Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K230178

    Validate with FDA (Live)

    Date Cleared
    2023-10-19

    (269 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyOne Sky spirometer is intended to conduct diagnostic spirometry testing on adults and pediatric patients starting at age 4. The EasyOne Sky spirometer is used by general practitioners, specialists and healthcare professionals. The EasyOne Sky is used in hospitals, clinical settings, and in occupational medicine.

    Device Description

    The product EasyOne Sky (EOS) is a medical electrical equipment for spirometry testing. It consists of the hand-held TrueFlow Sensor FT (SeNe) comprising an ultrasonic flow sensor that conducts air flow measurements, the application EasyOne Mobile (EOMA), and the breathing mouthpiece EasyOne FlowTube (EOFT). The EOS can be used with the optional accessories EasyOne Filter FT and a nose clip.

    Key functions:

    • . The key function of the TrueFlow Sensor FT (SeNe) is to acquire patient flow data by measuring transit-times of ultrasonic pulses and to send data wireless (via Bluetooth Low Energy) to the host system. The measurements and data transmission are performed by the firmware of the SeNe defined as MDSW with clinical functions.
    • . The key function of the application EasyOne Mobile (EOMA) is to trigger the start of the measurement with the TrueFlow Sensor FT, to communicate with it and to visualize the data received from it.
    • . The EasyOne FlowTube (EOFT) is an individually packaged, single-patient-use breathing tube and is intended to canalize patient breath through the flow sensor tube. The EOFT is an essential accessory to EOS.
    AI/ML Overview

    The EasyOne Sky spirometer, as described in the provided 510(k) summary, underwent various non-clinical tests to demonstrate its substantial equivalence to its predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CriteriaAcceptance CriteriaReported Device Performance (Summary)
    PerformanceComplies with spirometry standards: ATS/ERS recommendation and guidelines, ISO 23747:2015 (Accuracy, Repeatability and Linearity) and ISO 26782:2009 (13 waveforms)The device meets permissible margins given in ATS guidelines and ISO 26782. The parameters for basic spirometry tests are the same as defined in ATS/ERS 2019 Table 9.
    Measurement Accuracy (Volume)±2% or 0.050LReported as conforming to this during comparison with predicate.
    Measurement Accuracy (Flow, except PEF)±2% or 0.020L/sReported as conforming to this during comparison with predicate.
    Measurement Accuracy (Flow PEF)±5% or 0.2L/sReported as conforming to this during comparison with predicate.
    Cleaning, DisinfectionCleaning & disinfection validation support that there is no loss of functionality.Non-clinical testing demonstrated compliance (implied by "Non-clinical testing demonstrated that the proposed device is at least as safe and effective as the predicate").
    BiocompatibilityBiological Evaluation and toxicological risk assessment to evaluate device's biological safety for the intended use, in accordance with ISO 10993-series and FDA guidance. Specific endpoints considered (ISO 10993-18:2020 Physical and/or chemical information, ISO 10993-5:2009 Cytotoxicity, ISO 10993-10:2021 Sensitization, ISO 10993-23:2021 Irritation or intracutaneous reactivity). Gas path testing according to ISO 18562-1:2017 (Biocompatibility evaluation), ISO 18562-2:2017 (Particulate matter emissions), ISO 18562-3:2017 (VOCs emissions).The EasyOne FlowTube was introduced with K161536 and EasyOne Filter FT with K221250. Non-clinical testing demonstrated compliance for overall biocompatibility as per the general statement of safety and effectiveness.
    Electrical Safety, Electromagnetic CompatibilityComplies with IEC 60601-1:2020 (General Requirements), IEC 60601-1-2:2020 (Electromagnetic Disturbances), IEC 60601-1-6:2020 CSV (Usability).Non-clinical testing demonstrated compliance for overall safety and effectiveness.
    Software, CybersecuritySoftware developed according to IEC 62304:2006 and IEC 82304-1:2016. FDA Guidance: Final Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005), Final Guidance Off-The-Shelf Software Use in Medical Devices (September 2019), Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2014), Radio Frequency Wireless Technology in Medical Devices (August 2013).Non-clinical testing demonstrated compliance for overall safety and effectiveness.

    2. Sample size used for the test set and the data provenance

    The document specifies performance testing in accordance with ISO 26782:2009 for 13 waveforms. This refers to a standardized set of flow-volume and volume-time waveforms designed to test spirometer performance across a range of physiological conditions. The provenance of these waveforms is standardized and not patient-specific. The document does not specify additional clinical test set sizes or their provenance (e.g., country of origin, retrospective/prospective). The assessment relies heavily on compliance with recognized performance standards rather than a separate clinical test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the performance testing is established by international spirometry standards (ATS/ERS/ISO), which define the expected values for the 13 waveforms used in testing. These standards are developed by expert committees, but the document does not mention individual experts specifically establishing ground truth for this particular device's test set, as it relies on compliance with pre-defined standard waveforms.

    4. Adjudication method for the test set

    No specific adjudication method (like 2+1, 3+1) is mentioned for a test set. The evaluation is based on the device's technical performance against established international standards for spirometry.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not conducted or reported in this 510(k) summary. The device is a diagnostic spirometer, not an AI-assisted diagnostic tool that interprets images or signals requiring human reader input for comparison.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone performance assessment was effectively done. The spirometer's core function is to measure and calculate spirometry parameters based on patient breathing. The "Performance" section explicitly states compliance with established spirometry standards (ATS/ERS/ISO), which are machine-based performance tests using standardized waveforms. The device's firmware performs the measurements and calculates parameters independently, and this is what was tested against international accuracy standards.

    7. The type of ground truth used

    The ground truth used for performance testing (accuracy, repeatability, linearity) is based on established spirometry standards and guidelines (ATS/ERS/ISO). Specifically, ISO 26782:2009 includes defined "reference waveforms" for evaluating spirometers, which serve as the ground truth against which the device's measurements are compared.

    8. The sample size for the training set

    The document does not provide information about a "training set" in the context of machine learning. The device is a traditional medical device that performs measurements and calculations based on physics (ultrasonic transit-time measurement) and programmed algorithms derived from clinical standards, not an AI/ML-based device that requires a training set of data.

    9. How the ground truth for the training set was established

    As there is no mention of a machine learning "training set," the concept of establishing ground truth for it is not applicable here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221250

    Validate with FDA (Live)

    Device Name
    EasyOne Filter
    Date Cleared
    2022-11-22

    (204 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyOne Filter is intended to be used in combination with NDD breathing mouthpieces to reduce bacteria, viruses and other particulates from the patient's exhaled air while performing flow measurements, such as spirometry tests. The EasyOne Filter is a single-use device and intended for single-patient use only.

    Device Description

    The EasyOne Filter is a viral, bacterial filter designed to be used with NDD breathing mouthpieces during flow measurements, e.g. for spirometry testing with NDD spirometers. The EasyOne Filter exists in two variants which differ in the connection interface to enable mounting on the EasyOne FlowTube. The EasyOne Filter is composed of two injection molded polymeric parts which enclose a filter medium. The EasyOne Filter is an optional accessory and reduces viruses, bacteria, and other particles that may be released by the patient and thus contaminate the test environment. The EasyOne Filter is non-sterile and for single use.

    AI/ML Overview

    The provided text is a 510(k) summary for the EasyOne Filter, a breathing circuit bacterial filter. It details the device's characteristics, comparison to predicate devices, and the results of non-clinical, and usability tests. However, it does not describe a study involving an AI/algorithm or human readers for diagnostic purposes (e.g., radiologists interpreting images). The device itself is a physical filter for spirometry, not a diagnostic AI tool.

    Therefore, many of the requested points, such as MRMC studies, standalone algorithm performance, number of experts for ground truth, and training set details, are not applicable to this type of medical device submission.

    Here's an attempt to answer the applicable points based on the provided text, and explicitly state when information is not present or not relevant to this specific device:

    1. A table of acceptance criteria and the reported device performance

    TestPurposeAcceptance CriteriaResult
    BiocompatibilityEvaluate device's biological safety for the intended use, in accordance with ISO 10993-5, ISO 10993-10 and FDA's corresponding guidance document.Biological Evaluation per ISO 10993-5 and ISO 10993-10Justification for cytotoxicity, sensitization, and irritation tests based on results of ViroMax filter (K063526), same filter body and filter cap materials used for EasyOne Filter as well as VBMax filter (K000654).
    Gas Path Emissions (Particulate Matter)Assess if airborne particulate is emitted into the gas stream in accordance with ISO 18562Measured concentrations of particulate matter ≤ 2.5 µm and ≤ 10 µm were compared to air quality guidelines in ISO 18562-2.The minimum, maximum, and average particulate concentrations are all below acceptable limits.
    Gas Path Emissions (Volatile Organic Compounds)Assess if airborne VOCs are emitted from the filter into the gas stream per ISO 18562Gas path emissions of volatile organic compounds per ISO 18562-3; Risk assessment per ISO 10993-17 and ISO 18562-1Exposure of the individual VOCs released are unlikely to result in toxicological effects. (Implicitly passed)
    Filtration EfficiencyEvaluate aerosol bacterial and viral removalBacterial filtration 99.999% (BFE); Viral filtration 99.999% (VFE)All results were acceptable.
    Dead SpaceDetermination of dead space According to ATS/ERS standard 2005Dead space is within the ATS/ERS recommendation of < 300mlPassed
    ConnectorVerify that the connection between EasyOne Filter and the breathing mouthpieces Spirette and EasyOne FlowTubeMechanical force is within 6 - 55N; Hold in place during forced expiration (18L/s); Resistance higher 4800hPa*s/LPassed
    (ATS) Standardization of Spirometry Compliance (System test with EasyOne Filter and reference devices)Performance testing and comparative testing of NDD spirometers used with and without EasyOne Filter demonstrated compliance with spirometry standards ISO 26782 and ATS/ERS 2019.Test scenario according to ISO 26782 Annex B; Resistance ≤ 1.5 cm H2O/L/s at 14L/sNDD spirometers comply with ISO 26782 when used with the optional EasyOne Filter.
    Human test (System test with EasyOne Filter and reference devices)Human subjects performed a FVL test session using NDD spirometers with and without use of the EasyOne Filter. For each subject, the difference between the test session with and without filter were calculated for the following parameters: FVC, FIVC, FEV1, PEF According to ATS/ERS standard 1994FVC & FIVC: 6% or 200 ml whichever is greater; FEV1: 6% or 200 ml whichever is greater; PEF: 15% or 0.50 L/min whichever is greaterNDD spirometers provide the same spirometry test results when used with or without EasyOne Filter.
    Calibration Checks (System test with EasyOne Filter and reference devices)Confirm that the implemented calibration check meets specifications when NDD spirometers are used with EasyOne Filter.Pass rate > 90%Calibration check of NDD spirometers can be performed with EasyOne Filter.
    ConfigurationConfiguration of NDD spirometers for use with optional EasyOne FilterEasyOne Air firmware V1.18 or higher and EasyOne Connect V3.9.3 or higher allows setup of spirometery testing with or without EasyOne Filter.Passed

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Filtration Efficiency: Not explicitly stated, but implies lab testing.
    • Human Test (System test): The document mentions "Human subjects performed a FVL test session". The number of subjects is not specified.
    • Data Provenance: Not specified, but given the manufacturer is based in Switzerland, it's likely to be European or internationally sourced lab/human data. The tests are general performance tests for a physical device, not tied to specific patient populations or retrospective/prospective clinical data in the way an AI diagnostic would be.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This device is a physical filter, not a diagnostic tool requiring expert interpretation or ground truth establishment in the context of diagnostic accuracy. The tests are based on established engineering and medical device standards (e.g., ISO, ATS/ERS).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device. Its function is to filter exhaled air. The "human test" was to confirm the device doesn't negatively impact spirometry measurements, not to assess human diagnostic performance with/without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical filter, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for this device exists in the form of established engineering and medical standards (e.g., ISO 10993 for biocompatibility, ISO 18562 for gas path emissions, ATS/ERS standards for spirometry). The performance of the filter is measured against specified physical and biological parameters. Not ground truth from medical diagnoses.

    8. The sample size for the training set

    • Not applicable. This is a physical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161534

    Validate with FDA (Live)

    Date Cleared
    2017-02-22

    (264 days)

    Product Code
    Regulation Number
    868.1890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EasyOne Pro/LAB is designed for conducting lung function measurements in general or specialist practices or in hospitals.

    EasyOne Pro/LAB can also be used in clinical settings in occupational medicine for performing lung function screenings or measurements.

    EasyOne Pro/LAB is used to conduct lung function measurements on adults and children starting at age 4, except measurements of diffusing capacity of the lung based on CO (DLCO), which can be performed on adults and children starting at age 6.

    Device Description

    The EasyOne Pro Respiratory Analysis System consists of two different device models: EasyOne Pro and EasyOne Pro LAB.

    The EasyOne Pro Respiratory Analysis System devices are pulmonary function testing devices. Both EasyOne Pro Respiratory Analysis System device models provide Spirometry and Single Breath CO Diffusion (DLCO) tests including Lung Volume Parameters. The device model EasyOne Pro LAB additionally provides the Nitrogen Multiple Breath Washout (MBW) method. The EasyOne Pro Respiratory Analysis System devices meet the ATS recommendations for accuracy and precision for Spirometry, DLCO and MBW tests.

    The EasyOne Pro Respiratory Analysis System devices can be used as a stand-alone system and can be connected to a network.

    The EasyOne Pro Respiratory Analysis System devices use the following sensors: An ultrasonic flow sensor to measure flow velocity, volume and molar mass of the gases that the patient inhales and exhales; a molar mass sensor which determines the helium content of the respired air for the DLCO test and the nitrogen concentration for the multiple-breath washout (MBW) test; a CO sensor (EasyOne Pro) or a combined CO/CO₂ sensor (EasyOne Pro LAB) to determine CO and CO2 content in breathing gas; a temperature and humidity sensor to collect environmental data.

    The EasyOne Pro Respiratory Analysis System devices are used in combination with test gases (DLCO gas for DLCO tests and O2 for MBW tests).

    The EasyOne Pro Respiratory Analysis System devices are used in combination with the single-patient use breathing tube ndd Spirette. For DLCO and MBW tests, the additional single-patient use accessories, the ndd DLCO or FRC Barriettes, are used.

    The single-patient use accessories prevent cross-contamination between patients. The Spirette prevents the passage of microorganisms into the inside of the flow sensor; the Barriettes prevent the passage of microorganisms into gas supply tubing.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the EasyOne Pro Respiratory Analysis System. It describes the device, its intended use, and compares it to predicate devices to demonstrate substantial equivalence. The document summarizes the testing performed to ensure the modified device meets performance and safety standards.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the EasyOne Pro Respiratory Analysis System devices "meet the ATS recommendations for accuracy and precision for Spirometry, DLCO and MBW tests." While specific numerical acceptance criteria (e.g., ±X% accuracy) for each parameter are not explicitly listed in the provided text, the general acceptance criterion is compliance with ATS recommendations.

    Parameter/TestAcceptance Criteria (ATS Recommendations)Reported Device Performance
    SpirometryATS recommendations for accuracy and precision (specific numerical values not provided in this document, but typically involve thresholds for flow and volume measurements, e.g., ±3% or 50mL for FEV1/FVC, etc.)"The EasyOne Pro Respiratory Analysis System devices meet the ATS recommendations for accuracy and precision for Spirometry...""Simulated Spirometry, DLCO and MBW testing, as well as testing with humans, confirmed that the modified EasyOne Pro Respiratory Analysis System devices meet the recommendations for accuracy and precision for Spirometry..."
    DLCO (Single Breath CO Diffusion)ATS recommendations for accuracy and precision (specific numerical values not provided in this document)"...and Single Breath CO Diffusion (DLCO) tests... The EasyOne Pro Respiratory Analysis System devices meet the ATS recommendations for accuracy and precision for... DLCO...""Simulated Spirometry, DLCO and MBW testing, as well as testing with humans, confirmed that the modified EasyOne Pro Respiratory Analysis System devices meet the recommendations for accuracy and precision for... DLCO..."
    MBW (Nitrogen Multiple Breath Washout)ATS recommendations for accuracy and precision (specific numerical values not provided in this document)"...The device model EasyOne Pro LAB additionally provides the Nitrogen Multiple Breath Washout (MBW) method. The EasyOne Pro Respiratory Analysis System devices meet the ATS recommendations for accuracy and precision for... MBW tests.""Simulated Spirometry, DLCO and MBW testing, as well as testing with humans, confirmed that the modified EasyOne Pro Respiratory Analysis System devices meet the recommendations for accuracy and precision for... MBW tests."
    Electromagnetic Compatibility (EMC)IEC 60601-1-2 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests)"The modified devices were tested to demonstrate conformance with the requirements for medical electrical equipment basic safety and essential performance of standards IEC 60601-1 and IEC 60601-1-2."
    Basic Safety & Essential PerformanceIEC 60601-1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)"The modified devices were tested to demonstrate conformance with the requirements for medical electrical equipment basic safety and essential performance of standards IEC 60601-1 and IEC 60601-1-2."
    Software Verification & ValidationIEC 62304 (Medical device software - Software life cycle processes)"Software verification and validation in accordance with IEC 62304 confirmed that the modified EasyOne Pro Respiratory Analysis System meets the specified criteria."
    BiocompatibilityISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process"Biocompatibility was evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process."

    Note: The document states that "For the modified devices, some of the labelled specifications were updated to reflect the technical capabilities. No changes were made to the components the specifications relate to." This implies there might be specific numerical specifications in other documentation not provided here.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: The document mentions "testing with humans" for Spirometry, DLCO, and MBW, but does not specify the numerical sample size for the human test set. It also mentions "Simulated Spirometry, DLCO and MBW testing."
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    Not applicable / Not specified in the provided text. This device is a measurement system and its "ground truth" would generally be established by calibrated reference standards and established physiological measurements, rather than expert consensus on interpretive tasks.

    4. Adjudication Method for Test Set

    Not applicable / Not specified in the provided text. As noted above, the "ground truth" for this type of device (a physiological measurement system) is based on reference standards and calibrated instruments, not on human interpretation or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study assesses how human readers improve with AI assistance, which is not relevant for a device that measures lung function clinically. The EasyOne Pro is a diagnostic measurement device, not an AI-based interpretive aid for human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    Yes, the testing described appears to be a standalone performance evaluation of the device's measurement capabilities. The "Simulated Spirometry, DLCO and MBW testing" would assess the algorithm and hardware performance independently, and the "testing with humans" would validate its clinical measurement accuracy against the ATS recommendations. The device itself is designed to provide direct physiological measurements.

    7. Type of Ground Truth Used

    The ground truth for the EasyOne Pro Respiratory Analysis System is established by technical standards and physiological recommendations, specifically the "ATS recommendations for accuracy and precision for Spirometry, DLCO and MBW tests." This would involve:

    • Calibrated reference standards: For simulated testing (e.g., known flow and volume signals, known gas concentrations).
    • Established physiological measurement methods: When testing with human subjects, the device's measurements would be compared against expected values or other validated methods to confirm compliance with ATS standards.

    8. Sample Size for Training Set

    Not applicable / Not specified. This device is a measurement system, not a machine learning algorithm that requires a "training set" in the conventional sense of AI models. Its underlying principles are based on known physics and physiological models, not data-driven learning from previous cases.

    9. How Ground Truth for Training Set Was Established

    Not applicable. As explained above, the device does not use a "training set" in the context of an AI/ML model. Its design and calibration are based on established scientific principles and engineering practices to meet specific performance standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161536

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (216 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyOne Air spirometer is intended for prescription use only to conduct diagnostic spirometry testing of adults and pediatric patients over 4 years old. The EasyOne Air spirometer is used by general practitioners, specialists, and health care professionals, in hospitals and clinical settings in occupational medicine.

    Device Description

    The ndd EasyOne Air is a diagnostic spirometer.

    In order to conduct simple diagnostic spirometry testing, the EasyOne Air is used in combination with a commercially available disposable breathing tube with integrated mouthpiece (EasyOne Flow Tube).

    The sensor is an ultrasound flow sensor that measures pulse transit-time to determine gas flow velocity and volume, as well as molar mass of the gas. The collected data is used for pulmonary function evaluation and data management. The results of the testing are stored in a database and reports can be displayed or printed.

    AI/ML Overview

    This document describes the EasyOne Air Spirometer. However, it does not contain the specific information required to complete all parts of your request, especially regarding a clinical study with acceptance criteria and reported device performance in the format you provided. The document primarily focuses on demonstrating substantial equivalence to a predicate device for FDA 510(k) clearance.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    The document mentions "Dynamic wave form testing and comparative testing with a commercially available ndd Spirometer confirmed that the EasyOne Air Spirometer meets the spirometry recommendations for accuracy and precision published by the American Thoracic Society (ATS)."

    However, it does not provide a table of specific acceptance criteria (e.g., specific thresholds for FVC accuracy) or the numerical reported device performance in a format that could be directly extracted into such a table. It only states that the device "meets the spirometry recommendations."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described is "Dynamic wave form testing and comparative testing," which implies laboratory testing and comparison against a predicate device, rather than a clinical study with human subjects whose data provenance would be relevant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and therefore not provided in the document. The testing was against established spirometry recommendations and comparison to a predicate device, not against ground truth established by experts interpreting outputs from the device in a clinical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided. No adjudication method is described because there's no expert interpretation of device outputs.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and therefore not provided. The EasyOne Air Spirometer is a diagnostic device for measuring lung function, not an AI-powered image analysis tool that would typically involve human readers and efficacy studies like MRMC.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes the performance of the "EasyOne Air Spirometer" itself, which is a standalone device in the context of spirometry. It measures and processes data without human input affecting the measurement itself. The results are then displayed for interpretation by healthcare professionals. So, in a sense, its primary function is "standalone" (algorithm only) in generating the spirometry measurements. However, the term "standalone performance" often refers to an AI algorithm's performance before human review in a diagnostic workflow, which isn't directly analogous here.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance claims appears to be:

    • ATS (American Thoracic Society) spirometry recommendations for accuracy and precision.
    • Comparison to the predicate device (ndd EasyOne Spirometer).

    8. The sample size for the training set

    This information is not applicable and therefore not provided. The EasyOne Air Spirometer is a measurement device, not an AI/ML model that requires a "training set" in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1