Search Results
Found 7 results
510(k) Data Aggregation
(115 days)
DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, a patient's home, or an ambulatory setting. It is indicated for use with adults (18 years and above) in a clinical environment by or on the order of a medical professional.
DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotations of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); and infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional.
DeepRESP is a cloud-based software as a medical device (SaMD), designed to perform analysis of sleep study recordings, with and without EEG signals, providing data for the assessment and diagnosis of sleep-related disorders. Its algorithmic framework provides the derivation of sleep staging, including arousals, scoring of respiratory events, and key parameters such as the Apnea-Hypopnea Index (AHI) and Central Apnea-Hypopnea Index (CAHI).
DeepRESP (K252330) is hosted on a serverless stack. It consists of:
- A web Application Programming Interface (API) intended to interface with a third-party client application, allowing medical professionals to access DeepRESP's analytical capabilities.
- Predefined sequences called Protocols that run data analyses, including artificial intelligence and rule-based models for the scoring of sleep studies, and a parameter calculation service.
- A Result storage using an object storage service to temporarily store outputs from the DeepRESP Protocols.
Here's a breakdown of the acceptance criteria and study details for DeepRESP, based on the provided FDA 510(k) clearance letter:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria (e.g., "DeepRESP must achieve AHI ≥ 5 PPA% of at least X%"). Instead, it reports the performance of the device and compares it to predicate devices to demonstrate substantial equivalence and non-inferiority. The "Observed paired differences" columns, particularly those where the confidence interval does not cross zero, imply a comparison to show that the new DeepRESP v2.0 performs at least as well as the previous version or the additional predicate.
For the purpose of this analysis, I will present the reported performance of the Subject Device (DeepRESP v2.0) as the "reported device performance." Since no explicit acceptance criteria thresholds are given, the comparison to predicates and the demonstration of non-inferiority served as the implicit acceptance path for the FDA clearance.
Reported Device Performance (DeepRESP v2.0)
| Metric | Type I/II Studies (EEG) Reported Performance PPA% (NPA%, OPA%) | Type III HSAT-Flow Studies Reported Performance PPA% (NPA%, OPA%) | Type III HSAT-RIP Studies Reported Performance PPA% (NPA%, OPA%) |
|---|---|---|---|
| Severity Classification | |||
| AHI ≥ 5 | 87.7 (76.5, 87.3) | 91.0 (78.0, 90.6) | 93.7 (63.5, 92.8) |
| AHI ≥ 15 | 71.9 (94.8, 78.2) | 78.1 (93.9, 81.7) | 81.0 (91.1, 83.4) |
| CAHI ≥ 5 | 80.0 (98.0, 97.2) | 80.7 (98.0, 97.2) | 79.5 (97.6, 96.9) |
| Sleep Stages | |||
| Wake | 92.8 (95.8, 95.1) | 79.7 (96.6, 92.9) | 79.7 (96.6, 92.9) |
| REM | 82.5 (98.8, 96.5) | 77.0 (98.1, 95.2) | 77.0 (98.1, 95.2) |
| NREM1 | 43.1 (94.5, 91.7) | N/A (Only NREM total reported for Type III studies) | N/A (Only NREM total reported for Type III studies) |
| NREM2 | 78.1 (91.5, 85.3) | N/A (Only NREM total reported for Type III studies) | N/A (Only NREM total reported for Type III studies) |
| NREM3 | 87.5 (94.6, 93.7) | N/A (Only NREM total reported for Type III studies) | N/A (Only NREM total reported for Type III studies) |
| NREM (Total for Type III) | N/A | 94.2 (80.1, 89.1) | 94.2 (80.1, 89.1) |
| Respiratory Events | |||
| Respiratory events (overall) | 71.2 (93.2, 85.0) | 74.4 (92.0, 85.5) | 75.0 (90.7, 84.8) |
| All apnea | 83.7 (98.2, 97.1) | 84.5 (98.2, 97.0) | 81.1 (95.7, 94.5) |
| Central apnea | 79.3 (99.2, 99.0) | 77.5 (99.2, 99.0) | 78.8 (99.2, 99.0) |
| Obstructive apnea | 76.2 (98.4, 97.0) | 78.4 (98.4, 97.0) | 74.3 (96.0, 94.4) |
| Hypopnea | 60.1 (92.9, 83.5) | 63.9 (91.7, 83.3) | 58.9 (90.7, 81.0) |
| Desaturation | 98.5 (95.5, 96.1) | 98.8 (96.3, 96.9) | 98.8 (96.3, 96.9) |
| Arousal events | 62.1 (89.1, 81.5) | 64.0 (90.5, 83.1) | 64.0 (90.5, 83.0) |
PPA%: Positive Percent Agreement, NPA%: Negative Percent Agreement, OPA%: Overall Percent Agreement.
2. Sample Sizes and Data Provenance
The clinical validation was conducted using retrospective data.
-
Test Set Sample Size:
- Type I/II Scoring Validation: 4,030 PSG recordings
- Type III Scoring Validation: 5,771 sleep recordings
- This comprised 4,037 Type I recordings and 1,734 Type II recordings, processed as Type III by using only the relevant subset of signals.
-
Data Provenance:
- Country of Origin: United States.
- Data Type: Manually scored sleep recordings from sleep clinics, collected as part of routine clinical work for patients suspected of suffering from sleep disorders.
- Settings: Urban, suburban, and rural areas.
- Demographics: Included individuals in all age groups (18-21, 22-35, 36-45, 46-55, 56-65, >65) and all BMI groups (<25, 25-30, <30). The recording collection for Type I/II scoring consisted of 44% Females, and for Type III scoring, 35% Females. High level of race/ethnicity diversity (Caucasian or White, Black or African American, Other).
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not explicitly stated. The document refers to "manually scored sleep recordings" and "medical professional" for verification. It also mentions "board-certified sleep physicians" in the context of the training set. However, the specific number of experts used to establish the ground truth for the test set is not detailed.
- Qualifications of Experts: For the test set, it's implied that "medical professionals" performed the manual scoring, as the data originated from "routine clinical work." For the training set, "board-certified sleep physicians" were involved in establishing the ground truth.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for establishing the ground truth on the test set. It mentions "manually scored sleep recordings" but does not detail how potential disagreements between multiple scorers (if any were used) were resolved.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.
The study design was a retrospective data study comparing the automatic scoring of DeepRESP to manual scoring (ground truth) and also comparing DeepRESP's performance to two predicate devices (DeepRESP K241960 and Nox Sleep System K192469). This is a standalone performance evaluation against expert-derived ground truth, with a direct comparison to existing automated systems, not an MRMC study assessing human reader improvement with AI assistance.
6. Standalone Performance Study
Yes, a standalone performance study was done. The reported PPA, NPA, and OPA values for DeepRESP v2.0 (Subject Device) represent its performance as a standalone algorithm without human-in-the-loop assistance. The subsequent comparison to the predicate devices also evaluated their standalone performance. The document explicitly states: "All output is subject to verification by a medical professional," indicating that while the device is intended to aid in diagnosis, its performance evaluation was conducted on its automated output before any human review.
7. Type of Ground Truth Used
The ground truth used was expert consensus (manual scoring). The study used "manually scored sleep recordings" from "routine clinical work" as the reference standard against which DeepRESP's automatic scoring was compared.
8. Sample Size for the Training Set
The document does not report the sample size used for the training set. It only states the sample sizes for the validation (test) sets: 4,030 PSG recordings for Type I/II validation and 5,771 sleep recordings for Type III validation.
9. How the Ground Truth for the Training Set was Established
The document states that the ground truth for the training set "was established through a rigorous process involving multiple board-certified sleep physicians." This implies an expert-driven process, likely involving consensus or reconciliation among several highly qualified professionals. However, the exact methodology (e.g., number of physicians, adjudication rules) is not detailed beyond "rigorous process."
Ask a specific question about this device
(254 days)
DeepRESP is an aid in the diagnosis of various sleep disorders where subjects are often evaluated during the initiation or follow-up of treatment of various sleep disorders. The recordings to be analyzed by DeepRESP can be performed in a hospital, patient home, or an ambulatory setting. It is indicated for use with adults (22 years and above) in a clinical environment by or on the order of a medical professional.
DeepRESP is intended to mark sleep study signals to aid in the identification of events and annotation of traces; automatically calculate measures obtained from recorded signals (e.g., magnitude, time, frequency, and statistical measures of marked events); infer sleep staging with arousals with EEG and in the absence of EEG. All output is subject to verification by a medical professional.
DeepRESP is a cloud-based software as a medical device (SaMD), designed to perform analysis of sleep study recordings, with and without EEG signals, providing data for the assessment and diagnosis of sleep-related disorders. Its algorithmic framework provides the derivation of sleep staging including arousals, scoring of respiratory events and key parameters such as the Apnea-Hypopnea Index (AHI).
DeepRESP is hosted on a serverless stack. It consists of:
- A web Application Programming Interface (API) intended to interface with a third-party client application, allowing medical professionals to access DeepRESP's analytical capabilities.
- Predefined sequences called Protocols that run data analyses, including artificial intelligence and rule-based models for the scoring of sleep studies, and a parameter calculation service.
- A Result storage using an object storage service to temporarily store outputs from the DeepRESP Protocols.
Here's a breakdown of the acceptance criteria and the study details for the DeepRESP device, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria & Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" as a separate table, but it compares DeepRESP's performance against manual scoring and predicate devices. I've extracted the performance metrics that effectively serve as acceptance criteria given the "non-inferiority" and "superiority" claims against established devices.
| Metric (Against Manual Scoring) | DeepRESP Performance (95% CI) | Equivalent Predicate Performance (Nox Sleep System K192469) (95% CI) | Superiority/Non-inferiority Claim | Relevant Study Type |
|---|---|---|---|---|
| Severity Classification (AHI ≥ 5) | ||||
| PPA% | 87.5 [86.2, 89.0] | 73.6 [PPA% reported for predicate] | Superiority | Type I/II |
| NPA% | 91.9 [87.4, 95.8] | 65.8 [NPA% reported for predicate] | Non-inferiority | Type I/II |
| OPA% | 87.9 [86.6, 89.3] | 73.0 [OPA% reported for predicate] | Superiority | Type I/II |
| Severity Classification (AHI ≥ 15) | ||||
| PPA% | 74.1 [72.0, 76.5] | 54.5 [PPA% reported for predicate] | Superiority | Type I/II |
| NPA% | 94.7 [93.2, 96.2] | 89.8 [NPA% reported for predicate] | Non-inferiority | Type I/II |
| OPA% | 81.5 [79.9, 83.3] | 67.2 [OPA% reported for predicate] | Superiority | Type I/II |
| Respiratory Events | ||||
| PPA% | 72.0 [70.9, 73.2] | 58.5 [PPA% reported for predicate] | Non-inferiority (Superiority for OPA claimed) | Type I/II |
| NPA% | 94.2 [94.0, 94.5] | 95.4 [NPA% reported for predicate] | Non-inferiority | Type I/II |
| OPA% | 87.2 [86.8, 87.5] | 81.7 [OPA% reported for predicate] | Superiority | Type I/II |
| Sleep State Estimation (Wake) | ||||
| PPA% | 95.4 [95.1, 95.6] | 56.7 [PPA% reported for predicate] | Non-inferiority | Type I/II |
| NPA% | 94.6 [94.4, 94.9] | 98.1 [NPA% reported for predicate] | Non-inferiority | Type I/II |
| OPA% | 94.8 [94.6, 95.0] | 89.8 [OPA% reported for predicate] | Non-inferiority | Type I/II |
| Arousal Events | ||||
| ArI ICC (against Sleepware G3 K202142) | 0.63 [ArI ICC] | 0.794 [ArI ICC for additional predicate] | Non-inferiority | Type I/II |
| PPA% | 62.2 [61.2, 63.1] | N/A (Manual for primary predicate) | N/A | Type I/II |
| NPA% | 89.3 [88.8, 89.7] | N/A (Manual for primary predicate) | N/A | Type I/II |
| OPA% | 81.4 [81.1, 81.7] | N/A (Manual for primary predicate) | N/A | Type I/II |
| Type III Severity Classification (AHI ≥ 5) | ||||
| PPA% | 93.1 [92.2, 93.9] | 82.4 [PPA% reported for predicate] | Superiority | Type III |
| NPA% | 81.1 [75.1, 86.6] | 56.6 [NPA% reported for predicate] | Non-inferiority | Type III |
| OPA% | 92.5 [91.7, 93.3] | 81.1 [OPA% reported for predicate] | Non-inferiority | Type III |
| Type III Respiratory Events | ||||
| PPA% | 75.4 [74.6, 76.1] | 58.5 [PPA% reported for predicate] | Superiority | Type III |
| NPA% | 87.8 [87.4, 88.1] | 95.4 [NPA% reported for predicate] | Non-inferiority | Type III |
| OPA% | 83.7 [83.4, 84.0] | 81.7 [OPA% reported for predicate] | Superiority | Type III |
| Type III Arousal Events | ||||
| ArI ICC (against Sleepware G3 K202142) | 0.76 [ArI ICC] | 0.73 [ArI ICC for additional predicate] | Non-inferiority | Type III |
2. Sample Size Used for the Test Set and Data Provenance:
- Type I/II Studies (EEG present): 2,224 sleep recordings
- Type III Studies (No EEG): 3,488 sleep recordings (including 2,213 Type I recordings and 1,275 Type II recordings, processed to utilize only Type III relevant signals).
- Provenance: Retrospective study. Data originated from sleep clinics in the United States, collected as part of routine clinical work for patients suspected of sleep disorders. The patient population showed diversity in age, BMI, and race/ethnicity (Caucasian or White, Black or African American, Other, Not Reported) and was considered representative of patients seeking medical services for sleep disorders in the United States.
3. Number of Experts and Qualifications for Ground Truth:
The document explicitly states that the studies used "manually scored sleep recordings" but does not specify the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience"). It implicitly relies on the quality of "manual scoring" from routine clinical work in US sleep clinics as the ground truth.
4. Adjudication Method for the Test Set:
The document does not describe any specific adjudication method (e.g., 2+1, 3+1). It refers to "manual scoring" as the established ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a MRMC comparative effectiveness study was not reported. The study design was a retrospective data analysis comparing the algorithm's performance against existing manual scoring (ground truth) and established predicate devices. There is no information about human readers improving with AI vs. without AI assistance. The device is intended to provide automatic scoring subject to verification by a medical professional.
6. Standalone (Algorithm Only) Performance:
Yes, the study report describes the standalone performance of the DeepRESP algorithm. The reported PPA, NPA, OPA percentages, and ICC values represent the agreement of the automated scoring by DeepRESP compared to the manual ground truth. The device produces output "subject to verification by a medical professional," but the performance metrics provided are for the algorithmic output itself.
7. Type of Ground Truth Used:
The ground truth used was expert consensus (manual scoring). The document states "It used manually scored sleep recordings... The studies were done by evaluating the agreement in scoring and clinical indices resulting from the automatic scoring by DeepRESP compared to manual scoring."
8. Sample Size for the Training Set:
The document does not explicitly state the sample size used for the training set. The clinical validation study is described as a "retrospective study" used for validation, but details about the training data are not provided in this summary.
9. How the Ground Truth for the Training Set Was Established:
The document does not specify how the ground truth for the training set was established. It only describes the ground truth for the validation sets as "manually scored sleep recordings" from routine clinical work.
Ask a specific question about this device
(230 days)
Noxturnal Web is intended to be used for the diagnostic evaluation by a physician to assess sleep quality and as an aid for the diagnosis of sleep and respiratory-related sleep disorders in adults only.
Noxturnal Web is a software-only medical device to be used to analyze physiological signals and manually score sleep study results, including the staging of sleep, AHI, and detection of sleep disordered breathing events including obstructive apneas.
It is intended to be used under the supervision of a clinician in a clinical environment.
Noxturnal Web is a web-based software that can be utilized to screen various sleep and respiratoryrelated sleep disorders. The users of Noxturnal Web are medical professionals who have received training in the areas of hospital/clinical procedures, physiological monitoring of human subjects, or sleep disorder investigation. Users can input a sleep study recording stored on the cloud (electronic medical record repository) using their established credentials. Once the sleep study data has been retrieved, the Noxturnal Web software can be used to display, manually analyze, generate reports and print the prerecorded physiological signals.
Noxturnal Web is used to read sleep study data for the display, analysis, summarization, and retrieval of physiological parameters recorded during sleep and awake. Noxturnal Web facilitates a user to review or manually score a sleep study either before the initiation of treatment or during the treatment follow-up for various sleep and respiratory-related sleep disorders.
Noxturnal Web presents information from the input sleep study data in an organized layout. Multiple visualization layouts (e.g., Study Overview, Respiratory Signal Sheet, etc.) are available to allow the users to optimize the visualization of key data components. The reports generated by Noxturnal Web allow the inclusion of custom user comments, and these reports can then be viewed on the screen and/or printed.
The provided document is a 510(k) summary for the medical device Noxturnal Web. It states that clinical data were not relied upon for a determination of substantial equivalence. Therefore, there is no information in this document regarding a clinical study or a test set with expert-established ground truth.
However, the document does describe the performance expectations and how suitability was determined through non-clinical testing, specifically software verification and validation.
Here's the information based on the provided text, focusing on the non-clinical and comparative aspects:
1. A table of acceptance criteria and the reported device performance
The document does not present explicit quantitative acceptance criteria for performance in a table format with reported numerical device performance. Instead, it describes functional equivalence to the predicate device through comparative analysis and states that the software meets its pre-specified requirements and performs as intended.
The comparison table on pages 8-9 highlights the functional equivalences:
| Acceptance Criteria (Inferred from Functional Equivalence) | Reported Device Performance (as stated in document) |
|---|---|
| Aid/Assist in the diagnosis of sleep and respiratory-related sleep disorders | Yes (Same as predicates) |
| Arousal Scoring | Yes (Same as predicates) |
| Respiratory Events Scoring | Yes (Same as predicates) |
| Leg Movement Events Scoring | Yes (Same as predicates) |
| Sleep Study Scoring Method (Manual) | Manual (Same as primary predicate; additional predicate also has automatic) |
| Sleep Stage Scoring (W, N1/N2/N3, R) | Yes (Same as predicates) |
| Report Generation | Yes (Same as predicates) |
| Calculation of AASM standardized indices | Yes (Same as predicates) |
| Data Inputs (EEG, EOG, EMG, ECG, Chest/Abdomen movements, Airflow, Oxygen Saturation, Body Position/Activity) | All "Yes" (Same as predicates for all relevant inputs) |
| Software Type (Web-based) | Web-based (Same as additional predicate; primary predicate is computer program) |
| Physical Characteristics (Web-based operating in the cloud with Windows or Mac OS) | Web-based software operating in the cloud with Windows or Mac OS (Similar to additional predicate) |
| Standard of Scoring Manual | The American Academy of Sleep Medicine (AASM) Manual for the Scoring of Sleep and Associated Events (Same as predicates) |
| Backend implementation | Identical to corresponding qualitative and quantitative functionality implemented in the reference device (Nox Sleep System, K192469) |
| Cybersecurity controls | Implemented in accordance with FDA's Guidance "Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" |
The general acceptance criterion is that the Noxturnal Web is "as safe and effective as the predicate devices" and "meets its pre-specified requirements."
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical data were not relied upon for a determination of substantial equivalence." Therefore, there is no clinical test set of patient data with ground truth as would be used in a clinical study.
The testing performed was "Software verification and validation testing... to demonstrate safety and performance based on current industry standards," and "Verification and Validation testing of all requirement specifications defined for Noxturnal Web was conducted and passed." This implies that the 'test set' consisted of various software functions and their outputs, but not a large set of patient physiological recordings serving as a "test set" in the context of a clinical performance study. The data provenance and size of this kind of "test set" (software test cases) are not detailed in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Given that "Clinical data were not relied upon," there was no clinical test set requiring expert-established ground truth in the traditional sense for demonstrating substantial equivalence. The summary highlights that the device supports manual scoring completed by medical professionals who have received training in relevant areas (page 7). This implies that the human-in-the-loop performance is based on the expertise of the user, rather than the device itself establishing ground truth.
4. Adjudication method for the test set
Not applicable, as no clinical test set with established ground truth was used for assessing substantial equivalence.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done, as indicated by the statement "Clinical data were not relied upon for a determination of substantial equivalence." The device's primary function is to facilitate manual scoring by a clinician, not to provide AI-assisted automated interpretations that would then be compared to human-only interpretations via an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device is described as "software-only medical device to be used to analyze physiological signals and manually score sleep study results" and "It is intended to be used under the supervision of a clinician in a clinical environment." This indicates that the device is not intended for standalone (algorithm only) performance without human-in-the-loop interaction for interpretation and scoring. The comparative table also notes that both the subject device and the primary predicate "rely on manual scoring."
7. The type of ground truth used
For the purpose of regulatory clearance, the "ground truth" for the device's functionality was its ability to replicate the features and performance of legally marketed predicate devices, as demonstrated through "comparative analysis, software and performance testing." The ground truth for interpreting sleep studies using this device resides with the trained medical professional who manually scores the data according to the "American Academy of Sleep Medicine (AASM) Manual for the Scoring of Sleep and Associated Events."
8. The sample size for the training set
Not applicable, as this device appears to be a software tool for manual scoring and analysis, rather than an AI/ML algorithm that requires a "training set" in the context of deep learning or machine learning models. The summary makes no mention of AI/ML or training data; its emphasis is on providing tools for manual clinician review.
9. How the ground truth for the training set was established
Not applicable, for the same reasons as point 8.
Ask a specific question about this device
(65 days)
The Nox Sleep System is used as an aid in the diagnosis of different sleep disorders and for the assessment of sleep.
The Nox Sleep System is used to measure, record, display, organize, analyze, summarize, and retrieve physiological parameters during sleep and wake.
The Nox Sleep System allows the user to decide on the complexity of the study by varying the number and types of physiological signals measured.
The Nox Sleep System allows for generation of user/pre-defined reports based on subject's data.
The user of the Nox Sleep System are medical professionals who have received training in the areas of hospital/clinical procedures, physiological monitoring of human subjects, or sleep disorder investigation.
The intended environments are hospitals, institutions, sleep clinics, or other test environments, including patient's home.
The Nox Sleep System is intended for patients undergoing physiological measurements, for the assessment of sleep quality and the screening for sleep disorders.
The Nox Sleep System does not provide any alarms and is not intended to be used for continuous monitoring where failure to operate can cause injuries or death of the patient.
The basic Nox Sleep System consists of two recording/acquisition devices (Nox A1 Recorder and Nox C1 Access Point), a software running on a PC (Noxturnal PSG), an Android application (Noxturnal App) running on mobile platform, along with sensors and accessories. The system supports full Polysomnography (PSG) studies both in ambulatory and online/attended setups but also more simple sleep study setups, recording only few channels. The ambulatory sleep studies may take place in the clinic or in the home environment, but the online/attended sleep studies are only conducted in the clinical environment.
The Nox A 1 Recorder is a small battery-operated recording unit that is worn by the patient during the study. It records signals from patient applied sensors that connect to the unit but also supports recording of signals from auxiliary devices over Bluetooth. The Nox A1 Recorder allows for communication over Bluetooth with the Noxturnal App during ambulatory setup and with the Nox C1 Access Point during online setup. The recorder is intended to be worn over clothing.
New accessories and sensors as part of this submission are the Nox A 1 EEG 5 Lead Gold Electrode Cable and Nox A1 EEG Head Cable that are used for recording of EEG/EOG. These components are in direct contact with the patient.
The Nox C1 Access Point is a separate mains powered unit located remotely from the patient that allows for recording of signals from auxiliary devices. It supports communication over LAN/Ethernet to the Noxturnal PSG, and communication with the Nox A1 Recorder and Noxturnal App over Bluetooth. The Nox C1 Access Point is only used for online study setup and is thus not intended to be used in the home environment.
The Noxturnal App is used as a mobile interface to the Nox A1 Recorder and Nox C1 Access Point. The communications are via Bluetooth link. The app is normally used in the beginning of a sleep study, for basic tasks such as device configuration, starting a recording, checking the signal quality of signals being recorded and marking events during bio calibration.
The Noxturnal PSG is used for configuration of the Nox recording/acquisition devices, to download a study from ambulatory recording or to collect an online study. The software supports the viewing, retrieving, storing and processing of data recorded/collected, manual and automatic analysis and reporting on the results from the recorded studies. The purpose with the automatic scoring function in Noxturnal PSG is to assist the trained physician in the diagnosis of a patient. It is not intended to provide the trained physician with a diagnostic results. The type of automatic analysis events scored by Noxturnal PSG include: Sleep Stages (Wake, N1, N2, N3, REM), Apneas, Hypopneas, Apnea Cassification (Obstructive, Mixed and Central Apneas), Limb Movements, Periodic Limb Movements, SpO2 Desaturation Events. and potential Bruxism-Related Events.
The result of the automatic analysis/scoring must always be manually verified by the trained physician prior to diagnosis.
The Nox Sleep System is designed to aid in the diagnosis of sleep disorders and assess sleep quality by measuring, recording, displaying, organizing, analyzing, summarizing, and retrieving physiological parameters during sleep and wake. The system includes automatic scoring functionalities for various sleep events, which are intended to assist trained medical professionals in diagnosis.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The document presents separate sections for the performance of different automatic scoring algorithms rather than a single consolidated table. However, the information can be extracted and presented as follows:
| Automatic Scoring Algorithm | Acceptance Criteria (Safety Endpoint/Justification) | Reported Device Performance |
|---|---|---|
| Bruxism Analysis | Detect at least 90% of oromandibular movements considered by a human expert to be bruxism-related events with 95% confidence (Sensitivity). | Sensitivity: 95.7% (95% CI 93.2% - 97.4%)Specificity: 61.0% (95% CI 58.9% - 63.0%)PPV: 34.6% (95% CI 32.0% - 37.3%)NPV: 98.5% (95% CI 97.7% - 99.1%) |
| PLM Analysis | Interclass correlation (ICC) of 0.61 or greater and a bias unlikely to impact a diagnosis for the Periodic Limb Movement Index. | ICC for Periodic Limb Movement Index: 0.87 |
| Respiratory Flow Analysis (AHI) | Not classifying patients with an AHI below 5 as having an AHI greater than or equal to 15 (95% confidence), AND Not classifying patients with an AHI greater than or equal to 15 as having an AHI below 5 (95% confidence). Also, Cohen's kappa reported. | Cohen's Kappa for AHI (Cannula): 0.78Cohen's Kappa for AHI (RIP flow): 0.62 (95% CI 0.59-0.66)Cohen's Kappa for AHI (cRIP flow): 0.62 (95% CI 0.59-0.66) |
| Respiratory Flow Analysis (ODI) | (Implicitly similar to AHI, with Cohen's kappa reported) | Cohen's Kappa for ODI: 0.87 |
| Apnea Classification | ICC comparable to what has been reported in scientific literature for Central Apnea Index (0.46). | ICC for Central Apnea Index: 0.91Cohen's Kappa for Central Apnea Index: 0.89 |
| Sleep Staging Analysis | Average accuracy of at least 60% when scoring wake epochs to ensure total sleep time measurement with 10% error or less (assuming 80% sleep efficiency). | Cohen's Kappa: <=0.67Accuracy for Wake: 66.7%Accuracy for N1: 9.8%Accuracy for N2: 87.0%Accuracy for N3: 83.0%Accuracy for REM: 82.5% |
2. Sample size used for the test set and the data provenance
The document states that the clinical performance testing for the automatic scoring algorithms involved "retrospectively analyzing pre-existing clinical data from sleep studies that had already been collected and manually scored as part of routine clinical care."
- Bruxism Analysis: "adult Polygraphy recordings, including masseter EMG signal, recorded from multiple U.S. dental clinics."
- PLM Analysis: "adult Polysomnography recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
- Respiratory Flow Analysis (AHI & ODI): "adult Polysomnography recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
- Apnea Classification: "adult Polygraphy recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
- Sleep Staging Analysis: "adult recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
The exact sample size (number of patients/studies) for each test set is not explicitly provided in the given text. The provenance is retrospective and data was collected from U.S. dental clinics (for bruxism) and a national hospital (for other sleep studies).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The ground truth for the test set was established by "qualified polysomnographic technologists" who "followed the American Academy of Sleep Medicine (AASM) scoring guidelines."
The number of experts is not specified. Their qualification is: "qualified polysomnographic technologists." No details on years of experience are provided, but adherence to AASM guidelines suggests a standardized and recognized level of expertise.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document states, "All scorers were qualified polysomnographic technologists and followed the American Academy of Sleep Medicine (AASM) scoring guidelines." It then describes comparison of the automatic scoring results to "the results of the manually scored data."
This implies that the manual scoring by the polysomnographic technologists served as the ground truth. There is no explicit mention of an adjudication method (like 2+1 or 3+1) among multiple human scorers for establishing the ground truth. It seems to rely on single-expert scoring as the reference standard.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was described where human readers' performance with and without AI assistance was evaluated. The study focused on the standalone performance of the automatic scoring algorithms against human-scored ground truth. The device is intended as an aid for professionals, requiring manual verification.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, standalone performance was done. The clinical performance testing directly compared the output of the automatic scoring algorithms to the manually scored ground truth data. The results (sensitivity, specificity, PPV, NPV, ICC, Cohen's Kappa, and accuracy) reported are measures of the algorithm's performance on its own.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used was expert consensus (manual scoring by qualified polysomnographic technologists) following the American Academy of Sleep Medicine (AASM) scoring guidelines. This is a form of expert consensus, even if the "consensus" here refers to the adherence to established scoring rules rather than explicit multi-reader agreement on individual cases.
8. The sample size for the training set
The document details the "clinical performance testing," which involved "retrospectively analyzing pre-existing clinical data." This section describes the test set used for validation.
The sample size for the training set is not provided or mentioned in this document. Information about the training data used to develop the automatic scoring algorithms is absent.
9. How the ground truth for the training set was established
Since the document does not discuss the training set, there is no information provided on how its ground truth was established.
Ask a specific question about this device
(169 days)
The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems.
The Nox RIP Belts are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient 's home.
The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals.
The RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient´s home.
The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH.
The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient's home.
The Nox RIP Belts are respiratory effort sensors that are intended to function as an accessory with sleep/polysomnography (PSG) systems. The RIP Belts measure respiratory effort signals based on Respiratory Inductance Plethysmography (RIP) technology, which is the gold standard technology for respiratory effort belts.
Two RIP belts are used to measure the respiratory effort of the patient. One belt is placed around the patient's abdomen and the other around the patient's thorax. Both abdomen and thorax belts are identical.
The Nox RIP Belt Cables are used to connect between the respiratory effort sensor (RIP belts) and the applicable sleep recorder/polysomnography (PSG) system.
There are two product groups for the Nox RIP Belt Cables; RIP Belt Cables and Third Party RIP Belt Cables.
The RIP Belt Cables are designed for use with Nox recorders only. Those are abdomen cables only because the thorax belt is attached directly to the Nox recorders.
The Third Party RIP Belt Cables are designed for use with third party recorders. The Third Party RIP Belt Cables come in pairs for abdomen and thorax.
The provided document is a 510(k) premarket notification for Nox RIP Belts and Nox RIP Belt Cables. The core of this submission is to demonstrate substantial equivalence to a predicate device, not necessarily to set new performance acceptance criteria through a clinical study in the same way an AI-driven diagnostic might.
Therefore, the acceptance criteria and study described here are focused on demonstrating that the new devices perform equivalently to the predicate, and comply with relevant safety standards.
Here's an attempt to extract the information you requested, based on the provided text, while noting the differences in context for a traditional medical device accessory vs. an AI diagnostic:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission for device accessories (RIP belts and cables) demonstrating substantial equivalence to a predicate, the "acceptance criteria" are primarily related to meeting performance specifications and safety standards, and showing clinical equivalence to the predicate's signal quality. There aren't specific metrics like sensitivity/specificity for disease detection.
| Criteria Type | Acceptance Criteria (Met by) | Reported Device Performance (Demonstrated by) |
|---|---|---|
| Safety & Standards | Compliance with relevant standards. | Demonstrated compliance with: - ISO 14971 (Risk Management)- ISO 15223-1 (Symbols)- AAMI/ANSI/ES 60601-1 (Basic Safety & Essential Performance)- IEC 60601-1-2 (EMC)- AAMI/ANSI/IEC 62366 (Usability Engineering)- 21 CFR 898 (for Third Party RIP Belt Cables) |
| Functional | Conformance to design input/specifications. | Verification testing: Design output conforms to design input, fulfilling all physical characteristics, performance, functional, interface, packaging, labeling, safety, and reliability requirements. |
| Usability | Minimize use errors and risks. | Usability testing resulted in all usability goals passed. |
| Signal Quality | Clinically equivalent signal to predicate. | Signal integrity tests (signal-to-noise ratio, signal range, bandwidth, linearity) for new devices compared to predicate (QDC-PRO AND NOX-RIP) demonstrated clinical equivalence. |
| Risk Management | Risks appropriately managed. | Risk analysis performed according to ISO 14971; appropriate measures implemented and their effectiveness verified/validated. |
| Material/Physical Equivalence | Materials do not raise new safety/effectiveness concerns. | Verification testing and risk analysis show minor differences in material do not raise new questions about safety and effectiveness (for RIP Belt Cables). |
| Connector Equivalence | Different connectors do not raise new safety/effectiveness concerns. | Verification testing, signal integrity comparison, and risk analysis show different connectors do not raise new questions about safety and effectiveness (for RIP Belt Cables). |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify a "test set" in the context of patient data or clinical images for an algorithm. The testing described is primarily bench testing, engineering verification, and validation against product requirements and standards.
- Sample Size for Test Set: Not applicable in the context of patient-specific data for algorithm performance. The "samples" would be the manufactured devices (RIP Belts and Cables) themselves. The number of devices tested is not specified, but it implies a statistically sound sample for verification and validation activities.
- Data Provenance: Not applicable for patient data. The provenance of the testing results is "thorough internal testing" conducted by Nox Medical ehf.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. As a medical device accessory focused on signal measurement and equivalence to a predicate, the "ground truth" is established by direct physical/electrical measurements against known standards and the predicate device's performance, rather than expert interpretation of a clinical finding.
4. Adjudication Method for the Test Set
Not applicable. There is no expert adjudication mentioned, as the nature of the device (respiratory effort sensors) involves direct measurement and comparison, not subjective interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a sensor and cable accessory, not an AI diagnostic tool that assists human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
Not applicable. This device does not involve an algorithm with standalone performance. It measures respiratory effort signals, which are then used by sleep/polysomnography (PSG) systems, presumably to be interpreted by healthcare professionals.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the performance evaluation of these accessories primarily consists of:
- Engineering Specifications and Standard Compliance: Adherence to established ISO and IEC standards for medical devices (e.g., electrical safety, EMC, risk management, usability).
- Predicate Device Performance: The QDC-PRO AND NOX-RIP (K124062) device serves as the benchmark for "clinical equivalence" of signal quality. New devices' signals were compared against the predicate's signals using metrics like signal-to-noise ratio, signal range, bandwidth, and linearity.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this device does not involve machine learning or AI.
9. How the Ground Truth for the Training Set was Established
Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(191 days)
The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder.
The QDC-PRO device is indicated for use in patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments.
The ODC-PRO device is intended to function as an accessory for 3rd party PSG systems, delivering the respiratory signals needed in PSG studies along with position/activity signals.
Polysomnography (PSG) is a multi-parametric sleep study indicated for diagnosis of various sleep disorders. The ODC-PRO is therefore a part of a full PSG, the gold standard sleep study, and its main objectives is to provide signals used for diagnoses of sleep disordered breathing (SDB).
The signals measured by the ODC-PRO are provided as analog signals to a general 3rd party PSG amplifier that has DC inputs with characteristics matching the QDC-PRO device signal output specifications.
The output signals include:
- Abdomen respiratory effort (RIP) .
- . Thorax respiratory effort (RIP)
- . SUM of abdomen and thorax respiratory effort (RIP)
- Nasal pressure from nasal cannula .
- Snore signal from nasal cannula .
- . Body position
- . Activity
- Audio .
The respiratory effort is measured by the use of respiratory inductive plethysmography (RIP).
The QDC-PRO device provides calibration for the RIP signals by the use of Quantitative Diagnostic Calibration (QDC) technique. The calibrated RIP signals (Sum) represents the tidal volume of the respiration better than un-calibrated RIP signals.
The QDC-PRO contains a sensor unit, respiratory effort sensors (RIP belts) and cables. The QDC-PRO is worn by the patient. It measures signals from two respiratory effort sensors, audio via an inbuilt microphone, nasal pressure and snoring via a nasal cannula and patient 's position/activity data. The signals are processed within the device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder.
The QDC-PRO is powered with one AA (1.5V) battery and has a display for status indication, signal integrity, and buttons for control.
The provided 510(k) summary for the QDC-PRO device does not contain a specific table of acceptance criteria with reported device performance metrics in the format usually associated with clinical performance studies. Instead, it focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and a general summary of performance testing.
However, based on the provided text, we can infer the acceptance criteria for signal quality and the overall conclusion regarding device performance.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic / Acceptance Criteria | Reported Device Performance |
|---|---|
| Signal Quality (General) | "Signal comparison tests conducted for the QDC-PRO device confirm the quality of the signals being measured by the device. The results demonstrate that the QDC-PRO signals may be regarded as clinically equivalent to that of the predicates Respitrace QDC and NOX T3." |
| Quantitative Diagnostic Calibration (QDC) Function Performance | "Quantitative Diagnostic Calibration (QDC) function comparison tests conclude that the QDC-PRO performance may be regarded as clinically equivalent to that of the predicate Respitrace QDC." |
| Safety and Effectiveness | "Thorough internal and external testing has demonstrated that the QDC-PRO device is effective and safe for its intended use." "Based on the testing, risk analysis, verification and validation activates described above and Besed on the essang, how analysis) - chivices provided in Table 1 above, it is the conclusion of Nox Medical that the QDC-PRO device is substantially equivalent to the predicates NOX T3 from Nox Medical (K082113) and Respitrace QDC from SensorMedics Corporation (K903011) and presents no new concerns about safety and effectiveness." (FDA's substantial equivalence finding supports this.) |
| Compliance with Standards | "The ODC-PRO device complies with the applicable EMC and patient safety standards: IEC60601-1:2007 (Basic safety and essential performance), AAMI/ ANSI ES60601-1:2005 (Basic safety and essential performance), IEC62304 (Medical Device Software), ISO14971 (Risk Management), IEC 62366 (Usability Engineering)." |
| Usability | "Usability testing compliant with the standard: IEC 62366 - Medical devices - Application of usability engineering to medical devices, demonstrates that the QDC-PRO device is simple and safe to operate and minimizes the likelihood of errors and lapses." |
2. Sample Size Used for the Test Set and Data Provenance
The summary states that "Signal comparison tests" and "Quantitative Diagnostic Calibration (QDC) function comparison tests" were conducted, as well as "Clinical evaluation". However, it does not specify the sample size used for these test sets (e.g., number of patients, number of studies).
The data provenance is not explicitly stated as retrospective or prospective, nor are the specific countries of origin for the data provided. The overall context suggests internal testing and evaluation conducted by Nox Medical, with clinical evaluation comparing to publicly available data from predicate devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications used to establish ground truth for any test sets. The "clinical evaluation" mentions that signals were "comparable to the already validated and publicly available Respitrace QDC and Nox T3 device, respectively," implying a comparison to established performance of predicate devices rather than independent expert ground truth for novel data.
4. Adjudication Method for the Test Set
The document does not mention any adjudication method (e.g., 2+1, 3+1, none) for a test set.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done with human readers. The clinical evaluation focused on comparing the device's signals to those of predicate devices, not on assessing human reader performance with or without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies described are analogous to standalone performance studies for the device's signal acquisition and processing capabilities. The signal comparison and QDC function comparison tests assess the direct output of the QDC-PRO device against established benchmarks (predicate devices), without involving human interpretation as part of the core performance measurement against a ground truth. The device is a "sensor unit" providing output signals for a 3rd party PSG system, so its primary performance is in accurate signal generation.
7. The Type of Ground Truth Used
The ground truth used for performance validation was primarily:
- Established performance of predicate devices: The QDC-PRO's signals and QDC function were deemed "clinically equivalent" to those of the Respitace QDC and NOX T3. This implies that the performance of these legally marketed predicate devices serves as the benchmark or "ground truth" for comparison.
- Compliance with recognized standards: The device's safety, software development, risk management, and usability were validated against relevant IEC, AAMI/ANSI, and ISO standards.
8. The Sample Size for the Training Set
The document does not provide any information regarding a training set sample size. Given that this is a sensor unit that produces physiological signals, it is unlikely to involve "training" in the machine learning sense that would require a distinct training set (beyond perhaps internal calibration data or development data which are not specified). The focus is on the accuracy of the physical measurements and signal processing.
9. How the Ground Truth for the Training Set was Established
As no training set is mentioned or implied in the context of machine learning, there is no information on how ground truth for a training set was established.
Ask a specific question about this device
(102 days)
The Nox T3 device is intended for ambulatory recording of physiological signals during sleep. The recorded signals are then downloaded to a PC where the signals can be viewed and analyzed by use of the Nox T3 application (Noxturnal). The Nox T3 system is indicated for use in patients greater than 2 years of age.
The Nox T3 system is NOT intended for any patient monitoring or automatic diagnosis.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including patient's home.
The Nox T3 system is used for patients suspected of suffering from Sleep Disordered Breathing (SDB) or Periodic Limb Movement Disorder (PLMD).
The Nox T3 is an ambulatory recording system. It includes a recording device, respiratory effort sensors, clip straps, filter tube connector and an USB cable for data download and the Nox T3 application (Noxturnal).
The Nox T3 device is a pocket size battery powered digital recorder that incorporates electronics to record and store up to three nights of physiological parameters. The Nox T3 device is worn on the patient's chest by snapping it to the thoracic respiratory effort sensor belt and securing its position with the clip straps. It has a display for status indication, signal integrity and preliminary results, and buttons for control.
The Nox T3 device records signals from five external sensors and three built-in sensors. The external sensors that can be used with the device are abdominal and thoracic respiratory effort sensors, oximeter (via wireless transmission), and two leads of the following: ECG, EMG, EEG or EOG. The built-in sensors include a pressure transducer allowing either recording of nasal pressure (via nasal cannula) or mask pressure and measuring of snoring, a three dimensional acceleration sensor for measure of patient's position and activity, and a microphone for true audio recording capabilities.
The Nox T3 device includes a class II Bluetooth transmitter/receiver to allow for wireless transmission of data from Nonin´s Model 4100 Patient Oximeter Module.
The Nox T3 application (Noxturnal) is used to configure the device for recording, and downloading, viewing and analyzing of recorded data on a PC.
Here's an analysis of the acceptance criteria and study detailed in the provided document for the Nox T3 device:
The document (510(k) summary) describes the Nox T3 as an ambulatory recording system for physiological signals during sleep. The primary goal of the "performance testing summary" is to demonstrate substantial equivalence to predicate devices, rather than establishing specific performance metrics against a predefined acceptance criterion. The key performance claims for this device focus on its ability to record reliable signals and the accuracy of its automated scoring software (Noxturnal) compared to manual scoring.
Here's the breakdown as requested:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Signal Reliability & Usability | "The signals recorded with the Nox T3 system were compared to signals recorded with the predicate device Embla N7000. The result demonstrates the reliability and usability of all signals recorded with the Nox T3 system." |
| Automatic Scoring Accuracy (AHI) | "The analysis comparison result demonstrates that the Noxturnal application scores AHI...events in a substantially equivalent manner to the manual scoring on full PSG recordings obtained using Embla N7000 recorders." (No specific numerical target or statistical metric [e.g., agreement percentage, correlation coefficient] is provided in the summary.) |
| Automatic Scoring Accuracy (ODI) | "The analysis comparison result demonstrates that the Noxturnal application scores ...ODI values in a substantially equivalent manner to the manual scoring on full PSG recordings obtained using Embla N7000 recorders." (No specific numerical target or statistical metric is provided in the summary.) |
| Ease of Operation | "The general process from configuration of device to reading out the results and generating report was validated by having untrained person perform this actions. The results demonstrates that the Nox T3 system has meet its objective of being easy to operate, the Noxturnal interface guides the user appropriately, minimizing the likelihood of errors and lapses, and the design of the Nox T3 components and user instruction allows the hook-up to be performed by untrained people." |
| Compliance to Standards | The device underwent external testing to comply with applicable standards regarding EMC and patient safety (e.g., IEC60601-1, IEC60601-1-2, IEC 60601-2-25, IEC 60601-2-26, IEC 60601-2-40) and FCC requirements for R&TTE approval. (Compliance is stated; no specific results of these tests and how they meet criteria are detailed in this summary.) |
Note: The document explicitly states that the comparisons "do not raise new questions of safety and effectiveness and demonstrate the same effectiveness and safety as that of the predicates." This implies that the 'acceptance criteria' are primarily based on demonstrating equivalence rather than achieving absolute performance thresholds.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Automatic Scoring Comparison: 1057 Embla N7000 recordings.
- Data Provenance: The document does not explicitly state the country of origin. It indicates the data consisted of "Embla N7000 recordings." Embla is a brand associated with sleep diagnostics, and both Medcare Flaga (the predicate's manufacturer) and Nox Medical are based in European countries (Medcare Flaga in the past was associated with Iceland, though it might have had broader presence; Nox Medical is in Iceland). Given the context of a 510(k) submission to the FDA, it's highly probable that the data was collected from clinical settings, likely retrospectively as they were "recordings which had all gone through the process of being manually scored."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- Number of Experts: Not explicitly stated as a specific number. The ground truth was established by "Sleep Technicians with RPSGT certification" and "then reviewed by a Physician." This suggests at least one technician and one physician per recording, but doesn't give an aggregate count of unique individuals.
- Qualifications of Experts:
- Sleep Technicians: "RPSGT certification" (Registered Polysomnographic Technologist).
- Physician: "reviewed by a Physician." (Specific specialization like a sleep physician is implied but not explicitly stated.)
4. Adjudication Method for the Test Set
- The method described is sequential: "manually scored by Sleep Technicians with RPSGT certification, and then reviewed by a Physician." This implies a form of hierarchical review, where the technician's scoring is the initial pass, followed by a physician's oversight. It is not explicitly stated if there was a consensus process (e.g., 2+1, 3+1) if the physician disagreed with the technician, or if the physician's review constituted the final ground truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
- No, a MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not performed or reported in this summary. The study reported compared the device's standalone automatic scoring (Noxturnal application) against manual scoring by human experts. The study's focus was on the performance of the automated algorithm, not on improving human reader performance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
- Yes, a standalone performance evaluation of the algorithm was done. The "Analysis comparison was performed to validate the quality of the automatic scoring performed by the Noxturnal application compared to manual scoring of full PSG data." This directly assesses the algorithm's performance without a human in the loop during the scoring process.
7. The Type of Ground Truth Used
- The ground truth used was expert consensus/manual scoring. Specifically, "manual scoring of full PSG data" performed by "Sleep Technicians with RPSGT certification" and "then reviewed by a Physician."
8. The Sample Size for the Training Set
- The document does not specify a sample size for the training set. It only mentions the "recordings used consisted of 1057 Embla N7000 recordings" for the validation study. It is common for 510(k) summaries to omit details about training data, but its absence here means we cannot ascertain the training set size from this document.
9. How the Ground Truth for the Training Set was Established
- The document does not provide information on how the ground truth for the training set was established. Given the lack of detail on the training set itself, this information is not available in the provided text.
Ask a specific question about this device
Page 1 of 1