Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223575

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-06

    (280 days)

    Product Code
    Regulation Number
    886.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    General ophthalmic imaging including retinal, corneal, and external structures of the eye.

    Device Description

    The Phoenix ICON system is an updated cart based retinal imaging system covering the design changes to date on the predicate device, Phoenix ICON. The Phoenix ICON GO retinal imaging system is a portable version of the predicate device, Phoenix ICON (K170527) including the design changes in the Phoenix ICON system.

    Both the Phoenix ICON and Phoenix ICON GO are wide-field, handheld, high resolution, real-time retinal imaging devices. They are intended to be used for general ophthalmic imaging including retinal, corneal, and external structures of the eye. The intended users of the Phoenix ICON and Phoenix ICON GO are clinical imaging technicians, ophthalmic technicians, nurses, and physicians. The devices may be used in hospitals, medical clinics, and physician's offices.

    The Phoenix ICON platform consists of either a cart based (Phoenix ICON) or portable (Phoenix ICON GO) control box used in conjunction with a hand-held camera (Handpiece) using interchangeable LED based light sources (White and Blue light). The Phoenix ICON cart contains an AC mains power attachment, a battery module, a keyboard interface, a monitor, and a computer with Phoenix ICON software. The Phoenix ICON GO contains a portable control box with battery function and has an interface for attachment to a specified laptop computer which runs the Phoenix ICON software. Both systems may be used with a Foot Pedal, White Light Module (standard), Blue Light Module (FA) and/or Diffuser accessory.

    The Phoenix ICON Handpiece contains a wide-field, high resolution camera is used in three (3) modes, External Imaging (White Light), Retinal Imaging (White Light), and Fluorescein Angiography (Blue Light). For external imaging, the Diffuser accessory is placed over the lens tip to diffuse the light and provide for images of the outer surfaces of the eye. Both Retinal Imaging and Fluorescein Angiography are performed with the glass lens of the Handpiece coupled to the cornea via an imaging gel. In these imaging methods, LED light is emitted into the eye to illuminate the retina for image capture.

    Both the Phoenix ICON and Phoenix ICON GO are software-controlled systems which can capture either video or still images and store them on the control box (Cart computer or GO laptop) for later review. The Phoenix ICON system may be connected to IT networks under IT supervision.

    AI/ML Overview

    The provided document does not contain details about specific acceptance criteria for a device's performance in a clinical study or a study proving that the device meets those criteria. Instead, it is a 510(k) summary for the NeoLight Phoenix ICON and Phoenix ICON GO ophthalmic cameras, aimed at demonstrating substantial equivalence to a predicate device (K170527, Phoenix ICON by Phoenix Technology Group, LLC).

    The document focuses on comparing technological characteristics and safety testing, not on clinical performance acceptance criteria or a study to demonstrate such.

    However, based on the Performance Testing section (Table 5.3) related to Simulated Use, it states:
    Characteristic: Image Clarity - Comparison between subject and predicate images to ensure equivalent visual quality of the captured images.
    Results: Pass.

    While this indicates some form of performance assessment related to image quality, it does not provide the specific acceptance criteria (e.g., quantitative metrics, thresholds) or the detailed methodology of the study. It also doesn't present the "reported device performance" in a manner typical for clinical trials (e.g., sensitivity, specificity, or reader agreement scores).

    Therefore, a table of acceptance criteria and reported device performance, as well as several other requested details, cannot be fully extracted from the provided text.

    Here's an attempt to answer the questions based only on the available information, noting where information is missing:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Visual quality of captured images is equivalent to predicate devicePassed (Equivalent visual quality of captured images)

    Note: The document only provides a high-level "Pass" result for "Image Clarity - Comparison between subject and predicate images to ensure equivalent visual quality of the captured images." It does not specify quantitative acceptance criteria or detailed performance metrics.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "Simulated Use" testing for "Image Clarity - Comparison between subject and predicate images." However, it does not specify the sample size used for this comparison, nor does it provide any information on data provenance (e.g., country of origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    The document does not provide any information about the number or qualifications of experts used for establishing ground truth or evaluating image clarity.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The document does not specify any adjudication method used for the image clarity comparison.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document describes the device as an "Ophthalmic Camera" for "General ophthalmic imaging." It is a imaging acquisition device and does not include AI functionality. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers is not relevant to this submission, and no such study is mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This device is an ophthalmic camera. Its function is to capture images. It does not appear to incorporate an algorithm that independently processes or interprets images to provide a diagnosis or finding, nor does it claim AI capabilities. Therefore, a standalone algorithm performance study is not applicable and not mentioned. The "Simulated Use" test assesses the visual quality of the captured images, not the performance of an interpretive algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "Image Clarity" assessment, the implicit "ground truth" was a visual comparison to the predicate device's images to ensure equivalent visual quality. The document does not specify a separate, independent ground truth method like expert consensus on pathology, or outcomes data.

    8. The sample size for the training set

    The document concerns an ophthalmic camera, not an AI/ML algorithm requiring a training set. Therefore, this question is not applicable.

    9. How the ground truth for the training set was established

    As the device is an ophthalmic camera and not an AI/ML algorithm, there is no training set and therefore no ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170585

    Validate with FDA (Live)

    Device Name
    Skylife
    Manufacturer
    Date Cleared
    2017-10-27

    (241 days)

    Product Code
    Regulation Number
    880.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Skylife™ is intended for the treatment of neonatal unconjugated hyperbilirubinemia. It is designed to provide phototherapy treatment from underneath the baby. Skylife™ must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. Skylife™ can be used in a clinical setting or in the home.

    Device Description

    Skylife™ is a portable phototherapy light system that delivers a narrow band of high-intensity blue light via blue light emitting diodes (LEDs) to provide treatment for neonatal unconjugated hyperbilirubinemia. Skylife™ is designed to provide phototherapy treatment from underneath the baby. The system must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. Skylife™ utilizes blue LEDs to achieve intensities from 25 uW/cm²/nm to >55 uW/cm²/nm, emitting light in a narrow bandwidth between 430-475 nm. This light bandwidth corresponds to the spectral absorption of light by bilirubin. Skylife™ utilizes blue LEDs in this range to achieve peak intensities between 25 to 35, and over 55 uw/cm²/nm at low, high settings. Skylife™ at the Very High setting provides an average intensity of 56.3 uw/cm/nm over the treatment area with peak at 72.4 uw /cm-/nm. Treatment is intended to be applied until the bilirubin levels have dropped sufficiently that the child no longer suffers from jaundice. Phototherapy treatment of neonatal jaundice using Skylife™ can be applied at home or in a hospital. Skylife™ consists of the following components: Light bed and light modules, GelMat, disposable mattress cover (Cloud Swaddle), controller, and power supply.

    AI/ML Overview

    The NeoLight Skylife™ phototherapy unit for neonatal unconjugated hyperbilirubinemia underwent performance testing to demonstrate its substantial equivalence to predicate devices. The studies focused on electrical safety, electromagnetic compatibility, biocompatibility, and human factors usability.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Standard)Reported Device Performance
    Electrical SafetyIEC 60601-1: 2005, 3rd Edition (Medical Electrical Equipments for basic safety and essential performance)Device met the standard.
    IEC 60601-1-2: 2014, 4th Edition (Electromagnetic compatibility - Requirements and tests)Device met the standard.
    IEC 60601-2-50: 2009, 2nd Edition (Particular requirements for the safety of infant phototherapy equipment)Device met the standard.
    IEC 60601-1-11: 2015, 2nd Edition (Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment)Device met the standard.
    IEC 60601-1-8: 2006 (General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems)Device met the standard.
    BiocompatibilityISO 10993-1 (Surface Contact, Skin, Prolonged Contact (> 24 hours, < 30 days))Materials were found to be non-cytotoxic, non-sensitizers, and non-irritants.
    ISO 10993-5:2009 (Tests for in vitro cytotoxicity)Materials were found to be non-cytotoxic.
    ISO 10993-10:2010 (Tests for irritation and skin sensitization)Materials were found to be non-sensitizers and non-irritants.
    Human Factors UsabilityIEC 60601-1-6: 2010, 3rd Edition (General requirements for basic safety and essential performance Collateral standard: Usability)The study supported the instructions for successfully using the device as intended.
    IEC 62366:2007, 1st Edition (Medical devices -- Application of usability engineering to medical devices)The study supported the instructions for successfully using the device as intended and substantiated the acceptability of identified risks.
    Light Range Intensity (Device Specification)25-55 μW/cm²/nm (Low and High settings); >55 μW/cm²/nm (Very High setting, avg 56.3 μW/cm²/nm, max 72.4 μW/cm²/nm)This is a specification, not an acceptance criterion from a performance test against a standard in this section. However, the document states this range is comparable or higher than predicates and does not raise new questions of safety or effectiveness.
    Wavelength (Device Specification)430 – 475 nm (453 - 460 nm typical peak)This is a specification. Wavelength range and peak wavelengths overlap with predicate devices, thus not raising new safety/efficacy questions.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The provided document refers to various performance tests against international standards (IEC and ISO). These types of tests typically involve specific sample sizes prescribed by the standards for evaluating materials, electrical components, and hardware. However, the exact numerical sample sizes for each specific test (e.g., number of units tested for electrical safety, number of material samples for biocompatibility) are not explicitly stated in this summary.

    For the Human Factors Usability study:

    • Number of participants: 2 user groups
    • Data provenance: Not explicitly stated, but based on the nature of human factors usability, it would be prospective data collection (users interacting with the device).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For the technical performance tests (electrical safety, EMC, biocompatibility), the "ground truth" is established by adherence to the specified international standards. The evaluation is typically performed by trained engineers and technicians in accredited testing facilities. The document does not specify the number or qualifications of experts explicitly.

    For the Human Factors Usability study:

    • The document implies that the study was conducted by human factors experts. However, the number of experts used to establish "ground truth" (e.g., identifying usability issues or defining successful use) and their specific qualifications are not provided. The "ground truth" for usability is generally derived from user performance and expert observation against established usability principles.

    4. Adjudication Method for the Test Set

    The document does not describe specific adjudication methods for the performance tests in the context of clinical endpoints or expert review. For objective technical tests, the results either meet the standard's criteria or they don't, often without a need for "adjudication" in the sense of resolving inter-rater disagreement.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No, the provided document does not describe an MRMC comparative effectiveness study, nor does it mention AI in any context. This device is a phototherapy unit, not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, as mentioned, this device is a phototherapy unit and does not involve an algorithm or AI for standalone performance evaluation in that manner.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • For Electrical Safety, EMC, and Safety of Infant Phototherapy Equipment: The ground truth is established by the requirements and methodologies defined within the respective IEC standards. Compliance with these standards indicates safety and essential performance.
    • For Biocompatibility: The ground truth is established by the requirements and testing protocols defined within the ISO 10993 series of standards. Non-cytotoxicity, non-sensitization, and non-irritation are the "ground truth" criteria.
    • For Human Factors Usability: The ground truth is established by the requirements of IEC 60601-1-6 and IEC 62366, which focus on preventing use errors and ensuring that the device can be used safely and effectively as intended. User performance during the study, observed errors, and successful task completion against defined criteria form the basis of the ground truth.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a phototherapy unit and does not rely on a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1