    K Number
    K080180

    Validate with FDA (Live)

    Date Cleared
    2008-06-11

    (139 days)

    Product Code
    Regulation Number
    886.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Retinal Functional Imager (RFI) is intended to observe, capture, display, and store images of patients' fundus (retina) under mydriatic conditions to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed.

    Device Description

    The modified RFI, like the cleared RFI, is a mydriatic fundus imaging camera intended for taking red-free images. The modified RFI, as the cleared device, comprises of the following sub-assemblies:

    • An optical system for illuminating and imaging the retina. The optical imaging includes a stroboscopic light source for sequential rapid imaging of the retina.
    • A high resolution CCD camera.
    • An electronic unit for driving the light source.
    • A software package for operating the system, controlling the illumination, grabbing the images, data browsing and data analysis.
      The device is also capable of using 35 mm film or using a digital camera similar to the predicate devices. Visible light is used for observation. Alignment and focusing is manual via the fundus camera controls.
      Under red-free imaging, the modified RFI provides, through a series of multiple flashes, the ability to observe and register the blood flow velocity and path of motion.
    AI/ML Overview

    The Retinal Functional Imager 3000 (RFI 3000) is intended to observe, capture, display, and store images of patients' fundus (retina) under mydriatic conditions to aid in diagnosing or monitoring diseases of the eye. The device provides information on blood flow (velocity) and path of flow in retinal vessels, but its repeatability and trueness for retinal blood flow velocity measurements have not been clinically established.

    Here's an analysis of the provided information regarding acceptance criteria and performance studies:

    1. Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated through bench testing using a model eye simulation. The key metrics assessed were repeatability (standard deviation of repeated measurements) and accuracy (measured versus calculated results) of retinal blood flow velocity (RBFV) measurements. While specific acceptance criteria are not explicitly stated as numerical thresholds (e.g., "SD must be < X" or "Accuracy must be within Y%"), the study aims to characterize these performance aspects.

    Here's a table summarizing the reported device performance:

    Table: Bench Repeatability and Accuracy of Retinal Blood Flow Velocity Measurements

    MetricPerformance Range (across various velocities)Key Finding/Correlation
    RepeatabilityStandard Deviation (SD): 0.13 mm/sec to 1.22 mm/secVariability increased as velocity increased. %CV ranged from 8.3% to 19.1%.
    AccuracyMean Difference (RFI minus calculated): 0.04 mm/sec to 0.59 mm/secRFI measured velocity was consistently slightly higher than calculated velocity.
    % Difference: 3.3% to 14.8%
    Overall AccuracyAverage ratio between expected and measured velocity: 1.07RFI determined velocity is slightly higher (6.5%) than actual.
    Linear Regression Fit: Y = 1.07X + 0.064High correlation between RFI and calculated velocity.
    Square of Correlation Coefficient ($\text{R}^2$): 0.99Indicates a very strong linear relationship.
    Radiation LevelLower due to narrower band-pass filter.Meets ISO 15004-2:2007 requirements regarding exposure threshold limits.

    Note: The document states that "repeatability and trueness of the retinal blood flow velocity measurement has not been established clinically." The performance data presented above are from bench testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The "test set" for the RBFV measurements consisted of repeated measurements (6-8) at each of six different fixed flow rates. This means a total of approximately 36 to 48 individual measurements were taken.
    • Data Provenance: The RBFV data was generated from bench testing using a "model eye simulation by fixed flow rates of human blood through a pipette of diameter 80 micron inner diameter."
      • This is an in-vitro, controlled environment simulation, not from human subjects.
      • The "human blood" component suggests a biological material, but it's used in a simulated, non-clinical setup.
      • There is no country of origin for the data in the context of human trials, as no human trials were performed for these specific RBFV measurements.
      • The data is prospective in the sense that it was generated specifically for this performance evaluation study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The ground truth for the RBFV measurements was established by the calculated fluid dynamics of the human blood flowing through the pipette at fixed, controlled rates, which acted as the "actual" velocity. It was not based on expert interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was based on calculated physical parameters, not expert consensus requiring an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focused solely on the standalone performance of the device's measurement capabilities in a bench setting. There is no mention of human readers or AI assistance in this context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    Yes, a standalone performance evaluation of the algorithm's measurement capabilities was performed. The bench testing directly assessed the RFI device's ability to measure RBFV against known, calculated values without human intervention in the measurement process (beyond operating the device).

    7. Type of Ground Truth Used

    The ground truth used for the RBFV performance studies was calculated velocity based on the fixed flow rates of human blood through a pipette, which served as a controlled and known reference. This is essentially a form of "physical/engineering ground truth" derived from a calibrated setup rather than clinical pathology or outcomes data.

    8. Sample Size for the Training Set

    The document does not specify any training set size. This submission is a Special 510(k) for a modification (change of fundus camera base) to an already cleared device, not an initial submission for a novel AI algorithm that would typically require a training set. The descriptions of the software "Grab and Browse module" suggest standard image acquisition and processing software rather than a machine learning model that undergoes training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set is mentioned or implied for the performance evaluation described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062416

    Validate with FDA (Live)

    Date Cleared
    2006-11-17

    (92 days)

    Product Code
    Regulation Number
    886.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Retinal Functional Imager (RFI) is intended to observe, capture, display, and store images of patients' fundus (retina) under mydriatic conditions with the information of blood flow (velocity) and path of flow in a retinal vessel, to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed.

    Device Description

    The Retinal Functional Imager (RFI) is a mydriatic fundus imaging camera intended for taking red-free images. The RFI is comprised of the following subassys:

    • An optical system for illuminating and imaging the retina. The optical imaging includes a stroboscopic light source for sequential rapid imaging of the retina.
    • A high resolution CCD camera.
    • An electronic unit for driving the light source.
    • A software package for operating the system, controlling the illumination, grabbing the images, data browsing and data analysis.
      The device is also capable of using 35 mm film or using a digital camera similar to the predicate devices. Visible light is used for observation. Alignment and focusing is manual via the fundus camera controls,
      Under red-free imaging, the Retinal Functional Imager (RFI) provides, through a series of multiple flashes, the ability to observe and register the blood flow velocity and path of motion.
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Retinal Functional Imager (RFI) based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it describes general compliance and validation findings.

    Acceptance Criteria CategoryReported Device Performance
    Compliance with StandardsDetailed in Optical Imaging's IEC 60601-1 Technical Report, Risk Analysis.
    Effectiveness & PerformanceFound to perform as well as equivalent devices on the market.
    Blood Flow Velocity MeasurementsMeasurements obtained were within design specifications, repeatable, and accurate with reference to known velocity simulated data images. Subject data measurements were reasonable and consistent.
    Electromagnetic CompatibilityEvaluated as part of electrical safety test and verification activities.
    System Electrical SafetyEvaluated as part of electrical safety test and verification activities.
    Optical Radiation LevelEvaluated as part of electrical safety test and verification activities.
    Mechanical StabilityEvaluated as part of electrical safety test and verification activities.
    Software Risks (Communication Loss, Power Loss, Out-of-Limit Conditions)Evaluated as part of Software Test Procedure for validation of the software.
    User Interface Risks (Functional Clarity, Patient Data Integrity/Security, User Misuse)Evaluated as part of Software Test Procedure for validation of the software.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "known velocity simulated data images" were used for evaluating blood flow velocity measurements, and "subject data measurements" were taken. However, specific sample sizes for the test set (number of simulated images or human subjects) are not provided. The provenance (country of origin, retrospective/prospective) of the subject data is also not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not provided in the document.

    4. Adjudication Method for the Test Set:

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A MRMC comparative effectiveness study was not mentioned or conducted based on the provided text. The document focuses on the device's technical performance and equivalence to predicate devices, not on human-in-the-loop performance improvements.

    6. Standalone (Algorithm Only) Performance Study:

    Yes, a standalone performance study was done for the algorithm described. The device's ability to "map the blood flow through blood vessels in the eye and determine the velocity of the flow by tracking the motion of the blood cells" is a core function of the algorithm. The document states that blood flow velocity measurements were "repeatable and accurate with reference to known velocity simulated data images." This indicates a standalone evaluation of the algorithm's accuracy against a known truth.

    7. Type of Ground Truth Used:

    For blood flow velocity measurements, the ground truth used was "known velocity simulated data images." For "subject data measurements," it indicates consistency and reasonableness, implying a comparison against expected physiological norms or perhaps other measurement methods (though not explicitly stated as ground truth establishment).

    8. Sample Size for the Training Set:

    The document does not specify a separate training set or its sample size. The description of the device's capabilities implies that the software was developed and validated, but details about training data are absent.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided as a separate training set is not explicitly mentioned, nor is the method for establishing ground truth for any potential training data.