510(k) Data Aggregation

    K Number
    K252376

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-28

    (90 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Yomi S Robotic System (Yomi S) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi S Robotic System. The output of YomiPlan is to be used with the Yomi S Robotic System.

    Device Description

    The Neocis Yomi S is a modified next iteration of the Yomi Robotic System, designed to provide guidance for a dental surgeon during dental implant surgery. Yomi S is a dental stereotaxic medical device (Product Codes PLV, QRY) regulated under 21 CFR 872.4120. The device includes a YomiLink that is placed on the patient prior to the CT scan, and a fiducial array with fiducial markers that is placed on the YomiLink prior to the CT scan so the virtual plan can be related to the physical space of the system. The Guidance Arm secures a standard dental drill, allowing the surgeon to grip the drill as normal. The Guidance Arm does not move unless the surgeon applies a manual force to the drill. The Guidance Arm will constrain the surgeon to drill according to the prescribed surgical plan, preventing deviation. The surgeon is constantly in control of the drilling. The system has a mechanical feedback system that is connected to the YomiLink on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the patient movement, which will maintain the accuracy of the drill placement.

    The Yomi S Robotic System allows the user to plan the surgery virtually in YomiPlan, cleared for use alone on third-party PCs for preplanning. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in the planning software. The plan is used for the system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The Yomi S robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument.

    The patient tracking portion of Yomi S is comprised of linkages from the patient to Yomi S, which include the Patient Splint (YomiLink Teeth or YomiLink Bone), Tracker End Effector (TEE), and the Patient Tracker (PT). In cases where YomiLink Teeth is utilized, it is attached to the contralateral side of the patient's mouth over stable teeth using on-label dental materials prior to the presurgical CBCT scan. In cases where YomiLink Bone is utilized, it is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required), or after the scan when using the subject YomiLink Arch device.

    The subject of this submission is to: introduce the Yomi S Robotic System, a next-generation modification of the Yomi Robotic System, intended to assist dental surgeons by providing guidance during dental implant procedures.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251835

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-10

    (116 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Yomi Robotic System (Yomi) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.

    Device Description

    Yomi Robotic System is a dental stereotaxic instrument and a powered surgical device for bone cutting. Yomi Robotic System is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    The Yomi Robotic System allows the user to plan the surgery virtually in YomiPlan, cleared for use alone on third-party PCs for preplanning. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in the planning software. The plan is used for the system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The Yomi robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument.

    The patient tracking portion of Yomi is comprised of linkages from the patient to Yomi, which include the Patient Splint (YomiLink Teeth or YomiLink Bone), Tracker End Effector (TEE), and the Patient Tracker (PT). In cases where YomiLink Teeth is utilized, it is attached to the contralateral side of the patient's mouth over stable teeth using on-label dental materials prior to the presurgical CBCT scan. In cases where YomiLink Bone is utilized, it is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required), or after the scan when using the subject YomiLink Arch device.

    The subject of this submission is to: Integrating algorithms that provide automatic segmentation of maxillary sinuses, inferior alveolar nerve, and maxillary and mandibular bone. The integrated software, Relu Creator, was cleared in K233925. The software is not adaptive, it is trained at the manufacturer (Relu), and the weights are locked.

    Additionally, since the most recent clearance of Yomi Robotic System (K231018), minor modifications to the Yomi System include the following:
    • Planning software improvements
    • Restorative planning – Features to support customized crown design
    • Dual arch planning – Feature to enable the end user to plan multiple arches in a singe case and a singe scan
    • Patient work volume guidance improvements – Added guidance for the angulation of the patient chair
    • Added patient proximity for baseline
    • YomiLink Bone (YLB) planning – improved placement of the YLB
    • Added proximity threshold lower limit value
    • Improved alignment between CT scans and imported .stl objects
    • Added ability for user to designate soft tissue thickness to assist in bone reduction planning
    • Added max depth information to the implant cursor hover info
    • VTK Off-the-Shelf software version update
    • Added model details to implant selection
    • Added restorative planning case feedback option
    • Added additional implant models to the implant library

    • Control software and behavior improvements
    • Updates to handpiece interaction gestures, and optimization of the response of the control software to guide arm joint limits, singularities and potential wrist / base collisions.

    • Hardware improvements Tracker Arm Joint

    • Accessory improvements
    • Updates to the YomiLink Teeth and intraoral fiducial array

    • Minor bug fixes

    All other aspects of the Yomi Robotic System remain unchanged from prior clearances.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Yomi Robotic System focuses on the substantial equivalence of the modified device to its predicate. While it mentions the integration of an automatic segmentation algorithm (Relu Creator, K233925), it does not contain the detailed acceptance criteria or the specific study that proves the device meets those criteria for the automatic segmentation algorithm.

    The document primarily describes:

    • The indications for use.
    • A comparison of technological characteristics between the subject device (Yomi Robotic System with Automatic Segmentation Algorithm) and its predicate (Yomi Robotic System K231018) and a reference device (Relu Creator K233925).
    • General statements about software, cybersecurity, and usability verification and validation testing, but without specific performance metrics or study details.

    Therefore, many of the requested details about acceptance criteria, specific performance results, sample sizes, expert qualifications, and ground truth establishment for the automatic segmentation algorithm are not present in the provided text. The document refers to the Relu Creator (K233925) as having been cleared, implying its own performance evaluations would have been submitted in that separate clearance.

    Here's a breakdown of the information that can be extracted or inferred, and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not explicitly provided for the automatic segmentation algorithm (Relu Creator) in this document. The document states:

    • "Yomi Plan 2.7 with Automatic Segmentation Algorithm functionality was successfully verified and user validated."
    • "The software has been successfully verified to perform with the PC specifications of the Yomi Robotic System."
    • "All changes have been successfully verified and, therefore, not considered to affect the overall safety and efficacy profile of Yomi Plan."
    • "The combined testing and analysis of results provides assurance that the device performs as intended."

    These are general assurances of performance and validation but do not provide specific quantitative acceptance criteria or reported device performance metrics for the automatic segmentation algorithm itself.

    2. Sample size used for the test set and the data provenance

    Not provided in this document. The document mentions "Software verification and validation testing" and "User Validation testing" but does not specify the sample size of cases or the provenance (country of origin, retrospective/prospective) of the data used for testing the automatic segmentation algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not provided in this document.

    4. Adjudication method for the test set

    Not provided in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not provided in this document. The document does not mention an MRMC study or any results comparing human reader performance with and without AI assistance from the segmentation algorithm. The automatic segmentation algorithm is integrated into the planning software to assist (presumably by providing pre-segmented anatomy), but its impact on human reader performance is not quantified here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not explicitly detailed for the segmentation algorithm's performance. The document states, "The integrated software, Relu Creator, was cleared in K233925." This implies that the Relu Creator, which performs the automatic segmentation, underwent its own standalone performance evaluation as part of its original clearance (K233925). This current 510(k) focuses on its integration into the Yomi Robotic System, not its primary standalone performance evaluation.

    7. The type of ground truth used

    Not explicitly provided in this document for the automatic segmentation algorithm. For image segmentation algorithms, ground truth is typically established through manual segmentation by experts, often on a pixel/voxel level, sometimes validated by pathology or clinical outcomes. The document does not specify which method was used for the Relu Creator.

    8. The sample size for the training set

    Not provided in this document. The document states, "The software is not adaptive, it is trained at the manufacturer (Relu), and the weights are locked." This confirms that training occurred, but the size of the training dataset is not mentioned.

    9. How the ground truth for the training set was established

    Not provided in this document. Similar to item 7, the method for establishing ground truth for the training data is not detailed.