Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251408

    Validate with FDA (Live)

    Date Cleared
    2025-09-02

    (119 days)

    Product Code
    Regulation Number
    892.1171
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. OsteoSight™ provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report.

    Device Description

    OsteoSight is an AI-enabled software as a Medical Device (SaMD) that processes standard anteroposterior (AP) radiographs of the hip or pelvis in patients aged 50 years and older. The device operates via integration with a host integration platform that facilitates the secure transfer of DICOM images and associated metadata from clinical imaging systems.

    OsteoSight applies an AI-enabled algorithm pipeline to estimate bone mineral density (BMD) at the femoral neck. If an estimated BMD is below the pre-specified threshold, OsteoSight generates a report indicating that the patient is at risk of having low BMD. This report is delivered to the interpreting physician via the host integration platform. OsteoSight does not provide a diagnosis or treatment recommendation. If the estimated BMD is above or equal to the pre-specified threshold, no report is generated.

    The OsteoSight report is intended to inform physicians and support further clinical evaluation. Physicians may consider additional assessments, such as a comprehensive fracture risk assessment, in conjunction with other clinical information.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative thresholds in the document. However, the study aims to demonstrate that the device's sensitivity and specificity are sufficient to aid radiologists in identifying patients with possible low BMD. The reported device performance is as follows:

    MetricAcceptance Criteria (Implicit)Reported Device Performance (Intended Use Population)Reported Device Performance (Subset that produced a result)
    SensitivitySufficient to aid identification of patients with possible low BMD0.370 (95% CI 0.349-0.391)0.441 (95% CI 0.386-0.487)
    SpecificitySufficient to aid identification of patients with possible low BMD0.951 (95% CI 0.936-0.963)0.943 (95% CI 0.922-0.961)
    AUCNot applicable (for full Intended Use Population)n/a0.837 (95% CI 0.821-0.853)

    Notes on Acceptance Criteria: The document states, "All pre-specified acceptance criteria for the co-primary and secondary endpoints in the clinical studies were met." However, the document does not explicitly list these "pre-specified acceptance criteria" as quantitative targets for sensitivity, specificity, or AUC. The general implicit criterion is that the performance metrics (sensitivity, specificity, AUC) are deemed adequate to "aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): 3082 images
    • Data Provenance:
      • Country of Origin: United States (six independent cohorts from the Midwest, Northeast, South, and West).
      • Retrospective/Prospective: Retrospective analysis. The datasets included "previously acquired X-rays that included paired DXA T-scores."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set

    Not applicable, as expert adjudication was not used. The ground truth was established by DXA scans.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study focuses on the standalone performance of the AI device against a gold standard (DXA), not on how human readers improve with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The reported sensitivity, specificity, and AUC values are for the OsteoSight device's performance in identifying low BMD as determined by DXA, without human intervention in the diagnostic process beyond the device generating a report. The study measures the device's output (notification or no notification) against the ground truth.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Dual-Energy X-ray Absorptiometry (DXA)
    • Definition of Low BMD: Defined as a T-score of < -1.0 SD as measured at the femoral neck from a DXA scan acquired within 12 months of the X-ray.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It only describes the validation (test) set.

    9. How the Ground Truth for the Training Set was Established

    The document does not provide information on how the ground truth for the training set was established. It only discusses the ground truth for the validation (test) set, which was based on DXA scans.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1