510(k) Data Aggregation

    K Number
    K030443

    Validate with FDA (Live)

    Date Cleared
    2003-05-09

    (87 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MASCT device is intended for use in the collection of nipple aspirate fluid for cytological testing. The collected fluid can be used in the determination and/or differentiation of normal versus premalignant versus malignant cells.

    Device Description

    The Nastech MASCT device is similar to the non-powered breast aspirators manufactured by Nastech Pharmaceutical Company, Inc. and Pro-Duct Health, Inc. used to elicit and collect nipple aspirate fluid (NAF) from the excretory ducts (tubuli lactiferi or galactophori). The device is comprised of a rigid polycarbonate cup in which the sample collection container is inserted. Negative pressure (vacuum) is produced by nonpowered (hand) actuations of a lever like handle. The expressed NAF sample is collected on a membrane filter and washed into the sample container using an appropriate fixative solution.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Nastech Pharmaceutical Company, Inc. Mammary Aspiration Specimen Cytology Test (MASCT) device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the study's stated objective to demonstrate the device's ability to collect sufficient samples for cytological analysis and to support the claim that the collected cells can be used to identify differences associated with premalignant or malignant conditions.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Sample Collection SuccessAll breasts evaluated yield nipple aspirate fluid (NAF).100% of breasts evaluated yielded nipple aspirate fluid.
    Sample Sufficiency for Cytological AnalysisAll collected samples are clinically useful and sufficient for cytological analysis by a pathologist.All samples were sufficient for cytological analysis; the pathologist was able to read all samples and obtain cytology results.
    Cytological Data QualityThe device enables the collection of cells that can be used to identify differences (normal, premalignant, malignant) through cytological examination.97% of samples were Class I (normal/benign); 3% were Class IIa (benign). These results support the claim for identifying cell differences.
    Sample Weight (Implicit)While not an explicit "acceptance criterion," the viability of collecting samples above a minimal weight threshold for practical handling/analysis is implied.Mean sample weight was 6.6 mg, range <1 to 37 mg. Even samples reported as <1 mg due to weighing variability were sufficient for cytological analysis.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 60 samples (from 30 patients, right and left breasts).
    • Data Provenance: The study was a "clinical study conducted by Nastech," implying it was prospective and likely conducted in the country where Nastech is based (USA, as inferred from the address). No specific country of origin for the data is explicitly stated beyond "clinical study by Nastech."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: One pathologist is explicitly mentioned as reading all samples.
    • Qualifications of Experts: The text only states "pathologist" and "trained individuals" for cytological examinations. No specific experience level (e.g., "10 years of experience") is provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable/Not explicitly stated. The text indicates that "the pathologist was able to read all of the samples and cytology results were obtained," implying a single pathologist's assessment formed the basis of the results. There is no mention of multiple experts or an adjudication process for resolving discrepancies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This device is a fluid collection device, not an AI or image analysis tool. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not relevant or applicable to the stated purpose of this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: No. This device is a manual fluid collection tool. Standalone algorithm performance is not applicable. The device's output (nipple aspirate fluid) requires human intervention (a pathologist) for interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The ground truth for the interpretability of the samples was established by pathological evaluation (cytological analysis) of the collected nipple aspirate fluid. The study primarily aimed to prove that the device successfully collects samples suitable for such analysis, not to validate the accuracy of cytological interpretation against a "true" disease state (e.g., biopsy). It showed that the yielded cytology results were either Class I or Class IIa (both benign).

    8. The Sample Size for the Training Set

    • Training Set Sample Size: The document does not mention a training set. This is not an AI/machine learning device, so the concept of a "training set" for algorithm development is not applicable. The clinical study described served as the validation (test) set for the device's performance in collecting samples.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set: Not applicable, as there was no training set for an algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K012088

    Validate with FDA (Live)

    Date Cleared
    2001-12-18

    (168 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MASCT device is intended for use in the collection of nipple aspirate fluid for laboratory cytological testing.

    Device Description

    The Nastech MASCT device is similar to the non-powered breast pumps Manufactured by Windy Hill Technology used to elicit and collect nipple aspirate fluid (NAF) from the excretory ducts (tubuli lactiferi or galactophori). The device is comprised of a rigid polycarbonate cup in which the sample collection container is inserted. Negative pressure (vacuum) is produced by non-powered (hand) actuations of a lever like handle. The expressed NAF sample is collected on a membrane filter and washed into the sample container using an appropriate fixative solution.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Nastech Mammary Aspiration Specimen Cytology Test (MASCT) device. However, it does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and the comprehensive study that proves the device meets those criteria. The document explicitly states that "Nastech conducted in-vitro and clinical studies to verify and validate the device design. The results of that testing supported the conclusion that the MASCT device is safe and effective for its intended use." However, it does not provide specific details about the acceptance criteria, the reported device performance, the study design (e.g., sample sizes, ground truth establishment, expert qualifications, adjudication methods), or any comparative effectiveness studies with human readers.

    Therefore, the following response will reflect the information available and indicate where details are missing.


    Description of Acceptance Criteria and Proving Study for the Nastech Mammary Aspiration Specimen Cytology Test (MASCT)

    The provided 510(k) summary indicates that Nastech conducted studies to support the safety and effectiveness of the MASCT device and its substantial equivalence to a predicate device. However, the document does not explicitly define specific numerical acceptance criteria and does not detail the results of the studies in terms of specific performance metrics against such criteria. The basis for acceptance appears to be the successful demonstration of substantial equivalence to a legally marketed predicate device (Windy Hill Technology DucPrep™ Breast Aspirator).

    1. Table of Acceptance Criteria and Reported Device Performance

    As the document does not present specific acceptance criteria or quantitative performance metrics, a table cannot be fully populated. The primary "acceptance criterion" implied by the 510(k) process is demonstrating substantial equivalence to a predicate device in terms of safety, effectiveness, and technological characteristics.

    Acceptance CriterionReported Device Performance
    SafetyDemonstrated safe in in-vitro and clinical studies. (Specific metrics not provided)
    EffectivenessDemonstrated effective for its intended use in in-vitro and clinical studies. (Specific metrics not provided)
    Substantial Equivalence (to Windy Hill Technology DucPrep™ Breast Aspirator)Demonstrated to share similar design, material, and operating characteristics, and intended use as the predicate device.
    Intended Use FulfilledSuccessfully collects nipple aspirate fluid for laboratory cytological testing. (Specific success rates not provided)

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Nastech conducted in-vitro and clinical studies," but does not provide any specific sample sizes for these studies, nor does it detail the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide any information regarding the number of experts used, their qualifications, or how ground truth was established for any test sets.

    4. Adjudication Method for the Test Set

    The document does not provide any information regarding an adjudication method for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention or describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. Given the nature of a device for collecting biological samples, such a study focusing on human reader improvement with AI would not be directly applicable, as the device's function is collection, not interpretation via AI.

    6. Standalone (Algorithm Only) Performance Study

    The document does not describe a standalone performance study. The MASCT device is a physical instrument for sample collection, not an algorithm. Therefore, an "algorithm only" performance study is not relevant to this device.

    7. Type of Ground Truth Used

    The document does not specify the type of ground truth used for any studies. For a device collecting samples for cytological testing, relevant ground truth might involve the clinical diagnosis, pathology reports from biopsies, or long-term patient outcomes, but this is not detailed.

    8. Sample Size for the Training Set

    The document does not provide any information regarding a training set or its sample size. This is consistent with the device being a physical collection instrument rather than an AI/ML-based diagnostic algorithm.

    9. How Ground Truth for the Training Set Was Established

    As no training set is mentioned or implied for this device, the document does not provide information on how ground truth for a training set was established.