510(k) Data Aggregation

    K Number
    K221847

    Validate with FDA (Live)

    Date Cleared
    2024-04-11

    (657 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ni2 Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller Ni2 dental implants (Ø3.6, 4.0, 5.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger Ni2 Implants (Ø6.0, 7.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.

    Device Description

    The Ni2 Implant System is comprised of dental implants, superstructures, instruments for prosthetics and surgical instruments. The Ni2 Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions. They are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations.

    The Ni2 Implant System, fixtures are made of commercial pure titanium, grade 4 (ASTM F67) which have a S.L.A (Sand blasted large grit acid etched) treated surface and supplied sterile (gamma radiation). These fixtures can be used one stage surgery method or two stage surgery method. And that are surgically inserted into the upper and/or lower jawbone. The fixtures replace tooth root as providing a stable foundation for restorations.

    Geometrically, the implant is screw type. An abutment is connected to the implant through a tapered joint. The Ni2 fixtures are compatible with the abutments which have been previously cleared in K221866.

    The non-threaded section of the implant body is also SLA surface treated.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study proving a device meets them.

    Based on the provided document (K221847 for the Ni2 Implant System), there is NO information about AI/ML device performance or associated studies.

    The document is a 510(k) premarket notification for a dental implant system, and it focuses on demonstrating "substantial equivalence" to existing predicate devices through non-clinical testing. It explicitly states: "No clinical studies are submitted."

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI/ML device, as the provided text pertains to a traditional medical device (dental implant).

    To answer your request, if this were an AI/ML device, I would need a document that describes studies related to the AI/ML algorithm's performance.

    However, I can extract information relevant to the dental implant system's equivalence, which might be mistaken for "acceptance criteria" in a broader sense within the context of a 510(k) submission.

    If we interpret "acceptance criteria" as the parameters used to demonstrate substantial equivalence for this traditional dental implant device:

    Here's how the document addresses the device's characteristics and how its equivalence is proven:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't present a formal table of "acceptance criteria" and "reported device performance" in the way one would for an AI/ML algorithm (e.g., sensitivity, specificity thresholds). Instead, it demonstrates substantial equivalence by comparing the characteristics of the Ni2 Implant System to predicate and reference devices. The "performance" is implicitly deemed acceptable if it is substantially equivalent to legally marketed devices.

    Here's a summary of the comparative characteristics:

    CharacteristicNi2 Implant System (Subject)S-Plant Dental Implant System (Primary Predicate, K221866)DIO UF HSA Internal Sub-merged Implant System (Reference, K122519)Noris Medical Dental Implants System -Cortical (Reference, K210852)
    Regulation #21 CFR 872.364021 CFR 872.364021 CFR 872.364021 CFR 872.3640
    Regulation NameEndosseous Dental ImplantEndosseous Dental ImplantEndosseous Dental ImplantEndosseous Dental Implant
    ClassificationClass 2Class 2Class 2Class 2
    Indications for UseSurgical placement in upper/lower jaw for prosthetic attachment to restore chewing function. Smaller implants (Ø3.6, 4.0, 5.0 mm) for conventional two-stage with transmucosal healing or single-stage immediate loading (with good primary stability/appropriate occlusal loading). Larger implants (Ø6.0, 7.0 mm) for conventional two-stage (with transmucosal healing) in molar region with delayed loading.Surgical placement in upper/lower jaw for prosthetic attachment to restore chewing function. Smaller implants (Ø3.4, 3.6, 3.8, 4.2, 4.7, 5.2 mm) for conventional two-stage with transmucosal healing or single-stage immediate loading (with good primary stability/appropriate occlusal loading). Larger implants (Ø6.0, 7.0 mm) for conventional two-stage (with transmucosal healing) in molar region with delayed loading.Surgical placement in upper/lower jaw for prosthetic attachment to restore chewing function. Smaller implants (Ø3.8 ~ Ø5.5) for conventional two-stage with transmucosal healing or single-stage immediate loading (with good primary stability/appropriate occlusal loading). Larger implants (Ø6.0 ~ Ø7.0) for conventional two-stage (with transmucosal healing) in molar region with delayed loading.Intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices. One-stage or two-stage surgical operation. All implants appropriate for immediate loading (with good primary stability/appropriate occlusal loading).
    Prosthetic Interface ConnectionTapered conical hexTapered conical hexTapered conical hexInternal Hex
    Fixture Diameter & Endosseous Length (mm)Ø3.6 (8.5, 10.0, 11.5, 13.0, 15.0); Ø4.0 (7.0, 8.5, 10.0, 11.5, 13.0, 15.0); Ø5.0 (7.0, 8.5, 10.0, 11.5, 13.0, 15.0); Ø6.0 (7.0, 8.5, 10.0, 11.5); Ø7.0 (7.0, 8.5) (L1 total length)Ø3.4 (8.5, 10.0, 11.5, 13.0, 15.0); Ø3.6 (8.5, 10.0, 11.5, 13.0, 15.0); Ø3.8 (7.0, 8.5, 10.0, 11.5, 13.0, 15.0); Ø4.2 (7.0, 8.5, 10.0, 11.5, 13.0); Ø4.7 (7.0, 8.5, 10.0, 11.5, 13.0); Ø5.2 (7.0, 8.5, 10.0, 11.5); Ø6.0 (7.0, 8.5, 10.0, 11.5); Ø7.0 (7.0, 8.5) (total length)Ø3.8 (7.0, 8.5, 10, 11.5, 13.0, 15.0, 16.0); Ø4.0 (7.0, 8.5, 10, 11.5, 13.0, 15.0, 16.0); Ø4.5 (7.0, 8.5, 10, 11.5, 13.0, 15.0, 16.0); Ø5.0 (7.0, 8.5, 10, 11.5, 13.0, 15.0, 16.0); Ø5.5 (7.0, 8.5, 10, 11.5, 13.0, 15.0, 16.0); Ø6.0 (7.0, 8.5); Ø7.0 (7.0, 8.5) (total length)Ø4.0 (7.5, 9, 12, 14, 16); Ø5.0 (7.5, 9, 12); Ø6.0 (7.5, 9, 12) (total length)
    Non-Threaded Length (mm)2.0 mm for all lengths (within bone, SLA treated).N/A (implied by lack of entry, assumed fully threaded or similar to predicate)N/A (implied by lack of entry, assumed fully threaded or similar to predicate)For L7.5: 0.7; For L9: 1.0; For L12: 2.0; For L14: 3.0; For L16: 3.5 (within bone, surface treated).
    Implant MaterialCP Ti Grade 4 (ASTM F67)CP Ti Grade 4 (ASTM F67)CP Ti Grade 4 (ASTM F67)CP Ti Grade 4 (ASTM F67)
    Implant Endosseous SurfaceS.L.A.S.L.A.S.L.A.RBM (Resorbable Blasting Media)
    Implant PlacementBone LevelBone LevelBone LevelBone Level
    SterilizationGamma SterilizationGamma SterilizationGamma SterilizationGamma Sterilization
    Shelf Life8 years8 years5 years5 years

    Study to prove device meets the acceptance criteria (Substantial Equivalence):

    The "study" refers to the non-clinical tests performed to demonstrate that the Ni2 Implant System performs as safely and effectively as, or is substantially equivalent to, legally marketed predicate devices.

    2. Sample sized used for the test set and the data provenance

    • Sample size: Not explicitly stated in terms of number of physical implants tested for each mechanical or material property. The document refers to "non-clinical test data" validating substantial equivalence.
    • Data provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This would typically be a laboratory/bench-test setting.
    • Retrospective or prospective: The tests appear to be prospective, performed specifically for this 510(k) submission or leveraged from prior submissions (e.g., K221866).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable as there is no "test set" in the context of human expert review for an AI/ML algorithm. The "ground truth" for a traditional dental implant device is established through engineering specifications, material standards (ASTM, ISO), and validated test methods (e.g., fatigue testing, biocompatibility testing).

    4. Adjudication method for the test set

    • Not applicable as there is no "test set" requiring human adjudication in the context of an AI/ML algorithm.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is explicitly stated: "No clinical studies are submitted." An MRMC study is relevant for AI/ML diagnostic devices, not traditional dental implants.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an AI/ML algorithm.

    7. The type of ground truth used

    • For this traditional device, the "ground truth" is established by:
      • Engineering specifications and design parameters: Dimensions, tolerances, connection types.
      • Material standards: ASTM F67 (for CP Ti Grade 4).
      • Biocompatibility standards: ISO 10993-1.
      • Sterilization standards: ISO 11137-1, 2, 3.
      • Sterile barrier shelf-life standards: ASTM F1980.
      • Imaging analysis techniques: SEM (Scanning electron microscopy) and EDS (Energy Dispersive X-ray Spectroscopy) for surface characteristics.
      • Mechanical/Fatigue testing: Although specific results are not detailed, these are implied by the broad statement "non-clinical test data submitted for the Ni2 implant system validates its substantial equivalence."

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device.

    In conclusion, the provided document clearly indicates that the Ni2 Implant System is a traditional medical device (dental implant). The submission focuses on demonstrating substantial equivalence to predicate devices through non-clinical bench testing and adherence to recognized standards, not through AI/ML algorithm performance studies or clinical trials involving human readers/interpreters.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221866

    Validate with FDA (Live)

    Date Cleared
    2024-01-18

    (570 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S-Plant Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller S-Plant Dental Implants (03.4, 3.6, 3.8, 4.2, 4.7, 5.2 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger S-Plant Dental Implants (06.0, 7.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.

    Dual abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.

    Device Description

    The S-Plant Dental Implant System is comprised of dental implants, superstructures, instruments for prosthetics and surgical instruments. The S-Plant Dental Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions. There are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations.

    The S-Plant Dental Implant System fixtures are made of commercial pure titanium, grade 4 (ASTM F67) which have a S.L.A (Sand blasted large grit acid etched) treated surface and supplied sterile (gamma radiation). These fixtures can be used one stage surgery method or two stage surgery method. And that are surgically inserted into the upper and/or lower jawbone. The fixtures replace tooth roots providing a stable foundation for restorations. Geometrically, the implant is screw type. An abutment is connected to the implant through a tapered joint.

    The Abutment made of Ti-6AL-4V ELI alloy (ASTM F136) is intended for use as an aid in single or multiple-unit prosthetic restorations. It consists of Healing Abutment, and Abutment Screws. All abutments are supplied non-sterile and autoclaved by the end user.

    The Cover Screw made of Ti-6Al-4V ELI alloy (ASTM F136), is an essential component in dental implant procedures. This device safeguards the internal threads of dental implant fixtures during the healing phase, ensuring a sterile environment for successful osseointegration.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the S-Plant Dental Implant System, which is a Class II medical device. The submission aims to demonstrate substantial equivalence to legally marketed predicate devices. However, it does not contain information typically found in a study demonstrating the device meets acceptance criteria related to clinical performance or AI models.

    The provided text solely focuses on establishing substantial equivalence based on comparisons to predicate and reference devices regarding design, materials, indications for use, and non-clinical testing (biocompatibility, sterilization, surface modification, and MRI compatibility). There is no mention of an AI component, human-in-the-loop performance, or specific performance metrics like sensitivity, specificity, or AUC against a clinical ground truth for the S-Plant Dental Implant System itself.

    Therefore, I cannot fulfill your request for the following information as it is not present in the provided document:

    • A table of acceptance criteria and the reported device performance (for clinical performance/AI).
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth.
    • Adjudication method.
    • Multi Reader Multi Case (MRMC) comparative effectiveness study results or effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for clinical evaluation.
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    The provided text is a 510(k) summary for a dental implant system, which is a physical device, and not an AI/software device. The "acceptance criteria" discussed are related to regulatory conformance and manufacturing standards, not clinical performance metrics in the context of an AI study.