Search Results
Found 10 results
510(k) Data Aggregation
(170 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Sterile, powder free, examination gloves made of nitrile. The gloves are provided in sizes small, medium, and large. Sterile gloves are provided in dispenser boxes.
This looks像 a request for structured information extracted from the provided text about the K122727 device. Unfortunately, the provided text describes a Disposable Nitrile Examination Glove and not an AI/ML medical device.
Therefore, many of the requested fields, such as "Sample size used for the test set," "Number of experts used to establish the ground truth," "Adjudication method," "MRMC comparative effectiveness study," "Standalone performance," "Training set sample size," and "Ground truth for the training set," are not applicable to this type of device. These fields are specifically relevant to studies involving diagnostic or AI-powered devices that produce results requiring interpretation or comparison to a ground truth established by experts.
However, I can still extract the acceptance criteria and performance for the physical device characteristics from the provided text.
Here's the information that is applicable and can be extracted:
Acceptance Criteria and Device Performance (for a Nitrile Examination Glove)
1. A table of acceptance criteria and the reported device performance
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM Standard D6319-10 | Meets |
| Physical Properties | ASTM Standard D6319-10 | Meets |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151-06 | Meets |
| Powder Residual | ASTM Standard D6124-06 | Meets (values below 2mg of residual powder) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10:2010) | Gloves are non-irritating |
| Biocompatibility | Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | Gloves do not display any potential for sensitization |
Regarding the other requested points, they are not applicable to this medical device type:
- 2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for a physical glove's performance testing described here. Performance is measured against physical standards, not, for example, diagnostic data.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties is established by the ASTM and ISO standards themselves, not by expert consensus on specific cases.
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Performance is measured against objective standards.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical glove, not an AI diagnostic tool.
- 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is a physical glove.
- 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For physical properties, the "ground truth" is defined by the objective metrics and thresholds within the cited ASTM and ISO standards. For biocompatibility, it's the observed biological reactions in animal models as per ISO 10993-10:2010.
- 8. The sample size for the training set: Not applicable. There is no "training set" for a physical glove.
- 9. How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(128 days)
The examination glove is a disposuble device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Textured, White, Latex, Sterile Powder Free Examination Gloves with Protein labeling Claim (50µg/dm² or Less of Water Soluble Protein)
Here's an analysis based on the provided text, outlining the acceptance criteria and the study details:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard) | Device Performance |
|---|---|---|
| Dimension | ASTM Standard D3578 | Meets |
| Physical Properties | ASTM Standard D3578 | Meets |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151 | Meets |
| Powder Residual | ASTM Standard D6124 | Meets (values below 2mg of residual powder) |
| Protein Level | ASTM Standard D5712 | Meets (values below 50 mcg/gm) |
| Biocompatibility (Primary Skin Irritation) | ISO 10993-10 | Passes |
| Biocompatibility (Dermal Sensitization) | ISO 10993-10 | Gloves do not display any potential for sensitization |
| Sterilization | Radiation dose of 25 kGy (implied as "Same" to predicate) | Radiation dose of 25 kGy (from predicate) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for testing each characteristic. However, each standard (ASTM, ISO, 21 CFR) defines specific methodologies and sample sizes for their respective tests. For example, ASTM D5151 (Freedom from Pinholes) would specify a certain number of gloves to be tested.
The data provenance is not explicitly mentioned, but given the context of a 510(k) submission to the FDA, it is expected that the testing was conducted prospectively according to the listed standards. No information on country of origin for the data is provided, nor is it indicated if it's retrospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This information is not applicable (N/A) to this type of device (examination gloves). The "ground truth" for glove performance is established by objective measurements against harmonized standards (e.g., ASTM, ISO), not by expert consensus or interpretation of images/data. The tests are industrial and chemical, not interpretive.
4. Adjudication Method for the Test Set
This information is N/A for this device. Adjudication methods like 2+1 or 3+1 are typically used in studies where human interpretation of data (e.g., medical images) is involved and discrepancies need to be resolved. For objective measurements of physical, chemical, and biological properties, results are either within spec or not.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
This information is N/A for this device. MRMC studies are used to evaluate the impact of an AI algorithm on human reader performance, typically in diagnostic imaging. The device in question is a physical glove, not an AI system.
6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This information is N/A for this device. A "standalone study" typically refers to the performance of an AI algorithm independent of human interaction. The device is a physical product, not an algorithm.
7. The Type of Ground Truth Used
The ground truth for this device's performance is established by objective measurements against recognized national and international standards. These standards define the acceptable range or threshold for each characteristic (e.g., minimum elongation, maximum protein level, freedom from pinholes). There is no "expert consensus," "pathology," or "outcomes data" in the traditional sense for these performance criteria.
8. The Sample Size for the Training Set
This information is N/A for this device. The concept of a "training set" (used to train machine learning models) is not relevant to the evaluation of physical medical devices like examination gloves. Development of such gloves involves manufacturing processes and quality control, not AI model training.
9. How the Ground Truth for the Training Set Was Established
This information is N/A for this device, for the same reasons as point 8.
Ask a specific question about this device
(49 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free, examination gloves made of latex and colored blue. The gloves are provided in sizes small, medium, large and extra large.
The provided document describes the acceptance criteria and study results for the "Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50µg/dm2 Or Less of Water Soluble Protein)" in the context of an Abbreviated 510(k) Premarket Notification.
Here's the breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM D3578-05 | Meets |
| Physical Properties | ASTM D3578-05 | Meets |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151-06 | Meets |
| Powder Residual | ASTM D6124-06 | Meets (values below 2mg of residual powder) |
| Protein Level | ASTM D5712-10 | Meets (values below 50 mcg/g of protein) |
| Biocompatibility - Primary Skin Irritation | ISO 10993-10:2010 (rabbits) | Gloves are non-irritating |
| Biocompatibility - Dermal Sensitization | ISO 10993-10:2010 (guinea pig) | Gloves do not display any potential for sensitization |
2. Sample size used for the test set and the data provenance:
The document does not specify the exact sample sizes for each test listed (e.g., number of gloves tested for dimensions, pinholes, etc.). It only states that the device "Meets" the respective ASTM and ISO standards. These standards would inherently define the required sample sizes for testing, but the document does not explicitly report them.
Data provenance is not explicitly mentioned (e.g., retrospective or prospective, country of origin). However, given these are non-clinical (laboratory/material) tests conducted against established international standards (ASTM, ISO), they are typically conducted in a controlled testing environment, likely in the manufacturer's facility or a certified testing laboratory.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by whether they meet objective, quantifiable physical and chemical standards (ASTM, ISO, and 21 CFR). There are no human "experts" or clinical interpretation involved in establishing the performance criteria for these tests. The tests are designed to objectively measure properties like dimensions, strength, and chemical content.
4. Adjudication method for the test set:
This information is not applicable. Since the tests are objective measurements against defined standards, there is no need for an adjudication method. The results either meet the standard or they do not.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. This device is a medical glove, not an AI-assisted diagnostic or imaging tool. Therefore, MRMC studies and AI assistance are irrelevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This information is not applicable. This device is a medical glove, not an algorithm.
7. The type of ground truth used:
The ground truth used is based on established objective laboratory standards (ASTM and ISO standards) and regulatory requirements (21 CFR 800.20). These standards define specific thresholds and methodologies for evaluating the physical, chemical, and biocompatibility properties of examination gloves.
8. The sample size for the training set:
This information is not applicable as this is not a machine learning or AI-based device that would require a training set.
9. How the ground truth for the training set was established:
This information is not applicable as there is no training set for this type of device.
Ask a specific question about this device
(40 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Nitrile Powdered patient examination glove
The provided text is a 510(k) Premarket Notification for Nitrile Powdered Examination Gloves. It details the device's characteristics, intended use, and a comparison to a predicate device, focusing on nonclinical data.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM Standard D6319-10 | Meets |
| Physical Properties | ASTM Standard D6319-10 | Meets (Elongation 500%/500%, Tensile Strength 14MPa/14MPa for before/after aging) |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151-06 | Meets |
| Powder Residual | ASTM Standard D6124-06 | Meets (Results generated values below 10 mg/dm² of powder) |
| Biocompatibility: | ||
| Primary Skin Irritation | ISO 10993-10:2010 (in rabbits) | Gloves are non-irritating |
| Dermal Sensitization | ISO 10993-10:2010 (in guinea pig) | Gloves do not display any potential for sensitization |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes for each test set that led to the "Meets" determination. It references ASTM and ISO standards for testing, which would outline the appropriate sample sizes and methodologies for each specific test (e.g., how many gloves were tested for pinholes, how many animals for biocompatibility).
The data provenance is from nonclinical testing conducted by the manufacturer (Northstar Healthcare Holdings) or a designated testing facility on their behalf, designed to demonstrate compliance with established standards. No specific country of origin or retrospective/prospective nature of the data is mentioned beyond the standards themselves.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the device is a patient examination glove. The "ground truth" for this type of device is established by compliance with standardized physical performance and biocompatibility tests, not by expert interpretation of images or clinical data. The standards (ASTM, ISO) themselves define the criteria for "ground truth" or acceptable performance.
4. Adjudication method for the test set
This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers interpret data, often with conflicting opinions, requiring a consensus mechanism. For the nonclinical testing of gloves, the results are quantitative measurements against predefined thresholds specified in the ASTM and ISO standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This device is a physical medical device (examination glove), not an AI diagnostic or assistance tool. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the device is not an algorithm or an AI system. Its performance is evaluated through physical and biocompatibility testing.
7. The type of ground truth used
The ground truth for this device is based on established industry standards (ASTM and ISO), which define objective, measurable criteria for physical properties (dimensions, tensile strength, elongation, freedom from pinholes, powder residual) and biological safety (biocompatibility). The tests measure these characteristics directly against the specified thresholds.
8. The sample size for the training set
This information is not applicable. This device is not an AI/ML device that requires a training set. Its performance is assessed through direct testing against predetermined criteria.
9. How the ground truth for the training set was established
This information is not applicable for the same reason as point 8.
Ask a specific question about this device
(80 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation of Chemotherapy Drugs.
Textured, Blue, Latex Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs With Protein Labeling Claim (50µg/dm² Or Less Of Water Soluble Protein).
This document describes the acceptance criteria and performance of "Textured, Blue, Latex Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs With Protein Labeling Claim (50µg/dm² Or Less Of Water Soluble Protein)".
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard / Acceptance Criteria | Device Performance |
|---|---|---|
| Physical Properties | ||
| Dimension | ASTM Standard D3578-05 | Meets |
| Physical Properties | ASTM Standard D3578-05 | Meets |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151-06 | Meets |
| Powder Residual | ASTM Standard D6124-06 (Maximum of 2mg/glove) | Meets (results generated values below 2 mg of residual powder) |
| Protein Level | ASTM Standard D5712-10 (Maximum of 50 mcg/dm² of water soluble protein) | Meets (results generated values below 50 mcg/mg of protein) |
| Biocompatibility | ||
| Biocompatibility | Biological Evaluation of Medical Devices part 1: Evaluation and Testing (ISO 10993-1:2009) | Meets |
| Primary Skin Irritation | Primary Skin Irritation in rabbits (ISO 10993-10:2010) | Gloves are non-irritating |
| Dermal Sensitization | Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | Gloves do not display any potential for sensitization |
| Chemotherapy Permeation | ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation of Chemotherapy Drugs (Minimum Breakthrough Detection Time) | |
| Carmustine (BCNU) (3.3 mg/mL) | Minimum breakthrough detection time is specified by the standard for each drug. The device must meet or exceed this time. For specific claims, the measured time becomes the claim. | Average Breakthrough Detection Time: 15.4 minutes (Note: "extremely low permeation time of less than 30 minutes" is stated, indicating that 15.4 mins is the reported time which must be acceptable based on the standard and drug's known permeability) |
| Cisplatin (1.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Cyclophosphamide (Cytoxan) (20.0 g/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Dacarbazine (DTIC) (10.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Doxorubicin Hydrochloride (2.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Etoposide (20.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Fluorouracil (Adrucil) (50.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Methotrexate (25.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Mitomycin C (0.5mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Paclitaxel (Taxol) (6.0mg/mL) | No breakthrough up to 240 min (>240 minutes) | |
| Thiotepa (10.0 mg/mL) | Average Breakthrough Detection Time: 1.6 minutes (Note: "extremely low permeation time of less than 30 minutes" is stated, indicating that 1.6 mins is the reported time which must be acceptable based on the standard and drug's known permeability) | |
| Vincristine Sulfate (1.0 mg/mL) | No breakthrough up to 240 min (>240 minutes) |
2. Sample Size Used for the Test Set and Data Provenance
The provided document does not explicitly state the specific sample sizes used for each individual test (e.g., number of gloves tested for pinholes, physical properties, or chemotherapy permeation). However, it implicitly suggests that testing was conducted in accordance with the cited ASTM and ISO standards, which would define appropriate sample sizes for each evaluation.
The data provenance is not specified in terms of country of origin. The studies appear to be prospective laboratory tests conducted according to established international and US standards (ASTM, ISO, 21 CFR).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
For the type of device and tests conducted (physical, chemical, and biocompatibility), the "ground truth" is established by adherence to specified, objective, and standardized testing methodologies. This typically involves trained laboratory technicians and/or engineers performing the tests, rather than a panel of "experts" as seen in clinical diagnostic studies. The qualifications would be expertise in conducting the specific ASTM and ISO standard tests. The document does not specify the number or specific qualifications of these individuals.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 typically apply to studies where human interpretation of data (e.g., medical images) is involved and discrepancies need to be resolved. For objective laboratory tests like those performed on gloves, the results are quantitative and directly compared against the pass/fail criteria of the standard.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical data, often with and without AI assistance, to assess the impact of AI on diagnostic performance. This device is a medical glove, and its performance is evaluated through standardized physical, chemical, and biocompatibility tests, not human interpretation of medical cases.
6. If a Standalone Study Was Done
Yes, the studies described are standalone (algorithm/device only) in the sense that the performance of the glove itself was evaluated directly against the specified standards, without involving human users "in the loop" for the primary performance evaluations described. The performance metrics are objective measurements (e.g., tensile strength, pinhole count, breakthrough time, protein levels).
7. The Type of Ground Truth Used
The ground truth used for these studies is based on:
- Established Standards and Regulatory Requirements: The performance is measured against the criteria defined by ASTM standards (D3578-05, D5151-06, D6124-06, D5712-10, D6978-05), ISO standards (10993-1:2009, 10993-10:2010), and U.S. FDA regulations (21 CFR 800.20).
- Objective Laboratory Measurements: Data is generated through physical, chemical, and biological testing protocols designed to yield quantitative and qualitative results that are then compared to the acceptance criteria specified in the standards.
8. The Sample Size for the Training Set
Not applicable. This device is a physical product (medical glove), not an AI algorithm requiring a training set. The performance is assessed through direct testing of the manufactured product.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(69 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free blue nitrile patient exam glove, meets all the requirements of ASTM standard D6319, except for sterility requirements. Non-sterile, powder free, examination gloves made of nitrile and colored blue. The textured gloves are provided in sizes extra small, small, medium, large and extra large. Non-sterile gloves are provided in dispenser boxes.
The provided text describes a 510(k) summary for the "ETS Blue Powder Free Nitrile Patient Exam Glove," a Class I medical device. The summary focuses on demonstrating substantial equivalence to a predicate device by meeting established performance standards.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard) | Device Performance |
|---|---|---|
| Dimension | ASTM D6319 | Meets |
| Physical Properties | ASTM D6319 | Meets |
| Freedom from Pinholes | ASTM D6319 & ASTM D5151 | Meets |
| Powder Residual | ASTM D6124 | Meets (Results generated values below 2mg of residual powder) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10) | Gloves are non-irritating |
| Biocompatibility | Dermal Sensitization in the guinea pig (ISO 10993-10) | Gloves do not display any potential for sensitization |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the specific tests (e.g., number of gloves tested for dimensions, pinholes, etc.). It only indicates that the device "Meets" the specified ASTM and ISO standards, which inherently include sampling plans within those standards.
- Sample Size: Not explicitly stated for each test. The standards referenced (ASTM D6319, D5151, D6124, ISO 10993-10) would define the required sample sizes for demonstrating compliance.
- Data Provenance: Not explicitly stated. This information is typically generated by the manufacturer or a contracted testing lab during product development and quality assurance. It's retrospective testing of the manufactured product.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not applicable to this type of device. The "ground truth" for patient examination gloves is established by objective, verifiable physical and chemical properties measured against recognized industry standards (ASTM, ISO). There are no human "experts" establishing a subjective ground truth for these performance criteria in the way there would be for, for example, image interpretation in a diagnostic AI study.
4. Adjudication Method for the Test Set
This information is not applicable. As mentioned above, the "ground truth" is determined by objective physical and chemical testing against standards, not by expert consensus or adjudication of subjective interpretations.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No. This is a Class I medical device (patient examination glove). MRMC studies are relevant for diagnostic devices, particularly those involving human interpretation of complex data (like medical images), where the AI might assist a human reader. This device does not involve that type of interaction or assessment.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No. This is a physical product (glove), not an algorithm or software device. Standalone performance refers to the accuracy of an AI algorithm on its own.
7. The Type of Ground Truth Used
The ground truth used is based on objective measurements against established national and international standards. Specifically:
- Physical properties (dimensions, tensile strength, ultimate elongation) are measured against ASTM D6319-10.
- Freedom from pinholes is measured against ASTM D5151-06 and D6319-10.
- Residual powder is measured against ASTM D6124-06.
- Biocompatibility (skin irritation and sensitization) is assessed against ISO 10993-10.
8. The Sample Size for the Training Set
Not applicable. This product is a physical medical device. There is no "training set" in the context of machine learning or AI algorithms for this device. The manufacturing process is validated (which might involve statistical process control and testing of batches), but this is distinct from an AI training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. (See point 8).
In summary: The provided 510(k) summary for the ETS Blue Powder Free Nitrile Patient Exam Glove demonstrates its safety and effectiveness by showing that it meets recognized consensus standards (ASTM and ISO) for physical characteristics and biocompatibility, thereby establishing substantial equivalence to a legally marketed predicate device. The nature of this device means that many of the AI/algorithm-specific questions are not relevant.
Ask a specific question about this device
(49 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free black nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
Here's an analysis of the provided information, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard | Acceptance Criteria (from Standard) | Reported Device Performance |
|---|---|---|---|
| Dimension | ASTM D6319-10 | (Implicitly, compliance with all dimensional specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
| Physical Properties | ASTM D6319-10 | (Implicitly, compliance with all physical property specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
| Freedom from Pinholes | ASTM D5151-06 & ASTM D6319-10 | (Implicitly, compliance with pinhole requirements in ASTM D5151-06 and D6319-10) | Meets (meaning it complies with both ASTM D5151-06 and D6319-10) |
| Powder Residual | ASTM D6124-06 | (Implicitly, compliance with powder residual limits in ASTM D6124-06, specifically < 2mg) | Meets (Results generated values below 2mg of residual powder) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10:2010) | Non-irritating | Gloves are non-irritating |
| Biocompatibility | Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | No potential for sensitization | Gloves do not display any potential for sensitization |
2. Sample Size Used for the Test Set and Data Provenance
For the physical and chemical tests (Dimension, Physical Properties, Freedom from Pinholes, Powder Residual), the sample sizes are not explicitly stated in this summary. However, these standards (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06) typically define the sampling plans and quantities required for compliance testing. The data provenance is not explicitly stated in terms of country of origin but is implied to be from testing conducted by or for the device manufacturer to demonstrate compliance with international standards. These tests are generally prospective, in that they are performed specifically to assess the device's adherence to the standard.
For Biocompatibility, the sample sizes are also not explicitly stated in this summary, but ISO 10993-10:2010 would specify the minimum number of animals (rabbits for irritation, guinea pigs for sensitization) for testing. This data is prospectively generated for device testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of device (patient examination glove) does not typically involve expert review for "ground truth" in the way a diagnostic imaging device would. The "ground truth" for the performance criteria is established by the specifications defined within the referenced ASTM and ISO consensus standards. The "experts" would be the scientists and engineers who developed and validated these standards, and the laboratory personnel performing the tests according to standardized protocols. Qualifications of these personnel are not provided, but they would be expected to be trained and certified in laboratory testing procedures relevant to their specific roles.
4. Adjudication Method for the Test Set
Not applicable for this type of device and testing. The tests are objective and quantitative, based on standardized measurement protocols and predefined acceptance criteria in consensus standards. There is no subjective interpretation requiring adjudication among human reviewers.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is a medical device (glove) that does not involve "readers" or artificial intelligence. It's a physical barrier device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a medical device (glove) that does not involve algorithms.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the device's performance is the compliance with established, quantitative consensus standards (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010). These standards define objective chemical and physical properties as well as biological responses.
8. The Sample Size for the Training Set
Not applicable. This device does not use machine learning or AI models, so there is no training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(49 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free blue nitrile patient examination glove, tested for use with Chemotherapy drugs, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
The provided documentation describes the acceptance criteria and performance of the "Blue Powder Free Nitrile Patient Exam Glove," a Class I medical device. This device is a patient examination glove used to prevent contamination between patient and examiner.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard (Acceptance Criteria) | Device Performance |
|---|---|---|
| Dimension | ASTM D6319 | Meets |
| Physical Properties | ASTM D6319 | Meets |
| Freedom from Pinholes | ASTM D5151 & ASTM D6319 | Meets |
| Powder Residual | ASTM D6124 (acceptance criteria typically < 2mg/glove, implied by statement) | Meets (Results generated values below 2mg of residual powder.) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10) | Gloves are non-irritating |
| Biocompatibility | Dermal Sensitization in the guinea pig (ISO 10993-10) | Gloves do not display any potential for sensitization |
Note: The standard ASTM D6319, ASTM D6124, and ASTM D5151 inherently contain specific acceptance limits for metrics like tensile strength, elongation, dimensions, freedom from holes, and powder residual. The document states the device "Meets" these standards, indicating compliance with all their respective criteria. For Powder Residual, a specific value is provided (<2mg), which is typically the acceptance criterion for powder-free gloves.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes used for each specific test (Dimension, Physical Properties, Pinholes, Powder Residual, Biocompatibility). The tests refer to ASTM and ISO standards, which would outline the appropriate sampling plans and methodologies.
The "data provenance" (country of origin, retrospective/prospective) is not provided in the document. The applicant, Northstar Healthcare Holdings, is based in Dublin, Ireland, and the official correspondent (Cypress Medical Products LLC) is in McHenry, IL, USA. However, the location where the testing was conducted is not specified. It is likely prospective testing as it's a product performance evaluation for a 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This device is a patient examination glove, and its performance is evaluated against established physical and biocompatibility standards (ASTM D6319, D5151, D6124, ISO 10993-10) rather than requiring expert interpretation of diagnostic images or clinical outcomes. Therefore, the concept of "experts" establishing ground truth in the way it applies to diagnostic devices (e.g., radiologists interpreting images) is not applicable here. The "ground truth" is determined by objective measurements performed according to the specified standards by trained laboratory personnel.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The tests involve objective measurements and adherence to industry standards, not subjective interpretations requiring adjudicated consensus.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a disposable patient examination glove; it is not an AI-powered diagnostic or assistive technology.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical product, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is based on objective measurements against established national and international standards. These include:
- Physical Property measurements: e.g., tensile strength, elongation, dimensions as per ASTM D6319.
- Freedom from Pinholes: Determined by water leak tests as per ASTM D5151 and D6319.
- Powder Residual measurement: As per ASTM D6124.
- Biocompatibility testing: In-vivo assays (rabbit irritation, guinea pig sensitization) as per ISO 10993-10.
8. The sample size for the training set
Not applicable. This is a physical product, not a machine learning model, so there is no concept of a "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no training set for a machine learning model, there is no ground truth established for a training set.
Ask a specific question about this device
(43 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free blue nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
The provided text describes the regulatory filing for a medical device: "UTB Blue Powder Free Nitrile Patient Examination Gloves." The information focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of AI performance, clinical effectiveness, or diagnostic accuracy. Therefore, many of the requested categories related to AI studies, reader performance, and ground truth establishment are not applicable.
Here's the breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard (Acceptance Criteria) | Device Performance |
|---|---|---|
| Dimension | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from Pinholes | ASTM D5151-06 & D6319-10 | Meets |
| Powder Residual | ASTM D6124-06 | Meets; Results generated values below 2mg of residual powder |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10:2010), Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | Gloves are 'non-irritating'; Gloves do not display any potential for sensitization |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each test mentioned (e.g., number of gloves tested for pinholes, number of animals for biocompatibility). The data provenance is described by the standards used, which are international (ASTM, ISO). There is no explicit mention of the country of origin of the data or whether the tests were conducted retrospectively or prospectively, except that they generally refer to evaluations of the product itself for regulatory compliance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The "ground truth" here is compliance with established international standards for patient examination gloves, rather than interpretation of medical images or data by human experts. The tests are laboratory-based.
4. Adjudication method for the test set
Not applicable. This concept is relevant for studies involving human interpretation or subjective assessments. Here, the tests involve objective measurements against predefined standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a patient examination glove, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device. The tests performed are on the physical and biological properties of the glove.
7. The type of ground truth used
The ground truth used is the standards themselves (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, and ISO 10993-10:2010). The device performance is compared against the requirements stipulated in these recognized industry and international standards for each characteristic.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for this type of medical device.
Ask a specific question about this device
(34 days)
The patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder Free Latex Patient Exam Glove, Smooth and Textured Natural Color (Off White) with Protein Labeling Claim (50 µg/dm² or less of water soluble protein)
Acceptance Criteria and Device Performance for Powder Free Latex Patient Exam Glove
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM Standard D3578 | Meets |
| Physical Properties | ASTM Standard D3578 | Meets |
| Freedom from Pinholes | 21 CFR 800.20; ASTM D5151 | Meets |
| Powder Residual | ASTM Standard D6124 (specifically, < 2 mg per glove) | Meets (Results generated values below 2 mg of residual powder) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10); Dermal Sensitization in the guinea pig (ISO 10993-10) | Gloves are non-irritating; Gloves do not display any potential for sensitization |
| Residual Protein Levels | ASTM Standard D5712 (specifically, ≤ 50 µg/dm² of water soluble protein) | Meets (Protein Labeling Claim: 50 µg/dm² or less of water soluble protein) |
2. Sample Size and Data Provenance
The document does not explicitly state the specific sample sizes used for each test. However, the tests are conducted according to recognized ASTM and ISO standards, which typically specify appropriate sample sizes for such evaluations.
- Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted to meet the defined standards. The submission is for a Traditional 510(k), implying testing was performed to demonstrate substantial equivalence.
3. Number of Experts and Qualifications for Ground Truth
Not applicable. This device is a medical glove, and its performance is evaluated against engineering and biological standards, not through expert-driven image or data interpretation.
4. Adjudication Method
Not applicable. Performance is based on objective laboratory testing against established standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This device is a medical glove, and its performance is evaluated against engineering and biological standards, not typically through MRMC studies.
6. Standalone Performance (Algorithm Only)
Not applicable. This is a physical device (medical glove), not an algorithm or software.
7. Type of Ground Truth Used
The ground truth for this device is based on established industry standards and regulatory requirements. This includes:
- ASTM Standards: D3578 (dimensions, physical properties), D5151 (freedom from pinholes), D6124 (powder residual), D5712 (residual protein levels).
- ISO Standard: 10993-10 (biocompatibility - irritation and sensitization).
- CFR Regulations: 21 CFR 800.20 (freedom from pinholes).
8. Sample Size for the Training Set
Not applicable. As a physical medical device, there is no "training set" in the context of an AI or machine learning model. Performance is evaluated through laboratory testing of manufactured gloves.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for this device. The "ground truth" for evaluating the device's performance is defined by the requirements of the internationally recognized and regulatory standards listed above.
Ask a specific question about this device
Page 1 of 1