510(k) Data Aggregation

    K Number
    K213969

    Validate with FDA (Live)

    Date Cleared
    2022-10-07

    (291 days)

    Product Code
    Regulation Number
    878.3720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VisionAir Patient-Specific Airway Stent is indicated for the treatment of adults ≥22 years of age with symptomatic stenosis of the airway. The silicone stent is intended for implantation into the airway by a physician using the recommended deployment system or an equivalent rigid bronchoscope and stent placement system that accepts the maximum stent diameter being placed. The stent is intended to be in the patient up to 12 months after initial placement.

    Device Description

    The subject device, VisionAir Patient-Specific Airway Stent is comprised of a cloudbased software suite and the patient-specific airway stent. These two function together as a system to treat symptomatic stenosis of the airway per the indications for use. The implantable patient-specific airway stent is designed by a physician using a CT scan as a guide in the cloud-based software suite. The airway is segmented from the CT scan and used by the physician in designing a patient-specific stent. When design is complete, the stent is manufactured via silicone injection into a 3D-printed mold and delivered to the treating physician nonsterile, to be sterilized before use.

    The implantable patient-specific airway stent includes the following general features:

    • Deployed through a compatible rigid bronchoscope system
    • Made of biocompatible, implant-grade silicone
    • Steam sterilizable by the end user
    • Anti-migration branched design
    • Anti-migration studs on anterior surface of main branch
    • Single-use

    The cloud-based software suite has the following general features:

    • Upload of CT scans
    • Segmentation of the airway
    • Design of a patient specific stent from segmented airway
    • Order management of designed stents
    AI/ML Overview

    The provided text is a 510(k) Summary for the VisionAir Patient-Specific Airway Stent, which focuses on demonstrating substantial equivalence to a predicate device. It primarily discusses the device description, indications for use, technological characteristics, and a list of nonclinical performance and functional tests conducted.

    However, the document does not contain the detailed information required to fulfill the request regarding acceptance criteria and the study that proves the device meets those criteria. Specifically, it lacks:

    1. A table of acceptance criteria and reported device performance: While it lists types of tests, it does not provide specific quantitative acceptance criteria or the actual results from these tests.
    2. Sample size used for the test set and data provenance: No information is given about the sample size for any clinical or performance test, nor the origin or nature of the data (retrospective/prospective, country).
    3. Number of experts used to establish ground truth and qualifications: This information is completely absent.
    4. Adjudication method for the test set: Not mentioned.
    5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study details: No MRMC study is described; the testing mentioned is primarily non-clinical or related to software validation/verification, not human-AI comparative effectiveness.
    6. Standalone (algorithm-only) performance: While "Software Verification and Validation Testing" and "Airway Segmentation Process Testing" are mentioned, no specific standalone performance metrics (e.g., accuracy, precision for segmentation) or acceptance criteria are provided.
    7. Type of ground truth used: The document mentions "Airway Segmentation Process Testing" and refers to a predicate device (Mimics) for "performance reference specification" for dimensional testing of airway segmentation. This implies that the ground truth for segmentation would likely be derived from expert-reviewed segmentations or potentially from known anatomical measurements, but the method is not explicitly detailed.
    8. Sample size for the training set: There is no mention of a "training set" or any machine learning model that would require one. The software aspect described is for physician-guided design and semi-automated segmentation, not explicitly an AI/ML model that undergoes a training phase in the typical sense for medical image analysis.
    9. How the ground truth for the training set was established: Not applicable, as no training set is described.

    The document states: "Reference devices, Mimics (K073468) and Osirix MD (K101342) were used for reference software performance specifications." and "Dimensional Testing of Airway Segmentation (reference device Mimics K073468 used for performance reference specification)". These statements hint at software validation, especially for the segmentation component, but do not provide the detailed study design, acceptance criteria, or results.

    In summary, the provided text does not contain the necessary information to answer the request in detail, as it focuses on demonstrating substantial equivalence through non-clinical performance and functional testing rather than a clinical study with acceptance criteria for device performance based on human reader interaction or AI model performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182743

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-23

    (390 days)

    Product Code
    Regulation Number
    878.3720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Patient-Specific Airway Stent is indicated for the treatment of adults ≥22 years of age with symptomatic stenosis of the airway. The silicone stent is intended for implantation into the airway by a physician using the recommended deployment system or an equivalent rigid bronchoscope and stent placement system that accepts the maximum stent diameter being placed. The stent is intended to be in the patient up to 12 months after initial placement.

    Device Description

    The Patient-Specific Airway Stent is comprised of Web Software and the Patient-Specific Silicone Y-Stent. These two function together as a system to treat symptomatic stenosis of the airway. The Patient-Specific Silicone Stent is designed by a physician using a CT scan as a guide in the Web Software. The Web Software gives the user (physician) the ability to upload a scan, view the airway, and design a stent. The stent, after physician approval, is manufactured via silicone injection into a 3D-printed mold and delivered to the treating physician's medical center nonsterile.

    AI/ML Overview

    This document (K182743) is a 510(k) Premarket Notification for a Patient-Specific Airway Stent. It primarily focuses on demonstrating substantial equivalence to a predicate device (ENDOXANE, K971509) and the safety/effectiveness of the device.

    Based on the provided text, the device in question is a Patient-Specific Airway Stent System, which includes Web Software and the Patient-Specific Silicone Y-Stent. The software allows physicians to design the stent based on a CT scan, and then the stent is manufactured via silicone injection into a 3D-printed mold.

    The acceptance criteria and study that proves the device meets them are mostly related to non-clinical performance testing and software verification/validation, rather than a full clinical study with human patients evaluating the AI's diagnostic performance. Therefore, many of the typical acceptance criteria and study elements for an AI-powered diagnostic device are not explicitly detailed in this 510(k) summary.

    Here's an attempt to extract the relevant information based on your request, acknowledging the limitations of the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide a direct table of acceptance criteria with specific numerical performance metrics for the software's ability to accurately design the stent or for the final stent's intended clinical outcome in terms of patient-specific fit. Instead, it describes various tests performed to ensure the device functions as intended and is substantially equivalent to the predicate.

    The closest to "acceptance criteria" are the objectives of the non-clinical performance tests, and "reported device performance" is described qualitatively as "supports the claim" or "confirms."

    Please note: The "device performance" here refers to the engineering and functional performance of the stent and software, not clinical outcome metrics (e.g., patient improvement rates).

    Acceptance Criteria (Stated Objective of Test)Reported Device Performance (as described in document)
    Sterilization Testing: Ability to be steam sterilized to a SAL of 10-6.Confirms that the subject device can be steam sterilized to a SAL of 10-6, using a common cycle in medical centers.
    Material Equivalence (Tear Strength): Subject device material equivalent to predicate's in tear strength.Supports the claim that the subject device's material is equivalent to the predicate's material in tear strength.
    Fatigue Testing: Device does not fatigue when cyclically compressed over intended life (1 year).Supports the claim that the subject device does not fatigue when cyclically compressed over the intended life of the implant (1 year).
    Stent Deployment Testing: Able to be deployed by common applicator and rigid bronchoscope system.Supports the claim that the subject device is able to be deployed by a common applicator and rigid bronchoscope system.
    Biocompatibility Testing: Acceptable for implant up to one year.Confirmed that the Patient-Specific Silicone Stent is acceptable for use as a medical device following ISO 10993-1. (Specific tests: Cytotoxicity, Sensitization, Irritation, Toxicity, Pyrogenicity, Subacute/Sub-Chronic Toxicity, Genotoxicity, Chemical Characterization).
    Software Verification and Validation: Software functions as designed; risk mitigations are effective.Supports the claim that the software functions as designed, including any design mitigations.
    Human Factors and Usability Testing (Web Software): Software is safe and effective when used by intended users in its intended use-environment.Supports the claim that the software is as safe and as effective as the predicate device when used by its intended users in its intended use-environment.
    Dimensional Testing of Airway Segmentation: Accuracy of segmented airway rendering in proprietary software.Evaluated the accuracy of segmented airway rendering in proprietary segmentation software. (No specific metric provided, just that it was evaluated).
    Airway Segmentation Process Validation: Validation of the process of segmenting an airway from a CT scan in proprietary software.Validated the process of segmenting an airway from a CT scan in proprietary segmentation software. (No specific metric provided, just that it was validated).

    Note on Quantitative Acceptance Criteria: The document explicitly mentions some differences, such as the subject device having lower flat-plate compression strength than the predicate device. However, it states that "Any risks related to these technological differences have been mitigated to an acceptable level," implying that these differences did not prevent meeting overall safety/effectiveness. For the AI component (segmentation and design), specific quantitative acceptance criteria (e.g., Dice score for segmentation accuracy, deviation from ideal stent dimensions) are not provided in this summary document.

    2. Sample Size for Test Set and Data Provenance

    • The document does not specify a sample size for the "test set" in the context of typical AI performance evaluation (e.g., number of CT scans used to validate segmentation or design accuracy).
    • The closest mentions are "Dimensional Testing of Airway Segmentation" and "Airway Segmentation Process Validation." It's implied that some CT scan data was used for these, but neither the sample size nor the provenance (country, retrospective/prospective) of this data is mentioned.

    3. Number of Experts and Qualifications for Ground Truth

    • The document describes the device as a "Patient-Specific Airway Stent" where creation involves a "physician using a CT scan as a guide in the Web Software" and the stent is manufactured "after physician approval."
    • The "Web Software" allows "COS technicians to segment the airway and automatically calculate a centerline."
    • "Ground truth" for the AI component (segmentation, centerline calculation, stent design) is not explicitly defined in terms of expert consensus or pathological verification in this summary. Instead, it appears the software's output is reviewed and approved by a single physician for an individual patient.
    • Therefore, it's not a panel of experts establishing ground truth for a general test set, but rather an individual physician performing the critical review and approval step for each patient. No specific number of experts used to establish ground truth for a general test set is mentioned, nor are their qualifications.

    4. Adjudication Method for the Test Set

    • Given that the "ground truth" for the AI's output is implied to be physician review and approval for each specific case, there is no multi-reader adjudication method (like 2+1 or 3+1) described for a general test set. The process involves one physician approving the design.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study comparing human readers with and without AI assistance is mentioned or implied. The product is a custom-designed stent system, not a diagnostic AI intended to assist in interpreting medical images. The AI (software) assists in the design and manufacturing process, which is then approved by a physician.

    6. Standalone (Algorithm Only) Performance

    • No standalone (algorithm only) performance metrics are explicitly provided. The software acts as a design tool that then requires physician approval before manufacturing. The document describes "Software Verification and Validation Testing" and "Dimensional Testing of Airway Segmentation" which implies internal testing of the algorithm, but specific standalone metrics (e.g., segmentation accuracy against ground truth) are not reported in this summary.

    7. Type of Ground Truth Used

    • For the software's AI components (segmentation, centerline calculation), the implicit "ground truth" during real-world use is the physician's subjective review and approval based on the CT scan.
    • For the non-clinical tests (e.g., tear strength, fatigue), the ground truth relates to engineering specifications and established test methods.
    • No pathology or outcomes data is mentioned as ground truth for the software's performance, as this is a device design tool, not a diagnostic algorithm.

    8. Sample Size for the Training Set

    • The document does not specify the sample size for the training set used for any AI component (segmentation, centerline calculation). It only refers to a "proprietary software" used by "COS technicians" to segment the airway and calculate the centerline.

    9. How Ground Truth for Training Set Was Established

    • The document does not describe how ground truth for any potential AI training set was established.