510(k) Data Aggregation

    K Number
    K141771

    Validate with FDA (Live)

    Date Cleared
    2014-11-07

    (129 days)

    Product Code
    Regulation Number
    866.6040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prosigna® Breast Cancer Prognostic Gene Signature Assay is an in vitro diagnostic assay which is performed on the NanoString nCounter® Dx Analysis System using FFPE breast tumor tissue previously diagnosed as invasive breast carcinoma. This qualitative assay utilizes gene expression data, weighted together with clinical variables to generate a risk category and numerical score, to assess a patient's risk of distant recurrence of disease.

    The Prosigna Breast Cancer Prognostic Gene Signature Assay is indicated in female breast cancer patients who have undergone surgery in conjunction with locoregional treatment consistent with standard of care, either as:

    1. A prognostic indicator for distant recurrence-free survival at 10 years in postmenopausal women with Hormone Receptor-Positive (HR+), lymph node-negative, Stage I or II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors.

    2. A prognostic indicator for distant recurrence-free survival at 10 years in postmenopausal women with Hormone Receptor-Positive (HR+), lymph node-positive (1-3 positive nodes), Stage II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. The device is not intended for patients with 4 or more positive nodes

    Device Description

    The required components for the Prosigna™ Breast Cancer Prognostic Gene Signature Assay include the RNA Isolation kit (manufactured by Roche), Prosigna reagents (Reference Sample, CodeSet, Prep Pack, Cartridge(s) and Prep Plate) and the instruments that comprise the nCounter Dx Analysis System; the Prep Station and Digital Analyzer.

    The assay requires microdissection of tumor from FFPE biopsies, isolation of RNA using a Roche RNA isolation kit, transfer of RNA to PCR tubes for hybridization before placing onto the prep station. Two sets of probes specific to each of 58 RNAs are added to the hybridization reaction. These consist of biotin-labeled magnetic probes to purify the RNAs and capture them on the assay cartridge and fluorescent "barcode" probes to detect and quantify individual RNAs. The patient sample and probes are pipetted automatically into the Prosigna test cartridge by the Prep Station. The prep station uses magnetic bead capture and washing to remove excess RNA and un-hybridized probes. The isolated and hybridized RNA species are then bound via biotin on the capture probe randomly to streptavidin on the cartridge. The fluorescent molecules are then aligned on the cartridge by addition of an electric current. The cartridge is then transferred to the Digital Analyzer where the cartridge is scanned and digital analysis software is used to count the number of each RNA species present. The amount of each RNA is then put into a proprietary algorithm to produce a Prosigna score.

    The test output is a patient specific report which includes a Prosigna score (0-100) and risk category (low/intermediate/high).

    AI/ML Overview

    This document describes the FDA's decision to clear the Prosigna™ Breast Cancer Prognostic Gene Signature Assay (K141771) based on its substantial equivalence to a predicate device (K130010). The submission in question for K141771 relates to a modification of the device configuration and software, specifically introducing a 4-sample kit configuration and updating the instrument software. Therefore, the acceptance criteria and study data presented primarily focus on demonstrating that these modifications do not negatively impact the performance established by the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For K141771, the primary acceptance criterion related to device performance was to demonstrate that the new 4-sample kit configuration and updated software produced results substantially equivalent to the previously cleared 10-sample kit configuration.

    Acceptance CriteriaReported Device Performance (K141771)
    Equivalence of Prosigna Scores between 4-sample and 10-sample kit configurationsDeming Regression: Data was linear over the range of the assay with no outliers between the two methods, indicating substantial equivalence. No value deviated by more than 2 units from the average score when run using either configuration. Bland-Altman Analysis: Variations in data using the 4-kit versus the 10-kit configuration did not bias results.
    No change in risk categorization (Low/Intermediate/High) between 4-sample and 10-sample kit configurationsNo changes in risk categorization occurred between the 4-kit and 10-kit configurations.

    Note: The document explicitly states "See Predicate Device K130010 for Analytical Performance Data" and "See Predicate Device K130010 for Clinical Performance Data." This means that the core analytical and clinical performance acceptance criteria and their supporting studies were primarily established during the clearance of the predicate device (K130010). The K141771 submission focused on demonstrating that changes to the device (new kit configuration, software) did not degrade this established performance.

    2. Sample Size Used for the Test Set and Data Provenance

    For the specific comparison study outlined in K141771 (comparing the 4-sample and 10-sample kits):

    • Sample Size: Forty (40) samples of RNA previously extracted from FFPE breast tissue were tested with both configurations.
    • Data Provenance: The document does not specify the country of origin of the data. It indicates the samples were "previously extracted from FFPE breast tissue," implying the data was retrospective in nature for this specific comparison.

    For the original clinical performance data (referenced as part of K130010):

    • Sample Size: "a study with over 1007 patient samples associating Prosigna score with long-term outcome, defined by distance recurrence free survival at 10 years (DRFS) (Table 2)."
    • Data Provenance: Not specified in this document for the 1007 patient sample study, but it would have been part of the K130010 submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not directly applicable to the K141771 submission's comparison study, as the ground truth was the Prosigna score itself, as generated by the instrument. The study compared the scores generated by two different configurations of the same device.

    For the original clinical validation (referenced from K130010), the ground truth for "distant recurrence-free survival at 10 years (DRFS)" would typically be established based on patient follow-up data, which is objective outcomes data rather than subjective expert consensus.

    4. Adjudication Method for the Test Set

    Not applicable to the K141771 comparison study as there was no expert review or human adjudication involved in generating the Prosigna scores. The comparison was between two automated outputs of the device.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device (Prosigna Assay) is an in vitro diagnostic (IVD) test that provides an automated, objective score. It is not an imaging AI diagnostic system designed to assist human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    The Prosigna Assay is inherently a standalone algorithm. It takes gene expression data and clinical variables as input and produces a risk category and numerical score as output without human intervention in the scoring process. The performance of this standalone algorithm was presumably established and validated in the clinical studies for the predicate device (K130010), which involved "a study with over 1007 patient samples associating Prosigna score with long-term outcome, defined by distance recurrence free survival at 10 years (DRFS)."

    7. The Type of Ground Truth Used

    The ground truth for the clinical validation of the Prosigna Assay (for K130010) was outcomes data, specifically "distant recurrence free survival at 10 years (DRFS)". This is a definitive clinical outcome, not expert consensus or pathology review of the assay results themselves.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It mentions "Three risk categories (low, intermediate and high) were defined based on a study with over 1007 patient samples associating Prosigna score with long-term outcome...". This "study" likely encompassed both the development/training and validation of the algorithm, but the specific breakdown is not provided here.

    9. How the Ground Truth for the Training Set was Established

    The ground truth for the training set (or the 1007 patient study mentioned) was established based on long-term clinical outcomes, specifically "distance recurrence free survival at 10 years (DRFS)". This means that for each patient in the dataset, their disease recurrence status was tracked for 10 years after initial treatment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130010

    Validate with FDA (Live)

    Date Cleared
    2013-09-06

    (246 days)

    Product Code
    Regulation Number
    866.6040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prosigna™ Breast Cancer Prognostic Gene Signature Assay is an in vitro diagnostic assay which is performed on the NanoString nCounter® Dx Analysis System using FFPE breast tumor tissue previously diagnosed as invasive breast carcinoma. This qualitative assay utilizes gene expression data, weighted together with clinical variables to generate a risk category and numerical score, to assess a patient's risk of distant recurrence of disease.

    The Prosigna Breast Cancer Prognostic Gene Signature Assay is indicated in female breast cancer patients who have undergone surgery in conjunction with locoregional treatment consistent with standard of care, either as:

    1. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with Hormone Receptor-Positive (HR+), lymph node-negative, Stage I or II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors.

    2. A prognostic indicator for distant recurrence-free survival at 10 years in post-menopausal women with Hormone Receptor-Positive (HR+), lymph node-positive (1-3 positive nodes), Stage II breast cancer to be treated with adjuvant endocrine therapy alone, when used in conjunction with other clinicopathological factors. The device is not intended for patients with 4 or more positive nodes.

    Special Conditions for Use: Prosigna is not intended for diagnosis, to predict or detect response to therapy, or to help select the optimal therapy for patients.

    Device Description

    Used together, the Prosigna™ Breast Cancer Prognostic Gene Signature Assay and nCounter Dx Analysis System are a nucleic acid hybridization and image analysis system based upon coded probes designed to detect the messenger RNA transcribed from 58 genes. The test input is purified RNA from FFPE breast tumor specimens which are acquired from surgical resection. The Prosigna assay uses gene-specific probe pairs that hybridize directly to the mRNA transcripts in solution. The nCounter Dx Analysis System delivers direct, multiplexed measurements of gene expression through digital readouts of the relative abundance of the mRNA transcripts. Specifications are included as part of the Prosigna Assay to control for sample quality, RNA quality, and process quality. Prosigna simultaneously measures the expression levels of 50 genes used in the PAM50 classification algorithm (Parker et al., 2009), 8 housekeeping genes used for signal normalization, 6 positive controls, and 8 negative controls in a single hybridization reaction, using nucleic acid probes designed specifically to those genes. The Prosigna assay utilizes prototypical expression profiles (centroids) which are associated with and define each of the four PAM50 molecular subtypes of breast cancer. The software algorithm produces a Prosigna Score (referred to as ROR Score or Risk of Recurrence Score in the literature (Dowsett et al., 2013)) based on the similarity of the expression profile to each PAM50 molecular subtype, as well as the gross pathological tumor size and a proliferation score computed from a subset of genes. Three risk categories (low, intermediate and high) were defined based on a study with over 1007 patient samples associating Prosigna score with longterm outcome.

    The required components for the Prosigna Assay include the RNA Isolation kit (manufactured by Roche), Prosigna reagents (Reference Sample, CodeSet, Prep Pack, Cartridge(s) and Prep Plate) and the instruments that comprise the nCounter Dx Analysis System; the Prep Station and Digital Analyzer.

    The test output is a patient specific report which includes a Prosigna score (0-100) and risk category (low/intermediate/high).

    AI/ML Overview

    Let's break down the acceptance criteria and the study that proves the device meets those criteria for the Prosigna™ Breast Cancer Prognostic Gene Signature Assay, based on the provided FDA 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary for the Prosigna assay demonstrates clinical performance relative to its intended use as a prognostic indicator for distant recurrence-free survival (DRFS) at 10 years. The acceptance criteria are implicit in the statistical significance and magnitude of the prognostic information provided by the Prosigna Score, both as a continuous variable and when categorized into risk groups.

    Acceptance Criteria (Implicit)Reported Device Performance (ABCSG-8 Study)
    Primary Clinical Performance: Prosigna Score must add "significant prognostic information" for DRFS over and above existing clinical and treatment variables (CTS).For DRFS at 10 years: - Prosigna Score as a continuous variable: Added significant prognostic information (p < 0.0001) over and above CTS (ΔLR χ² = 53.49, critical value 3.84 for df=1). - Prosigna Score using risk groups: Added significant prognostic information (p < 0.0001) over and above CTS (ΔLR χ² = 34.12, critical value 5.99 for df=2).
    Risk Group Separation (Node-Negative): Pre-defined Prosigna Score cutoffs must separate node-negative patients into three risk groups (Low, Intermediate, High) with statistically different outcomes at 10 years DRFS.Node-Negative Population: - Intermediate vs. Low Prosigna Score: Hazard Ratio = 2.60 (95% CI: 1.44 - 4.70, p = 0.0015). Statistically significantly greater than 1. - High vs. Low Prosigna Score: Hazard Ratio = 3.96 (95% CI: 2.18 - 7.20, p < 0.0001). Statistically significantly greater than 2. - Estimated Percent Without Recurrence at 10 years: - Low Risk: 96.6% [94.4% - 97.9%] - Intermediate Risk: 90.4% [86.3% - 93.3%] - High Risk: 84.3% [78.4% - 88.6%] The low-risk group had 10-year DRFS well above 90% and was separated from the high-risk group by more than a 10% probability of recurrence.
    Risk Group Separation (Node-Positive, 1-3 nodes): Pre-defined Prosigna Score cutoffs must separate node-positive patients into two risk groups (Low, High) with statistically different outcomes at 10 years DRFS.Node-Positive (1-3 nodes) Population: - High vs. Low Prosigna Score: Hazard Ratio = 4.22 (95% CI: 1.98 - 9.00, p = 0.0002). Statistically significantly greater than 2. - Estimated Percent Without Distant Recurrence at 10 years: - Low Risk: 94.2% [88.1% - 97.2%] - High Risk: 75.8% [68.9% - 81.4%] The low-risk group had 10-year DRFS well above 90% and was separated from the high-risk group by more than a 10% probability of recurrence.
    Analytical Precision/Reproducibility: Device must reliably measure differences in Prosigna Score.- Standard deviation of Prosigna Score from 5 pooled RNA samples: < 1 Prosigna Score unit across 3 sites, 3 reagent lots, and 108 measurements. - Total variability (tissue and RNA Processing): 2.9 Prosigna Score units. This demonstrates the assay can reliably measure a difference of 6.75 Prosigna Score units with 95% confidence. - Concordance in categorical risk classifications across 43 tissue samples in Tissue Reproducibility study: average > 90%.
    Clinical Utility (Small Score Changes): Small changes in Prosigna Score (5-10 units) should be statistically significant for time to distant recurrence.C-index analysis showed statistically significant utility for small changes in Prosigna Score (P<0.05) for 5-10 Prosigna Score units. A difference of 7 Prosigna Score units is both statistically reproducible and clinically meaningful.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Validation):
      • The primary clinical validation study was ABCSG-8, with 1,478 patients available for analysis after QC failures (out of 1,620 tissues initially available).
      • A prior study, TransATAC, was used to select the Prosigna Score cut-offs for risk categories. This study also demonstrated continuous relation to DRFS.
    • Data Provenance: The 510(k) summary states that the validation population for the Prosigna Assay utilized "Treatment arms from a randomized trial conducted in Europe" with a "prospective retrospective study design". Both TransATAC and ABCSG-8 samples were independent from those used to train the Prosigna algorithm.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test set. However, a "pre-defined pathology review criteria for adequate tumor" was applied to the ABCSG-8 tissue samples, which would typically involve expert pathologists. The clinical endpoints (distant recurrence, distant recurrence-free survival) are outcome data derived from patient follow-up within the clinical trial, rather than ground truth established by expert consensus on images.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method for establishing specific diagnoses or outcomes on a case-by-case basis. The outcomes (distant recurrence, death) are clinical events recorded during patient follow-up, typical of a clinical trial. Pathology review criteria were applied to screen samples, but this is a quality control step, not an adjudication of a device output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No MRMC comparative effectiveness study involving human readers and AI assistance is mentioned. The Prosigna assay is a gene expression profiling test system, providing a numeric score and risk category directly. It is not designed to assist human readers in interpreting images or other data in the way an AI diagnostic tool might. Its primary purpose is to add prognostic information to clinical variables.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the clinical performance described is a standalone (algorithm only) performance. The Prosigna Assay generates a risk category and numerical score directly from gene expression data and clinical variables. The studies evaluate the prognostic value of this score and risk category on patient outcomes (DRFS) without human interpretation of the assay's raw output. The output is intended for healthcare professionals to consider alongside other clinicopathological factors in treatment decisions.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for the clinical validation studies (TransATAC and ABCSG-8) was patient outcomes data, specifically:

    • Distant Recurrence-Free Survival (DRFS) at 10 years.
    • Incidence of distant recurrence (events).
      The clinical endpoints were derived from long-term follow-up of patients in these retrospective clinical trials.

    8. The Sample Size for the Training Set

    The document explicitly states that "Both the TransATAC and ABCSG-8 study samples were independent from those samples used to train the Prosigna algorithm." However, the exact sample size for the training set (i.e., the cohort(s) used to develop the PAM50 classification algorithm and subsequently to define the Prosigna Score and risk categories prior to its validation) is not specified in this 510(k) summary. It references "Parker et al., 2009" for the PAM50 classification algorithm and "Dowsett et al., 2013" for the Prosigna Score literature, which would contain details on the training cohorts used for algorithm development and initial risk stratification. The summary notes that "Three risk categories (low, intermediate and high) were defined based on a study with over 1007 patient samples associating Prosigna score with long-term outcome." This "study with over 1007 patient samples" likely refers to the TransATAC study, which was used for defining cut-offs and therefore might be considered part of an extended training/calibration phase before independent validation in ABCSG-8.

    9. How the Ground Truth for the Training Set Was Established

    Similar to the validation set, the ground truth for the (unspecified) training set would have been established using patient outcomes data, specifically clinical endpoints such as:

    • Distant Recurrence-Free Survival
    • Overall Survival
    • Other breast cancer-specific endpoints.

    The "PAM50 classification algorithm" itself was "based on intrinsic subtypes" (Parker et al., 2009), implying that molecular profiling of tumors was correlated with clinical outcomes to identify and define these subtypes, which then formed the basis of the gene signature. The "study with over 1007 patient samples associating Prosigna score with long-term outcome" (likely TransATAC) would also have used patient outcomes data for this association and for defining risk category cut-offs.