510(k) Data Aggregation

    K Number
    K191179

    Validate with FDA (Live)

    Device Name
    Myo Plus
    Date Cleared
    2019-09-04

    (125 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Myo Plus is to be used exclusively for exoprosthetic fittings of the upper limbs.

    Device Description

    MYO PLUS measures the electromyographic control signals of the user and assigns them to the movements of the prosthesis, such as the hand or rotation unit. The product is the central control unit of a myoelectric prosthesis. By calibrating via the Myo Plus App, the control unit learns to assign muscle signals from the user to the various types of movement. This calibration can be performed by the users themselves, and repeated at regular intervals if necessary.

    Description of the components:
    Myo Plus TR
    The central control unit is installed in the forearm of the prosthesis and is supplied with energy by the battery system of the prosthesis. Up to 8 remote electrodes can be connected. The electronics handles the signal processing of the electrode signals and controls the compatible prosthesis components.

    Remote Electrode
    The remote electrodes take the muscle signal from the skin surface and transmit it to MyoPlus TR. The electrode contacts are located on a flexible cable and are individually screwed into the socket by the prosthesist.

    Myo Plus App
    The Myo Plus App is the central interface for the user and/or professionals to make settings on the control system or monitor operating states. The data transfer to mobile setting systems on Android or mobile iOS devices shall be carried out by a wireless connection according to the state of the art.

    Mvo Cuff
    The Myo Cuff is used to evaluate the muscle signals of a potential user of a Myo Plus controller. The Myo Cuff forms a flexible wristband which is temporarily applied to the user's forearm. Thus, the signal patterns of the user can be evaluated with the help of the app without the preparation of a complex test socket. If a user has problems separating the patterns at the beginning, he can also use the Myo Cuff for practicing with an professional.

    Myo Plus trial kit
    The Myo Plus trial kit provides the components of Myo Plus as a loaner kit for the prosthesist.

    AI/ML Overview

    The provided text describes the Myo Plus device and its substantial equivalence to a predicate device, but it does not contain specific details about acceptance criteria or a study that proves the device meets those criteria in the context of a comparative effectiveness study involving AI assistance for human readers or standalone algorithm performance.

    The document primarily focuses on demonstrating substantial equivalence based on:

    • Identical Indications for Use
    • Identical Principles of Operation
    • Similar Technological Characteristics
    • Compliance with recognized safety and performance standards (electrical safety, EMC, biocompatibility, battery safety)
    • Results from general functional and performance tests.

    There is no mention of an AI component, human readers, effect sizes, multi-reader multi-case studies, or the types of ground truth typically established for AI-driven diagnostic or assistive devices. The "Pattern recognition" mentioned refers to the device's ability to interpret muscle signals for prosthetic control, not image analysis or AI interpretation of medical data in a diagnostic context.

    Therefore, I cannot extract the requested information from the provided text for the following reasons:

    1. Acceptance criteria and reported device performance related to AI/human reader studies: Not present. The acceptance criteria mentioned are related to functional performance (e.g., "Functional testing: Passed," "Pattern recognition: Passed") and compliance with safety standards.
    2. Sample size for test set and data provenance: Not mentioned.
    3. Number of experts and their qualifications for ground truth: Not mentioned.
    4. Adjudication method for test set: Not mentioned.
    5. MRMC comparative effectiveness study and effect size: No such study is described.
    6. Standalone (algorithm only) performance: No such performance evaluation is described. Myo Plus is a physical device (control unit for prosthetic limbs) that uses signal processing, not a standalone AI algorithm for medical image analysis.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for the device's function would be the user's intended muscle movements and the device's accurate interpretation and execution of those movements, but this is not defined in terms of a clinical study or labeled dataset.
    8. Training set sample size and how ground truth was established for training: Not mentioned. The device "learns to assign muscle signals from the user to the various types of movement" through a calibration process via the Myo Plus App, which can be done by users themselves or with a professional. This is a user-specific calibration, not a large-scale AI model training process using a distinct training dataset.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141812

    Validate with FDA (Live)

    Date Cleared
    2015-04-02

    (269 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease.

    During gait, the stimulation system sends electric stimuli to muscles in the affected leg, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.

    Device Description

    The MyGait® Stimulation System is a battery-operated two-channel stimulator and was developed exclusively for everyday use during walking in an everyday environment.

    The patient wears a wireless heel switch in a special sock. If the user is not able to put sufficient weight on the affected leg, he or she can wear the heel switch on the other foot. The heel switch registers heel lift and ground contact and thus the swing and stance phase. It transmits the information to the stimulator wirelessly.

    The stimulator is located in a cuff that holds the electrodes. It is easy to put on, even with only one hand. The correct placement of the electrodes is achieved by the cuff. The stimulator delivers the electrical stimulation to the common peroneal nerve. The nerve stimulates for controlled dorsiflexion of the foot during the swing phase.

    The second channel can be used independently of the first channel to provide stimulation to the knee flexors or knee extensors in patients with thigh muscle weakness.

    Using the wireless remote control, the patient can control and adjust the stimulator settings. When the patient sits for an extended period, the sleep mode helps save energy.

    The adjustment tool enables qualified personnel to adjust the stimulation parameters to the patient's individual needs. Stimulation parameters are used to affect movements in the foot. Examples of stimulation parameters are current, pulse width and shape, frequency, stimulation timing,

    The "MyGait Soft" PC software is used to manage and analyze gait analysis data and stimulation parameters. The data is loaded from the stimulator to the adjustment tool. The stimulation parameters can be restored to the stimulator from the PC.

    AI/ML Overview

    The Otto Bock Healthcare Products GmbH MyGait Stimulation System, referenced in K141812, did not report a clinical study on human subjects to determine device performance against acceptance criteria. Instead, substantial equivalence was established by comparing the technical characteristics and indications for use of the MyGait Stimulation System to three legally marketed predicate devices: WalkAide (K123972), Empi 300pv (K021100), and NESS L300 Plus System (K103343). The document primarily focuses on demonstrating that the new device's specifications fall within the range or are similar to those of the predicate devices, and that it complies with relevant electrical safety and electromagnetic compatibility standards.

    Therefore, the following information cannot be fully provided as a traditional performance study was not conducted or reported in this 510(k) summary:

    • A table of acceptance criteria and the reported device performance
    • Sample size used for the test set and the data provenance
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    • Adjudication method for the test set
    • Multi reader multi case (MRMC) comparative effectiveness study findings
    • Standalone (i.e. algorithm only without human-in-the loop performance) findings
    • Type of ground truth used
    • Sample size for the training set
    • How the ground truth for the training set was established

    However, based on the provided document, here's a summary of the technical specifications and how they align with the substantial equivalence claim:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Since this is a 510(k) for substantial equivalence based on technical comparison, there isn't a direct "acceptance criteria" table from a performance study. Instead, the "acceptance criteria" are implied by the ranges and characteristics of the predicate devices. The "reported device performance" are the MyGait's technical specifications.

    Feature / Criterion (Implied by Predicate Devices)MyGait Stimulation System Reported Performance
    Indications for UseAnkle dorsiflexion or extension; thigh muscle weakness; upper motor neuron injury/disease; improve gait; prevent/retard atrophy; facilitate muscle reeducation; maintain/improve range of motion; promote local blood circulation.
    Power SourceBuilt-in Lilon battery (3.7V, 1100mAh)
    Patient Leakage Current (normal condition)< 1 µA (tested per IEC 60601-1)
    Patient Leakage Current (single fault condition)< 1 µA (tested per IEC 60601-1)
    Number of Output Channels2
    Output ModesSynchronous or alternating
    Control BasisMicroprocessor controlled, constant current
    Automatic Overload TripYes
    Automatic No-Load TripYes
    Patient Override ControlON/OFF Button
    Indicator Display (On/Off Status)Yes (LED)
    Indicator Display (Low Battery)Yes (LED)
    Indicator Display (Current Level)Displayed on remote control
    Timer Range5 to 120 minutes (in 5-min steps)
    AccessoriesWireless foot switch (with failsafe), wireless remote control
    Stimulation Parameter SetupWireless via MyGait Adjustment Tool
    WaveformBiphasic symmetrical or asymmetrical
    Waveform ShapeRectangular
    Max Output Voltage (@ 500Ω)45V +/- 10%
    Max Output Voltage (@ 1kΩ)90V +/- 10%
    Max Output Voltage (@ 2kΩ)120V +/- 10%
    Max Output Current (@ 500Ω)90mA (+/- 10%)
    Max Output Current (@ 1kΩ)90mA (+/- 10%)
    Max Output Current (@ 2kΩ)60mA (+/- 10%)
    Pulse Width (Symmetrical)50 to 400µs (in 50µs steps)
    Pulse Width (Asymmetrical, positive)50 to 400µs (in 50µs steps)
    Pulse Width (Asymmetrical, negative)200 to 1600µs (in 100µs steps)
    Frequency10 to 80Hz (in 1Hz steps)
    Net Charge0µC (balanced pulses)
    Max Pulse Charge (@ 500Ω)36µC
    Max RMS Current Density (@ 500Ω)1.51 mA/cm²
    Max Power Density (@ 500Ω)0.024 W/cm²
    On Time (training mode)1 to 4 sec (in 0.1sec steps)
    Off Time (training mode)1 to 30 sec (in 1 sec steps)

    2. Sample size used for the test set and the data provenance
    No clinical test set of human subjects was reported for performance evaluation against acceptance criteria. The evaluation was based on a comparison of technical specifications to predicate devices and compliance with international standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable, as no human subject test set requiring ground truth establishment by experts was reported.

    4. Adjudication method for the test set
    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a functional electrical stimulator, not an AI-assisted diagnostic or imaging device for which MRMC studies would typically be conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. The device is a physical stimulator used on a patient. Its functionality involves direct interaction with the human body and is not an "algorithm only" device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable. The ground truth for this submission is based on engineering specifications and compliance with recognized international safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-2).

    8. The sample size for the training set
    Not applicable, as this is not a machine learning/AI device requiring a training set in that context. The device's operation is based on established electrical stimulation principles, not learned from data.

    9. How the ground truth for the training set was established
    Not applicable.