510(k) Data Aggregation

    K Number
    K141368

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-11

    (49 days)

    Product Code
    Regulation Number
    882.5250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuroPace® Burr Hole Cover is intended for use following cranial surgery to cover a 14 mm bur hole. Secondarily, the NeuroPace® Burr Hole Cover also can be used to support a 1.3 mm indwelling Jead.

    Device Description

    The NeuroPace® Burr Hole Cover (model 8110) includes a base (also referred to as a "retainer") that is screwed to the cranium (skull) using bone screws. The cap is pressed into the base covering the opening in the base and securing a single 1.3 mm lead. The Burr Hole Cover requires three bone screws (1.5 to 1.8 mm). The screws and driver are not included in the device's packaging. The contents of the unopened, undamaged package are sterile and non-pyrogenic. The NeuroPace® Burr Hole Cover is provided sterile (for single-use only) and consists of one model / size. The device is meant to be a permanent implant. The Burr Hole Cover incorporates materials commonly found in medical devices that are known to be biocompatible. The base is made from a synthetic polymer and the cap is made from silicone. The Burr Hole Cover is MR/CT scanning compatible and is sterilized using ethylene oxide gas (EtO). The Burr Hole Cover does not include any software, incorporate any medicinal substances or contain any color additives.

    AI/ML Overview

    The provided document, K141368, is a 510(k) premarket notification for the NeuroPace® Burr Hole Cover, Model 8110. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting extensive clinical trial data or detailed performance studies with acceptance criteria in the same way a PMA (Premarket Approval) might.

    Therefore, the information regarding acceptance criteria, specific study design elements (like sample sizes, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, training set details), is not explicitly present in this summary document. The document primarily relies on non-clinical testing and comparison to the predicate device to establish substantial equivalence.

    Here's a breakdown of what can be inferred or directly stated from the provided text, and what is not available:

    Acceptance Criteria and Device Performance

    The document doesn't explicitly list numerical acceptance criteria or performance metrics in a table format that would typically be associated with AI/software device evaluation. Instead, it states that the modified device was subjected to various non-clinical tests to demonstrate functional equivalence and safety.

    Category of TestingReported Device Performance/Conclusion
    Dimensional / GeometrySuccessfully performed, indicating the device meets its design specifications.
    Functionality at OperationSuccessfully performed, indicating the device functions as intended during use.
    Lead CompatibilitySuccessfully performed, confirming compatibility with the 1.3 mm indwelling lead.
    Lead MovementSuccessfully performed, indicating appropriate control or restriction of lead movement.
    RetentionSuccessfully performed, indicating the device maintains its position.
    Cranial RigiditySuccessfully performed, indicating the device provides adequate structural support to the cranium.
    BiocompatibilityMaterials are biocompatible and are the same as used in the predicate device.
    SterilityEvaluations confirmed non-pyrogenicity and product sterility, and that bioburden and residual levels demonstrated compliance with recognized performance standards. (Sterile (SAL 1 x 10^-6)).
    MR/CT CompatibilityYes.
    Shelf Life1 year.
    Substantial EquivalenceThe modified NeuroPace® Burr Hole Cover is substantially equivalent to the legally marketed predicate NeuroPace® Burr Hole Cover (K123163). No new issues of safety or effectiveness are raised.

    Study Details (Not Applicable/Available for this submission type)

    Since this is a 510(k) submission for a physical medical device (burr hole cover) and not a software or AI/ML-based device, most of the requested details related to performance evaluation against ground truth, expert consensus, and sample sizes for diagnostic accuracy are not applicable or not provided in this type of regulatory document. The focus is on non-clinical testing demonstrating physical and material properties, and functionality consistent with its intended use and predicate device.

    1. Sample size used for the test set and the data provenance: Not applicable in the context of a diagnostic dataset. The "test set" here refers to physical devices undergoing engineering and material testing. No specific sample sizes for these tests are detailed in this summary, nor is data provenance in the sense of patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic accuracy is not relevant for this device.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. "Ground truth" for this device would be established engineering specifications and material properties, against which the physical device is tested.
    7. The sample size for the training set: Not applicable, as no algorithm is being "trained."
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the K141368 document demonstrates substantial equivalence through non-clinical testing of the physical, material, and functional characteristics of the NeuroPace® Burr Hole Cover, Model 8110, compared to its predicate device. It does not involve AI/ML components, clinical studies for diagnostic accuracy, or associated performance metrics and ground truth evaluations often seen in software as a medical device (SaMD) submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123163

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-18

    (101 days)

    Product Code
    Regulation Number
    882.5250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuroPace® Burr Hole Cover is intended for use following cranial surgery to cover a 14 mm burr hole. Secondarily, the NeuroPace Burr Hole Cover also can be used to support a 1.3 mm indwelling lead.

    Device Description

    The NeuroPace® Burr Hole Cover (model 8110) (also referred to as "the device") includes a base (also referred to as a "retainer") that is screwed to the cranium (skull) using bone screws. The cap is an assembly comprised of a cap and a gasket. The cap is pressed into the base covering the opening in the base and securing a single 1.3 mm lead. The Burr Hole Cover requires three bone screws (1.5 to 1.8 mm).

    The screws and driver are not included in the device's packaging. The contents of the unopened, undamaged package are sterile and non-pyrogenic.

    The Burr Hole Cover is provided sterile (for single-use only) and consists of one model / size. The device is meant to be a permanent implant.

    The Burr Hole Cover incorporates materials commonly found in medical devices that are known to be biocompatible. The cap and base are made from a synthetic polymer and the gasket is made from silicone.

    The Burr Hole Cover is MR/CT scanning compatible and is sterilized using ethylene oxide gas (EtO).

    The Burr Hole Cover does not include any software, incorporate any medicinal substances or contain any color additives.

    AI/ML Overview

    The NeuroPace® Burr Hole Cover (model 8110) is a medical device designed to cover a 14 mm burr hole following cranial surgery and to support a 1.3 mm indwelling lead. The device's acceptance criteria and performance were assessed through non-clinical testing rather than clinical studies with human subjects.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes performance testing conducted to demonstrate compliance with product requirements. The acceptance criteria are implicit in the successful completion of these tests, indicating the device met the specified engineering and safety standards. Since the document does not present quantitative metrics for "performance" as one might expect from a diagnostic AI, the performance is stated as "compliance" or "demonstrated."

    Acceptance Criteria CategoryReported Device Performance
    Dimensional / GeometryCompliant (demonstrated through performance testing)
    Device UsabilityCompliant (demonstrated through performance testing)
    Lead CompatibilityCompliant (demonstrated through performance testing)
    Lead MovementCompliant (demonstrated through performance testing)
    Fixation / RetentionCompliant (demonstrated through performance testing)
    Cranial RigidityCompliant (demonstrated through performance testing)
    BiocompatibilityDemonstrated in accordance with ISO 10993-1
    SterilityCompliant (demonstrated through performance testing)
    PyrogenicityCompliant (demonstrated through performance testing)
    MR/CT Scanning CompatibilityCompliant (demonstrated through performance testing)
    Packaging / Shelf LifeCompliant (demonstrated through performance testing)

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical testing was deemed necessary to support substantial equivalence." This means there was no human test set in the traditional sense of a clinical study. The "test set" consisted of physical devices undergoing various laboratory and engineering tests. The data provenance is from non-clinical laboratory testing conducted by NeuroPace Inc.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Since no clinical testing was performed, there were no human experts establishing ground truth for a patient-based test set. The "ground truth" here is based on engineering specifications, regulatory standards, and established material science properties. The experts involved would have been the engineers, material scientists, and regulatory affairs personnel responsible for designing, testing, and submitting the device, operating under the established guidelines and standards (e.g., ISO 10993-1).

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical test set was used, there was no need for an adjudication method involving experts making clinical decisions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This device is a passive implantable medical device, not an AI or diagnostic software tool. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The device is a physical implant, not an algorithm or AI.

    7. The Type of Ground Truth Used

    The ground truth used for this device's evaluation was based on:

    • Engineering Specifications and Design Requirements: The device was designed to meet specific dimensions, material properties, and functional criteria (e.g., lead compatibility, fixation strength).
    • Regulatory Standards: Compliance with international and national standards for medical devices, such as ISO 10993-1 for biocompatibility, ISO 11607-1 for packaging, and various ASTM and AAMI standards for material testing and sterilization.
    • Predicate Device Characteristics: The device's performance was also assessed against the known characteristics and approved uses of the predicate device (Image-Guided Neurologics Burr Hole Cover, K001096) to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI or machine learning device, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    In summary, the NeuroPace® Burr Hole Cover is a physical implantable device, and its acceptance criteria and "performance" were established and demonstrated through rigorous non-clinical laboratory testing against established engineering specifications and regulatory standards, rather than through clinical trials, expert consensus, or AI performance metrics. Its substantial equivalence to a legally marketed predicate device was a key aspect of its clearance.