Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K201196

    Validate with FDA (Live)

    Date Cleared
    2021-06-25

    (417 days)

    Product Code
    Regulation Number
    890.3800
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The Scooter (Models: FDB01) is an indoor/outdoor electric scooter that is intended to be used by individuals that are able to walk but suffer from mobility limitations. It has a base with metal alloy frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, an adjustable steering column, a tiller console, an electric motor, an electromagnetic brake, 2 rechargeable Lithium-Ion Battery with an off-board charger. The movement of the scooter is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter only can be operated on the flat road.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a Scooter (Model: FDB01). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than efficacy or clinical performance in terms of disease detection or diagnostic accuracy. Therefore, many of the typical acceptance criteria and study elements associated with AI/ML-driven diagnostics will not be present.

    Based on the information provided, here's a breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, the "acceptance criteria" are compliance with various international standards for wheelchairs and scooters, and the "reported device performance" is the statement that the device meets these standards. The device is not an AI/ML diagnostic and therefore does not have acceptance criteria in terms of sensitivity, specificity, or similar metrics.

    Acceptance Criteria (Applicable Standard)Reported Device Performance
    Biocompatibility
    ISO 10993-5: 2009 (Tests For In Vitro Cytotoxicity)All user directly contacting materials are compliant with ISO 10993-5 requirements.
    ISO 10993-10: 2010 (Tests For Irritation And Skin Sensitization)All user directly contacting materials are compliant with ISO 10993-10 requirements.
    Electrical / Electromagnetic Compatibility (EMC)
    ISO 7176-21:2009 (Electromagnetic compatibility)The EMC performance results meet the requirements of ISO 7176-21.
    Performance (Mechanical & Functional)
    ISO 7176-1: 2014 (Static stability)The static stability has been determined after testing according to ISO 7176-1, and test results meet its design specification.
    ISO 7176-2:2017 (Dynamic stability of Powered Wheelchairs)The dynamic stability has been determined after testing according to ISO 7176-2, and test results meet its design specification.
    ISO 7176-3: 2012 (Effectiveness of brakes)The effectiveness of brakes has been determined after testing according to ISO 7176-3, and test results meet its design specification.
    ISO 7176-4: 2008 (Energy consumption for theoretical distance range)The theoretical distance range has been determined after testing according to ISO 7176-4, and test results meet its design specification.
    ISO 7176-5: 2008 (Overall dimensions, mass and manoeuvring space)The dimensions and mass have been determined after testing according to ISO 7176-5.
    ISO 7176-6: 2018 (Maximum speed, acceleration and deceleration)The maximum speed, acceleration, and deceleration have been determined after testing according to ISO 7176-6.
    ISO 7176-7 (Measurement of seating and wheel dimensions)The seating and wheel dimensions have been determined after testing according to ISO 7176-7.
    ISO 7176-8:2014 (Static, impact and fatigue strengths)All test results meet the requirements in Clause 4 of ISO 7176-8.
    ISO 7176-9:2009 (Climatic tests for Powered Wheelchairs)The test results shown that the device could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9.
    ISO 7176-10:2008 (Obstacle-climbing ability)The obstacle-climbing ability of the device has been determined after testing according to ISO 7176-10.
    ISO 7176-11:2012 (Test dummies)The test dummies used in the testing of ISO 7176 series meet the requirements of ISO 7176-11.
    ISO 7176-13: 1989 (Coefficient of test surfaces)The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved.
    ISO 7176-14:2008 (Power and control systems)All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14.
    ISO 7176-15:1996 (Information disclosure, documentation and labeling)The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15.
    ISO 7176-16:2012 (Resistance to ignition of postural support devices)The performance of resistance to ignition met the requirements of ISO 7176-16.
    ISO 7176-25:2013 (Batteries and chargers for powered wheelchairs)The performance of batteries and charger of device met the requirements in Clause 5 and 6 of ISO 7176-25.
    Labeling
    Conforms to FDA Regulatory requirements for Label and LabelingConforms to FDA Regulatory.

    Study Information:

    The study described is a non-clinical test program designed to demonstrate that the Scooter (Model: FDB01) meets established performance and safety standards, and is substantially equivalent to predicate devices. There are no clinical studies mentioned in this submission.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated as "sample size" in the context of patients or cases, as this is a non-clinical device performance study. The testing refers to one Scooter (Model: FDB01) being subjected to various tests.
    • Data Provenance: The tests were conducted to verify compliance with international standards (ISO series). The manufacturer is Nanjing Jin Bai He Medical Apparatus Co., Ltd, located in China. The testing itself is non-clinical, implying laboratory or controlled environment testing of the device specimen(s).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Experts: Not applicable. For non-clinical device performance testing against established standards, "ground truth" is determined by whether the device's physical and functional properties meet the numerical or qualitative requirements of the standard, as measured by calibrated equipment and documented test protocols. This isn't based on expert judgment in the way a diagnostic imaging study would be.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. Device performance against ISO standards is typically assessed via objective measurements and adherence to specified test procedures, not through expert adjudication in the traditional sense of consensus reading or arbitration.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This is a non-clinical submission for a medical device (scooter), not an AI/ML diagnostic or imaging device used by human readers for interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable. This device is a motorized scooter, not a standalone AI algorithm. It is a physical device operated by a human user.

    7. The Type of Ground Truth Used:

    • Ground Truth: The "ground truth" for this device's performance is defined by the requirements and test methods outlined in the cited ISO 7176 series standards and ISO 10993 series standards. These standards specify how to measure various performance characteristics (e.g., static stability, brake effectiveness, dimensions, EMC) and the acceptable limits or conditions for those characteristics. The tests verify that the device's measured performance falls within these specified "ground truth" parameters.

    8. The Sample Size for the Training Set:

    • Training Set Sample Size: Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Training Set Ground Truth: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170787

    Validate with FDA (Live)

    Date Cleared
    2017-08-09

    (146 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

    Device Description

    The powered wheelchair DYW30A(D09) is driven by two DC motors, and is suitable for indoor uses. It is an individual transportation vehicle for disabled and elderly people that experience difficulties in walking. The powered wheelchair DYW30A(D09) comes as preassembled, designed to be lightweight, easily maneuvered, highly portable, and most importantly, safe and comfortable.

    It consists of two foldable armrests, a seat belt, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, two Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller.

    The device is powered by two 12 volt / 6 Ah Lithium-ion batteries with 11.2 miles travel range that can be recharged by an off-board battery charger that can be plugged into an AC outlet (110-220 V, 50-60 Hz), when the device is not in use.

    The user can activate the joystick to move in the direction of the joystick is actuated. When the user releases the joystick, the device slows to stop and the brakes are automatically re-engaged. The maximum weight capacity of DYW30A(D09) is 264 lbs and its maximum speed is 3.75 mph.

    AI/ML Overview

    The provided document is a 510(k) summary for a Powered Wheelchair (DYW30A(D09)). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing a detailed study of an AI/algorithm-based device and its acceptance criteria as requested in the prompt.

    Therefore, many of the specific questions regarding acceptance criteria and study methodologies for an AI/algorithm-based device (e.g., sample size for test set, data provenance, number of experts for ground truth, MRMC study, standalone performance, training set details) cannot be answered from the provided text.

    The document primarily addresses the safety and effectiveness of a physical medical device (a powered wheelchair) through performance testing against established ISO and IEC standards, and a comparison to a predicate device.

    However, I can extract the information relevant to the performance testing and comparison aspects that serve as the "acceptance criteria" for this specific medical device.

    Here's what can be extracted based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria for this medical device are based on compliance with a comprehensive set of international standards (ISO and IEC) for wheelchairs. The document reports that the device "passed" these tests, implying compliance with the requirements within those standards. The specific performance values are not always listed as "reported device performance" in a separate row but are implicitly confirmed by passing the tests and in the comparison table.

    Acceptance Criteria (Standard / Test)Reported Device Performance (Implied by "Passed" / Stated in Comparison)
    ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stabilityPassed
    ISO 7176-2:2001 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairsPassed
    ISO 7176-3:2013 Wheelchairs - Part 3: Determination of effectiveness of brakesPassed (Minimum braking distance: 1m, which is smaller/better than predicate's 1.5m)
    ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance rangePassed (Cruising Range: 11.2 miles (18 km), which is larger than predicate's 9.3 miles)
    ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering spacePassed (Dimensions: 37.4" x 22.6" x 36.2"; Weight: 51.8 lbs / 23.5 kg)
    ISO 7176-6:2001 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairsPassed (Max Speed Forwards: 3.75 mph (6 km/h); Max Speed Backward: 1.86 mph (3.0 km/h))
    ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensionsPassed (Front wheel: 7"x1.77"; Rear wheel: 12.5"x2.25")
    ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengthPassed
    ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairsPassed
    ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairsPassed (Maximum obstacle climbing: 1.36" (34.5 mm))
    ISO 7176-11:2012 Wheelchairs - Part 11: Test dummiesPassed
    ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfacesPassed
    ISO 7176-14:2008 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methodsPassed (Controller: Changzhou Billon Electronic Appliance Co., Ltd., WS-1)
    ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labelingPassed
    ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devicesPassed
    ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scootersPassed
    ISO 7176-25:2013 Wheelchairs - Batteries and chargers for powered wheelchairsPassed (Battery: Lithium-ion, 6 Ah x 24 VDC x 2 pcs; Charger: High Power Technology Inc., HP0060W(L2))
    IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and testsPassed
    IEC 662133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applicationsPassed
    Biocompatibility Testing (ISO 10993-5 and ISO 10993-10) for patient-contacting parts (Seat, Backrest, Safety belt, Joystick, Armrest)Conducted (implies passed for PVC Vinyl and PU Foam, with rationale based on common use and no adverse claims for thousands of units sold in market).
    Maximum Safe Operational Incline8 degrees (restricted from predicate's 12 degrees, deemed sufficient for indoor use)
    Max Loading264 lbs (compared to predicate's 250 lbs)
    Turning Radius32.5" (833 mm) (compared to predicate's 31.5", difference deemed minor inconvenience, not safety concern)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the performance tests. Typically, device testing standards involve a certain number of units or test cycles, but this is not detailed in the summary.
    • Data Provenance: The testing was conducted by or on behalf of Nanjing Jin Bai He Medical Apparatus Co., Ltd. in Nanjing, Jiangsu, China. The document is a summary of findings, not raw data. The testing is assumed to be prospective as it's to demonstrate compliance for a new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the device is a physical product being tested against engineering standards, not an AI/algorithm-based diagnostic tool requiring expert consensus for ground truth. The "ground truth" is established by the specifications and measurement methodologies defined in the listed ISO and IEC standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable for physical device performance testing against engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a powered wheelchair, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    The "ground truth" for this device comes from the established and recognized international standards (ISO and IEC) for wheelchair performance and safety. These standards define the test methods, acceptance limits, and measurement criteria.

    8. The sample size for the training set:

    Not applicable. This is not an AI/algorithm-based device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1