Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K222644

    Validate with FDA (Live)

    Date Cleared
    2023-05-31

    (272 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.

    Device Description

    The NKV-440 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for neonate through adult patients. The NKV-440 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for the Nihon Kohden NKV-440 Ventilator System. It details the device's indications for use, technological characteristics, and comparison to a predicate device (Nihon Kohden NKV-550 Ventilator System).

    The document does not include information about AI/ML algorithm performance, acceptance criteria for such algorithms, or studies involving human readers or ground truth established by experts for image analysis. Therefore, I cannot address most of the specific points requested regarding acceptance criteria and the study proving the device meets them, as it relates to AI/ML device performance.

    This 510(k) pertains to a physical medical device (a ventilator), and its substantial equivalence is demonstrated through engineering performance testing against established standards and comparison of technical characteristics with a legally marketed predicate device.

    However, I can extract the general acceptance criteria for this type of device based on the information provided:

    Summary of Acceptance Criteria and Device Performance (Based on the provided document for a Ventilator):

    Since this is a ventilator and not an AI/ML diagnostic tool, the "acceptance criteria" discussed are related to performance against established engineering and safety standards, and functional equivalence to a predicate device, rather than diagnostic accuracy metrics.

    Acceptance Criteria (General for Ventilators)Reported Device Performance (Nihon Kohden NKV-440 Ventilator System)
    Intended Use Equivalence: Provide continuous ventilation for adult, pediatric, and neonatal patients requiring invasive or noninvasive respiratory support, including mandatory and spontaneous ventilation modes and respiratory monitoring, for use in hospitals and hospital-type facilities, as well as in-hospital transport.Met: "The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation." (Identical to predicate)
    Clinical Conditions Equivalence: Use in hospitals, hospital-type facilities, and in-hospital transportation for patients needing ventilation therapy.Met: "Hospitals, hospital-type facilities and in-hospital transportation for patients who need ventilation therapy." (Identical to predicate)
    Anatomical Site Equivalence: Patient airways.Met: "Patient airways." (Identical to predicate)
    Target Population Equivalence: Adult, pediatric, and neonatal patients.Met: "Adult, pediatric and neonatal patients." (Identical to predicate)
    User Equivalence: Trained clinicians, not lay users.Met: "Trained clinicians, not lay users." (Identical to predicate)
    Compliance with ISO 80601-2-12 (Essential Performance of Critical Care Ventilator):Met: "Met ISO 80601-2-12 requirements on essential performance of critical care ventilator." (Identical to predicate)
    Waveform Comparison: Comparable waveform test result to predicate device.Met: "Comparable waveform test result."
    Chemicals Deployed to Patient Equivalence: Air and oxygen.Met: "Air and oxygen." (Identical to predicate)
    Delivery Method to Patient Equivalence: Positive pressure.Met: "Positive pressure." (Identical to predicate)
    Energy Used for Device Equivalence: AC Power and internal DC Power (Li-ion battery).Met: "AC Power and internal DC Power (Li-ion battery)." (Identical to predicate, noting "Exactly same battery packs are used in both NKV-440 and NKV-550.")
    Control Principle Equivalence: Time-cycled, volume-constant, pressure-controlled.Met: "time-cycled, volume-constant, pressure-controlled." (Identical to predicate)
    Breathing Modes Equivalence: All specified invasive and non-invasive modes (A/CMV-PC, A/CMV-VC, A/CMV-PRVC, SIMV-PC-PS, SIMV-VC-PS, SIMV-PRVC-PS, SPONT-CPAP, SPONT-PS, SPONT-VS, APRV, CPAP, O2 Therapy).Met: All listed modes are identical to the predicate device.
    Compliance with Relevant Standards: (e.g., ANSI AAMI ES 60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62133, IEC 62304, IEC 62366-1, ISO 10993-1, ISO 14971, ISO 18562-1, ISO 80601-2-12, ISO 80601-2-55, ISO 80601-2-61).Met: The device "has been tested in compliance with the following standards" (a comprehensive list is provided).
    Other Non-Clinical Performance: Performance of Ventilation Modes and Control Settings, Device/Software Functionality, Power performance, Essential Performance and Worst Case VBS, Environmental, Product Endurance/Reliability, Biocompatibility, Cleaning & Disinfection, EMC and Electrical Safety, Human Factors/Usability, Risk Management, Compatibility with 3rd Party Devices.Demonstrated: "Performance of the Nihon Kohden NKV-440 Ventilator was demonstrated by the following..." (followed by the list above).

    Regarding the points specific to AI/ML device studies, the document states "Not Applicable" for clinical and animal performance data, and does not mention any AI/ML components inherent to the device or its assessment.

    Therefore, the following points cannot be answered from the provided text:

    1. Sample sizes used for the test set and the data provenance: Not applicable, as this is not an AI/ML clinical study. Performance is demonstrated through engineering tests and comparison to a predicate device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    In conclusion, the supplied document describes the regulatory clearance for a conventional medical ventilator by demonstrating substantial equivalence to a predicate device through non-clinical performance data and adherence to recognized standards. It does not involve AI/ML technology or human reader studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213521

    Validate with FDA (Live)

    Date Cleared
    2022-07-01

    (240 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-330 Ventilator is intended to provide ventilation and oxygen concentration for patients who are breathing spontaneously but need partial ventilation support due to respiratory failure or chronic respiratory insufficiency. It is intended for children weighing 12.5 kg or greater to adult patients. It offers noninvasive ventilation, invasive ventilation, and respiratory monitoring. The NKV-330 is intended for use in hospital-type facilities, and in-hospital transportation by qualified and trained users under the directions of a physician.

    Device Description

    The NKV-330 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for pediatric through adult patients. The NKV-330 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.

    AI/ML Overview

    The provided document is a 510(k) summary for the Nihon Kohden NKV-330 Ventilator System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that a device meets specific acceptance criteria through a clinical study or detailed performance metrics.

    Therefore, the document does not contain the specific acceptance criteria, reported device performance in those terms, details of a specific study proving it, sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details as requested.

    The document primarily states that the device's technical characteristics are "substantially equivalent" to a predicate device (Philips/Respironics V60 Ventilator) and lists various non-clinical performance data and standards compliance to support this claim. It explicitly states "Clinical performance data was not required to demonstrate substantial equivalence."

    Ask a Question

    Ask a specific question about this device

    K Number
    K192307

    Validate with FDA (Live)

    Date Cleared
    2020-01-10

    (137 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.

    Device Description

    The Nihon Kohden NKV-550 Series Ventilator System consists of a graphic user interface (GUI) and a breath delivery unit (BDU). The GUI allows clinicians to set ventilator control parameters such as tidal volume and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, to view waveform and loops, and to operate various features through the apps. The BDU contains a microprocessor that receives inputs from the electronic system and controls the pneumatic system for breath delivery to the patient. It also provides various alarms, a safety valve, and other design features to maximize patient safety.

    AI/ML Overview

    This document, K192307, is a 510(k) premarket notification for a medical device (Nihon Kohden NKV-550 Series Ventilator System). It asserts substantial equivalence to a predicate device. As such, the FDA does not require a clinical study with human patients to demonstrate device performance against specific acceptance criteria for a new AI/ML algorithm. Instead, the focus is on demonstrating that the revised device meets the same performance and safety standards as the predicate device.

    Therefore, the requested information about "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/ML algorithm (including sample sizes, ground truth establishment, expert adjudication, MRMC studies, and standalone performance) is not applicable to this submission. This submission is for a ventilator, not an AI/ML diagnostic or therapeutic device.

    The document discusses non-clinical performance data and states that clinical performance data was not required to demonstrate substantial equivalence.

    Here's a breakdown of why the specific questions about an AI/ML study can't be answered from this document:

    • 1. A table of acceptance criteria and the reported device performance: The document does not provide a table of performance acceptance criteria in the context of an AI/ML algorithm's output. It mentions "Met ISO 80601-2-12 requirements on essential performance of critical care ventilator," which refers to international safety and performance standards for ventilators, not an AI's diagnostic accuracy.
    • 2. Sample sizes used for the test set and the data provenance: Not applicable. There is no AI/ML test set.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications: Not applicable. There is no AI/ML ground truth establishment.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    • 8. The sample size for the training set: Not applicable. There is no AI/ML training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    What the document does provide regarding device performance:

    The document states that the performance of the Nihon Kohden NKV-550 Series Ventilator System was demonstrated by various non-clinical tests performed in compliance with Design Controls:

    • Software Verification
    • Electrical Safety & EMC Testing
    • Device Functionality Testing
    • Performance of Therapy Types and Ventilation Modes
    • Environmental Testing
    • Cleaning & Disinfection
    • Risk Management
    • Accessory Compatibility
    • Regression Testing

    The key "performance" criterion mentioned is meeting ISO 80601-2-12 requirements on essential performance of critical care ventilators. This is a general safety and effectiveness standard for the device's physical and functional operation, not related to an AI's analytical accuracy.

    In summary, this 510(k) submission is for a conventional medical device (a ventilator) with minor modifications, and as such, the detailed clinical study and AI/ML-specific acceptance criteria information requested is not part of this type of regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181695

    Validate with FDA (Live)

    Date Cleared
    2018-12-07

    (163 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.

    Device Description

    The Nihon Kohden NKV-550 Series Ventilator System consists of a graphic user interface (GUI) and a breath delivery unit (BDU). The GUI allows clinicians to set ventilator control parameters such as tidal volume and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, to view waveform and loops, and to operate various features through the apps.

    The BDU contains a microprocessor that receives inputs from the electronic system and controls the pneumatic system for breath delivery to the patient. It also provides various alarms, a safety valve, and other design features.

    AI/ML Overview

    The Nihon Kohden NKV-550 Series Ventilator System's acceptance criteria and the study proving it meets these criteria are described below.

    1. Table of Acceptance Criteria and Reported Device Performance

    Technical CharacteristicAcceptance Criteria (Predicate Device: Draeger Evita Infinity V500, K093633)Reported Device Performance (Nihon Kohden NKV-550 Series Ventilator System)Comparison
    Indication for UseIntended for the ventilation of adult, pediatric and neonatal patients. Offers mandatory ventilation modes and ventilation modes for spontaneous breathing support and airway monitoring. Intended for use in different medical care areas (hospitals, medical rooms) or for patient transportation within the hospital.Intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. Offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. Intended for use in hospitals and hospital-type facilities, as well as for in-hospital transportation.Same
    Environment of UseHospitals, hospital-type facilities and in-hospital transportation for patients who need ventilation therapyHospitals, hospital-type facilities and in-hospital transportation for patients who need ventilation therapySame
    Anatomical SitePatient airwaysPatient airwaysSame
    Target PopulationAdult, pediatric and neonatal patientsAdult, pediatric and neonatal patientsSame
    PerformanceMet ISO 80601-2-12 requirements on essential performance of critical care ventilatorMet ISO 80601-2-12 requirements on essential performance of critical care ventilatorSame
    Waveform ComparisonComparable waveform test resultComparable waveform test resultSubstantially equivalent
    DesignConsists of a graphic user interface to set and monitor ventilation, breath delivery unit, breathing circuit; Controls air and oxygen deliveries by proportional valves through microprocessorsConsists of a graphic user interface to set and monitor ventilation, breath delivery unit, breathing circuit; Controls air and oxygen deliveries by proportional valves through microprocessorsSubstantially equivalent
    Chemicals Delivered to PatientMedical Air and OxygenMedical Air and OxygenSame
    Delivery method to PatientPositive pressurePositive pressureSame
    Energy Used for DeviceAC Power and DC Power (battery)AC Power and DC Power (battery)Same
    Control principleTime-cycled, volume-constant, pressure-controlledTime-cycled, volume-constant, pressure-controlledSame
    Therapy TypesIntubated, Non-invasive, O2 TherapyInvasive, Non-invasive, O2 TherapySame

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state a "test set" in the context of clinical data for performance evaluation. Instead, the performance was demonstrated through non-clinical testing against various standards and comparisons to a predicate device. Therefore, a specific sample size for a test set of patient data is not applicable here as clinical performance data was not required for substantial equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as clinical performance data was not required, and thus, no ground truth established by experts on patient data was utilized for the device's clearance. The "ground truth" for the non-clinical testing was established by the requirements and thresholds set by the relevant international standards (e.g., ISO 80601-2-12, ANSI AAMI ES60601-1, etc.).

    4. Adjudication method for the test set:

    Not applicable, as clinical performance data and expert-established ground truth were not used for the device's clearance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a ventilator system, not an AI-assisted diagnostic or interpretation tool that would involve human readers or MRMC studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device's performance was evaluated in a standalone manner by testing against established international standards. The listed "Summary of Non-Clinical Performance Data" includes:

    • Software Verification
    • Electrical Safety & EMC Testing
    • Agency Testing to Applicable Standards
    • Device Functionality Testing
    • Performance of Therapy Types and Ventilation Modes
    • Product Endurance
    • Power Performance
    • Environmental
    • Cleaning & Disinfection
    • Gas Path Biocompatibility
    • Human Factors/Usability Testing
    • Risk Management
    • Waveform Comparison with Predicate
    • Accessory Compatibility

    These tests assess the device's intrinsic functions and compliance with safety and performance requirements, which can be considered standalone performance.

    7. The type of ground truth used:

    For the non-clinical performance evaluation, the "ground truth" was derived from:

    • International Standards: e.g., ISO 80601-2-12 (essential performance of critical care ventilators), ANSI AAMI ES60601-1 (general requirements for basic safety and essential performance), IEC 60601-1-2 (electromagnetic compatibility), IEC 60601-1-6 (usability), IEC 60601-1-8 (alarm systems), IEC 62133 (batteries), ISO 80601-2-55 (respiratory gas monitor), ISO 10993-1 (biological evaluation), ISO 18562-1 to 3 (biocompatibility of breathing gas pathways).
    • Predicate Device Characteristics: The Nihon Kohden NKV-550 Series Ventilator System was compared to the predicate device (Draeger Evita Infinity V500) based on its documented technical characteristics and performance, aiming for substantial equivalence.

    8. The sample size for the training set:

    Not applicable, as this device does not appear to involve machine learning or AI models that would require a training set of data. Its performance is based on engineering design and meeting pre-defined physical and functional specifications.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no mention of a training set or machine learning components in the provided documentation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1