510(k) Data Aggregation

    K Number
    K201010

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-14

    (119 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infinix™ Universal Composite:

    • Restorations in the posterior region (Class I and II)
    • Anterior restorations (Class III, IV)
    • Class V restorations (cervical caries, root erosion, wedge-shaped defects)
    • Veneering of discolored anterior teeth
    • Splinting of mobile teeth
    • Extended fissure sealing in molars and premolars
    • Repair of composite/ceramic veneers
      The addition of the QASi particles to the Infinix™ Universal Composite reduces demineralization, which is part of the caries-formation process.

    Infinix™ Flowable Composite:

    • Class III and V restorations
    • Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations)
    • As base/liner under direct or indirect restorations
    • Repair of small defects in aesthetic indirect restorations
    • Pit and fissure sealant
    • Blocking out of undercuts
    • Repair of resin and acrylic temporary materials
      The addition of the QASi particles to the Infinix™ Flowable Composite reduces demineralization, which is part of the caries-formation process.

    Infinix™ Bulk Fill Flow Composite:

    • Base under Class I and II direct restorations
    • Liner under direct restorative materials
    • Pit and fissure sealant
    • Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations)
    • Class III and V restorations
    • Blocking out of undercuts
    • Repair of small enamel defects
    • Repair of small defects in esthetic indirect restorations
    • Repair of resin and acrylic temporary materials
    • As a core build-up where at least half the coronal tooth structure is remaining to provide structural support for the crown
      The addition of the OASi particles to the Infinix™ Bulk Fill Flow Composite reduces demineralization, which is part of the caries-formation process.
    Device Description

    The Infinix™ Family of Dental Composites include the following light-cured, resinbased composite materials:

    • Universal Composite
    • Flowable Composite
    • Bulk Fill Flow Composite
      Some of the silica-based filler particles (max 1.2% wt/wt of the total formulation) contained in all the Infinix™ composites are decorated with quaternary ammonium functional groups that are covalently bound to the filler particles' silica core via a silane linker (herein referred to as "QASi"). Following light induced polymerization, QASi particles (produced by Nobio) remain permanently entrapped within the composite restoration and destroy microorganisms by contact. As the cell membrane of microorganisms is negatively charged and the QASi particles have strong positive charges, microorganisms are electrostatically attracted to the positively charged cured material's surface resulting in disruption of their cell membrane and immediate microbial lysis.
    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding dental composites. It does not contain information about an AI/ML medical device. Therefore, I cannot provide details on acceptance criteria and study proving device meets criteria as requested in the prompt, as the provided text doesn't describe such a device or study.

    The document discusses the substantial equivalence of the "Infinix™ Family of Dental Composites" to their predicate devices, emphasizing a new claim related to the reduction of demineralization due to the addition of QASi particles. The clinical performance section describes an in-situ clinical study comparing a composite with QASi particles to one without, focusing on mineral loss in enamel slabs. This study is for dental materials, not an AI/ML algorithm.

    Therefore, I cannot fulfill the request as it pertains to an AI/ML device and its acceptance criteria, ground truth establishment, sample sizes for training/test sets, expert adjudication, or MRMC studies, because this information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183391

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-24

    (291 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Novidia Universal Bond is indicated for:

      1. Direct restorations using light-cured composite resin or compomer
      1. Cavity sealing as a pretreatment for indirect restorations
      1. Treatment of hypersensitive teeth and/or exposed root surfaces
      1. Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin
      1. Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin
      1. Core build-ups using light- or dual-cured composite resin
      1. Cavity sealing under amalgam restorations
    • For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
    Device Description

    Novidia™ Universal Bond is a resin-based bond system consisting of a self-etch primer and a light-cured bond. It is designed to enhance penetration into prepared enamel and dentin surfaces and to establish strong adhesion between mineralized dental surfaces and the restorative materials such as resin composites, resin cements, metals, and ceramics (e.g., porcelain, lithium disilicate, zirconia, or hybrid ceramics) restorations. The chemical composition of the Universal Bond's primer consists of adhesion promoter and solvent, and the Universal Bond consists of resin, solvent, initiators and antibacterial insoluble particles, which contain covalently bound quaternary ammonium (QA).

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Novidia™ Universal Bond, a resin tooth bonding agent. It outlines the device's indications for use, its technical characteristics, and a comparison to predicate and reference devices. The core of the submission relies on bench testing to demonstrate substantial equivalence, rather than extensive clinical or human-in-the-loop studies commonly seen with AI/software as a medical device (SaMD).

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text, while noting the limitations in addressing all aspects of your request (particularly those related to AI/SaMD studies):

    Acceptance Criteria and Reported Device Performance

    Table 1: Acceptance Criteria and Novidia™ Universal Bond Performance

    TestAcceptance CriteriaNovidia™ Universal Bond Performance
    Polymerization Conversion Degree$\ge 70%$Passed
    Viscosity$550 \pm 450 \text{ cP}$Passed
    Shear Bond Strength (Zirconia)$\ge 10 \text{ MPa}$Passed
    Shear Bond Strength (Glass Ceramics)(Implicitly $\ge 10 \text{ MPa}$ based on the Zirconia criterion being the general requirement for most materials)Passed
    Shear Bond Strength (Enamel)(Implicitly $\ge 10 \text{ MPa}$ based on the Zirconia criterion being the general requirement for most materials)Passed
    Shear Bond Strength (Dentin)(Implicitly $\ge 10 \text{ MPa}$ based on the Zirconia criterion being the general requirement for most materials)Passed
    Shear Bond Strength (Titanium)(Implicitly $\ge 10 \text{ MPa}$ based on the Zirconia criterion being the general requirement for most materials)Passed
    Shear Bond Strength (Composite)(Implicitly $\ge 10 \text{ MPa}$ based on the Zirconia criterion being the general requirement for most materials)Passed
    Shear Bond Strength (Amalgam)$\ge 7 \text{ MPa}$Passed
    Micro LeakagePenetration levels: 0 (no penetration) to 1 (only shallow enamel surface penetration).Passed
    Antibacterial ActivityReduction of the viable bacteriaPassed

    Note: The acceptance criteria for Shear Bond Strength for Glass Ceramics, Enamel, Dentin, Titanium, and Composite are not explicitly stated, but are implied to be $\ge 10 \text{ MPa}$ given the general structure of the table and the explicit criterion for Zirconia.

    Study Details

    The provided document describes a bench testing study to demonstrate substantial equivalence, rather than a clinical trial or AI model validation study.

    1. Sample Size and Data Provenance for Test Set:

      • The document does not specify the exact sample sizes used for each bench test (e.g., number of samples for polymerization conversion, viscosity, or shear bond strength). It only states that "Non-clinical and biological testing was completed."
      • The data provenance is not explicitly mentioned in terms of country of origin or whether it was retrospective/prospective. As these are physical/chemical tests, data provenance in the context of geographical clinical data or time-based data collection methods (retrospective/prospective) is typically not applicable.
    2. Number of Experts and Qualifications for Ground Truth:

      • This is not applicable to this type of study. The ground truth for the performance of a dental bonding agent, in this context, is established through standardized laboratory procedures and measurements (e.g., ISO standards for bond strength, chemical analysis for polymerization conversion). Expert consensus, common in diagnostic AI, is not relevant here.
    3. Adjudication Method for Test Set:

      • Not applicable. Bench tests are objective measurements, not subjective evaluations requiring adjudication.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, a MRMC study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, typically in diagnostic imaging. The Novidia™ Universal Bond is a physical dental material, not an AI/SaMD.
    5. Standalone Performance (Algorithm Only without Human-in-the-Loop):

      • Not applicable in the AI/SaMD sense. The "performance" demonstrated is the physical and chemical properties of the material itself (e.g., bond strength, antibacterial activity) as measured in a lab, which is inherently "standalone" from a human interpretative loop.
    6. Type of Ground Truth Used:

      • The ground truth was established through objective laboratory measurements based on established scientific principles and potentially ISO standards for material testing (e.g., ISO 10993 for biocompatibility, ISO 7405 for dental materials). It's not expert consensus, pathology, or outcomes data as would be the case for a diagnostic AI.
    7. Sample Size for Training Set:

      • There is no mention of a training set because this is not an AI/machine learning device. The "training" for such a device would be the R&D and formulation process to develop the chemical composition and manufacturing procedures that yield the desired performance characteristics.
    8. How the Ground Truth for the Training Set Was Established:

      • As there is no "training set" in the AI sense, this question is not applicable. The development of the ground truth (i.e., the target performance characteristics) for the material would be based on dental material science, clinical needs, and performance of existing predicate devices.

    In summary, the provided document details the regulatory pathway for a dental material (Novidia™ Universal Bond) based on substantial equivalence demonstrated through rigorous non-clinical, bench testing, and biocompatibility assessments, adhering to established industry standards for dental materials. It largely differs from the typical safety and effectiveness studies required for AI/Software as a Medical Device (SaMD) which often involve clinical data, human reader performance studies, and complex expert ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182921

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-13

    (329 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InfinixTM Universal Composite is indicated for:

    • · Restorations in the posterior region (Class I and II)
    • · Anterior restorations (Class III, IV)
    • · Class V restorations (cervical caries, root erosion, wedge-shaped defects)
    • · Veneering of discolored anterior teeth
    • · Splinting of mobile teeth
    • · Extended fissure sealing in molars and premolars
    • · Repair of composite/ceramic veneers
    Device Description

    Infinix™ Universal Composite is a packable (high viscosity) dental composite filling material, intended for direct anterior and posterior restorations, class V restorations, veneering of discolored anterior teeth, and repair of composite/ceramic veneers. The resin matrix of the Infinix™ Universal Composite contains Urethane dimethacrylate (UDMA), Bisphenol-a glycidyl dimethacrylate (Bis-GMA) and triethylene glycol dimethacrylate (TEGDMA). The inorganic filler of the Infinix™ Universal Composite is a mix of particles of alumino-silicate-based glasses, silica dioxide and pigments. A small percentage of the silica-based filler particles are covalently bound to quaternary ammonium residues (OASi), added to maintain the integrity of the restoration. The Infinix™ Universal Composite is provided in syringes (4 g) and in single-dose capsules (0.2 g) and is available in five shades (A1, A2, A3, B1 and translucent).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a dental resin material named "Infinix™ Universal Composite". This submission focuses on demonstrating substantial equivalence to a predicate device based on material properties and non-clinical testing. It does not describe a study involving algorithms, human readers, or image analysis for diagnostic purposes.

    Therefore, most of the requested information regarding acceptance criteria, device performance in a diagnostic context, sample sizes for test/training sets, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance cannot be extracted from this document as it pertains to a different type of medical device and evaluation.

    However, I can extract the non-clinical performance criteria and reported performance from the document.

    1. A table of acceptance criteria and the reported device performance

    The document lists various non-clinical tests performed on the Infinix™ Universal Composite to establish substantial equivalence to the predicate device, Tetric EvoCeram®, and compliance with ISO 4049 and FDA guidance. The acceptance criteria are implicitly those defined by these standards and guidance documents. The document states that the device "performs as intended based on the bench testing per ISO 4049 and FDA guidance document". Specific numerical results for "reported device performance" are not provided in this summary, but the general conclusion is that the device met these criteria.

    Acceptance Criteria (Implied by Standard/Guidance)Infinix™ Universal Composite Performance (Reported)
    Biocompatibility (ISO 10993, ISO 7405)Biocompatible (based on assessment of Cytotoxicity, Oral Mucosal Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Bacterial Reverse Mutation, Mouse Lymphoma Assay, Biological Risk Assessment)
    Flexural Strength (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Elastic ModulusPerformed as intended
    Compression StrengthPerformed as intended
    Shade and Color StabilityPerformed as intended
    Polymerization Conversion DegreePerformed as intended
    ViscosityPerformed as intended
    Spontaneous Polymerization Sensitivity at Ambient Light (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Water Solubility (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Water Sorption (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Depth of Cure (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Radio-opacity (ISO 4049:2009)Performed as intended (Met ISO 4049:2009 criteria)
    Knoop HardnessPerformed as intended
    Preservation of Surface IntegrityPerformed as intended
    Intensity for curing$\geq550 \text{ mW/cm}^2$
    Wavelength for curing$430-490 \text{ nm}$
    Curing time$20 \text{ sec.}$

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided. The study pertains to non-clinical bench testing and biocompatibility assessments of a dental material, not a diagnostic device with test sets of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. Ground truth in the context of this device would relate to the physicochemical properties and biocompatibility of the material, not expert interpretation of diagnostic data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Adjudication methods are relevant for expert review of human data, not for material property testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was done or is described. This device is a dental filling material, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No such study was done or is described. This device is a dental filling material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation is based on established scientific and regulatory standards for material properties and biocompatibility as outlined in ISO 4049, ISO 10993, ISO 7405, and FDA guidance for dental composite resin devices. It's objective, quantitative measurements against these standards.

    8. The sample size for the training set

    This information is not applicable and not provided. There is no "training set" in the context of material property evaluation for a dental composite.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182714

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-07-12

    (288 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Novidia™ Bulk Fill Flow Composite is indicated for:

      1. Base under Class I and II direct restorations
      1. Liner under direct restorative materials
      1. Pit and fissure sealant
      1. Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations)
      1. Class III and V restorations
      1. Blocking out of undercuts
      1. Repair of small enamel defects
      1. Repair of small defects in esthetic indirect restorations
      1. Repair of resin and acrylic temporary materials
    1. As a core build-up where at least half the coronal tooth structure is remaining to provide structural support for the crown
    Device Description

    Novidia™ Bulk Fill Flow Composite is a low viscosity, visible-light activated, radiopaque, flowable composite indicated for minimally invasive cavity preparations as well as various other indications. The resin matrix of the Novidia™ Bulk Fill Flow Composite contains Urethane dimethacrylate (UDMA), Bisphenol-a glycidyl dimethacrylate (Bis-GMA) and triethylene glycol dimethacrylate (TEGDMA). The inorganic filler of the Novidia™ Bulk Fill Flow Composite is a mix of particles of alumino-silicate-based glasses, silica dioxide and pigments. A small percentage of the silica-based filler particles are covalently bound to quaternary ammonium residues (QASi), added to maintain the integrity of the restoration. The Novidia™ Bulk Fill Flow Composite is provided in syringes (2 g) and in single-dose capsules (0.2 g) and is available in two shades (A2 and A3).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental device, "Novidia™ Bulk Fill Flow Composite." The information focuses on demonstrating substantial equivalence to a predicate device ("Filtek™ Bulk Fill Flowable Restorative") primarily through non-clinical performance testing (bench testing and biocompatibility assessments).

    Therefore, the requested information regarding acceptance criteria and the study that proves the device meets the acceptance criteria will reflect this focus on non-clinical, in vitro performance, rather than clinical study data involving human readers or AI algorithms.

    Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for Novidia™ Bulk Fill Flow Composite are implicitly aligned with the performance characteristics of its predicate device, Filtek™ Bulk Fill Flowable Restorative, as well as the requirements of the ISO 4049:2009 standard for polymer-based restorative materials.

    FeatureAcceptance Criteria (based on ISO 4049:2009 & Predicate)Reported Device Performance (Novidia™ Bulk Fill Flow Composite)
    Flexural Strength≥80 MPa (met by predicate)Met acceptance criteria (explicitly stated)
    Radio-opacity>1.0 mm Al (met by predicate)Met acceptance criteria (explicitly stated)
    Depth of Cure≥2mm at an intensity of ≥ 550 mW/cm2 for 20 seconds (met by predicate)Met acceptance criteria (explicitly stated)
    Water Sorption≤ 40 µg/mm3 (met by predicate)Met acceptance criteria (explicitly stated)
    Water Solubility≤ 7.5 µg/mm3 (met by predicate)Met acceptance criteria (explicitly stated)
    Spontaneous Polymerization Sensitivity at Ambient Light60 ± 5 sec. (physically homogenous by visual inspection) (met by predicate)Met acceptance criteria (explicitly stated)
    Intensity for CuringSimilar to predicate's range (400-1000 mW/cm²)≥550 mW/cm²
    Wavelength for CuringSimilar to predicate's range (400-500 nm)430-490 nm
    Curing TimeSimilar to predicate's (20 sec; 40 sec for opaque shades)20 sec.
    BiocompatibilityPass ISO 10993 and ISO 7405Passed (stated to be biocompatible)
    Other Bench TestsPerforms as intendedPerformance assessed for Elastic modulus, Compression strength, Shade and color stability, Polymerization conversion degree, Viscosity, Knoop hardness, Preservation of Surface Integrity

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (e.g., number of specimens tested for flexural strength) for the non-clinical bench tests. However, it indicates that testing was conducted "per ISO 4049 and FDA guidance document 'Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions'." These standards and guidance documents typically specify minimum sample sizes for material properties testing.

    The data provenance is from non-clinical laboratory testing performed to support the 510(k) submission. No specific country of origin for the testing lab is mentioned, but the submitter, Nobio Ltd., is based in Kadima, Israel. This is retrospective in the sense that the testing was completed prior to the submission for regulatory clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable in the context of this 510(k) submission. The "ground truth" for the device's performance is established by objective, standardized laboratory measurements against established international standards (ISO 4049:2009) and FDA guidance, not by expert consensus or human interpretation of images/data for diagnostic purposes.

    4. Adjudication Method for the Test Set

    Not applicable. This device is a material, and its performance is evaluated through physical and chemical property testing, not through human adjudication of diagnostic outcomes.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental restorative material, not an AI-assisted diagnostic or imaging device. No MRMC study was performed or is relevant. The document explicitly states: "Animal and clinical performance data was not included." and "Clinical studies demonstrating that the presence of QASi in this device improves clinical outcomes have not been conducted."

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on objective, standardized measurements of its physical, chemical, and biological properties as defined by:

    • International standards (e.g., ISO 4049:2009 for polymer-based restorative materials, ISO 10993 for biocompatibility).
    • FDA guidance documents for dental composite resin devices.
    • Comparison to the established performance characteristics of the legally marketed predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI/ML device, there is no training set or associated ground truth establishment process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182580

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-07-03

    (287 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Novidia™ Flowable Composite is indicated for:

    1. Class III and V restorations
    2. Restoration of minimally invasive cavity preparations (including small, non-stress bearing occlusal restorations)
    3. As base/liner under direct or indirect restorations
    4. Repair of small defects in esthetic indirect restorations
    5. Pit and fissure sealing
    6. Blocking out of undercuts
    7. Repair of resin and acrylic temporary materials
    Device Description

    Novidia™ Flowable Composite is a low viscosity, visible light activated, radiopaque, flowable composite indicated for minimally invasive cavity preparations as well as various other indications. The resin matrix of the Novidia™ Flowable Composite contains Urethane dimethacrylate (UDMA), Bisphenol-a glycidyl dimethacrylate (Bis-GMA) and triethylene glycol dimethacrylate (TEGDMA). The inorganic filler of the Novidia™ Flowable Composite is a mix of particles of alumino-silicate-based glasses, silica dioxide and pigments. A small percentage of the silica-based filler particles are covalently bound to quaternary ammonium residues (OASi), added to maintain the integrity of the restoration. The Novidia™ Flowable Composite is provided in syringes (2 g) and in single-dose capsules (0.2 g) and is available in three shades (A1, A2 and A3).

    AI/ML Overview

    The provided text describes a 510(k) submission for a dental device, the Novidia™ Flowable Composite. The document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed clinical study proving acceptance criteria for a new AI/medical image analysis device. Therefore, much of the requested information regarding acceptance criteria, study details, and AI-specific metrics is not available in the provided text.

    However, I can extract and structure the available information regarding performance data, which in this case refers to non-clinical testing.

    1. Table of acceptance criteria and the reported device performance:

    The document outlines a series of non-clinical (bench) tests performed on the Novidia™ Flowable Composite. For several key physical properties, it states that the device was tested according to ISO 4049:2009 standards, which would implicitly define the acceptance criteria, as the statement of "performing as intended" implies meeting these standards. However, the specific numerical acceptance criteria (e.g., minimum flexural strength value) are not explicitly stated in the text, nor are the precise numerical results for the Novidia™ Flowable Composite.

    Acceptance Criteria (Standard Reference)Reported Device Performance (Implicitly Meets Standard)
    Flexural Strength (ISO 4049:2009)Tested and performs as intended
    Elastic ModulusTested and performs as intended
    Compression StrengthTested and performs as intended
    Shade and Color StabilityTested and performs as intended
    Polymerization Conversion DegreeTested and performs as intended
    ViscosityTested and performs as intended
    Spontaneous Polymerization Sensitivity (Ambient Light) (ISO 4049:2009)Tested and performs as intended
    Water Solubility (ISO 4049:2009)Tested and performs as intended
    Water Sorption (ISO 4049:2009)Tested and performs as intended
    Depth of Cure (ISO 4049:2009)Tested and performs as intended
    Radio-opacity (ISO 4049:2009)Tested and performs as intended
    Knoop HardnessTested and performs as intended
    Preservation of Surface Integrity (Resistant to microbial degradation)Novidia Flowable Composite with 1.2% (wt/wt) QASi filler is resistant to microbial degradation evidenced by preserved surface roughness in presence of continuous microbial challenge.

    Biocompatibility Testing:

    Biocompatibility Test (Standard Reference)Reported Device Performance (Implicitly Meets Standard)
    Cytotoxicity (ISO 10993 and ISO 7405)Biocompatible
    Oral Mucosal Irritation Test (ISO 10993 and ISO 7405)Biocompatible
    Acute Systemic Toxicity (ISO 10993 and ISO 7405)Biocompatible
    Material Mediated Pyrogenicity (ISO 10993 and ISO 7405)Biocompatible
    Bacterial Reverse Mutation (ISO 10993 and ISO 7405)Biocompatible
    Mouse Lymphoma Assay (ISO 10993 and ISO 7405)Biocompatible
    Biological Risk Assessment (ISO 10993 and ISO 7405)Biocompatible

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified for any of the bench tests.
    • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective. Given these are bench tests rather than clinical studies, these terms are less applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable and not provided. The data comes from physical/chemical bench testing, not expert interpretation of medical images or conditions.

    4. Adjudication method for the test set:

    • This information is not applicable and not provided. Adjudication is relevant for clinical or image-based studies where multiple readers interpret data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable and not provided. The device is a dental material, not an AI or medical image analysis device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable and not provided. The device is a dental material, not an AI or algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the non-clinical performance and biocompatibility data is established by objective measurements and standardized protocols (e.g., ISO 4049, ISO 10993, ISO 7405 benchmarks for material properties and biological responses). For the "Preservation of Surface Integrity" test, the ground truth was the measured surface roughness under microbial challenge.

    8. The sample size for the training set:

    • This information is not applicable and not provided. This is a material science submission, not an AI/machine learning submission, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    • This information is not applicable and not provided, as there is no training set mentioned or implied.

    In summary, the provided document describes the substantial equivalence of a dental material based on its physical, chemical, and biological properties evaluated through non-clinical bench testing. It does not contain information relevant to AI/medical image analysis device evaluation criteria such as clinical study design, expert consensus, or reader performance metrics.