510(k) Data Aggregation

    K Number
    K201224

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-09-18

    (135 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electrosurgical Generator 8070 is indicated for monopolar or bipolar surgery to achieve cut or coagulation for the tissue. It is intended to be used with monopolar handpiece and dispersive electrode or bipolar handpiece and footswitches.

    Device Description

    8070 is an Electrosurgical Generator with Monopolar Pure cut, Monopolar Bland Cut, Bipolar function. 8070 consists of HF generator and accessories. The device generates HF power, which is transmitted to the tissue through the output Pencil. Operator selects the output function and power with the front panel. The indicator is used to indicate the setup status and output status of the output is controlled by foot or hand switch.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for an Electrosurgical Generator and Accessories. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than fulfilling specific acceptance criteria for a novel device's performance in a clinical setting like an AI algorithm would.

    Therefore, the document does not contain the information requested regarding acceptance criteria for an AI device, a study proving it meets those criteria, sample sizes for test/training sets, expert adjudication methods, MRMC studies, or specific ground truth establishment for AI.

    Instead, the document details:

    • Acceptance Criteria (for this device's type): Demonstrated safety and effectiveness through compliance with recognized standards for electromedical devices (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2) and FDA guidance for electrosurgical devices.
    • Study Proving Acceptance: Non-clinical (bench) performance testing, electrical safety testing (IEC 60601-1), electromagnetic compatibility testing (IEC 60601-1-2), high-frequency safety testing (IEC 60601-2-2), and usability testing (IEC 60601-1-6). The document states these tests were "successfully completed."
    • Performance Comparison: The subject device's technological characteristics (e.g., maximum power, voltage output, waveforms for different modes) are compared in a table to its predicate device (Medrange Corporation MB 8010). The differences are presented, and the conclusion is that these differences "do not raise new questions of safety and effectiveness."

    In summary, the information requested in your prompt (especially points 2-9) is typical for AI/ML device submissions, which often involve clinical performance studies. The provided document is for an electrosurgical generator, a hardware device, and thus the basis for its clearance is non-clinical testing and comparison to a predicate device, not in-depth clinical studies with human readers or AI ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122092

    Validate with FDA (Live)

    Date Cleared
    2012-10-23

    (99 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EES 100 with Accessories is intended to deliver argon gas for argon plasma coagulation on tissue when used in conjunction with an Electrosurgical Generator and applicators or probes.

    Device Description

    The EES 100 is an expansion system for the electrosurgical unit (ESU). The EES 100 is not a generator. It is designed for the enhanced control of bleeding and the electrosurgical destruction of tissue in multispecialty procedures in the operating room or endoscopy suite. The EES 100 equipped with the argon module is used in combination with a high-frequency generator and an electrosurgical handpiece and the users can select appropriate operation modes: Pulse, Endo or Ar-cut mode to fulfill various argon surgical requirements.

    The device consists of four major assemblies integrated into one system: Main Unit, Universal Cart, the Argon Module, Accessories. The EES 100 provides finer maneuverability and fits a variety of surgical equipments, endoscopes and instruments. The Software is equipped in the device. Dispense with setting or modification for the user. Dispense with training before using the device.

    AI/ML Overview

    This 510(k) summary for the Medrange Corporation's Electrosurgical Expansion System (EES 100) does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices based on intended use, principles of operation, technological characteristics, and functional capabilities. It primarily discusses compliance with various safety and performance standards, but it does not report on a specific study with defined acceptance criteria for the device's performance.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I answer questions about sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies, as this information is absent from the provided text.

    Here's a breakdown of what is mentioned in relation to performance and testing:

    • Performance Data Section:
      • States: "The Hazard Analysis of the device is in accordance with ISO 14971 Medical Application of Risk Management to Medical Devices."
      • States: "The performance of product has been tested and verified by users." (This is a general statement and does not describe a formal study with acceptance criteria).
      • States: "The Safety of General Requirement, EMC, Software, Usability and Alarm of the device are in compliance with following standards: IEC60601-1, IEC60601-1-2, IEC 60601-1-4, IEC60601-1-6, IEC60601-1-8, And the device is also in compliance with ANSI/AAMI HF18: Standard for Electrosurgical Instruments." (This indicates compliance with standards, not a study proving specific acceptance criteria laid out for the device's performance).

    In summary, the provided document focuses on demonstrating that the EES 100 is substantially equivalent to existing, cleared devices by complying with general safety and performance standards for electrosurgical equipment. It does not detail a specific performance study with acceptance criteria as envisioned in your prompt for medical imaging devices or AI-powered diagnostics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102114

    Validate with FDA (Live)

    Date Cleared
    2010-11-22

    (117 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MB8010 Electrosurgical generator is indicated for monopolar cutting and coagulation and bipolar coagulation in surgical procedures and is intended to be used with monopolar handpieces and ground pad or bipolar handpieces and footswitches.

    Device Description

    Medrange MB 8010 ESU is an electrosurgical generator with monopolar pure cut; blend cut, coagulation and bipolar coagulation functions which meet surgical demands. Those functions are achieved by front panel selection of waveforms and power level. All selection is effected through push buttons and lamps which give the operator feedback of status. Power level for each mode is indicated by front panel digital displays which also show status of self-test and monitoring. The final output power control is made through foot or hand switches. Both monopolar and bipolar electrodes are provided. It is designed to comply with international safety standards.

    AI/ML Overview

    This document is a 510(k) summary for the Medrange MB8010 Electrosurgical Unit (ESU). It describes the device, its intended use, and claims substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Medrange MB8010 ESU are primarily based on its conformance to various international safety and performance standards. The "reported device performance" in this context refers to the device being found to meet these standards and thus demonstrating substantial equivalence to its predicate devices.

    Acceptance CriterionReported Device Performance
    Conformance to ANSI/AAMI HF-18 (American National Standard for Electrosurgical Devices)The device "conforms to" this standard.
    Conformance to IEC 60601-1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)The device "conforms to" this standard.
    Conformance to IEC 60601-1-2 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests)The device "conforms to" this standard.
    Conformance to IEC 60601-1-4 (Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems)The device "conforms to" this standard.
    Conformance to IEC 60601-1-8 (Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems)The device "conforms to" this standard.
    Conformance to IEC 60601-2-2 (Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories)The device "conforms to" this standard.
    Conformance to ISO 14971 (Medical devices – Application of risk management to medical devices)The device "conforms to" this standard (risk management).
    Substantial Equivalence to Predicate Devices (ConMed Excalibur Plus PC K953007 and Valleylab FX™ K944602)"Sufficient data were obtained to show that the device is substantially equivalent to the predicate devices, and meets safety and effectiveness criteria."
    Functions as Intended"Performance testing was performed to ensure that MB 8010 electrosurgical generator functions as intended..."
    Meets Design Specifications"...and meets design specifications."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for performance testing or study (e.g., number of units tested, number of simulated procedures, or patient data). It only mentions "Performance testing was performed to ensure that MB 8010 electrosurgical generator functions as intended, and meets design specifications."

    The data provenance is not specified regarding country of origin or whether it was retrospective or prospective. Given the nature of a 510(k) for an electrosurgical unit, the testing would typically involve laboratory benches tests and potentially animal or cadaveric studies rather than human patient data in the same way a diagnostic AI device might.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. For electrosurgical units, ground truth is typically established through engineering specifications, regulatory standards, and objective physical measurements rather than expert consensus on clinical findings.

    4. Adjudication Method

    An adjudication method is not applicable or mentioned in this context. Adjudication typically applies to assessments of subjective clinical findings, which are not detailed here for an electrosurgical generator.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported. MRMC studies are primarily relevant for diagnostic devices where human readers interpret medical images or data, and AI assistance can be evaluated for improving their performance. This device is an electrosurgical unit, not a diagnostic imaging AI.

    6. Standalone Performance Study (Algorithm Only)

    The concept of "standalone performance (algorithm only without human-in-the-loop performance)" is not directly applicable to this device. The MB8010 ESU is a hardware device with embedded control systems. Its "performance" refers to its electrical output characteristics, safety features, and functional behavior as designed, not an AI algorithm that operates independently. The performance testing evaluated the device's ability to meet its technical specifications and regulatory standards.

    7. Type of Ground Truth Used

    The ground truth used for this device's performance evaluation would primarily be:

    • Engineering Specifications and Design Requirements: The device's output power levels, waveform characteristics, safety interlocks, and other functional parameters must adhere to predefined engineering designs.
    • Regulatory Standards: Conformance to standards like ANSI/AAMI HF-18, IEC 60601-1 series, and ISO 14971 establishes the ground truth for safety and essential performance. This is the explicit ground truth mentioned in the document.
    • Predicate Device Performance: Implicitly, the performance of the predicate devices (ConMed Excalibur Plus PC and Valleylab FX™) serves as a benchmark for "substantial equivalence."

    8. Sample Size for the Training Set

    The concept of a "training set" is not applicable here. This device is an electrosurgical unit, not an AI/ML-based diagnostic or predictive algorithm that requires a training set of data. Its functionality is based on established electrical engineering principles and control systems, not machine learning from data.

    9. How Ground Truth for the Training Set Was Established

    As there is no training set for this type of device, this question is not applicable.