Search Results
Found 133 results
510(k) Data Aggregation
(246 days)
Ask a specific question about this device
(99 days)
The SCOUT MD Reflector is intended to be placed percutaneously in soft tissue (>30 days) to mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (SCOUT MD System) the SCOUT MD Reflector is located and surgically removed with the target tissue. The SCOUT MD System is intended only for the non-imaging detection and localization of the SCOUT MD Reflector that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.
The SCOUT MD Surgical Guidance System consists of the following components:
- SCOUT MD Delivery System, which includes two components:
- SCOUT MD Reflector
- SCOUT MD Delivery Device
- SCOUT MD Guide
- SCOUT MD Single-Use Handpiece
- SCOUT MD Console
The SCOUT MD Delivery System is used to implant the preassembled SCOUT MD Reflector. The needle of the Delivery System is percutaneously advanced into tissue to the site to be marked for biopsy or surgical removal. Needle placement is confirmed under imaging technique (radiographic, ultrasound). The Reflector (tissue marker) is deployed at the target site and the Delivery Device is removed from the patient and discarded. The Reflector, a passive implant, remains in situ and, if surgical removal of the target tissue is necessary, the Reflector is located at the time of surgery (intraoperatively) by the SCOUT MD Surgical Guidance System. The SCOUT MD Guide/Handpiece connected to the SCOUT MD Console is used to detect the SCOUT MD Reflector but it does not contact tissue. SCOUT MD Guides are used with sterile SCOUT Guide Sheath and/or sterilized per the instructions in the IFU. When the SCOUT MD System detects the Reflector, the Console emits audible feedback that increases in cadence as the Guide/Handpiece is placed closer to the Reflector. The distance between the distal end of the Guide/Handpiece and the detected Reflector, in millimeters, is displayed on the Console. If necessary, the Reflector is removed from the patient during a subsequent surgical procedure along with the tissue of interest or the Reflector can be left in-situ.
N/A
Ask a specific question about this device
(228 days)
The Merit Syringe is used to inject fluids into, and withdraw fluids from, the body.
The Merit Syringe a standard three-piece piston syringe intended for general purpose injection and aspiration by a qualified clinician. The syringe is constructed of a calibrated, hollow, clear polycarbonate barrel into which is inserted a closely fitted movable ABS plunger with black silicone rubber piston seal (tip). The barrel is marked with standard graduations per ISO 7886-1 requirements and contains an ISO 80369-7 compliant fixed male luer, small-bore connector. The subject device is single use and is available in either an ethylene oxide (EO) or gamma-sterile configuration.
N/A
Ask a specific question about this device
(90 days)
The Resilience Fully Covered Esophageal Stent System is intended for maintaining esophageal luminal patency in patients with esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and for occlusion of esophageal fistulae.
The MERIT ENDOTEK® Resilience™ Fully Covered Esophageal Stent System is a through-the-scope system comprised of two components: the radiopaque self-expanding nitinol stent and the delivery system. The stent is completely covered with a biocompatible silicone membrane. Stent expansion results from the physical properties of the metal and the proprietary geometry. The stent is designed with unique geometrical features at the distal and proximal ends to anchor the stent and resist migration. The proximal and distal ends of the stent are threaded with suture intended for use in repositioning and removal of the stent.
The stents come in nine (9) sizes, comprising all combinations of three lengths (5cm, 6cm, and 7cm) and three midbody diameters (14mm, 17mm, and 20mm). The midbody lengths are 1cm, 2cm, and 3cm with the flared ends comprising the remaining length. The ends of the stents are larger in diameter than the midbody.
There is one 10.5Fr delivery system used to deploy all 9 stents.
The provided FDA 510(k) clearance letter and summary for the Resilience Fully Covered Esophageal Stent System describes non-clinical testing to demonstrate substantial equivalence to a predicate device. However, it does not include information about clinical studies involving human readers or AI algorithms. The clearance is based on the device itself (the stent system), not an AI algorithm performing diagnosis or analysis.
Therefore, many of the requested points, especially those pertaining to AI performance, human reader studies, and ground truth establishment for AI, cannot be answered from the provided text.
Here's an analysis based on the information provided, focusing on the device's non-clinical testing:
Acceptance Criteria and Reported Device Performance (Non-Clinical)
The study described is a series of non-clinical, in-vitro tests on the mechanical and material properties of the Resilience Fully Covered Esophageal Stent System and its delivery system.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Stent Verification Testing | |
| Dimensions & Condition (post-deployment & removal) | Passed predetermined specifications |
| Suture purse-stringing ability | Passed predetermined specifications |
| Stent spring-back after purse-stringing | Passed predetermined specifications |
| Repositioning with forceps | Passed predetermined specifications |
| Suture withstands forceps | Passed predetermined specifications |
| Removal force | Passed predetermined specifications |
| Flare retention force | Passed predetermined specifications |
| Radial pressure forces (compression/expansion) | Passed predetermined specifications |
| Tensile strength | Passed predetermined specifications |
| Food flow | Passed predetermined specifications |
| Post-Fatigue Stent Testing | |
| Condition (post-fatigue) | Passed predetermined specifications |
| Removal force (post-fatigue) | Passed predetermined specifications |
| Flare retention force (post-fatigue) | Passed predetermined specifications |
| Radial pressure forces (compression/expansion, post-fatigue) | Passed predetermined specifications |
| Tensile strength (post-fatigue) | Passed predetermined specifications |
| Delivery System Testing | |
| Guidewire compatibility | Passed predetermined specifications |
| Catheter working length | Passed predetermined specifications |
| Catheter outer diameters | Passed predetermined specifications |
| Safety functionality | Passed predetermined specifications |
| Stent deployment force | Passed predetermined specifications |
| Stent repositioning after partial deployment | Passed predetermined specifications |
| Stent expansion after deployment | Passed predetermined specifications |
| Deployment accuracy | Passed predetermined specifications |
| Delivery system condition (post-deployment & removal) | Passed predetermined specifications |
| Tip-to-guidewire lumen bond strength | Passed predetermined specifications |
| Outer sheath-to-connector bond strength | Passed predetermined specifications |
| Egress plug-to-guidewire lumen bond strength | Passed predetermined specifications |
| Material/Environmental Testing | |
| Corrosion testing | Met all predetermined corrosion performance requirements |
| MRI compatibility | MR conditional in accordance with FDA's guidance |
| Biocompatibility testing | Met requirements of FDA Final Guidance ISO 10993-1 |
| Overall Design Validation | All design validation requirements were met |
Information Not Found in the Provided Text:
Since this is a clearance for a medical device (esophageal stent) and not an AI/software device, the following points related to AI studies are not applicable or not present in the provided document.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- This device clearance is based on non-clinical, in-vitro testing. It does not utilize "test sets" in the context of clinical data for AI analysis. The "test sets" here refer to batches of manufactured stents and delivery systems for mechanical and material property evaluation. Specific sample sizes for each non-clinical test are not detailed in this summary, but the summary states "all of the above testing passed the predetermined specifications" and "All design validation requirements were met."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable as this is not an AI/software clearance based on expert-labeled clinical data. Ground truth for non-clinical device testing is typically based on engineering specifications and direct physical measurements/observations.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-assisted device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical testing, the "ground truth" or "reference standard" would be the pre-defined engineering specifications and performance criteria derived from established medical device standards and the predicate device's performance benchmarks. For biocompatibility, the ground truth is compliance with ISO 10993-1.
-
The sample size for the training set:
- Not applicable, as this is not an AI/software device. There is no "training set" in the context of machine learning.
-
How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(30 days)
The Merit Prelude Wave Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.
The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guidewires.
The Prelude Wave™ Hydrophilic Sheath Introducer consists of a sheath introducer with compatible vessel dilator that snaps securely into the sheath introducer hub. The sheath hub contains an integral hemostasis valve. The sheath tubing is coated with a hydrophilic coating. A sidearm is affixed to the sheath hub and has a 3-way stopcock at its proximal end. The Prelude Wave™ Hydrophilic Sheath Introducer is available in 11cm and 23cm lengths, (French sizes 4F, 5F, 6F and 7F) and is designed to accept 0.018", 0.021" and 0.025" diameter guidewires. The Prelude Wave™ Hydrophilic Sheath Introducer is marketed with any of the following components, depending on the product configuration: guidewire, metal access needle, access needle with inner metal needle and outer plastic cannula and SnapFix™ securement device.
The provided FDA 510(k) clearance letter and summary for the Prelude Wave Hydrophilic Sheath Introducer (K250909) do not contain the specific information requested in the prompt regarding acceptance criteria, device performance, sample sizes for training and testing, expert qualifications, or detailed descriptions of ground truth establishment for an AI/ML medical device.
This document describes a traditional medical device (a catheter introducer), not an AI/ML-based device. Therefore, the types of studies and acceptance criteria typically associated with AI/ML device clearance (as outlined in your prompt questions) are not applicable here.
The document primarily focuses on:
- Substantial Equivalence: Demonstrating that the new device is as safe and effective as a legally marketed predicate device (Prelude IDEAL Hydrophilic Sheath Introducer, K173750). This is a common pathway for traditional medical devices.
- Performance Testing: Verification and validation studies to ensure the device meets its specifications and performs as intended, based on established international standards (ISO 10555-1, EN ISO 11070, ISO 10993-1, ISO 14971). These tests are standard for physical medical devices and include mechanical, functional, and biocompatibility assessments.
- Design Changes: Highlighting minor design modifications from the predicate, such as sheath material, sheath/dilator tip fit, and inclusion of an adhesive fixation device.
Therefore, I cannot populate the table or answer the specific questions about AI/ML device study parameters (like MRMC, standalone AI performance, expert labeling, training data) based on the provided text. The provided text simply does not contain this information because it's a submission for a physical medical device, not an AI software.
Ask a specific question about this device
(169 days)
The Merit Micro Ace™ Gold Advanced Micro Access System is intended for percutaneous placement of a 0.035" (0.89mm) or 0.038" (0.97mm) guidewire into the vascular system.
The Merit Micro Ace™ Gold Advanced Micro Access System (hereafter referred to as Micro Ace™ Gold) provides access to the vascular system and facilitates the placement of a 0.035" (0.89mm) or 0.038" (0.97mm) guide wires. The system is available in 4 French and 5 French sizes – and includes the coaxial introducer-dilator pair, a 21 gauge needle and a 0.018” (0.46mm) guide wire.
The introducer-dilator assembly that is included in each kit is 10 cm effective length. The system will be offered with optional guidewires: Some catalog codes will include a nitinol wire with platinum guidewire tips; Some catalog codes will be offered with a stainless steel wire with platinum guidewire tips.
The provided text is a 510(k) summary for the Merit Micro Ace™ Gold Advanced Micro Access System. It describes the device, its intended use, comparison to a predicate device, and the safety and performance testing conducted.
However, this document does not describe a study proving a device meets acceptance criteria related to an AI/Software as a Medical Device (SaMD) product. Instead, it pertains to a physical medical device (vessel dilator for percutaneous catheterization) and describes the verification and validation tests for its physical characteristics and biocompatibility.
Therefore, many of the requested items, such as "effect size of how much human readers improve with AI vs without AI assistance," "standalone (i.e. algorithm only without human-in-the-loop performance)," "number of experts used to establish ground truth," and "sample size for training set," are not applicable to this type of device and review.
The document focuses on demonstrating substantial equivalence to a predicate device through physical and material property testing, rather than performance of an AI algorithm based on data analysis.
Given this, I will answer the applicable questions based on the provided text, and explicitly state when a question is not applicable.
Acceptance Criteria and Device Performance (for a physical medical device):
The document lists "Design Verification Studies" and "Design Validation Studies" as evidence that the device meets "pre-established performance criteria" and addresses "unacceptable risks." The acceptance criteria are implicitly the successful completion of these tests as per relevant international standards.
1. A table of acceptance criteria and the reported device performance:
The document broadly states: "All tested samples met pre-established performance criteria and were deemed acceptable." It does not provide a quantitative table of specific acceptance criteria values and the measured performance results for each test. Instead, it lists the types of tests performed.
| Acceptance Criteria Category (Implicit via test type) | Test Performed | Reported Device Performance |
|---|---|---|
| Physical Dimensions | Dilator Effective length | All tested samples met pre-established performance criteria and were deemed acceptable. |
| Dilator Outer Diameter | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Dilator Tip Inner Diameter | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Mechanical Properties | Insertion Force | All tested samples met pre-established performance criteria and were deemed acceptable. |
| Pushability and Kink Test | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Shaft Tensile | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Hub Tensile | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Dilator Tip Bend | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Dilator Tip Fracture | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Material/Surface Characteristics | Particulate test | All tested samples met pre-established performance criteria and were deemed acceptable. |
| Lubricity test | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Biocompatibility | Cytotoxicity | All tested samples met pre-established performance criteria and were deemed acceptable. |
| Sensitization | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Irritation | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Acute Systematic Toxicity | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Pyrogenicity | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Hemolysis | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Partial Thromboplastin Time | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Heparinized Blood Platelet and Leukocyte Count | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Complement Activation | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Extract Particulates | All tested samples met pre-established performance criteria and were deemed acceptable. | |
| Design Validation | Clinician feedback assessment | Not specified, but implied satisfactory as part of overall acceptability statement. |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document states that "The tests were performed on both 4F and 5F Micro Ace™ Gold products." It also mentions "All samples were manufactured in accordance with existing and validated processes, and when sterile, are representative of product that Merit Medical intends to commercialize." It does not specify the exact number of samples per test for the test set.
- Data Provenance: The data provenance is from internal testing conducted by Merit Medical Systems, Inc. The document does not specify country of origin for the data (beyond the company's location in South Jordan, Utah, USA), nor does it classify the study as retrospective or prospective in the context of clinical data, as this is a device verification/validation study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable in the context of AI/SaMD ground truth establishment.
- For the "Clinician feedback following assessment of design changes" in the Design Validation Studies, the specific number and qualifications of clinicians are not provided. This feedback typically informs the practical usability and design, rather than establishing a 'ground truth' for an AI algorithm.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. This concept is relevant for reconciling discrepancies in expert annotations for AI ground truth, which is not what this document describes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study is for evaluating the impact of AI on human reader performance, and is not applicable to a physical vessel dilator device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This applies to AI/SaMD performance, which is not the subject of this submission.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not Applicable in the sense of 'ground truth' for AI performance. The "ground truth" for this device's performance is defined by its compliance with engineering specifications, industry standards (ISO 10555-1, EN ISO 11070 2014/A1:2018, ISO 10993-1, ISO 14971), and successful completion of pre-established physical, mechanical, and biocompatibility criteria.
8. The sample size for the training set:
- Not Applicable. This document describes a physical medical device, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established:
- Not Applicable. As above, this document does not describe an AI algorithm or its training set.
Ask a specific question about this device
(114 days)
The Ventrax™ Delivery System is intended to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart.
The Ventrax™ Delivery System is designed to provide a conduit to deliver diagnostic and therapeutic catheters to specific heart chambers and locations. It provides support for positioning and maintaining the position of catheters at specific locations in the heart. The sheath may be used for percutaneous entry. The system consists of four components: a sheath, a pigtail dilator, a straight dilator and a J-tipped Amplatz guidewire.
VENTRAX™ DELIVERY SYSTEM COMPONENTS
A. 8.5F Guiding Sheath Introducer: provides a conduit to deliver diagnostic and therapeutic catheters to specific heart chambers and locations. The sheath has an integrated valve to restrict blood loss, and a sideport for flushing and withdrawing blood.
B. Mating Pigtail Dilator: designed to conform to the sheath introducer inner diameter, has a tapered tip, has a pigtail at the distal end to assist aortic valve crossing, has an integrated valve to restrict blood loss, and has a sideport for flushing.
C. Mating Straight Dilator: designed to conform to the sheath introducer inner diameter and has a tapered tip. Usage of this straight dilator is optional. This straight dilator is intended to be used only when access is unsuccessful after using the mating pigtail dilator.
D. 0.035" X 220cm J-tipped Amplatz guidewire: provide path for sheath and dilator advancement.
This looks like a 510(k) summary for a medical device called the Ventrax™ Delivery System. However, this submission does not contain information about software with AI/ML components. The document focuses on the substantial equivalence of the Ventrax Delivery System to a previously cleared predicate device with respect to its physical design, materials, and mechanical performance. There is no mention of accepting criteria related to AI/ML device performance, or any studies designed to prove AI/ML performance.
Therefore, I cannot provide an answer based on the prompt's request for acceptance criteria and study details related to AI/ML. The provided text does not contain any of that information.
Ask a specific question about this device
(195 days)
The Prelude Small O.D. Introducer Guide Wire is intended to facilitate the placement of introducer sheaths during diagnostic and interventional procedures.
The Prelude Small O.D. Introducer Guide Wire is indicated for use in the peripheral vasculature only.
The Prelude® Small O.D. Introducer Guide Wire is composed of a stainless-steel core wire inside a stainless-steel coil. The guide wire is welded at the distal and proximal tips and has a polished weld finish. The stainless-steel construction provides radiopacity and visibility to the user under X-ray. The wires are available in outer diameters 0.018", 0.021" and 0.025", lengths from 45 cm to 80 cm, straight and double-ended J 3mm / straight tip shapes, Standard and Firm body wire stiffness profiles, with a flexible atraumatic tip.
The provided text is a 510(k) summary for a medical device called the "Prelude Small O.D. Introducer Guide Wire." It details the device's characteristics, intended use, and a comparison to a predicate device, along with a list of non-clinical performance and biocompatibility tests conducted.
However, the request asks for specific information regarding acceptance criteria and a study that proves the device meets the acceptance criteria, particularly in the context of an AI-powered device or a study involving human readers and AI assistance.
The provided document does not describe an AI-powered device, a study involving human readers, a test set with expert ground truth, or an MRMC comparative effectiveness study. The tests listed are for a physical medical guide wire, focusing on its mechanical properties, material integrity, and biocompatibility.
Therefore, I cannot fulfill the request for information on acceptance criteria and study proving device meets the criteria for an AI device or a study involving human readers and AI assistance, as the provided text pertains to a traditional, non-AI medical device.
Here's a breakdown of why I cannot answer the specific questions based on the provided text:
- 1. A table of acceptance criteria and the reported device performance: While the document states "The subject Prelude Small O.D. Introducer Guide Wire met the predetermined acceptance criteria," it does not list the specific numerical or qualitative acceptance criteria for each test (e.g., "Tensile strength must be >X Newtons"). It also does not report the exact performance results (e.g., "Tensile strength was Y Newtons"). It only confirms that the criteria were met.
- 2. Sample size used for the test set and the data provenance: Not applicable. The "test set" here refers to the physical guide wires tested, not a dataset of images or clinical cases. Sample sizes for each mechanical and biocompatibility test are not provided. Data provenance (country of origin, retrospective/prospective) is irrelevant for this type of device testing.
- 3. Number of experts used to establish the ground truth... and qualifications: Not applicable. Ground truth as typically defined for AI/imaging studies (e.g., expert radiological reads) is not relevant for the physical testing of a guide wire.
- 4. Adjudication method for the test set: Not applicable. There's no "adjudication" for mechanical and biocompatibility tests in the way there would be for expert interpretations of medical images.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done... effect size: Not applicable. This type of study involves human readers interpreting medical images, usually with and without AI assistance. The device in question is a physical guide wire, not an AI algorithm for imaging.
- 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. The device is not an algorithm.
- 7. The type of ground truth used: Not applicable. Ground truth in the context of AI/imaging studies (expert consensus, pathology, outcomes data) is not relevant here. The "ground truth" for this device's performance is derived from standardized physical and chemical tests (e.g., measuring tensile strength against a specification).
- 8. The sample size for the training set: Not applicable. This device does not involve a training set as it's not an AI model.
- 9. How the ground truth for the training set was established: Not applicable, as there's no training set.
In summary, the provided document is for a conventional medical device (a guide wire) and does not contain the information requested regarding AI device performance studies.
Ask a specific question about this device
(236 days)
The Impress Angiographic Catheter is designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures. Angiographic catheters with marker bands may also be used for anatomical measurements.
The Merit Impress Angiographic Catheters with/without hydrophilic coating are intravascular diagnostic catheters that are intended for the administration of contrast for conducting fluoroscopic studies. The catheters are available in a variety of 4F and 5F wire-braided and non-braided configurations ranging in length from 40cm to 125cm. The device may include a marker band to assist anatomical measurements. The devices are available in a variety of tip shapes to cater to variations in physician preference and patient anatomy.
The catheters consist of a shaft with a molded hub assembly. It is the shaft that may be offered with or without a wire-braided reinforcement. The distal tip of the device is tapered for efficient volume flow and dispersion of the contrast media. The distal tip is flexible so as to minimize the potential for vessel trauma.
The provided text is a 510(k) premarket notification summary for the Impress Angiographic Catheter. It outlines the device's characteristics, indications for use, comparison to predicate devices, and the testing conducted to demonstrate its safety and performance. However, this document does not contain information about an AI/algorithm-based medical device study.
The device in question, the "Impress Angiographic Catheter," is a physical medical device used for delivering radiopaque media in the vascular system. The testing described (design verification, design validation, and biocompatibility studies) are standard for physical medical devices and do not involve AI or software performance evaluation against an acceptance criterion that would be typical for an AI/ML product.
Therefore, I cannot provide the requested information regarding acceptance criteria and study proving an AI device meets those criteria based on the provided text. The document focuses on the substantial equivalence of a physical catheter, primarily due to changes in its material composition.
Ask a specific question about this device
(92 days)
The Siege Vascular Plug is indicated for arterial embolization in the peripheral vasculature.
The Siege Vascular Plug is intended for therapeutic embolization to reduce or obstruct blood flow.
The subject Siege Vascular Plug ("Device") is a self-expanding braided Nitinol vascular embolization implant that is supplied with components used for implantation. The Device has radiopaque marker bands attached to each end and a screw attachment for connection to the Delivery Wire. The Device is packaged collapsed within a Loader and attached to a 225 cm Delivery Wire that is provided within a hoop dispenser. The Delivery Wire has a lubricious, hydrophilic coating at its distal end (approximately 26.5cm long). A Tuohy Borst Valve is provided for flushing and maintaining hemostasis: a Torque Device is provided for releasing the Device; a Non-Vented Luer Cap is provided for device preparation.
The subject Siege Vascular Plug has been designed with a material, size, configuration, and shape that allows introduction through recommended 0.021" or 0.027" inner diameter commercial microcatheters for the embolization of arterial blood vessels in the peripheral vasculature. The Siege Vascular Plug Devices are provided in three different outer diameters (4.5 mm, 6.5 mm, 8.5 mm) to treat different artery diameter ranges (1.5 - 2.5 mm, 2.5 - 4 mm, 4 - 6 mm) in the peripheral vasculature.
This document describes the 510(k) premarket notification for the Siege Vascular Plug, a medical device intended for arterial embolization in the peripheral vasculature. The device's manufacturer, Merit Medical Systems, Inc., demonstrated substantial equivalence to a legally marketed predicate device (Siege Vascular Plug [K212817]) through a series of non-clinical performance tests and a GLP animal study.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document broadly states that "All test results were comparable to the predicate Siege Vascular Plug and the subject Siege Vascular Plug met the predetermined acceptance criteria." However, specific numerical acceptance criteria and detailed performance results for each test are not explicitly provided in the furnished text. Instead, it lists the types of tests conducted:
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Simulated Use | Met predetermined acceptance criteria. |
| Set (Tensile) Strength | Met predetermined acceptance criteria. |
| Loader Heat Shrink Tensile | Met predetermined acceptance criteria. |
| Fatigue Testing | Met predetermined acceptance criteria. |
| Nickel Leach | Met predetermined acceptance criteria. |
| Radial Force | Met predetermined acceptance criteria. |
| Nitinol Austenite Finish Temperature | Met predetermined acceptance criteria. |
| MRI Compatibility | Met predetermined acceptance criteria. |
| Corrosion Testing | Met predetermined acceptance criteria. |
| Dimensional Testing/Size Designation | Met predetermined acceptance criteria. |
| Particulate Testing | Met predetermined acceptance criteria. |
| Radiopacity Testing | Met predetermined acceptance criteria. |
| Packaging | Met predetermined acceptance criteria. |
| Sterilization | Met predetermined acceptance criteria. |
| Pyrogenicity | Met predetermined acceptance criteria. |
| Shelf Life | Met predetermined acceptance criteria. |
| Biocompatibility | Met predetermined acceptance criteria. |
| GLP Animal Study | Demonstrated acute and chronic safety, performance, and handling comparable to a commercially available Reference Device in a porcine model. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes used for each non-clinical test (e.g., number of plugs tested for tensile strength or fatigue).
For the GLP Animal Study:
- Sample Size: Not specified (only mentions "a porcine model").
- Data Provenance: Prospective (as it's a study explicitly conducted to evaluate the subject device). The "country of origin" is not specified.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
This information is not applicable and therefore not provided in the document. The acceptance criteria were based on "requirements outlined in guidance's and industry standards," and the performance was evaluated through physical and biological testing, not through expert-established ground truth in the context of diagnostic performance.
4. Adjudication Method for the Test Set:
This information is not applicable and therefore not provided in the document. Adjudication methods are typically used for evaluating human-in-the-loop diagnostic performance, not for the technical and biological performance tests described here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. The evaluation focused on the physical, mechanical, and biological performance of the device itself, not on human reader performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No, a standalone (algorithm only) performance study was not done. The device is a physical vascular plug, not an AI algorithm.
7. The Type of Ground Truth Used:
The "ground truth" for the non-clinical performance tests was adherence to established industry standards, guidance documents, and pre-defined specifications. For the animal study, the ground truth was the observable acute and chronic safety, performance, and handling within the porcine model, compared to a reference device. This is equivalent to performance against established benchmarks/standards and direct observation of biological safety/effectiveness.
8. The Sample Size for the Training Set:
This information is not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable, as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 14