Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K150607

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-09

    (30 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osseoflex® SB is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutaneous vertebral augmentation, such as kyphoplasty.

    Device Description

    The Osseoflex® SB is designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter. After the bone is disrupted, PMMA is injected through an Osseoflex® needle to fill the previously created void(s).

    An access channel is required for Osseoflex® SB placement. The Osseoflex® SB device does not create an access channel; the Osseoflex® SB is designed to follow a pre-existing channel created by an access channel device.

    AI/ML Overview

    This document describes the performance testing for the Osseoflex SB, 10 Gauge Straight Balloon.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Bench Tests PerformedAcceptance CriteriaReported Device Performance (Implied from Conclusion)
    Balloon Profile and Tamp (Catheter) Working Length (TM-003)• Balloon profile ≤ 2.79 mm (0.109 in) • Balloon catheter working length > 16.5 cmMet or exceeded requirements
    Balloon Compliance (TM-004)• Balloon working length (L) ≤ 24 mm • Balloon diameter (D) ≤ 19 mmMet or exceeded requirements
    Maximum Pressure (TM-006)The inflatable bone tamp exceeds the maximum inflation pressure, 27 atm (~400 psi) without failure.Met or exceeded requirements (device withstood pressure without failure)
    Bond Tensile Strengths (TM-007)Bond tensile strength ≥ 15 N (3.37 lbf). The tensile force specification was adopted directly from ISO 10555 (Single Use Intravascular Catheters). This tensile force may be applied to the device during use when the balloon is deflated and retracted back through the access cannula.Met or exceeded requirements
    Balloon Maximum Volume (TM-008)Maximum inflation volume to be greater than the maximum recommended volume with 95% confidence and 90% reliability.Met or exceeded requirements
    Balloon Fatigue, Unconstrained (TM-009)Inflate to maximum recommended volume of, hold for 30 seconds / deflate; without leaks for 20 cycles.Met or exceeded requirements (device performed without leaks for 20 cycles)
    Balloon Inflation, Deflation Time (TM-010)Deflation times to be < 5 seconds with 90/90% confidence/ reliability.Met or exceeded requirements (device deflated in under 5 seconds with specified confidence/reliability)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for each specific test (TM-003, TM-004, etc.) within the "PERFORMANCE DATA" section. It broadly states that "The results of the non-clinical tests show that the Osseoflex SB, 10 Gauge Straight Balloon meets or exceed all performance requirements."

    The data provenance is for non-clinical bench tests. These are laboratory tests and do not involve human or animal subjects, so country of origin or retrospective/prospective does not apply in the typical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. These are non-clinical bench tests, and the "ground truth" is established by the predefined engineering and performance specifications and international standards (like ISO 10555 mentioned for tensile strength). No human experts are used to establish ground truth for this type of testing.

    4. Adjudication Method for the Test Set

    Not applicable. As these are bench tests with objective measurements against predefined criteria, an adjudication method for conflicting expert opinions is not relevant.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This document describes non-clinical bench testing of a device (inflatable bone tamp) and not an AI or diagnostic imaging system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the performance data presented is for standalone device performance (algorithm only, if one considers the device's mechanical function as an "algorithm") without human interaction during the test execution itself, beyond operating the testing equipment.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on engineering specifications, design requirements, and relevant international standards (e.g., ISO 10555 for bond tensile strength). These criteria define what constitutes acceptable performance for the device.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a training set as this document describes non-clinical performance testing for a medical device, not a machine learning model.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set mentioned, the establishment of its ground truth is irrelevant to this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141930

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-21

    (36 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osseoflex® SB is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutaneous vertebral augmentation, such as kyphoplasty.

    Device Description

    The Osseoflex® SB is designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter. After the bone is disrupted, PMMA is injected through an Osseoflex® needle to fill the previously created void(s). An access channel is required for Osseoflex® SB placement. The Osseoflex® SB device does not create an access channel; the Osseoflex® SB is designed to follow a pre-existing channel created by an access channel device. The articulating or steering feature of the device assists the clinician in directing the device to the pre-existing channel. The Osseoflex® SB knob can be turned clockwise to aid in directing the distal portion of the device. Turning the knob counter-clockwise will relax the device and allow the device to be returned to its start position. The device should be manipulated only while under fluoroscopic observation with radiographic equipment that provides high quality images.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of a study for the Osseoflex SB device. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedAcceptance CriteriaReported Device Performance
    Balloon Profile and Tamp (Catheter) Working Length (TM-003)Balloon profile ≤ 2.79 mm (0.109 in) Balloon catheter working length > 16.5 cmThe device meets or exceeds all performance requirements (implying the reported performance met these criteria, though specific values are not provided).
    Balloon Compliance (TM-004)Balloon working length (L) ≤ 24 mm Balloon diameter (D) ≤ 19 mmThe device meets or exceeds all performance requirements.
    Maximum Pressure (TM-006)The inflatable bone tamp exceeds the maximum inflation pressure, 27 atm (~400 psi) without failure.The device meets or exceeds all performance requirements.
    Bond Tensile Strengths (TM-007)Bond tensile strength ≥ 15 N (3.37 lbf). The tensile force specification was adopted directly from ISO 10555 (Single Use Intravascular Catheters) requirements. This tensile force may be applied to the device during use when the balloon is deflated and retracted back through the access cannula.The device meets or exceeds all performance requirements.
    Balloon Maximum Volume (TM-008)Maximum inflation volume to be greater than the maximum recommended volume with 95% confidence and 90% reliability.The device meets or exceeds all performance requirements.
    Balloon Fatigue, Unconstrained (TM-009)Inflate to maximum recommended volume of, hold for 30 seconds / deflate; without leaks for 20 cycles.The device meets or exceeds all performance requirements.
    Balloon Inflation, Deflation Time (TM-010)Deflation times to be < 5 seconds with 90/90% confidence/reliability.The device meets or exceeds all performance requirements.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes for each non-clinical test. The tests are non-clinical, implying they were conducted in a laboratory setting, not on human subjects. Therefore, there is no country of origin or retrospective/prospective data as it relates to patient data. The provenance is internal laboratory testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. The tests are non-clinical engineering and performance characteristic tests of a medical device, not clinical studies requiring expert ground truth establishment for diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Not applicable, as no expert adjudication was involved in these non-clinical tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. The studies described are non-clinical performance evaluations of a medical device, not AI-assisted diagnostic studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The device is a physical medical instrument (inflatable bone tamp), not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for these non-clinical tests is based on established engineering and material science principles, international standards (e.g., ISO 10555 for bond tensile strength), and device design specifications. The acceptance criteria themselves serve as the 'ground truth' for performance evaluation.

    8. The Sample Size for the Training Set

    Not applicable. This is a non-clinical device performance study, not an AI model training study.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model. For the non-clinical tests, the "ground truth" (acceptance criteria) was established based on engineering requirements and relevant industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140937

    Validate with FDA (Live)

    Device Name
    OSSEOFLEX SB
    Manufacturer
    Date Cleared
    2014-06-19

    (69 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osseoflex® SB is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during perculancous vertebral augmentation. This system is to be used with cleared spinal polymethymethacrylate (PMMA) bone cements for use during percutancous vertebral augmentation, such as kyphoplasty.

    Device Description

    The Osseoflex® SB is designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter. After the bone is disrupted, PMMA is injected through an Osseoflex® needle to fill the previously created void(s). An access channel is required for Osseoflex® SB · placement. The Osseoflex® SB device does not create an access channel; the Osseoflex® SB is designed to follow a pre-existing channel created by an access channel device. The articulating or steering feature of the device assists the clinician in directing the device to the pre-existing channel. The Osseoflex® SB knob can be turned clockwise to aid in directing the distal portion of the device. Turning the knob counter-clockwise will relax the device and allow the device to be returned to its start position.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text:

    Device: Osseoflex® SB (2mL)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance Test (TM #)Acceptance CriteriaReported Device Performance
    Balloon Profile and Tamp (Catheter) Working Length (TM-003)Balloon profile ≤ 3.48 mm (0.137 in) Balloon catheter working length > 16.5 cm (length of access cannula)The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". Specific numerical results for this test are not provided.
    Balloon Compliance (TM-004)Balloon working length (L) is 16 mm (reference) at the maximum recommended volume 2 mL. Balloon diameter (D) is 15 mm maximum at the maximum recommended volume 2 mL. (15 mm diameter ensures balloon does not pass through vertebral end plates).The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". Specific numerical results for this test are not provided.
    Maximum Pressure (TM-006)The inflatable bone tamp exceeds the maximum inflation pressure, 27 atm (~400 psi) without failure.The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". Specific numerical results for this test are not provided.
    Bond Tensile Strengths (TM-007)Bond tensile strength ≥ 15 N (3.37 lbf). (Adopted from ISO 10555 requirements).The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". Specific numerical results for this test are not provided.
    Balloon Maximum Volume (TM-008)Maximum inflation volume 2 mL with 95% confidence and 90% reliability.The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". T specific numerical result verifying the 95% confidence and 90% reliability is not provided.
    Balloon Fatigue, Unconstrained (TM-009)Inflate to maximum recommended volume of 2 mL, hold for 30 seconds / deflate; without leaks for 20 cycles.The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". Specific numerical results confirming 20 cycles without leaks are not provided.
    Balloon Inflation, Deflation Time (TM-010)The Osseoflex SB 2ml samples deflation times to be clinically equivalent to other marketed inflatable bone tamps.The document states that the "results of the non-clinical tests show that the Osseoflex SB, 2ml meet or exceed all performance requirements". While clinical equivalence is the criterion, specific data comparing deflation times is not presented.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for each non-clinical performance test. It only summarizes that "non-clinical tests" were conducted. The data provenance is from non-clinical laboratory testing, performed by Osseon LLC, as indicated by the "Summary of Non-Clinical Tests Conducted." There is no indication of country of origin for the data (beyond the company being US-based) or whether it was retrospective or prospective in nature (as it's laboratory testing, it would inherently be prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. The tests conducted were non-clinical performance tests for a medical device (inflatable bone tamp), not a diagnostic or AI-assisted system requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable, as this was non-clinical laboratory testing against pre-defined engineering specifications, not a study requiring human adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "N/A – No clinical test were conducted for this submission."

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Not applicable. This is a medical device (inflatable bone tamp), not an algorithm or AI system. The performance tests are for the physical properties and functionality of the device itself.

    7. The Type of Ground Truth Used:

    The "ground truth" for these tests consisted of pre-defined engineering specifications and performance requirements (acceptance criteria) based on device design and relevant standards (e.g., ISO 10555 for bond tensile strength).

    8. The Sample Size for the Training Set:

    Not applicable. This is a medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1