510(k) Data Aggregation

    K Number
    K142040

    Validate with FDA (Live)

    Device Name
    OCS-B COLLAGEN
    Manufacturer
    Date Cleared
    2015-07-31

    (368 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OCS-B Collagen® is recommended for:

    • Augmentation or reconstructive treatment of the alveolar ridge.
    • Filling of infrabony periodontal defects.
    • Filling of defects after root resection, apicoectomy, and cystectomy
    • Filling of extraction sockets to enhance preservation of the alveolar ridge
    • Elevation of maxillary sinus floor
    • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
    Device Description

    OCS-B Collagen® is a combination of purified cancellous bone mineral granules (OCS-B") and 10% collagen in a block form in a blister and cylinderic form in a syringe and blister. It is sterilized by yirradiation.

    AI/ML Overview

    The provided text describes the OCS-B Collagen® device, a bone grafting material, and its substantial equivalence to predicate devices, primarily Bio-Oss Collagen®. The information requested falls under the category of performance testing for medical devices, which typically includes detailed studies to prove efficacy and safety. However, this document is a 510(k) summary, which focuses on demonstrating "substantial equivalence" to existing legally marketed devices rather than performing extensive novel clinical trials as would be required for a novel device or a PMA.

    Therefore, the study described here is primarily a comparative study against predicate devices, supported by biocompatibility testing and animal studies, rather than a standalone clinical efficacy study with specific acceptance criteria in terms of performance metrics like sensitivity, specificity, or improvement with AI assistance.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in terms of quantitative performance metrics (e.g., specific percentages for success rates, bone growth, etc.). Instead, the acceptance criteria are implicitly that the OCS-B Collagen® device is "substantially equivalent" to its predicate devices (Bio-Oss® Collagen and OCS-B®) in terms of:

    Acceptance Criterion (Implicit)Reported Device Performance
    Same Indications for UseOCS-B Collagen® has the same indications for use as Bio-Oss® Collagen, including augmentation/reconstructive treatment of alveolar ridge, filling of infrabony periodontal defects, defects after root resection/apicoectomy/cystectomy, extraction sockets, elevation of maxillary sinus floor, and filling of periodontal/peri-implant defects in conjunction with GTR/GBR.
    Same Intended UseOCS-B Collagen® has the same intended use as Bio-Oss® Collagen and OCS-B® (adjective therapy in restoring bony defects).
    Similar Physical and Chemical StructureBoth OCS-B Collagen® and Bio-Oss® Collagen are porous, biocompatible bone grafts facilitating new bone formation. They have the same component ratio of bone mineral granules and collagen, same source of bone (bovine bone), and collagen (porcine skin). Physical and chemical characteristics were found to be comparable.
    Comparable BiocompatibilityOCS-B Collagen® was subjected to a full range of biocompatibility tests (ISO 10993, cytotoxicity, hemolysis, acute systemic injection, intracutaneous reactivity, skin sensitization, genotoxicity, oral mucosa irritation, implantation) and confirmed equivalence to predicates.
    Similar Performance in Animal Studies (Bone Growth and Resorption)In several animal studies, both OCS-B Collagen® and predicate devices were found to grow new bone and be subsequently resorbed at similar rates.
    Successful Clinical Outcomes (Case Series)A clinical case series using OCS-B Collagen® resulted in defect healing and formation of new bone of sufficient quality to allow dental implant placement.
    Same Sterilization ProcessSterile by Gamma Irradiation (SAL 1 x 10⁻⁶).
    Same Shelf-Life36 Months.
    Compliance with Relevant Standards (ASTM F1581-99, FDA "Medical Device Materials Derived from Animal Sources", ISO 11137, ASTM F1980-99)OCS-B Collagen® was tested in accordance with these standards. (Compliance implies meeting the requirements set forth in these standards, but specific results against numerical criteria are not detailed in this summary).
    Controlled RisksOCS-B Collagen® presents the same types of potential risks as the predicate and controls these risks in a similar manner (e.g., manufacturing processes for removal of organic impurities, protein content limits, TSE/virus inactivation studies).

    2. Sample Size Used for the Test Set and Data Provenance

    • Animal Studies: "several animal studies" are mentioned, but no specific sample sizes or species are provided in this summary document.
    • Clinical Case Series: "a clinical case series" is mentioned, but no specific number of patients (sample size) is provided.
    • Data Provenance: Not explicitly stated for animal studies or the clinical case series (e.g., country of origin, retrospective/prospective). However, the sponsor is NIBEC Co., Ltd. from Korea. Given the K-number, the submission is to the U.S. FDA, so the data would have been submitted to the U.S. regulatory body. The document does not specify if these studies are retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This type of information is generally not applicable to the studies described for this device. The "ground truth" here is the biological and clinical outcome (bone formation, resorption, defect healing, suitability for implant placement), which would be assessed by standard histological, imaging, and clinical follow-up methods by trained professionals (e.g., veterinarians for animal studies, dentists/oral surgeons for clinical case series). The document does not specify an "expert panel" for establishing ground truth in the context of, for example, image interpretation.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this 510(k) summary. Adjudication methods (like 2+1, 3+1) are typically used for studies where multiple human readers interpret data (e.g., images) and consensus is needed, often in AI comparative effectiveness studies. The studies mentioned here focus on biological outcomes rather than interpretive tasks.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not conducted as described in this summary. The comparison is between the device and predicate devices based on physical/chemical properties, biocompatibility, and general performance in animal studies and a clinical case series, not a comparative study of human reader performance with and without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an AI algorithm's performance, which is not relevant for this bone graft material. The device itself is a physical implant, not a diagnostic algorithm.

    7. The Type of Ground Truth Used

    The ground truth used in the animal studies and clinical case series would be based on:

    • Histology: Direct examination of tissue to confirm new bone formation and resorption.
    • Imaging: X-rays, CT scans, or other imaging modalities to assess defect healing and bone density/volume.
    • Clinical Outcomes: Assessment of defect healing, absence of adverse events, and functional outcomes, such as successful dental implant placement.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical bone graft, not a machine learning algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133507

    Validate with FDA (Live)

    Device Name
    GUIDOSS
    Manufacturer
    Date Cleared
    2014-11-05

    (356 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GuidOss® is recommended for:

    • Simultaneous use of membrane (GuidOss®) and implants : 1.
    • Augmentation around implants placed in immediate extract sockets; 2.
      1. Augmentation around implants placed in delayed extraction sockets;
      1. Localized ridge augmentation for later implantation;
      1. Alveolar ridge reconstruction for prosthetic treatment
      1. Filling of bone defects after root resection, cystectomy, removal of retained teeth.
      1. Guided bone regeneration in dehiscence defects; and
      1. Guided tissue regeneration procedures in periodontal defects.
    Device Description

    GuidOss® is made from pure type I collagen membrane obtained by a standardized controlled manufacturing process. The type I collagen is obtained from veterinary certified porcine and is carefully purified to avoid immunological reactions. GuidOss® membrane is manufactured by fibrillogenesis and crosslinking. The collagen fibers are self-assembled by the process of fibrillogenesis. The crosslinked collagen fibers exhibited increased mechanical strength, thermal stability and increased resistance to pepsin digestion compared to non-crosslinked collagen.

    AI/ML Overview

    The provided document is a 510(k) summary for the medical device GuidOss®, a resorbable collagen membrane. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance acceptance criteria through a dedicated study with defined metrics.

    Here's an analysis of the provided information, addressing your questions where possible:

    1. A table of acceptance criteria and the reported device performance

    The document does not present specific quantitative acceptance criteria or reported device performance metrics in the way one would typically find for an AI/device performance study (e.g., sensitivity, specificity, AUC values against a predefined threshold). Instead, it relies on demonstrating substantial equivalence to predicate devices. The "performance" is implicitly defined by the intended use and the predicate devices' established performance.

    Property/CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (GuidOss®)
    Intended UseSame or similar to predicate devices: Used for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus; and other specified indications (simultaneous use of membrane and implants, augmentation around implants, localized ridge augmentation, alveolar ridge reconstruction, filling of bone defects, guided bone regeneration in dehiscence defects, guided tissue regeneration).Used for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus; and all other specified indications (see pages 2-3 for full list).
    Target PopulationHuman oral, periodontalHuman oral, periodontal
    MaterialType I collagen (similar to predicates); Porcine sourceType I collagen; Purified from veterinary certified porcine skin
    BiocompatibilityMust be biocompatible, as demonstrated by a range of tests (genotoxicity, micronucleus, intracutaneous reactivity, maximization/sensitization, pyrogen, acute systemic injection, cytotoxicity, implantation). Non-pyrogenic.Biocompatible, demonstrated by: Genotoxicity testing, Micronucleus Test for Genetic Toxicology, Intracutaneous reactivity testing, Maximization and sensitization testing, Pyrogen testing, Acute systemic injection testing, Cytotoxicity testing, Implantation testing. Non-Pyrogenic: Yes. Confirmed product safety in accordance with ISO 10993 and FDA's "Class II Special Controls Guidance Document."
    PerformancePeriodontal Regeneration (comparable to predicate devices)Periodontal Regeneration. Animal studies showed growth of new bone and similar resorption rates to predicate. Clinical case series resulted in defect healing and new bone formation sufficient for implant placement.
    SterilizationSterile by Gamma Irradiation (similar to predicates)Sterile by Gamma Irradiation to SAL 1 x 10⁻⁶.
    Shelf-Life36 Months (similar to predicates)36 Months
    Safety EfficacyRaises no new issues of safety and efficacy compared to predicate devices.Demonstrates substantial equivalence, expected to be safe and effective for intended uses.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "several animal studies" and "a clinical case series."

    • Animal Studies: Sample size is not specified. Provenance not specified but likely conducted by the manufacturer or a contracted lab. These would be prospective.
    • Clinical Case Series: Sample size is not specified. Provenance not specified. This would typically be prospective, showcasing real-world use.

    The document does not detail a "test set" in the context of an AI device's validation with specific performance metrics against a ground truth.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a bone grafting material, not an AI device that requires expert adjudication of images or data for ground truth establishment. The "ground truth" for its performance would be histological evaluation of bone regeneration in animal models or clinical outcomes in human subjects, as observed by clinicians.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. See point 3. This type of adjudication is typically for image-based diagnostics where human experts generate a consensus label.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical device (resorbable collagen membrane), not an AI-assisted diagnostic tool. Therefore, MRMC studies and "human readers improving with AI" are not relevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. See point 5.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance of the GuidOss® device, the "ground truth" would be:

    • Pathology/Histology: For the animal studies, this would confirm new bone formation and resorption rates.
    • Clinical Outcomes Data: For the clinical case series, this would include observations of defect healing, formation of new bone, and subsequent successful dental implant placement.
    • Biocompatibility Testing Standards: Adherence to established ISO 10993 standards for confirming safety (e.g., cytotoxicity tests, implantation tests).

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.