    K Number
    K210025

    Validate with FDA (Live)

    Device Name
    INVU by Nuvo
    Manufacturer
    Date Cleared
    2021-05-28

    (144 days)

    Product Code
    Regulation Number
    884.2730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INVU by Nuvo™ is a maternal-fetal monitor that non-invasively measures and displays fetal heart rate (FHR), maternal heart rate (MHR) and Uterine Activity (UA).
    The INVU Sensor Band™ acquires the fetal heart electrocardiogram and maternal heart electrocardiogram signals from abdominal surface electrodes and the fetal phonocardiogram and the maternal phonocardiogram signals from surface acoustic sensors. The FHR, MHR and UA tracings are derived from these signals and presented.
    INVU by Nuvo™ is indicated for use by pregnant women who are in their 32nd week of gestation (or later), with a singleton pregnancy.
    The INVU by Nuvo™ maternal-fetal monitor is intended for use by healthcare professionals in health care facilities and by the patient in the patient's home, on the order of a physician.
    The INVU by Nuvo™ is indicated for antepartum fetal surveillance (i.e. non-stress testing).
    This system does not prevent the onset of preterm labor nor will it prevent the occurrence of preterm birth.

    Device Description

    INVU by Nuvo™ is a non-invasive medical device that acquires and displays vital signs of the pregnant woman and her fetus. It measures and processes signals picked up on the abdominal surface using highly sensitive sensors, special electronic circuitry, and processing software. Two types of sensors pick up the signals: ECG-like sensors that pick-up bio-potential signals, and acoustic sensors. The bio-potential (ECGlike) sensors pick up the fECG (of the fetus), the mECG signals (of the pregnant woman) and the acoustic sensors measure the sounds from the pregnant woman's abdomen (PCG -phonocardiogram and fPCG fetal PCG). The FHR, MHR, and UA tracings are derived from these signals and presented.
    lt is an integrated platform that uses a signal acquisition tool to provide input related to fetal heart rate (FHR), maternal heart rate (MHR), to two separate software applications, one for the patient (INVU App) and one for the physician (INVU-Pro application), which also provides uterine activity (UA) tracings. Sensors are incorporated in a belt that is worn on the abdomen of the pregnant woman, where it acquires both biopotential and acoustic signals. The signals are processed at a cloud-server level.
    The sensors are attached to the pregnant woman's abdomen by a belt (INVU by Nuvo™ sensory band). An integrated electronic module receives, processes and transmits the measured data to algorithmic modules at cloud servers, where the inputs from the sensors are processed, merged, and sent to the INVU by Nuvo™ (patient) application to display the average MHR after a 30 minute (minimum) monitoring session and to the INVU by Nuvo™ Pro (physician) application to display the complete FHR, MHR, and uterine activity data from a monitoring session can only be scheduled by a health care provider.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device: INVU by Nuvo™ (Maternal-fetal monitor)

    Performance Metrics of Interest for this Documentation: Uterine Activity (UA) detection. The text mentions FHR and MHR performance was addressed in a prior K-submission (K191401), so this summary focuses on UA as per the provided study details.


    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter/MetricAcceptance Criteria (Performance Goal)Reported Device Performance (INVU by Nuvo™)Context for Comparison (TOCO Device)
    Positive Percent Agreement (PPA) with IUPC (Gold Standard)Lower 95% Confidence Bound of PPA > 75%84.80% (95% CI: [81.58%; 88.02%])37.50% (95% CI: [28.23%; 46.77%])
    False Positive Rate (FPR) with IUPC (Gold Standard)Not explicitly stated as a numerical acceptance criterion, but implicitly evaluated for comparability.24.28% (95% CI: [20.46%; 28.11%])10.69% (95% CI: [5.65%; 15.72%])

    Conclusion from Table: INVU by Nuvo™ met its performance goal for Positive Percent Agreement. The False Positive Rate was higher than the TOCO device, but the overall conclusion stated that INVU™ is comparable to the standard of care TOCO device and meets the required accuracy for its intended use.


    2. Sample Size and Data Provenance

    • Test Set Sample Size: 80 (Validation Phase)
    • Data Provenance: Not explicitly stated regarding country of origin. The study was a "pivotal clinical study," suggesting it was prospective.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. (The ground truth was established by a medical device, IUPC, not human experts for this specific test.)

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable/None mentioned for the UA data itself. The ground truth for Uterine Activity was established directly by an Internal Uterine Pressure Catheter (IUPC), which is the "gold standard" device. There's no indication of human adjudication of this instrumental ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. This study focuses on the device's accuracy in detecting Uterine Activity compared to an instrumental gold standard and a standard-of-care device. It does not evaluate human reader improvement with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    • Was a standalone performance study done? Yes. The study directly assessed the INVU by Nuvo™ platform's ability to collect and display UA data and compared its performance (specifically the algorithm's output) against the IUPC (gold standard) and TOCO devices.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Instrumental Gold Standard. Uterine activity was measured by an Internal Uterine Pressure Catheter (IUPC), which is explicitly referred to as the "gold standard" for uterine activity measurement.

    8. Sample Size for the Training Set

    • Training Set Sample Size: 40

    9. How the Ground Truth for the Training Set was Established

    • Method for Training Set Ground Truth: The text states, "During the training phase, uterine activity data was collected to 'train' the UA algorithm to identify uterine contractions." While not explicitly stated, it is highly implied that the same method for ground truth establishment as the validation phase (IUPC) would have been used to provide accurate Uterine Activity data for algorithm training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191401

    Validate with FDA (Live)

    Device Name
    PregSense
    Manufacturer
    Date Cleared
    2020-03-27

    (308 days)

    Product Code
    Regulation Number
    884.2730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PregSense is a maternal-fetal monitor that non-invasively measures and displays fetal heart rate (FHR) and maternal heart rate (MHR). The PregSense acquires and displays the FHR and MHR tracings from abdominal surface electrodes that pick up the fetal heart biopotential and maternal heart biopotential signal, and from surface acoustic sensors that pick up the fetal PCG (fPCG; phonocardiogram) and the maternal PCG (mPCG; phonocardiogram) signals.

    PregSense is indicated for use by pregnant women who need documentation of fetal heart rate activity, and who are in their 32nd week of gestation (or later), with a singleton pregnancy. PregSense is intended to be used for a maximum of five minutes.

    The PregSense maternal-fetal monitor is intended for use in the antepartum period by healthcare professionals in health care facilities and by the patient in the patient's home, on the order of a physician.

    The PregSense is not intended for use in critical care situations or in laboring patients hospitalized for or suspected to have preterm labor.

    PregSense is not intended to be used for antepartum monitoring (e.g., non-stress testing).

    Device Description

    PregSense™ is a non-invasive medical device that acquires and displays vital signs of the pregnant woman and of her fetus. It measures and processes signals picked up on the abdominal surface using sensors, electronic circuitry and processing software. Two types of sensors pick up the signals: electrocardiogram (ECG)-like sensors that capture bio-potential signals, and acoustic sensors. The biopotential (ECG-like) sensors capture fECG (heartrate of the fetus) and mECG signals (heartrate of the pregnant woman). The acoustic sensors measure the sounds from the pregnant woman's abdomen, (PCG -phonocardiogram and fPCG -fetal PCG). Monitoring of the fetal and maternal heart rate using PregSense is limited to a five-minute session.

    PregSense™ is an integrated platform that uses a signal acquisition tool to provide input to two separate software applications, one for the patient (PregSense™ ME) and one for the physician (PregSense™ MD). The sensors are incorporated in a belt (PregSense™ Belt) that is worn on the abdomen of the pregnant woman, where it acquires both biopotential and acoustic signals are processed at the cloud-server level where the inputs from the sensors are processed, merged and downloaded to the mobile devices of the pregnant woman and her health care provider. The PregSense™ ME application (for the patient) allows the pregnant woman to view the average maternal and fetal heart rate after a five minute session has been completed and PregSense™-MD application (for the health care provider) allows the health care provider to view the complete fetal and maternal heart rate data from the five minute session online and remotely via the internet. A monitoring session can only be initiated by a health care provider.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Fetal Heart Rate (FHR) Measurement AccuracyFor a 30-minute session, 97.08% of all differences between PregSense and a "standard of care" monitoring device (CTG) lie between the 95% agreement limits of [-8.84, 8.24] bpm. For a 5-minute session (relevant to the device's indicated use): The lower limit of the 95% confidence interval of the lower agreement bound is -7.47 bpm and the upper limit is 8.16 bpm. Furthermore, 96.64% of the differences fall between the limits of agreement.
    Maternal Heart Rate (MHR) Measurement AccuracyFor a 30-minute session, 95.31% (48241/50616) of all differences between PregSense and a "standard of care" monitoring device (CTG) lie between the 95% agreement limits of [-5.30, 5.86] bpm.
    Overall FHR Detection Percentage (Feasibility Study)> 70% for the overall population. 90% (±11.3) for subjects at least 32 weeks of gestation.
    Device Usability by Lay UsersSuccessfully demonstrated through a usability study.
    SafetyNo adverse events (procedure-related or device-related) reported during the feasibility study with 76 subjects. No adverse events reported in the pivotal clinical testing with 149 subjects.

    Study Details:

    1. Sample sized used for the test set and the data provenance:

      • Feasibility Study: 76 subjects. The text mentions "two in the United States and two outside the United States" for the pivotal study settings, which suggests a global data provenance, but it's not explicitly stated for the feasibility study. The feasibility study seems to have been retrospective based on the reporting ("510 recording sessions were executed").
      • Pivotal Study: 149 subjects. The study was conducted in four healthcare settings (two in the United States and two outside the United States), indicating a prospective, multi-site approach.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test set. However, the pivotal study compared PregSense's performance to a "standard of care" monitoring device (CTG), implying that the CTG readings served as the reference standard, which would typically be interpreted by trained healthcare professionals.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      The document does not specify an adjudication method like 2+1 or 3+1. The comparison was directly against a "standard of care" CTG system.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study focuses on the standalone performance of the PregSense device in measuring FHR and MHR compared to a standard, rather than evaluating how human readers' performance changes with AI assistance.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, the performance data presented is for the standalone device (PregSense™ system, including its embedded sensors and associated software for analysis) compared to a "standard of care" CTG. The "PregSense ME App" and "PregSense MD App" are for displaying data, not for human interpretation during the measurement process that would alter the algorithm's output. The usability study involved lay users, but that was to ensure proper device handling, not to assess a human-in-the-loop diagnostic process.

    6. The type of ground truth used:
      The ground truth was established by comparison to a "standard of care" monitoring device, specifically CTG (Cardiotocography). This is a widely accepted clinical method for monitoring fetal heart rate.

    7. The sample size for the training set:
      The document does not provide information regarding the sample size used for the training set for the PregSense algorithms. It only discusses the clinical studies conducted for performance validation.

    8. How the ground truth for the training set was established:
      The document does not provide information on how the ground truth for the training set was established, as details about the training set itself are absent.