Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K162371

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-15

    (113 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pulsaderm ACE-All Blue is indicated to treat mild to moderate acne on the face.

    Device Description

    The Pulsaderm ACE-All Blue is a device that uses Blue Light Emitting Diodes to provide LED Light therapy to the skin. Blue light is emitted for the treatment of mild to moderate acne.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for the Pulsaderm ACE-All Blue device, which is an LED light therapy system for treating mild to moderate acne. This type of document focuses on establishing substantial equivalence to existing legally marketed devices rather than presenting a standalone study with detailed clinical performance metrics against specific acceptance criteria.

    Therefore, the document does not contain the requested information about acceptance criteria and a study proving the device meets those criteria in the typical sense of a clinical trial demonstrating efficacy against a defined endpoint. Instead, it relies on demonstrating equivalence in technological characteristics and safety to predicate devices.

    However, I can extract the information that is present concerning performance testing and a lay-user study, and explicitly state what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Pulsaderm ACE-All Blue Technical Specifications
    Wavelength (nm)N/A (Compared to Predicates)Blue light 415 +/- 5 nm
    Dose Rate (mW cm2)N/A (Compared to Predicates)Blue light: 25 mW cm2
    Dose (J/cm2)N/A (Compared to Predicates)12 J/cm2
    Lay-User Study
    Lay users able to properly use device by reading instructions without assistanceNot explicitly stated as a numerical acceptance criterion, but implied goal is high success rate.98% of lay users were able to properly use the device by reading instructions in the user manual without any assistance.
    Lay users able to read and comprehend user manual and package labeling.Not explicitly stated, but implied goal is successful comprehension.The conclusion was that an average lay user can read and comprehend correctly the user manual and package labeling.
    Safety and Conformity
    Compliance with IEC 60601-1:2012YesConforms
    Compliance with IEC 60601-1-2:2014YesConforms
    Compliance with IEC 62471:2006YesConforms
    Biocompatibility testing to ISO 10993YesTested to ISO 10993
    Risk Management System conforming to ISO 14971:2007YesDesigned and developed under a Risk Management System conforming to ISO 14971:2007

    Missing Information: The document focuses on technical specifications comparison, safety standards compliance, and a lay-user study for usability, rather than clinical efficacy against a pre-defined acceptance criteria for acne treatment effectiveness.


    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample size for the "lay-user study". It only mentions "study data was collected".
    • Data Provenance: The lay-user study was described as being conducted "with Institutional Review Board (IRB) approval and oversight." No specific country of origin is mentioned, but the company is based in Fort Myers, FL, USA, suggesting a US-based study. The study was prospective in nature, as it involved observing users reading instructions and attempting to use the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • This information is not provided as the "ground truth" for the lay-user study was based on observed successful use of the device by the participants themselves, as per the instructions. There were no "experts" establishing a disease-related ground truth, as the study was about usability of instructions.

    4. Adjudication Method for the Test Set

    • This information is not provided. Given the nature of the lay-user study (observing proper usage based on instructions), an adjudication method in the traditional sense (e.g., 2+1 reads of medical images) would not be applicable. The assessment was likely direct observation of user interaction with instructions and the device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is irrelevant for this device, which is a standalone light therapy device, not an AI-assisted diagnostic or interpretation tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The device itself is a "standalone" light therapy device. The non-clinical testing included a "lay-user study" which assessed the ability of users to follow instructions to operate the device correctly. This is a form of standalone performance assessment in terms of usability and safety, but not clinical efficacy. No "algorithm only" performance was assessed as it's not an AI/software device.

    7. The Type of Ground Truth Used

    • For the lay-user study, the "ground truth" was defined by the ability of participants to correctly follow the instructions and use the device as intended. This is an operational or usability ground truth, not a clinical one (like pathology or outcome data).

    8. The Sample Size for the Training Set

    • This information is not applicable and not provided. The Pulsaderm ACE-All Blue is an LED light therapy device; it does not involve machine learning algorithms that require a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161941

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-19

    (96 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pulsaderm Acne Device is intended to emit energy in the red and blue region of the spectrum, specifically indicated to treat mild to moderate acne on the face.

    Device Description

    The Pulsaderm Acne Device is a device that uses Blue and Red Light Emitting (LED) Diodes to provide LED Light therapy to the skin. A combination of red light and blue light is emitted for the treatment of mild to moderate acne.

    AI/ML Overview

    The Pulsaderm Acne Device is compared against predicate devices to demonstrate substantial equivalence, rather than establishing specific acceptance criteria for performance metrics like accuracy, sensitivity, or specificity commonly seen in diagnostic devices. The acceptance criteria and supporting studies focus on safety, proper use by lay users, and observable clinical benefits relevant to its intended use for mild to moderate acne.

    Here's an analysis based on the provided document:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    SafetyCompliance with electrical safety standards IEC 60601-1, IEC 60601-1-2, and photobiological safety standard IEC 62471. Biocompatibility according to ISO 10993.The Pulsaderm Acne Device "has been tested and found in conformance with test standards: IEC 60601-1, IEC 60601-1-2, and IEC 62471. Biocompatibility testing to ISO 10993." The device was "designed and developed under a quality management system conforming to ISO 14971," indicating a structured approach to risk management.
    Proper Use by Lay UsersLay users should be able to properly self-select for the device using box labeling and correctly use the device by following instructions in the user manual without assistance."100% of lay users were able to properly self-select themselves using the box labeling and 98% of lay users were able to properly use the device by reading instructions in the user manual without any assistance." Conclusion: an "average lay user can read and comprehend correctly the user manual and package labeling."
    Clinical EffectivenessThe device should be able to significantly decrease the size and redness of facial acne lesions. This translates to a statistically significant reduction in lesion size and redness compared to baseline, specifically after 24 hours (for redness), 48 hours (for combined lesions), and one week (for lesion size, redness, and combined lesions)."The Pulsaderm acne device significantly decreased lesion size and redness when compared to baseline. Treated lesions significantly decreased in size after one week. Patients saw a significant reduction in redness after 24 hours, 48 hours and one week. Subjects had significant reduction in combined lesions (redness and size) after 48 hours and one week." The data described indicates that the device met these clinical effectiveness criteria by demonstrating statistically significant improvements in the measured parameters over the specified timeframes. While raw data or specific statistical values (e.g., p-values, effect sizes) are not provided in this summary, the qualitative statements affirm that the device met these endpoints.
    Substantial EquivalenceThe device should have the same intended use, be as safe and effective, and employ similar technological characteristics (energy type, wavelength, etc.) as legally marketed predicate devices, with any differences not affecting safety or efficacy. The comparison table directly addresses this by showing similarities in intended use, energy type, OTC status, and similar wavelengths/dose rates/treatment schedules.The document extensively argues for substantial equivalence by comparing the Pulsaderm Acne Device to three predicate devices (Tanda Mini Skincare System K124042, illuMask Acne Light Therapy Mask K123999, Omnilux Clear-U K081307). It states, "Pulsaderm Acne Device has the same intended use as the predicate devices... is as safe and (commensurate to number of diodes) effective as K124042... uses a combination of red and blue light as K123999 and K081307... has same treatment schedule as K124042." The comparison table supports that, despite some differences in dose rate and treatment schedule details, the overall technological characteristics are very similar. The conclusion is that "no significant differences exist between this device and the predicates."

    2. Sample Size Used for the Test Set and Data Provenance

    • Lay-user study (for proper usage): The document does not explicitly state the sample size for the lay-user study. It mentions the study was conducted to "determine if lay users could read the product labeling and then self-assess if the Pulsaderm acne device would be beneficial for them to use."
    • Clinical Study (for effectiveness): 26 male and female subjects
    • Data Provenance: Both studies were conducted with "Institutional Review Board (IRB) approval and oversight." This indicates a prospective study design. The country of origin of the data is not specified, but the submission is to the U.S. FDA, suggesting the studies were likely conducted in the U.S. or to international standards acceptable to the FDA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Lay-user study: No experts were explicitly mentioned for establishing ground truth regarding proper usage. The "ground truth" here was the observed ability of lay users to follow instructions and self-select.
    • Clinical Study: An "expert clinical grading of facial lesion redness" was performed.
      • Number of experts: Not specified.
      • Qualifications of experts: Only "expert clinical grading" is mentioned. Specific qualifications (e.g., dermatologist, years of experience) are not provided in this summary.

    4. Adjudication Method for the Test Set

    • Lay-user study: Not applicable. The assessment was based on direct observation of user behavior and self-assessment of selection criteria.
    • Clinical Study: Not explicitly stated. For "expert clinical grading," a single expert could have performed the grading, or multiple experts might have been involved without a specified adjudication method like 2+1 or 3+1. However, the summary does not provide these details.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a light therapy device, not an AI-powered diagnostic or assistive tool, so such a study would not be relevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • This question is not applicable as the Pulsaderm Acne Device is a physical light therapy device, not an algorithm or software. Its performance is inherent in its physical emission of light, with the subject ("human-in-the-loop") being the direct recipient of treatment.

    7. Type of Ground Truth Used

    • Lay-user study: The ground truth for proper use was established by observing user behavior against defined instructions and their self-assessment based on provided labeling.
    • Clinical Study:
      • Lesion size: Measured by an "electronic caliper," which can be considered an objective measurement providing clinical ground truth.
      • Redness: Established by "expert clinical grading," indicating expert consensus/assessment.
      • Combined lesions: A combination of objective measurements and expert assessment.
      • Consumer perception questionnaires: Used to gather subjective patient feedback, which is also a form of data collection in clinical assessment but distinct from objective or expert-derived ground truth.

    8. Sample Size for the Training Set

    • Not applicable. The Pulsaderm Acne Device is a hardware medical device and does not involve AI or machine learning algorithms that require a "training set" in the computational sense. The studies described are performance validation studies for the device itself.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable for the reasons stated in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-10

    (317 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nutra Face LIFT Model PE8050 device is intended for facial stimulation and indicated for over-the-counter cosmetic use.

    Device Description

    The Nutra Face Lift is a hand-held device intended to apply low level electrical impulses to strategic locations on the face. The Nutra Face Lift probes are designed for optimal contact with the face. The device continually alternates between the positive and negative probes and allows the user to adjust the settings from 0 to 400 microamps for personalized comfort level by pressing the up/down button. The Nutra Face Lift requires the use of a conductive derma gel.

    The Nutra Face Lift unit contains a power supply and internal Ni-MH rechargeable battery. The enclosure is made of medical grade biocompatibility plastics and the output contacts (Probes) consist of chrome-plated spheres. The on/off button on the unit directly allows the user to immediately remove all power from the unit.

    AI/ML Overview

    The provided document, a 510(k) summary for the "Nutra Face LIFT Model PE8050", primarily focuses on establishing substantial equivalence to a predicate device and does not contain detailed information about the acceptance criteria for a specific performance study in the way a clinical trial might, nor does it present the study for proving those criteria.

    However, it does describe a "Usability and Product Self Selection Study" and states that "All four goals of the Study were met." This implies that the 'acceptance criteria' for this specific study were the successful achievement of these stated goals. The document also mentions nonclinical performance data related to compliance with FDA performance standards and electrical safety.

    Here's an analysis based on the provided text, addressing your questions where possible:


    Description of Acceptance Criteria and Study

    The primary study mentioned, the "Usability and Product Self Selection Study," had four goals that served as its acceptance criteria. The document states that all four goals were met, thereby proving the device met these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Study Goals)Reported Device Performance
    1. To attract participants that represented the "intended users" of the device.Met
    2. To determine if consumers could correctly self-select using the Packaging labeling only.Met
    3. To test consumer knowledge of the Packaging labeling and Instruction Manual.Met
    4. To have consumers demonstrate their ability to adhere to what they had read in the Instruction Manual and actually operate and care for the device correctly.Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 80 participants
    • Data Provenance: Not explicitly stated, but the submission is to the US FDA, suggesting the study was likely conducted in the United States. The study is referred to as a "Usability and Product Self Selection Study," indicating a prospective design where data was collected specifically for this evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts being used to establish a 'ground truth' in the context of this Usability and Product Self Selection Study. The study focused on consumer interaction, self-selection, and knowledge, not on expert medical evaluation of outcomes.

    4. Adjudication Method for the Test Set

    Not applicable for this type of self-selection and usability study. The criteria were based on participants' actions and knowledge, which would likely have been directly observed or assessed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The device is a "Transcutaneous electrical nerve stimulator for pain relief" (classified as such, but indicated for "facial stimulation and is indicated for-over-the counter cosmetic use"), not an AI-powered diagnostic tool requiring human reader interpretation of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the loop performance) was done

    This is not applicable, as the device is a physical electrostimulation device for cosmetic use, not an algorithm. Its performance is evaluated through its physical and electrical characteristics, and user interaction.

    7. The Type of Ground Truth Used

    For the "Usability and Product Self Selection Study," the "ground truth" was essentially the observed behavior and stated understanding of the participants regarding the device's packaging and instruction manual, and their ability to operate the device correctly. This is a form of behavioral or self-reported data against pre-defined success metrics for usability and comprehension.

    The document also mentions compliance with various engineering and electrical safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993) for which the "ground truth" would be the successful adherence to the requirements of these published technical standards, verified through objective laboratory testing.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150370

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-22

    (159 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nutra Light Advanced is intended/indicated for over the counter use for the treatment of mild to moderate acne and for the treatment of periorbital wrinkles and rythides.

    Device Description

    Nutra Light Advanced is a visible light and/or heat source with high spectral purity. The device is made of Lustran ABS 348 with clear cover lenses covering the LED light sources. The device has two exchangeable heads; with one head emitting narrow band blue light at 410mm for the treatment of mild to moderate acne vulgaris, the other head emits red light at 650mm for the treatment of periorbital wrinkles and rhytides. The device consists of a basic-hand-held, battery operated control unit and two optional (Blue and Red LED) interchangeable heads, which allows the operator to select either the blue or red operation. The STOP button directly on the unit allows the user to immediately remove all power to the unit. The device does not allow the operator to select the red and blue light simultaneously.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nutra Light Advanced device, focusing on demonstrating substantial equivalence to predicate devices rather than presenting a detailed study with specific acceptance criteria and performance metrics of the device itself.

    Therefore, many of the requested points cannot be answered from the provided document as it does not describe a clinical study in the way typically required for a novel device proving its efficacy against acceptance criteria. Instead, it relies on demonstrating similar technical characteristics to already approved devices.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The document states that the device has been tested and found in compliance with performance standards (e.g., IEC 60601-1, EMC 60601-1-2, IEC 60601-2-57, ISO 10993-5, ISO 10993-10) but does not list specific acceptance criteria (e.g., a certain percentage reduction in acne lesions) or the device's measured performance against such criteria. The basis for approval is substantial equivalence to predicate devices, implying their performance is accepted.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. As no clinical study is described, there's no test set in the context of device efficacy. The "tests" mentioned are nonclinical performance data related to electrical safety, EMC, and biocompatibility, not clinical outcomes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. Ground truth would be relevant for a clinical study assessing efficacy, which is not described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. Adjudication methods are typically used in clinical trials to resolve discrepancies in expert assessments, which is not applicable here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. The device is an LED light device for acne and wrinkles, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided. This question is not applicable to a physical light therapy device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided. As no clinical efficacy study is detailed, no ground truth related to clinical outcomes is discussed. The "ground truth" for the nonclinical performance data would be the defined standards themselves (e.g., passing a specific electrical safety test).

    8. The sample size for the training set

    This information is not provided. Training sets are relevant for machine learning algorithms, which is not what this device is.

    9. How the ground truth for the training set was established

    This information is not provided. Not applicable for this device type.


    Summary of available information regarding acceptance criteria and "study" (nonclinical data):

    The "study" presented is a series of nonclinical performance tests to ensure the device meets safety and engineering standards and demonstrates substantial equivalence based on technical specifications rather than a clinical efficacy trial.

    Acceptance Criteria CategorySpecific Criteria (from referenced standards)Reported Device Performance
    Electrical SafetyIEC 60601-1:2005; ANSI/AAMI ES60601-1: 2005 + CORR. 1 (2009) + A2 (2010); CAN/CSA C22.2 No.60601-1-08 incorporates Corrigendum 2; June 2011: General requirements for basic safety and essential performance of medical electrical equipment.Tested and found in compliance.
    Electromagnetic Compatibility (EMC)EMC 60601-1-2:2014 Medical Electrical Equipment- Part 1-2: General Requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances-Requirements and tests.Tested and found in conformity.
    Specific Light Source SafetyIEC 60601-2-57, Medical electrical equipment-Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.Tested and found in compliance.
    Biocompatibility (In Vitro Cytotoxicity)ISO 10993-5, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity, 1999.Tested and found in compliance.
    Biocompatibility (Irritation & Hypersensitivity)ISO 10993-10. Biological Evaluation of medical devices-Part 10: Tests for irritation and delayed-type hypersensitivity, 1996.Tested and found in compliance.
    Substantial Equivalence (Primary "Criteria")The device's technical specifications and intended use are similar enough to predicate devices such that it raises no new questions of safety and effectiveness. This includes matching wavelengths (410nm and 650nm), similar treatment duration, and similar treatment regimen to the predicate devices (Nutra Light Blue K132838 and Nutra Light Red K141308). The only difference (interchangeable heads) must not raise new safety/effectiveness concerns.Wavelengths: 410nm (blue light) and 650nm (red light) - Matches predicates. Treatment Duration: Same as predicates. Treatment Regimen: Similar to predicates. Interchangeable Heads: This feature was deemed not to raise new issues of safety and effectiveness. Conclusion: Substantially equivalent to predicate devices.

    In essence, the "acceptance criteria" for K150370 were primarily met by demonstrating compliance with recognized safety and performance standards for medical electrical equipment and establishing substantial equivalence to previously approved devices based on similar technology and intended use, rather than a de novo clinical efficacy study with defined clinical endpoints.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141588

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-01-08

    (209 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nutra Stim Hair Laser Comb is indicated to treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, LL-1, LL-2, or frontal and Fitzpatrick Skin Types L to IV.

    Device Description

    The Nutra Stim Hair Laser Comb is a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp.

    The Nutra Stim Hair Laser Comb unit contains the power supply, internal NI-MH rechargeable battery and the control unit. The enclosure and laser lens are made of medical grade biocompatibility plastics via injection molding. The STOP button directly on the unit allows the user to immediately remove all power to the unit. The Nutra Stim Hair Laser Comb does not use any software.

    AI/ML Overview

    The provided text describes the Nutra Stim Hair Laser Comb, a low-level laser device intended for promoting hair growth and treating androgenetic alopecia in specific female populations. However, the document does not contain a study that establishes acceptance criteria for device performance in terms of clinical efficacy (hair growth), nor does it provide performance data against such criteria.

    The document focuses on demonstrating substantial equivalence to a predicate device (HairMax Pro 12 LaserComb, K112524) for regulatory clearance, rather than presenting a clinical trial for performance.

    Here's an analysis based on the information provided, highlighting what is present and what is missing:


    1. A table of acceptance criteria and the reported device performance

    This information is not present in the document. The document discusses regulatory and safety testing but does not provide clinical performance acceptance criteria (e.g., minimum percentage increase in hair count, or specific visual improvement scores) or reported device performance against such criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size (Clinical): Not applicable for clinical efficacy. The document mentions an "OTC Usability Study" with "80 participants." This study was focused on usability and consumer understanding, not clinical performance for hair growth.
    • Data Provenance: Not specified for any clinical performance data, as none is presented. For the OTC Usability Study, the provenance is not mentioned.
    • Study Type: The OTC Usability Study was likely prospective, as it involved attracting participants to test the device's labeling and operation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not present. The usability study did not involve establishing ground truth for efficacy using experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not present. No such adjudication method is mentioned, as there was no clinical performance data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device is a physical laser comb, not an AI-assisted diagnostic or interpretation tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance Study: No, the device is not an algorithm. What was done were non-clinical performance tests for electrical safety, EMC, and biocompatibility, as listed in Section 8. These are tests of the device's physical and electrical properties, not its clinical standalone efficacy for hair growth.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not present for clinical efficacy. For the OTC Usability Study, the "ground truth" was whether participants could correctly self-select, understand labeling, assemble, and operate the device. This "ground truth" would have been established by observing participant actions and responses against predefined correct answers or procedures.

    8. The sample size for the training set

    This information is not present. The document does not describe the development of an algorithm or model that would require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no mention of a training set or an algorithm being developed.


    Summary of Device and Regulatory Context (based on the provided text):

    • Device Name: Nutra Stim Hair Laser Comb Model NSL-6318
    • Intended Use/Indications: Treats androgenetic alopecia and promotes hair growth in females with Ludwig (Savin) Scale I-4, LL-1, LL-2, or frontal, and Fitzpatrick Skin Types I to IV.
    • Regulatory Class: Class II
    • Product Code: OAP (Lamp, nonheating, for promotion of hair growth)
    • Regulation Number: 21 CFR 890.5500
    • Predicate Device: HairMax Pro 12 LaserComb (K112524)
    • Basis for Clearance: Substantial Equivalence to the predicate device.
    • Supportive Studies Mentioned:
      • OTC Usability Study: 80 participants, focused on self-selection, understanding labeling/manual, assembly, and operation. All four goals of this study were reportedly met. This study did not assess clinical efficacy.
      • Nonclinical Performance Data: Risk analysis, verification, and validation tests (specifics not detailed in the summary, but referenced as being in Section 10). Compliance with electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11) and laser safety (IEC 60825-1) standards. Biocompatibility testing per ISO 10993.

    In conclusion, the provided document is an FDA 510(k) clearance letter and its accompanying summary, which primarily establishes substantial equivalence for marketing purposes. It highlights compliance with safety and usability aspects but does not present a clinical study with acceptance criteria and performance data related to the device's efficacy in promoting hair growth. The efficacy is implicitly accepted based on the substantial equivalence to the predicate device, which would have had its own supporting data for clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141596

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-01-08

    (209 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nutra Stim Hair Laser Comb is indicated to treat androgenetic alopecia, promote hair growth and prevent further hair loss in males who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and Fitzpatrick Skin Types I to IV.

    Device Description

    The Nutra Stim Hair Laser Comb is a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The Nutra Stim Hair Laser Comb unit contains the power supply, internal NI-MH rechargeable battery and the control unit. The enclosure and laser lens are made of medical grade biocompatibility plastics via injection molding. The STOP button directly on the unit allows the user to immediately remove all power to the unit. The Nutra Stim Hair Laser Comb does not use any software.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Nutra Stim Hair Laser Comb. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed acceptance criteria and performance data for a new device's efficacy.

    Therefore, many of your requested details, particularly those related to a clinical study establishing performance against specific acceptance criteria for a novel device, are not directly available in this document. The document primarily relies on the predicate device's established effectiveness and safety, and direct comparison of technological characteristics.

    However, I can extract the information that is present:

    1. Table of Acceptance Criteria and Reported Device Performance

    This specific table, comparing acceptance criteria to performance, is not provided in the document. The document describes an "OTC Usability Study" and lists its goals, which could be interpreted as acceptance criteria for usability, but it does not present quantifiable performance metrics against those goals.

    Acceptance Criteria (Implied from Usability Study Goals)Reported Device Performance
    Attract participants representing "intended users" of the deviceMet (80 participants)
    Consumers could correctly self-select using packaging labeling onlyMet
    Consumer knowledge of packaging labeling and user manual, and ability to assemble, operate, and care for the device correctlyMet

    2. Sample Size for Test Set and Data Provenance

    The document mentions an "OTC Usability Study" with:

    • Sample Size: 80 participants
    • Data Provenance: Not explicitly stated, but it's an "OTC Usability Study," implying it was conducted with potential end-users for over-the-counter use, likely within the US, where the FDA is regulating it. It is a prospective study as participants were actively recruited to test the device's usability.

    3. Number of Experts and Qualifications for Ground Truth

    • This information is not provided as the usability study did not involve establishing a medical ground truth in the way a diagnostic device study would. The "ground truth" for the usability study was whether participants could successfully interact with the device and understand its labeling.

    4. Adjudication Method for Test Set

    • This information is not provided. Given it's a usability study where success was measured against understanding and operating the device, a formal adjudication process as seen in clinical trials for diagnostic accuracy is unlikely to have been implemented or necessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices where human readers interpret results, often with and without AI assistance. The Nutra Stim Hair Laser Comb is a therapeutic device for hair growth, not a diagnostic one.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone performance study (algorithm only) was not conducted. The device is a physical laser comb, not an AI algorithm. The document mentions "The Nutra Stim Hair Laser Comb does not use any software," further confirming the irrelevance of an algorithm-only study.

    7. Type of Ground Truth Used

    • For the "OTC Usability Study," the "ground truth" was essentially observational success/failure in performing tasks (self-selection, understanding instructions, operating the device) and participant feedback. It was not based on expert consensus, pathology, or outcomes data in a medical sense for efficacy.

    8. Sample Size for Training Set

    • This information is not applicable and therefore not provided. The Nutra Stim Hair Laser Comb is a hardware device for therapeutic use; it does not involve AI or machine learning algorithms that require a "training set."

    9. How Ground Truth for Training Set Was Established

    • This information is not applicable and therefore not provided for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141308

    Validate with FDA (Live)

    Device Name
    NUTRA LIGHT RED
    Manufacturer
    Date Cleared
    2014-08-20

    (93 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nutra Light Red is a non-invasive LED light device is intended/indicated for over- the -counter use for the treatment of periorbital wrinkles, and rhytides

    Device Description

    The NutraLuxe Red, Model NL-2212 is a device that utilizes Light Emitting Diodes to provide LED light to the body. The NutraLuxe Red is a visible light source of high spectral purity. They provide uniform or "hot-spot" free illumination. The output is pre-tuned to one wavelength with a narrow spectral bandwidth. The output wavelength of Red is 650 +/- 5 at 80 mW. The NutraLuxe Red unit contains the power supply, internal NI-MH rechargeable battery and the control unit. The enclosure and LED lens are made of medical grade biocompatibility plastics via injection molding. It has the capabilities to deliver the light to the skin as the unit is moved over the STOP button directly on the unit allows the user to immediately remove all power to the unit. The Nutra Light Red does not use any software.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on acceptance criteria and study details for the Nutra Light Red device:

    Important Note: The provided text does not describe acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity, accuracy) for an AI/algorithm-based medical device. Instead, it details an OTC Usability Study and compliance with electrical, EMC, photobiological safety standards, and biocompatibility for a light-emitting diode (LED) device. This is a crucial distinction. The questions in your prompt are more aligned with the evaluation of algorithmic diagnostics, which this device is not.

    However, I will extract the information as best as possible to align with your request, interpreting "acceptance criteria" here as the goals of the usability study and "reported device performance" as the outcome of that study.


    Acceptance Criteria and Study Details for Nutra Light Red (Based on provided text)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Study Goals)Reported Device Performance (Study Outcome)
    To attract participants that represented the "intended users" of the deviceMet
    To determine if consumers could correctly self-select using the packaging labeling onlyMet
    To test consumer knowledge of the packaging labeling and user manualMet
    To accurately assemble, operate, and care for the device correctlyMet
    Compliance with Electrical Safety (IEC 60601-1:2012)Complies
    Compliance with EMC (IEC 60601-1-2)Complies
    Compliance with Photobiological Safety (IEC 62471)Complies
    Compliance with Basic Safety and Performance of Non-Laser Light (IEC 60601-2-57)Complies
    Biocompatibility (ISO 10993)Demonstrated compliance

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 80 participants
    • Data Provenance: Not explicitly stated, but the study was an "OTC Usability Study," implying it was conducted with representative consumers. No country of origin is specified. It is a prospective study as it involved participants performing tasks with the device.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable for this device and study. This device is an LED light therapy device, and the study was a usability study, not a diagnostic accuracy study requiring expert-established ground truth for a test set. The "ground truth" here was the successful completion of tasks by consumers.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no ambiguity in "ground truth" (i.e., whether tasks were performed correctly). The study aimed to see if participants could correctly perform actions and understand labeling, not to make a diagnostic determination that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This was not an AI/algorithm-based device, nor was it a diagnostic tool where human readers would be evaluated with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • No. The device does not contain an "algorithm" in the sense of an AI model that would have standalone performance. It is a hardware device. The "performance data" presented relates to its physical and usability characteristics. The document explicitly states: "The Nutra Light Red does not use any software."

    7. Type of Ground Truth Used

    • For the usability study: The ground truth was the observed behavior and understanding of the 80 participants when interacting with the device, packaging, and user manual. Whether they could correctly perform actions (assemble, operate, care for) and understand labeling constituted the "ground truth" of usability.
    • For safety and compliance: The ground truth was established international standards and regulatory requirements (e.g., IEC, ISO).

    8. Sample Size for the Training Set

    • Not applicable. This device does not use an AI or machine learning algorithm, therefore there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set, there is no ground truth for it.

    Summary of Device and Study Type:

    The Nutra Light Red is a non-invasive LED light device for treating periorbital wrinkles and rhytides, intended for over-the-counter use. The "study" described in the 510(k) summary is primarily an OTC Usability Study to ensure consumers can understand and safely operate the device without professional assistance. Other "performance data" relates to compliance with electrical, electromagnetic compatibility (EMC), photobiological safety, and biocompatibility standards, which are standard requirements for medical devices. This document does not describe the evaluation of an AI or algorithmic diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132838

    Validate with FDA (Live)

    Device Name
    NUTRA LIGHT BLUE
    Manufacturer
    Date Cleared
    2014-05-23

    (255 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nutra Light Blue is a non-invasive LED light device is intended/indicated for over- the-counter use. The device delivers low intensity light emitting diodes (LED's) in the blue region at a wavelength of 410 +/- 5 nanometers, for the treatment of mild to moderate inflammatory acne.

    Device Description

    The Nutra Light Blue Model: NL-2213, is deviccs that utilize Light Emitting Diodes to provide LED light to the body. The Nutra Light Blue is a visible light source of high spectral purity. They provide uniform or "hot-spot" free illumination. The output is pre-tuned to one wavelength with a narrow spectral bandwidth. The output wavelength of Blue is 410 +/- 5 at 75 mW. The Nutra Light Blue unit contains the power supplies, internal NI-MH rechargeable battery and the control unit. The enclosure and LED lens are made of medical grade biocompatility plastics via injection molding. It has the capabilities to deliver the light to the skin as the unit is moved over the skin surface. The STOP button directly on the unit allows the user to immediately remove all power to the unit. The Nutra Light Blue does not use any software.

    AI/ML Overview

    The provided text describes the Nutra Light Blue device, which is an LED light device intended for the treatment of mild to moderate inflammatory acne. It is important to note that this document is a 510(k) Summary of Safety and Effectiveness, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical study report with acceptance criteria and comprehensive performance metrics.

    Based on the provided information, I can extract the following:

    1. Table of acceptance criteria and the reported device performance:

    The document does not explicitly state numerical acceptance criteria for clinical performance (e.g., a specific percentage reduction in acne lesions). Instead, it focuses on demonstrating that an "OTC Usability Study" met its goals. The "reported device performance" in this context refers to the outcomes of that usability study.

    Acceptance Criteria (Goals of OTC Usability Study)Reported Device Performance
    To attract participants that represented the "intended users" of the device.All four goals of the study were met.
    To determine if consumers could correctly self-select using the packaging labeling only.All four goals of the study were met.
    To test consumer knowledge of the packaging labeling and user manual and actually assembly, operate and care for the device correctly.All four goals of the study were met.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 80 participants.
    • Data Provenance: The study was an "OTC Usability Study," implying it was conducted for the purpose of this submission and likely prospective, focusing on how consumers interact with the device. The country of origin is not explicitly stated, but given the FDA submission, it's typically understood to be within the US or compliant with US standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The study was a "Usability Study" and not a clinical efficacy trial with expert-assessed outcomes like lesion counts. Therefore, the concept of "ground truth" established by experts in the context of acne treatment efficacy (e.g., dermatologists assessing lesion counts) does not appear to apply to this specific usability study.

    4. Adjudication method for the test set:

    This information is not provided. As it was a usability study, it's unlikely to have involved the kind of clinical adjudication typically seen in efficacy trials.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. The Nutra Light Blue device is a non-AI, LED light therapy device for acne treatment. The document does not mention any AI component or human-in-the-loop performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided. The Nutra Light Blue device does not use an algorithm for diagnosis or treatment decision-making; it delivers light therapy directly. The document explicitly states: "The Nutra Light Blue does not use any software."

    7. The type of ground truth used:

    For the usability study, the "ground truth" would be the observed user behavior and responses to survey questions regarding packaging understanding, assembly, operation, and care. It's not a clinical ground truth like pathology or physician diagnosis. The document states the goals were to determine if consumers could "correctly self-select" and "correctly assembly, operate and care for the device." The "all four goals... were met" indicates the usability performance was deemed successful.

    8. The sample size for the training set:

    This information is not applicable and not provided. The Nutra Light Blue device is a hardware device (LED light therapy) and does not involve AI or machine learning models that require a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1