Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K050237

    Validate with FDA (Live)

    Date Cleared
    2005-03-18

    (44 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PoleStar system is a magnetic resonance diagnostic device intended for general diagnostic use to present images, which reflect the spatial distribution and/or magnetic resonance spectra, which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance.

    The StarShield V3 is intended to create an area of substantially reduced Electromagnetic Interference (EMI) in the area enclosed by the RF Tent during image acquisition by the PoleStar system.

    Device Description

    The StarShieldV3 is an accessory of Odin's PoleStar Intraoperative Magnetic Resonance system. It is intended to create an area of substantially reduced Electromagnetic Interference (EMI) in the area enclosed by the RF Tent, required Elootromagnotion The RF Tent is a portable modular structure that consist of four principal components:

    • The foldable frame .
    • The cover .
    • The control unit .
    • The floor .

    The RF Tent is not connected to the PoleStar gantry or the OR table.

    AI/ML Overview

    The provided text is a 510(k) summary for the StarShield V3 device, which is an accessory to an Intraoperative Magnetic Resonance system. The submission aims to demonstrate substantial equivalence to a predicate device (Odin's RF Tent). The primary function of the StarShield V3 is to reduce Electromagnetic Interference (EMI) during MRI image acquisition.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" for the StarShield V3 in the typical sense of performance metrics with thresholds. Instead, it provides a "Substantial Equivalence Comparison Table" that lists parameters of the StarShield V3 and compares them to its predicate device, the RF Tent (K022157). The implicit acceptance criterion is that the StarShield V3 performs equivalently to the predicate device in terms of EMI reduction.

    ParameterPredicate Device (RF Tent - K022157)StarShield V3Acceptance/Performance
    ApplicationTo create an area of substantially reduced Electromagnetic Interference (EMI) in the area enclosed by the RF Tent during image acquisition by the PoleStar systemSameEquivalent
    Materials - FrameAluminum alloy AlMg Si 0.5SameEquivalent
    Materials - Cover(Polyethylene Terephthalate) PET incorporating Cu+Ni, laminated with flame-retardant PVC.SameEquivalent
    Dimensions - Length (Folded)950 mmSameEquivalent
    Dimensions - Length (Extended)3500 mmSameEquivalent
    Dimensions - Width/Base1250 mmSameEquivalent
    Dimensions - Maximum height1650 mmSameEquivalent
    Clearance between the RF tent and the operating table - Foot side (height)200 mmSameEquivalent
    Clearance between the RF tent and the operating table - Width (each side)100 mmSameEquivalent
    Clearance between the RF tent and the operating table - Head side (distance from scanner)70 mmSameEquivalent
    Average Attenuation>40dB at 100 MHzSameEquivalent
    FloorMatte aluminum alloy sheet measuring 3.8 x 1.5 mSameEquivalent
    Energy usedPneumaticElectricDifferent (Minor)
    Where UsedOperating RoomSameEquivalent
    BiocompatibilityDoes not contact patientSameEquivalent

    Note regarding "Acceptance/Performance": For most parameters, "Same" indicates the StarShield V3 meets the performance of the predicate by having identical specifications. For "Average Attenuation," the ">40dB at 100 MHz" is a specific performance metric that the StarShield V3 also claims as "Same," implying it meets or exceeds this level of attenuation. The only notable difference is "Energy used," which changed from pneumatic to electric, but the document implicitly suggests this change does not affect the substantial equivalence in terms of the primary function (EMI reduction).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not provide information on a test set, its sample size, or data provenance. This 510(k) submission is for a physical medical device (an EMI shielding tent), not a software or diagnostic algorithm that would typically require a dataset to evaluate its performance. The "study" here is primarily a comparison of specifications and physical attributes to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. As this is a physical device intended for EMI reduction, there is no "ground truth" derived from expert consensus on medical images or diagnoses. The performance is based on engineering specifications (e.g., attenuation levels).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There is no test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-powered diagnostic tool, and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant to its purpose.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The StarShield V3 is a physical accessory, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the StarShield V3's performance, as implied by the comparison table, is based on engineering specifications and measurements related to EMI attenuation. The predicate device (RF Tent) established a benchmark of ">40dB at 100 MHz" for average attenuation. The StarShield V3 is declared to have the "Same" attenuation, meaning it is expected to meet this engineering standard.

    8. The sample size for the training set

    Not applicable. This device is not a machine learning model, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032541

    Validate with FDA (Live)

    Device Name
    POLESTAR N-20
    Date Cleared
    2003-09-12

    (25 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.

    Anatomical regions: extremities and selected sections of the head.

    Nuclei excited: H-1

    Diagnostic uses: T1, T2, T2* and density weighted imaging.

    The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.

    Device Description

    The PoleStar N-20 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N-20 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N-20 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.

    AI/ML Overview

    This document is a 510(k) summary for the Odin Medical Technologies PoleStar N-20, an intraoperative MRI device. The primary claim for substantial equivalence is based on a comparison to the predicate device, Odin's PoleStar N-10. This filing does not describe a study to prove acceptance criteria for device performance. Instead, it focuses on demonstrating that the PoleStar N-20 is substantially equivalent to an existing device (the PoleStar N-10) based on similar technical specifications and identical intended use.

    Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and statistical analysis is not present in the provided text.

    Here's a breakdown of what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide explicit "acceptance criteria" for performance that would typically be seen in a clinical validation study for a new diagnostic device. Instead, it offers a "Substantial Equivalence Comparison Table" that lists technical specifications of the new device (PoleStar N-20) against its predicate (PoleStar N-10). The "performance" in this context refers to these specifications matching or being comparable to the predicate.

    ParameterPoleStar N-10 (Predicate, K010850)PoleStar N-20 (New Device)Performance/Comparison
    Clinical applicationExtremities and selected sections of the headExtremities and selected sections of the headIdentical
    Magnet typePermanentPermanentIdentical
    Field strength0.12T0.13TSlight increase, deemed comparable
    5 gauss fringe field (radial/axial, m)1.52.2Minor difference, likely considered acceptable for similar use
    ShimmingPassive, activePassive, activeIdentical
    Gradient subsystem Strength mT/m2522Minor difference, likely considered acceptable
    Gradient subsystem Rise time to 10mT/m msec<1<1Identical
    Computer system - CPUPentium 586Pentium 586Identical
    Computer system - Memory size [MB]6464Identical
    Array processor4xDSP C44 TI4xDSP C44 TIIdentical
    Array processor - Memory size [MB]40004000Identical
    Storage mediaMagnetic disk, floppy diskMagnetic disk, floppy diskIdentical
    Number of images stored50005000Identical
    Imaging modesSingle, multislice, volume studySingle, multislice, volume studyIdentical
    Reconstruction time - single slice, sec<3/slice<3/sliceIdentical
    Reconstruction time - multislice, sec<3/slice<3/sliceIdentical
    Reconstruction time - volume sec<20/volume<20/volumeIdentical
    Cardiac gating (ECG/peripheral)NoNoIdentical
    Respiratory gatingNoNoIdentical
    AngiographyOptionalOptionalIdentical
    SpectroscopyNoNoIdentical
    Imaging - pulse sequenceSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DIdentical
    Imaging - repetition time, msec10-5000 increments of 110-5000 increments of 1Identical
    Imaging - echo time, msec3-1503-150Identical
    Imaging - inversion time, msecN/AN/AIdentical
    Imaging - slice thickness, mm2-102-10Identical
    Imaging - scan orientationTransverse, coronal, sagittal, obliqueTransverse, coronal, sagittal, obliqueIdentical
    Imaging - measuring matrix64x64 to 256x256 steps of 1 in phase encoding64x64 to 256x256 steps of 1 in phase encodingIdentical
    Imaging - display matrix1024x7681024x768Identical
    Imaging - pixel intensity0-40950-4095Identical
    Surface coilsKnee, extremity, headKnee, extremity, headIdentical
    Bore diameter or WxH, cm24.5x3925.2x42Minor difference, likely considered acceptable
    Bore featuresOpen access to patientOpen access to patientIdentical
    Cooling system typeClosed loop water cooling (Gradients only)Closed loop water cooling (Gradients only)Identical
    Cryogen useNoNoIdentical
    Magnet weight, kg330400Minor difference, likely considered acceptable
    HxWxD, cm145x96x120153x97x120Minor difference, likely considered acceptable
    Field Of View (FOV), cm5-185-20Minor difference, likely considered acceptable
    Dicom 3.0 interfaceYesYesIdentical
    Power requirements3x208V, 16 Kva, <10000 BTU/hr A/C3x208V, 15 Kva, <10000 BTU/hr A/CMinor difference, likely considered acceptable

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable. This document does not describe a clinical study with a test set of data. The substantial equivalence is based on technical specifications and intended use comparison.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. No ground truth establishment is described as there is no clinical test set in this filing. The filing mentions that "When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis," which refers to the eventual use of the device, not a study performed for this 510(k).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes an MRI diagnostic device, not an AI or CAD system, and therefore, an MRMC study related to AI assistance would not be relevant or expected for this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a hardware device (MRI scanner) which requires human operation and interpretation. The concept of standalone algorithm performance does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable. As no clinical study or test set is described, no ground truth was established for this 510(k filing for the PoleStar N-20.**

    8. The sample size for the training set:

    • Not Applicable. This document does not describe a machine learning or AI-based device that would require a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set or ground truth for a training set is described.

    In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence of the PoleStar N-20 to its predicate, the PoleStar N-10, by comparing technical specifications and intended use. It does not include information about performance acceptance criteria validated through a clinical study with a test set, expert ground truth establishment, or AI-related metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022157

    Validate with FDA (Live)

    Device Name
    RF TENT
    Date Cleared
    2002-08-28

    (57 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RF Tent is intended to create an area of substantially reduced Electromagnetic Interference (EMI) in the area enclosed by the RF Tent during image acquisition by the PoleStar N-10 (K002242).

    Device Description

    To acquire high-quality MR images. it is necessary to isolate the MRI scanner from external sources of EMI (Electromagnetic Interference). Generally, copper sheeting is installed in the OR walls to isolate the OR from external interference. This solution is relatively expensive, and also requires the OR to be out of use during the installation process. Once the copper sheeting is installed, all procedures involving the PoleStar N-10 intraoperative MRI must be performed in the dedicated shielded room.

    The RF Tent presented in this premarket notification expands the capability of the PoleStar N-10 (K002242), by providing a low cost, simple and effective alternative to the fixed copper shielding. Because it is portable, it also allows the PoleStar N-10 to be used in more than a single dedicated OR. Thereby increasing the flexibility of the use of the system.

    AI/ML Overview

    The provided text describes the "RF Tent" device, which is an accessory designed to reduce Electromagnetic Interference (EMI) during MRI image acquisition. The document focuses on its substantial equivalence to existing solutions (installed copper sheeting) rather than providing detailed acceptance criteria and a comprehensive study report in the format typically seen for algorithm-based devices.

    Therefore, many of the requested fields cannot be directly populated from the provided text because the device is a physical RF shielding enclosure, not an AI/algorithm-based diagnostic device. The performance data section is very brief.

    Here's an attempt to extract the available information and indicate where data is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Create an area of substantially reduced Electromagnetic Interference (EMI)"Similar attenuation levels" to installed copper sheeting.
    Acquire high-quality MR images by isolating the MRI scanner from external sources of EMI"Signal to Noise results are provided in Attachment 2 of the 510(k) submission." (Specific values not provided in this document)
    Proper function of the RF Tent"We believe that the RF Tent can be safely and effectively used."

    2. Sample size used for the test set and data provenance

    • Test Set Sample Size: Not explicitly stated. The "Performance Data" section refers to "Signal to Noise results" (plural), suggesting multiple data points or tests were conducted, but the number of tests or specific scenarios is not given.
    • Data Provenance: Not explicitly stated. The tests were performed by "Odin Medical Technologies." It's likely conducted in a controlled lab or clinical setting, but country of origin or whether it's retrospective/prospective is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

    • This information is not applicable and not provided. The "ground truth" for an RF shielding device would be based on objective physical measurements of EMI attenuation and Signal-to-Noise Ratio (SNR) in MR images, not expert interpretation of diagnostic images.

    4. Adjudication method for the test set

    • Not applicable and not provided. See point 3 above.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an RF shielding enclosure, not an AI or diagnostic algorithm that human readers would interact with. No MRMC study was performed or discussed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used

    • The implicit ground truth is based on physical measurements of Electromagnetic Interference (EMI) attenuation and resulting Signal-to-Noise Ratio (SNR) in Magnetic Resonance (MR) images. The text mentions "Signal to Noise results" as performance data.

    8. The sample size for the training set

    • Not applicable. This is a physical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is a physical device, not a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020601

    Validate with FDA (Live)

    Date Cleared
    2002-04-24

    (61 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A magnetic resonance diagnostic device is intended for general diagnostic use to present images, which reflect the spatial distribution and/or magnetic resonance spectra, which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance.

    The FlexiCoil - malleable receive-only surface coil expands the capability of the PoleStar N-10 MR Imaging System (K002242). It improves imaging of selected sections of the head and extremities.

    Device Description

    The FlexiCoil is a malleable receive-only surface coil intended to facilitate intraoperative MR imaging of selected sections of the head and extremities. This component is a malleable circular enclosed coil that consists of copper straps covered with Medical Foam Tape and an electronic tuning box, which can be placed around the anatomy of interest for MR imaging. The malleable structure of the coil allows adjustment of size and shape by pulling/releasing the plastic holders on both sides of the coil, providing the highest possible contact between the coil and the area to be imaged.

    Due to its malleability the FlexiCoil can cover a wider area than the presently used coils (K991243) resulting in better image uniformity and a better overall SNR when imaging structures near the surface of the patient's head or extremity. The improved SNR is a result of the coil being closer to the signal-emitting anatomy, and being fully tuned and matched to the signal and receive channel. In addition the coil can be easily placed near the anatomy to be imaged with little or no patient post procedure discomfort. The FlexiCoil is for single use only and is supplied non-sterile.

    AI/ML Overview

    The provided 510(k) summary for the FlexiCoil - malleable receive-only surface coil does not contain a detailed study report with acceptance criteria and specific performance metrics. Instead, it focuses on demonstrating substantial equivalence to predicate devices based on intended use and technological characteristics, and mentions "Performance Data: Sample images were provided in attachment 4 of the 510(k) submission." This indicates a qualitative assessment or side-by-side comparison of image quality rather than quantitative performance against predefined acceptance criteria.

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method for the test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study, its effect size, or human reader improvement with AI.
    • Standalone (algorithm only) performance.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for the test set.
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    Based on the provided document, the device met the acceptance criteria by demonstrating substantial equivalence, not through a formal study with predefined quantitative acceptance metrics.

    The document states:
    "Substantial equivalence was based on performance data. Sample images were provided in attachment 4 of the 510(k) submission."
    And:
    "The PoleStar N-10 operated with the FlexiCoil - malleable receive-only surface coil adressed in this premarket notification, has the same intended use and technological characteristics as the same system operated with the legally marketed coil cleared in K991243 and does not affect the PoleStar N-10 system safety parameter specification. The FlexiCoil also has the the same intended use and technological characteristics as the other OEM predicate coil."

    This indicates that the "performance data" likely involved a comparison of images produced by the FlexiCoil with those from the predicate devices to show similar image quality and diagnostic usefulness, rather than meeting specific quantitative performance thresholds. The FDA's clearance is based on this demonstration of substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K010850

    Validate with FDA (Live)

    Device Name
    POLESTAR N-10
    Date Cleared
    2001-12-17

    (271 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1