510(k) Data Aggregation

    K Number
    K142104

    Validate with FDA (Live)

    Device Name
    NANOFUSE DBM
    Date Cleared
    2015-01-21

    (173 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NanoFUSE® DBM is indicated to be gently placed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure (i.e., the extremities, pelvis and posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. NanoFUSE® DBM must be used with autograft as a bone graft extender in the posterolateral spine. The product provides a bone graft substitute that remodels into the recipient's skeletal system.

    Device Description

    NanoFUSE® DBM is a malleable, putty-like, bone-void filler for use in general orthopedic applications. The product is comprised of human demineralized bone matrix (DBM) and synthetic calcium phosphor-silicate particulate material particles (45s5 bioactive glass), both coated with gelatin derived from porcine skin. These coated particles are packaged dry in a single use, polypropylene syringe (20 cc or 3 cc), double-wrapped in peel-back pouches, and final packaged in a dust cover paperboard carton. The 20 cc syringe will be filled with either of two different fill quantities of dry powder, identified as 10 cc or 5 cc final product volume. The 3 cc syringe will be filled with dry powder, identified as 2 cc final product volume. NanoFUSE® DBM is intended for single patient use only. At point of use, the surgeon will reconstitute the product with an appropriate sterile solution of his/her choice (sterile saline, water for injection, or autologous whole blood). The coated particles rehydrate in less than 30 seconds and do not require mixing to form a uniform paste or putty. The material is then gently extruded by the surgeon into the appropriate bone voids. NanoFUSE® DBM is progressively resorbed and replaced by host bone during the osteo-remodeling process.

    AI/ML Overview

    Here's an analysis of the provided text regarding the NanoFUSE® DBM device, focusing on acceptance criteria and supporting studies.

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device for marketing clearance. It does not typically include detailed clinical trial results or performance metrics in the same way a PMA (Pre-Market Approval) or a more extensive clinical study report would. Therefore, much of the requested information, particularly quantitative performance metrics and details about ground truth establishment, training sets, and MRMC studies, is not present in this document.


    Description of Acceptance Criteria and Study to Prove Device Meets Criteria

    The NanoFUSE® DBM device is a bone void filler intended for use in general orthopedic applications, specifically in bony voids or gaps of the skeletal system (extremities, pelvis, and posterolateral spine). The document describes a 510(k) submission, which aims to demonstrate that a new device is as safe and effective as a legally marketed predicate device. Therefore, the "acceptance criteria" are primarily related to showing substantial equivalence to the predicate, with specific focus on biological safety, performance characteristics, and the new indication for use in the posterolateral spine.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the 510(k) submission, the "acceptance criteria" are implied by the comparison to the predicate and the specific tests conducted. Quantitative performance metrics as commonly seen in diagnostic device evaluations are not explicitly stated for this type of device in this document.

    Acceptance Criteria (Implied by 510(k) Submission)Reported Device Performance (Summary from Document)
    Technological Characteristics EquivalenceThe applicant's NanoFUSE® DBM is stated to be identical to the currently legally marketed NanoFUSE® DBM with respect to technological characteristics (Form, Materials of Construction, Comparable Sizes). It is comprised of human demineralized bone matrix (DBM) and synthetic calcium phosphor-silicate particulate material particles (45s5 bioactive glass), both coated with gelatin derived from porcine skin. It is provided dry and reconstituted at the point of use into a paste-like, malleable form.
    BiocompatibilityBiocompatibility testing was conducted according to ISO 10993 to evaluate biological safety. (Specific results are not provided in this summary, but the implication is that it met the standards for substantial equivalence).
    Bench Testing PerformanceIn vitro bench testing was conducted to evaluate performance characteristics according to ISO 10993. (Specific results are not provided, but the implication is that it met the standards for substantial equivalence). The rehydration time (less than 30 seconds to form a fluid paste that progressively thickens and sets into a rubbery mass) is described as a performance characteristic.
    OsteoinductivityFor the predicate device, a "Cell bioassay" was used. For the applicant device, "in vivo athymic rat implant or cell bioassay" was specified. The intent is to demonstrate that the device retains osteoinductive potential.
    Sterility"Yes - Radiation" for both predicate and applicant, implying the sterility method and outcome are equivalent.
    Effectiveness for New Anatomic SiteThe primary focus of this 510(k) appears to be the expanded indication for use in the "posterolateral spine." This was evaluated in vivo. The document states it was "found to be substantially equivalent to cleared predicate devices" for this indication. (Specific clinical outcomes or direct comparative measures are not detailed in this summary).
    History of Safe and Effective Clinical UseThe device "has a history of safe and effective clinical use" (referring to the existing, similar NanoFUSE® DBM, likely the predicate itself).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in this 510(k) summary. The document mentions an "in vivo" evaluation for the posterolateral spine indication, and possibly an "in vivo athymic rat implant or cell bioassay" for osteoinductivity, but no details on the number of subjects/animals are provided.
    • Data Provenance: Not specified. Given the nature of a 510(k), it likely includes a combination of internal bench testing, animal studies, and potentially a summary of existing clinical data for the predicate device. Country of origin for data is not mentioned, nor is a definitive statement on retrospective or prospective nature of all data, though "in vivo" studies would be prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts & Qualifications: Not applicable/not specified. This document is not evaluating a diagnostic device that requires expert interpretation for a "ground truth" test set. The evaluation is primarily based on material properties, biocompatibility, and in vivo assessment of bone formation/integration.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This document does not describe a process that would require expert adjudication of a test set, as might be found in a diagnostic imaging study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, this type of study is not mentioned as it is typically relevant for diagnostic devices where human readers interpret medical images or data. NanoFUSE® DBM is a therapeutic/device implant, not a diagnostic tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Not applicable. This concept applies to AI algorithms. NanoFUSE® DBM is a physical medical device (bone void filler). The "device performance" refers to its biological and mechanical properties, not an algorithmic output.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for a bone void filler generally involves:
      • Histology/Pathology: For in vivo studies, this would confirm bone formation, material degradation, and biocompatibility in animal models. The document mentions "in vivo" evaluation for the spine and "in vivo athymic rat implant" for osteoinductivity, implying such histological assessment.
      • Biocompatibility Standards: Compliance with ISO 10993 standards serves as a "ground truth" for biocompatibility.
      • Bench Test Results: Measurable physical/chemical properties from in vitro bench testing.
      • Clinical Outcomes/Radiographic Evidence: While not detailed in this summary, a full clinical evaluation for a similar device would typically involve radiographic evidence of bone healing/integration and clinical follow-up for adverse events.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable/not specified. This concept is for AI/machine learning models. NanoFUSE® DBM is a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable. (See #8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K120279

    Validate with FDA (Live)

    Device Name
    NANOFUSE DBM
    Manufacturer
    Date Cleared
    2012-04-06

    (67 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NanoFUSE® DBM is indicated to be gently placed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. It is indicated to be placed into bony voids or gaps of the skeletal structure (e.g., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. The product provides a bone graft substitute that remodels into the recipient's skeletal system.

    Device Description

    NanoFUSE® DBM is a malleable, putty-like, bone-void filler. The product is comprised of human demineralized bone matrix (DBM) and synthetic calcium phosphor-silicate particulate material particles (45s5 bioactive glass), both coated with gelatin derived from porcine skin. These coated particles are packaged dry in a single use, polypropylene syringe (20 cc or 3 cc), double-wrapped in peel-back pouches, and final packaged in a dust cover paperboard carton. The 20 cc syringe will be filled with either of two different fill quantities of dry powder, identified as 10 cc or 5 cc final product volume. The 3 cc syringe will be filled with dry powder, identified as 2 cc final product volume. NanoFUSE® DBM is intended for single patient use only. At point of use, the surgeon will reconstitute the product with an appropriate sterile solution of his/her choice (sterile saline, water for injection). The coated particles rehydrate in less than 30 seconds and do not require mixing to form a uniform paste or putty. The material is then gently extruded by the surgeon into the appropriate bone voids. NanoFUSE® DBM is progressively resorbed and replaced by host bone during the osteo-remodeling process.

    AI/ML Overview

    Note that the provided text is a 510(k) summary for a medical device (NanoFUSE® DBM) seeking substantial equivalence to predicate devices, not a clinical study report. Therefore, it does not contain the detailed information about acceptance criteria, specific performance metrics, sample sizes for test sets, expert qualifications, adjudication methods, or direct comparative effectiveness studies with human readers that are typically found in a clinical study demonstrating performance against predefined acceptance criteria.

    The submission focuses on demonstrating substantial equivalence based on similar materials, design, intended use, and technological characteristics, not on proving specific performance metrics against a predetermined acceptance criterion through a clinical trial or a standalone performance study as described in your request.

    However, I can extract the relevant information that is present regarding the device's characteristics and the changes discussed in the 510(k).

    Here's an attempt to answer your questions based only on the provided document, highlighting where information is not present or not directly applicable to your request format:


    1. A table of acceptance criteria and the reported device performance

    This document does not present quantitative acceptance criteria or corresponding reported device performance metrics in the way a clinical study would. The summary focuses on substantial equivalence to a predicate device by comparing technological characteristics. The primary "performance" discussed is related to osteoinductivity testing, which is a screening method for the DBM component, not a final device performance metric with acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified in terms of a clinical test set. The document mentions testing "each lot of DBM" for osteoinductivity, using either an in vitro bioassay or an in vivo athymic rat model. There's no sample size given for these tests in the context of device performance.
    • Data Provenance: The DBM is human-derived, but the country of origin for the data (if it were a clinical study) is not mentioned. The athymic rat model suggests animal testing, not human clinical data directly. This is a regulatory submission from a US-based company (Alachua, FL).
    • Retrospective/Prospective: Not applicable to this type of regulatory submission. The osteoinductivity testing described is part of quality control for the raw material (DBM).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The document discusses in vitro bioassays and in vivo athymic rat models for osteoinductivity, not human expert assessment of a test set. There's no "ground truth" established by human experts in the context of device performance as you describe.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no human expert assessment or adjudication described in this document for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone void filler, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The closest to "ground truth" mentioned pertains to the osteoinductivity of the DBM component:

    • Osteoinductivity: Assessed through in vitro bioassays or an in vivo athymic rat model. The document states, "Results from this bioassay were correlated to the athymic rat model." This suggests these animal/lab models serve as the "truth" for osteoinductive potential.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that uses a training set.

    9. How the ground truth for the training set was established

    Not applicable.


    Summary of Device and Regulatory Context:

    The NanoFUSE® DBM is a bone void filler indicated for use in bony voids or gaps of the skeletal system. The 510(k) submission (K120279) primarily addresses a change in the osteoinductivity testing method for the DBM component used in the device. The applicant seeks to use either the original in vitro bioassay or an in vivo athymic rat model. The FDA's determination is that this change does not alter the fundamental scientific technology and thus the device remains substantially equivalent to its predicate. The document focuses on material composition, intended use, and manufacturing processes rather than performance metrics from a comparative clinical trial.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110976

    Validate with FDA (Live)

    Date Cleared
    2011-05-03

    (26 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Origen DBM® with Bioactive Glass (NanoFUSE® DBM) is indicated to be gently placed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. It is indicated to be placed into bony voids or gaps of the skeletal structure (e.g., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. The product provides a bone graft substitute that remodels into the recipient's skeletal system.

    Device Description

    Origen DBM® with Bioactive Glass (also registered with the FDA as NanoFUSE® DBM) is a malleable, putty-like, bone-void filler for use in general orthopedic applications. The product is comprised of human demineralized bone matrix (DBM) and synthetic calcium phosphor-silicate particulate material particles (45s5 bioactive glass), both coated with gelatin derived from porcine skin. These coated particles are packaged dry in a single use, polypropylene syringe (20 cc or 3 cc), double-wrapped in peel-back pouches, and final packaged in a dust cover paperboard carton. The 20 cc syringe will be filled with either of two different fill quantities of dry powder. identified as 10 cc or 5 cc final product volume. The 3 cc syringe will be filled with dry powder, identified as 2 cc final product volume. Origen DBM® with Bioactive Glass (NanoFUSE® DBM) is intended for single patient use only. At point of use, the surgeon will reconstitute the product with an appropriate sterile solution of his/her choice (sterile saline, water for injection). The coated particles rehydrate in less than 30 seconds and do not require mixing to form a uniform paste or putty. The material is then gently extruded by the surgeon into the appropriate bone voids. Origen DBM® with Bioactive Glass (NanoFUSE® DBM) is progressively resorbed and replaced by host bone during the osteo-remodeling process.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Origen DBM® with Bioactive Glass / NanoFUSE® DBM). It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not include detailed information about specific acceptance criteria and a study demonstrating the device meets those criteria, as typically found in clinical study reports or more detailed verification and validation documents.

    The document primarily focuses on demonstrating that the modified manufacturing process for the applicant device does not affect its safety or effectiveness compared to the predicate device. It states, "The minor change in manufacturing process does not alter the fundamental scientific technology of the device and has been shown to have no effect on the safety or effectiveness of the device."

    Therefore, based solely on the provided text, I cannot complete many of the requested sections as the specific "study that proves the device meets the acceptance criteria" in the way one might expect for a diagnostic or AI-driven device is not present. The document focuses on showing equivalence rather than a new standalone performance study with detailed metrics.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of quantitative metrics for performance. The overall acceptance criterion is "substantial equivalence" to the predicate device, implying that its safety and effectiveness are comparable.
    • Reported Device Performance: The document only states that the modification "has been shown to have no effect on the safety or effectiveness of the device." No specific performance metrics (e.g., strength, resorption rate, osteoinductivity quantification) or their values are reported in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable/Not mentioned. The document describes a manufacturing process change and asserts equivalence. It does not refer to a test set of data in the context of a performance study with a specific sample size.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not mentioned. Ground truth establishment is not relevant to this type of 510(k) submission focused on a manufacturing change for a bone void filler.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a bone void filler, not a diagnostic imaging device involving human readers or AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable/Not mentioned.

    8. The sample size for the training set

    • Not applicable/Not mentioned. This is not a machine learning model.

    9. How the ground truth for the training set was established

    • Not applicable/Not mentioned.

    Summary regarding the "Study that proves the device meets the acceptance criteria":

    The core of this 510(k) submission is to demonstrate Substantial Equivalence to a legally marketed predicate device, specifically after a minor manufacturing process modification. The "study" (or rather, the justification) provided is that this manufacturing change results in a thicker paste-like form at initial extrusion but does not fundamentally alter the scientific technology or affect the safety or effectiveness of the device.

    The document states:

    • "The subject of this Special 510(k), the applicant version of Origen DBM® with Bioactive Glass (NanoFUSE® DBM), is produced under a modified manufacturing process to the legally marketed Origen DBM® with Bioactive Glass."
    • "Origen DBM® with Bioactive Glass (NanoFUSE® DBM) has the same intended uses and indications and similar technological characteristics as the currently cleared version of Origen DBM® with Bioactive Glass."
    • "The minor change in manufacturing process does not alter the fundamental scientific technology of the device and has been shown to have no effect on the safety or effectiveness of the device."
    • It also notes that "Each lot of DBM used in manufacturing the applicant version of Origen DBM® with Bioactive Glass (NanoFUSE® DBM) will be screened for osteoinductivity in an in vitro assay," which implies an ongoing quality control measure, not a specific performance study for the 510(k) itself to demonstrate new acceptance criteria.

    In essence, the "study" for this Special 510(k) is the demonstration that the manufacturing change does not negatively impact the previously established safety and effectiveness of the predicate device, thereby maintaining substantial equivalence. Specific quantitative performance data comparing the new and predicate devices is not detailed in this summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062459

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-02-26

    (187 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Origen™ DBM with Bioactive Glass is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be placed into bony voids or gaps of the skeletal system (e.g., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone graft substitute that remodels into the recipient's skeletal system.

    Device Description

    Origen™ DBM with Bioactive Glass is a malleable, putty-like bonevoid filler. The product is comprised of human demineralized bone matrix (DBM) and synthetic calcium phosphor-silicate particulate material particles, both coated with gelatin derived from porcine skin. These coated particles are packaged dry in a single use, polypropylene syringe (20 cc or 5 cc), double-wrapped in peel-back pouches, and final packaged in a dust cover paperboard carton. The 20 cc syringe will be filled with either of two different fill quantities of dry powder, identified as 10 cc or 5 cc final product volume. The 5 cc syringe will be filled with two additional fill quantities of dry powder, identified as 2.5 cc or 1 cc final product volume. Origen™ DBM with Bioactive Glass is intended for single patient use only. At point of use, the surgeon will reconstitute the product with an appropriate sterile solution of his/her choice (WFI, sterile normal saline). The coated particles rehydrate in less than 30 seconds and do not require mixing to form a uniform paste or putty. The material is then gently extruded by the surgeon into the appropriate bone voids. Origen™ DBM with Bioactive Glass is progressively resorbed and replaced by host bone during the osteo-remodeling process.

    AI/ML Overview

    The provided text (K062459) is a 510(k) Summary for a medical device called "Origen™ DBM with Bioactive Glass." This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing specific acceptance criteria and proving performance through a standalone clinical study with defined endpoints and statistical significance.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets them, particularly in the context of clinical performance metrics like sensitivity, specificity, accuracy, or human reader improvement, is not present in this document.

    Here is what the document does describe in relation to non-clinical performance and substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This document does not present a table of acceptance criteria for a specific performance metric (e.g., accuracy, sensitivity) and then report the device's performance against it. Instead, it asserts substantial equivalence based on material composition, intended use, and comparative non-clinical studies.

    Acceptance Criterion (Not explicitly stated as such, but inferred from the equivalence argument)Reported Device Performance (as presented for equivalence)
    Osteoinductivity: DBM component must exhibit osteoinductive potential.Each lot of DBM is screened for osteoinductivity in an in vitro assay, which has been correlated to the athymic rat model.
    Bone Formation & Healing: Performance comparable to autogenous bone graft and predicate devices in promoting bone formation and healing.GLP compliant study comparing bone formation and subsequent healing with Origen™ to autogenous bone graft and predicate devices (Novabone K021336 and Osteofil Paste K043420). The study "indicates that Origen™ DBM with Bioactive Glass is substantially equivalent to the predicate devices."
    Viral Reduction: Demineralization process effectively eliminates or reduces viruses.A viral reduction study demonstrated the demineralization process effectively eliminates or reduces viruses tested (BVDV, HAV, HIV-1, PPV, PrV).
    Biocompatibility: Human demineralized bone used in the product must be biocompatible.Stated that "Human demineralized bone used in Origen™ DBM with Bioactive Glass is single-donor processed and biocompatible."

    2. Sample Size Used for the Test Set and Data Provenance:

    • For the "bone formation and healing" study: The document does not specify the sample size used in this GLP compliant study.
    • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective for the "bone formation and healing" study, beyond it being a "GLP compliant study."
    • For the "viral reduction study": The "test set" would be the viral models used. The document lists five specific viruses: bovine diarrhea virus (BVDV); hepatitis A virus (HAV); human immunodeficiency virus type 1 (HIV-1), porcine parvovirus (PPV); and pseudorabies virus (PrV).
    • Data Provenance (Viral Reduction): Conducted by a CLIA certified testing laboratory (location not specified).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This information is not provided in the document. The studies referenced are non-clinical (animal model, in vitro assay, viral reduction). Clinical ground truth established by experts is not described.

    4. Adjudication Method for the Test Set:

    • This information is not provided as the studies are not clinical trials requiring adjudication of endpoints by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done as this is a bone void filler device, not an imaging or diagnostic AI device that would involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • This concept is not directly applicable. Origen™ DBM with Bioactive Glass is a physical medical device (bone void filler) used by a surgeon, not an algorithm. The "performance" is its biological effect (bone formation/healing), not an algorithmic output. The studies mentioned (osteoinuctivity assay, bone formation/healing study, viral reduction) are "standalone" in the sense they evaluate the biological/physical properties of the product itself.

    7. Type of Ground Truth Used:

    • For osteoinductivity: In vitro assay correlated to the athymic rat model.
    • For bone formation and healing: Comparison to autogenous bone graft and predicate devices, implying assessment of osseous tissue formation or healing outcomes in a relevant animal model (though the specific outcome measures are not detailed).
    • For viral reduction: Direct measurement of viral load reduction/elimination following the demineralization process.

    8. Sample Size for the Training Set:

    • This concept is not applicable as Origen™ DBM with Bioactive Glass is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • This concept is not applicable for the reasons stated above.

    In summary: The provided 510(k) summary focuses on demonstrating "substantial equivalence" of the Origen™ DBM with Bioactive Glass to existing legally marketed predicate devices based on material composition, intended use, and non-clinical studies. It does not present detailed acceptance criteria for clinical performance or a standalone clinical study to prove those criteria are met in the way one would for a diagnostic AI device.