medical implant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No, a standalone evaluation was not done. This concept applies to AI/ML algorithms that might provide a diagnostic output without human intervention. This document is for a physical medical implant.

    7. The type of ground truth used:

    • The "ground truth" for the Oneday Mini Implant System is based on:
      • Adherence to recognized international standards: e.g., ISO 14801:2016 for fatigue, ISO 11137 for sterilization, ISO 10993 for biocompatibility, etc.
      • Demonstrated equivalence to a legally marketed predicate device: The predicate device (UF(II) Narrow Implant System, K161987) has already been determined to be safe and effective by the FDA. The substantial equivalence argument posits that since the new device is fundamentally similar and meets performance parameters, it is also safe and effective.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set" as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1