Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221498

    Validate with FDA (Live)

    Device Name
    Miha Bodytec II
    Manufacturer
    Date Cleared
    2022-10-07

    (137 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    miha bodytec II is a device which performs electronic muscle stimulation based on EMS technology. The device is specifically designed as an addition to other sports and for training muscles.

    miha bodytec II is intended to stimulate muscles in order to improve or facilitate muscle performance. In addition it is indicated for the following conditions:

    • Re-educating muscles
    • Relaxation of muscle spasm
    • Retarding or preventing disuse muscle atrophy

    The miha bodytec II electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    miha bodytec II may only be used by persons above the age of 21.

    Device Description

    miha bodytec II is a transcutaneous electrical muscle stimulation (EMS) device which stimulates motor nerves by means of electrical impulses transmitted by electrodes. These excitations of motor nerves are transmitted to the muscle fibers where they stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, pulse duration, pulse width, pulse rise, pause time, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    miha Bodytec II consists of a control unit mounted on a stand for the selection of programs, setting the parameters and starting/stopping the device, the i-body® electrode vest for applying electrodes to the upper body, i-body® straps for applying electrodes to the arm and legs and the i-body® belt for applying electrodes to the buttocks.

    The device encompasses the following variants:

    • "Miha bodytec II" control unit connected to the electrodes via cable
    • "Miha bodytec II" control unit connected via Bluetooth through the "i-body connect wireless" device"
    • "Miha bodytec m.ove" control unit for mobile use connected via Bluetooth through the "i-body connect wireless" device"

    The "miha bodytec m.ove" is a lighter version of the "miha bodytec II" device. Almost all electrical components and circuit boards are identical. It shall mainly be used as a mobile solution either with "travel station m.ove" or with "work station m.ove" in medical facilities. It must be connected with the electrode system via the additional wireless stimulation equipment "i-body connect wireless" and has no cable connection.

    The "travel station m.ove" is an equipment for the "miha bodytec m.ove" device (not for miha bodytec II) and is supposed to be used as a mobile solution for personal trainers all around the world. It will be used as a "bag" or "suitcase" where the device and the electrodes can be safely stored while travelling and will be used as a stand while the training with the customer is in progress. It has no electronics built in and is mainly made out of textiles, plastics and aluminum.

    The "work station m.ove" is an additional equipment for the "miha bodytec m.ove" (not for miha bodytec II) and is supposed to be mainly used in hospitals, physiotherapy facilities and homes for the elderly. It will be used as a transport solution where the device and the electrodes can be safely stored while moving through the facilities and will be used as a stand while the training with the patient is in progress. It has no electronics built in and is mainly made out of wood, plastics and aluminum/steel.

    The "i-body connect wireless" is a portable device, which is supposed to be worn at the patient's body. It has a built in stimulation circuit board and a Bluetooth receiver and produces the stimulation itself while the main device only acts as a remote control. This way a wireless training / treatment of the trainee / patient is possible. On the one hand, it is a mandatory equipment for the "miha bodytec m.ove" device due to the missing cable connection. On the other hand, it is an additional equipment for the "miha bodytec II" device, which can be used instead of the main connection cable. Through a corresponding software update, the second generation of the device (primary predicate device, K201975) can be changed into a third generation version (subject device, K221498), enabling the wireless functions of the device and thus compatibility with the i-body connect wireless.

    miha bodytec II must be used in a professional setting incl. professional sport setting and stationary in closed rooms (clinics, hospitals, nursing homes, doctor's offices, physical therapists' private offices). The device must be operated by a trainer who has received a full training from the manufacturer. Before the training, the trainer selects the accessories in the correct size, applies the electrodes to the athlete/patient i.e. by wearing the electrode vest and connects the straps and belt via cable to the vest and via the "i-body connect wireless" device to the control unit. The trainer can choose between several training programs on the control unit for impulse familiarization basic/advanced, muscular endurance and body relax. The intensity can be adjusted by the trainer at the UI of the control unit separately for each channel. Complete body training which addresses all muscle groups is possible with up to 10 pairs of electrodes. Each athlete/patient receives an RFID transponder card for storing training results and individually adjusted programs. Once the training is started, the control unit generates and transmits the electrical signals wirelessly to the "i-body connect wireless" device and thus to the electrodes. miha bodytec II uses bipolar pulses and supplies all channels equally during all programs.

    During pulse application, the trainer instructs the athlete/patient on specific exercises to perform. The training can be stopped anytime by pressing the multi-function / stop button.

    AI/ML Overview

    This document is a 510(k) summary for the Miha Bodytec II, a powered muscle stimulator. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed performance study with acceptance criteria in the typical sense of a clinical trial for diagnostic devices. Therefore, much of the requested information regarding sample sizes, expert ground truth, adjudication methods, and comparative effectiveness studies (MRMC) would not be found in this type of submission.

    The "acceptance criteria" and "reported device performance" provided are primarily related to safety, electrical output specifications, and regulatory compliance, demonstrating that the device functions as intended within established safety limits and industry standards.

    Here's a breakdown of the available information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is derived from the "Output Specifications" table (Table 2) and the "Performance Testing" section. Due to the nature of a 510(k) for a powered muscle stimulator, the "acceptance criteria" are generally compliance with international safety standards and performance within specified technical ranges, rather than clinical efficacy metrics like sensitivity or specificity.

    CharacteristicAcceptance Criteria (Implicit from Predicate & Standards)Reported Device Performance
    WaveformSymmetric biphasic (matching predicate)Symmetric biphasic
    ShapeRectangular (matching predicate)Rectangular
    Max Output VoltageWithin ranges comparable to predicate devicesmiha bodytec II (cable): ≤ 74Vp @ 500Ω, ≤ 152Vp @ 2kΩ, ≤ 152Vp @ 10kΩ miha bodytec II (wireless): ≤ 76.2Vp @ 500 Ω, ≤ 111Vp @ 2 kΩ, ≤ 117Vp @ 10 kΩ miha bodytec m.ove: ≤ 76.2Vp @ 500 Ω, ≤ 111Vp @ 2 kΩ, ≤ 117Vp @ 10 kΩ
    Max Output CurrentWithin ranges comparable to predicate devicesmiha bodytec II (cable): < 148mAp @ 500Ω, < 76mAp @ 2kΩ, < 15 mAp @ 10kΩ miha bodytec II (wireless): < 152.4mAp @ 500 Ω, < 55.5mAp @ 2 kΩ, < 11.7mAp @ 10 kΩ miha bodytec m.ove: < 152.4mAp @ 500 Ω, < 55.5mAp @ 2 kΩ, < 11.7mAp @ 10 kΩ
    Pulse Width50 - 400 µs (matching primary predicate)50 - 400 µs
    Frequency (Hz)Cable: 2 - 150 Hz (matching primary predicate) Wireless: 2 - 100 Hzmiha bodytec II (cable): 2 - 150 Hz miha bodytec II (wireless): 2 - 100 Hz miha bodytec m.ove: 2 - 100 Hz
    Symmetrical PhasesYes (matching predicate)Yes
    Phase Duration25 - 200 µs (matching primary predicate)25 - 200 µs
    Maximum Phase ChargeComparable to predicate devicesmiha bodytec II (cable): <32 µC @ 500Ω miha bodytec II (wireless): < 30.48 µC @ 500Ω miha bodytec m.ove: < 30.48 µC @ 500Ω
    Maximum Current DensityComparable to predicate devicesmiha bodytec II (cable): 0.64 mA/cm² @ 500Ω miha bodytec II (wireless): 0.42 mA/cm² @ 500 Ω miha bodytec m.ove: 0.42 mA/cm² @ 500 Ω
    Maximum Power DensityComparable to predicate devicesmiha bodytec II (cable): 0.82 mW/cm² @ 500Ω miha bodytec II (wireless): 5.47 mW/cm² @ 500 Ω miha bodytec m.ove: 5.47 mW/cm² @ 500 Ω
    Burst ModeContraction: 1 - 10 s, Relaxation: 0.0 - 10 s (matching primary predicate)Contraction time: 1 - 10 s, Relaxation time: 0.0 - 10 s
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-2-10Tested according to and in compliance with recognized standards.
    EMCCompliance with IEC 60601-1-2Tested according to and in compliance with recognized standards.
    Software ValidationCompliance with IEC 62304 and FDA guidanceVerified and validated; met software system requirements, full system validation including FDA "Guidance Document for Powered Muscle Stimulator 510(k)s" recommendations.
    Usability ValidationDevice meets intended use safely and effectively by specified usersOverall system validated to confirm safe and effective use.
    Electrode Uniform CurrentUniform current distribution (dispersion testing)Bench testing performed (4 batches, 8 lateral back, 1 upper back, 1 abdomen electrode per batch; 8 measure points per chest/back, 5 per abdomen). All tests successfully passed.
    Electrode Shelf Life3-year storage demonstratedAccelerated aging at 60°C for 10 weeks simulated for 3-year storage. 10 electrodes and 6 contact points tested. All tests successfully passed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Dispersion Testing: 4 batches of electrodes. From each batch, 8 lateral back-electrodes, 1 upper back-electrode, and 1 abdomen-electrode were selected. Specific numbers of measurement points were defined (8 points from each chest/lateral/upper back electrode, 5 points from each abdomen electrode).
    • Sample Size for Shelf Life Testing: 10 electrodes and 6 contact points.
    • Data Provenance: The nature of these tests (bench testing, climate chamber simulation) indicates they were conducted in a laboratory setting, likely in Germany, given the manufacturer's location. The tests are "retrospective" in the sense that they are conducted on manufactured components to verify design specifications, not on real-world patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of submission. The tests performed are engineering and electrical measurements, and compliance with standards. Ground truth for device performance in this context is established by the specifications in the standards themselves and the measurable physical outputs and characteristics of the device.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic evaluations where interpretative differences between human readers need to be resolved. The tests described are objective physical and electrical measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a powered muscle stimulator, not an AI-powered diagnostic tool, so an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The "Performance Testing" section describes validation of the firmware and electrical output characteristics, which can be considered a form of standalone performance evaluation for the device's electrical functions. This involves oscilloscope tracings of output waveforms and other electrical/thermal tests. The term "algorithm only" is not appropriate as this is a hardware device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the performance tests described:

    • Electrical Safety and EMC: Ground truth is defined by the limits and methodologies specified in the cited international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10).
    • Software and System Validation: Ground truth is defined by the software system requirements document and the specifications in the "Guidance Document for Powered Muscle Stimulator 510(k)s" and IEC 62304.
    • Bench Testing of Electrodes: Ground truth for uniform current distribution (dispersion) and electrical resistance for shelf life is based on the device's design specifications and the FDA's requirements for such tests.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set in that context. The "software" refers to firmware for device control, validated through traditional software engineering processes.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the machine learning sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201975

    Validate with FDA (Live)

    Device Name
    miha bodytec II
    Manufacturer
    Date Cleared
    2021-03-11

    (238 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    miha bodytec II (MBT II) is a machine with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles.

    miha bodytec II is intended to stimulate muscles in order to improve or facilitate muscle performance. In addition it is indicated for the following conditions:

    • · Re-educating muscles
    • · Relaxation of muscle spasm
    • · Retarding or preventing disuse muscle atrophy

    The miha bodytec II electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    Device Description

    miha bodytec II is a transcutaneous electrical muscle stimulation (EMS) device which stimulates motor nerves by means of electrical impulses transmitted by electrodes. These excitations of motor nerves are transmitted to the muscle fibers where they stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, pulse duration, pulse width, pulse rise, pause time, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    miha Bodytec II consists of a control unit mounted on a stand for the selection of programs, setting the parameters and starting/stopping the device, the i-body® electrode vest for applying electrodes to the upper body, i-body® straps for applying electrodes to the arm and legs and the i-body® belt for applying electrodes to the buttocks. The control unit and the i-body® accessories are connected via cables and must be worn on top of optionally available undergarment.

    AI/ML Overview

    This document is a 510(k) summary for the "miha bodytec II" powered muscle stimulator. It focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a performance study with acceptance criteria in the typical sense of a clinical trial or algorithm validation study. Therefore, most of the requested information (regarding expert readers, MRMC studies, ground truth, training sets, etc.) is not applicable to this type of regulatory submission.

    However, I can extract the information related to the device's technical acceptance criteria and the engineering and bench testing performed to prove it meets those criteria.

    Here's a breakdown of what can be inferred and what is not applicable based on the provided text:

    Key Takeaway: This document is a regulatory submission for device clearance based on substantial equivalence, not a clinical study demonstrating performance against specific disease detection acceptance criteria. The "acceptance criteria" here relate to engineering performance, safety standards, and functional specifications, not diagnostic accuracy metrics.


    Acceptance Criteria and Reported Device Performance (Engineering/Safety)

    The "acceptance criteria" in this context are primarily defined by compliance with recognized electrical safety, electromagnetic compatibility, and software validation standards, as well as functional equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Feature/Criterion (Generally Implied by Standards Compliance & Comparison)Acceptance Criteria (Implied by Standards/Predicate)Reported Device Performance (as stated in document)
    Electrical SafetyCompliance with AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012, IEC 60601-2-10:2016 (limits for current, voltage, leakage, etc.)Tested according to and in compliance with recognized standards. Specific values mentioned in comparison tables are within these limits.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2014Tested according to and in compliance with recognized standards.
    Software ValidationCompliance with IEC 62304 and FDA's Guidance: General Principles of Software Validation; meeting software system requirements.Firmware verified and validated. Software validation demonstrated firmware met software system requirements.
    UsabilityMeets intended use safely and effectively by specified users in specified environment (human factors and usability requirements).Overall system validated. Usability tests successfully conducted following FDA's guidance.
    Shelf Life of ElectrodesDemonstrated 1-year shelf life and 1-year storage.Tested according to ASTM F1980-16 with accelerated aging (60°C for 4 weeks). 72 electrodes tested. All tests successfully passed.
    Current Distribution (Dispersion Testing) of ElectrodesUniform current distribution.Bench testing performed according to FDA's requirements. 4 batches of electrodes tested (8 chest, 8 lateral back, 1 upper back, 1 abdomen). Measured prior to and after cleansing, and after accelerated aging. All tests successfully passed.
    Output WaveformsSymmetric biphasic, rectangular shape, specific voltage/current ranges (within IEC 60601-2-10 limits).Oscilloscope tracings obtained under max supported voltage/pulse widths at 500 Ω, 2 kΩ, and 10 kΩ. Values provided in "Output Specifications" table, all stated to comply and be within limits of IEC 60601-2-10.
    Operational Safety FeaturesShort-circuit monitoring, watchdog monitoring, no load trip, onload trip, immediate shut-off button, redundant hardware error monitoring, firmware self-tests.All these features are listed as present in the "miha bodytec II" and the primary predicate.

    Study Details (Applicable to Engineering/Bench Testing)

    2. Sample size used for the test set and the data provenance:

    • Test Set (Electrodes for Shelf Life/Dispersion):
      • Dispersion Testing: 4 batches of electrodes. From each batch, 8 chest, 8 lateral back, 1 upper back, and 1 abdomen electrode were selected. (Total: 4 x (8+8+1+1) = 4 x 18 = 72 electrodes).
      • Shelf Life Testing: 72 electrodes (from the same batches, implied).
    • Data Provenance: The testing was "bench testing" and conducted "in a climate cabinet" (for accelerated aging). The location (country of origin) is not explicitly stated for these specific tests but the manufacturer is based in Germany. The testing is prospective for the purpose of demonstrating compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the human-expert-reader sense. The ground truth for this type of submission is defined by the compliance of the device's technical specifications and performance against recognized international and national standards (e.g., IEC, ISO, ASTM) and FDA guidance documents. These are objective engineering measurements and validations.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This refers to clinical study methodologies for interpreting results (e.g., medical image interpretation). The "test set" here refers to engineering measurements against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is a powered muscle stimulator, not an AI-assisted diagnostic device. The submission does not involve human-in-the-loop diagnostic performance studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm-only diagnostic device. Its performance is related to its physical outputs (electrical impulses) and safety features.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the engineering and safety performance, the "ground truth" is defined by established international and national standards (e.g., IEC 60601 series, ISO 14971, ASTM F1980-16) and FDA guidance documents.
      • For current distribution: FDA's requirements for bench testing.
      • For software: IEC 62304 and FDA guidance for software validation.
      • For electrical safety/EMC: IEC 60601 series.
      • For usability: IEC 62366-1 and FDA guidance.

    8. The sample size for the training set:

    • Not Applicable. This device does not use an AI model that requires a "training set" in the machine learning sense. The software validation is for the device's operational firmware, not for an adaptive or learning algorithm.

    9. How the ground truth for the training set was established:

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182519

    Validate with FDA (Live)

    Device Name
    Miha Bodytec II
    Manufacturer
    Date Cleared
    2019-06-07

    (267 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    miha bodytec II is a machine with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for healthy muscles and clients, not be used for rehabilitation purposes.

    miha bodytec II is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. miha bodytec II is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the miha bodytec II training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.

    The miha bodytec II electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    The various types of muscle work that the miha bodytec II can impose on the stimulated muscles are able to improve or facilitate muscle performance. The miha bodytec II may therefore be considered a technique of muscle training.

    Device Description

    miha bodytec II is a transcutaneous electrical muscle stimulation (EMS) device which stimulates motor nerves by means of electrical impulses transmitted by electrodes. These excitations of motor nerves are transmitted to the muscle fibers where they stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, pulse intensity, pulse duration, pulse width, pulse rise, pause time, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    miha Bodytec II consists of a control unit mounted on a stand for the selection of programs, setting the parameters and starting/stopping the device, the i-body® electrode vest for applying electrodes to the upper body, i-body® straps for applying electrodes to the arm and legs and the i-body® belt for applying electrodes to the buttocks. The control unit and the i-body® accessories are connected via cables and must be worn on top of optionally available undergarment.

    miha bodytec II must be used in a professional sport setting. The device must be operated by a trainer who has received a full training from the manufacturer. Before the trainer selects the accessories incl. electrodes in the correct size, applies the athlete i.e. by wearing the electrode vest and connects the straps und belt via cable to the vest and to the control unit. The trainer can choose between several training programs on the control unit for impulse familiarization, invigoration basic/advanced, muscular endurance and body relax. The intensity can be adjusted by the trainer at the UI of the control unit separately for each channel. Complete body training which addresses all muscle groups is possible with up to 10 pairs of electrodes. Each athlete receives an RFID transponder card for storing training results and individually adjusted programs. Once the training is started, the control unit generates and transmits the electrical signals to the electrodes via cable. miha bodytec II uses bipolar pulses and supplies all channels equally during all programs.

    AI/ML Overview

    The provided document is a 510(k) Summary for the miha bodytec II device, which is a powered muscle stimulator. It details various aspects of the device and its comparison to a predicate device (E-Fit EF-1280) to demonstrate substantial equivalence for FDA clearance.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:

    Important Note: This document is a 510(k) Summary for a Powered Muscle Stimulator, not an AI/ML-enabled diagnostic device for medical imaging. Therefore, many of the typical "acceptance criteria" and "study" elements related to AI algorithm performance (like sensitivity, specificity, MRMC studies, expert consensus for ground truth on imaging data, sample sizes for training/test sets, etc.) as you might expect for an AI diagnostic tool, are not applicable here. The acceptance criteria for this device are focused on electrical safety, electromagnetic compatibility, software validation, usability, biocompatibility, and bench testing to demonstrate functional characteristics and shelf life.

    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Tests ConductedReported Device Performance (Compliance/Results)
    Electrical SafetyCompliance with AAMI ANSI ES 60601-1_2005/(R)2012 And A1:2012In compliance.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2014In compliance.
    Software/Firmware ValidationCompliance with IEC 62304:2015 and FDA's "General Principles of Software Validation"Verified and validated; firmware met software system requirements.
    System Validation (Functional)Oscilloscope tracings of output waveforms under various loads (500 Ω, 2 kΩ, 10 kΩ) with maximum supported voltage and pulse widths.Tracings obtained (results not explicitly detailed but implied compliance with expected output).
    System Validation (Functional, Other)Electrical tests of interfaces, thermographic inspections, climate chamber tests, shock and vibration tests.Conducted (results not explicitly detailed but implied compliance).
    Usability ValidationCompliance with IEC 62366-1:2015 and FDA's guidance on usability. Confirmation that device meets intended use safely and effectively by specified users in specified environment, considering human factors.Overall system validated; UI tested for usability according to FDA's guidance.
    BiocompatibilitySkin contacting material tested to ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2009 (irritation/sensitization).Tested to these standards (results implied acceptable).
    Shelf LifeAccelerated aging simulation (60°C for 4 weeks) with typical training current applied to electrodes; tested for 1-year shelf life and 1-year storage.All tests successfully passed.
    Dispersion Testing (Electrodes)Uniform current distributions bench testing using 4 batches of electrodes (8 chest, 8 lateral back, 1 upper back, 1 abdomen electrode from each batch). Measurements taken prior to and after cleansing, and after accelerated aging.All tests successfully passed.
    Output Specifications (Comparison to Predicate)Max Output Voltage: miha bodytec II: <= 74Vp @ 500Ω, <= 152Vp @ 2kΩ, <= 152Vp @ 10kΩ. E-Fit EF-1280: 36V @500Ω. Max Output Current: miha bodytec II: < 148mAp @ 500Ω, < 76mAp @ 2kΩ, < 15 mAp @ 10kΩ. E-Fit EF-1280: 72mA @500Ω. Max Power Density: miha bodytec II: 0.82 mW/cm² @ 500Ω. E-Fit EF-1280: 6.3 mW/cm² @ 500Ω.Differences were noted but considered not to raise new questions regarding safety and effectiveness, as both devices comply with IEC 60601-2-10 limits.
    Compliance with General ControlsAnnual registration, listing, labeling, GMP, prohibitions against misbranding and adulteration.The FDA letter reminds the applicant of these ongoing requirements.

    Study Details (as applicable to this type of device)

    As this is a 510(k) for a powered muscle stimulator, the "study" is primarily a series of engineering, biological, and bench tests demonstrating compliance with recognized standards and substantial equivalence to a predicate device, rather than a clinical study involving human subjects or an AI model's performance on a dataset of medical images.

    2. Sample size used for the test set and the data provenance:

    • Electrical Safety & EMC, Software, Usability, System Validation: These tests are typically performed on a representative number of manufactured units of the device (e.g., a few units, or internal test benches). The document doesn't specify a "sample size" in the context of a dataset of patient data, as it's not applicable. Provenance isn't relevant as it's testing the device itself.
    • Biocompatibility: Testing was performed on the "skin contacting material." No specific "sample size" of patients or data is provided as it's material testing.
    • Shelf Life & Dispersion Testing:
      • Dispersion: "4 batches of electrodes were tested. From each batch 8 chest-electrodes (yellow), 8 lateral back-electrodes (blue), 1 upper back-electrode (green) and 1 abdomen-electrode (red) were selected." This translates to a total of (8+8+1+1) * 4 = 72 electrodes for dispersion testing.
      • Shelf Life: "Altogether 72 electrodes were tested."
      • Data Provenance: This is bench testing of the physical electrodes/device components. It's not human patient data, so concepts like country of origin or retrospective/prospective don't apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is not an AI/ML diagnostic tool requiring human expert annotation to establish ground truth on medical images. The "ground truth" for this device's performance is objective measurements (electrical parameters, material properties, software function validated against requirements, etc.) and compliance with engineering standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This relates to human expert consensus or resolving discrepancies in diagnostic interpretations, which is not relevant for the type of testing described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a powered muscle stimulator, not a diagnostic AI device to be used by human readers for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. While the device does have firmware/software, its purpose is to control the muscle stimulation, not to perform an independent diagnostic analysis. The performance is assessed based on its ability to generate the specified electrical impulses safely and effectively when operated by a trained user.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the usual sense for AI diagnostics. For this device, "ground truth" is established by:
      • Reference Standards: Adherence to recognized national and international standards (IEC 60601 series, ISO 10993 series, etc.) for electrical safety, EMC, software, and biocompatibility.
      • Engineering Specifications: The device's output (voltage, current, pulse width, frequency) is measured against its own design specifications and the limits set by relevant standards.
      • Functional Benchmarks: Performance testing to confirm the device operates as intended (e.g., uniform current distribution in electrodes, successful completion of aging tests).

    8. The sample size for the training set:

    • Not applicable. This device does not use machine learning in a way that requires a "training set" of data in the typical sense of AI/ML development for diagnostic tasks. Its firmware is validated as traditional software, not trained on data.

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no "training set" for AI/ML, there's no ground truth established in that context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1