Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K141244

    Validate with FDA (Live)

    Date Cleared
    2014-09-09

    (118 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NAMSA Biological Indicator Spore Strip (single species Geobacillus stearothermophilus ATCC® 7953, product code STS-05R, or dual species Geobacillus stearothermophilus ATCC® 7953 and Bacillus atrophaeus ATCC® 9372. product code STNS-65R) is intended for use in testing the efficacy of chemiclave sterilization.

    Performance characteristics are established for exposure at 132ºC for 20 minutes in a MDT Harvey Chemiclave model EC5500 chemiclave sterilizer. Vapo-Steril Solution is the sterilant utilized in the MDT Harvey Chemiclave sterilizer. The solution was cleared under 510(k) number K984270. The validation load used for evaluation of the device performance characteristics consisted of stainless steel dental instruments in a wrapped tray with a mass of 1 kg.

    A reduced incubation time of 72 hours for chemiclave sterilization at 132°C has been validated when the Biological Indicator Spore Strips are used in conjunction with Tryptic Soy Broth (TSB) modified with Bromocresol Purple and incubated at 58° - 62°C.

    Device Description

    The Biological Indicator Spore Strips are typically used in a dental office but may be used by other small healthcare offices which utilize chemiclave processes to sterilize instruments. The spore strips are utilized to verify the chemiclave exposures were effective at killing the Geobacillus stearothermophilus bacterial spores present on the strips in high volume.

    Modification to K113221 are to include chemiclave sterilization at 132°C for 20 minutes in a Harvey model EC5500. The predicate device was not specific to sterilizer type or model number but rather is legally marketed for use in monitoring the chemiclave process. Specificity of the chemiclave model in no way affects the safety or effectiveness of the device.

    NAMSA Biological Indicator Spore Strips are easy to use; simply place the spore strips in the most difficult area to sterilize and process the load as normal. After exposure, remove the spore strips and aseptically transfer to growth medium and incubated. After incubation if no signs of bacterial growth are present, the sterilization cycle was effect.

    The spore strips may be aseptically transferred to standard media such as Tryptic Soy Broth (TSB) and incubated for a minimum of 7 days. Growth will be indicated by the presence of turbidity. Conversely, the strips may be cultured using TSB which has been modified with a pH indicator (Bromocresol Purple) for a reduced incubation time of 72 hours. Growth will be indicated by a change in color of the media from purple to yellow and/or presence of turbidity.

    Biological Indicator Spore Strips consist of a 1.25" x 0.25" filter paper strip which has been inoculated with either single species (Geobacillus stearothermophilus ATCC® 7953 at a population level of 105 per strip), or dual species (Geobacillus stearothermophilus ATCC® 7953 at a population level of 105 and Bacillus atrophaeus ATCC® 9372 at a population level of 106 per strip). The spore strips are individually packaged in a 30# glassine pouch.

    AI/ML Overview

    This document describes the NAMSA Biological Indicator Spore Strip, a device intended to test the efficacy of chemiclave sterilization. However, the provided text does not contain a typical acceptance criteria table with reported device performance metrics in the format usually associated with AI/ML device studies. Instead, it outlines the technological characteristics of the device and a list of testing performed to demonstrate substantial equivalence to a predicate device.

    Given the nature of this submission (a 510(k) for a biological indicator, not an AI/ML device), the "acceptance criteria" discussed are primarily about meeting performance characteristics analogous to existing predicate devices, rather than predefined statistical thresholds for a new analytical algorithm.

    Here's an attempt to extract the relevant information based on the provided text, reformatted to fit the request where possible, keeping in mind the device is a biological indicator, not an AI algorithm.


    1. Table of Acceptance Criteria (Technological Characteristics) and Reported Device Performance

    The acceptance criteria are implicitly defined by the properties of the predicate device and the new device's ability to demonstrate substantial equivalence in key performance characteristics. The document presents a comparison of the new device's technological characteristics against the predicate device.

    CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (NAMSA Spore Strip)
    Product TypeBiological IndicatorBiological Indicator
    Intended UseChemiclave at 132°CChemiclave 132°C for 20 minutes, Harvey Model EC5500
    ConstructionPaper spore strip in glassinePaper spore strip in glassine
    OrganismG. stearothermophilus, ATCC® 7953G. stearothermophilus, ATCC® 7953
    Viable Spore Population10^5^/strip10^5^/strip (single species) or 10^5^ G. stearothermophilus and 10^6^ B. atrophaeus (dual species)
    Certified Resistance CharacteristicsDvalue at 132°C: 1.0 to 3.0 Minutes; Survival: 5 min; Kill: 15 minDvalue at 132°C: 1.0 to 3.0 Minutes; Survival: 5 min; Kill: 15 min
    Incubation Temperature60°±2°C60°±2°C
    Readout Time in Modified Culture Medium72 hours72 hours
    Shelf LifeMinimum of 18 Months18 Months

    Study Proving Acceptance Criteria is Met:

    The document describes a series of studies designed to demonstrate that the NAMSA Biological Indicator Spore Strip is substantially equivalent to the predicate device and performs as intended. These include:

    • Survival-Kill Window Verification: Demonstrated equivalence with the predicate device at the manufacturer's certified chemiclave resistance characteristics (5-minute survival, 15-minute kill). This was also verified at the end of the device's shelf life.
    • Population Evaluation: Confirmed equivalent population levels to the predicate device.
    • Readout Time Evaluation: Confirmed equivalent reduced incubation times to the predicate device.
    • Resistance Characterization (D-value Determination and Verification): Determined D-values at 124°C, 132°C, and 140°C using the Fraction Negative Method. Verified D-values using expired Vapo-Steril solution to account for its full shelf life.
    • Load/Chamber Evaluation Studies: Verified device performance in a full chamber with a standard load (stainless steel dental instruments in a wrapped tray with a mass of 1 kg) to confirm no negative influence on sterilization conditions.
    • Quantitative Determination of Pre-Exposure Phase Lethality: Quantified spores killed prior to the exposure phase across various load conditions and temperature extremes.
    • Carrier and Primary Packaging Material Evaluation: Assessed inhibitory properties and degradative changes of materials.
    • Holding Time Assessment: Verified that the Biological Indicators (BIs) were not negatively impacted if left at room temperature for 96 hours post-exposure prior to incubation.
    • Recovery Protocols - Reduced Incubation Time (RIT): Determined that the 72-hour RIT claim remains valid even with a 96-hour delay in incubation post-exposure.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (number of biological indicators or test runs) for each individual test or for a single "test set" in the context of an algorithm. However, testing involved:

    • Evaluation of "NAMSA's Biological Indicator Spore Strips" for population and readout time.
    • "Three lots of the device" were evaluated for D-values with expired Vapo-Steril solution.
    • "Standard chemiclave sterilizer" at "standard 132°C cycle for 20 minutes."
    • "Validation load...consisted of stainless steel dental instruments in a wrapped tray with a mass of 1 kg."
    • "BI's not processed...immediately following exposure are not negatively impacted" if left at room temperature for 96 hours.

    The provenance of data is not explicitly stated in terms of country of origin, but all testing appears to be conducted by NAMSA (North American Science Assoc., Inc.) to support their 510(k) submission to the US FDA. The studies are prospective in nature, as they involve performing specific tests on the device under defined conditions.

    3. Number of Experts and Qualifications for Ground Truth

    This type of submission for a biological indicator does not typically involve human experts establishing "ground truth" for a test set in the way an AI/ML diagnostic device does. The ground truth for biological indicators is established through objective biological and physical measurements:

    • Bacterial spore population counts: Measured via standard microbiological techniques.
    • Survival/kill determination: Based on the presence or absence of bacterial growth after incubation, which is an objective observation (turbidity or color change).
    • D-value calculations: Derived mathematically from experimental data on spore survival at different exposure times.
    • Sterilization efficacy: Determined by the physical/chemical process itself and validated by the BI's response.

    Therefore, the concept of "number of experts used to establish ground truth" and their "qualifications" is not directly applicable in the same way it would be for medical image interpretation or similar diagnostic tasks. The "experts" in this context would be the microbiologists and engineers who designed and executed the validated test methods.

    4. Adjudication Method for the Test Set

    Not applicable. As explained above, the "ground truth" for a biological indicator is based on objective biological and physical measurements rather than expert human interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI/ML diagnostic device that would be used by human readers. Its purpose is to indicate the success of a sterilization process.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This is a physical biological indicator, not an algorithm. Its performance is inherent in its biological and physical properties.

    7. Type of Ground Truth Used

    The ground truth used is based on:

    • Biological viability: Growth or no growth of Geobacillus stearothermophilus spores.
    • Quantitative microbiology: Spore population counts and D-value calculations.
    • Physical sterilization parameters: Exposure time, temperature, and sterilant concentration, which are controlled experimental variables against which the BI's performance is measured.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that uses training data. The device's performance is based on its inherent physical and biological characteristics, which are then validated through empirical testing, not learned from a dataset.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132291

    Validate with FDA (Live)

    Date Cleared
    2014-05-20

    (301 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NAMSA Chemical Process Indicator Strip for Steam, REF TST4-S is intended for use with individual materials (i.e. pouches, pack, tray) to demonstrate that the material has been exposed to a steam sterilization process to distinguish between processed and unprocessed goods. The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals as a process indicator:

    • 121°C for 30 minutes (gravity cycle) .
    • 134℃ for 3 minutes (pre-vacuum) .
    Device Description

    The NAMSA Chemical Process Indicator Strip for Steam utilizes NAMSA indicating ink SSI-10-YBI that when exposed to high temperature steam turns from an initial yellow color to a permanent dark brown/black signal color. The intended use for the NAMSA Chemical Process Indicator Strip for Steam processes is a Process Indicator intended for use by health care facilities to distinguish between processed and unprocessed goods. The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a permanent dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals:

    • 121℃ for 30 minutes (gravity cycle) ●
    • 134℃ for 3 minutes (pre-vacuum) .
    AI/ML Overview

    This document describes the NAMSA Chemical Process Indicator Strip for Steam and its performance testing to demonstrate substantial equivalence to a predicate device.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from ANSI/AAMI/ISO 11140-1:2005 guidelines for Class 1 Process Indicators for STEAM. The reported device performance indicates that the NAMSA Chemical Process Indicator Strip for Steam met these criteria.

    Test ConditionTest TimeTest TemperatureAcceptance Criteria: No change or a change markedly different from specifiedAcceptance Criteria: Visible change as specifiedReported Device Performance (Implied by equivalence statement)
    Saturated Steam3.0 min ± 5 sec121°C (+3/0°C)Acceptable resultUnacceptable resultMet: No change / Markedly different from specified color change (yellow remains yellow)
    Saturated Steam10.0 min 0/-5 sec121°C (+3/0°C)Unacceptable resultAcceptable resultMet: Visible change to dark brown/black
    Saturated Steam20 sec134°C (+3/0°C)Acceptable resultUnacceptable resultMet: No change / Markedly different from specified color change (yellow remains yellow)
    Saturated Steam0.5 min ± 5 sec134°C (+3/0°C)Acceptable resultUnacceptable resultMet: No change / Markedly different from specified color change (yellow remains yellow)
    Saturated Steam2 min +5/0 sec134°C (+3/0°C)Unacceptable resultAcceptable resultMet: Visible change to dark brown/black
    Dry Heat30 min ± 1 min140°C (+2/0°C)Acceptable resultUnacceptable resultMet: No change / Markedly different from specified color change (yellow remains yellow)

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that testing was performed using "multiple lots of NAMSA Chemical Process Indicator Strip for Steam." However, it does not specify the exact sample size for the test set (number of strips from each lot) nor the country of origin of the data. The studies are prospective in nature, as they involve testing the device under controlled conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used or their qualifications for evaluating the color changes. Given the nature of a chemical indicator, the "ground truth" for a color change is typically determined by visual observation against a known standard or a reference indicator by trained personnel. It is inferred that the evaluation was performed by qualified personnel involved in the testing, likely those trained in quality control or laboratory procedures for medical device sterility indicators.

    4. Adjudication Method for the Test Set

    The document does not describe a specific adjudication method (like 2+1 or 3+1). The nature of the test (color change) implies a direct observation against a pass/fail criterion. It's likely that a single observer or
    an internal quality control process determined the outcome.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices where human interpretation of images or data is involved, often with an AI assist. The NAMSA Chemical Process Indicator Strip is a simple chemical indicator with a direct visual color change, not an AI-driven diagnostic tool. Therefore, the question of human readers improving with AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm-only performance study was not done. This device is a passive chemical indicator that relies on a visual color change in response to sterilization conditions, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for this device is based on pre-defined physical and chemical changes as specified by the manufacturer and validated against the ANSI/AAMI/ISO 11140-1:2005 standard. The "ground truth" is the expected behavior (color change or no change) of the indicator when subjected to specific time and temperature intervals under saturated steam or dry heat conditions. This is a direct physical measurement/observation against established criteria.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" as typically understood in machine learning. For chemical indicators, the "training" analogous to development involves:

    • Developing the indicating ink (NAMSA formulation SSI-10-YBI)
    • Formulating the design (substrate, ink application, over-laminate)
    • Testing numerous iterations during the R&D phase to achieve the desired performance characteristics (i.e., the correct color change at the specified parameters and no change outside those parameters).

    The document states the "Indicating Ink was developed and validated for compliance with ANSI/AAMI/ISO 11140-1:2005 guideline," implying an iterative development and testing process.

    9. How the Ground Truth for the Training Set Was Established

    As there isn't a "training set" in the machine learning sense, the "ground truth" for the development and validation of the indicating ink and device design would have been established through:

    • Reference Standards: Adherence to the critical parameters and performance requirements outlined in the ANSI/AAMI/ISO 11140-1:2005 standard, which defines what constitutes a "visible change" and "no change" for a Class 1 process indicator.
    • Controlled Laboratory Conditions: Exposing prototype indicators to precisely controlled steam sterilization cycles (time, temperature, saturated steam) and dry heat conditions in a laboratory setting.
    • Visual Assessment: Comparing the resulting color changes (or lack thereof) against the predetermined "acceptable" and "unacceptable" results as per the standard for each test condition. This iterative process allows for refinement of the ink formulation and device composition until the desired performance is consistently achieved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113221

    Validate with FDA (Live)

    Date Cleared
    2012-04-27

    (178 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NAMSA Biological Indicator Spore Strip (single species Geobacillus stearothermophilus ATCC® 7953, or dual species Geobacillus stearothermophilus ATCC® 7953 and Bacillus atrophaeus ATCC® 9372) · is intended for use in testing the efficacy of steam sterilization for single species and steam and ethylene oxide sterilization for dual species.

    Performance characteristics are established in accordance with ANSI/AAMI/ISO 11138 and USP for 121°C steam gravity displacement for 30 minutes and ethylene oxide at 600 mg/L, 60% RH and 55°C for 2 hours, 10 minutes.

    A reduced incubation time of 24 hours for steam sterilization has been validated when the Biological Indicator Spore Strips are used in conjunction with Tryptic Soy Broth (TSB) modified with Bromocresol Purple.

    Device Description

    The NAMSA Biological Indicator Spore Strip consists of a 1.25" x 0.25" filter paper strip inoculated with either a single species (Geobacillus stearothermophilus ATCC® 7953, 105) or dual species (Geobacillus stearothermophilus ATCC® 7953, 106 and Bacillus atrophaeus ATCC® 9372, 106) bacterial spores. The strip is packaged in a 30# glassine pouch.

    Operational Principles: The NAMSA Biological Indicator Spore Strip (single species Geobacillus stearothermophilus ATCC® 7953, or dual species Geobacillus stearothermophilus ATCC® 7953 and Bacillus atrophaeus ATCC® 9372) is intended for use in testing the efficacy of steam sterilization for single species and steam and ethylene oxide sterilization for dual species.

    Performance characteristics are established in accordance with ANSI/AAMI/ISO 11138 and USP for 121°C steam gravity displacement for 30 minutes and ethylene oxide at 600 mg/L, 60% RH and 55°C for 2 hours, 10 minutes.

    The media containing the spore strip should be incubated at the organism's growth temperature. Media should be monitored daily for visible signs of growth and results recorded.

    When standard media is utilized, incubate strips for a minimum of 7 days. Growth will be indicated by the presence of turbidity. A reduced incubation time of 24 hours for steam sterilization has been validated when the Biological Indicator Spore Strips are used in conjunction with Tryptic Soy Broth (TSB) modified with Bromocresol Purple.

    AI/ML Overview

    The provided document describes the NAMSA Biological Indicator Spore Strips and their performance, focusing on their substantial equivalence to predicate devices for assessing sterilization efficacy.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state a table of acceptance criteria with numerical targets. Instead, it refers to compliance with recognized standards. The performance characteristics are established according to:

    • ANSI/AAMI/ISO 11138: This is a series of international standards for biological indicators.
    • USP (United States Pharmacopeia): This outlines pharmaceutical and medical device standards.

    The performance is reported in the context of these standards for specific sterilization parameters:

    Acceptance Criteria (Based on ANSI/AAMI/ISO 11138 and USP)Reported Device Performance
    For Steam Sterilization (Single Species: Geobacillus stearothermophilus ATCC® 7953): - Efficacy for 121°C steam gravity displacement for 30 minutes. - Required resistance characteristics (e.g., D-value, Z-value, Survival/Kill Windows) must meet standard specifications. - Required Total Viable Spore Count. - Medium suitability for organism growth. - Holding time assessment must confirm stability. - Recovery protocols must be effective. - Reduced incubation time (24 hours with specific media) validation (if applicable).Demonstrates efficacy in testing steam sterilization for Geobacillus stearothermophilus ATCC® 7953. - Testing performed for Total Viable Spore Count. - Testing performed for Resistance Characteristics Studies including D-value, Z-value and Survival/Kill Windows. - Compliance with ANSI/AAMI/ISO 11138 and USP for 121°C steam gravity displacement for 30 minutes. - Holding Time Assessment performed. - Recovery Protocols evaluated. - Medium Suitability evaluated. - Reduced incubation time of 24 hours for steam sterilization has been validated when used with Tryptic Soy Broth (TSB) modified with Bromocresol Purple.
    For Steam and Ethylene Oxide (EO) Sterilization (Dual Species: Geobacillus stearothermophilus ATCC® 7953 and Bacillus atrophaeus ATCC® 9372): - Efficacy for 121°C steam gravity displacement for 30 minutes. - Efficacy for ethylene oxide at 600 mg/L, 60% RH and 55°C for 2 hours, 10 minutes. - Required resistance characteristics (e.g., D-value, Z-value, Survival/Kill Windows) must meet standard specifications. - Required Total Viable Spore Count. - Medium suitability for organism growth. - Holding time assessment must confirm stability. - Recovery protocols must be effective. - Reduced incubation time (24 hours with specific media) validation (if applicable).Demonstrates efficacy in testing steam and ethylene oxide sterilization for Geobacillus stearothermophilus ATCC® 7953 and Bacillus atrophaeus ATCC® 9372. - Testing performed for Total Viable Spore Count. - Testing performed for Resistance Characteristics Studies including D-value, Z-value and Survival/Kill Windows. - Compliance with ANSI/AAMI/ISO 11138 and USP for 121°C steam gravity displacement for 30 minutes and ethylene oxide at 600 mg/L, 60% RH and 55°C for 2 hours, 10 minutes. - Holding Time Assessment performed. - Recovery Protocols evaluated. - Medium Suitability evaluated. - Reduced incubation time of 24 hours for steam sterilization has been validated when used with Tryptic Soy Broth (TSB) modified with Bromocresol Purple.
    General Acceptance Criteria: - Total viable spore count within specified limits. - Carrier and Primary Packaging Materials Evaluation to ensure integrity and compatibility.- Total Viable Spore Count testing was performed. - Carrier and Primary Packaging Materials Evaluation was performed.

    The study described indicates that "testing was performed for steam for single specie spore strips and steam and Ethylene Oxide (EO) sterilization processes for dual species," validating the performance characteristics against these standards. The conclusion is that the device is "substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document states that "Multiple lots of NAMSA Biological Indicator Spore Strips were utilized" for testing. However, it does not provide a specific numerical sample size for the test set (e.g., how many strips per lot, how many lots).
    • Data Provenance: The document does not specify the country of origin of the data. It is a submission to the FDA in the US for a US-based company (NAMSA). The data is undoubtedly retrospective as it describes testing already performed to demonstrate device performance for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document. Biological indicator testing focuses on measurable biological and physical parameters (spore count, D-value, survival/kill) rather than human interpretation. Therefore, "experts" in the context of ground truth establishment as it applies to image interpretation or diagnostic test reading would not typically be relevant here. The ground truth is intrinsically tied to the viability of the spores after exposure to sterilization processes, assessed by laboratory culture methods.

    4. Adjudication Method for the Test Set:

    • This information is not applicable/provided. Adjudication methods are typically employed in studies where human readers or multiple assessments are involved (e.g., radiology studies). For biological indicators, the outcome (growth/no growth, quantitative spore count) is a direct, objective measurement.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., medical imaging AI). Biological indicators are laboratory-based tools with objective outcomes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable in the context of an "algorithm." This device is a biological indicator, not a software algorithm. The "standalone performance" is the inherent performance of the spore strip itself as it responds to sterilization processes, which is what the testing describes.

    7. The Type of Ground Truth Used:

    • The ground truth in this context is based on direct biological measurement and adherence to established physical/chemical parameters of sterilization. Specifically:
      • Culture-based viability: Growth or no growth of spores after exposure to sterilization, assessed by incubating the strips in suitable media. This determines survival.
      • Quantitative Spore Count: Measurable number of viable spores.
      • Resistance Characteristics (D-value, Z-value, Survival/Kill Windows): These are quantitative measures directly derived from the resistance of the spores to specific sterilization conditions, which are established scientific benchmarks for biological indicators.
      • Adherence to ANSI/AAMI/ISO 11138 and USP standards: These standards define the accepted performance criteria for effective sterilization monitors.

    8. The Sample Size for the Training Set:

    • This information is not provided and is not applicable in the context of this device. Biological indicators do not involve machine learning algorithms that require training sets. Their performance is inherent to their biological and physical composition and is validated through experimental testing.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As stated above, there is no "training set" for a biological indicator. The ground truth for validating their performance is established through rigorous laboratory testing against defined sterilization cycles and using culture methods to confirm spore viability, all under the guidelines of recognized standards like ANSI/AAMI/ISO 11138 and USP.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113302

    Validate with FDA (Live)

    Date Cleared
    2012-01-20

    (73 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NAMSA SCBI is intended for use in monitoring the efficacy of saturated steam sterilization processes. Performance characteristics are for pre-vacuum and gravity displacement 121°C steam processes. Additional saturated steam sterilization temperatures are also included on the Certificate of Analysis. NAMSA Steam SCBIs are also appropriate for use in saturated steam processes of 132°C, 134°C and 135°C.

    Device Description

    The activated SCBI should be incubated at 58-62°C for a minimum of 24 hours. The SCBI should be monitored for visible signs of growth. Growth will be indicated by a color shift from purple to yellow and/or the presence of turbidity. The absence of growth indicates the exposure was effective.

    AI/ML Overview

    1. Table of acceptance criteria and reported device performance:

    The provided document does not explicitly state quantitative acceptance criteria or a dedicated section for "reported device performance." However, based on the testing descriptions and the nature of a biological indicator, the performance is implicitly linked to the successful demonstration of specific resistance characteristics and proper recovery.

    Acceptance Criteria (Implied)Reported Device Performance/Testing Description
    Viability of Spores"Total Viable Spore Count" - This test is conducted to ensure the SCBIs contain a sufficient and consistent number of viable spores (Geobacillus stearothermophilus) to accurately monitor sterilization. Specific counts are generally specified by standards, but not detailed here.
    Resistance to Sterilization Process"Resistance Characteristic Studies including D value, z value, Survival/Kill Windows" - This demonstrates the SCBI's ability to resist specific steam sterilization cycles, quantifying its resistance (D-value) and the temperature coefficient (z-value) for demonstrating effectiveness. The survival/kill windows confirm that the SCBI shows growth when conditions are insufficient for sterility and no growth when conditions are sufficient.
    Material Compatibility/Integrity"Carrier and Primary Packaging Materials Evaluation" - This ensures the materials housing the spores and growth medium do not interfere with the sterilization process or the indicator's performance.
    Stability Over Time (Shelf Life)"Holding Time Assessment" - This likely refers to testing performed over the product's shelf life to ensure the SCBI maintains its performance characteristics. The document also states the device has "the same or similar shelf life" as the predicate.
    Accurate and Timely Detection of Growth/No Growth"Recovery Protocols Reduced Incubation Time Studies" & "Medium Suitability" - These tests ensure that surviving spores can be accurately and reliably recovered and grow within the specified incubation time (24 hours in this case), and that the growth medium supports this. The "color shift from purple to yellow and/or the presence of turbidity" indicates growth.
    Monitoring Efficacy in Various Steam ProcessesStated for "pre-vacuum and gravity displacement 121°C steam processes" and "saturated steam processes of 132°C, 134°C and 135°C." The testing would have evaluated the performance at these specific temperatures and exposure times.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact sample size for each test. It broadly states "multiple lots of NAMSA Steam Self-Contained Biological Indicators over the range of the shelf life" were used for the various tests.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given it's a 510(k) submission to the FDA, the testing was likely conducted in accordance with recognized FDA consensus standards and guidance documents, implying prospective studies for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to this type of device (biological indicator). The "ground truth" for a biological indicator is established by standard microbiological and sterilization testing methods, not by expert consensus on visual interpretation of results. The ground truth would be the known efficacy of the sterilization cycle itself, determined by validated physical and chemical parameters, against which the BI's performance is measured.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation of images or clinical data, where there might be inter-reader variability. For biological indicators, the results (growth/no growth, D-value, Z-value) are objectively measured based on established microbiological and sterilization science principles.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a biological indicator, not an AI-powered diagnostic tool requiring human interpretation. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a single-use, self-contained biological indicator. Its function is to provide a visual indication of sterilization efficacy (color change and/or turbidity), which is then interpreted by a human. There is no "algorithm only" component to assess.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for a biological indicator's performance is established by standardized microbiological methods and established sterilization parameters.

    • For D-value and z-value determination: This involves exposing the BIs to precisely controlled sterilization cycles, often using resistance test equipment, to determine the exact time/temperature required to kill a certain percentage of spores. The "ground truth" is the mathematically derived logarithmic reduction of microorganisms based on kill-time studies.
    • For survival/kill windows: The ground truth is determined by running cycles that are known (validated) to achieve sterility and cycles that are known (validated) to fail sterility. The BI's performance (growth/no growth) must align with these known outcomes.
    • For total viable spore count: The ground truth is the actual number of viable spores determined by standard plate counting methods.

    8. The sample size for the training set:

    Not applicable. This device does not involve machine learning or AI, so there is no "training set." The performance is based on the intrinsic biological and physical properties of the indicator.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983451

    Validate with FDA (Live)

    Date Cleared
    1999-04-23

    (205 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RSI Process Challenge Device is intended to be used in conjunction with RSI Rapid Indicators for Monitoring Steam Sterilization to monitor the effectiveness of wrapped loads at 132°C/134°C vacuum assisted steam sterilization cycles.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) letter dated April 23, 1999, for the "RSI® Process Challenge Device for Monitoring Steam Sterilization Process" does not contain the detailed information required to answer your request about acceptance criteria and a study proving device performance.

    The letter primarily:

    • Acknowledges receipt and review of the 510(k) submission.
    • States that the device is substantially equivalent to pre-amendment devices.
    • Informs the manufacturer that they can market the device, subject to general controls and good manufacturing practices.
    • Provides contact information for further inquiries.
    • Includes an "Indications for Use Statement" for the device.

    It does not include:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes or data provenance for a test set.
    3. Details on experts used for ground truth or adjudication methods.
    4. Information about MRMC or standalone studies.
    5. The type of ground truth used.
    6. Sample size for a training set or how its ground truth was established.

    To obtain this information, one would typically need to refer to the full 510(k) submission document itself, which would contain the detailed testing protocols and results that the FDA reviewed to make its substantial equivalence determination. The letter provided is simply the FDA's decision communication.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962649

    Validate with FDA (Live)

    Date Cleared
    1997-09-26

    (445 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RSI Rapid Indicator, a multiple, interactive, bacterial enzyme indicator, is used for monitoring saturated steam sterilization processes operating at 121°C gravity, 132°C vacuum assisted 134°C vacuum assisted cycles.

    Device Description

    RSI, Rapid Indicator

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for the "RSI™ Rapid Indicator for Monitoring Steam Sterilization." This type of document does not contain the detailed study information or acceptance criteria requested.

    A 510(k) clearance primarily establishes substantial equivalence to a predicate device already on the market. It confirms that the new device is as safe and effective as a legally marketed device and does not typically delve into the specifics of performance studies in the way you've outlined for AI or diagnostic devices.

    Therefore, I cannot provide the requested information from this document. The document confirms the device name, its use for monitoring steam sterilization processes, and its substantial equivalence determination, but no detailed performance metrics, study designs, sample sizes, or ground truth establishment methods are present.

    To obtain such information, one would typically need to refer to the original 510(k) submission document itself (which is usually much more extensive than the clearance letter) or published studies related to the device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1