510(k) Data Aggregation

    K Number
    K192669

    Validate with FDA (Live)

    Date Cleared
    2020-07-24

    (303 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The condom is used for contraception and for prophylactic purposes (preventing transmission of sexually transmitted infections).

    Device Description

    The device is a condom that is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. This device is a smooth-surfaced, straight-walled, teat-ended, silicone-lubricated condom with nominal length 180 ± 2mm, nominal width 53.5 ± 2mm, and nominal thickness of 0.04 mm. This condom conforms to the recognized standards ASTM D3492-16 and ISO 4074:2015.

    AI/ML Overview

    The provided text describes the 510(k) summary for the "Extremely Thin 003, ZERO ZERO THREE" male latex condom. It discusses the device's characteristics, its comparison to a predicate device, and performance testing data.

    This document does not describe an AI/ML medical device. Therefore, a study proving an AI/ML device meets acceptance criteria, including specifics like sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance, cannot be extracted from this text. The acceptance criteria and performance data pertain only to the physical properties and biological compatibility of condoms.

    However, I can extract the acceptance criteria and reported device performance for the condom itself based on the provided text.


    Acceptance Criteria and Device Performance (for the condom)

    Based on the provided text, the device is a condom, not an AI/ML-driven medical device. Therefore, the acceptance criteria and performance data relate to the physical and biological properties of the condom.

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance Criteria (Standard & Values)Reported Device Performance
    Biocompatibility- Cytotoxicity: Per ISO 10993-5:2009Demonstrated to be biocompatible
    - Sensitization: Guinea pig maximization sensitization testing per ISO 10993-10:2010Demonstrated to be biocompatible
    - Vaginal Irritation: Per ISO 10993-10:2010Demonstrated to be biocompatible
    - Acute Systemic Toxicity: Per ISO 10993-11:2006Demonstrated to be biocompatible
    Physical Testing- Airburst specifications: Met ISO 4074:2015 - Natural rubber latex male condoms - Requirements and test methods, and ASTM D3492:2016 - Standard Specification for Rubber Contraceptives (Male Condoms).Met airburst specifications
    Shelf-Life Stability5 years shelf-life. Per 21 CFR §801.435. All samples must meet predefined acceptance criteria.All samples met predefined acceptance criteria. (5 years)
    Dimensions (Nominal)- Nominal Width: 53.2 ± 2 mm (Proposed Device Specification)53.2 ± 2 mm (implicitly met as per device specification)
    - Nominal Length: 180 ± 10 mm (Proposed Device Specification)180 ± 10 mm (implicitly met as per device specification)
    - Nominal Thickness (30mm): 0.039 ± 0.0013 (Proposed Device Specification)0.039 ± 0.0013 (implicitly met as per device specification)
    - Nominal Thickness (90mm): 0.039 ± 0.001 (Proposed Device Specification)0.039 ± 0.001 (implicitly met as per device specification)
    - Nominal Thickness (150mm): 0.039 ± 0.001 (Proposed Device Specification)0.039 ± 0.001 (implicitly met as per device specification)
    Bursting Pressure1.42 ± 0.10 kPa (Proposed Device Specification)1.42 ± 0.10 kPa (implicitly met as per device specification)
    Bursting Volume35.4 ± 3.83 dm3 (Proposed Device Specification)35.4 ± 3.83 dm3 (implicitly met as per device specification)

    Study Details (for the condom):

    As the product is a condom and not an AI/ML device, the following points address what can be inferred or what is not applicable from an AI/ML perspective.

    2. Sample size used for the test set and the data provenance:
    * Sample Size: For physical testing, "Three lots of devices were tested at baseline." The specific number of condoms per lot is not stated. For biocompatibility, typical regulatory testing involves a sufficient number of samples as per the respective ISO standards, but the exact count is not provided.
    * Data Provenance: Not explicitly stated, however, the testing was conducted to meet international (ISO) and US (ASTM, CFR) standards, implying a controlled laboratory environment typically adhering to Good Manufacturing Practices. The origin isn't specified (e.g., country), but it's for a US-marketed product. The data collection would be prospective, as it's part of the premarket submission process.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * Not Applicable: For a condom, "ground truth" in the context of expert review (like for image interpretation) is not relevant. Performance is determined by objective physical and biological tests against established standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    * Not Applicable: Adjudication methods are relevant for subjective assessments, typically in clinical readings or image interpretations. Condom performance is assessed through objective measurements against specified criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * Not Applicable: This is an AI/ML related study type. The product is a physical device (condom).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Not Applicable: This is an AI/ML related study type. The product is a physical device (condom).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    * The "ground truth" for the condom's performance is established by objective measurements and chemical/biological assays according to recognized consensus standards (ASTM D3492-16, ISO 4074:2015, and ISO 10993 series) and regulatory requirements (21 CFR §801.435). This includes measurements for dimensions, bursting pressure/volume, and results from biocompatibility tests (cytotoxicity, sensitization, irritation, systemic toxicity).

    8. The sample size for the training set:
    * Not Applicable: This pertains to AI/ML model development. For a non-AI/ML device, there is no "training set."

    9. How the ground truth for the training set was established:
    * Not Applicable: This pertains to AI/ML model development. For a non-AI/ML device, there is no "training set" or corresponding ground truth establishment in this manner.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141256

    Validate with FDA (Live)

    Date Cleared
    2014-07-23

    (70 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Okamoto Studded Condom is used for contraceptive and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

    Device Description

    This condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. This condom is textured surface (dotted), tapered shape with reservoir tip, silicone lubricated condom with nominal length 180 mm, nominal flat width 54 mm measured at 30 mm from the open end, a maximum width of 58 mm at the closed end, and nominal thickness of 0.070 mm. The air burst pressure is ≥ 1 kPa and air burst volume is ≥ 17 dm3. It is lubricated with silicone (viscosity 200 cps) and cornstarch is used as a dressing material. This condom conforms to current established national standard ASTM D3492: 2008. The term Studded refers to the irregular surface of the irregular surface is not applied externally but is achieved in the molding process. The glass mold has an engraved pattern of semispherical pits replicated as a pattern of small round tips on the surface of the condom. This results in an exterior surface that has a pattern of small raised areas best described as a dotted pattern. The new condom is defined as a condom of increased thickness relative to some of other Okamoto condoms (e.g., Crown condom, K893039).

    AI/ML Overview

    The provided text is a 510(k) summary for the Okamoto Studded Condom. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with acceptance criteria and device performance results in the context of an AI/medical device study. Therefore, most of the requested information regarding acceptance criteria and study details cannot be extracted from this text.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document indicates that the device "Complied with ASTM D3492:2008". The specific acceptance criteria within this standard are not explicitly listed in the 510(k) summary, nor are detailed quantitative performance results for each criterion. The document states that "The air burst pressure is ≥ 1 kPa and air burst volume is ≥ 17 dm3," which are performance metrics, but not framed as "acceptance criteria" in a table alongside reported device performance.

    Acceptance Criteria (Implied by ASTM D3492:2008 & mentioned specifics)Reported Device Performance (from summary)
    Compliance with ASTM D3492:2008 (general)Complied with ASTM D3492:2008
    Air burst pressure≥ 1 kPa
    Air burst volume≥ 17 dm3
    Biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity)Non-toxic, non-sensitizing, and non-irritating (conforms to ISO 10993)
    Shelf life5 years (established on similar condoms)

    2. Sample Size for Test Set and Data Provenance:

    The document does not detail specific sample sizes for particular tests (e.g., air burst pressure, biocompatibility). It mentions "studies conducted on the similar condoms" for shelf life. Data provenance (country of origin, retrospective/prospective) is not mentioned.

    3. Number of Experts and Qualifications for Ground Truth:

    This information is not applicable to this type of device submission. The evaluation is based on compliance with established standards and direct physical testing, not expert consensus on interpretations.

    4. Adjudication Method:

    Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This device is not an AI algorithm for diagnostic interpretation where human reader improvement would be measured.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. This is a physical medical device (condom), not an algorithm.

    7. Type of Ground Truth Used:

    The "ground truth" for this device is based on objective measurements conforming to established industry standards (ASTM D3492:2008, ISO 10993). For example, air burst pressure is a direct physical measurement.

    8. Sample Size for Training Set:

    Not applicable, as this is a physical device, not an AI model.

    9. How Ground Truth for Training Set was Established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140379

    Validate with FDA (Live)

    Date Cleared
    2014-07-01

    (137 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Mega Big Boy condom is used for contraceptive and prophylactic purpose (to help prevent pregnancy, HIV/ADS and the transmission of sexually transmitted infections).

    Device Description

    This condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. This condom is smooth surface, flared at the closed end with reservoir tip, silicone lubricated condom with nominal length 200 mm, nominal flat width 57 mm, and nominal thickness of 0.065 mm. The air burst pressure is > 1 kPa and air burst volume is > 19 dm3. It is lubricated with silicone (viscosity 350 cps) and cornstarch is used as a dressing material. This condom conforms to current established national standard ASTM D3492: 2008.

    AI/ML Overview

    The provided text is a 510(k) summary for a male latex condom called "Mega Big Boy Condom." It does not describe a study involving an AI/Machine Learning device or a diagnostic/screening technology. Therefore, I cannot extract the information required to populate the requested sections about acceptance criteria, study design, ground truth establishment, or human-in-the-loop performance, as this type of information is not present in the provided document.

    The document focuses on demonstrating substantial equivalence of the "Mega Big Boy Condom" to a predicate device (TROJAN EXTRA LARGE LATEX CONDOM) based on technological characteristics and performance standards for condoms (e.g., air burst pressure, air burst volume, shelf life, biocompatibility testing).

    Therefore, I must state that the requested information (acceptance criteria for an AI device, study details for its performance, expert adjudication, etc.) cannot be extracted from the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090259

    Validate with FDA (Live)

    Date Cleared
    2009-03-31

    (57 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections).

    Device Description

    This condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. This device is smooth surfaced, straight-walled, teat-ended, silicone lubricated condom with minimum length 160mm, maximum width 54mm, and minimum thickness of 30μm as required by ASTM Standard D3492.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a new condom product, OKAMOTO ULTRA THIN Male Natural Latex Condom. However, it does not contain a detailed study report with specific acceptance criteria, sample sizes for test or training sets, ground truth establishment, or multi-reader multi-case studies as requested in the prompt.

    The document primarily focuses on demonstrating substantial equivalence to existing predicate devices based on technological characteristics and adherence to ASTM standards for condom manufacturing.

    Here's a breakdown of what can be extracted and what information is missing based on your request:

    1. A table of acceptance criteria and the reported device performance

    The document states the device's design is in conformance with ASTM Standard D3492-03 Specification for Rubber Contraceptives (Male Condoms). While the specific acceptance criteria from this standard are not explicitly listed in the text, the document implies that the device performance meets these standards.

    The key performance characteristic mentioned for the new device is:

    • Nominal thickness: 0.04 +/- 0.01 mm

    For predicate devices, the nominal thickness was 0.05 +/- 0.01 mm.

    Missing Information: The explicit breakdown of all acceptance criteria from ASTM Standard D3492-03 (e.g., burst pressure, freedom from holes, dimensions, etc.) and the precise reported performance values for each of these criteria for the OKAMOTO ULTRA THIN condom are not present in this document.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Missing Information: The document does not specify any sample size for a test set, nor does it describe data provenance (country of origin, retrospective/prospective). This type of detail would typically be found in a more comprehensive testing report.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Missing Information: This information is not applicable and therefore not present. The device is a physical product (condom), and its performance is evaluated against engineering standards rather than expert-established ground truth like in medical imaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Missing Information: Not applicable. Performance testing for condoms against ASTM standards does not typically involve adjudication methods like those used for expert consensus in medical diagnostics.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Missing Information: This type of study is entirely irrelevant to a physical product like a condom and is therefore not present. It's a concept used for AI-assisted diagnostic tools.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Missing Information: This is also irrelevant to a physical product and is not present.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance would be the objective measurements taken on the condoms themselves against the specifications defined in ASTM Standard D3492-03. This standard outlines various tests (e.g., burst tests, visual inspections for holes,
    dimensional measurements) and acceptable limits.

    8. The sample size for the training set

    Missing Information: This is not applicable to the development of a physical product like a condom. There is no "training set" in the context of an AI/algorithm. The "training" in manufacturing would relate to process control and quality assurance, not a dataset for an algorithm.

    9. How the ground truth for the training set was established

    Missing Information: Not applicable for the reasons stated above.

    In summary, the provided text is a regulatory submission for premarket clearance of a physical medical device (condom) and therefore does not contain the types of information (AI study data, expert consensus, MRMC studies) that would be relevant for software/AI-based diagnostic devices. The "study" referenced implicitly is the adherence to the ASTM standard for condom manufacturing and performance.