510(k) Data Aggregation

    K Number
    K251977

    Validate with FDA (Live)

    Date Cleared
    2025-10-22

    (117 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The electric breast pump (models FZ-P8, YH-P8, YH-P8X) is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

    Device Description

    The electric breast pump (models FZ-P8, YH-P8, YH-P8X) is an electrical wearable breast pump powered by a rechargeable Li-ion battery. It consists of a Pump, Diaphragm, Cup, Valve, Silicone breast shield (flange) and dust cap:

    The pump is the core component of the electric breast pump. It generates a cyclic negative pressure (vacuum) to simulate a baby's sucking motion, drawing milk from the breast into the collection system. The pump is electrically driven and controlled via a motor and control unit, typically offering adjustable suction levels and mode. The diaphragm separates the milk collection system from the pump mechanism, preventing breast milk from entering the pump motor and tubing. The cup is the container that collects and stores the expressed breast milk. The valve controls the one-way flow of breast milk. It opens and closes in sync with the suction cycles to allow milk to flow into the cup while preventing backflow. The silicone breast shield, also known as the flange, is the part that comes into direct contact with the breast. It creates a seal around the nipple and areola to enable effective suction. The dust cap covers the opening of the breast shield or milk collection port when the electric breast pump is not in use. It prevents dust, dirt, and other contaminants from entering the breast pump

    The electric breast pump has 2 (Primary mode and secondary mode) x 3 modes (Massage Mode, Expression Mode, and Mixed Mode), each mode has 10 adjustable levels:

    • Massage Mode provides a vacuum range of 40–190 mmHg and a frequency of 52–110 cycles per minute;
    • Expression Mode provides a vacuum range of 80-280 mmHg and a frequency of 19–75 cycles per minute;
    • Mixed Mode provides a vacuum range of 80-280 mmHg and a frequency of 6–63 cycles per minute.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The electric breast pumps are intended to be used by lactating women to express and collect milk from their breasts. They are intended for a single user. The electric breast pumps are intended for at home use.

    Device Description

    The Youha, Bebebao, and Yiyadodo electric breast pumps are single-user, powered breast pump intended to be used by lactating women to express and collect milk from their breasts. THE ONE, YH-8012, BB-5020 and YY-5030 have both single (pumping from one breast) and double pumping (pumping from both breasts) configurations. YY-8011 and YH-8019 have a single pumping configuration only. A reciprocating diaphragm vacuum pump, driven by a microprocessor, generates the suction to extract the milk at vacuum levels up to 280 mmHg. The user interface for the THE ONE, YH-8012, BB-5020 and YY-5030 models consists of a front panel keypad and LCD display or LED display in which the user switches between modes and controls the vacuum pressure. The YY-5030 model does not have any visual indicators on the pump housing. The device has three modes of operation: Massage mode: Suction patterns with fast cycles and low vacuum to start milk flowing; Low expression mode: Suction patterns with slow cycles and high vacuum to express more milk gently and efficiently; High expression mode: Suction patterns with slower cycles and higher vacuum to express more milk efficiently. THE ONE, YH-8012 and BB-5020 have 3 modes including massage mode, low expression mode and high expression mode. YY-5030, YH-8011, and YH-8019 have 2 modes including massage mode and high expression mode. The device is electrically powered from either an internal lithium ion rechargeable battery or an external AC power adapter. The external adapter also charges the battery.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information provided in the document for the Youha Electric Breast Pump (and related models):

    Important Note: The provided document is a 510(k) Summary for a powered breast pump. It is not a document for an AI/ML-driven device. Therefore, many of the requested categories related to AI/ML device studies (e.g., expert adjudication, MRMC study, training set ground truth, effect size) are not applicable to this type of medical device. The information below will focus on the applicable criteria and studies described for a non-AI medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a breast pump and not an AI/ML device, the "acceptance criteria" are related to performance specifications rather than diagnostic accuracy.

    Acceptance Criteria (Performance Specification)Reported Device Performance (Subject Device)
    Suction Strength (Expression Mode)Low: THE ONE: 75-250 mmHg, YH-8012: 75-250 mmHg, BB-5020: 75-250 mmHg
    High: THE ONE: 110-280 mmHg, YH-8012: 110-280 mmHg, YY-5030: 110-280 mmHg, YH-8011: 120-280 mmHg, BB-5020: 110-280 mmHg, YH-8019: 120-280 mmHg
    Suction Strength (Massage Mode)THE ONE: 35-190 mmHg, YH-8012: 35-190 mmHg, YY-5030: 35-190 mmHg, YH-8011: 50-190 mmHg, BB-5020: 35-190 mmHg, YH-8019: 50-190 mmHg
    Adjustable Suction LevelsYes
    Suction Settings (High/Low Expression Mode)THE ONE: 6, YH-8012: 9, YY-5030: 9, YH-8011: 9, BB-5020: 9, YH-8019: 9
    Suction Settings (Massage Mode)THE ONE: 6, YH-8012: 9, YY-5030: 9, YH-8011: 9, BB-5020: 9, YH-8019: 9
    Cycle Speed (Expression Mode)Low: THE ONE: 41-69 cycles/min, YH-8012: 39-77 cycles/min, BB-5020: 39-77 cycles/min
    High: THE ONE: 16-34 cycles/min, YH-8012: 16-35 cycles/min, YY-5030: 26-54 cycles/min, YH-8011: 30-57 cycles/min, BB-5020: 16-35 cycles/min, YH-8019: 30-58 cycles/min
    Cycle Speed (Massage Mode)Not explicitly listed, but inferred to be within acceptable range based on overall performance statement.
    Back Flow ProtectionYes (Demonstrated to meet specifications)
    Performance DurationMaintain specifications for vacuum level, cycle rate, and backflow protection for the duration of a pumping session (identified as 30 minutes) under conditions of single and double pumping mode with varying power sources (AC/DC power vs. battery power). Battery and pump use life demonstrated device maintains specifications throughout its use life under varying power sources (AC, battery).
    Electrical SafetyComplies with IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012.
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2:2014.
    Home Use SafetyComplies with IEC 60601-1-11:2015.
    BiocompatibilityNo new testing required as all patient-contacting materials are identical to those used in cleared Youha electric breast pumps (K163136).
    Software ValidationComplies with IEC 62304:2016 and FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005).
    Risk AnalysisDeveloped in accordance with ISO 14971:2007.

    The document states that "The test results demonstrated that the proposed device complies with the following standards and guidance documents" and "The specifications were met for vacuum level, cycle rate, and backflow protection for the duration of a pumping session... These specifications were maintained under conditions of single and double pumping mode with varying power sources."

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical performance testing of the device itself (hardware and software), rather than a clinical study with human subjects or a dataset derived from humans. Therefore, the concept of a "test set" and "data provenance" (country of origin, retrospective/prospective) as typically applied to AI/ML performance studies, does not directly apply here. The testing involved measuring the device's physical and electrical properties against its design specifications and recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. As this is a non-AI/ML medical device, no expert clinical review for ground truth establishment on a test set (e.g., image interpretation) was performed. The "ground truth" for the device's physical performance was its design specifications and compliance with international standards, which would be verified by qualified engineers and technicians.

    4. Adjudication Method for the Test Set

    Not applicable for a non-AI/ML medical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Its design specifications (e.g., target vacuum levels, cycle rates).
    • Compliance with international and national standards for medical device safety and performance (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 14971, IEC 62304, FDA biocompatibility guidance).

    8. The Sample Size for the Training Set

    Not applicable. This is a non-AI/ML medical device, so there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is a non-AI/ML medical device.