Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K112269

    Validate with FDA (Live)

    Date Cleared
    2011-10-20

    (73 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSTAR USB Blood Pressure Monitor Upper Arm Type : P200 and Wrist Type : M100, are noninvasive blood pressure measurement systems intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist or upper arm.

    The cuff circumference is limited to be 5.5'~7.8'(14 cm ~20 cm) for wrist type, and 9'~13'(24cm ~32cm) for Upper Arm Type.

    Device Description

    Basically the measuring system were composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at arm, pneumatic pump, inflation and deflation system, housing, display LCD, and measuring software ... and so on.

    OSTAR USB Blood Pressure Monitor Upper Arm Type : P200 and Wrist Type : M100 use the Oscillometric method to measure blood pressure. The Oscillometric method is adopted clinically to measure the blood pressure recently. It is not needed to ise the stethoscope, as in the traditional measuring method, to monitor the Korothkov sound when deciding the systolic and diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating stop decreasing. In addition, the heart beat rate will be also measured and displayed on LCD as part of measurement result.

    For P200 Upper Arm Type, all the measuring circuit, measuring devices, measuring and display software are all mounted on the device housing. It can be operated individually or be operated by the PC installed with measurement software through USB connection port to control and transfer the measuring records, and then stored the records in device.

    For M100 Wrist Type. all the measuring circuit, measuring devices, software are all mounted on the device housing except for the measurement control keys and display. It can be operated only by the PC installed with measurement software through USB connection port to control and transfer the measuring records, and then stored the records in device.

    AI/ML Overview

    The provided text describes the Ostar USB Blood Pressure Monitor and details its classification, intended use, and the non-clinical tests performed to demonstrate its substantial equivalence to predicate devices. However, it does not explicitly state acceptance criteria or provide a study with detailed performance metrics against those criteria. The document states that the device is compliant with certain standards, including ANSI/AAMI, SP 10-2002/A 1:2003, for blood pressure monitors. These standards typically contain detailed accuracy requirements that would serve as acceptance criteria. Without explicit mention of those criteria in the document or the results against them, I can only indicate general compliance.

    Here's the information based on the provided text, with limitations noted due to the absence of specific acceptance criteria and detailed study results:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states compliance with ANSI/AAMI SP 10-2002/A 1:2003. This standard dictates accuracy requirements for automated sphygmomanometers.

    • Acceptance Criteria (Based on ANSI/AAMI SP 10-2002/A 1:2003 for automated sphygmomanometers) - Inferred, not explicitly stated in the document:

      • Mean difference between device and reference standard: ≤ ±5 mmHg
      • Standard deviation of the difference: ≤ 8 mmHg
    • Reported Device Performance:
      The document states: "verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device."
      This broadly confirms that the device meets the necessary performance standards, including those for accuracy as defined by ANSI/AAMI SP 10-2002/A 1:2003, but specific numerical results are not provided in this summary.

    Acceptance Criteria (Inferred from ANSI/AAMI SP 10-2002/A 1:2003)Reported Device Performance (Summary Statement)
    Mean difference between device and reference standard ≤ ±5 mmHg; Standard deviation of the difference ≤ 8 mmHg"Verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device." (Indicates compliance with relevant standards)

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It generally refers to "verification and validation tests."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number of experts or their qualifications for establishing ground truth. For blood pressure measurements, ground truth would typically come from auscultatory measurements by trained observers using a reference sphygmomanometer, often with mercury.

    4. Adjudication method for the test set

    The document does not describe any adjudication method used for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a blood pressure monitor, not an AI-assisted diagnostic imaging device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the blood pressure monitor is a standalone device that measures blood pressure using an oscillometric method. Its performance is inherent to the device's algorithms and hardware, without requiring human interpretation for the measurement itself.

    7. The type of ground truth used

    The type of ground truth for blood pressure monitors, while not explicitly stated in the document, is universally established by comparison to a reference standard, typically auscultatory measurements performed by trained clinicians using a mercury sphygmomanometer or an equivalent validated non-invasive method.

    8. The sample size for the training set

    The document does not provide information about a "training set." This term is usually associated with machine learning or AI models, which is not directly applicable to the functional description of this blood pressure monitor. The oscillometric method is an established physiological measurement technique.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" in the context of an AI/ML model, this question is not applicable. The device relies on established oscillometric principles and internal calculations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050680

    Validate with FDA (Live)

    Date Cleared
    2005-05-13

    (58 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ostar A2, and K7 are indicated for trained medical professionals to operate on the patient at clinical settings to collect and measure blood pressure, pulse rate noise index and pressure waveform diagram (including heartbeat waveform and noise waveform). In addition, the Ostar P2 and X2 are indicated for patients at home who are capable and willing to self-administrate this device upon prescription of their healthcare provider. P2 and X2 do not display the measuring waveform.

    The devices do not send any real-time alarms. Clinical judgment and experience are required to check and interpret the information delivered.

    Device Description

    Basically the measuring system were composite of blood pressure measuring circuit via Dacillometric method, pressure sensor, measuring cuff at arm, pneumatic pump, inflation and deflation system, housing, display LCD, and measuring software ... and so on.

    The main operation for the blood pressure measurement is carried out in such a way that the measuring cuff at arm is inflated to the estimated pressure level, then deflated to zero measuring our at unfine inflation and deflation, the pressure change with respective of time were recorded as the data base of measurement. Then the following measuring results will be calculated against the measurement data base :

    • Blood pressure information including systolic and diastolic pressure (calculated via Oscillometric method)
    • Heart beat rate.
    • Heart beat noise calculated via FFT ( Fast Fourier Transformation).

    For the display of measuring results and operation interface, the following difference were provided for the following different models :

    • For model A2, RS 232 is provided for the connection with PC for the measurement operation and display of the whole measuring pressure waveform, noise waveform, systolic and diastolic pressure, heat beat rate, and noise index.
    • For model K7, the operation and display details are completely identical to that of model A2.
    • For model X2 and P2, all the measurement components are mounted within a compact housing with operation key and display LCD. These two models do not display the measurement waveform and noise waveform, but they could display the measurement results of systolic and diastolic pressure, heat beat rate, and noise index.
    AI/ML Overview

    The provided text is a 510(k) summary for a blood pressure monitor. It includes information about the device's intended use, comparison to a predicate device, and non-clinical tests performed. However, it does not contain detailed information about specific acceptance criteria for performance metrics (like accuracy, precision, sensitivity, specificity, etc.) nor does it report the results of a specific study proving the device meets said criteria in the format requested.

    The document states that the substantial equivalence determination is based on the device having the "same intended use and technological characteristics" as the predicate device (DynaPulse 500G) and that "verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device."

    It mentions compliance with voluntary standards:

    • ANSI/AAMI SP 10-1992: "American National Standard, Manual, electronic, or automated sphygmomanometers" - This standard sets accuracy requirements for blood pressure monitors.
    • EN 60601-1: "Medical electrical equipment – Part 1: General requirements for basic safety and essential performance"
    • EN 60601-1-2: "Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests"

    It also mentions that "software verification has been carried out according to the FDA software guidance."

    Since the document does not provide the specific "acceptance criteria" and "reported device performance" in a quantitative manner, nor does it detail a specific study with sample sizes, expert qualifications, and ground truth methodologies, I cannot fill in the table or the other requested points directly from the provided text.

    Therefore, many of the requested fields will be marked as "Not provided in the text."


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Metric)Value (If provided)Reported Device Performance (Value)
    Compliance with ANSI/AAMI SP 10-1992 (Accuracy, etc.)MetNot explicitly detailed
    Compliance with EN 60601-1 (Basic safety & essential performance)MetNot explicitly detailed
    Compliance with EN 60601-1-2 (Electromagnetic compatibility)MetNot explicitly detailed
    Software verification according to FDA software guidancePerformedNot explicitly detailed
    Same safety and effectiveness as predicate device (DynaPulse 500G)Verified by "verification and validation tests contained in this submission"Not explicitly detailed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not provided in the text.
    • Data provenance: Not provided in the text. The manufacturer is from Taiwan, R.O.C., but the location of the clinical data (if any) is not mentioned.
    • Retrospective or prospective: Not provided in the text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not provided in the text. The document refers to "verification and validation tests" but does not detail the methodology for establishing ground truth or the involvement of experts for a test set. Blood pressure monitors are typically validated against a reference standard (like a mercury sphygmomanometer) during device testing, observing agreement within specified limits rather than relying on expert consensus for "ground truth" in the same way an imaging algorithm might.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not provided in the text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a blood pressure monitor, not an AI-assisted diagnostic imaging device requiring human reader interpretation or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • The device itself is a standalone blood pressure monitor. The document implies its performance was evaluated against standards and compared to a predicate device, which would inherently be a "standalone" evaluation of the device's measurement capabilities. However, specific performance metrics are not given.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • While not explicitly stated, for blood pressure monitors, the "ground truth" for accuracy testing is typically established by simultaneously measuring blood pressure using a validated reference method (e.g., a trained observer using a mercury sphygmomanometer following a standardized protocol) on the same subjects, and comparing the device's readings to these reference measurements. The document mentions compliance with ANSI/AAMI SP 10-1992, which specifies such comparison methods.

    8. The sample size for the training set

    • Not applicable/Not provided. This device is a traditional medical device (oscillometric blood pressure monitor), not explicitly an AI/ML algorithm that requires a "training set" in the common sense. Its "algorithm" is fixed and based on oscillometric principles.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided, as there is no mention of a "training set" for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1