Search Results
Found 6 results
510(k) Data Aggregation
(225 days)
Product Name: nSet+ Stabilization Set
Part Number: 6426733
The nSet+ Stabilization Set is for single use only. The nSet+ Extension Set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. The nSet+ Extension Set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10 ml per second.
Product Name: nSet+ Stabilization Set with nSyte Needle Free Connector (NFC)
Part Number: 6426731
The nSet+ Stabilization Set with nSyte needle free connector is for single use only. The nSet+ Extension Set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. The nSet+ Stabilization Set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10 ml per second.
Product Name: nSyte Needle Free Connector (NFC)
Part Number: 6426727
The nSyte needle free connector is for single use only. The nSyte needle free connector may be used for direct injection, intermittent infusion, continuous infusion or aspiration. The nSyte needle free connector may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10 ml per second.
Product Name: nSet+ Stabilization Set
Part Number: 6426733
Description: The nSet+ Stabilizing Set is a sterile, single patient use, add-on medical device used in peripheral vascular access systems. The nSet+ Stabilizing Set includes a stabilization base, an extension tube, a tube pinch clamp, and terminates with locking ISO 80369-7 connectors on each end, capped. It is sold individually blister packed with fifty (50) total units per carton and is labeled with a shelf life of one year.
Product Name: nSet+ Stabilization Set with nSyte Needle Free Connector (NFC)
Part Number: 6426731
Description: The nSet+ Stabilizing Set with nSyte Needle Free Connector is a sterile, single patient use, add-on medical device used in peripheral vascular access systems. The nSet+ Stabilizing Set includes a stabilization base, an extension tube, a tube pinch clamp, an nSyte Needle Free Connector, and terminates with a locking ISO 80369-7 male connector, capped. It is sold individually blister packed with fifty (50) total units per carton and is labeled with a shelf life of one year.
Product Name: nSyte Needle Free Connector (NFC)
Part Number: 6426727
Description: The nSyte Needle Free Connector is a sterile, single patient use, add-on medical device used in peripheral vascular access systems.
N/A
Ask a specific question about this device
(442 days)
The nPulse™ neutral displacement connector is a single patient use, sterile, nonpyrogenic device for needleless access to the IV line and/or IV catheter during IV therapy. The nPulse™ connector can be used for direct injection, intermittent infusion, continuous infusion, or aspiration.
The nPulse K150 Neutral Displacement Needle Free Connector is a single patient use, sterile, non-pyrogenic device for needleless access to the IV line and/or IV catheter during IV therapy. The nPulse K150 Neutral Displacement Needle Free Connector can be used for direct injection, intermittent infusion, continuous infusion, or aspiration. The nPulse K150 valve is a normally closed, luer-activated, valved connector which has an open mode that permits bi-directional fluid flow and a closed mode that prevents fluid flow.
The provided text describes a medical device, the NP Medical nPulse™ K150 Neutral Displacement Needle Free Connector, and its substantial equivalence to a predicate device (NP Medical K100 Neutral Displacement Needle Free Connector) for FDA 510(k) clearance.
The document highlights performance testing conducted to support this equivalence. Here's extraction of relevant information regarding acceptance criteria and study details:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria alongside performance data in a direct row-by-row comparison. Instead, it lists the performance tests conducted and generally states that the device "has been evaluated using the appropriate methodology as specified in the FDA recognized consensus standards and in conjunction with NP Medical established test protocols" and that differences from the predicate "demonstrated applicable performance" or "aligns with current best practices."
However, specific performance claims and how they relate to the predicate and established standards are mentioned for certain differences:
| Performance Characteristic | Acceptance Criteria (Implied / Achieved) | Reported Device Performance (Subject Device - NP Medical nPulse K150) |
|---|---|---|
| Multiple Activations | Applicable performance following 200 activations based on Gland Height and Integrity, Back Pressure, and Activated Pressure Resistance. | Demonstrated applicable performance following 200 activations. |
| Use Days | Microbial ingress performance supporting a 7-day change protocol with 0 CFU recovered. | Aligning with current best practices for 7-day use; 0 CFU recovered for all test device replicates. |
| Shelf Life | Performance after 5 years of aging, including microbial ingress testing with 0 CFU recovered. | Demonstrated performance after 5 years of aging; 0 CFU recovered for all test device replicates. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes used for each of the performance tests (e.g., number of devices tested for Air Bolus, Repeat Insertion, etc.). It mentions "all test device replicates" for microbial ingress testing but doesn't quantify this.
The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated in the provided text. It implies internal testing by NP Medical, Inc., located in Clinton, MA, USA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable to this type of device and study. The testing for the NP Medical nPulse™ K150 connector involves technical performance benchmarks (e.g., flow rates, pressure resistance, sterility) rather than clinical evaluations requiring expert interpretation of diagnostic images or patient outcomes.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable. Adjudication methods are typically used in clinical studies, particularly when interpreting subjective data or reaching consensus on diagnoses, which is not the nature of the bench testing described here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. MRMC studies and AI assistance are relevant to diagnostic tools or image analysis, not to a needle-free connector's performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. This device is a physical medical device, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance tests conducted on the NP Medical nPulse™ K150 connector and its predicate device is established by specified functional requirements and consensus standards. For example:
- Microbial ingress testing: The ground truth is the absence of microbial growth (0 CFU recovered), demonstrating sterility barrier integrity.
- Pressure ratings, flow rates, etc.: The ground truth is compliance with pre-defined engineering specifications and relevant FDA recognized consensus standards (e.g., ISO standards) for intravascular administration sets.
8. The sample size for the training set
This information is not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable. As in point 8, this device does not utilize machine learning or AI requiring a training set with established ground truth.
Ask a specific question about this device
(162 days)
The NP Medical Extension Set with stabilizing base is a sterile single patient use device.
The NP Medical Extension Set is intended for direction, intermittent infusion, continuous infusion or aspiration for all patients for which IV therapy is prescribed.
The NP Medical Extension Set incorporates a normally closed bidirectional pressure-activated valve that prevents low pressure retrograde flow and is intended for power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.
The NP Medical Extension Set with Stabilizing Base is a sterile, single use, nonpyrogenic intravenous administration extension set comprised of a side repositionable tube bonded to a connector hub on one end and a female luer on the other end with a clamp in between. The connector hub has a locking male luer and a rotating tube connection joint. The hub houses a bidirectional Pressure-Activated Valve (PAV) and an integral stabilizer under the connector hub. The stabilizing base component (along with IV tape / adhesive) is intended to help anchor the extension set to the patient. This device is not made with plasticizers Diethylhexylphthalate (DEHP). The NP Medical Extension Set with Stabilizing Base may be used with compatible Vascular Access Devices (VAD), Needle Free Connectors, and other commercially available devices.
Based on the provided text, the device in question is the "NP Medical nCompass Extension Set with Stabilizing Base," which is an intravascular administration set. The document is a 510(k) summary for premarket notification to the FDA, demonstrating substantial equivalence to a predicate device.
The acceptance criteria and study proving the device meets these criteria are primarily based on bench performance testing and biocompatibility testing, as no clinical data was required or provided.
Here's a breakdown of the requested information:
1. Table of acceptance criteria and the reported device performance:
The document doesn't provide specific numerical acceptance criteria for each test with their corresponding reported performance values. Instead, it states that the device "met all requirements" and "satisfy the acceptance criteria specified by the above applicable standards."
However, we can infer some criteria and the general reported performance:
| Performance Test / Acceptance Criteria Category | Reported Device Performance (as stated in the document) | Inferable Acceptance Criteria (based on predicate comparison and industry standards) |
|---|---|---|
| Device Flow Rate (gravity) | "Equal to or greater than 3 L/hr." | ≥ 3 L/hr. (matching predicate) |
| Device Flow Rate (pressurized) | "Equal to or greater than 10ml/second from the device distal end at a pressure at or below 325psi." | ≥ 10 ml/second at ≤ 325 psi (matching predicate) |
| Retrograde Pressure Resistance | Met requirements | Prevent low pressure retrograde flow (as per Indications for Use) |
| Dynamic Hemolysis | Met requirements | Demonstrate substantial equivalence (per USP <788>) |
| Positive Pressure | Met requirements | Withstand positive pressure without failure |
| 325psi Burst Resistance (rotating tube connection and other joints) | Met requirements | No burst at 325 psi |
| Extension tube bond integrity | Met requirements | Maintain integrity/no leaks |
| Stabilization pad bond integrity | Met requirements | Maintain integrity/proper adhesion |
| Extension tubing port (rotating connection) interfaces, sub-atm. pressure air leakage | Met requirements | No significant air leakage |
| Resistance to overriding | Met requirements | Function correctly when components are connected/operated |
| ESC resistance to lipids | Met requirements | Maintain integrity/functionality when exposed to lipids |
| PAV Robustness | Met requirements | Maintain functionality over repeated use |
| Particulate Contamination (sub-visible particles) | Met requirements | Conform to USP <788> for particulate limits |
| Biocompatibility (Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility, Chemical Characterization, Allowable Limits Establishment) | "All the Biocompatibility study results satisfy the acceptance criteria specified by the above applicable standards and proposed device has been found biocompatible for its intended use/indications for use." | Conformance with ISO 10993-1 and specific endpoints. |
| Sterilization (SAL) | "Sterility Assurance Level (SAL) of 10-6" | Achieve an SAL of 10-6 (as per ISO 11135-2014) |
| Sterile Barrier Packaging | Met requirements | Conform to ISO 11607-1:2019 for maintaining sterility |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: The document does not specify the sample size for any of the performance tests. It states that tests were conducted using "appropriate methodology" and "NP Medical established test protocols."
- Data Provenance: The data is generated from bench testing conducted by NP Medical. The country of origin for the testing itself is not explicitly stated, but NP Medical is based in Clinton, Massachusetts, USA. The studies appear to be prospective in nature, as they were conducted specifically for the purpose of demonstrating substantial equivalence for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This device, being a physical medical device (extension set), does not involve diagnostic interpretations or algorithms that would require expert human review for ground truth establishment in the traditional sense of AI/imaging studies.
- The "ground truth" for its performance is established through adherence to recognized consensus standards (e.g., ISO, USP) and validated bench testing protocols. The experts involved would be technical professionals (engineers, quality assurance personnel, lab technicians) performing the tests and interpreting results against the objective criteria of these standards, rather than clinical experts establishing a "truth" from patient data. Their qualifications would be in metrology, engineering, and quality systems for medical devices. The document does not specify the number or specific qualifications of these individuals.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or AI performance evaluations where multiple human readers interpret data, and a consensus or tie-breaking mechanism is needed for ground truth. For bench testing of physical device performance against objective criteria, such methods are not used. The results are quantitative measurements against predefined specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC comparative effectiveness study is not applicable as this is a physical medical device (extension set), not an AI algorithm for diagnosis or interpretation. The document explicitly states: "VIII. CLINICAL DATA: Not applicable."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This relates to AI/software as a medical device. This product is a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the performance of this device is established by:
- Objective measurements against physical and chemical standards (e.g., flow rate, burst pressure, particulate count, material composition).
- Adherence to recognized consensus standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, ISO 11607-1 for packaging) and FDA guidance documents.
- Comparison to the performance characteristics of the legally marketed predicate device.
8. The sample size for the training set:
- Not applicable. This device is a physical product and does not involve machine learning or AI, and therefore no "training set" in the context of data for algorithm development.
9. How the ground truth for the training set was established:
- Not applicable. See point 8.
Ask a specific question about this device
(96 days)
The nPro surgical masks are intended for operating room personnel health care workers to protect both patients and healthcare workers from the transfer of microorganisms, blood and particulate materials. nPro Surgical Masks are single use, disposable devices provided non-sterile.
The nPro surgical mask meets ASTM F2100-19 Level 1 performance requirements for bacterial filtration efficiency, differential pressure, sub-micron particulate filtration efficiency, resistance to penetration by synthetic blood and flame spread.
The three-ply pleated mask fabric can be vertically adjusted to cover the user's mouth and nose. The inner and outer layers are constructed of non-woven fabric. The middle layer is constructed of melt-blown negative charge non-woven fabric and includes a flexible nose wire to the user's nose and reduce air gaps. The three layers of the mask body are collated and ultrasonically welded around the edges to enclose the filter media and nose wire. The polyester-spandex ear loops are ultrasonically welded to the mask body. All materials are not made with natural rubber latex. The nPro Surgical Masks are single use, disposable devices that are provided non-sterile in a dispensing carton.
This document is a 510(k) premarket notification for a surgical mask, not an AI-powered medical device. Therefore, the information typically requested regarding AI/ML device performance (like MRMC studies, training set details, or multiple expert ground truth establishment) is not applicable here.
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the non-clinical performance testing conducted on the surgical mask to demonstrate its substantial equivalence to a predicate device.
Here's the breakdown of the relevant information provided:
1. A table of acceptance criteria and the reported device performance
| Performance Requirements (units) | Acceptance Criteria | Reported Device Performance (Results) |
|---|---|---|
| Bacterial filtration efficiency (%) | ≥95% | Pass, 99.1% |
| Differential pressure (H2O / cm²) | <5.0 mm H2O / cm² | Pass, 4.5 mm H2O / cm² |
| Sub-Micron particulate filtration efficiency (%) | ≥95% | Pass, 99.4% |
| Resistance to penetration by synthetic blood (mm Hg) | ≥80 mm Hg (Accept on 3, reject on 4 per lot, AQL=4%) | Pass (32/32, 30/32, 32/32 passed for 3 lots) |
| Flammability Class | Class 1 Rating | Pass, Class 1 Rating |
| Structural Integrity Failure | Ear loops must not detach when exposed to a 5N load when tested individually (Accept on 0, reject on 1) | Pass, Greater than 5N |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size:
- Bacterial filtration efficiency, Differential pressure, Sub-Micron particulate filtration efficiency, Resistance to penetration by synthetic blood, Flammability Class: 32 each from 3 lots
- Structural Integrity Failure: 23 each from 3 lots
- Data Provenance: Not specified in terms of country of origin or retrospective/prospective. The testing appears to be standard laboratory/engineering testing for product performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as this pertains to a surgical mask and its physical performance characteristics, not an AI/ML device requiring expert interpretation of medical images or data. The "ground truth" here is established by the standardized test methods (e.g., ASTM F2101-19, EN 14683:2019, ASTM F2299-03, ASTM F1862-17, 16 CFR Part 1610, NP-TM-2019-0188-1) and their associated measurement protocols.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods are relevant for human interpretation of data, typically in clinical studies or AI ground truth establishment. For physical product testing, results are typically objective measurements against defined acceptance criteria.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. MRMC studies are for evaluating AI-assisted medical imaging devices impacting human reader performance. This document is about a surgical mask's physical performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. This refers to AI algorithm performance. The "standalone" performance here is the mask's ability to perform against physical and biological challenges.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the surgical mask's performance is established by standardized laboratory test methods defined by organizations like ASTM, EN, and CFR. These methods provide quantifiable metrics for properties like filtration efficiency, differential pressure, and resistance to synthetic blood penetration. For example, a 99.1% bacterial filtration efficiency is a direct measurement result from the ASTM F2101-19 test method, serving as its own "ground truth" against the acceptance criterion.
8. The sample size for the training set
This is not applicable. This device is a physical product (surgical mask), not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for this device.
Ask a specific question about this device
(90 days)
The K100 Neutral Displacement Connector is a single patient use, sterile, non-pyrogenic device for needleless access to the IV line and/or IV catheter during IV therapy. The K100 connector can be used for direct injection, intermittent infusion, continuous infusion, or aspiration.
The K100 Neutral Displacement Needleless Connector is a sterile, single patient use, swabable, normally closed, luer-activated, valved connector. The K100 valve is comprised of a: a housing (inlet and outlet), a fluid channel, a translating member (or lower gland) secured to a longitudinal wall of the housing and bounding a variable volume region of the fluid channel, a post having a lumen terminating at a distal head of the post and biased in the proximal direction by the lower gland when in the closed mode, and an upper seal having a sealing ring and secured to the longitudinal wall of the housing, the sealing ring contacting the distal head of the post when in the closed mode to prevent flow through the fluid channel, Moving components within the assembly are lubricated with silicone oil.
The provided text is a 510(k) summary for the K100 Neutral Displacement Needle Free Connector. It outlines the device's description, intended use, comparison to a predicate device, and performance testing. However, it does not contain acceptance criteria for specific performance metrics nor detailed study results that prove the device meets these criteria in a quantitative manner.
Instead, it states that "All test results meet the acceptance criteria, and support that the devices are appropriately designed for their intended use." without providing the criteria themselves or the specific data showing compliance.
Therefore, I cannot populate the table with acceptance criteria and reported device performance, nor can I provide specific details on sample sizes, ground truth establishment, or clinical study types (like MRMC or standalone) because this information is not present in the provided document.
Here's an overview of what is available in the document related to performance claims:
1. Table of acceptance criteria and the reported device performance:
This information is not explicitly provided in a quantifiable manner. The document lists the tests performed and states that "All test results meet the acceptance criteria." without detailing what those criteria are or the specific results.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not specified | Not specified beyond "Pass" |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not specified for any of the performance tests.
- Data Provenance: The tests were "conducted by NP Medical," indicating internal testing. The country of origin and whether the data is retrospective or prospective are not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not specified. These are bench tests, not involving human interpretation or subjective assessments typically requiring expert consensus as ground truth.
4. Adjudication method for the test set:
- Not applicable/Not specified. This pertains to expert review of ambiguous cases, which is not relevant for bench testing of medical devices like this connector.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study is for evaluating AI systems in diagnostic imaging or similar fields. This document describes a medical device (needle-free connector) and its mechanical/physical performance, not an AI system.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No. This question is relevant for AI algorithms. The document describes a physical medical device. The "standalone" performance here refers to the device's functional integrity as a component.
7. The type of ground truth used:
- For the bench tests listed (e.g., Air Bolus, Flow Rate, Hydraulic Burst Leak), the "ground truth" would be established by physical measurement against a predefined engineering specification or standard. The document states that "All test results meet the acceptance criteria," implying such specifications were used.
8. The sample size for the training set:
- Not applicable. This device is not an AI algorithm that requires a "training set."
9. How the ground truth for the training set was established:
- Not applicable. This device is not an AI algorithm that requires a "training set."
Summary of available performance claims:
The document lists the following bench tests that were conducted, and for each, it explicitly states that "All tests met acceptance criteria."
- Air Bolus and Bubble Free Priming
- Repeat Insertion
- Blood Flushing Evaluation
- Bolus Back Pressure
- Docking Stability
- Flow Rate
- Fluid Displacement
- Gland Height Determination
- Hydraulic Burst Leak
- ISO 594 Test Methods
- Lipid Resistance
- Priming Volume
- Residual Volume
- Stress Resistance to Swabbing Chemicals (IPA and CHG).
- Torque Testing
- Microbial Ingress testing
- Sterilization Validation
- Packaging/Shelf Life Validation
The comparison to the predicate device also highlights some performance aspects:
- Multiple Activations: K100 (96 intermittent) vs. Predicate (96 intermittent)
- Indwell: K100 (96 hours) vs. Predicate (96 hours)
- Pressure Rating: K100 (325 psi) vs. Predicate (> 60 psi)
- Priming Volume: K100 (0.12 mL) vs. Predicate (0.25 mL)
- Microbial Ingress Test: K100 (Pass) vs. Predicate (Pass)
The document concludes that "The K100 Neutral Displacement Needle Free Connector is substantially equivalent to the identified predicate system based on the indication for use, design features, operating principles, performance tests and material composition."
Ask a specific question about this device
(146 days)
The Capless Luer Activated Valve, incorporating a luer activated valve, is intended for use in facilitating needleless fluid delivery and may be swabbed with antiseptic just prior to use, thereby eliminating the need for capping between uses.
The Capless Luer Activated Valve is a two-part device consisting of a Gland Housing Assembly and a Gland/Center Post Assembly, as described below:
The Gland is a valve/gasket which provides a seal against the syringe/luer connector when the device is being utilized. The Gland incorporates a slit to accept the syringe/luer connector. The Gland is also the swabbable surface of the Capless Luer Activated Valve. The valve can be easily swabbed per hospital protocol before each connection.
The Center Post mechanically supports the Gland and serves as the primary high pressure seal to keep the fluid path closed during the resting state.
The Gland and the Center Post are mechanically press-fit together to form the Gland/Center Post Assembly.
In the resting state, the Center Post is flush with the walls of the Gland Housing, ensuring that there is no fluid path. As the device is activated by a syringe/male luer connector, the flexible Gland is forced open and the Center Post is pushed down. As the Center Post is forced further down into the Gland Housing, the fluid path is established.
The provided text describes a 510(k) submission for a medical device, the "NP Medical Capless Luer Activated Valve." However, it does not contain the detailed information necessary to fully answer all aspects of your request, particularly regarding specific numerical acceptance criteria and the results of a study that quantifies device performance against those criteria. The submission states that performance requirements were met but does not provide the specific metrics or values.
Here's a breakdown of the available information based on your requested points:
1. A table of acceptance criteria and the reported device performance
The document mentions that mechanical, biocompatibility, and microbial challenge testing were performed and that the device "meets their performance requirements" or "passed all" tests. However, it does not specify the exact acceptance criteria or the quantitative results achieved. Without specific numerical values for acceptance criteria and device performance from the document, a table cannot be constructed with the requested level of detail.
Illustrative Table (Based on available qualitative information):
| Acceptance Criterion (Hypothetical - Not explicitly stated in document) | Reported Device Performance (As stated in document) |
|---|---|
| Mechanical Integrity: Maintain structural integrity under anticipated forces (e.g., pressure, luer connection cycles). | "The results of the Mechanical Testing demonstrate that the NP Medical Capless Luer Activated Valves meet their performance requirements." |
| Biocompatibility: Absence of adverse biological reactions (e.g., cytotoxicity, sensitization, irritation). | "Full biocompatibility testing has been performed ... The materials passed all of the biocompatibility tests." |
| Microbial Barrier/Sterility Maintenance: Prevent microbial ingress into the fluid pathway under specified challenge conditions. | "The results of the testing demonstrate that, in response to excessive microbial challenge conditions, the sterility of the fluid pathway was maintained." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes used for any of the tests (Mechanical, Biocompatibility, Microbial Challenge). It also does not explicitly state the provenance of the data (country of origin, retrospective or prospective). Given that NP Medical, Inc. is located in Clinton, MA, and the submission is to the FDA, it is highly probable the testing was conducted in the USA by or for the manufacturer. The testing described (mechanical, biocompatibility, microbial challenge) is typically prospective and laboratory-based.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable and therefore not available in the provided document. The device is a physical medical device (a luer activated valve), not an AI or diagnostic imaging device that requires expert review for ground truth establishment. The performance is assessed through laboratory testing (mechanical, chemical, microbial) rather than interpretation by human experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and therefore not available in the provided document. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human expert assessment where there might be disagreement in interpretations (e.g., in radiology studies). This submission describes laboratory-based engineering and biological tests for a physical device, not an interpretive process.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and therefore not available in the provided document. An MRMC study is relevant for evaluating diagnostic tools, often AI-powered, where human readers interpret patient cases. The Capless Luer Activated Valve is a physical device for fluid delivery, not a diagnostic tool, and involves no "human readers" in its intended function or performance testing described.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable and therefore not available in the provided document. "Standalone performance" in this context usually refers to the performance of an AI algorithm alone. The NP Medical Capless Luer Activated Valve is a physical medical device, not an algorithm. Its performance is inherent in its physical and material properties.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the mechanical testing, the "ground truth" would be defined by engineering specifications and established test methods (e.g., pressure resistance, flow rates, number of activation cycles without failure). For biocompatibility, the "ground truth" is determined by standardized biological assays according to ISO 10993. For microbial challenge testing, the "ground truth" is the absence of microbial growth in the fluid pathway after being subjected to a known microbial load, following established microbiological testing protocols.
8. The sample size for the training set
This information is not applicable and therefore not available in the provided document. "Training set" refers to data used to train machine learning models. As this is a physical medical device, there is no "training set" in the context of AI. The device's design and manufacturing rely on engineering principles and material science, not machine learning.
9. How the ground truth for the training set was established
This information is not applicable and therefore not available in the provided document. As there is no AI training set for this physical device, there is no ground truth to be established for it.
Ask a specific question about this device
Page 1 of 1