Search Results
Found 100 results
510(k) Data Aggregation
(267 days)
Models FD-410LR & FD-412LR:
The Single Use Electrosurgical Hemostatic Forceps FD-410LR/412LR are intended to be used to cauterize and coagulate or perform hemostasis using high-frequency current within the upper digestive tract in combination with an endoscope.
Model FD-411UR:
The Single Use Electrosurgical Hemostatic Forceps FD-411UR is intended to be used to cauterize and coagulate or perform hemostasis using high-frequency current within the lower digestive tract in combination with an endoscope.
The Electrosurgical Hemostatic Forceps FD-400 series is comprised of three (3), sterile, single-use, electrosurgical hemostatic forceps of varying forceps opening widths (4 – 6.5 mm). The subject device was designed to be used with compatible endoscopes to cauterize and coagulate or to perform hemostasis using high-frequency current in the upper (Models FD-410LR and FD-412LR) and lower digestive tract (FD-411UR).
The hemostatic forceps are attached to an electrosurgical unit via an A-cord to deliver high-frequency current to cauterize and coagulate or to perform hemostasis on the target tissue. The forceps are manufactured from stainless steel.
The Electrosurgical Hemostatic Forceps FD-400 series is to be used with compatible electrosurgical units, A cords, and endoscopes.
N/A
Ask a specific question about this device
(94 days)
The SINGLE USE DISTAL COVER MAJ-2315 has been designed to be attached to OLYMPUS endoscopes to cover the distal end of the insertion tube and around the forceps elevator.
The Single Use Distal Cover MAJ-2315 has been designed to be attached to OLYMPUS endoscopes to cover the distal end of the insertion tube and around the forceps elevator. The MAJ-2315 is used as the distal cover for the EVIS EXERA III Duodenovideoscope Olympus TJF-Q190V. The MAJ-2315 is provided sterile and is discarded after use. The 510(k) communicates the change of the addition of the Attachment Tool to the MAJ-2315 and a minor manufacturing process change. The MAJ-2315 is pre-loaded into the Attachment Tool and is discarded once the MAJ-2315 is placed onto the distal end of the endoscope.
The subject device MAJ-2315 is minimally changed from the predicate device MAJ-2315 (K220587).
N/A
Ask a specific question about this device
(196 days)
The VISERA ELITE III Surgical Imaging System is intended to provide real-time visible and near infrared fluorescence imaging.
Upon intravenous administration and use of ICG consistent with its approved label, VISERA ELITE III Surgical Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the VISERA ELITE III Surgical Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Upon interstitial administration and use of ICG consistent with its approved label, the VISERA ELITE III Surgical Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
The VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 is intended to process electronic signals transmitted from a video endoscope/camera head and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, light sources, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
The VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 is intended to provide light to an endoscope/video endoscope in order to process electronic signals transmitted from them and output image signal to monitor, and to be used with endoscopes, video endoscopes, camera heads, video system centers, monitors and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
The 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA is intended to be used with endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
The VISERA ELITE III Surgical Imaging System is intended to be used with ancillary equipment for endoscopic diagnosis, treatment, and observation and supports the function of high definition (HD) videoscopes and is Camera Head (CH) compatible.
The following subject devices of the OLYMPUS VISERA ELITE III Surgical Imaging System are identical to and unchanged from the VISERA ELITE III Surgical Imaging System (K242067):
-
VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 (Model Number: OLYMPUS OTV-S700) – OLYMPUS OTV-S700 is a video system center that processes electronic signals transmitted from a video endoscope or a camera head and outputs the image signal to a monitor.
- OTV-S700 3D UPGRADE PACK (Model Number: OLYMPUS MAJ-2511) – A function activation portable memory key accessory that unlocks the 3D software function when connected with VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700 to enable 3D observation mode.
-
VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700 (Model Number: OLYMPUS CLL-S700) – A LED light source that provides examination light to a video endoscope and a camera head.
-
4K Camera Head OLYMPUS CH-S700-XZ-EA (Model Number: OLYMPUS CH-S700-XZ-EA) – A 4K Inline camera head is intended to be used with Olympus endoscopes, the video system center, and other ancillary equipment for the visualization of internal organs (Endoscopic diagnosis), treatment and observation.
N/A
Ask a specific question about this device
(64 days)
The Single-use Fine Needle Biopsy (FNB) device NA-U210H has been designed to be used with an ultrasound endoscope for sampling of submucosal and extramural lesions within or adjacent the gastrointestinal tract.
The Single-Use Fine Needle Biopsy (FNB) Device NA-U210H (SecureFlex) is a sterile, single-use fine needle biopsy device intended for use with an ultrasound endoscope to sample submucosal and extramural lesions within or adjacent to the gastrointestinal tract. It consists of:
- Biopsy Needle Assembly: Handle, needle (19G, 22G, or 25G), and coiled sheath.
- Aspiration System: VACLOK syringe and stopcock (FDA-cleared under K994253).
The device features a bifurcated needle tip for enhanced tissue acquisition, echo-enhanced regions for ultrasound visibility, and is compatible with Olympus GF and TGF endoscopes with ≥2.8 mm channel diameter.
N/A
Ask a specific question about this device
(203 days)
The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire.
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use Sphincterotome KD-VC, KD-V, and KD Series consists of the following devices:
- Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC600 Series (aka CleverCut3V)
- Single Use Sphincterotome V (Distal Wireguided) KD-VC400 Series (aka CleverCut3V)
- Single Use 3-Lumen Sphincterotome V KD-V Series (aka CleverCut3V)
- DISPOSABLE TRIPLE Lumen SPHINCTEROTOME KD Series (aka FlowCut)
The subject devices have separate lumens for guidewire, cutting wire and injection of contrast medium for papillotomy. The knives are pre-curved and have a tapered tip to facilitate insertion into the papilla of Vater. The insertion portion of the KD-V, and KD-VC Series have a V-marking, which the relative insertion length into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope and feature an integrated C-Hook to attach to the endoscope. The distal end of the sphincterotome is coated with CleverCut coating ("Olympus' PFA (perfluoroalkoxy) coating designed to prevent thermal injury to non-target tissue while cutting the papilla of Vater. The KD-VC600 Series offers models that are preloaded with VISIGLIDE guidewires.
The subject devices are provided to the user sterilized by ethylene oxide and intended for single use only.
N/A
Ask a specific question about this device
(94 days)
The EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V has been designed to be used with a video system center, light source, documentation equipment, monitor, endo therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
The EVIS EXERA III DUODENOVIDEOSCOPE TJF-Q190V has been designed to be used with a video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. The TJF-Q190V is compatible with Olympus system "Video System Center OLYMPUS CV-190 and XENON LIGHT SOURCE OLYMPUS CLV-190 (K112680). TJF-Q190V is also compatible with Olympus system "EVIS X1 Video System Center OLYMPUS CV-1500" (K222584).
The subject device consists of a flexible insertion section, control section and endoscope connector section with equipped charge-coupled device (CCD) chip which delivers images.
The light from the light source travels through the light guide to the light guide lens at the distal end. The light source can offer both white light for normal observation and narrow band imaging (NBI). The CCD chip transduces the incident light from the objective lens to electrical signal. The video processor transduces the electrical signal to video signal.
There is an instrument channel located inside of the flexible insertion section. EndoTherapy accessories can be inserted through the instrument channel. A forceps elevator is located at the distal end of the insertion section to elevate endo therapy accessories for endoscopic treatment.
A sterile, single-use distal cover (MAJ-2315) has been designed to be attached to the OLYMPUS TJF-Q190V to cover the distal end of the insertion tube and fit around the forceps elevator. MAJ-2315 is to be discarded after clinical use. MAJ-2315 and TJF-Q190V were previously cleared under 510(k)s K193182, K202661, K220587, and K250701.
The subject device has the same technological characteristics and similar design as the predicate device.
N/A
Ask a specific question about this device
(94 days)
EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XP190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-P190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XT190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
The EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-XP190, OLYMPUS BF-P190, and BF-XT190) and BRONCHOVIDEOSCOPE BF-H1100 and BF-1TH1100 are used for endoscopic diagnosis and treatment within the respiratory organs. These endoscopes consist of three parts: the control section, the insertion section, and the connector section.
The provided FDA 510(k) clearance documentation for the Olympus Bronchovideoscopes (K250862) primarily demonstrates substantial equivalence based on technological characteristics and bench testing. It does not contain information about comparative effectiveness studies (like MRMC studies for AI devices), expert ground truth establishment, or typical performance metrics associated with AI/software-as-a-medical-device (SaMD) clearances.
This submission focuses on:
- Technological Equivalence: Showing that the devices (BF-XP190, BF-P190, BF-XT190, BF-H1100, and BF-1TH1100) are fundamentally the same as their predicates, with the primary change being compatibility with a new video system center (CV-1500) and updated labeling related to laser/high-frequency/APC systems.
- Bench Testing: Verification of physical and imaging performance parameters (e.g., thermal safety, color performance, resolution, noise, video latency) to ensure they meet specifications, particularly when combined with the new video system.
- Animal Testing: To assess the imaging modes (WLI, NBI, TXI, BAI-MAC) with the new video processor.
Therefore, the requested information about "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/software performance (e.g., sensitivity, specificity, MRMC studies, ground truth establishment by experts) is not present in this document. The document explicitly states that "Software Testing and Cybersecurity" was "not performed" due to "no design, material, sterilization, reprocessing, packaging, shelf life, or software changes" (Page 29). This implies that the device itself is a hardware endoscope, and any software associated with it is considered an intrinsic part of its established functionality, not a new or significantly changed software component requiring a separate performance study with clinical endpoints or AI evaluation.
The "acceptance criteria" here relate to the successful completion of the listed bench (and limited animal) tests, demonstrating that the new combination (endoscope + CV-1500) functions as intended and safely, similar to the predicate combinations.
Below is a table summarizing the "acceptance criteria" and "reported device performance" as derived from the document's comparison tables and performance data section, which are primarily about technical specifications and functional verification, not software/AI performance metrics.
Acceptance Criteria and Reported Device Performance (Summary based on provided text)
Since this 510(k) is for existing endoscopes with a new video system and updated labeling, the "acceptance criteria" are implied by the extensive comparison tables (Tables 1-5) which show the subject devices having nearly identical technical specifications to their predicate devices, and the successful completion of specified bench and animal testing. The performance data section doesn't list specific quantitative acceptance criteria for each test but rather states that tests were conducted to "ensure that the subject device performs as intended and meet design specifications."
| Acceptance Criteria Category | Specific Acceptance Criteria (Implied/Defined by Predicate Equivalence) | Reported Device Performance (as stated or implied by equivalence) |
|---|---|---|
| Indications for Use | Identical to predicate devices. | The Indications for Use for the Subject Device is exactly the same as the Indications for Use for the equivalent Predicate Device (Page 9). |
| Material Composition | Identical materials used for key components (e.g., Insertion Tube, Bending Section Rubber). | Materials are identical to predicates (e.g., Fluoro Resin, Fluoro Rubber, Epoxy Glue, Polytetrafluoro-ethylene, Polysulfone, Glass/Sapphire) (Pages 12-13, 16-17, 20-21). |
| Physical Dimensions | Maintain specific dimensions (e.g., Outer Diameter, Working Length, Instrument Channel inner diameter). | Dimensions are identical to predicates across all models (e.g., BF-P190 Outer Diameter of Distal End Φ4.2mm, Working Length 600mm, Instrument Channel inner diameter φ2.0; BF-XP190 Outer Diameter of Distal End Φ3.1mm; BF-XT190 Outer Diameter of Distal End Φ6.1mm; BF-H1100 Distal end outer diameter ø 4.9 mm; BF-1TH1100 Distal end outer diameter ø 5.8 mm) (Pages 12, 16, 19-20, 23, 26). |
| Optical/Imaging Performance | Maintain specific optical characteristics (e.g., Depth of Field, Field of View, Pixels, CCD type). | Identical optical characteristics to predicates (e.g., Depth of Field 2-50mm/3-100mm, Field of View 110°/120°, Total Number of Pixels 50,621 / 465,588, Color CCD) (Pages 11-12, 15, 19, 22, 25). |
| Software Functionality | Identical software functions (e.g., pixel defect correction, scope info, NBI, Pre-freeze). | All listed software functions are "Available" and match the predicate devices (Pages 13, 17, 21, 24, 27-28). The RDI, TXI, and BAI-MAC modes are explicitly stated as "Available" for the subject devices when previously "Not Available" for some predicates, implying this is the new feature enabled by the CV-1500 system, and their performance was assessed. |
| Safety | Meet safety standards for thermal, photobiological, and ignition factors (especially with laser/HF/APC compatibility). | Bench testing conducted for: Thermal Safety, Photobiological Safety, Ignition Factors Safety Study (Laser, High-Frequency, APC, Scope Comparison) to ensure performance and safety (Page 28). |
| Functional Verification | Performance as intended with the new CV-1500 video system center. | Bench testing (Color Performance, Resolution, Noise and Dynamic Range, Image Intensity Uniformity, Video Latency, RDI, TXI, BAI-MAC, Automatic Brightness Adjustment, Pre-Freeze) was conducted to ensure intended performance (Page 28). Animal testing performed to assess imaging modes (WLI, NBI, TXI, BAI-MAC) with CV-1500 (Page 29). |
| Risk Management | Risks identified and mitigated in accordance with ISO 14971:2019. | Risk management performed, and design verification tests and acceptance criteria identified as a result (Page 29). |
Study Details (based on provided text)
The document describes performance testing rather than a comparative clinical study for AI/software-as-a-medical-device.
-
A table of acceptance criteria and the reported device performance:
- See table above. The acceptance criteria are largely implied by the equivalence to the predicate devices in terms of physical, optical, and functional characteristics, and the successful completion of specified bench and animal tests. Quantitative metrics for these tests are not provided in this summary but would have been part of the full submission.
-
Sample sizes used for the test set and the data provenance:
- Test set sample size: Not specified. The document mentions "bench testing" and "animal testing." For bench tests, it typically refers to a small number of devices or engineered test setups. For animal testing, the number of animals or studies is not provided.
- Data provenance: Not specified. It can be inferred that the testing was conducted by or on behalf of Olympus Medical Systems Corp. in Japan, given the manufacturing site and submitter location. It does not state if the data is retrospective or prospective, or from which country/region the "animal" data would originate.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not specified. This type of information is typically provided for studies evaluating AI algorithms or diagnostic accuracy, where human experts establish a ground truth for imaging interpretation. The present submission is for a hardware endoscope system, and its performance evaluation relies on engineering specifications and functional testing, not expert interpretation of diagnostic images.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable / Not specified. Adjudication methods are relevant for human reader studies or expert ground truth establishment, which are not detailed here.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, not performed / Not applicable. This submission is not for an AI companion diagnostic or an AI-software-as-a-medical-device. It's for an endoscope system. The document explicitly lists "Software Testing and Cybersecurity" and "Clinical" as "not performed" because there were no fundamental software changes beyond integration with a new video system, which itself does not constitute an AI component in the context of this 510(k) summary. The NBI, RDI, TXI, and BAI-MAC modes are imaging enhancements, not AI algorithms.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This relates to AI/software performance, which is not the focus of this 510(k).
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable. For this device (endoscope), ground truth typically relates to engineering specifications (e.g., accurate measurements, clear image quality as determined by reproducible test patterns, proper functionality of mechanical parts). For the animal testing of imaging modes, the "ground truth" would be the direct observation of the animal's internal anatomy via the endoscope itself and comparison to expected/known characteristics, not an independent "pathology" or "outcomes" ground truth in a diagnostic sense.
-
The sample size for the training set:
- Not applicable. This submission is not for an AI system that requires a "training set."
-
How the ground truth for the training set was established:
- Not applicable. As above, no AI training set is described.
Ask a specific question about this device
(90 days)
The EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V has been designed to be used with a video system center, light source, documentation equipment, monitor, endo therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
The EVIS EXERA III DUODENOSCOPE TJF-Q190V has been designed to be used with an video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. The TJF-Q190V is compatible with Olympus system "Video System Center OLYMPUS CV-190 and XENON LIGHT SOURCE OLYMPUS CLV-190 (K112680)." TJF-Q190V is also compatible with Olympus system "EVIS X1 Video System Center OLYMPUS CV-1500 (K222584)."
The subject device consists of a flexible insertion section, control section and endoscope connector section with equipped charge-coupled device (CCD) chip which delivers images.
The light from the light source travels through the light guide to the light guide lens at the distal end. The light source can offer both white light for normal observation and narrow band imaging (NBI). The CCD chip transduces the incident light from the objective lens to electrical signal. The video processor transduces the electrical signal to video signal.
There is an instrument channel located inside of the flexible insertion section. EndoTherapy accessories can be inserted through the instrument channel. A forceps elevator is located at the distal end of the insertion section to elevate endo therapy accessories for endoscopic treatment.
A sterile, single-use distal cover (MAJ-2315) has been designed to be attached to the OLYMPUS TJF-Q190V to cover the distal end of the insertion tube and fit around the forceps elevator. MAJ-2315 is to be discarded after clinical use. MAJ-2315 and TJF-Q190V were previously cleared under 510(k)s K193182, K202661 and K220587.
The provided 510(k) clearance letter for the Olympus TJF-Q190V Duodenoscope indicates that the clearance is based on adhesive modifications to the device. This implies that the acceptance criteria and the study proving the device meets these criteria would primarily relate to the durability and safety of the new adhesive material and its impact on the overall performance of the endoscope, particularly concerning reprocessing and material integrity.
However, the provided text does not contain the detailed information typically found in a clinical study report or a summary of non-clinical performance data that would explicitly list acceptance criteria and device performance in the format requested. The document states:
- "Verification/validation activities were performed subsequent to a risk assessment evaluation of the device modifications per the Olympus Quality Management System."
- "Results of the following testing demonstrate that the changes to the device do not adversely affect device performance: Performance Testing - Bench, Sterilization and Shelf-Life - Residual Toxicity of Reprocessing Chemicals, Biocompatibility Evaluation."
- "No clinical data were collected."
This means the clearance was based on non-clinical (bench) testing. Therefore, I cannot provide information on clinical performance metrics, multi-reader multi-case studies, or the establishment of ground truth by human experts, as these were not part of the submission for this particular change.
Given the limited information, I will infer the implied acceptance criteria based on the described non-clinical tests.
Implied Acceptance Criteria and Reported Device Performance
Based on the provided information, the acceptance criteria would be related to the non-clinical performance of the device with the new adhesive, ensuring it is equivalent or superior to the predicate device, especially concerning durability against reprocessing and biocompatibility.
1. Table of Acceptance Criteria and Reported Device Performance (Inferred)
| Acceptance Criterion (Inferred) | Reported Device Performance (Inferred from "demonstrate that the changes... do not adversely affect device performance") |
|---|---|
| Material Durability/Mechanical Performance (Bench Testing): The modified device, including the new adhesive, must maintain its structural integrity and functionality after repeated reprocessing cycles, comparable to or better than the predicate device. | The changes to the device (adhesive modifications) were verified/validated, and results demonstrate that they "do not adversely affect device performance" in terms of mechanical integrity and functionality after testing. The device performs "as well as or better than the predicate device." |
| Sterilization and Shelf-Life - Residual Toxicity of Reprocessing Chemicals: The new adhesive must not retain toxic residues from reprocessing chemicals above acceptable limits, nor should it degrade in a way that compromises safety or effectiveness over its shelf-life. | Testing for residual toxicity of reprocessing chemicals was performed, and results indicate no adverse effect, confirming safety and chemical compatibility. |
| Biocompatibility: The new adhesive material must be biocompatible according to established standards, ensuring it does not elicit adverse biological responses when in contact with tissue. | Biocompatibility evaluation was performed, and results demonstrate no adverse effect, meaning the new adhesive material meets biocompatibility standards. |
| Overall Performance Equivalence: No adverse changes to optical or electrical performance, or other device specifications, compared to the predicate device. | "There are no changes in contents of the indications for use, conditions of use, compatible components or accessories to be marketed/used with the device, or device specifications for TJF-Q190V, including optical or electrical performance." This implies equivalence was maintained. |
Study Proving Device Meets Acceptance Criteria:
The study involved non-clinical, bench-level verification and validation activities.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated. For bench testing, this would typically involve a defined number of devices or components subjected to various tests (e.g., durability cycles, chemical exposure). The exact number of units or test replicates is not provided in the summary.
- Data Provenance: The manufacturing site is Aizu Olympus Co., Ltd., Japan. The testing would have been conducted in a controlled lab environment. This was a retrospective evaluation of a design modification, and the data would be laboratory-generated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This question is not applicable as "No clinical data were collected." The "ground truth" for non-clinical bench testing would be defined by engineering specifications, material standards, and validated test methods, rather than expert human interpretation of medical images or patient outcomes.
4. Adjudication Method for the Test Set:
- Not applicable. As no clinical data or human evaluations were involved for this specific submission.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, a MRMC study was not done. The document explicitly states, "No clinical data were collected." Therefore, there is no information on how human readers might improve with or without AI assistance, as AI is not mentioned and no human reading study was conducted.
6. Standalone (Algorithm Only) Performance:
- Not applicable. This device is a duodenoscope, a physical medical instrument, not an AI algorithm. There is no mention of a software algorithm or standalone performance in the context of AI.
7. Type of Ground Truth Used:
- For the non-clinical testing, the "ground truth" would be based on:
- Engineering Specifications: Device design parameters, material specifications.
- Regulatory Standards: ISO standards for biocompatibility (e.g., ISO 10993), sterilization, and other relevant performance standards for endoscopes.
- Validated Test Methods: Established laboratory protocols for evaluating material durability, chemical resistance, and device functionality.
- Predicate Device Performance Baseline: The performance characteristics of the legally marketed predicate device (K202661) served as a benchmark for substantial equivalence.
8. Sample Size for the Training Set:
- Not applicable. This submission focuses on a hardware modification (adhesive) to an existing device, not the development or training of a software algorithm.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. As there was no training set for a software algorithm.
Ask a specific question about this device
(90 days)
The Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q are intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series is comprised of nine (9) sterile, single-use, cannulas designed to inject contrast medium in the biliary or pancreatic duct when used in conjunction with a compatible endoscope.
Each device has two sections: the handle (proximal portion) and the insertion portion. The insertion portion is introduced into the biliary or pancreatic ducts through an endoscope. The distal end of the insertion portion is designed for smooth cannulation of the papilla of Vater or the minor papilla. All models are visible under fluoroscopy and feature a distal marking system.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series models are to be used with compatible endoscopes.
The provided 510(k) Premarket Notification document describes a medical device, the "Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series," and its comparison to a predicate device for demonstrating substantial equivalence.
However, the document does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the context of device performance as typically expected for software-enabled devices or those with diagnostic capabilities.
This submission is for a physical medical device (cannulas) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing of its physical properties and biocompatibility. Therefore, many of the requested categories (like sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable to the information provided.
Based on the document, here's what can be extracted and what cannot:
1. Table of Acceptance Criteria and Reported Device Performance:
The document outlines performance data that was provided to demonstrate substantial equivalence, rather than specific and quantitative acceptance criteria with reported numerical device performance against those criteria. The "Analysis" column in the comparison table broadly states "Substantially equivalent" or "Identical," but doesn't provide the detailed numbers that would typically be associated with acceptance criteria for a diagnostic or algorithmic device.
| Performance Data Category | Description |
|---|---|
| Biocompatibility | Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity |
| Sterilization Validation | Per ISO 11135:2014 |
| Ethylene Oxide Residuals | Per ISO 10993-7:2008 |
| Packaging Validation & Shelf Life | Per ISO 11607-1:2019 and ASTM F1980-21 |
| Mechanical Testing & Comparative Testing | Insertion force/Withdrawal force, Insertion w/ Stylet, Attachment and detachment of the hook, Contrast medium infusion, Connection strength, Visibility |
| Human Factors Testing | Verification of device performance |
Acceptance Criteria and Reported Performance (General statement from the document):
The document states: "Non-clinical testing demonstrates that the slight differences in device design do not alter the safety, efficacy, or performance of the subject devices when compared to the predicate devices." and "The non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device." This is a qualitative conclusion of meeting equivalence rather than presenting specific numerical acceptance criteria.
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified in the provided document. The performance data refers to various non-clinical tests, and the sample size for these individual tests (e.g., number of cannulas tested for insertion force) is not detailed.
- Data Provenance: Not applicable in the context of clinical data. The tests are non-clinical (laboratory/bench testing).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth as typically understood for diagnostic performance (e.g., disease presence/absence) is not relevant for these non-clinical, physical device performance tests. "Human Factors Testing" is mentioned, which would involve experts, but the number and qualifications are not provided, nor is it the type of "ground truth" establishment usually refers to in the context of diagnostic AI.
4. Adjudication method for the test set:
- Not applicable. This is typically used for clinical study endpoints or image interpretation, not for physical performance tests of a cannula.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a physical cannula, not an AI or diagnostic imaging system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable in the conventional sense. The "ground truth" for the non-clinical tests would be the physical properties and functional performance measured against predefined specifications or predicate device performance.
8. The sample size for the training set:
- Not applicable for a physical device where "training set" doesn't apply in the context of machine learning.
9. How the ground truth for the training set was established:
- Not applicable for a physical device.
Ask a specific question about this device
(90 days)
The Disposable Distal Attachment D-201 Series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope's view field within the gastrointestinal tract.
The Disposable Distal Attachment D-201 Series is comprised of seven (7), sterile, single-use, distal attachments of varying diameters. Each distal attachment is a short transparent plastic tube which is attached to the distal end of an endoscope to facilitate observation of tissues during endoscopic procedures. A subset of the models includes a drainage side hole that allows egress of water and blood during use. The Disposable Distal Attachment D-201 Series is to be used with compatible endoscopes.
The provided FDA 510(k) clearance letter and summary for the "Disposable Distal Attachment D-201 Series" does not describe a study involving an AI/Medical Imaging device that requires ground truth, human readers, or a test set as described in the prompt.
Instead, this document describes a traditional medical device (a physical accessory for an endoscope) and the performance testing conducted to demonstrate its substantial equivalence to a predicate device. The performance data focuses on biocompatibility, sterilization, packaging, shelf-life, mechanical testing, and human factors. There is no mention of an algorithm, image analysis, or clinical performance in terms of diagnostic accuracy or reader improvement.
Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria using the provided information, as the requested elements (ground truth, expert consensus, MRMC studies, training/test sets for an algorithm, etc.) are not applicable to this type of device and its clearance process.
The document states: "Clinical data is not required to demonstrate substantial equivalence." This further confirms that no studies involving analysis of images or AI performance were conducted or necessary for this device's clearance.
Ask a specific question about this device
Page 1 of 10